Communication

Validation of a Loop-Mediated Isothermal Amplification-Based Kit for the Detection of Legionella pneumophila in Environmental Samples According to ISO/TS 12869:2012

1
U.O.C. of Microbiology and Virology, ARNAS “Civico Di Cristina and Benfratelli”, 90127 Palermo, Italy
2
Legionella Reference Laboratory, Department of Medical, Surgical Sciences and Advanced Technologies “G.F. Ingrassia”, University of Catania, Via Santa Sofia 87, 95123 Catania, Italy
3
Hygiene Complex Operative Unit, A.O.U. Policlinico—Vittorio Emanuele, Via S. Sofia 87, 95123 Catania, Italy
4
Legionella Reference Laboratory, University of Messina, 98125 Messina, Italy
5
Department of Biomedical and Dental Sciences and Morphofunctional Imaging, University of Messina, 98125 Messina, Italy
6
Legionella Reference Laboratory, University of Palermo, 90127 Palermo, Italy
7
Department of Health Promotion, Mother and Child Care, Internal Medicine and Medical Specialties, University of Palermo, 90127 Palermo, Italy
8
Enbiotech s.r.l., Via Aquileia 34, 90144 Palermo, Italy
9
Sicilian Health Department, Public Health and Environmental Risks Service, 90127 Palermo, Italy
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
Microorganisms 2024, 12(5), 961; https://doi.org/10.3390/microorganisms12050961
Submission received: 29 January 2024 / Revised: 20 April 2024 / Accepted: 22 April 2024 / Published: 10 May 2024

Abstract

:
Legionella pneumophila is a freshwater opportunistic pathogen and the leading cause of severe pneumonia known as Legionnaires’ disease. It can be found in all water systems and survives in biofilms, free-living amoebae, and a wide variety of facilities, such as air conditioning and showers in hospitals, hotels and spas. The reference cultural method allows for the isolation and identification in many days, and in addition, it does not detect viable but rather non-culturable bacteria, increasing the risk of infection. In this context, a new LAMP-based (loop-mediated isothermal amplification) kit was developed, allowing for the rapid, sensitive, and labor-saving detection of L. pneumophila. The kit, “Legionella pneumophila Glow”, was validated according to ISO/TS 12869:2012, testing sensitivity, inclusivity and exclusivity, and kit robustness. Sensitivity showed that the “Legionella pneumophila Glow” kit can detect up to 28 plasmid copies/µL. Robustness tests showed consistent results, with both contamination levels and the matrices used giving reproducible results. Furthermore, real samples were evaluated to compare the performance of the two methods. The LAMP kit “Legionella pneumophila Glow” proved a useful option for the rapid, efficient, and labor-saving screening of different typologies of water samples, offering significant advantages over the traditional method, as it is characterized by a high sensitivity, ease of use for laboratory testing, and a large reduction in analysis time, making it an asset to official controls.

1. Introduction

L. pneumophila is a waterborne opportunistic pathogen, causing severe pneumonia called Legionnaires’ disease. Among all the species, L. pneumophila serogroup 1 is the causative agent of at least 70% of all legionellosis cases in the United States and Europe, making it the most clinically relevant [1]. It is found ubiquitously in natural freshwater environments, and in man-made water systems where it is able to survive within biofilms and intracellularly, in free-living amoebae. Indeed, the role of biofilms is being increasingly recognized as relevant for the establishment and maintenance of a chronic colonization within water distribution and plumbing systems, eluding disinfection treatments [2]. It has also been observed that high percentages of Legionella populations in water systems resist in a viable but non-culturable state (VNBC), despite still being infectious [3]. Furthermore, surveillance of Legionnaires’ disease has shown an increasing trend in illness incidence in Europe from 2011 to 2015 [4], which may have been potentially influenced by changes in temperature and rainfall rates due to climate change [5].
Up to 70% of outbreaks are observed in community establishments [6], at healthcare facilities and prisons, even though L. pneumophila infections have been generally underestimated because of a lack of clinical awareness and variances in diagnostic procedures [7,8]. The available cultural method requires more than a week to determine both positive and negative results and is laborious and time-consuming. In addition, it does not detect viable but non-culturable bacteria, consequently increasing the risk of false negative results. Molecular methods are more rapid and have an interesting potential for screening, revealing a preliminary result, even if the cultural method remains essential to isolate the strain for further characterization. Since 2015, molecular methods such as real-time PCR were already introduced, and an official reference states that the “Italian guidelines for the prevention and control of legionellosis” aim to obtain the rapid identification of negative samples. Today, PCR is poorly used, both because it detects live and dead bacteria, and because, based on the previous ISO [9], it requires the collection of two liters of water, one liter for molecular testing and the other for cultivation in case of positive molecular results. Recently, a quicker and easier molecular method has been proposed: loop-mediated isothermal amplification (LAMP). In this study, a L. pneumophila LAMP assay was validated. LAMP is an innovative specific and cost-effective nucleic acid amplification method for bacterial detection and identification. This novel method is made up of six primers which specifically recognize eight different regions on the target gene [10].
LAMP shows many advantages compared to PCR and real-time PCR as it does not require complex and expensive instruments for carrying out the analysis. In addition, the system validated in this study allows for the automatic interpretation of the results and it is characterized by ready-to-use reagents that simplify the workflow of the analysis which can be carried out by semi-skilled staff [11]. Moreover, LAMP detection sensitivity is 10–100-fold more sensitive and displays a higher amplification efficiency, with higher amounts of amplification products than PCR [12,13,14]. The enzyme involved in the LAMP amplification process, Bst DNA polymerase, allows for strand displacement and amplification under isothermal conditions, and it is more tolerant to PCR-interfering substances such as hemin and anticoagulants [14]. For these reasons, it may represent an ideal candidate for point-of-care diagnosis and when rapid results are needed, such as in the case of L. pneumophila facility contamination [15].
This study aimed to carry out a validation of a new commercial system, using the LAMP method, for the detection and quantification of L. pneumophila in water samples, according to the ISO [9].

2. Materials and Methods

2.1. Study Design

The kit “Legionella pneumophila Glow” (Enbiotech srl, Palermo, Italy) has been under validation in compliance with ISO [9] “Water quality—Detection and quantification of Legionella spp. and/or Legionella pneumophila by concentration and genic amplification by quantitative polymerase chain reaction (qPCR)”. The validation was carried out between March and December 2018 at three Regional Reference Laboratories for the control of legionellosis in Sicily, in Palermo, Catania, and Messina.
The kit “Legionella pneumophila Glow” (Enbiotech Cat. No. EBT621) includes rapid preliminary DNA extraction from membrane filters (PES, 0.45 μm, 47 mm, PALL) after 1 L water filtration, genetic amplification using LAMP technology, and the detection of the results using the dedicated device ICGENE mini (Enbiotech Cat. No. EBT801). The kit is composed of a DNA extraction buffer, acting through chemical lysis and all ready-to-use amplification reagents, such as the master mix, mineral oil, tubes strip containing lyophilized primers, and controls.
Therefore, different parameters were analyzed accordingly to the ISO. In particular, the DNA from 1–15 serogroups of L. pneumophila was tested (inclusivity), and for exclusivity, the DNA from 18 microorganisms not belonging to the species L. pneumophila or the genus Legionella spp. was tested, as specified in Table 1. Some strains among those tested were isolated from clinical samples, identified, and confirmed by biochemical identification (Phoenix, BD).

2.2. Bacterial Growth Conditions

A total of 25 Legionella spp. strains were supplied by the DSMZ (Braunschweig, Germany) and ATCC (American Type Culture Collection, Manassas, VA, USA), and rehydrated, collected, and maintained in culture from the laboratory of the Department of Health Promotion, Mother and Child Care, Internal Medicine and Medical Specialties (University of Palermo), as recommended ISO [9].
Briefly, the 29 strains included in the exclusivity test were chosen based on their presence in the same ecological niche and/or phylogenetic correlation, according to Olabarria et al. [16]. All 25 Legionella spp. strains used in this study (Table 1) were rehydrated in a nutrient broth and cultured on a Legionella CYE agar base supplemented with L-cysteine HCl (Oxoid Ltd., Hampshire, UK) incubated at 35 ± 2 °C, 3% CO2 for 5 days and examined for atypical presence of Legionella colonies. Nineteen non-Legionella bacteria were rehydrated in a nutrient broth and cultured according to their appropriate growth agar media and temperature.
L. pneumophila and Legionella spp. strains were maintained on cryogenic beads at −20 °C; before use, the beads were placed on buffered charcoal yeast extract (BCYE) plates and incubated at 37 °C for 2–5 days in humidified atmosphere (air with 5% CO2). Other bacterial strains were placed in tryptone soy agar (TSA) or Columbia blood agar (CBA), incubated at 37 °C for 18–24 h and maintained at 4 °C for two weeks.

2.3. Sensitivity and Robustness

Sensitivity was established through serial dilutions of L. pneumophila plasmid (DSM 27564) DNA, starting from 1 ng/μL. The last positive dilution was tested with 30 independently laced replicates. Of the replicates, 90% had to be positive to establish the limit of detection (LOD).
In addition, the method robustness was also studied. It consisted of 10 independent samples for three different matrices at two levels of contamination, evaluating whether the detection of the pathogen is not affected by the matrix tested. The following types of matrices were analyzed according to AFNOR NF148 (2015): mineral water, domestic hot water, and cooling tower water. Analyses were carried out under intra-laboratory reproducibility conditions (different days and operators), with a blank sample in addition.

2.4. Types of Water Samples

Hence, in total, 60 water samples and 15 blank samples were tested. Firstly, these matrices were tested free from Legionella spp. and were then artificially spiked, reaching a final contamination of 105 and 107 CFU/L of L. pneumophila (DSM 27564). Lastly, a performance comparison between LAMP and the reference method provided by the ISO [17] was also carried out. Specifically, 72 samples were tested using both methods.

3. Results

3.1. Inclusivity and Exclusivity

As well as inclusivity and exclusivity, all 15 L. pneumophila serogroups showed amplification. The other species tested for exclusivity were negative, except for L. parisiensis and L. tucsonensis, which showed DNA amplification, as listed in Table 1.
The last positive dilution was 10−7 (i.e., 0.0000001 ng/μL) and 27 replicates out of 30 showed amplifications, respecting the 90% confidence interval. Sensitivity was obtained through the following equation, based on the average weight of a base pair (bp) of 650 Daltons (Copy number calculator [Internet]. Rhode Island; available online: http://sciprim.com/html/copyNumb.v2.0.html, accessed on 20 April 2024):
Number   of   copies = n g × ( 6.022 × 10 23 ) k b × (   660 × 10 9 )
  • ng = the amount of DNA (in ng) of the last dilution positive for the tested method
  • 6.022 × 1023 = Avogadro’s number
  • kb = plasmid kilobases
  • 660 = mean weight of a nucleobase (dalton)
  • 1 × 109 = conversion factor to ng
Thus, the lowest number of copies detected by the LAMP method is 28 plasmid copies/μL.

3.2. Robustness (or Strength)

The results proved 100% of robustness (or strength). The positive samples detected at both contamination levels, 105 and 107 CFU/mL, showed confirmed reproducibility, as shown in Table 2.

3.3. Sensitivity and Specificity

In the comparative study, the cultural method and “Legionella pneumophila Glow” obtained the same results, showing full agreement: out of 72 samples, 16 were confirmed positive and 56 were negative. The relative accuracy (AC), relative specificity (SP), and sensitivity (SE) were obtained as follows:
AC = (TP + TN)/(PA + NA + FN + FP) × 100%
SP = (TN)/(TN + FP) × 100%
SE = (TP)/(TP + FN) × 100%
TP = true positive samples; TN = true negative samples; FN = false negative samples; and FP = false positive samples (Table 3).

4. Discussion

L. pneumophila investigation requires rapid isolation and identification to prevent potential infections. A rapid response time could help diagnose and treat water for warding off outbreak events. Thus, the use of alternative methods that rapidly identify the source of contamination is important. In the case of Legionella infection, fast identification is mandatory considering its slow-growing bacteria sources that could reach areas or facilities with high population densities. In addition, cultural methods may misdiagnose VBNC cells, which are common environmental forms of Legionella.
In this study, all the criteria and parameters tested were achieved according to the ISO requirements, with 100% inclusivity and robustness. The specificity was set at 92%. Legionella species such as L. parisiensis and L. tucsonensis that are rarely isolated showed amplification. L. parisiensis was isolated by [18]; only two human isolations have been reported: a liver transplant patient with pneumonia in France [19] and an immunosuppressed patient in Germany [20]. Otherwise, L. tucsonensis was isolated from a pleural fluid specimen from an immunosuppressed patient [21]; after that, it has never been sampled, despite environmental and clinical surveillance [22].
Therefore, the obtained data support the suitability of this kit for water matrices. In particular, “Legionella pneumophila Glow” proved to be useful for ease of use and significant time reduction in the analytical process. In addition, its analysis time of about 90 min and high sensitivity (28 plasmid copies/μL) may be implemented routinely by competent authorities and laboratories as an effective screening tool for L. pneumophila, helping prevent legionnaire.

5. Conclusions

The data in this study support the suitability of the “Legionella pneumophila Glow” kit for commercial use for different water samples.
This molecular method aims to improve the reliability of the results, making it a promising tool that should be used in addition to cultural analysis. LAMP, as a screening tool, offers a lot of advantages in terms of specificity, sensitivity, and a considerable reduction in time analysis, useful in timely corrective actions. In conclusion, we would like to underline that the investigated method has been validated, showing appropriate performances in accordance with the international standard [9]. Although culturing represents the reference method, as suggested by other authors, PCR should be used simultaneously to culture rather than as an alternative.

Author Contributions

Conceptualization, G.C., M.A.C., P.L. and T.F.; methodology, G.C., I.A. and G.A.; validation, G.C., M.A.C., P.L. and T.F.; investigation, G.C., I.A., M.A.C., P.L., T.F. and A.G.; data curation, M.P. and A.G.; writing—original draft preparation, G.C.; writing—review and editing, G.C., M.A.C., P.L., T.F. and A.G.; visualization, G.C., M.A.C., P.L., P.D.C. and T.F.; funding acquisition, M.P. and A.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by Regione Siciliana, Azienda Ospedaliera Universitaria Policlinico “P. Giaccone” of Palermo. Progetto Obiettivo di Piano Sanitario Nazionale 2016. Azione “4.9.2—Attività di coordinamento dei CIO (Comitato Infezioni Ospedaliere) per il controllo e la diffusione dei microrganismi Multi Drug Resistant (MDR)”.

Data Availability Statement

Data are contained within the article.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Berjeaud, J.-M.; Chevalier, S.; Schlusselhuber, M.; Portier, E.; Loiseau, C.; Aucher, W.; Lesouhaitier, O.; Verdon, J. Legionella pneumophila: The Paradox of a Highly Sensitive Opportunistic Waterborne Pathogen Able to Persist in the Environment. Front. Microbiol. 2016, 7, 486. [Google Scholar] [CrossRef] [PubMed]
  2. Simões, L.C.; Simões, M.; Vieira, M.J. Influence of the diversity of bacterial isolates from drinking water on resistance of biofilms to disinfection. Appl. Environ. Microbiol. 2010, 76, 6673–6679. [Google Scholar] [CrossRef] [PubMed]
  3. Dietersdorfer, E.; Kirschner, A.; Schrammel, B.; Ohradanova-Repic, A.; Stockinger, H.; Sommer, R.; Walochnik, J.; Cervero-Aragó, S. Starved viable but non-culturable (VBNC) Legionella strains can infect and replicate in amoebae and human macrophages. Water Res. 2018, 141, 428–438. [Google Scholar] [CrossRef] [PubMed]
  4. Beauté, J.; on Behalf of the European Legionnaires’ Disease Surveillance Network. Legionnaires’ disease in Europe, 2011 to 2015. Eurosurveillance 2017, 22, 30566. [Google Scholar] [CrossRef] [PubMed]
  5. Sakamoto, R. Legionnaire’s disease, weather and climate. Bull World Health Organ. 2015, 93, 435–436. [Google Scholar] [CrossRef] [PubMed]
  6. Correia, A.M.; Ferreira, J.S.; Borges, V.; Nunes, A.; Gomes, B.; Capucho, R.; Gonçalves, J.; Antunes, D.M.; Almeida, S.; Mendes, A.; et al. Probable Person-to-Person Transmission of Legionnaires’ Disease. N. Engl. J. Med. 2016, 374, 497–498. [Google Scholar] [CrossRef] [PubMed]
  7. Fasciana, T.; Mascarella, C.; Distefano, S.A.; Calà, C.; Capra, G.; Rampulla, A.; Di Carlo, P.; Palermo, M.; Giammanco, A. Cluster of Legionnaires’ Disease in an Italian Prison. Int. J. Environ. Res. Public Health 2019, 16, 2062. [Google Scholar] [CrossRef] [PubMed]
  8. Arrigo, I.; Galia, E.; Fasciana, T.; Diquattro, O.; Tricoli, M.R.; Serra, N.; Palermo, M.; Giammanco, A. Four-Year Environmental Surveillance Program of Legionella spp. in One of Palermo’s Largest Hospitals. Microorganisms 2022, 10, 764. [Google Scholar] [CrossRef] [PubMed]
  9. ISO/TS 12869:2019 (en); Water quality—Detection and Quantification of Legionella spp. and/or Legionella pneumophila by Concentration and Genic Amplification by Quantitative Polymerase Chain Reaction (qPCR). ISO (International Organization for Standardization): Geneva, Switzerland, 2019.
  10. Notomi, T.; Okayama, H.; Masubuchi, H.; Yonekawa, T.; Watanabe, K.; Amino, N.; Hase, T. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000, 28, e63. [Google Scholar] [CrossRef] [PubMed]
  11. Martzy, R.; Kolm, C.; Brunner, K.; Mach, R.L.; Krska, R.; Šinkovec, H.; Sommer, R.; Farnleitner, A.H.; Reischer, G.H. A loop-mediated isothermal amplification (LAMP) assay for the rapid detection of Enterococcus spp. in water. Water Res. 2017, 122, 62–69. [Google Scholar] [CrossRef] [PubMed]
  12. Parida, M.; Posadas, G.; Inoue, S.; Hasebe, F.; Morita, K. Real-time reverse transcription loop-mediated isothermal amplification for rapid detection of West Nile virus. J. Clin. Microbiol. 2004, 42, 257–263. [Google Scholar] [CrossRef] [PubMed]
  13. Besuschio, S.A.; Murcia, M.L.; Benatar, A.F.; Monnerat, S.; Cruz, I.; Picado, A.; Curto, M.d.L.; Kubota, Y.; Wehrendt, D.P.; Pavia, P.; et al. Analytical sensitivity and specificity of a loop-mediated isothermal amplification (LAMP) kit prototype for detection of Trypanosoma cruzi DNA in human blood samples. PLOS Neglected Trop. Dis. 2017, 11, e0005779. [Google Scholar] [CrossRef]
  14. Tomita, N.; Mori, Y.; Kanda, H.; Notomi, T. Loop-mediated isothermal amplification (LAMP) of gene sequences and simple visual detection of products. Nat. Protoc. 2008, 3, 877–882. [Google Scholar] [CrossRef] [PubMed]
  15. Francois, P.; Tangomo, M.; Hibbs, J.; Bonetti, E.-J.; Boehme, C.C.; Notomi, T.; Perkins, M.D.; Schrenzel, J. Robustness of a loop-mediated isothermal amplification reaction for diagnostic applications. FEMS Immunol. Med. Microbiol. 2011, 62, 41–48. [Google Scholar] [CrossRef] [PubMed]
  16. Olabarria, G.; Eletxigerra, U.; Rodriguez, I.; Bilbao, A.; Berganza, J.; Merino, S. Highly sensitive and fast Legionella spp. in situ detection based on a loop mediated isothermal amplification technique combined to an electrochemical transduction system. Talanta 2020, 217, 121061. [Google Scholar] [CrossRef] [PubMed]
  17. ISO 11731:2017; Water quality-Enumeration of Legionella. ISO (International Organization for Standardization): Geneva, Switzerland, 2017.
  18. Brenner, D.J.; Steigerwalt, A.G.; Gorman, G.W.; Wilkinson, H.W.; Bibb, W.F.; Hackel, M.; Tyndall, R.L.; Campbell, J.; Feeley, J.C.; Thacker, W.L.; et al. Ten new species of Legionella. Int. J. Syst. Evol. Microbiol. 1985, 35, 50–59. [Google Scholar] [CrossRef]
  19. Presti, F.L.; Riffard, S.; Vandenesch, F.; Reyrolle, M.; Ronco, E.; Ichai, P.; Etienne, J. The first clinical isolate of Legionella parisiensis, from a liver transplant patient with pneumonia. J. Clin. Microbiol. 1997, 35, 1706–1709. [Google Scholar] [CrossRef] [PubMed]
  20. Igel, L.; Helbig, J.H.; Lück, P.C. Isolation and characterization of a nonfluorescent strain of Legionella parisiensis. J. Clin. Microbiol. 2004, 42, 2877–2878. [Google Scholar] [CrossRef] [PubMed]
  21. Thacker, W.L.; Benson, R.F.; Schifman, R.B.; Pugh, E.; Steigerwalt, A.G.; Mayberry, W.R.; Brenner, D.J.; Wilkinson, H.W. Legionella tucsonensis sp. nov. isolated from a renal transplant recipient. J. Clin. Microbiol. 1989, 27, 1831–1834. [Google Scholar] [CrossRef] [PubMed]
  22. Pearce, M.M.; Theodoropoulos, N.; Mandel, M.J.; Brown, E.; Reed, K.D.; Cianciotto, N.P. Legionella cardiaca sp. nov., isolated from a case of native valve endocarditis in a human heart. Int. J. Syst. Evol. Microbiol. 2012, 62 Pt 12, 2946–2954. [Google Scholar] [CrossRef] [PubMed]
Table 1. L. pneumophila serogroups tested for inclusivity. and strains tested for exclusivity (non-target microorganisms).
Table 1. L. pneumophila serogroups tested for inclusivity. and strains tested for exclusivity (non-target microorganisms).
SpeciesCode
L. pneumophila serogroup 1DSM 27564
L. pneumophila serogroup 2DSM 25071
L. pneumophila serogroup 3Not available
L. pneumophila subsp. fraseri serogroup 4ATCC 33156
L. pneumophila serogroup 5DSM 24991
L. pneumophila serogroup 6Not available
L. pneumophila serogroup 7Not available
L. pneumophila serogroup 8Not available
L. pneumophila serogroup 9DSM 25001
L. pneumophila serogroup 10Not available
L. pneumophila serogroup 11DSM 25063
L. pneumophila serogroup 12DSM 25224
L. pneumophila serogroup 13Not available
L. pneumophila serogroup 14Not available
L. pneumophila serogroup 15ATCC 35251
Legionella longbeachaeDSM 25315
Legionella londiniensisNot available
Legionella anisaNot available
Tatlockia micdadei (o Legionella micdadei)DSM 16640
Fluoribacter gormanii (o L. gormanii)DSM 25296
Fluoribacter dumoffii (o L. dumoffii)DSM 17625
Fluoribacter bozemanae (o L.bozemanae o L. bozemanii)DSM 16523
Legionella jordanisDSM 19212
Legionella parisiensisDSM 19216
Legionella tucsonensisDSM 19246
Flavobacterium aquicolaDSM 100880
Burkholderia cepaciaNot available
Bacillus subtilisBCS51
Aeromonas hydrophilaATCC 35650
Clostridium perfringensATCC 13124
Listeria monocytogenesATCC 7684
Staphylococcus aureusDSM 2569
Alcaligenes faecalisNot available
Enterobacter aerogenesNot available
Escherichia coliNot available
Klebsiella oxytocaNot available
Proteus vulgarisNot available
Pseudomonas aeruginosaNot available
Pseudomonas fluorescensNot available
Pseudomonas putidaNot available
Serratia marcescensNot available
Stenotrophomonas maltophiliaNot available
Enterococcus faecalisNot available
Klebsiella pneumoniaeNot available
Table 2. Robustness of the method with two contamination levels, 105 and 107 CFU/mL. P = positive result, N = negative result.
Table 2. Robustness of the method with two contamination levels, 105 and 107 CFU/mL. P = positive result, N = negative result.
Matrices105 CFU/mL
(n. Samples)
Results107 CFU/mL
(n. Samples)
ResultsBlank Samples
(n. Samples)
Results
Control mineral water10P10P5N
Domestic hot water10P10P5N
Cooling tower water10P10P5N
TOT.30 30 15
Table 3. Relative accuracy, sensitivity, and specificity of the alternative method.
Table 3. Relative accuracy, sensitivity, and specificity of the alternative method.
Legionella pneumophila GlowISO 11731:2017 [17]
Positive (+)Negative (−)
Water samplesPositive (+)160
Negative (−)056
Total samples9951
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Caruso, G.; Coniglio, M.A.; Laganà, P.; Fasciana, T.; Arcoleo, G.; Arrigo, I.; Di Carlo, P.; Palermo, M.; Giammanco, A. Validation of a Loop-Mediated Isothermal Amplification-Based Kit for the Detection of Legionella pneumophila in Environmental Samples According to ISO/TS 12869:2012. Microorganisms 2024, 12, 961. https://doi.org/10.3390/microorganisms12050961

AMA Style

Caruso G, Coniglio MA, Laganà P, Fasciana T, Arcoleo G, Arrigo I, Di Carlo P, Palermo M, Giammanco A. Validation of a Loop-Mediated Isothermal Amplification-Based Kit for the Detection of Legionella pneumophila in Environmental Samples According to ISO/TS 12869:2012. Microorganisms. 2024; 12(5):961. https://doi.org/10.3390/microorganisms12050961

Chicago/Turabian Style

Caruso, Giorgia, Maria Anna Coniglio, Pasqualina Laganà, Teresa Fasciana, Giuseppe Arcoleo, Ignazio Arrigo, Paola Di Carlo, Mario Palermo, and Anna Giammanco. 2024. "Validation of a Loop-Mediated Isothermal Amplification-Based Kit for the Detection of Legionella pneumophila in Environmental Samples According to ISO/TS 12869:2012" Microorganisms 12, no. 5: 961. https://doi.org/10.3390/microorganisms12050961

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop