Rapid Paper-Based System for Human Serum Creatinine Detection

1
Graduate Institute of Materials Engineering, National Pingtung University of Science and Technology, Pingtung 912, Taiwan
2
Department of Biomechatronics Engineering, National Pingtung University of Science and Technology, Pingtung 912, Taiwan
3
Department of Engineering Science, National Cheng Kung University, Tainan 701, Taiwan
4
Department of Internal Medicine, College of Medicine, National Cheng Kung University and Hospital, Tainan 704, Taiwan
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Inventions 2018, 3(2), 34; https://doi.org/10.3390/inventions3020034
Submission received: 9 April 2018 / Revised: 21 May 2018 / Accepted: 21 May 2018 / Published: 28 May 2018
(This article belongs to the Special Issue Microfluidics and Nanofluidics)

Abstract

:
An integrated system consisting of a paper-based chip and a smart detection device is proposed for determining the human serum creatinine concentration based on Jaffé reaction theory. In the proposed approach, the reaction zone of the paper-based chip is implanted with picric acid and NaOH reagent and dried at 35 °C for 20 min. Human serum creatinine is dripped onto the reaction zone of the chip. A Jaffé reaction is induced by heating the chip at 37 °C for 5 min and the creatinine concentration is then derived by analyzing the RGB (red, green and blue) intensity of the resulting Janovsky complex using self-written analysis software installed on a smartphone. The validity of the proposed method is demonstrated using control samples with creatinine concentrations ranging from 0.2~8 mg/dL. The detection results obtained for 32 real-world creatinine samples are shown to be in excellent agreement with those obtained using a standard macroscale method (R2 = 0.9994). Overall, the results show that the proposed system provides a compact, low-cost and reliable approach for human serum creatinine concentration detection.

Graphical Abstract

1. Introduction

Creatinine is an end-product of creatine metabolism and is one of the most useful analytes for the assessment of renal function. In normal healthy individuals, the typical reference ranges for serum creatinine are 0.5 to 1.0 mg/dL for women and 0.7 to 1.2 mg/dL for men [1,2]. In healthy humans, creatinine is continuously excreted by the kidneys at a rate of around 1.6~1.7% per day [3]. Thus, increasing levels of creatinine in serum (or decreasing levels in urine) provide a possible indication of kidney failure. The creatinine concentration is also employed as a correction factor for fluctuations in the urine volume and is thus useful in accurately determining the micro-albumin/creatinine ratio. Consequently, sensitive and accurate assays for measuring the creatinine level in blood and urine samples are of significant interest in the clinical diagnostics field.
Numerous approaches for creatinine detection have been reported in the literature, including Jaffé kinetic assays [4,5,6], enzyme reactions [7], capillary electrophoresis [8], chemiluminescence [9], chromatography [10], molecular imprinted polymer (MIP) assays [11], spectrophotometry [12,13], liquid chromatography-tandem mass spectrometry (LC-MS/MS) [14], potentiometric sensors [15,16], electrochemical sensors [17,18], colorimetry [19,20], pH sensors [21,22], amperometric sensors [23,24] and nanoparticle-enhanced sensors [25,26,27]. Although the results of these common methods provide accurate and reliable results, while almost all these instruments are bulky, expansive and require complicated operations.
Recent years have seen a growing interest in micro-total analysis systems (μTAS) and lab-on-a-chip (LoC) devices. Microfluidic systems and microfluidic paper-based devices (µPADs) offer many advantages over traditional devices, including a reduced reagent consumption, a faster analysis time, a lower cost, an improved portability and better disposability. As a result, they are regarded as extremely promising candidates for point of care testing (POCT) and disease screening applications in the developing world [28,29,30,31,32]. Various microfluidic systems for the monitoring and diagnosis of disorders such as cardiovascular disease, diabetes, cancer and renal disease have been proposed in the literature [32,33,34,35,36]. Furthermore, several integrated microfluidic devices (or LoC devices) [37,38,39,40,41] and µPADs [42,43,44,45] have been presented for determining the creatinine concentration. For example, Dosso et al. [41] developed an integrated SIMPLE-based biosensor (named Creasensor) for the detection of creatinine in blood plasma samples. The platform combined fast bioassay on a self-powered SIMPLE microfluidic cartridge with an integrated signal processing unit and a simple but robust colorimetric read-out. The platform was validated on field in a single-blind study by measuring the creatinine levels in 16 creatinine-spiked plasma samples. An excellent agreement was observed between the measured concentration values and the spiked concentrations, over the creatinine range of 0.76~20 mg/dL. Sununta et al. [45] presented a μPAD for determining the creatinine concentration in urine samples by means of a Jaffé reaction process followed by colorimetric detection. The device was shown to achieve a wide linear range of 0.2–1 mM with a limit of detection of 0.08 mM.
The present study proposes an integrated system for rapid human serum creatinine concentration measurement consisting of a paper-based chip and a detection system comprising a heating module, a voltage regulator module, a cooling module, a CMOS camera, a USB connector and a smartphone. In the proposed detection process, a Jaffé reaction is induced between the creatinine and alkaline picrate reagent implanted in the detection zone of the chip by heating the chip at 37 °C for 5 min in the detection box. The color intensity of the resulting yellow-orange reaction complex is captured by the CMOS camera and transferred to the smartphone using the USB connector. Finally, the concentration of the creatinine sample is evaluated by analyzing the RGB (red, green and blue) color intensity of the reaction image using self-written software installed on the smartphone in the form of an app. Overall, the proposed system provides celerity, convenience, low-cost, miniaturization and automation for human serum creatinine concentration measurement.

2. Materials and Methods

2.1. Reagent Preparation

Creatinine powder and reagent (R1: sodium hydroxide; R2: picric acid) were purchased from Roche calibrator (Roche Diagnostics, Taipei, Taiwan) and Roche/Hitachi Creatinine plus (Roche Diagnostics, Germany). Creatinine detection was performed using a reagent solution consisting of 0.1 M sodium hydroxide (R1, pH > 13.5) and 50 mM picric acid (R2, pH = 6.5). Phosphate buffered saline (PBS) solution was prepared comprising 0.2 M sodium dihydrogen phosphate (NaH2PO4), de-ionized (DI) water and 0.15 M sodium chloride (NaCl). Finally, a standard 1000 mg/dL creatinine (C4H7N3O) stock solution was prepared by mixing 0.1 g creatinine powder with 10 ml 0.1 M HCl. Control samples with creatinine concentrations in the range of 0.2~8 mg/dL were then prepared by diluting the stock solution with appropriate quantities of PBS solution.

2.2. Fabrication of Paper-Based Chip

Figure 1a–d show the main steps, including (a) pattern design, (b) wax printing, (c) heating wax process and (d) reagent and sample dropped, in the fabrication process used to prepare the paper-based chip. The chip was fabricated on Advantech qualitative filter paper (No. 1, Toyo Roshi Kaisha Ltd., Tokyo, Japan; pore size: 6 µm; thickness: 0.2 mm) and consisted of a circular reaction zone with a diameter of 3 mm bounded by an impermeable wax ring with an outer diameter of 5 mm. The chip was designed using CorelDrawR Graphics Suite X5 software and printed using a commercial wax printer (Fuji Xerox ColorQube 8750, Japan). After printing, the chip was heated for 3 min at 120 °C in a benchtop furnace (Vulcan A-550, Taiwan) to melt the wax and form a hydrophobic barrier through the thickness of the paper. Finally, the paper-based chip was bonded with a cardboard substrate using 3 M double-side tape (V1205, 3 M Comp., Taipei, Taiwan).

2.3. Detection System and Jaffé Chemical Reaction Process

As shown in Figure 2, the main components of the detection system include a power source, a temperature controller, a voltage controller, a cooling module, a detection box and a smartphone (Xperia Z5, Sony Co., Tokyo, Japan) The detection box further contains a CMOS camera (1920 × 1080 pixels, Dmatek Co. Ltd., Taichung, Taiwan), two LED light sources, a microheater, a chip holder and a connector. A more detailed and the function of the module in the detection system can be referred to our previous study [46].
In performing the creatinine detection process, 3 µL of sample solution was dropped onto the reaction zone of the chip (reagent implanted with R1:R2 = 1 µL:2 µL and dried at a temperature of 35 °C for 10 min). The chip was then inserted into the chip holder of the detection box and heated at 37 °C for 5 min to induce a Jaffé reaction (i.e., Creatinine + Picric acid a l k a l i n e yellow-orange complex (see Scheme 1)). Images of the creatinine-picric complex were acquired by a built-in CMOS camera and transferred to a smartphone by means of the USB connector. Note that the smartphone was only applied to calculate the RGB values without catching images. Finally, the human serum creatinine concentration was derived using a self-written RGB color analysis app installed on the phone.

3. Results and Discussion

3.1. Effects of Reaction Time and Reagent Concentration

A series of preliminary experiments were performed to investigate the effects of the Jaffé reaction time on the G (green) + B (blue) intensity of the reacted yellow-orange complex. Figure 3a presents the corresponding results obtained for creatinine control samples with concentrations of 0.2, 4.0 and 8.0 mg/dL given reaction times of 1~8 min and a constant reaction temperature of 37 °C. For reaction times of 1~3 min, the G (green) + B (blue) intensity steadily increases with an increasing reaction time. In other words, the reaction process between the reagent and the creatinine sample is incomplete. However, for reaction times of more than 5 min, the G (green) + B (blue) intensity remains stable for all three samples. In other words, the reaction process is complete. Thus, in performing all the remaining experiments, the reaction time was set as 5 min.
For RGB color analysis, a wide color distribution range is essential in improving the resolution and accuracy of the detection results. Figure 3b shows the G (green) + B (blue) intensity values of the reaction complexes obtained for 0.20 and 8 mg/dL creatinine control samples given NaOH reagent concentrations of 0.05~0.25 M, a reaction time of 5 min and a reaction temperature of 37 °C. For NaOH concentrations of less than 0.1 M, the G + B values of the two samples differ by only ~30 (G + B). However, as the NaOH concentration increases, the difference in the R+B values of the samples also increases. For example, given an NaOH concentration of 0.15 M, the interval between the G+B values of the two samples is equal to approximately 55 (G + B). In other words, the interval between the two values is more than 1.8 times higher than that obtained using 0.1 M NaOH. Consequently, the NaOH concentration was set as 0.15 M in all of the remaining detection tests.

3.2. Linearity of RGB Intensity Values and Calibration of Detection System

To determine the optimal RGB signal for detection purposes, the intensities of the R (red), G (green) and B (blue) signals were observed for a reaction temperature of 37 °C, a reaction time of 5 min and creatinine samples with concentrations in the range of 0.2~8 mg/dL. Figure 4a shows the corresponding results. It is seen that the R (red) intensity signal has a correlation coefficient of just R2 = 0.8603 and is thus unsuitable for detection purposes. By contrast, the G (green) and B (blue) intensity signals both have correlation coefficients of more than 0.99. In other words, both signals are highly correlated with the creatinine concentration and thus provide a reliable basis for creatinine detection.
To calibrate the proposed system, the G (green) + B (blue) intensity values were measured for five standard creatinine samples with concentrations ranging from 0.2~8 mg/dL. For each sample, five separate measurements of the G (green) + B (blue) intensity were obtained. The measurements were then averaged in order to obtain a representative value for the sample. In accordance with the results described above, the reaction process was performed using an NaOH concentration of 0.15 M, a reaction temperature of 37 °C and a reaction time of 5 min. Figure 4b shows the intensity values obtained for the various control samples. Applying a regression analysis technique to the experimental results, the G (green) + B (blue) intensity value (Y) is found to be related to the creatinine concentration (X) as Y = −9.4150X + 344.837. Moreover, the correlation coefficient is equal to R2 = 0.9934. In other words, the G (green) + B (blue) intensity varies linearly with the creatinine concentration over the normal creatinine range in human blood. Thus, the validity of the proposed system is confirmed.

3.3. Creatinine Detection in Real Samples

The applicability of the proposed platform for clinical diagnostic purposes was investigated using human serum creatinine samples obtained from 32 adult volunteers (patients at National Cheng Kung University Hospital in Taiwan). These human serum creatinine samples are obtained from whole blood samples via a separation and purification process at the hospital. The detection results were compared with those obtained using a macroscale human serum creatinine assay technique (a spectrophotometry method performed using an automatic biochemistry analyzer/integrated system 7600, Hitachi High-Technologies Co., Tokyo, Japan) at a major university hospital in Taiwan (National Cheng Kung University Hospital, Tainan, Taiwan). Figure 5a presents a scatter-plot of the results obtained for the 32 samples by the two methods. As shown in Figure 5b, the two sets of results are related as Y = 0.9907X + 0.0001, with a correlation coefficient of 0.9994. In other words, the ability of the proposed paper-based human serum creatinine system to detect the creatinine concentration in real-world samples is confirmed.

4. Conclusions

This study has presented a simple method based on Jaffé reaction theory for performing creatinine concentration detection using a paper-based chip and a smart detection system comprising a detection box, a temperature controller, a voltage regulator, a power source, a USB connector and a smartphone. The detection box further comprises a CMOS camera, two LED light sources, a chip holder and a hotplate. In the proposed system, the reaction region of the paper chip is implanted with 50 mM picric acid reagent and 0.15 M NaOH and the chip is then dried at 35 °C for 20 min. In the detection process, the serum creatinine sample is dripped onto the reaction zone of the chip and the chip is then transferred to the hotplate in the detection system, where a Jaffé reaction is induced by heating at 37 °C for 5 min. The experimental results have shown that the creatinine concentration measurements obtained by the proposed platform for 32 adult volunteers are in excellent agreement with those obtained using a conventional macroscale technique (R2 = 0.9994). In other words, the feasibility of the proposed paper-based system for practical clinical purposes is confirmed. Overall, the proposed method provides a simple, portable and low-cost approach for human serum creatinine concentration detection. Thus, it represents an ideal solution for a wide range of diagnostic and POCT applications.

Author Contributions

L.-M.F., C.-C.T. and W.-J.J. conceived and designed the experiments; W.-J.J. performed the experiments; L.-M.F. and R.-J.Y. analyzed the data; W.-J.J. contributed reagents/materials/analysis tools; L.-M.F. wrote the paper.

Acknowledgments

The authors would like to thank the Ministry of Science and Technology of Taiwan for the financial support of this study under Grant Nos. MOST-104-2221-E-006-154-MY3, MOST 106-2314-B-020-002-MY3, MOST 106-2221-E-020-019-MY3, MOST 106-2622-B-020-001-CC2, MOST 107-2622-B-020-003-CC2 and 107TFDA-A-106.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Pundir, C.S.; Yadav, S.; Kumar, A. Creatinine sensors. Trends Anal. Chem. 2013, 50, 42–52. [Google Scholar] [CrossRef]
  2. Songjaroen, T.; Maturos, T.; Sappat, A.; Tuantranont, A.; Laiwattanapaisal, W. Portable microfluidic system for determination of urinary creatinine. Anal. Chim. Acta 2009, 647, 78–83. [Google Scholar] [CrossRef] [PubMed]
  3. Randviir, E.P.; Banks, E. Analytical methods for quantifying creatinine within biological media. Sens. Actuators B Chem. 2013, 183, 239–252. [Google Scholar] [CrossRef]
  4. Wang, A.B.; Fang, P.H.; Su, Y.C.; Hsieh, Y.W.; Lin, C.W.; Chen, Y.T.; Hsu, Y.C. A novel lab-on-a-chip design by sequential capillary–gravitational valves for urinary creatinine detection. Sens. Actuators B Chem. 2016, 222, 721–727. [Google Scholar] [CrossRef]
  5. Chen, S.; Song, Y.; Shi, F.; Liu, Y.; Ma, Q. Sensitive detection of picric acid based on creatinine-capped solid film assembled by nitrogen-doped graphene quantum dots and chitosan. Sens. Actuators B Chem. 2016, 231, 634–640. [Google Scholar] [CrossRef]
  6. Blel, A.; Orven, Y.; Pallet, N.; Chasse, J.F.; Vedie, B.; Loriot, M.A.; Paul, J.L.; Narjoz, C. Pegylated liposomal doxorubicin (Caelyx®) interference with the spectrophotometric Jaffemethod for quantitative determination of creatinine in human plasma. Clin. Biochem. 2017, 50, 455–457. [Google Scholar] [CrossRef] [PubMed]
  7. Duong, H.D.; Rhee, J.I. Development of ratiometric fluorescent biosensors for the determination of creatine and creatinine in urine. Sensors 2017, 17, 2570. [Google Scholar] [CrossRef] [PubMed]
  8. Vitali, L.; Gonçalves, S.; Rodrigues, V.; Fávere, V.T.; Micke, G.A. Development of a fast method for simultaneous determination of hippuric acid, mandelic acid and creatinine in urine by capillary zone electrophoresis using polymer multilayer-coated capillary. Anal. Bioanal. Chem. 2017, 409, 1943–1950. [Google Scholar] [CrossRef] [PubMed]
  9. Hanif, S.; John, P.; Gao, W.; Saqib, M.; Qi, L.; Xu, G. Chemiluminescence of creatinine/H2O2/Co2+ and its application for selective creatinine detection. Biosens. Bioelectron. 2016, 75, 347–351. [Google Scholar] [CrossRef] [PubMed]
  10. Langsi, V.K.; Ashu-Arrah, B.A.; Ward, N.; Glennon, J.D. Synthesis and characterisation of non-bonded 1.7 μm thin-shell (TS1.7-100 nm) silica particles for the rapid separation and analysis of uric acid and creatinine in human urine by hydrophilic interaction chromatography. J. Chromatogr. A 2017, 1506, 37–44. [Google Scholar] [CrossRef] [PubMed]
  11. Diouf, A.; Motia, S.; El Alami, N.; El Bari, N.; Bouchikhi, B. Development and characterization of an electrochemical biosensor for creatinine detection in human urine based on functional molecularly imprinted polymer. J. Electroanal. Chem. 2017, 788, 44–53. [Google Scholar] [CrossRef]
  12. Saidi, T.; Moufid, M.; Zaim, O.; Bari, N.E.; Bouchikhi, B. Voltammetric electronic tongue combined with chemometric techniques for direct identification of creatinine level in human urine. Measurement 2018, 115, 178–184. [Google Scholar] [CrossRef]
  13. Krishnegowda, A.; Padmarajaiah, N.; Anantharaman, S.; Honnur, K. Spectrophotometric assay of creatinine in human serum sample. Arab. J. Chem. 2017, 10, S2018–S2024. [Google Scholar] [CrossRef]
  14. Suzuki, M.; Furuhashi, M.; Sesoko, S.; Kosuge, K.; Maeda, T.; Todoroki, K.; Inoue, K.; Min, J.Z.; Toyo’oka, T. Determination of creatinine-related molecules in saliva by reversed-phase liquid chromatography with tandem mass spectrometry and the evaluation of hemodialysis in chronic kidney disease patients. Anal. Chim. Acta 2016, 911, 92–99. [Google Scholar] [CrossRef] [PubMed]
  15. Guinovart, T.; Hernández-Alonso, D.; Adriaenssens, L.; Blondeau, P.; Martínez-Belmonte, M.; Rius, F.X.; Andrade, F.J.; Ballester, P. Recognition and sensing of creatinine. Angew. Chem. Int. Ed. 2016, 55, 1–7. [Google Scholar] [CrossRef] [PubMed]
  16. Guinovart, T.; Hernández-Alonso, D.; Adriaenssens, L.; Blondeau, P.; Rius, F.X.; Ballester, P.; Andrade, F.J. Characterization of a new ionophore-based ion-selective electrode for the potentiometric determination of creatinine in urine. Biosens. Bioelectron. 2017, 87, 587–592. [Google Scholar] [CrossRef] [PubMed]
  17. Raveendran, J.; Resmi, P.E.; Ramachandran, T.; Nair, B.G.; Babu, T.G.S. Fabrication of a disposable non-enzymatic electrochemical creatinine sensor. Sens. Actuators B Chem. 2017, 243, 589–595. [Google Scholar] [CrossRef]
  18. Hooshmand, S.; Es’haghi, Z. Microfabricated disposable nanosensor based on CdSe quantum dot/ionic liquid-mediated hollow fiber-pencil graphite electrode for simultaneous electrochemical quantification of uric acid and creatinine in human samples. Anal. Chim. Acta 2017, 972, 28–37. [Google Scholar] [CrossRef] [PubMed]
  19. Sutariya, P.G.; Pandya, A.; Lodha, A.; Menon, S.K. A simple and rapid creatinine sensing via DLS selectivity, using calix[4] arene thiol functionalized gold nanoparticles. Talanta 2016, 147, 590–597. [Google Scholar] [CrossRef] [PubMed]
  20. Alula, M.T.; Karamchand, L.; Hendricks, N.R.; Blackburn, J.M. Citrate-capped silver nanoparticles as a probe for sensitive and selective colorimetric and spectrophotometric sensing of creatinine in human urine. Anal. Chim. Acta 2018, 1007, 40–49. [Google Scholar] [CrossRef] [PubMed]
  21. Pal, S.; Lohar, S.; Mukherjee, M.; Chattopadhyay, P.; Dhara, K. A fluorescent probe for the selective detection of creatinine in aqueous buffer applicable to human blood serum. Chem. Commun. 2016, 52, 13706–13709. [Google Scholar] [CrossRef] [PubMed]
  22. Roy, S.; Roy, A.; Panja, R.; Samanta, S.; Chakrabarti, S.; Yu, P.L.; Maikap, S.; Cheng, H.M.; Tsai, L.N.; Qiu, J.T. Comparison of resistive switching characteristics by using e-gun/sputter deposited SiOx film in W/SiOx/TiN structure and pH/creatinine sensing through iridium electrode. J. Alloy Compd. 2017, 726, 30–40. [Google Scholar] [CrossRef]
  23. Nieh, C.H.; Tsujimura, S.; Shirai, O.; Kano, K. Amperometric biosensor based on reductive H2O2 detection using pentacyanoferrate-bound polymer for creatinine determination. Anal. Chim. Acta 2013, 767, 128–133. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  24. Kumar, P.; Jaiwal, R.; Pundir, C.S. An improved amperometric creatinine biosensor based on nanoparticles of creatininase, creatinase and sarcosine oxidase. Anal. Biochem. 2017, 537, 41–49. [Google Scholar] [CrossRef] [PubMed]
  25. Du, H.; Chen, R.Y.; Du, J.J.; Fan, J.L.; Peng, X.J. Gold nanoparticle-based colorimetric recognition of creatinine with good selectivity and sensitivity. Ind. Eng. Chem. Res. 2016, 55, 12334–12340. [Google Scholar] [CrossRef]
  26. Parmar, A.K.; Valand, N.N.; Solanki, K.B.; Menon, S.K. Picric acid capped silver nanoparticles as a probe for colorimetric sensing of creatinine in human blood and cerebrospinal fluid samples. Analyst 2016, 141, 1488–1498. [Google Scholar] [CrossRef] [PubMed]
  27. Kasap, B.O.; Marchenko, S.V.; Soldatkin, O.O.; Dzyadevych, S.V.; Kurc, B.A. Biosensors based on nano-gold/zeolite-modified Ion selective field-effect transistors for creatinine detection. Nanoscale Res. Lett. 2017, 12, 162–172. [Google Scholar] [CrossRef] [PubMed]
  28. Lee, C.Y.; Fu, L.M. Recent advances and applications of micromixers. Sens. Actuators B Chem. 2018, 259, 677–702. [Google Scholar] [CrossRef]
  29. Yeh, E.; Fu, C.; Hu, L.; Thakur, R.; Feng, J.; Lee, L.P. Self-powered integrated microfluidic point-of-care low-cost enabling (SIMPLE) chip. Sci. Adv. 2017, 3, e1501645. [Google Scholar] [CrossRef] [PubMed]
  30. Swain, B.; Hong, M.H.; Kang, L.; Kim, B.S.; Kim, N.H.; Lee, C.G. Optimization of CdSe nanocrystals synthesis with a microfluidic reactor and development of combinatorial synthesis process for industrial production. Chem. Eng. J. 2017, 308, 311–321. [Google Scholar] [CrossRef]
  31. Kong, X.; Chong, X.; Squire, K.; Wang, A.X. Microfluidic diatomite analytical devices for illicit drug sensing with ppb-Level sensitivity. Sens. Actuators B Chem. 2018, 259, 587–595. [Google Scholar] [CrossRef] [PubMed]
  32. Kaur, G.; Tomar, M.; Gupta, V. Development of a microfluidic electrochemical biosensor: Prospect for point-of-care cholesterol monitoring. Sens. Actuators B Chem. 2018, 261, 460–466. [Google Scholar] [CrossRef]
  33. Frank, P.; Haefner, S.; Elstner, M.; Richter, A. Fully-programmable, low-cost, “do-it-yourself” pressure source for general purpose use in the microfluidic laboratory. Inventions 2016, 1, 13. [Google Scholar] [CrossRef]
  34. Ynag, R.J.; Fu, L.M.; Hou, H.H. Review and perspectives on microfluidic flow cytometers. Sens. Actuators B Chem. 2018, 266, 26–45. [Google Scholar] [CrossRef]
  35. Han, T.; Zhang, L.; Xu, H.; Xuan, J. Factory-on-chip: Modularised microfluidic reactors for continuous mass production of functional materials. Chem. Eng. J. 2017, 326, 765–773. [Google Scholar] [CrossRef]
  36. Chung, J.; Hwang, H.Y.; Chen, Y.; Lee, T.Y. Microfluidic packaging of high-density CMOS electrode array for lab-on-a-chip applications. Sens. Actuators B Chem. 2018, 254, 542–550. [Google Scholar] [CrossRef]
  37. Huang, C.J.; Lin, J.L.; Chen, P.H.; Syu, M.J.; Lee, G.B. A multi-functional electrochemical sensing system using microfluidic technology for the detection of urea and creatinine. Electrophoresis 2011, 32, 931–938. [Google Scholar] [CrossRef] [PubMed]
  38. Wang, S.; Li, X.; Yang, J.; Yang, X.; Hou, F.; Chen, Z. Rapid determination of creatinine in human urine by microchip electrophoresis with LED induced fluorescence detection. Chromatographia 2012, 75, 1287–1293. [Google Scholar] [CrossRef]
  39. Lin, C.C.; Hsu, J.L.; Tseng, C.C.; Lee, G.B. An integrated microfluidic system for the determination of microalbuminuria by measuring the albumin-to-creatinine ratio. Microfluid. Nanofluid. 2011, 10, 1055–1067. [Google Scholar] [CrossRef]
  40. Lin, Y.H.; Wang, S.H.; Wu, M.H.; Pan, T.M.; Lai, C.S.; Luo, J.D.; Chiou, C.C. Integrating solid-state sensor and microfluidic devices for glucose, urea and creatinine detection based on enzyme-carrying alginate microbeads. Biosens. Bioelectron. 2013, 43, 328–335. [Google Scholar] [CrossRef] [PubMed]
  41. Dosso, F.D.; Decrop, D.; Pérez-Ruiz, E.; Daems, D.; Agten, H.; Al-Ghezi, O.; Bollen, O.; Breukers, J.; Rop, F.D.; Katsafadou, M.; et al. Creasensor: SIMPLE technology for creatinine detection in plasma. Anal. Chim. Acta 2018, 1000, 191–198. [Google Scholar] [CrossRef] [PubMed]
  42. Talalak, K.; Noiphung, J.; Songjaroen, T.; Chailapakul, O.; Laiwattanapaisal, W. A facile low-cost enzymatic paper-based assay for the determination of urine creatinine. Talanta 2015, 144, 915–921. [Google Scholar] [CrossRef] [PubMed]
  43. Sittiwong, J.; Unob, F. Paper-based platform for urinary creatinine detection. Anal. Sci. 2016, 32, 639–643. [Google Scholar] [CrossRef] [PubMed]
  44. Tambaru, D.; Rupilu, R.H.; Nitti, F.; Gauru, I.; Suwari. Development of paper-based sensor coupled with smartphone detector for simple creatinine determination. AIP Conf. Proc. 2017, 1823, 020095. [Google Scholar]
  45. Sununta, S.; Rattanarat, P.; Challapakul, O.; Praphairaksit, N. Microfluidic paper-based analytical devices for determination of creatinine in urine samples. Anal. Sci. 2018, 34, 109–113. [Google Scholar] [CrossRef] [PubMed]
  46. Liu, C.C.; Wang, Y.N.; Fu, L.M.; Chen, K.L. Microfluidic paper-based chip platform for benzoic acid detection in food. Food Chem. 2018, 249, 162–167. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Schematic illustration showing main steps in fabrication procedure for paper-based chip.
Figure 1. Schematic illustration showing main steps in fabrication procedure for paper-based chip.
Inventions 03 00034 g001
Figure 2. Photograph and schematic illustrations of: (a) main components of the smart detection device; (b,c) show detailed components of the detection box.
Figure 2. Photograph and schematic illustrations of: (a) main components of the smart detection device; (b,c) show detailed components of the detection box.
Inventions 03 00034 g002
Scheme 1. Jaffé reaction between creatinine sample and picric acid/NaOH indicator.
Scheme 1. Jaffé reaction between creatinine sample and picric acid/NaOH indicator.
Inventions 03 00034 sch001
Figure 3. (a) Reaction time stability of samples with various creatinine concentrations given reaction temperature of 37 °C; (b) G + B values of creatinine samples with concentrations of 0.2 mg/dL and 8 mg/dL given different NaOH concentrations and constant reaction time of 5 min.
Figure 3. (a) Reaction time stability of samples with various creatinine concentrations given reaction temperature of 37 °C; (b) G + B values of creatinine samples with concentrations of 0.2 mg/dL and 8 mg/dL given different NaOH concentrations and constant reaction time of 5 min.
Inventions 03 00034 g003
Figure 4. (a) RGB intensity values of creatinine samples and (b) variation of G (green) + B (blue) intensity (and associated color images) with creatinine concentration in the range of 0.2~8 mg/dL. (Note that the reaction temperature and reaction time are 37 °C and 5 min, respectively.).
Figure 4. (a) RGB intensity values of creatinine samples and (b) variation of G (green) + B (blue) intensity (and associated color images) with creatinine concentration in the range of 0.2~8 mg/dL. (Note that the reaction temperature and reaction time are 37 °C and 5 min, respectively.).
Inventions 03 00034 g004
Figure 5. Comparison of human serum creatinine detection results obtained using proposed paper-based system and traditional macroscale system (a) a scatter-plot and (b) a related-plot by the two methods.
Figure 5. Comparison of human serum creatinine detection results obtained using proposed paper-based system and traditional macroscale system (a) a scatter-plot and (b) a related-plot by the two methods.
Inventions 03 00034 g005

Share and Cite

MDPI and ACS Style

Fu, L.-M.; Tseng, C.-C.; Ju, W.-J.; Yang, R.-J. Rapid Paper-Based System for Human Serum Creatinine Detection. Inventions 2018, 3, 34. https://doi.org/10.3390/inventions3020034