Chopper Stabilized Low Resistance Comparator

by
Radojle M. Radetic
1 and
Dragan R. Milivojevic
2,*
1
Serbian Transmission System / Nade Dimic 40, 19210 Bor, Serbia
2
Mining and Metallurgy Institute, Department of Informatics / Zeleni Bulevar 35, 19210 Bor, Serbia
*
Author to whom correspondence should be addressed.
Sensors 2009, 9(4), 2491-2497; https://doi.org/10.3390/s90402491
Submission received: 18 March 2009 / Revised: 6 April 2009 / Accepted: 7 April 2009 / Published: 9 April 2009
(This article belongs to the Section Chemical Sensors)

Abstract

:
The paper describes an improvement of the chopper method for elimination of parasitic voltages in a low resistance comparison and measurement procedure. The basic circuit diagram along with a short description of the working principle are presented and the appropriate low resistance comparator prototype was designed and realized. Preliminary examinations confirm the possibility of measuring extremely low voltages. Very high accuracy in resistance comparison and measurement is achieved (0.08 ppm for 1,000 attempts). Some special critical features in the design are discussed and solutions for overcoming the problems are described.

1. Introduction

Standard methods for measurement and low resistance comparison are based on Thompson (Kelvin) bridges, DC current comparators, potentiometers, etc [1]. The new generation high quality multimeters can also be used for this kind of measurement [2,3]. In this paper an original low resistance comparator design is presented.
The circuit diagram of the proposed chopper stabilized comparator is given in Figure 1. The current flow through two serial connected resistors (measured RX and reference RR) provides the adequate voltages in them, UX and UR, respectively. As known, the voltage ratio on the resistor terminals is equal to their resistances ratio. In that case, the value of measured resistance RX could be expressed as:
R X = R R   U X U R
The maximum relative measurement error is:
| δ R X | = | Δ R R R RT | + | Δ U X U XT | + | Δ U R U RT | = | δ R R | + | δ U X | + | δ U R |
where δRR is reference resistance error (could be less than 10−6) and δUX, δUR are measurement errors for the appropriate voltages.
In low resistance measurement (four-terminal resistors [4]), the voltage on the potential terminals should theoretically be RI, but in a real measurement environment, except for the RI value, parasitic voltages can occur, such as thermo-electric, noise etc. All these unwanted influences can cause measurement errors and need to be eliminated.
Thermo-electric or Peltier voltage is generated at the thermocouple junctions of different metals. Even when all the junctions are at the same temperature, the thermoelectric voltage can reach a value of about 0.1μV/°C. The most significant disturbances are a consequence of offset voltages of the operational amplifiers and can be higher than 50 μV. These are direct current (DC) parasitic voltages. Alternating current (AC) unwanted voltages can also occur. The AC parasitic voltages are a consequence of AC power supply inductive or capacitive influence, noise, etc. AC parasitic voltages cause dispersion of measured results around the mean value. The mains influence (inductive and capacitive) is periodic and can be efficiently decreased (shielding, filtering, etc). Noise is a random occurrence with a zero mean and its disturbance may be reduced to acceptable levels by filtration. DC parasitic voltages cause systematic measurement errors and need to be removed from the measurement voltage as much as possible. One solution for minimization of the influence of these parasitic voltages is presented in this paper.

2. Results and Discussion

There are several known methods for removing DC parasitic voltages. The main idea of DC voltage elimination is presented in Figure 2.
According to the figure, the voltages can be expressed as:
U 1 = R × I + U P
where I is the measuring current and UP parasitic DC voltage. When the measuring current is zero, the voltage should be:
U 2 = U P
The measured voltage U is independent of parasitic voltage UP:
U = U 1 U 2 = R × I
It is possible to use integrated operational amplifiers with chopper stabilized input voltage offset, such as ICL7650 [5]. The chopper stabilization includes the inner gates only, just to their input pins. The input offset voltages are reduced to 1 μV with temperature coefficient (TC) of 0.01 μV/°C, but sometimes there is a need to decrease the offset value below 1 μV. The improvement of the present solutions is, in fact, the main goal of design and realization of our electrical circuit with chopper elimination of DC parasitic voltage.
The principle circuit diagram of the low resistance comparator is shown in Figure 1. Both resistances (measured and reference), RX and RR are serially connected. The current circuit supplies them with a current of about 1 A. There is no need for high and long-term current stability. Since the current supply circuit is galvanically separated, the reference potentials are connected with an appropriate analog switch.
The maximal value of resistances RX and RR for the chosen measuring range is about 10 mΩ and the RI voltages on their potential terminals are about 10 mV. Both voltages (UX and UR) are amplified by the same amplifier (G=1,000) and give maximal outputs of about 10 V (Figure 3). Besides the amplifier, there is a control circuit, circuit for chopper elimination of parasitic voltages (correction circuit) and output sample and hold circuits.
The controller switches the measuring current on and off and controls the functions of the voltage circuit analog switch. It is adjusted so that the duration of current pulse of 1 A is 60% of one controller cycle. During the remaining 40% of a cycle the current is switched off. While the measuring current is switched off, the amplifier's output should be zero, but parasitic voltage at the amplifier input occurs and it is amplified 1,000 times, as well.
Using correction and feed back circuits, this amplified voltage can be reduced to an acceptable value, below 10 μV. This is done for both resistances (RX and RR), sequentially. These correction voltages (annulling voltage, Figure 3.) are memorized on the corresponding capacitors and used while the measuring current is switched on. In a new cycle, when the measuring current is switched on again, both voltages (UX and UR) are amplified consecutively and the earlier memorized correction voltages are activated and included. The voltage without the component of parasitic unwanted voltage occurs for every resistor on the amplifier output. These two voltage outputs are separated by a demultiplexer and led to two independent sample and hold circuits, giving output voltages UOR and UOX, respectively. Their measurement or their difference gives the possibility of calculation of the resistance with very high accuracy using the above expressions.
The described solution for low resistance measurement allows a very useful possibility: to perform the measurement with a known value of resistor RR, then repeat the measurement with the same resistor (RX = RR). Let us call this a self-comparison mode (SCM). In this case, the measured value of both output voltages should be the same. But, in practice, these are slightly different. This difference is exactly equal to the comparator error. For the realized instrument, the average value of this difference for voltages of 10 V is less then 20 μV, or 2 ppm. The remaining AC voltages cause a dispersion of the measured values of up to 50 μV (5 ppm). For output voltages of about 10 V these voltage differences are acceptable.
In order to reach high measurement accuracy the instrument must have extremely high sensitivity. In such cases unwanted influences can occur. Some critical points of design, construction and practical realization are listed below:
  • For a complete elimination of error caused by common mode rejection ratio (CMRR) input voltage, a special way of switching was applied. A third switch was added to connect the reference potential of the voltage circuit with the negative resistor potential terminal.
  • The influence of transient processes was avoided by the use of appropriate length of dead time in controller cycle (pause, Figure 3).
  • The leakage current of output sample and hold circuits should be extremely small because the voltage drop down mustn’t exceed 10 μV. The controller cycle is synchronized in such a way that the cycle step duration is a multiple of the main frequency period to be able to reach these conditions and thus eliminate the capacitive and inductive disturbances.
  • In order to reduce the mains supply influence the controller cycle is synchronized with the mains supply frequency.
  • Excellent quality operational amplifiers with very high open loop gain are used in the design.
  • To achieve very high linearity, the complete amplification is realized with three-stage amplifiers with low gain (10 times each). The described solution allows the possibility of not only the resistance comparison., but with voltage ratio measurement (UX/UR) and high quality reference resistor RR (standard resistor for example) it is possible to measure the resistance RX with very high accuracy (milliohm meter).

3. Experimental Section

A 6½ digits A/D converter is embedded in the realized prototype instrument. The applied resistance of RR = 10 mΩ gives a measurement range of about 10 mΩ □ □ and 10 nΩ resolution. Excellent experimental results, presented in Table 1, were achieved in the SCM mode [6], where:
  • - N - number of independent repeated measuring attempts
  • - UOX, UOR ≈ 10 V (corresponding RX, RR =10 mΩ)
  • - Mean - error mean value
  • - sd - standard deviation
  • - u - standard deviation of the mean (Type A standard uncertainty)
The realized prototype instrument is shown in Figure 4. It has shown excellent results: for the measurement with a current of 1 A for a resistance of about 10 mΩ, the measurement deviation was no greater than 50 nΩ in every single attempt [6].

4. Conclusions

An improvement of the chopper method for elimination of parasitic voltages in a procedure of a small resistance comparison and measurement are presented in this paper. The realized instrument prototype had to overcome many practical difficulties. Some of them were solved empirically (i.e. choosing appropriate operational amplifiers) [7]. The nstrument is working as a prototype at this moment, but we have the intention of integrating it into laboratory equipment. All experiments were performed at the Department of Electrical Measuring of the Faculty of Technical Sciences in Novi Sad and confirmed by the official institution in Belgrade. Very high accuracy in resistance comparison and measurement is achieved (0.08 ppm for 1,000 independent measurement attempts and 0.26 ppm for 100 attempts). Although these are excellent results, there is room for further improvements. The main activities should be focused on design and realization of more different measurement ranges with the possibility of auto calibration. An interface for direct PC connection should be also developed, for the proposed comparator to become an intelligent instrument [8].

References and Notes

  1. Kible, B.P.; Legg, D.J. A generalized Wheatstone bridge for comparing 1-Ω resistors. IEEE Trans. Instrum. Meas 1985, 34, 282–284. [Google Scholar]
  2. Wavetek Corporation. Wawetek, Model 1281/1271 Data Sheet 1997.
  3. Keithley Instruments Inc. 8½-digit Digital Multimeters Data Sheet 2002.
  4. Nuckolls, B. Two-Point and Four-Point Methods for Measuring Small Resistances, Practical Low Resistance Measurements. Aero Electric Connection. Available online: http://www.aeroelectric.com/articles/LowOhmsAdapter_3.pdf, August 2004.
  5. Intersil Corporation. ICL7650 Data Sheet, File Number 2920.5 2000.
  6. Radetic, R.; Zupunski, I. Low resistance comparator test results. Proc. of the Congress of Metrologists of Yugoslavia, Novi Sad, Serbia, 2003.
  7. Radetic, R.; Zupunski, I. Low resistance comparator. Proc. of the Congress of Metrologists of Yugoslavia, Novi Sad, Serbia, 2000.
  8. Barney, G.C. Intelligent Instrumentation; Prentice Hall International: 66 Wood Lane End, Hemel Hempstead, HP2 4RG, United Kingdom, 1988. [Google Scholar]
Figure 1. Circuit diagram of chopper stabilized comparator.
Figure 1. Circuit diagram of chopper stabilized comparator.
Sensors 09 02491f1
Figure 2. Parasitic voltage elimination.
Figure 2. Parasitic voltage elimination.
Sensors 09 02491f2
Figure 3. The controllers timing diagram.
Figure 3. The controllers timing diagram.
Sensors 09 02491f3
Figure 4. Low resistance comparator prototype.
Figure 4. Low resistance comparator prototype.
Sensors 09 02491f4
Table 1. The experimental results.
Table 1. The experimental results.
NMean (μV) (ΔUO=UOX-UOR)ΔR (nΩ)sd (μV)u (ppm)
10011,211,225,60,26
10004,54,524,30,08

Share and Cite

MDPI and ACS Style

Radetic, R.M.; Milivojevic, D.R. Chopper Stabilized Low Resistance Comparator. Sensors 2009, 9, 2491-2497. https://doi.org/10.3390/s90402491

AMA Style

Radetic RM, Milivojevic DR. Chopper Stabilized Low Resistance Comparator. Sensors. 2009; 9(4):2491-2497. https://doi.org/10.3390/s90402491

Chicago/Turabian Style

Radetic, Radojle M., and Dragan R. Milivojevic. 2009. "Chopper Stabilized Low Resistance Comparator" Sensors 9, no. 4: 2491-2497. https://doi.org/10.3390/s90402491

Article Metrics

Back to TopTop