Next Article in Journal
Combination of On-line pH and Oxygen Transfer Rate Measurement in Shake Flasks by Fiber Optical Technique and Respiration Activity MOnitoring System (RAMOS)
Next Article in Special Issue
Application of Multiplexed FBG and PZT Impedance Sensors for Health Monitoring of Rocks
Previous Article in Journal
An Overview of Label-free Electrochemical Protein Sensors
Previous Article in Special Issue
Tempo-Spatial Patterns of Land Use Changes and Urban Development in Globalizing China: A Study of Beijing
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Use of a Dynamic Enclosure Approach to Test the Accuracy of the NDIR Sensor: Evaluation Based on the CO2 Equilibration Pattern

1
Dept. of Earth & Environmental Sciences, Sejong University, Seoul, Korea
2
Dept. of Applied Mathematics, Sejong University, Seoul, Korea
*
Author to whom correspondence should be addressed.
Sensors 2007, 7(12), 3459-3471; https://doi.org/10.3390/s7123459
Submission received: 7 November 2007 / Accepted: 19 December 2007 / Published: 20 December 2007
(This article belongs to the Special Issue Sensors for Urban Environmental Monitoring)

Abstract

:
As part of a quality assurance (QA) study for sensor systems, an enclosure approach is applied to assess the accuracy of non-dispersive infrared (NDIR)-based CO2 sensors. To examine the performance of the sensor system, an enclosure chamber containing six sensor units of the two model types (B-530 and H-500) was equilibrated with calibrated CO2 standards at varying concentration levels. Initially, the equilibration pattern was analyzed by CO2-free gas (0 ppm) at varying flow rates (i.e., 100, 200, 500, and 1000 mL min-1). Results of the test yielded a highly predictable and quantifiable empirical relationship as a function of such parameters as CO2 concentration, flow rate, and equilibration time for the enclosure system. Hence, when the performance of the NDIR-method was evaluated at other concentrations (i.e., 500 and 1000 ppm), all the sensor units showed an excellent compatibility, at least in terms of the correlation coefficients (r > 0.999, p = 0.01). According to our analysis, the NDIR sensor system seems to attain an overall accuracy near the 5% level. The relative performance of the NDIR sensor for CO2 analysis is hence comparable with (or superior to) other methods previously investigated. The overall results of this study indicate that NDIR sensors can be used to provide highly accurate and precise analyses of CO2 both in absolute and relative terms.

1. Introduction

It is widely understood that carbon dioxide plays an important role as a greenhouse gas for its ability to absorb a wide range of the infrared light along with a long residence time in the Earth's atmosphere [1]. Because of increasing anthropogenic activities (such as the combustion of fossil fuels and deforestation), the concentration of atmospheric carbon dioxide has been gradually raised to reach a global mean of 380 ppm, which is about 35% higher than that seen at the beginning of the industrialization age [2]. It is hence estimated that atmospheric CO2 is currently responsible for up to 63% of radiative forcing with a prominent rise in its concentration level [3].
Because of the unique physicochemical properties of CO2, its distribution in the atmosphere is found to maintain a consistent but dynamic variation on both spatial and temporal scales. The concentration of CO2, while experiencing a gradual increase through the years, tends to be subject to a strong seasonality. CO2 levels, when compared on a spatial scale, also exhibit a unique pattern in which enhanced values are prevalent in urban areas relative to rural areas [4]. Likewise, CO2 levels in indoors are commonly found to be several times higher than those in the outdoor atmosphere [5]. Because of the environmental significance of CO2 pollution under both outdoor and indoor conditions, it is highly desirable to acquire accurate information concerning the real concentration levels of CO2.
The non-dispersive infrared (NDIR) sensors have been used widely in the real time monitoring of CO2 in air [6, 7]. In a recent study, we investigated the relative performance of these sensors for the real-near time analysis of CO2 under open air conditions in the laboratory [8]. Based on the concurrent analysis of several sensors, we were able to demonstrate that these sensors can be employed to yield fairly unbiased data between different sensor units. In the present study, we intended to examine the accuracy of the NDIR-based sensor method as a continuation of our efforts to validate the application of such a sensor system. To this end, a series of experiments were conducted by equilibrating an enclosure system with calibrated CO2 gaseous standards at known concentrations. The results of these calibration experiments were then used to explore the basic properties of the NDIR-based sensor method.

2. Materials and methods

2.1. Schematic of the analytical setup

In this study, the analytical performance of CO2 sensors was evaluated using two types of the NDIR sensor model (B-530 and H-550: ELT Co., Korea). Each model is distinguished in terms of their detection range and response time. The maximum detection ranges are 10,000 and 50,000 ppm, respectively. In addition, their initial response times are less than 120 and 30 sec, respectively. More details concerning the sensor types, instrumental set-up, and quality control procedures have been described elsewhere [8]. As a simple means to examine the performance of the sensor units, an enclosure system was developed in which NDIR sensors were exposed to constant levels of CO2 standards prepared at varying concentration ranges (0, 500, and 1000 ppm) (refer to Fig 1). The CO2 gaseous standards used in this study were purchased at three concentration levels of 0, 497.9, and 999.9 μmol mol-1 (Deok Yang Energen Co., Korea). As a platform to conduct the accuracy test, a vacuum enclosure built to serve as a sensor repository system was used to concurrently retrieve CO2 data from six individual sensors (i.e., three identical sensor units of each model introduced above) (Fig 1). The enclosure box inhabiting all the sensor units was built with nearly 35 L capacity (a dimension of 47.8×38.0 ×19.4 cm) and connected to the standard gas cylinder to supply CO2 gases at a constant flow rate. Once the sensor units are connected to the data acquisition system, the punched holes in the enclosure were completely sealed with silicon adhesives. To provide the constant flow of CO2 gases into this enclosure box, flow rates of standard gases were regulated by the union valve system in the inlet.

2.2. Experimental scheme

After establishing all the experimental set-up for the enclosure system, the basic performance of this system was tested by supplying pure air (CO2-free air) contained in a cylinder to the inlet of the enclosure box at four different flow rates (i.e., 100, 200, 500 and 1000 mL min-1). By flushing the container (initially equilibrated with the laboratory air at CO2 concentrations of approx. 400-450 ppm) with CO2-free air, the general circulation cycle to attain equilibrium was assessed as a function of flow rate and time (Table 1). The concentration of CO2 in the container decreased gradually with the mixing of the CO2-free air until the 0 ppm was recorded (Fig 2). Based on our initial experiment with CO2-free air, the time required for the system equilibration was quantified at each individual flow rate.
Upon completing the initial test with CO2-free air, the experiments proceeded to the next stage with two known concentrations of CO2 standards (500 and 1000 ppm). For all experiments in the second stage, the NDIR sensors were tested at a single flow rate of 500 mL min-1 (Table 1). The moderate flow rate of 500 mL min-1 was selected because it was plausible to attain the system equilibration under reasonably long durations (Fig 3). The measurements for each of the two CO2 standards (500 and 1000 ppm levels) were initiated immediately after equilibrating the container with the CO2-free air (To make a parallel comparison between different experiments, CO2 values for each experiment were forced to rise from the zero ppm level). Knowing that the duration for equilibration at a flow rate 500 mL min-1 requires approx. 300 min, the experiments for those two concentration levels were conducted well above the predicted equilibration time (i.e., up to 360 min duration). This was intended to ensure the complete equilibration of the enclosure system at given CO2 concentration levels.
The CO2 data sets for each experiment were initially retrieved at each 3-second interval but analyzed after being converted into 10-min averaged values (by pooling them together for each 10-min interval). This 10-min conversion was made to optimize the handling of CO2 data with the Excel program from raw data sets with enormously large number. To examine the reliability of this 10-min conversion, the variabilities of two different data sets (primary 3-sec vs. converted 10-min data) were compared using the results of a 1000-ppm CO2 analysis. The data sets with two different intervals, taken after the equilibration (i.e., after 300 min), were compared in terms of the relative standard error (RSE %) values of all the six sensor units, the results were highly comparable between the raw (3.02 ± 0.10%) and the converted data sets (2.68 ±0.02%). When the reproducibility of our individual sensor model is assessed based on the repetitive measurement of an equilibrated CO2 sample, it ranged from 0.08 (H1) to 0.17% (H3) for H-500 and 0.10 (B1) to 0.13% (B3) for the B-530 model. Hence, all of our analysis of these sensor systems was mainly made using these 10-min converted data sets.

3. Results and discussions

3.1. The general pattern of CO2 circulation in the container

In order to investigate the dynamic equilibration pattern of the enclosure system investigated in this study, a series of preliminary tests were conducted initially using CO2-free air. As shown in Table 1, the results of this test conducted at four different flow rates (i.e., 100, 200, 500, and 1000 mL min-1) demonstrated that all the sensor units exhibited highly systematic patterns at each individual flow rate (Fig 2). As shown in Fig 2, log (CO2) values tend to maintain strong inverse correlations with time. Hence, to learn more about the fundamental features of CO2 circulation in an enclosure system, these experimental results were utilized to derive the empirical relationship between all the experimental variables.
In order to describe the exchange dynamics of CO2 gas in a quantitative manner, the experimental data acquired from the CO2-free gas were used to fit the non-linear equations via a trial and error. As a result, the equation (1) was derived to equally consider the major parameters involved in this exchange process:
C = [ CO 2 ( t ) / CO 2 ( i ) ] = { 1 0.00015 * t 0.00000629 * t * FR } 3.5724 ,
where [1] the concentration of CO2 expressed as the ratio between a given time (CO2 (t)) and the initial time (CO2 (i)) (unitless), [2] flow rate (FR) of CO2-free gas (mL min-1), and [3] time (t) for the equilibration (min).
From this equation, the equilibration time to reach the CO2 concentration (of zero or any other concentration values) at a given flow rate (mL min-1) can be computed by the following equation:
t = 1 / ( 0.00015 + 0.00000629 FR )
According to the above formula, the equilibration times for all four flow rates (100, 200, 500, and 1000 mL min-1) are computed as 1283, 710, 303, and 155 minutes, respectively (Table 2).
To evaluate the efficacy of these model-fit equations, a parallel comparison was made between observed and predicted values of CO2 ratio (Fig 3). As seen in the Fig 3, the observed route of CO2 equilibration in the enclosure system shows an excellent agreement with the predicted pattern at each of all flow rates.

3.2. The accuracy of NDIR-CO2 analysis

In an attempt to investigate the analytical bias arising from the application of the NDIR-system, the accuracy of the NDIR method was assessed using CO2 standards prepared at the two concentration levels of 500 and 1000 ppm. Based on our initial equilibration experiments using CO2-free air, all experiments representing CO2 levels of 500 and 1000 ppm were conducted at a moderate flow rate of 500 mL min-1. As shown in Fig 4, the concentration of CO2 increased systematically with time to reach their respective equilibration concentrations (i.e., near 500 and 1000 ppm) after the equilibration point (approx. 300 min). The observed pattern suggests that the NDIR sensors behave systematically to detect the changes in the CO2 concentrations. As summarized in Table 2A, the results taken from the 500 ppm CO2 standard gas showed an average CO2 concentration of 477 ± 44.9 ppm with a range of 427-536 ppm after the equilibration time (using the data taken between 300 and 360 min). Likewise, the experimental data for the 1000 ppm CO2 standard, taken for the similar duration, averaged 954 ± 62.7 ppm with a range of 885-1031 ppm.
The CO2 data obtained from these comparative experiments were used to compute the percent deviation (PD) values of each sensor unit. The PD values were calculated by subtracting the CO2 concentration value for a given unit by the known value of CO2 for the respective gaseous standard:
P D = [ ( CO 2 ( obs ) CO 2 ( known ) ] [ CO 2 ( known ) ] × 100
where CO2(obs) is the mean CO2 value observed by all sensor units, and CO2(known) is the actual CO2 concentration of standard (i.e., 500 and 1000 ppm). The PD values obtained using the CO2 values in the equilibrium stage (during the 300 to 360 min of the study duration) are summarized in Table 2.
As shown in Table 2B, the sensor units of the model type B-530 showed a negative bias for the 500 ppm CO2 measurement in terms of their PD values (ranging from -14.6 % (B3) to -9.39 % (B1)). In contrast, H-500 units generally exhibited a positive bias, with PD values of +7.30 (H1) and +4.83% (H2) (with the exception of the H3 unit). The results of 1000 ppm experiment in fact showed patterns that are highly analogous to the 500 ppm experiment.
As another means to assess the analytical bias of NDIR sensor units, the percent recovery (PR) rates of each sensor unit were computed and plotted for CO2 standards at both concentration levels (500 and 1000 ppm) (Fig 5). The PR is calculated by dividing the mean CO2 concentration (after the equilibration time) with the known standard concentration (i.e., after approx. 300 min onwards for 60 min; refer to Table 2 for details). The PR value of the 500 ppm CO2 standard averaged 95.4 ± 8.95 % with a range of about. 86 to 107 %. Similarly, the PR values for the 1000 ppm test averaged 95.4 ± 6.27 % with a range of about 88 to 103 %. As such, the accuracy of NDIR sensor investigated in this study seems to approach the 5% level, if the accuracy is expressed as the difference between the unity and the recovery ratio. Although the absolute accuracy of the NDIR-based CO2 sensors seemed to improve slightly at the 1000 ppm level, their variabilities do not appear to be significant with increasing concentration levels. The results of the present study thus indicate that the performance of the NDIR method is fairly stable and reproducible in the tested concentration ranges in terms of its precision and accuracy.
To evaluate the compatibility of each sensor unit in the CO2 measurement, a correlation analysis was performed using the CO2 concentration data acquired by all sensor units at each flow rate. The results of the correlation analyses for three different concentration levels (i.e., 0, 500, and 1000 ppm) are summarized in Table 3. As shown in Table 3, all the sensor units showed highly significant correlations with each other in terms of their correlation strengths (i.e., R2 > 0.99, p = 0.01). Although the correlation results of the CO2-free air were the strongest of all, the results of high CO2 levels (500 or 1000 ppm) were also indistinguishable in that respect; all sensor units showed correlation coefficients approaching the unity (> 0.999), irrespective of the model types. Hence, as seen in our earlier study based on side-by-side monitoring of ambient CO2 levels under the laboratory conditions [8], we were able to obtain highly comparable patterns between different sensor units. The findings of relatively enhanced compatibility between different sensor units or with CO2 standard gases may be ascribable to the use of a stable equilibrium system developed for the comparative test.

3.3 Comparison between different detection techniques

As a simple means to assess the analytical reliability of our sensor method in the CO2 analysis, the previous records of diverse measurement techniques involved in CO2 analysis were explored in terms of accuracy and precision (Table 4). According to this analysis, it is found that many scientists relied on the Gas chromatographic (GC) method for the CO2 analysis. For instance, Ekeberg et al (2004) attempted to quantify the precision and accuracy of certified CO2 standards by GC with mass spectrometry (GC-MS) [9]. The results of their GC-MS analysis revealed accuracy near ± 4% for a 977 μL/L certified CO2 standard; their results appear to be highly comparable to the overall mean accuracy observed in this study (i.e., 4.60%), which was in fact determined from a comparable CO2 concentration range (i.e., 1000 ppm) using all six sensor units (Table 2B). It is also found that the overall mean precision of the repetitive analysis made by GC-MS (2.30%) is highly comparable with those derived by our sensor method (2.68%). In comparison, the GC-TCD method for CO2 analysis has shown an accuracy of ± 5.3% in the range of 2270-10000 ppm [10]. The precision of the GC-TCD and GC-FIA techniques was also estimated by the repetitive measurement of CO2 at a constant concentration level (approx. 340 ppm), and the results were found to be 6.20% and 5.98%, respectively [11].
In a continuous-flow isotope ratio mass spectrometric (CF-IRMS) analysis of CO2, the accuracy was estimated in the 0.18-0.38 range as differences between the means of two systems, i.e., dual-inlet isotope ratio mass spectrometry/nondispersive infrared gas analysis system (DI-IRMS/NDIR) and automated system of CF-IRMS in the analysis of five CO2 test gases covering a concentration range of 328 to 603 μmol mol-1 [12]. The overall precision of CF-IRMS was estimated at 0.04% for the five CO2 test gases covering a concentration range of 328 to 603 μmol mol-1 [12]. A comparative study of a CO2 analysis (in the 800-1000 ppm range), conducted between three brands of detector tubes (Draeger CH 30801, Kitagawa 126, and Gastec 2LL), yielded relative standard error values in the 5-7% range [13]. In comparison, the relative error in the CO2 measurement was reported at -3.0% by the continuous-flow method using a conductimetric detector [14].
Considering the performance of various measurement techniques introduced in the previous studies, the NDIR method investigated in the present study appears to be fairly reliable and capable of the near real-time analysis of CO2 in terms of its accuracy and precision. In view of the fact that other techniques (such as GC-TCD, GC-MS, and CF-IRMS, etc.) should consider the delicate arrangement for instrumental management (complicated operating procedures, high cost, duration for the data acquisition, etc), the simple, economic NDIR sensor technique can be comparatively advantageous from a number of aspects (e.g., affordability, handling, and rapid measurement time).

4. Conclusions

In the present study, an effort was made to evaluate the accuracy of the NDIR sensor by equilibrating an enclosure system with CO2 gases at known concentration levels. As the basic tool for such a purpose, a dynamic enclosure system was built to attain CO2 equilibrium at different CO2 concentration levels. The status of the equilibrium was measured by 6 sensor units with two different model types. By acquiring the equilibrium concentration data of CO2 by the enclosure system, we were able to derive empirical equations in which the CO2 concentration can be predicted at a given time and flow rate. The accuracy of the NDIR-sensor was hence assessed by comparing the observed and calibrated CO2 concentration levels derived based on our experimental design.
According to the analysis of calibrated CO2 gases at 500 and 1000 ppm, the NDIR system was found to maintain approx. 5% accuracy. Moreover, when the CO2 data obtained by all the sensor units were compared, an excellent compatibility was maintained throughout the entire side-by-side analysis. The results of correlation analysis indicated that all sensor units had a correlation coefficient approaching the unity (> 0.999 at 0.01 level), regardless of model type or concentration range. The overall results of this study thus suggest that NDIR sensors can be used to produce highly reliable data sets for CO2 analysis in both absolute and relative senses. It should also be pointed out that the concentration range examined in this study (0 to 1000 ppm) covers the current atmospheric CO2 levels (approx. 380 ppm), as reported by the WMO and the IPCC. Hence, one can explore the possibility of utilizing these sensors in the real time monitoring of atmospheric CO2, as they have been proven to be competent in terms of experimental performance, like other instrumental methods with proven reliability.

Acknowledgments

This work was supported by a grant from the Korea Research Foundation (KRF-2006-341-C00026) funded by the Korean government (MOEHRD). All authors appreciate the excellent comments raised by reviewers.

References and Notes

  1. Forster, P.; Ramaswamy, V.; Artaxo, P.; Berntsen, T.; Betts, R.; Fahey, D.W.; Haywood, J.; Lean, J.; Lowe, D.C.; Myhre, G.; Nganga, J.; Prinn, R.; Raga, G.; Schulz, M.; Van Dorland, R. 2007; Changes in Atmospheric Constituents and in Radiative Forcing; In: Climate Change: The Physical Science Basis. Contribution of Working Group I to the Fourth Assessment Report of the Intergovernmental Panel on Climate Change; Solomon, S., Qin, D., Manning, M., Chen, Z., Marquis, M., Averyt, K.B., Tignor, M., Miller, H.L., Eds.; Cambridge University Press: Cambridge, United Kingdom and New York, NY, USA, 2007.
  2. WMO. The 13th WMO/IAEA Meeting of Experts on Carbon Dioxide Concentration and Related Tracers Measurement Techniques, TD No. 1359. 2006. Boulder, Colorado, USA, 19-22 September 2005.
  3. Raupach, M.R.; Marland, G.; Ciais, P.; Le Que´ re´, C.; Canadell, J.G.; Klepperand, G.; Field, C.B. Global and regional drivers of accelerating CO2emissions. PNAS 2007. 10.1073.pnas.0700609104. [Google Scholar]
  4. Nemitz, E.; Hargreaves, K. J.; McDonald, A. G.; Dorsey, J. R.; Fowler, D. Micrometeorological measurements of the urban heat budget and CO2 emissions on a city scale. Environmental Science and Technology 2002, 36, 3139–3146. [Google Scholar]
  5. Kovesi, T.; Gilbert., N.L.; Stocco, C.; Fugler, D.; Dales, R.E.; Guay, M.; Miller, D.J. Indoor air quality and the risk of lower respiratory tract infections in young Canadian Inuit children. Canadian Medical Association or its Licensors 2007, 177, 155–60. [Google Scholar]
  6. Wong, J. Y. NDIR Gas Sensor. US Patent No. 5,444,249, 22 1995. Aug 1995.
  7. Yi, S.H.; Park, Y.W.; Han, S.O.; Min, N.; Kim, E.S.; Ahn, T.W. Novel NDIR CO2 sensor for indoor Air quality monitoring. The 13th International Conference on Solid-State Sensors, Actuators and Microsystems, Seoul, Korea, June 5-9, 2005; 2005. [Google Scholar]
  8. Pandey, S.K.; Kim, K.H. The Relative Performance of NDIR-based Sensors in the Near Real-time Analysis of CO2 in Air. Sensors 2007, 7, 1683–1696. [Google Scholar]
  9. Ekeberg, D.; Ogner, G.; Fongen, M.; Jonerc, E.J.; Wickstrømc, T. Determination of CH4, CO2 and N2O in air samples and soil atmosphere by gas chromatography mass spectrometry, GC-MS. Journal of Environmental Monitoring. 2004, 6, 621–623. [Google Scholar]
  10. NIOSH manual of analytical methods (NMAM), fourth edition; Carbon dioxide: Method 6603; Issue 2, 1994.
  11. Satienperakul, S.; Cardwell, T.J.; Cattrall, R.W.; McKelvie, I.D.; Taylor, D.M.; Kolev, S.D. Determination of carbon dioxide in gaseous samples by gas diffusion-flow injection. Talanta 2004, 62, 631–636. [Google Scholar]
  12. Schauer, A.J.; Lott, M.J.; Cook, C.G.; Ehleringer, J.R. An automated system for stable isotope and concentration analyses of CO2 from small atmospheric samples. Rapid. Commun. Mass. Spectrom. 2005, 19, 359–362. [Google Scholar]
  13. Norback, D.; Ancker, K.; Johanson, G. Field Evaluation of CO2 Detector Tubes for Measuring Outdoor Air Supply Rate in the Indoor Environment. Indoor Air 1992, 2, 58–64. [Google Scholar]
  14. Calegario, F.F.; Cosso, R.G.; Almeida, F.V.; Vercesi, A.E.; Jardim, W.F. Determination of the respiration rate of tomato fruit using flow analysis. Postharvest Biology and Technology 2001, 22, 249–256. [Google Scholar]
Figure 1. A schematic diagram of the enclosure system for the evaluation of CO2 equilibration pattern based on the NDIR. sensor method: [A] closed view and [B] transparent view showing the sensor components of the system.
Figure 1. A schematic diagram of the enclosure system for the evaluation of CO2 equilibration pattern based on the NDIR. sensor method: [A] closed view and [B] transparent view showing the sensor components of the system.
Sensors 07 03459f1
Figure 2. A plot for CO2 equilibration pattern derived by an enclosure system investigated in this study: tests were conducted at four different flow rates (FR (mL min-1)) of CO2-free air shown in the legend box. Results are drawn as log (CO2) concentration vs. time (linear regression equations derived at each flow rate).
Figure 2. A plot for CO2 equilibration pattern derived by an enclosure system investigated in this study: tests were conducted at four different flow rates (FR (mL min-1)) of CO2-free air shown in the legend box. Results are drawn as log (CO2) concentration vs. time (linear regression equations derived at each flow rate).
Sensors 07 03459f2
Figure 3. A parallel comparison of the observed and predicted equilibration pattern of CO2 concentration ratio; results are shown as the concentration ratio for the derivation of empirical formula The predicted CO2 values are shown as lines.
Figure 3. A parallel comparison of the observed and predicted equilibration pattern of CO2 concentration ratio; results are shown as the concentration ratio for the derivation of empirical formula The predicted CO2 values are shown as lines.
Sensors 07 03459f3
Figure 4. The performance of the CO2 sensors between different sensor units; tests were made with two known concentrations of CO2 gaseous standards at [A] 500 and [B] 1000 ppm. All dotted lines show the absolute concentration values of calibrated CO2 standard for the given test.
Figure 4. The performance of the CO2 sensors between different sensor units; tests were made with two known concentrations of CO2 gaseous standards at [A] 500 and [B] 1000 ppm. All dotted lines show the absolute concentration values of calibrated CO2 standard for the given test.
Sensors 07 03459f4
Figure 5. Percent recovery (PR) computed for each sensor unit based on the NDIR-based analysis of CO2. Dotted line denotes the theoretical recovery rate (100%). The standard error bars for the repetitive measurement of CO2 are shown for each individual sensor unit.
Figure 5. Percent recovery (PR) computed for each sensor unit based on the NDIR-based analysis of CO2. Dotted line denotes the theoretical recovery rate (100%). The standard error bars for the repetitive measurement of CO2 are shown for each individual sensor unit.
Sensors 07 03459f5
Table 1. Experiment schedule in this study for the analysis of CO2 equilibrium using a dynamic enclosure system
Table 1. Experiment schedule in this study for the analysis of CO2 equilibrium using a dynamic enclosure system
[A] Initial test to evaluate the equilibrium pattern of he enclosure system using CO2-fee air

ExperimentCO2 concentration (ppm)Flow rate (mL min-1)Exp dateTotal duration (min)Predicted duration for equilibration (min)1]
10100007. Sep 071201283
2050007. Sep 073001283
3020008. Sep 07650303
4010009. Sep 071140155
[B] Comparison with the calibrated CO2 standards at two concent:
550050013. Sep 07360303
6100050014. Sep 07360303
1]Calculation based on empirical equation by considering total volume of the box (i.e., approx. 35 L) containing all the sensors.
Table 2. Comparison of analytical bias of the NDIR sensor system: Results are shown in terms of both [A] CO2 concentration level and [B] percent deviation (PD) values derived by all 6 sensor units employed in the analysis of CO2.
Table 2. Comparison of analytical bias of the NDIR sensor system: Results are shown in terms of both [A] CO2 concentration level and [B] percent deviation (PD) values derived by all 6 sensor units employed in the analysis of CO2.
[A] CO2 concentration data acquired after the equilibration time (300-360 min duration) 1]

Concentration levelSensor unitsGrand mean

B12]B2B2H13]H2H3
[A] 500 ppmMean ± SD (Median)453 ± 1.11 (453)441 ± 1.41 (441)427 ± 1.90 (427)536 ± 1.23 (536)524 ± 1.00 (524)483 ± 2.16 (483)477 ± 44.9 (468)
(N=6)Range451-454439-443424-430534-537522-525481-486427-536

[B] 1000 ppmMean ± SD (Median)919 ± 2.42 (920)903 ± 2.59 (903)885 ± 2.87 (885)1031 ± 2.07 (1031)1025 ± 3.22 (1025)963 ± 3.98 (964)954 ± 62.7 (941)
(N=6)Range915-921900-906880-8881028-10341020-1029958-969885-1031
[B] PD values derived after the equilibration time (300-360 min duration) 1]

Concentration levelSensor unitsGrand mean4]

B12]B2B3H13]H2H3

[A] 500 ppmMean ± SD (Median)-9.39 ± 0.22 (-9.33)-11.7 ± 0.28 (-11.7)-14.6 ± 0.38 (-14.6)7.30 ± 0.10 (7.33)4.83 ± 0.08 (4.85)-3.30 ± 0.18 (-3.38)8.52 ± 4.25 (8.34)
(N=6)Range-9.78 ∼ -9.17-12.2 ∼ -11.4-15.1 ∼ -13.96.93-7.554.50-5.06-3.69 ∼ -2.663.30-14.6

[B] 1000 ppmMean ± SD (Median)-8.03 ± 0.24 (-7.91)-9.65 ± 0.26 (-9.65)-11.5 ± 0.29 (-11.5)3.15 ± 0.21 (3.15)2.53 ± 0.32 (2.59)-3.62 ± 0.40 (-3.58)6.41 ± 3.81 (5.83)
(N=6)Range-8.46 ∼ -7.85-10.0 ∼ -9.35-11.9 ∼ -11.22.88-3.462.03-2.89-4.15 ∼ -3.052.53-11.5
Table 3. Results of correlation analysis between CO2 concentration data (10-min converted data) derived at 3 concentration ranges: [A] 0, [B] 500, and [C] 1000 ppm.
Table 3. Results of correlation analysis between CO2 concentration data (10-min converted data) derived at 3 concentration ranges: [A] 0, [B] 500, and [C] 1000 ppm.
[A] Results of CO2- free air (CO2 = 0 ppm)

B1B2B3H1H2H3
B11.000
B21.0001.000
B31.0001.0001.000
H10.9970.9980.9981.000
H20.9970.9980.9981.0001.000
H30.9950.9960.9971.0001.0001.000
Table 3. Results of correlation analysis between CO2 concentration data (10-min converted data) derived at 3 concentration ranges: [A] 0, [B] 500, and [C] 1000 ppm.
[B] Results of the intermediate range (CO2 = 500 ppm)

B1B2B3H1H2H3
B11.000
B21.0001.000
B31.0001.0001.000
H10.9990.9990.9991.000
H20.9990.9990.9991.0001.000
H30.9990.9990.9991.0001.0001.000
Table 3. Results of correlation analysis between CO2 concentration data (10-min converted data) derived at 3 concentration ranges: [A] 0, [B] 500, and [C] 1000 ppm.
[C] Results of the highest range (CO2 = 1000 ppm)

B1B2B3H1H2H3
B11.000
B21.0001.000
B31.0001.0001.000
H11.0001.0001.0001.000
H21.0001.0001.0001.0001.000
H31.0001.0001.0001.0001.0001.000
All correlations are significant at 0.01 level (N= 30)
All correlations are significant at 0.01 level (N= 36)
All correlations are significant at 0.01 level (N= 36)
Table 4. A comparative evaluation of various detection methods in the CO2 analysis; all data compared in terms of precision and accuracy.
Table 4. A comparative evaluation of various detection methods in the CO2 analysis; all data compared in terms of precision and accuracy.
MethodMeasurement conditionMean (ppm)SDNAccuracyPrecision RSE (%)Reference
NDIR-sensorEnclosure system95462.764.602.68This study
NDIR-sensorEnclosure system47744.964.603.84This stduy
NDIR-sensorLaboratory429336NA2.33Pandey and Kim [8]
GC-MSCertified CO2 standard101674103.992.30Ekeberg et al.[9]
CF-GC/IRMSTest cylinders328-6030.33-0.61100.18-0.380.33-0.61Schauer et al.[12]
GC-TCDCO2 gas sampling bags2270-10,000NANA5.30.014NIOSH [10]
GD-FIAIn open air (Laboratory)338353NA5.98Satierperakul [11]
GC-TCDIn open air (Laboratory)335363NA6.2Satierperakul [11]
GD-FIAUndercover car parking56593NA0.92Satierperakul [11]
GC-TCDUndercover car parking554153NA1.56Satierperakul [11]
Detector tube anlysisIndoor air800-1000NANANA5-7Norback [13]
1] Accuracy is expressed as the difference between the unity and recovery ratio.
2] Data not available.
3] Accuracy is expressed in terms of the differences between the means of two systems, i.e., DI-IRMS/NDIR and CF-IRMS.
4] Precision is expressed as the standard deviation of the samples analyzed on the automated systems.

Share and Cite

MDPI and ACS Style

Pandey, S.K.; Kim, K.-H.; Lee, S.-H. Use of a Dynamic Enclosure Approach to Test the Accuracy of the NDIR Sensor: Evaluation Based on the CO2 Equilibration Pattern. Sensors 2007, 7, 3459-3471. https://doi.org/10.3390/s7123459

AMA Style

Pandey SK, Kim K-H, Lee S-H. Use of a Dynamic Enclosure Approach to Test the Accuracy of the NDIR Sensor: Evaluation Based on the CO2 Equilibration Pattern. Sensors. 2007; 7(12):3459-3471. https://doi.org/10.3390/s7123459

Chicago/Turabian Style

Pandey, Sudhir Kumar, Ki-Hyun Kim, and Sun-Ho Lee. 2007. "Use of a Dynamic Enclosure Approach to Test the Accuracy of the NDIR Sensor: Evaluation Based on the CO2 Equilibration Pattern" Sensors 7, no. 12: 3459-3471. https://doi.org/10.3390/s7123459

Article Metrics

Back to TopTop