Custom Array Comparative Genomic Hybridization: the Importance of DNA Quality, an Expert Eye, and Variant Validation

1
Dipartimento di Scienzedella Salute, Sezione di Biostatistica, Università degli Studi di Genova, Via Pastore 1, 16132 Genoa, Italy
2
Struttura Complessa Laboratorio Genetica Umana, E.O. Ospedali Galliera, Via Volta 6, 16128 Genoa, Italy
3
Department of Medical Genetic and Laboratories, University Hospitals of Geneva, Bâtiment de Base 8C-3-840.3, 4 Rue Gabrielle-Perret-Gentil, 1211 Geneva 14, Switzerland
4
UOC Genetica Medica, Istituto Giannina Gaslini, L. go G. Gaslini 5, 16148 Genoa, Italy
*
Author to whom correspondence should be addressed.
Int. J. Mol. Sci. 2017, 18(3), 609; https://doi.org/10.3390/ijms18030609
Submission received: 7 February 2017 / Revised: 1 March 2017 / Accepted: 7 March 2017 / Published: 10 March 2017
(This article belongs to the Section Biochemistry)

Abstract

:
The presence of false positive and false negative results in the Array Comparative Genomic Hybridization (aCGH) design is poorly addressed in literature reports. We took advantage of a custom aCGH recently carried out to analyze its design performance, the use of several Agilent aberrations detection algorithms, and the presence of false results. Our study provides a confirmation that the high density design does not generate more noise than standard designs and, might reach a good resolution. We noticed a not negligible presence of false negative and false positive results in the imbalances call performed by the Agilent software. The Aberration Detection Method 2 (ADM-2) algorithm with a threshold of 6 performed quite well, and the array design proved to be reliable, provided that some additional filters are applied, such as considering only intervals with average absolute log2ratio above 0.3. We also propose an additional filter that takes into account the proportion of probes with log2ratio exceeding suggestive values for gain or loss. In addition, the quality of samples was confirmed to be a crucial parameter. Finally, this work raises the importance of evaluating the samples profiles by eye and the necessity of validating the imbalances detected.

Graphical Abstract

1. Introduction

Array-based comparative genomic hybridization (aCGH) has provided a new impulse to cytogenetic diagnostics and has proved to be a valuable tool in the clinical management of patients with developmental delays and multiple congenital anomalies [1]. This approach has also allowed the identification of novel chromosomal syndromes [2,3], helped to define the clinical variability associated with several genomic disorders [4], and led to the discovery of polymorphic copy number variants (CNVs) in the human genome [5,6,7].
Early CGH arrays were composed of large-insert bacterial artificial chromosome (BAC) clones and later evolved to microarray-based technology oligoarray CGH (aCGH).
There were concerns about procedure variability and interpretation criteria for the clinical application of early versions of targeted BAC clone array because of the presence of false negative results [8]. Two studies attempted to estimate the false positive (FPR) and false negative (FNR) rates. Wong et al. [9] analyzed six repeated experiments on 95 individuals. Given the very low binomial probability of detecting by chance the same clone twice within six experiments, they assumed that any clone detected twice or more in their experiments was a true CNV [9]. In this way, they calculated a FNR of 45.3% and a FPR of 0.23%. Following a similar method, de Smith et al. [10] calculated an estimate of FNR of 0.16 [10]. The FPR was instead estimated to be 0.05 for multi-probe calls by using three self-self hybridizations of the reference sample and comparing the average number of variant interval calls with that calculated for each sample. However, a few studies validated BAC clone arrays by examining well ascertained CNVs [11,12].
The more recent development of oligonucleotide aCGH has led to a greater resolution in CNV identification. Concordance for oligo aCGH with BAC array was shown [13,14], with around 99% sensitivity and 99% specificity [11,15] and superior performance of oligonucleotide aCGH over BAC clone aCGH [13,16].
However, all these studies based their confirmatory results on imbalances of standard cytogenetic size (Mbs). In addition, high probe density might generate more noise in aCGH data. Using two different Agilent CGH microarrays, it has been shown empirically that subject-to-subject variance is almost twice as large as array-to-array and dye-to-dye variance, supporting results reliability. However, the same study showed that the array-to-array variability was more than 10 times larger than both subject-to-subject and dye-to-dye variance for a custom microarray [17]. This observation was suggested to be ascribable to the fact that to achieve the highest possible density coverage might have lead to include less reliable probes.
The use of oligo array, though increasing resolution, has lead to a lower specificity and higher potential for noise, with need for several adjacent probes to confidently identify CNV regions and a large amount of data analysis and result interpretation. Each of the necessary steps of data transformation, normalization and summarization involves algorithm parameters that directly affect the sensitivity and specificity of the aCGH assay and represents a source of potential Type I and Type II errors [18]. The analysis can be complicated by the presence of platform and method artifacts including GC-waves [19,20] and by centralization methods [21].
More recently, the growing use of custom arrays, which are based on libraries of validated synthetic probes that can interrogate relevant genomic regions, have further enhanced the resolution capabilities of targeted regions [22,23,24,25,26]. In addition, single nucleotide polymorphism (SNP) arrays have been exploited to search for CNVs: not feasible for single-exon resolution throughout the genome, they nonetheless have the advantage to provide genotypes and detect regions of absence of heterozygosity (AOH) thus also allowing the identification of uniparental isodisomies (UPD) and genetic identity by descent [27]. However, the amount of data produced by SNP arrays is computationally challenging and requires a burden of analysis and filters to allow interpretation of results, and it is not yet well described whether the widely used SNP-array-based CNV calling methods can provide sufficient concordance with CGH in CNV detection [28]. In addition, SNP array were shown to not outperform oligonucleotides aCGH in a study that carried out a CNV search by the Affymetrix 6.0 SNP array on patients with developmental disorders already found negative by oligo aCGH at higher resolution [29]. Combining SNPs and oligo arrays in a single assay is increasingly being employed, with the advantage to obtain genotypes and a higher resolution with respect to aCGH data alone [30]. However, if the final goal is limited to the search of CNVs, oligo aCGH remains the most cost effective and straightforward method.
Several quality metric variables can be used to evaluate the quality of the oligo array and dataset, such as probe-to-probe log2ratio noise, signal intensity, background noise channels, signal to noise and reproducibility. In addition, CNVs are generally claimed when several probes are indicative of CNVs, although this reduces the array resolution. However, false positive and false negative results are still an issue.
A few studies regarding preimplantation genetic screening in human assisted reproduction showed a not infrequent presence of false positive and false negative results from aCGH [31,32]. In particular, Capalbo A. et al. [31] compared aCGH and qPCR on 120 aneuploid blastocysts, finding that 18.3% of embryos gave a discordant result for at least one chromosome and that most of these were due to aCGH false positive results.
A small number of other studies mention the presence of false positive and false negative results detected, for instance, by the use of different aberration detection methods [24], or the finding of possible aberration missed at aCGH when reanalyzing the results by Next Generation Sequencing (NGS) [26]. However, in our opinion, the presence of false and positive results is not addressed enough in the results reported in the literature.
Abnormal results should be confirmed and various strategies have been described to follow-up analysis, including repeated aCGH testing, fluorescence in situ hybridization (FISH), microsatellite analysis, multiplex ligation-dependent probe amplification (MLPA) and, in particular, real-time quantitative PCR (qPCR). Recently, NGS has also been suggested [26].
Nonetheless, aCGH remains the first-tier testfor CNV detection, due to its genome wide applicability. Validation on a large number of patient and control samples following aCGH analysis is not practical, and it is not always reported in large screenings [33,34], or only few interesting candidates are validated [26].
We took advantage of a custom aCGH, recently carried out on 59 patients affected by Hirschsprung Disease (HSCR) to search for imbalances in genes and loci candidate for HSCR [35], to analyze its performance, the use of several aberrations detection algorithms, and the presence of false positive and false negative results.

2. Results

2.1. Sample Quality and Design Reliability

DLRSs (derivative log ratio spread) and the other metrics of the final 59 samples analyzed for the search of aberrations are shown in Figure 1. The DLRS, in particular, is a measure of the log ratio noise for each sample, calculated as the standard deviation (spread) of the log ratio differences between consecutive probes.
To assess the reliability of the results, we correlated the log2ratios between replicates as reported elsewhere [36], finding low correlations (mean r = 0.18 across 37 comparisons), though higher than among random sample pairs (mean r = 0.07, p = 0.0040) (Table 1). Such a low correlation is not unexpected since log2ratios not different from zero are supposed to vary randomly. As a matter of fact, considering only log2ratios with absolute values above 0.3, the mean correlations improved in both replicated and random sample pairs (mean r = 0.42, p = 1.8 × 10−9 and 0.14, p = 0.0036, respectively), but at a much higher rate for the replicates (p = 4.8 × 10−5). These results reassure on the design quality and suggest that the application of a minimum log2ratio values criterion, such as the mean absolute log2ratio > 0.30 (hereafter referred as MALR > 0.30), is reasonable and advisable. Not surprisingly, the level of correlation is dependent on the derivative log ratio spread (DLRS): correlations between replicates and random pairs were significantly different only when at least one sample had DLRS ≤ 0.2 (p = 6.8 × 10−6 on |log2ratio| ≥ 0.3). Correlations were much lower and not significant for the 13 pairs with both samples with DLRS > 0.2 (r = 0.21 for replicates and r = 0.09 for random pairs).
Of note, although we could not find any correlation between the DLRS and the log2ratio values, besides a faint negative trend, correlations between pairs in which at least one sample had DLRS ≤ 0.2 were significantly higher than correlations between pairs without any sample with DLRS ≤ 0.2, for replicates but not for random pairs (i.e., r = 0.53 vs. 0.21, p = 0.0009 for replicates and r = 0.17 vs. 0.09, p = 0.2188 for random pairs on |log2ratio| ≥ 0.3) (Table 1).
Following the method described elsewhere [10], we estimated the FPR for the Aberration Detection Method 2 (ADM-2) at threshold 6 to be 0.130 for single-probe calls and 0.184 for single-probe calls, higher than that estimated by de Smith et al. [10]. However, all the aberrations called in the three self-self test regarded the high density region around RET (10q11.2), confirming that this sub-centromeric region is problematic, and presented with MALR < 0.3, so that applying such a filter the FPR virtually dropped to 0. Of note, all the calls on this gene, including those that seemed likely or possible, were not confirmed at validation.
We also estimated the FNR in a manner similar to that described by Wong et al. and de Smith et al. [9,10]. In the four replicated experiments, 13 putative variant intervals were observed twice and then considered true calls, yielding an estimate of FNR of 0.50. We also evaluated three samples assayed in triplicate, finding an average FNR of 0.213. The FNR found in the four replicates experiments was slightly higher than that estimated by Wong et al. [9] and definitely higher than that estimated by de Smith et al. [10], while it was closer to their estimates for the three replicates experiments. However, we calculated that if Wong et al. [9] had performed four (or three) replicates, they would have estimated a FNR of 0.3629 and 0.2111, respectively, similar to those obtained by us.
The studies by both de Smith et al. [10] and Wong et al. [9] made use of BAC-based CGH, so that a direct comparison with our results might be misleading. Nevertheless, a high FNR (above 20%), with calls that were missed in one or more replicates, and a very low FPR were confirmed. This latest estimate is based on a reference DNA with very high quality (DLRS < 0.15) and we cannot exclude that FPR would have been higher with worse quality samples. Accordingly, few imbalances could not be validated or confirmed in replicates.

2.2. High Density Design Performance

Notwithstanding the possibility that high density regions might present low replicability and display worse profiles than less high density region, when zooming in the regions we did not observe any higher variability in the profiles between high density probes regions and the rest of the genome (Figure 2).
Accordingly, we found that the number of calls in the selected regions correlated with the number of probes analyzed (r = 0.773) as expected, while it did not correlate with either the size or the probe density of the region selected. These observations were confirmed considering only calls sustained by at least two probes or considering the number of probes called instead of the number of calls. When only calls with MALR > 0.3 were considered, no correlation could be detected at all (Figure 3).
Redundancy of probes covering the same target sequence might instead create problems in the hybridization and thus false results, as we could observe in a preliminary design.

2.3. Comparison between Algorithms and Filters

Applying the ADM-2 algorithm with a threshold of 6, the CG correction and the centralization algorithms, and excluding the positive control regions, we got 572 aberration calls on autosomal chromosomes (for a median of seven calls per sample, range 1–34), 393 of which sustained by at least two consecutive probes (five median calls per sample, range 1–17). We also repeated the analysis without the GC correction but no difference could be detected in the results.
However, at the samples profiles visual inspection, most of the calls seemed to us as false positives, for this reason we have also applied two additional filters, MALR > 0.30 and thresholde > 0.33. With the first filter we obtained 75 calls, six of which corresponding to the already known chromosomal alterations (for these latter we got nine calls, but five were overlapping for the two chromosomal RET deletions on the two sample used as controls for this region and are not reported in Table 2). Two CNVs were not found in the best quality sample firstly evaluated but were added to the list because found in the two additional good quality replicates. These 77 aberrations reduced to 52 considering the thresholde > 0.33, an empirical threshold based on the number of probes with specific log2ratio ranges that we have applied to the present data (see the methods) (Table 2).
Forty-two calls were detected with the same two filters applying the Fuzzy-zero algorithm, while 22 aberrations were called applying the ADM-2 algorithm with a more stringent threshold of 8.
The visual inspection of the sample profiles allowed us to add six calls to the list of aberrations, for a total of 83 aberrations in 61 different chromosomal locations (including controls) in 44 samples.
Excluding aberrations previously reported on DGV and controls, 51 aberrations were detected in 25 patients, two of which repeated in three patients each. The variants called reduced to 25 aberrations in 17 patients when applying the thresholde > 0.33, to 15 aberrations in nine patients applying the more stringent threshold of 8 and to 24 aberrations in 16 patients applying the fuzzy-zero algorithm (Table 2).
For 15 of the novel aberrations we had at least one replicate sample available, although for two samples the replicate was of low quality (DLRS ≥ 0.3). While one replicate was inconclusive because of a too noisy profile (a low quality replicate), six were definitely not replicated, four were likely although not called by the software and four were clearly replicated, although two showed a different size.

2.4. Software Algorithms Calls and Visual Inspection

The visual evaluation of the log2ratio sample profiles allowed the identification of additional putative aberrations, not detected by the software, suggesting the substantial chance of false negative results. Of the 44 CNVs classified as likely or possible based on visual inspection and further verified, 39 resulted to be true (confirmed at the validation, detected also on a second replicate or reported on DGV and thus assumed as true, in addition to the six known aberrations all very well visible). Among those classified as unlikely, instead, only four could be claimed as true while eight were excluded at the validation (p = 0.0004), suggesting that the visual inspection of the sample profile is crucial, and even more reliable than the use of algorithms for the variant calls (Table 3).
To note, by comparing the visual inspection results with the software calls under various scenario (a more stringent threshold of 8 for the ADM-2 algorithm, the application of the Fuzzy zero algorithm and thresholde > 0.33 filters), under the application of the MALR > 0.3 with at least two probes filter, the ADM-2 algorithm with threshold 6 together with the application of the thresholde > 0.33 filters resulted to be the most comparable with the visual inspection, and the thresholde > 0.33 in general resulted to be a good discriminatory filter (Table 3). Both the ADM-2 algorithm with a threshold of 8 and the Fuzzy zero algorithm missed too many true calls (25 and 15, respectively).

3. Discussion

The search for CNVs in genes and loci candidate for HSCR in a panel of individuals affected by the disease has provided the opportunity to investigate in more detail the quality of our custom aCGH design and to address some general remarks. First of all, we could confirm that the use of a high density design does not seem to increase the error in variants detection, neither lead to worse sample profiles, thus confirming the validity of this strategy of searching for small imbalances, otherwise undetectable.
The size of the imbalances that can be detected depends upon the density of probes targeting the regions of interest and the criteria set for software-generated calls (i.e., minimum two probes as applied here). Redundancy of probes covering the same target sequence might create problems in the hybridization and thus false results. In addition, not all probes perform equally well. However, the selection of probes and the density in the coverage of the array, which in our design was up to one probe every 250 nt (for RET), might greatly increase the CNV search resolution, provided that there is no probe overlap. We could detect a few imbalances that are less than 5 kb that were successfully confirmed by other techniques.
Not surprisingly, the quality of samples is confirmed to be a crucial step. There is no correlation between the number of calls detected by the algorithm software and the DLRS of the samples, as it is expected since the algorithm already took into account the sample quality. However, the log2ratio correlation between replicates was greatly improved when DLRS were excellent (≤0.2). To note, among calls with MALR > 0.3, those detected in samples with excellent DLRS were classified as “likely” or “possible” more frequently than those detected in sample with worse DLRS. “Likely” and “possible” variants could be grouped, although “likely” variants seemed to be true variants more often than “possible” variants (30 vs. 4 compared to 4 vs. 1).
Above all, our study highlighted the importance of the visual inspection by an expert eye. The human eye can take into account several factors such as the general profile of the sample, the specific region profile, and the single log2ratio values involved in the putative aberration qualitatively better than any algorithm. Of course, in the case of a large screening, the use of algorithms becomes a pivotal tool. In this case, we noticed that the ADM-2 algorithm with the threshold of 6 suggested by the Agilent company performed quite well, provided that a minimum absolute log2ratio for the region is taken into account, such as a MALR > 0.30, as already applied in several studies. In addition, we suggest that also the number of probes concordant should be considered to call the imbalance. For instance, we applied an empirical filter that evaluated if at least one thirds of the probes were above (for gain) or below (for loss) a certain cutoff value, here chosen to be +0.5 and −0.8, finding that the chance to discriminate between true and false calls was greatly improved, especially with the ADM-2 detection algorithm.
A more stringent threshold value for the ADM-2 algorithm did not seem to improve the detection of true imbalances, while, conversely, was at risk of missing the call of several possibly true variants. The same can be said for the Fuzzy Zero algorithm. It is useful for large regions that however are easily detectable as false positive also by the visual inspection of the samples profile.
Very evident large calls, clearly visible at visual inspection and with a neat discrimination above the baseline, were identified with a high level of confidence. They were also easily replied in replicates, including very low quality replicates. This happened for instance for the loss and gain CNVs we included as controls and for two Down syndrome samples. Similarly, the first studies that investigated the reliability of aCGH were mostly based on this sort of gold standard and achieved a complete or almost complete concordance. It remains to explain why other smaller calls seem to be less evident. One possibility is that smaller calls rely on a small number of probes and are therefore more dependent on local probes quality. In addition, problematic regions (such as those closed to centromere) had variability: the RET region, known to be difficulty at amplification and screening, was the one to show most false results. In addition, mosaicism cannot be excluded to explain this observation.
It has been suggested that an additional source of variability is given by bench bias [17]. We did not presently investigate such a matter, but from our observation we cannot exclude that correlation and concordance in aberrations calls is dependent on the array and time at which the experiment has been carried out.
Finally, it has also been shown that the fluorescent dyes commonly used in array CGH, fundamentally the red dyes, are sensitive to ozone, and that ozone has a very strong effect on array data especially during the post-hybridization step [37]. We did not specifically address the possibility of false results in this context; however, the arrays were washed and dried under laminar flow hoods in a semi-darkness environment to reduce ozone exposure. Accordingly, we did not observe a difference in gain/loss ratio among the different groups of CNVs (likely, possible, unlikely, true or not true), with the exception of the variants defined as unlikely, basically due to CNVs bigger than 5 kb. The four unlikely true variants (thus possible false negative) were all gains (red dyes predominant) bigger than 5 kb, while among the eight not true variants (possible false positive), the three variants smaller than 5 kb were all gains, and the five variants bigger than 5 kb were all losses. Thus, we cannot exclude that ozone has affected our data, but if this is the case it seems to affect less the smaller aberrations.
In accordance with the impression we got by visual inspection, we have estimated high false positive and false negative rates for the software calls, though these figures need to be taken cautiously. Indeed, we had some concerns given to the fact that few variants detected were then excluded by alternative methods (such as three unlikely and one likely losses and gains excluded by qPCR) and, on the other hand, we were able to detect variants missed by the software and that were successively confirmed to be true (such as three likely losses and gains confirmed by qPCR). Similarly, a deletion found in a sample for which we had two additional replicates, was also detected on the bad quality replicate but not on the good one. In addition, we found two true aberrations in a sample that was in triplicate that were not detected in the best quality replicate and would have thus gone undetected.
Therefore, our observations strengthen the need to validate the results by means of other techniques, among which qPCR is particularly recommended. A priori use of a combined oligo aCGH and SNP array might have avoided some false results, too. Supposed deletions could have been excluded based on heterozygote genotypes in the same region, for instance. However, the non-uniform distribution of informative SNPs throughout the genome might nullify such advantage in specific regions, especially in the case of small CNVs.
Most of the observations reported here come from visual classification, thus lacking of a proved gold standard. Nevertheless, our study provides a confirmation that the high density design of aCGH does not generate more noise than lower density designs and, in addition, does reach a better resolution, with the finding of validated imbalances smaller than 5 kb. In conclusion, our design proved to be reliable, provided that some filters are applied such as MALR ≥ 0.3. We also propose an additional filter, treshholde > 0.33, which takes into account the proportion of probes with log2ratios exceeding suggestive values for gain or loss. In any case, besides the use of additional filters, we would like to stress the importance of paying a great attention to the observation of the samples profiles and the necessity of validating the imbalances detected.

4. Materials and Methods

4.1. Microarray Design

The sample was constituted by 59 Italian sporadic HSCR patients, six of whom carrying known chromosomal aberrations at the karyotype level: three chromosome 21 trisomies, an invdup(22)(q11) and two interstitial deletions in 10q11.21. The clinical features, selection and processing of the samples are described elsewhere [35].
We have designed a high-density custom array (8X15K SurePrint G3 Human Kit, Agilent Technologies, Santa Clara, CA, USA) through the Agilent eArray web portal. Genomic DNA (test) and sex-matched controls (Promega, Madison, WI, USA) were labelled and hybridized following the protocols provided by the manufacturers. Spot intensities were processed by Agilent Feature Extraction software and the text file outputs were imported into Agilent Genomic Workbench v. 5.0.14 software (Agilent Technologies, Santa Clara, CA, USA) distributed by the vendor.
The microarray consisted of 8333 probes at a high density, selected to cover 20 HSCR candidate genes as described elsewhere, and 3130 probes scattered along the genome, with a probe density of around 1 probe every 900 kb, that constituted the backbone together with 1262 normalization probes (13 of which located in the selected regions), 301 probes replicated five times (1505 probes), and 1482 control probes provided by Agilent (Table 4).
Gene and locus positions are based on the Human Genome GRCh37 (hg19) assembly of UCSC genome browser [38].

4.2. Data Analysis and Structural Variant Detection

To investigate genomic imbalances, we applied the ADM-2 algorithm. A threshold of 6 was set, as recommended by Agilent, and a more stringent threshold of 8 has also been tried.
We applied the centralization algorithm and the GC correction algorithm, and we repeated the aberration detection call both applying and not applying the Fuzzy Zero algorithm.
Finally, we have considered as aberrant only those regions with a minimum of 2 probes and with minimum absolute average log2ratio for region >0.3. In addition, we evaluated the effect of an additional criteria, the thresholde > 0.33, namely that at least one third of probes in the putative imbalanced interval (at least 2 probes in the case of aberrations based on 3 probes) must present log2ratio above 0.5 or below −0.8 for gains and loss respectively (based on log2(3/2) = 0.58, log2(1/2) = −1).
We also evaluated the samples profiles by visual inspection, reviewed by a second well-trained operator. Loci with nearby gain or loss intervals and an intervening region of more than 2 probes were considered two separate CNVs, as well as those differing for 2 probes presenting inconsistent log2ratio (opposite direction, that is log2ratio < −0.3 for gain and >0.3 for deletions).

4.3. Statistical Analysis

The Agilent Feature Extraction (FE) processes the data, calculates signal log2ratios, estimates errors, and provides basic QC metrics. In particular, we have evaluated: (i) the DLRSpread (derivative log2ratio spread), which is a measure of the log2ratio noise for each sample; (ii) the BGNoise (background noise), which is a measure of the background fluorescence for each channel (Red and Green); (iii) signal intensity; (iv) signal to noise; and (v) the reproducibility for each channel. Measures were considered as excellent, good or to be evaluated, based on Agilent’s guideline (Figure 1).
Newly extracted or purified DNA was run on an additional array for 10 samples with bad profiles, which are those with DRLS ≥ 0.3, and for 16 samples arbitrarily selected. Moreover, four samples were replicated three times (two of which had the third replicate of bad quality), and one sample was replicated four times. When more replicates were available, the variants search was performed in the sample with the lowest DRLS, or, in case of very similar DRLS among replicates, in the one with an overall better quality. In any case, aberrations detected were evaluated also in the replicated samples, when available.
The FPR was determined comparing the average number of variant calls in self-self tests with the average number of variant calls for each sample, using three self-self hybridizations of a reference sample, as described by de Smith et al. [10]. We estimated the FNR using replicated experiments as described by Wong et al. and de Smith et al. [9,10] and adjusted their estimates based on BAC array by changing number of clones with number of probes. While they based their estimate on 6 and 4 replicates, respectively, we used both four and three replicates. Of note, if Wong et al. [9] had performed 4 replicates only, some calls detected more than once would have been lost (virtually present in one of the two additional replicates no more present). For this reason, we calculated by permutation the expected number of calls replicated more than twice by Wong et al. [9] in the case of four (or three) replicates, assuming a random distribution on the replicates experiments. We considered aberrant intervals revealed in different experiment as identical if the overlap among probes was ≥0.90.
We also calculated the correlation between replicates and between random sample pairs selected to have similar DLRS (<0.2, ≥0.2 and <0.3 or ≥0.3). We repeated the analysis for both all log2ratios and log2ratio exceeding the threshold of absolute 0.3 (considering that those with log2ratio not different from 0 are not expected to correlate at all) and among all the replicated samples and among only those with DRLS below the two cutoffs of 0.3 and 0.2. Moreover, we investigated whether the number of aberrations called by the software correlated with the number of probes called, the size, or the probe density, excluding from the analysis the aberrations used as controls and considering the aberrations that overlapped the high density probes regions and the genome as belonging to the high density group. We assumed the aberration size to be the mean between the inner and the outer probes called.
Finally, we tried to investigate whether there was an association between the calls obtained with the software under different criteria and the visual inspection results.

4.4. Detected Variants Classification and Validation

The visual inspection of the aCGHsample profiles has allowed us to classify the aberrations detected as known (the controls), likely, possible but not convincing (possible), or unlikely (Table 2).
Aberrations were compared with CNVs observed in the normal population and reported in the Database of Genomic Variants (DGV) [39] and with the CNVs reported in the DECIPHER database of phenotypes, v8.7 released [40]. The comparison between different platforms and techniques is tricky, exact boundaries of the aberrations detected by arrays are not known but only assumed to be between the last “normal” probe (outer) and the first “aberrant” probe (inner) and depend on the average coverage. However, we considered aberrations as consistent with those already reported in the databases if they showed an overlap ≥80%, did not differ for more than two probes with compatible log2ratios (that is ≥|0.3|), and were of the same kind (gain or loss). The frequency of the CNV or the number of individuals in the database in which the variant is reported was not a selection criterion but is reported.
We arbitrarily selected the most promising regions (those classified as likely and not reported on the DGV database) and those more interesting for us (i.e., on the RET gene) for validation with other molecular biology techniques, and parental check, as described elsewhere [35] and reported in Table 2.
We considered as true calls those detected in the controls (already assessed with other techniques or trisomy 21 in patients also affected by Down syndrome), the aberrations confirmed at the validation, the calls detected also on a second replicate and CNVs reported on DGV.

5. Conclusions

Our study provides a confirmation that the high density design of aCGH does not generate more noise than lower density designs and, in addition, does reach a better resolution. However, false positive and false negative results are not trivial. For this reason, we suggest that some filters are applied such as the MALR ≥ 0.3 and the treshholde > 0.33, this latter taking into account the proportion of probes with log2ratios exceeding suggestive values for gain or loss. We have also shown the importance of visual inspection of results and the necessity of validating the imbalances detected.

Acknowledgments

We wish to thank the numerous patients, their families and referring physicians that have participated in these studies in the Gaslini Institute. We also thank Chiara Viaggi, Manuela Mosconi and Alessio Piniprato for their help in interpreting results, and Herve Chaulet from Agilent for his suggestions with the custom array design. This work was supported by the Italian Ministry of Health: “Cinque per mille” and Ricerca Corrente to the Gaslini Institute and by the Italian Ministry of Health: “GR-2011-02347381”.

Author Contributions

Francesca Lantieri and Isabella Ceccherini conceived and designed the experiments; Michela Malacarne and Stefania Gimelli performed the experiments; Domenico Coviello provided help for the validation experiments; Francesca Lantieri analyzed the data; Giuseppe Santamaria prepared the samples; and Francesca Lantieri and Isabella Ceccherini wrote the paper.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Shaffer, L.G.; Bejjani, B.A.; Torchia, B.; Kirkpatrick, S.; Coppinger, J.; Ballif, B.C. The identification of microdeletion syndromes and other chromosome abnormalities: Cytogenetic methods of the past, new technologies for the future. Am. J. Med. Genet. C Semin. Med. Genet. 2007, 145C, 335–345. [Google Scholar] [CrossRef] [PubMed]
  2. Ballif, B.C.; Hornor, S.A.; Jenkins, E.; Madan-Khetarpal, S.; Surti, U.; Jackson, K.E.; Asamoah, A.; Brock, P.L.; Gowans, G.C.; Conway, R.L.; et al. Discovery of a previously unrecognized microdeletion syndrome of 16p11.2-p12.2. Nat. Genet. 2007, 39, 1071–1073. [Google Scholar] [CrossRef] [PubMed]
  3. Sharp, A.J.; Hansen, S.; Selzer, R.R.; Cheng, Z.; Regan, R.; Hurst, J.A.; Stewart, H.; Price, S.M.; Blair, E.; Hennekam, R.C.; et al. Discovery of previously unidentified genomic disorders from the duplication architecture of the human genome. Nat. Genet. 2006, 38, 1038–1042. [Google Scholar] [CrossRef] [PubMed]
  4. Coppinger, J.; McDonald-McGinn, D.; Zackai, E.; Shane, K.; Atkin, J.F.; Asamoah, A.; Leland, R.; Weaver, D.D.; Lansky-Shafer, S.; Schmidt, K.; Feldman, H.; et al. Identification of familial and de novo microduplications of 22q11.21-q11.23 distal to the 22q11.21 microdeletion syndrome region. Hum. Mol. Genet. 2009, 18, 1377–1383. [Google Scholar] [CrossRef] [PubMed]
  5. Iafrate, A.J.; Feuk, L.; Rivera, M.N.; Listewnik, M.L.; Donahoe, P.K.; Qi, Y.; Scherer, S.W.; Lee, C. Detection of large-scale variation in the human genome. Nat. Genet. 2004, 36, 949–951. [Google Scholar] [CrossRef] [PubMed]
  6. Perry, G.H.; Ben-Dor, A.; Tsalenko, A.; Sampas, N.; Rodriguez-Revenga, L.; Tran, C.W.; Scheffer, A.; Steinfeld, I.; Tsang, P.; Yamada, N.A.; et al. The fine-scale and complex architecture of human copy-number variation. Am. J. Hum. Genet. 2008, 82, 685–695. [Google Scholar] [CrossRef] [PubMed]
  7. Redon, R.; Ishikawa, S.; Fitch, K.R.; Feuk, L.; Perry, G.H.; Andrews, T.D.; Fiegler, H.; Shapero, M.H.; Carson, A.R.; Chen, W.; et al. Global variation in copy number in the human genome. Nature 2006, 444, 444–454. [Google Scholar] [CrossRef] [PubMed]
  8. Shearer, B.M.; Thorland, E.C.; Gonzales, P.R.; Ketterling, R.P. Evaluation of a commercially available focused aCGH platform for the detection of constitutional chromosome anomalies. Am. J. Med. Genet. A 2007, 143A, 2357–2370. [Google Scholar] [CrossRef] [PubMed]
  9. Wong, K.K.; de Leeuw, R.J.; Dosanjh, N.S.; Kimm, L.R.; Cheng, Z.; Horsman, D.E.; MacAulay, C.; Ng, R.T.; Brown, C.J.; Eichler, E.E.; et al. A comprehensive analysis of common copy-number variations in the human genome. Am. J. Hum. Genet. 2007, 80, 91–104. [Google Scholar] [CrossRef] [PubMed]
  10. De Smith, A.J.; Tsalenko, A.; Sampas, N.; Scheffer, A.; Yamada, N.A.; Tsang, P.; Ben-Dor, A.; Yakhini, Z.; Ellis, R.J.; Bruhn, L.; et al. Array CGH analysis of copy number variation identifies 1284 new genes variant in healthy white males: Implications for association studies of complex diseases. Hum. Mol. Genet. 2007, 16, 2783–2794. [Google Scholar] [CrossRef] [PubMed]
  11. Cheung, S.W.; Shaw, C.A.; Yu, W.; Li, J.; Ou, Z.; Patel, A.; Yatsenko, S.A.; Cooper, M.L.; Furman, P.; Stankiewicz, P.; et al. Development and validation of a CGH microarray for clinical cytogenetic diagnosis. Genet. Med. 2005, 7, 422–432. [Google Scholar] [CrossRef] [PubMed]
  12. Yu, W.; Ballif, B.C.; Kashork, C.D.; Heilstedt, H.A.; Howard, L.A.; Cai, W.W.; White, L.D.; Liu, W.; Beaudet, A.L.; Bejjani, B.A.; et al. Development of a comparative genomic hybridization microarray and demonstration of its utility with 25 well-characterized 1p36 deletions. Hum. Mol. Genet. 2003, 12, 2145–2152. [Google Scholar] [CrossRef] [PubMed]
  13. Shen, Y.; Irons, M.; Miller, D.T.; Cheung, S.W.; Lip, V.; Sheng, X.; Tomaszewicz, K.; Shao, H.; Fang, H.; Tang, H.S.; et al. Development of a focused oligonucleotide-array comparative genomic hybridization chip for clinical diagnosis of genomic imbalance. Clin. Chem. 2007, 53, 2051–2059. [Google Scholar] [CrossRef] [PubMed]
  14. Yu, S.; Bittel, D.C.; Kibiryeva, N.; Zwick, D.L.; Cooley, L.D. Validation of the Agilent 244 K oligonucleotide array-based comparative genomic hybridization platform for clinical cytogenetic diagnosis. Am. J. Clin. Pathol. 2009, 132, 349–360. [Google Scholar] [CrossRef] [PubMed]
  15. Xiang, B.; Li, A.; Valentin, D.; Nowak, N.J.; Zhao, H.; Li, P. Analytical and clinical validity of whole-genome oligonucleotide array comparative genomic hybridization for pediatric patients with mental retardation and developmental delay. Am. J. Med. Genet. A 2008, 146, 1942–1954. [Google Scholar] [CrossRef] [PubMed]
  16. Neill, N.J.; Torchia, B.S.; Bejjani, B.A.; Shaffer, L.G.; Ballif, B.C. Comparative analysis of copy number detection by whole-genome BAC and oligonucleotide array CGH. Mol. Cytogenet. 2010, 3, 11. [Google Scholar] [CrossRef] [PubMed]
  17. McDonnell, S.K.; Riska, S.M.; Klee, E.W.; Thorland, E.C.; Kay, N.E.; Thibodeau, S.N.; Parker, A.S.; Eckel-Passow, J.E. Experimental designs for array comparative genomic hybridization technology. Cytogenet. Genome Res. 2013, 139, 250–257. [Google Scholar] [CrossRef] [PubMed]
  18. Przybytkowski, E.; Aguilar-Mahecha, A.; Nabavi, S.; Tonellato, P.J.; Basik, M. Ultradense array CGH and discovery of micro-copy number alterations and gene fusions in the cancer genome. Methods Mol. Biol. 2013, 973, 15–38. [Google Scholar] [PubMed]
  19. Carter, N.P. Methods and strategies for analyzing copy number variation using DNA microarrays. Nat. Genet. 2007, 39, S16–S21. [Google Scholar] [CrossRef] [PubMed]
  20. Leo, A.; Walker, A.M.; Lebo, M.S.; Hendrickson, B.; Scholl, T.; Akmaev, V.R. A GC-wave correction algorithm that improves the analytical performance of aCGH. J. Mol. Diagn. 2012, 14, 550–559. [Google Scholar] [CrossRef] [PubMed]
  21. Commo, F.; Ferté, C.; Soria, J.C.; Friend, S.H.; André, F.; Guinney, J. Impact of centralization on aCGH-based genomic profiles for precision medicine in oncology. Ann. Oncol. 2015, 26, 582–588. [Google Scholar] [CrossRef] [PubMed]
  22. Boone, P.M.; Bacino, C.A.; Shaw, C.A.; Eng, P.A.; Hixson, P.M.; Pursley, A.N.; Kang, S.H.; Yang, Y.; Wiszniewska, J.; Nowakowska, B.A.; et al. Detection of clinically relevant exonic copy-number changes by array CGH. Hum. Mutat. 2010, 31, 1326–1342. [Google Scholar] [CrossRef] [PubMed]
  23. Hackmann, K.; Kuhlee, F.; Betcheva-Krajcir, E.; Kahlert, A.K.; Mackenroth, L.; Klink, B.; di Donato, N.; Tzschach, A.; Kast, K.; Wimberger, P.; et al. Ready to clone: CNV detection and breakpoint fine-mapping in breast and ovarian cancer susceptibility genes by high-resolution array CGH. Breast Cancer Res. Treat. 2016, 159, 585–590. [Google Scholar] [CrossRef] [PubMed]
  24. La Cognata, V.; Morello, G.; Gentile, G.; D’Agata, V.; Criscuolo, C.; Cavalcanti, F.; Cavallaro, S. A customized high-resolution array-comparative genomic hybridization to explore copy number variations in Parkinson‘s disease. Neurogenetics 2016, 17, 233–244. [Google Scholar] [CrossRef] [PubMed]
  25. Tayeh, M.K.; Chin, E.L.; Miller, V.R.; Bean, L.J.; Coffee, B.; Hegde, M. Targeted comparative genomic hybridization array for the detection of single- and multiexon gene deletions and duplications. Genet. Med. 2009, 11, 232–240. [Google Scholar] [CrossRef] [PubMed]
  26. Yoshikawa, Y.; Emi, M.; Hashimoto-Tamaoki, T.; Ohmuraya, M.; Sato, A.; Tsujimura, T.; Hasegawa, S.; Nakano, T.; Nasu, M.; Pastorino, S.; et al. High-density array-CGH with targeted NGS unmask multiple noncontiguous minute deletions on chromosome 3p21 in mesothelioma. Proc. Natl. Acad. Sci. USA 2016, 113, 13432–13437. [Google Scholar] [CrossRef] [PubMed]
  27. Wiszniewska, J.; Bi, W.; Shaw, C.; Stankiewicz, P.; Kang, S.H.; Pursley, A.N.; Lalani, S.; Hixson, P.; Gambin, T.; Tsai, C.H.; et al. Combined array CGH plus SNP genome analyses in a single assay for optimized clinical testing. Eur. J. Hum. Genet. 2014, 22, 79–87. [Google Scholar] [CrossRef] [PubMed]
  28. Zhang, X.; Du, R.; Li, S.; Zhang, F.; Jin, L.; Wang, H. Evaluation of copy number variation detection for a SNP array platform. BMC Bioinform. 2014, 15, 50. [Google Scholar] [CrossRef] [PubMed]
  29. Siggberg, L.; Ala-Mello, S.; Linnankivi, T.; Avela, K.; Scheinin, I.; Kristiansson, K.; Lahermo, P.; Hietala, M.; Metsähonkala, L.; Kuusinen, E.; et al. High-resolution SNP array analysis of patients with developmental disorder and normal array CGH results. BMC Med. Genet. 2012, 13, 84. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  30. Wajnberg, G.; Carvalho, B.S.; Ferreira, C.G.; Passetti, F. Combined analysis of SNP array data identifies novel CNV candidates and pathways in ependymoma and mesothelioma. BioMed Res. Int. 2015, 2015, 902419. [Google Scholar] [CrossRef] [PubMed]
  31. Capalbo, A.; Treff, N.R.; Cimadomo, D.; Tao, X.; Upham, K.; Ubaldi, F.M.; Rienzi, L.; Scott, R.T. Comparison of array comparative genomic hybridization and quantitative real-time PCR-based aneuploidy screening of blastocyst biopsies. Eur. J. Hum. Genet. 2015, 23, 901–906. [Google Scholar] [CrossRef] [PubMed]
  32. Tiegs, A.W.; Hodes-Wertz, B.; McCulloh, D.H.; Munné, S.; Grifo, J.A. Discrepant diagnosis rate of array comparative genomic hybridization in thawed euploid blastocysts. J. Assist. Reprod. Genet. 2016, 33, 893–897. [Google Scholar] [CrossRef] [PubMed]
  33. Zanda, M.; Onengut-Gumuscu, S.; Walker, N.; Shtir, C.; Gallo, D.; Wallace, C.; Smyth, D.; Todd, J.A.; Hurles, M.E.; Plagnol, V.; et al. A genome-wide assessment of the role of untagged copy number variants in type 1 diabetes. PLoS Genet. 2014, 10, e1004367. [Google Scholar] [CrossRef] [PubMed]
  34. Zhou, Q.; Wu, S.Y.; Amato, K.; di Adamo, A.; Li, P. Spectrum of cytogenomic abnormalities revealed by array comparative genomic hybridization on products of conception culture failure and normal karyotype samples. J. Genet. Genom. 2016, 43, 121–131. [Google Scholar] [CrossRef] [PubMed]
  35. Lantieri, F.; Gimelli, S.; Viaggi, C.; Stathaki, E.; Malacarne, M.; Santamaria, G.; Grossi, A.; Coviello, D.; Ceccherini, I. Copy Number Variations in Candidate Regions Confirm Genetic Heterogeneity in Hirschsprung Disease. Unpublished work. 2017. [Google Scholar]
  36. Jiang, Q.; Ho, Y.Y.; Hao, L.; Nichols Berrios, C.; Chakravarti, A. Copy number variants in candidate genes are genetic modifiers of hirschsprung disease. PLoS ONE 2011, 6, e21219. [Google Scholar] [CrossRef] [PubMed]
  37. Byerly, S.; Sundin, K.; Raja, R.; Stanchfield, J.; Bejjani, B.A.; Shaffer, L.G. Effects of ozoneexposure during microarray posthybridization washes and scanning. J. Mol. Diagn. 2009, 11, 590–597. [Google Scholar] [CrossRef] [PubMed]
  38. Kent, W.J.; Sugnet, C.W.; Furey, T.S.; Roskin, K.M.; Pringle, T.H.; Zahler, A.M.; Haussler, D. The human genome browser at UCSC. Genome Res. 2002, 12, 996–1006. [Google Scholar] [CrossRef] [PubMed]
  39. MacDonald, J.R.; Ziman, R.; Yuen, R.K.; Feuk, L.; Scherer, S.W. The database of genomic variants: A curated collection of structural variation in the human genome. Nucleic Acids Res. 2014, 42, D986–D992. [Google Scholar] [CrossRef] [PubMed]
  40. Firth, H.V.; Richards, S.M.; Bevan, A.P.; Clayton, S.; Corpas, M.; Rajan, D.; van Vooren, S.; Moreau, Y.; Pettett, R.M.; Carter, N.P. DECIPHER: Database of chromosomal imbalance and phenotype in humans using ensembl resources. Am. J. Hum. Genet. 2009, 84, 524–533. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Quality control metrics: Distribution of the sample quality controls is reported as box plots and as statistics. In particular, sample metrics are highlighted as excellent, good or poor (evaluate) and how many samples are in each category is also reported. Solid circles and asterisks in the box plot graphs represents the outliers: solid circles are cases with values more than 1.5 times the InterQuartile (IQ) range, asterisks are cases with values more than 3 times the IQ range.
Figure 1. Quality control metrics: Distribution of the sample quality controls is reported as box plots and as statistics. In particular, sample metrics are highlighted as excellent, good or poor (evaluate) and how many samples are in each category is also reported. Solid circles and asterisks in the box plot graphs represents the outliers: solid circles are cases with values more than 1.5 times the InterQuartile (IQ) range, asterisks are cases with values more than 3 times the IQ range.
Ijms 18 00609 g001
Figure 2. Sample profiles. An example of four samples selected for excellent quality, good, evaluate and very bad quality. For each, the profile at chromosome 9 is shown, including a region of probes scattered across the genome and two high density regions. The upper high density region inside the blue box in the left panel is zoomed in into the central panel (inside the large blue box) and the specific region inside the yellow box is further zoomed in into the right panel (inside the large yellow box).
Figure 2. Sample profiles. An example of four samples selected for excellent quality, good, evaluate and very bad quality. For each, the profile at chromosome 9 is shown, including a region of probes scattered across the genome and two high density regions. The upper high density region inside the blue box in the left panel is zoomed in into the central panel (inside the large blue box) and the specific region inside the yellow box is further zoomed in into the right panel (inside the large yellow box).
Ijms 18 00609 g002
Figure 3. Aberration calls and probes correlations. Correlation between the number of calls detected in each high density region and the number of probes selected in each region (upper) or the probe density (number of probes/size) of the selected region (bottom) considering any calls, including: single probe calls (A); or only multi probes calls with MAAD > 0.3 (B).
Figure 3. Aberration calls and probes correlations. Correlation between the number of calls detected in each high density region and the number of probes selected in each region (upper) or the probe density (number of probes/size) of the selected region (bottom) considering any calls, including: single probe calls (A); or only multi probes calls with MAAD > 0.3 (B).
Ijms 18 00609 g003
Table 1. Log2ratios correlations.
Table 1. Log2ratios correlations.
Selected PairsGroups ComparisonsNAll Log2ratiosLog2ratios > |0.3|p-Value
Mean rMean r
All samplesreplicated370.180.421.8 × 10−9
random370.070.140.0036
p-value 0.0044.8 × 10−5
Only pairs with at least one excellent quality sample (DLRS ≤ 0.2)replicated240.23 *0.53 §2.01 × 10−8
random240.09 **−0.17 §§0.0057
p-value 0.00186.8 × 10−6
Pairs with no excellent quality sample (DLRS > 0.2)replicated130.09 *0.21 §0.003
random130.05 **0.09 §§0.1594
p-value 0.26350.1492
* p = 0.0069; ** p = 0.2320; § p = 0.0009; §§ p = 0.2188.
Table 2. Aberrations detected.
Table 2. Aberrations detected.
Sample IDDLRSChromosomal Region (chr:start–end)CNV Type# ProbesDetection AlgorithmFuzzy ZeroVisual Inspection ClassificationReported on DGVReported on DecipherValidatedReplicateTrue Variants
ADM-2, Threshold 6ADM-2, Threshold 8
HSCR0000.1489:110381888–110401999gain9YYYlikelyNNYconfirmedyes
HSCR0000.14810:43435867–60812533loss849YYYknownNNknownconfirmedknown
HSCR0000.14810:43572551–43573368gain3NNNunlikelyNN not confirmedno
HSCR0370.12010:43589687–62786887loss544YYYknownNNknown known
HSCR0050.2267:84217007–84225649loss4Y--likelyY (freq < 1%)NY yes
HSCR0050.22610:43679892–43680816loss5Y-YlikelyNNN no
HSCR005 *0.22621:9833187–11096086loss4N-NpossibleNN unknown
HSCR0060.27610:43679612–43680816loss6N--likelyNNN no
HSCR0060.27610:43685614–43715348gain78NN-unlikelyNN unknown
HSCR0060.27619:5822193–5832504gain13Y--unlikelyNN unknown
HSCR0090.17610:43691613–43713132gain50NNNunlikelyNN unknown
HSCR0090.17619:5825458–5831976gain9YYYunlikelyNN unknown
HSCR010 *0.21115:20848460–22432687gain5Y--likelyY (freq ≥ 5%)N not excludedyes
HSCR0140.2218:32532001–32532545gain2Y-YunlikelyNN unknown
HSCR014 *0.22110:29939955–30822470gain3Y-YpossibleNN unknown
HSCR014 *0.22112:80226392–80589429gain2Y-YpossibleNN unknown
HSCR0140.22122:22417683–23228483loss15YYYlikelyY (freq ≥ 5%)N yes
HSCR0160.1175:69288477–70309855gain3Y--likelyY (freq ≥ 5%)N not excludedyes
HSCR0160.11722:25672585–25892401gain5Y-YlikelyY (freq ≥ 5%)Y (3 inds.) not excludedyes
HSCR018 §0.1729:109336464–109348467gain6---likelyNNY yes
HSCR019 *0.1221:146638075–147824207loss4YYYlikelyNY (1q21.1 recurrent microdel)Yconfirmed with a different sizeyes
HSCR033*0.22915:21162691–22173977loss3YYYlikelyY (freq ≥ 5%)N yes
HSCR0360.17722:22781091–23228483loss8YYYlikelyY (freq ≥ 5%)N yes
HSCR0390.2173:51458492–51665134loss62NN-unlikelyNN not confirmedno
HSCR0390.2176:148651353–150170473loss52NN-unlikelyNN not confirmedno
HSCR0390.2179:110130442–110370427loss99NN-unlikelyNN not confirmedno
HSCR043 §0.1759:109273643–109275694loss2---likelyNNY yes
HSCR045 §0.2717:84594683–84607065loss6---unlikelyNNN no
HSCR045 §0.2718:32597644–32598929loss3---likelyNNY yes
HSCR0450.27110:43679612–43680816loss6Y-YlikelyNNN no
HSCR0450.27119:5819037–18310693gain25Y--unlikelyNN unknown
HSCR0580.24322:18661724–18920001gain7Y-YunlikelyY (freq ≥ 5%)N not evaluableyes
HSCR064 *0.19215:20848460–22173977loss4YYYlikelyY (freq ≥ 5%)N yes
HSCR1260.17619:4205366–18310693gain26N--unlikelyNN unknown
HSCR146 *0.12215:58257674–59009890gain2YYYlikelyNNY yes
HSCR1460.12219:30888070–30891329gain2Y-YlikelyNNN no
HSCR160 *0.20015:20848460–22173977gain4Y-YlikelyY (freq ≥ 5%)N yes
HSCR162 *0.1849:43659247–43659512loss2---likelyY (freq ≥ 5%)N confirmed with a different sizeyes
HSCR181 *0.15015:20848460–22432687loss5N--possibleY (freq ≥ 5%)N not excludedyes
HSCR1810.15021:14629063–48080926gain245YYYknownNNknownconfirmedknown
HSCR1830.13822:22781091–23228483loss8YYYlikelyY (freq ≥ 5%)N yes
HSCR1950.1589:112078131–112089193loss5Y--likelyNNinconclusiveconfirmed with a different sizeyes
HSCR2170.16816:82200334–82202467gain2Y-YlikelyNNY yes
HSCR228 §0.15822:25672585–25892401gain5---likelyY (freq ≥ 5%)Y (3 inds.) not excludedyes
HSCR231*0.16415:21162691–22432687gain4Y-YunlikelyY (freq ≥ 5%)N yes
HSCR3120.2153:50161771–50618134gain143N--unlikelyNN unknown
HSCR3120.2154:41748211–41753993gain16N--unlikelyNN unknown
HSCR3120.21510:43550696–43621994gain196N--unlikelyNN unknown
HSCR3120.21510:43684681–43718450gain86NNNunlikelyNN unknown
HSCR3120.21514:36983123–36994136gain14Y--unlikelyNN unknown
HSCR3120.21519:5821171–5832504gain15NNNunlikelyNN unknown
HSCR3230.25313:78465278–78484576gain30N--unlikelyNN unknown
HSCR3310.17219:5822193–5832928gain14N--unlikelyNN not excludedunknown
HSCR335 *0.18315:20848460–22173977gain4Y--possibleY (freq ≥ 5%)N not excludedyes
HSCR3350.18322:18628019–18807881gain6Y-YunlikelyNN not excludedunknown
HSCR3350.18322:20345868–20499789gain4Y-YunlikelyY (freq ≥ 5%)N not excludedyes
HSCR3350.18322:21494163–21704972gain5Y-YunlikelyNN not excludedunknown
HSCR3490.2203:51452049–51647312loss59NN-unlikelyNN unknown
HSCR349 *0.2207:63449575–75986814loss25N--unlikelyNN unknown
HSCR3490.22010:43573685–43574005gain2Y-YunlikelyNNN no
HSCR3740.26610:43473690–43474033gain4Y-YunlikelyNNN no
HSCR3800.12322:16054691–18807881gain23YYYknownNNknown known
HSCR3800.12322:20345868–20659606gain5YYYunlikelyNN unknown
HSCR3800.12322:21494163–21704972gain5YYYunlikelyNN unknown
HSCR3820.23510:43474436–43483543loss29N--unlikelyNN unknown
HSCR3820.23510:43630181–43636329gain31N--unlikelyNN unknown
HSCR382 *0.23515:20190548–22173977gain5Y--possibleY (freq ≥ 5%)N yes
HSCR3910.17321:14629063–48080926gain245YYYknownNNknownconfirmed with a different sizeknown
HSCR403 §§0.1114:41746863–41751291loss11N--likelyNNY yes
HSCR403 *,§§0.1119:43659247–43659512gain2Y-YlikelyY (freq ≥ 5%)N yes
HSCR4030.11122:18661724–18807881gain5Y--possibleNN not excludedunknown
HSCR4030.11122:21494163–21704972gain5Y-YunlikelyNN confirmed and not excludedyes
HSCR4030.11122:23056562–23228483loss3Y-YlikelyY (freq ≥ 5%)N confirmed with a different sizeyes
HSCR409 *0.13915:20848460–22173977gain4YYYlikelyY (freq ≥ 5%)N yes
HSCR4120.20422:20345868–21778882loss26N--unlikelyNN not confirmedno
HSCR414 *0.15615:20848460–22432687loss5N--possibleY (freq ≥ 5%)N yes
HSCR4150.1959:113025039–113029430loss2YYYlikelyY (freq ≥ 5%)Y (1 ind.) yes
HSCR421 *0.1669:43659247–43659512loss2YYYlikelyY (freq ≥ 5%)N confirmedyes
HSCR4210.16622:25672585–25892401loss5YYYlikelyY (freq ≥ 5%)Y (3 inds.) not excludedyes
HSCR426 *0.1119:43659247–43659512loss2Y-YlikelyY (freq ≥ 5%)N not confirmed and confirmedyes
HSCR481 *0.2485:7656467–8124532loss2Y-YpossibleNN not confirmedno
HSCR4810.24819:31954093–31966036loss5Y--likelyNNYnot evaluableyes
HSCR4810.24821:14629063–48080926gain245YYYknownNNknownconfirmedknown
True (yes) = if either already reported on DGV, validated with different methods or confirmed on at least one replicate; (no) if not validated and/or not confirmed on replicate(s); known = selected controls or known chromosomal rearrangements; unknown = not possible to discriminate between true yes or no; * probes not located in the selected high density regions; § aberration not detected by the software call, but identified by visual inspection; Y = percentage of probes with absolute high log2ratio (≥0.5 for gains and ≤−0.8 for loss) above 33.3%; N = percentage ≤ 33.3%, - = not called by the algorithm; deletion reported as CNV with pathogenicity unknown, reported in an individual with aganglionic megacolon (another name for HSCR), intellectual disability and short stature; §§ aberrations assumed as detected because identified in two additional replicates.
Table 3. Detection filters comparison.
Table 3. Detection filters comparison.
Comparison GroupsTrue CallsNot ConfirmedUnknownTotalp-Value Likely/Possible vs. Unlikely or Called vs. Not Called by the Software *p-Values Thresholde ≥ 0.33 vs. below *
Likely/possible3954480.0003
Unlikely482335
ADM-2_th6 ≥ 0.33335612531.00000.0033 ††
ADM-2_th6 < 0.333361524
NO ADM-2_th6 (visual only) *5106
ADM-2_th8 ≥ 0.3331803210.53460.0001 ††
ADM-2_th8 < 0.3330459
NO ADM-2_th82591953
Fuzzy ≥ 0.3332868420.52300.2000
Fuzzy < 0.3330145
NO Fuzzy1561536
Total43132783
True calls include controls, aberrations reported on DGV, aberrations confirmed in at least a replicate and aberrations confirmed at validation. Not confirmed calls include aberrations not confirmed at validation and not found in the available replicate. Unknown includes calls not validated and not reported on DGV, for which a replicate sample was not available and that have not been evaluated in the statistical test. * p-value for true vs. not confirmed calls; likely/possible have a significantly higher chance of being true than those unlikely; †† the thresholde ≥ 0.33 filter has a better chance to discriminate between true and false calls, significant for the ADM-2 detection algorithm.
Table 4. Regions mapped on the aCGH and probe density.
Table 4. Regions mapped on the aCGH and probe density.
Kind of ProbesCandidate RegionLocus# of Features *# of Unique Probes *Average Space (nt) *
SelectedRET10q11.28138333312
9q319q3118242501
9p24.19p24.11423521
PHOX2B4p1349508
NRG18p12473501
SEMA3A/SEMA3D7q21.114682506
rs1270768240500
6q25.16q25.17143501
21q2221q2220248,297
3p213p2111413503
19q1219q1210853502
NRTN19p13.318806
16q23.316q23.37143501
NKX2-114q1317812
SOX1022q1327823
22q11.222q11.216249,383
ECE11p36.1103806
ZEB22q22.3165923
EDNRB13q22112804
GDNF5p13.1-p1242810
EDN320q13.2-q13.344808
Genome 31493130971,074
Replicates 301 × 5 = 1505301
Normalization 12621262
Agilent controls 1482
Total 15,74813,026
* Twenty-two probes selected among the high density panel were also included in the normalization set or in the replicates set and are not reported among the # of unique probes selected, but considered for the average coverage. Nineteen probes selected in the rest of the genome had already been selected for the high density regions (10) or already part of the normalization set (9).

Share and Cite

MDPI and ACS Style

Lantieri, F.; Malacarne, M.; Gimelli, S.; Santamaria, G.; Coviello, D.; Ceccherini, I. Custom Array Comparative Genomic Hybridization: the Importance of DNA Quality, an Expert Eye, and Variant Validation. Int. J. Mol. Sci. 2017, 18, 609. https://doi.org/10.3390/ijms18030609

AMA Style

Lantieri F, Malacarne M, Gimelli S, Santamaria G, Coviello D, Ceccherini I. Custom Array Comparative Genomic Hybridization: the Importance of DNA Quality, an Expert Eye, and Variant Validation. International Journal of Molecular Sciences. 2017; 18(3):609. https://doi.org/10.3390/ijms18030609

Chicago/Turabian Style

Lantieri, Francesca, Michela Malacarne, Stefania Gimelli, Giuseppe Santamaria, Domenico Coviello, and Isabella Ceccherini. 2017. "Custom Array Comparative Genomic Hybridization: the Importance of DNA Quality, an Expert Eye, and Variant Validation" International Journal of Molecular Sciences 18, no. 3: 609. https://doi.org/10.3390/ijms18030609

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop