Systematic Review

The Effect of Different Electric Toothbrush Technologies on Interdental Plaque Removal: A Systematic Review with a Meta-Analysis

School of Medicine and Dentistry, University of Central Lancashire, Preston PR1 2HE, UK
*
Author to whom correspondence should be addressed.
Healthcare 2024, 12(10), 1035; https://doi.org/10.3390/healthcare12101035
Submission received: 24 March 2024 / Revised: 12 May 2024 / Accepted: 13 May 2024 / Published: 16 May 2024

Abstract

:
The removal of dental plaque from the gingival margins of the teeth is essential to maintaining periodontal health. Whilst it has been established that electric toothbrushes demonstrate a greater ability to remove plaque, no systematic review has specifically investigated which technology is better for removing plaque from the interdental tooth surfaces, where plaque control may be more difficult. Three databases were searched until October 2023: MEDLINE and DOSSS via EBSCOhost and Embase. Data extraction was carried out on studies which met the inclusion criteria, and a risk of bias assessment was completed. The study findings were combined via a narrative synthesis and a meta-analysis where appropriate. A total of 77 studies were found, out of which 14 were selected and included in the analysis. The mean difference in interproximal plaque reduction, measured using the Rustogi Modified Naval Plaque Index (RMNPI) at 8 weeks, was 0.09 (p < 0.00001) in favor of the oscillating–rotating toothbrush. At 6 and 12 weeks, the mean difference in plaque reduction (RMNPI) was 0.05 (p = 0.0008) and 0.04 (p = 0.0001) in favor of the oscillating–rotating toothbrush, respectively. The studies show a tendency for oscillating–rotating toothbrushes to remove more interproximal plaque than oscillating toothbrushes, especially in a short time (8 weeks).

1. Introduction

The association between dental plaque, gingivitis and periodontitis has long been established [1,2]. Dental plaque contains bacteria which irritate the gingival tissues, causing them to become inflamed, leading to gingivitis. If left to progress, gingivitis can lead to the breakdown of the periodontium [3,4]. Furthermore, periodontitis has been linked with other systemic chronic inflammatory conditions, meaning its prevention can have a positive impact on overall health as well as oral health [4]. Lang et al. [5] demonstrated that plaque formation begins at the interdental sites. Despite this, Marchesan et al. [6] found that nearly one-third of respondents based in the United States reported that they do not use any form of interdental cleaning aid. This study also found that those who did report interdental cleaning as part of their oral hygiene routine at home were more likely to exhibit periodontal health. It is clear, therefore, that whilst the removal of interdental plaque is important in the prevention of diseases of the periodontium, interdental cleaning is not routinely carried out by a sizeable proportion of patients [7,8]. If interdental cleaning is not ubiquitous within the population, advocating for the type of toothbrush which is most effective in removing interdental plaque may go some way towards addressing this issue, ensuring that patients who are perhaps reluctant or have low motivation to carry out designated interdental plaque removal still reap the maximum benefit from the toothbrush they use.
A consensus report by Chapple et al. [4] concluded that mechanical plaque removal is recommended to maintain gingival health and that the evidence base suggests that electric toothbrushes demonstrate a greater reduction in plaque than manual brushes both in the short (up to 3 months) and long term (over 3 months), which is also reflected by a reduction in gingival inflammation. This consensus report is supported by the findings of a Cochrane review [9], which again supports the use of electric toothbrushes over manual toothbrushes for plaque reduction. However, the analysis did not directly compare different electric toothbrush types with one another, despite their different modes of action and required brushing techniques. Patients often look to their dental clinician for guidance when it comes to oral hygiene, and such a comparison may help to guide clinical recommendations as to which type of tool should be used by patients in a more performant way. Whilst it can be concluded from this review that electric toothbrushes are more effective than manual toothbrushes in removing plaque, directly comparing different electric toothbrush modalities to one another may provide clinicians with more certainty when providing toothbrush recommendations to their patients, should one modality demonstrate a greater cleaning ability. A recent systematic review [10] directly compared different electric toothbrush modalities. The authors concluded that oscillating–rotating electric toothbrushes were the most efficacious at removing plaque when compared to manual and sonic toothbrushes, but the review did not investigate their effects on interproximal plaque reduction, despite the interproximal sites being highly at risk of plaque formation [5]. Thomassen et al. [11] also compared manual, oscillating–rotating and sonic electric toothbrushes and found a marginal benefit in plaque removal with the oscillating–rotating brush compared to the other two modalities. Again, however, this study makes no specific reference to interproximal plaque reduction.
Zou et al. [12] compared different electric toothbrush modalities in their systematic review, which found that oscillating–rotating electric toothbrushes removed statistically significantly more interproximal plaque than both manual and sonic toothbrushes. However, this study was not independently carried out, as the authors are affiliated with the manufacturer of an oscillating–rotating toothbrush.
To date, there has been no independent systematic review published which focuses specifically on the efficacy of interproximal plaque reduction using different electric toothbrush modalities. The present systematic review aims to address this.
The research question of the present review was “Does the use of an oscillating–rotating electric toothbrushing remove more interdental plaque compared to an oscillating (sonic) electric toothbrushing in adults?”.

2. Materials and Methods

2.1. Study Registration

The current systematic review was registered on PROSPERO on 5 April 2022, with the number CRD42022319371.

2.2. Literature Search

Following the recommendations of the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement [13], a PICOS framework was used to develop and define the research question:
P (Population): Adults performing oral hygiene.
I (Intervention): Any electric toothbrush in which the bristles move according to oscillating–rotating action.
C (Comparator): Any electric toothbrush which employs an oscillating “side-to-side” movement, with no rotational component, often termed “sonic” toothbrushes.
O (Outcomes): Interproximal plaque index variation from baseline examination.
S (Study design): Randomised controlled trials only.
Three databases were searched for relevant articles in January 2022: MEDLINE and Dentistry and Oral Sciences Source (via EBSCOhost) and Embase (via Ovid).
The searches were carried out and adapted as appropriate for each database (Table 1). Boolean operators were used to combine the searches. Each individual PICO component was combined with OR. The final search combined the PICO terms with AND. A manual search was carried out on field-related journals such as Journal of Periodontology, Journal of Clinical Periodontology, International Journal of Dental Hygiene and the American Journal of Dentistry. There are two commonly used brands of electric toothbrush which utilise either an oscillating–rotating or sonic mode of action, as can be seen from Amazon’s “Best Sellers in Electric Toothbrushes” webpage [14]. Representatives from these two brands, namely Philips and Oral-B, were contacted by email to see whether they could provide any gray literature which may address the research question.

2.3. Literature Selection

The eligibility criteria were set out as follows:

2.3.1. Inclusion Criteria

The following table (Table 2) shows the inclusion criteria of the studies.
The included studies had to have been published from the year 2000 onwards, as it is reasonable to assume that electric toothbrush technology has changed considerably in that time [15], and any conclusions drawn from this review should be based on the electric toothbrush technologies available to the public today.

2.3.2. Exclusion Criteria

Table 3 shows the exclusion criteria for the present review.

2.4. Literature Screening

The studies generated by the search were first screened by title by two of the authors (R.L. and F.P.). Relevant studies from the database searches were imported into RefWorks [16]. Duplicates were excluded, and the studies were then checked by the same two authors through two levels of screening (titles and abstracts) and the inter-agreement score was recorded according to Cohen’s Kappa score to check the consistency [17] (Table S1). “Maybes” were not counted as disagreements. Any disagreement was discussed and, where appropriate, moderated by a third reviewer (N.C.). The final searches were all completed in October 2023. The results were summarised into a PRISMA flowchart (Figure 1) [13].

2.5. Outcome Measures

The outcome measure of interest was the reduction in the reported interproximal plaque level using the Rustogi Modified Naval Plaque Index (RMNPI). This index extends the scoring of plaque to the interproximal (mesial and distal) tooth areas and to the gumline (marginal gingival) region, as well as the total tooth. It divides the buccal and lingual surfaces into nine areas, which are scored on the presence or absence of dental plaque [19]. Alternative plaque indices were also considered. Turesky modification of the Quigley–Hein index (TMQH) considers the amount of the tooth surface covered by plaque as a fraction of the total tooth surface and splits the buccal and lingual tooth surfaces into mesial, distal and mid [19]. The wet weight of interproximal plaque was also used as an outcome measure.

2.6. Data Extraction

A data extraction table was used for each study so that relevant information could be laid out in an accessible format for later comparison (Table 4). Development of the data extraction table was an iterative process based on the findings from the first few studies, which were screened to ensure that all relevant data were extracted. Of particular interest were the study design, outcome measures and the results on the interproximal plaque reduction. The data extraction table was based on the Cochrane guidelines [20].

2.7. Assessment of Heterogeneity

A thorough assessment of the heterogeneity across the included studies was detailed according to the following factors:
(a)
Study design,
(b)
Trial duration and evaluation timeline,
(c)
Population sample and demographics,
(d)
Baseline oral health features across the test and control groups,
(e)
Smoking habits,
(f)
Previous use of manual or electric toothbrushes.

2.8. Quality Assessment

Once the data extraction was completed, the papers were critically appraised using the Cochrane Risk of Bias 2 (RoB 2) tool, which was developed to allow for a more standardised way of assessing bias compared to its predecessor [35] and the Jadad tool [36].
RoB 2’s methodological quality was evaluated according to the five risk of bias domains of randomisation, deviation from the intended intervention, missing outcome data, measurement of the outcome and the selection of the reported result. The overall and domain-related judgement for each included paper was reported in tabular format, displaying the risk of bias in different colors (red for high-risk, amber for some concerns and green for low-risk) according to the Cochrane RoB 2 resources tool [37]. The Jadad tool is a numeric scale based on three items: randomisation, blinding and drop-out reporting. The appraisal was carried out independently by two reviewers (R.D.L. and F.P.) and measured according to Cohen’s Kappa score to check the consistency. It was agreed to include the studies at moderate risk of bias in the qualitative and quantitative syntheses due to the expected small number of retrievable studies.

2.9. Data Synthesis and Publication Bias Check

Following data extraction and critical appraisal, the included papers were analysed by way of narrative synthesis. To guide the narrative synthesis process, the flowchart by Rodgers et al. [38] was consulted, which provides guidance on how to synthesise the data, look for patterns within them, establish the strength of these patterns and then make conclusions based on these patterns.
To identify any potential publication bias, a funnel plot analysis was performed [39].
Where appropriate, dependent on the studies having similar characteristics and methodologies, a meta-analysis was carried out on the outcome data, displayed as the Weighted Mean Difference (WMD). For continuous variables, mean differences and 95% confidence intervals were used to summarise the data for each study. Forest plots and funnel plots were created to illustrate the effects of different studies and the overall average results. Review Manager (RevMan) version 5.3. for macOs from Cochrane collaboration was used for all the analyses. Statistical significance was set as a p value < 0.05.
The statistical heterogeneity among the studies was assessed using the I2 test [40]. A random-effects model was adopted as per the hypothesis of a population of studies with possible variations.

3. Results

3.1. Study Selection

From an initial batch of 156 articles retrieved via databases and registers, 90 were removed as duplicates. Eleven studies were yielded from the manual search and from a recently published systematic review [10].
Figure 1 shows a PRISMA flowchart [18] which documents the transition from the initial results to the final ones for data extraction.
Seventeen studies were eligible for full-text review; however, four were not retrievable. The remaining 13 studies were retrieved, fully assessed and included. From the manual search, 6 out of 10 were available for the full-text assessment, of which all but 1 [29] were excluded, due to a lack of reporting on the interproximal plaque level (3/6), to the sonic toothbrush being multi-directional (1/6) or to the lack of a sonic toothbrush control group in the study (1/6). Fourteen studies were included at the end of the selection process.
To test the agreement in the process performance between the two reviewers (R.L. and F.P.), Cohen’s Kappa statistics were used at different stages of the screening [17]. The mean overall calculated score was 0.92, which indicated the substantial reliability of the overall procedure [41]. Disagreements arose when there was some ambiguity within the abstracts of the articles as to whether interproximal plaque was specifically reported on. Following discussion, where uncertainty remained, the papers were retrieved so that the full text could be analysed and either included or excluded as appropriate based on the aforementioned study selection criteria.
The extraction of data from each study was carried out by one supervised single author (R.L.) using a standardised tabular set. The format reflected the PICOS template, providing specific demographic data, the main features of the intervention and comparative groups and the main outcome characteristics in terms of the interproximal plaque variation from the baseline to the staged timelines as means (M) and standard deviation (SD)/standard error (SE). Once complete, the data ultimately suitable for the quantitative analysis were checked, grouped by time and processed with the use of statistical software (RevMan 5.3). Table 4 shows the study characteristics of the included studies.

3.2. Study Methodologies

Almost all the studies (10) described a stratification process for randomisation. One study [22] used a set of two Latin squares for randomisation, while four [27,29,30,31] did not clearly describe their process beyond stating that participants were randomised.
The participant numbers within the studies ranged from 148 [31] to 45 [25]. Twelve of the studies assessed plaque reduction using the Rustogi Modified Naval Plaque Index (RMNPI). One study [26] used the Turesky modified Quigley–Hein plaque index (TMQH). One [21] used the wet weight of plaque to measure the reduction.

3.3. Risk of Bias

The Cochrane Risk of Bias 2 (RoB 2) tool [37] was used to assess each study and to assign the overall risk of bias, while the Jadad tool [36] was used as a reference. The results of this were depicted on a colour-coded chart for easy consultation (Table 5).
Overall, three of the studies [33,35,36] were deemed to be at moderate risk of bias due to their adherence to the intervention category, as the test and control groups differed in their use of apps or brushing guides to help ensure thorough toothbrushing. It is not clear to what extent these brushing aides affected the results between the test and control groups, but as the results fall in line with the results from most of the papers, a decision was made to score them as moderate and still include them in the analysis.
All of the remaining studies were deemed to have a low risk of bias. Comparative analysis of the RoB 2 and Jadad tools helped to clarify whether the existence of reported dropouts would affect the quality of the study, as per the possible underpowered sample used by Cchauana Vasquez et al. [31].
To test the agreement in the process performance between the two reviewers (R.L. and F.P.), a Cohen’s Kappa statistical test was used to calculate the different quality assessments performed independently.
The mean calculated Kappa score was 0.57, which indicated the substantial discrete reliability of the overall procedure [41]. Disagreements in the risk of bias score were due to the subjective nature of the RoB 2 tool. They were resolved following discussion between the two reviewers until an agreement was made.
A decision was made to still include the studies with a moderate risk of bias and test them for heterogeneity where appropriate in the meta-analysis, thereby generating as large a dataset as possible from which to draw conclusions.

3.4. Categorisation of Studies

The included studies were categorised into different groups according to the follow-up interval with the aim of making them comparable for the quantitative meta-analysis. Four groups were available for the analysis at 6, 8, 12 and 24 weeks.

3.5. Qualitative Synthesis

The included trials were all randomised controlled trials, 6 with a crossover and 8 with a parallel design.
The earliest study was published in 2004 [21], and the most recent one was published in 2021 [34]. Most of the studies allowed the use of the test and control toothbrushes at home, although one did not [22]. The studies had a range of follow-up times, from four days to 24 weeks. Four of the studies provided results for 12 weeks from the baseline [33,34,35,38], which was the most common follow-up period. Two studies reported findings at 24 weeks [38,40], two at 8 weeks [37,39] and three at 6 weeks [33,35,36] from the baseline. All the studies used a single, blinded examiner, reducing the risk of examiner bias in the pre- and post-brushing plaque assessments. All of the papers used an Oral-B oscillating–rotating toothbrush as the oscillating–rotating brush, and the majority (12) of them used a Philips Sonic toothbrush (Sonicare). Only one study [28] did not include the Philips sonic electric toothbrush, instead using one made by Colgate with the same technology. One study [32] investigated the GEVILAN GET011 sonic toothbrush as well as the Philips Sonicare brush.

3.6. Interproximal Plaque Reduction

3.6.1. Overall Comparison of OR TBs and Sonic TBs

Overall, most of the studies found that the oscillating–rotating electric toothbrush removed more interproximal dental plaque than the sonic toothbrush. Only two studies [27,38] found more beneficial effects from the sonic toothbrush. Based on the RMNPI as the interproximal plaque outcome measure, the largest reduction from the baseline for an oscillating–rotating toothbrush was found by Strate et al. [23] (mean RMNPI reduction 0.966 (SD = 0.06)). However, this study compared the test and control toothbrushes only once. The largest reduction over a more prolonged period was found by Biesbrock et al. [25], 12 days after the baseline assessment (mean RMNPI reduction 0.884 ± 0.013). For a sonic toothbrush, the largest reduction was found by Goyal et al. [34] (mean RMNPI reduction = 0.794 ± 0.0101) at just day 1 following the baseline assessment, although it was less than that of the oscillating–rotating brush, while the next highest reduction was found by Biesbrock et al. [25] at 12 days (mean RMNPI reduction = 0.724 ± 0.013).

3.6.2. Time-Wise Comparison of OR TBs and Sonic TBs

In order to visualise the comparison, a time-wise analysis according to the different milestones has been reported (Table 6).

3.7. Quantitative Analysis

3.7.1. Meta-Analysis

Where appropriate, the data were synthesised according to a meta-analysis (Table 7). A concomitant funnel plot (Figure 2) was created to check the eligibility of the comparative analysis. This demonstrated an overall low risk of publication bias.
The results of the meta-analyses (Table 7) show that at 6, 8 and 12 weeks, the oscillating–rotating electric toothbrushes removed more interproximal plaque (RMNPI) than the sonic toothbrushes. The difference was greatest at eight weeks. No statistically significant difference was established at six months.

3.7.2. Heterogeneity Assessment

The I2 statistics showed substantial heterogeneity at the majority of the follow-up time periods.
Appropriate studies were grouped by follow-up time. As can be seen from the forest plots (Figure 3), the results were found to be in favour of the oscillating–rotating toothbrush in most of the studies.

4. Discussion

The results of this systematic review demonstrate that the majority of studies which address the research question have found an oscillating–rotating toothbrush to be more effective than a sonic toothbrush in removing interproximal plaque. This effect tends to be greatest in the shorter term, with the highest plaque reduction for the oscillating–rotating brush at four days following the baseline [23] and the highest for the sonic brush at one day following the baseline [34]. However, the findings were generally consistent across all time periods. At all time periods, the difference in the interproximal RMNPI reduction between the two toothbrushes was low.
It may be that the bristles of the oscillating–rotating toothbrush work best in the short term when the patients’ technique is better and they become less effective in these areas if the patients show a tendency to “scrub” their teeth with the brush over the longer term, preventing the bristles from adequately accessing the interproximal areas.
The RMNPI and TMQH have previously been compared and were both demonstrated to be adequate for assessing plaque reduction differences between different toothbrushes [19]. Weighing plaque to measure this reduction has been shown to be less performant in demonstrating the differences between pre- and post-brushing plaque levels when compared to a plaque index [42].
The meta-analyses were generally found to be in favour of an oscillating–rotating brush. However, there were a limited number of studies available for inclusion in each analysis, and only the six-week meta-analysis demonstrated an I2 value below 50%, suggesting some heterogeneity between the studies. This limits the reliability of the present meta-analysis for providing firm conclusions as to which electric toothbrush modality is best. Furthermore, the results of the meta-analyses, whilst statistically significant, may not represent clinically important reductions in interproximal plaque levels. Whilst the meta-analysis does show a tendency towards an increased interproximal plaque reduction with oscillating–rotating toothbrushes, firmer conclusions cannot be made without more homogeneous data and longer-term studies which may demonstrate more clinical significance.
This systematic review agrees with a Cochrane review which concluded that oscillating–rotating toothbrushes had the largest amount of research to support their plaque removal ability compared to other electric toothbrush designs [9]. It also supports the conclusion made by Grender, Adam and Zou [10] that oscillating–rotating toothbrushes are superior at removing plaque to sonic toothbrushes.
Three of the fourteen included studies were found to be of moderate risk of bias, while the remainder were all at low risk. There appears to be no available literature on this topic which is not supported in some way by a toothbrush manufacturer, and it must be recognised that any results published are likely to favour the supporting company’s toothbrush. The low number of published studies supported by sonic toothbrush manufacturers may suggest, however, that there have been few carried out which demonstrate that sonic brushes are better. This may also be the reason the studies which supported sonic toothbrushes used a different plaque index compared to all the others. Nevertheless, the funnel plots (Figure 2) which were generated to check for the existence of any publication bias suggested that there was not any in the present systematic review and analysis. Moreover, Heinemann et al. [43] argued that industry-sponsored studies must meet more rigorous standards than purely academic studies and that authors of both types of studies are still compelled to publish data which demonstrate statistically significant differences between the test and control groups. The quality check and the parallel use of Jadad with RoB 2 are effectively supportive of strict adherence to the study structure and results.
It was often not clear how similar the test and control groups were at baseline, and so there is the possibility that confounding factors may have accounted for the results if the two groups were not adequately similar. Participants may have even been more familiar with one of the toothbrushes being investigated, which could have led to an unfair advantage if they already felt comfortable using one toothbrush over another.
Another element to reflect on was the use of dedicated apps supporting patients in efficient toothbrushing. It is unclear how much the toothbrush mechanism, and not the effectiveness of the apps, was responsible for the differences in interproximal plaque reduction.
The recruitment process of many of the studies was unclear. The majority were carried out at a research facility, but there is no mention as to who the participants were or whether they received any financial reward for their participation in the studies. Of particular concern were the two papers by Klukowska et al. [35,36], which both reported data at six weeks from the baseline. It is not clear whether the studies represent entirely different study populations or whether there is some overlap. This is a problem, as it could mean that one study finding is then represented twice in the narrative synthesis and meta-analysis.
There are a number of limitations to this systematic review, as follows:
  • There were a limited number of studies suitable for meta-analysis.
  • There was considerable heterogeneity between the studies.
  • Some of the studies were found to be of some concern for risk of bias.
  • The longest follow-up period was six months, meaning there is a lack of long-term data available on this subject. Those studies with shorter follow-up periods are of limited value to this review, as their validity over a longer period was not established. Despite this, their findings remain in agreement with most of the longer-term studies.
Due to the widespread use of electric toothbrushes and the improvement of their technology, there is potential for future research to be carried out by independent researchers to compare these two modalities, so that the data can be published no matter which type is shown to be better. Longer-term studies could also be carried out, perhaps assessing plaque reduction up to a year from the baseline and with only a small number of assessments in between, which may be a better reflection of how effective the “real-world” interproximal plaque reduction using either toothbrush type would be.
Despite the limitations discussed above, there is evidence to suggest that oscillating–rotating toothbrushes are the more effective toothbrush type for removing interproximal plaque, as demonstrated by 12 of the 14 studies. This means that clinical recommendations based on the findings of this review can be made with some confidence, which, in turn, will help patients to make an informed choice when purchasing a toothbrush and will hopefully lead to a greater reduction in interproximal plaque for them, facilitating the maintenance of gingival health.

5. Conclusions

Oscillating–rotating electric toothbrushes appear to be more effective at removing dental plaque from the interproximal areas of teeth than sonic toothbrushes when used by adults. Across the relevant literature available on this topic, the tested oscillating–rotating toothbrushes demonstrated a tendency towards an increased reduction in interproximal plaque from the baseline compared to the sonic toothbrushes. This may be due to the generally smaller toothbrush head size allowing for better bristle placement between the teeth or due to the technique for using an oscillating–rotating brush being more easily learned by the patients. However, this difference has only been demonstrated over a relatively short time period, and the increased interproximal plaque reductions are small. Further, independent studies would be useful, perhaps reporting on a longer-term follow-up of participants, to help to overcome this issue and allow firm recommendations to be made.

Supplementary Materials

The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/healthcare12101035/s1. Table S1: Cohen’s Kappa agreement scores, excluded studies.

Author Contributions

Conceptualisation, F.P. and R.D.L.; methodology, F.P., R.D.L., K.T. and N.C.; software, F.P.; validation, R.D.L., N.C., S.K. and M.K.; formal analysis, F.P., N.C., M.K. and R.D.L.; data curation, F.P., N.C., M.K., S.K. and K.T.; writing—original draft preparation, R.D.L. and F.P.; writing—review and editing, R.D.L., F.P., N.C., M.K., K.T. and S.K.; supervision, F.P., N.C. and S.K. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflicts of interest.

Abbreviations

OR TB: oscillating–rotating toothbrush; RMNPI: Rustogi Modified Naval Plaque Index; TMQH: Turesky modification of the Quigley–Hein index.

References

  1. Ash, M.M.; Gitlin, B.N.; Smith, W.A. Correlation between plaque and gingivitis. J. Periodontol. 1964, 35, 424–429. [Google Scholar] [CrossRef]
  2. Listgarten, M.A. Pathogenesis of periodontitis. J. Clin. Periodontol. 1986, 13, 418–425. [Google Scholar] [CrossRef] [PubMed]
  3. Murakami, S.; Mealey, B.L.; Mariotti, A.; Chapple, I.L. Dental plaque–induced gingival conditions. Parodontologiya 2019, 24, 244–252. [Google Scholar]
  4. Chapple, I.L.; Van der Weijden, F.; Doerfer, C.; Herrera, D.; Shapira, L.; Polak, D.; Madianos, P.; Louropoulou, A.; Machtei, E.; Donos, N.; et al. Primary prevention of periodontitis: Managing gingivitis. J. Clin. Periodontol. 2015, 42, S71–S76. [Google Scholar] [CrossRef] [PubMed]
  5. Lang, N.P.; Cumming, B.R.; Löe, H. Toothbrushing frequency as it relates to plaque development and gingival health. J. Periodontol. 1973, 44, 396–405. [Google Scholar] [CrossRef] [PubMed]
  6. Marchesan, J.T.; Morelli, T.; Moss, K.; Preisser, J.S.; Zandona, A.F.; Offenbacher, S.; Beck, J. Interdental cleaning is associated with decreased oral disease prevalence. J. Dent. Res. 2018, 97, 773–778. [Google Scholar] [CrossRef] [PubMed]
  7. Asadoorian, J.; Locker, D. The impact of quality assurance programming: A comparison of two canadian dental hygienist programs. J. Dent. Educ. 2006, 70, 965–971. [Google Scholar] [CrossRef] [PubMed]
  8. Gufran, K.; Alanazi, K.M.; Alanazi, A.K.; Alqwiri, A.S.; Alsubaie, F.M.; Alotaibi, N.M. Self-reported knowledge and practice of interdental aids among people of Riyadh, Saudi Arabia—A cross-sectional study. J. Pharm. Bioallied Sci. 2021, 13 (Suppl. 1), S280–S283. [Google Scholar] [CrossRef] [PubMed]
  9. Yaacob, M.; Worthington, H.V.; Deacon, S.A.; Deery, C.; Walmsley, A.D.; Robinson, P.G.; Glenny, A.M. Powered versus manual toothbrushing for oral health. Cochrane Database Syst. Rev. 2014, CD002281. [Google Scholar] [CrossRef]
  10. Grender, J.; Adam, R.; Zou, Y. The effects of oscillating-rotating electric toothbrushes on plaque and gingival health: A meta-analysis. Am. J. Dent. 2020, 33, 3–11. [Google Scholar]
  11. Thomassen, T.; Van der Weijden, F.; Slot, D. The efficacy of powered toothbrushes: A systematic review and network meta-analysis. Int. J. Dent. Hyg. 2022, 20, 3–17. [Google Scholar] [CrossRef]
  12. Zou, Y.; Grender, J.; Adam, R.; Levin, L. A meta-analysis comparing toothbrush technologies on gingivitis and plaque. Int. Dent. J. 2024, 74, 146–156. [Google Scholar] [CrossRef]
  13. Moher, D.; Liberati, A.; Tetzlaff, J.; Altman, D.G.; Prisma Group. Preferred reporting items for systematic reviews and meta-analyses: The PRISMA statement. Int. J. Surg. 2010, 8, 336–341. [Google Scholar] [CrossRef]
  14. Amazon. 2 Screens. 2022. Available online: https://www.amazon.co.uk/Best-Sellers-Health-Personal-Care-Electric-Toothbrushes/zgbs/drugstore/3012128031/ref=zg_bs_nav_d_4_74141031 (accessed on 20 November 2022).
  15. Tadinada, A.; Kilham, J.; Bysani, P.; Gopalakrishna, A. The evolution of a toothbrush: From antiquity to present-A mini-review. J. Dent. Health Oral Disord. Ther. 2015, 2, 127–130. [Google Scholar] [CrossRef]
  16. Refworks [Internet]. 2022. Available online: https://refworks.proquest.com (accessed on 20 November 2022).
  17. McHugh, M.L. Interrater reliability: The kappa statistic. Biochem. Medica 2012, 22, 276–282. [Google Scholar] [CrossRef]
  18. Page, M.J.; McKenzie, J.E.; Bossuyt, P.M.; Boutron, I.; Hoffmann, T.C.; Mulrow, C.D.; Shamseer, L.; Tetzlaff, J.M.; Akl, E.A.; Brennan, S.E.; et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ 2021, 372, n71. [Google Scholar] [CrossRef] [PubMed]
  19. Cugini, M.; Thompson, M.; Warren, P.R. Correlations between two plaque indices in assessment of toothbrush effectiveness. J. Contemp. Dent. Pr. 2006, 7, 1–9. [Google Scholar] [CrossRef]
  20. Cochrane. Chapter 5: Collecting Data [Internet]. United Kingdom: The Cochrane Collaboration. 2022. Available online: https://training.cochrane.org/handbook/current/chapter-05#section-5-3-5 (accessed on 13 May 2024).
  21. Sjögren, K.; Lundberg, A.B.; Birkhed, D.; Dudgeon, D.J.; Johnson, M.R. Interproximal plaque mass and fluoride retention after brushing and flossing-a comparative study of powered toothbrushing, manual toothbrushing and flossing. Oral Health Prev. Dent. 2004, 2, 119–124. [Google Scholar]
  22. Sharma, N.C.; Goyal, C.R.; Qaqish, J.G.; Cugini, M.A.; Thompson, M.C.; Warren, P.R. Single-use plaque removal efficacy of three power toothbrushes. J. Dent. 2005, 33, 11–15. [Google Scholar] [CrossRef]
  23. Strate, J.; Cugini, M.A.; Warren, P.R.; Qaqish, J.G.; Galustians, H.J.; Sharma, N.C. A comparison of the plaque removal efficacy of two power toothbrushes: Oral-B Professional Care Series versus Sonicare Elite. Int. Dent. J. 2005, 55, 151–156. [Google Scholar] [CrossRef]
  24. Biesbrock, A.R.; Bartizek, R.D.; Walters, P.A.; Warren, P.R.; Cugini, M.; Goyal, C.R.; Qaqish, J. Clinical evaluations of plaque removal efficacy: An advanced rotating-oscillating power toothbrush versus a sonic toothbrush. J. Clin. Dent. 2007, 18, 106–111. [Google Scholar]
  25. Biesbrock, A.R.; Walters, P.A.; Bartizek, R.D.; Goyal, C.R.; Qaqish, J.G. Plaque removal efficacy of an advanced rotation-oscillation power toothbrush versus a new sonic toothbrush. Am. J. Dent. 2008, 21, 185–188. [Google Scholar]
  26. Putt, M.S.; Milleman, J.L.; Jenkins, W.; Schmitt, P.; Master, A.S.; Strate, J. A randomized crossover-design study to investigate the plaque removal efficacy of two power toothbrushes: Philips Sonicare Flexcare and Oral-B Triumph. Compend. Contin. Educ. Dent. 2008, 29, 56–58. [Google Scholar]
  27. Klukowska, M.; Grender, J.M.; Goyal, C.R.; Mandl, C.; Biesbrock, A.R. 12-week clinical evaluation of a rotation/oscillation power toothbrush versus a new sonic power toothbrush in reducing gingivitis and plaque. Am. J. Dent. 2012, 25, 287–292. [Google Scholar]
  28. Klukowska, M.; Grender, J.M.; Conde, E.; Goyal, C.R. A 12-week clinical comparison of an oscillating-rotating power brush versus a marketed sonic brush with self-adjusting technology in reducing plaque and gingivitis. J. Clin. Dent. 2013, 24, 55–61. [Google Scholar]
  29. Klukowska, M.; Grender, J.M.; Conde, E.; Ccahuana-Vasquez, R.A.; Goyal, C.R. A randomized 12-week clinical comparison of an oscillating-rotating toothbrush to a new sonic brush in the reduction of gingivitis and plaque. J. Clin. Dent. 2014, 25, 26–31. [Google Scholar]
  30. Klukowska, M.; Grender, J.M.; Conde, E.; Goyal, C.R.; Qaqish, J. A six-week clinical evaluation of the plaque and gingivitis efficacy of an oscillating-rotating power toothbrush with a novel brush head utilizing angled CrissCross bristles versus a sonic toothbrush. J. Clin. Dent. 2014, 25, 6–12. [Google Scholar]
  31. Ccahuana-Vasquez, R.A.; Conde, E.; Grender, J.M.; Cunningham, P.; Qaqish, J.; Goyal, C.R. An Eight-Week Clinical Evaluation of an Oscillating-Rotating Power Toothbrush with a Brush Head Utilizing Angled Bristles Compared with a Sonic Toothbrush in the Reduction of Gingivitis and Plaque. J. Clin. Dent. 2015, 26, 80–85. [Google Scholar]
  32. Lv, J.; Guo, B.; Ling, J. A 6-month clinical evaluation of a high frequency sonic toothbrush in comparison with an oscillating-rotating power toothbrush and a traditional sonic toothbrush in reducing gingivitis and plaque. Am. J. Dent. 2018, 31, 171–176. [Google Scholar]
  33. Adam, R.; Ram Goyal, C.; Qaqish, J.; Grender, J. Evaluation of an oscillating-rotating toothbrush with micro-vibrations versus a sonic toothbrush for the reduction of plaque and gingivitis: Results from a randomized controlled trial. Int. Dent. J. 2020, 70, S16–S21. [Google Scholar] [CrossRef] [PubMed]
  34. Goyal, C.R.; Adam, R.; Timm, H.; Grender, J.; Qaqish, J. A 6-month randomized controlled trial evaluating a novel smart-connected oscillating-rotating toothbrush versus a smart-connected sonic toothbrush for the reduction of plaque and gingivitis. Am. J. Dent. 2021, 34, 54–60. [Google Scholar] [PubMed]
  35. Sterne, J.A.; Savović, J.; Page, M.J.; Elbers, R.G.; Blencowe, N.S.; Boutron, I.; Cates, C.J.; Cheng, H.Y.; Corbett, M.S.; Eldridge, S.M.; et al. RoB 2: A revised tool for assessing risk of bias in randomised trials. BMJ 2019, 366, l4898. [Google Scholar] [CrossRef] [PubMed]
  36. Jadad, A.R.; Moore, R.A.; Carroll, D.; Jenkinson, C.; Reynolds, D.J.; Gavaghan, D.J.; McQuay, H.J. Assessing the quality of reports of randomized clinical trials: Is blinding necessary? Control. Clin. Trials 1996, 17, 1–12. [Google Scholar] [CrossRef] [PubMed]
  37. Cochrane. Risk of Bias 2 (RoB 2) Tool. 2023. Available online: https://methods.cochrane.org/risk-bias-2 (accessed on 20 February 2023).
  38. Rodgers, M.; Sowden, A.; Petticrew, M.; Arai, L.; Roberts, H.; Britten, N.; Popay, J. Testing methodological guidance on the conduct of narrative synthesis in systematic reviews: Effectiveness of interventions to promote smoke alarm ownership and function. Evaluation 2009, 15, 49–73. [Google Scholar] [CrossRef]
  39. Egger, M.; Smith, G.D.; Schneider, M.; Minder, C. Bias in meta-analysis detected by a simple, graphical test. BMJ 1997, 315, 629–634. [Google Scholar] [CrossRef] [PubMed]
  40. Higgins, J.P.; Thompson, S.G.; Deeks, J.J.; Altman, D.G. Measuring inconsistency in meta-analyses. BMJ 2003, 327, 557–560. [Google Scholar] [CrossRef] [PubMed]
  41. Landis, J.R.; Koch, G.G. An application of hierarchical kappa-type statistics in the assessment of majority agreement among multiple observers. Biometrics 1977, 33, 363–374. [Google Scholar] [CrossRef]
  42. McCracken, G.I.; Preshaw, P.M.; Steen, I.N.; Swan, M.; DeJager, M.; Heasman, P.A. Measuring plaque in clinical trials: Index or weight? J. Clin. Periodontol. 2006, 33, 172–176. [Google Scholar] [CrossRef]
  43. Heinemann, L. Are all clinical studies sponsored by industry not valid? J. Diabetes Sci. Technol. 2008, 2, 1161–1163. [Google Scholar] [CrossRef]
Figure 1. PRISMA flowchart with details of the screening stages [18]. * Consider, if feasible to do so, reporing the number of records identified from each database or register searched (rather than the total number across all databases/registers). ** If automation tools were used, indicate how many records were excluded by a human and how many were excluded by automation tools.
Figure 1. PRISMA flowchart with details of the screening stages [18]. * Consider, if feasible to do so, reporing the number of records identified from each database or register searched (rather than the total number across all databases/registers). ** If automation tools were used, indicate how many records were excluded by a human and how many were excluded by automation tools.
Healthcare 12 01035 g001
Figure 2. Funnel plots grouped by study follow-up intervals. The circles represent a single study. The y-axis shows standard error. The x-axis shows mean difference. The vertical dotted line represents the overall effect.
Figure 2. Funnel plots grouped by study follow-up intervals. The circles represent a single study. The y-axis shows standard error. The x-axis shows mean difference. The vertical dotted line represents the overall effect.
Healthcare 12 01035 g002
Figure 3. Forest plots of follow-up periods at 6, 8, 12 and 24 weeks. OR: oscillating–rotating TB; SON: sonic TB. Green boxes represent the study results. Horizontal line represents the 95% confidence interval. Diamonds represent the average of the studies within each subgroup [27,28,29,30,31,32,33,34].
Figure 3. Forest plots of follow-up periods at 6, 8, 12 and 24 weeks. OR: oscillating–rotating TB; SON: sonic TB. Green boxes represent the study results. Horizontal line represents the 95% confidence interval. Diamonds represent the average of the studies within each subgroup [27,28,29,30,31,32,33,34].
Healthcare 12 01035 g003
Table 1. Search strategy and terms. MEDLINE via EBSCOhost.
Table 1. Search strategy and terms. MEDLINE via EBSCOhost.
Main ConceptSearch NumberSearch TermBoolean Operator
PROBLEM
Interdental Plaque
S1MH “Dental plaque”
S2MH “Dental plaque index”
S3“dental plaque”
S4“dental plaque index”
S5“plaque index”
S6“dental biofilm”
S7MH “Biofilms”
S8biofilms
S9MH “Dental Deposits”
S10“Dental deposits”
S11MH “toothbrushing”
S12toothbrushing
S13MH “oral hygiene”
S14“oral hygiene”
S15 Combine S1–S14 with OR
S16interdental
S17inter-dental
S18interproximal
S19inter-proximal
S20proximal
S21approximal
S22 Combine S16–S21 with OR
FINAL PROBLEM SEARCHS23 Combine S15 and S22 with AND (search A)
INTERVENTION
Toothbrushing with an oscillating toothbrush
S24elect* N3 toothbrush*
S25power* N3 toothbrush*
S26battery N3 toothbrush*
S27 Combine S24–S26 with OR (search B)
S28oscillat*
S29sonic
S30MH “sonication”
S31sonication
S32“side-to-side”
S33MH “Vibration”
S34vibrat*
S35Sonicare
S36 Combine S28–S35 with OR (search C)
FINAL INTERVENTION SEARCHS37 Intervention final search = (search B) AND (search C)
COMPARISON
Toothbrushing with an oscillating–rotating toothbrush
S38oscillat* N3 rotat*
S39Oral B
S40 Combine S39–S40 with OR (search D)
FINAL COMPARISON SEARCHS41 Comparison final search = (search B) AND (search D)
FINAL SEARCHS42 Final search = (Final problem search) AND (Final intervention search) AND (Final comparison search)
The search was adapted for the other databases. MEDLINE: Interdental plaque search terms: MH “Dental plaque”; MH “Dental plaque index”; “dental plaque”; “dental plaque index”; “plaque index”; “dental biofilm”; MH “Biofilms”; biofilms; MH “Dental Deposits”; “Dental deposits”; MH “toothbrushing”; toothbrushing; MH “oral hygiene”; “oral hygiene”; interdental; inter-dental; interproximal; inter-proximal; proximal; approximal. Electric toothbrush search terms: elect* N3 toothbrush*; power* N3 toothbrush*; battery N3 toothbrush*. oscillating toothbrush search terms: oscillat*; sonic; MH “sonication”; sonication; “side-to-side”; MH “Vibration”; vibrat*; Sonicare. Oscillating–rotating toothbrush search terms: oscillat* N3 rotat*; Oral-B. DOSS: Interdental plaque search terms: DE “Dental plaque”; “dental plaque”; “plaque index”; DE ”BIOFILMS”; “dental biofilm”; DE “DENTAL deposits”; “dental deposits”; toothbrushing; DE “ORAL hygiene”; “oral hygiene”; interdental; inter-dental; interproximal; inter-proximal; proximal; approximal. Electric toothbrush search terms: electr* N3 toothbrush*; power* N3 toothbrush*; battery N3 toothbrush*. Oscillating toothbrush search terms: DE “OSCILLATIONS”; oscillat*; sonic; DE “SONICATION”; sonication; DE “VIBRATION (Mechanics)”; vibrat*; Sonicare; “side-to-side”. Oscillating–rotating toothbrush search terms: oscillat* N3 rotat*; Oral-B. Embase: Interdental plaque search terms: tooth plaque/; “tooth plaque”; plaque index/; “plaque index”; “dental plaque”; biofilm/; biofilm; dental deposit/; “dental deposit”; tooth brushing/; toothbrushing; mouth hygiene/; “mouth hygiene”; “oral hygiene”; interdental; inter-dental; interproximal; inter-proximal; proximal; approximal. Electric toothbrush search terms: electr* adj3 toothbrush*; power* adj3 toothbrush*; battery adj3 toothbrush*. Oscillating toothbrush search terms: oscillation/; oscillat*; sonic; Sonication; vibration/; vibrat*; Sonicare; “side-to-side”. Oscillating–rotating toothbrush search terms: oscillat* adj3 rotat*; Oral-B.
Table 2. Inclusion criteria of the studies.
Table 2. Inclusion criteria of the studies.
Inclusion Criteria
Adult participants (18 years of age or older)
Studies measuring a change in interproximal plaque, measured using the Rustogi Modified Naval Plaque Index (RMNPI), Turesky modification of the Quigley–Hein index (TMQH) or wet weight of plaque.
Studies investigating self-performed oral hygiene
Studies with a randomised controlled trial design
Studies published since the year 2000
Table 3. Exclusion criteria for the present review.
Table 3. Exclusion criteria for the present review.
Exclusion Criteria
Children (under 18 years of age)
Participants with fixed or removable orthodontic appliances
Participants with removable oral prostheses, such as a denture
Participants unable to carry out self-performed oral hygiene
Studies not written or translated into English
Studies which are not randomised controlled trials
Table 4. Study characteristics and results. OR, oscillating–rotating; RMNPI, Rustogi Modified Naval Plaque Index; MGI, Modified Gingival Index; GBI, Gingival Bleeding Index; TMQH, Turesky modified Quigley–Hein plaque index; SE, standard error.
Table 4. Study characteristics and results. OR, oscillating–rotating; RMNPI, Rustogi Modified Naval Plaque Index; MGI, Modified Gingival Index; GBI, Gingival Bleeding Index; TMQH, Turesky modified Quigley–Hein plaque index; SE, standard error.
Authors, YearCountry/Setting/PopulationStudy Design (All Randomised Controlled Trials)InterventionComparatorOutcome Measure(s)Follow-Up DurationNumber of Participants (Dropouts in Brackets)Results for Interproximal Plaque
Sjögren, K., Lundberg, A., Birkhed, D., Dudgeon, D. and Johnson, M., 2004 [21]Sweden,
adults
Four-period crossover design Four regimens: 1. Manual brushing only with Jordan Active Tip toothbrush. 2. Manual brushing after flossing (as above, plus waxed floss from Johnson & Johnson). 3. Oral-B Ultra Plaque Remover (D9). 4. Sonicare Plus (now renamed Sonicare Advance) See interventionWet weight inter proximal plaque (mg) 8 days for each regimen 47 (unclear if any dropouts) Wet weight inter proximal plaque (mg) mean (SD)
Day 1: OR brush = 14.8 (11.5); sonic brush = 5.7 (2.5)
Day 8: OR brush = 17.2 (10.8); sonic brush = 6.0 (2.6)
Sharma, N., Goyal, C., Qaqish, J., Cugini, M., Thompson, M. and Warren, P., 2005 [22]Clinic, adultsRandomised, examiner-blind, three-arm, single-use crossover design OR/pulsating brush: Oral-B ProfessionalCare 7000, 340 Hz pulsation, 73 Hz oscillation OR/pulsating brush: Oral-B 3D Excel (3DE), 340 Hz pulsation, 63 Hz oscillation; sonic brush: Sonicare Advance toothbrush, 260 Hz, Easy Start feature deactivated, used with “normal” settingsRMNPI, reduction safety Minimum 8 days 79 (0)Post-brushing RMNPI scores and plaque reduction (mean (SD))
PC7000: Post-brushing RMNPI = 0.33, (0.12) plaque reduction = 0.68 (0.12); 3DE: Post-brushing RMNPI = 0.32 (0.11), plaque reduction = 0.68 (0.11); sonic brush: Post-brushing RMNPI = 0.41 (0.15), plaque reduction = 0.59 (0.15)
Strate, J., Cugini, M., Warren, P.R., Qaqish, J., Galustians, H. and Sharma, N., 2005 [23]Research facility, adults Randomised, examiner-blind, crossover design with two study visits OR toothbrush: Oral-B Professional Care Series, oscillation angle of 45 degrees, 73 Hz, with a pulsation frequency of 340 Hz Sonic toothbrush: Sonicare Elite (260 Hz, Easy Start feature deactivated) RMNPI, reduction safety Minimum 4 days 61 (0) Post-brushing RMNPI scores and mean plaque reduction (mean (SD)): OR toothbrush: Post-brushing = 0.034 (0.06), mean plaque reduction = 0.966 (0.06), % plaque removal = 96.6%;
sonic toothbrush: Post-brushing = 0.271 (0.18), mean plaque reduction = 0.729 (0.18), % plaque removal = 72.9% p = 0.0001
Biesbrock, A., Bartizek, R., Walters, P., Warren, P., Cugini, M., Goyal, C. and Qaqish, J., 2007 [24]AdultsTwo studies used a randomised, examiner-blind, two-treatment, crossover design Oscillating–rotating brush: Oral-B Triumph. This power brush has a round brush head and three-dimensional motion (rotation–oscillation plus pulsation). It operates at 8800 oscillations/40,000 pulsations per minute. The toothbrush was fitted with a FlossAction brush head. Sonic brush: Sonicare Elite 7300. This power brush has a conventionally shaped brush head and side-to-side motion. It operates at a frequency of 260 Hz. The Easy Start feature was deactivated prior to use, and the brush was used with the normal settings. This toothbrush can be used with either the standard, full-size brush head or a newly designed compact brush head intended for smaller mouths and precision cleaning. Whole-mouth plaque reduction (RMNPI)
Marginal and interproximal plaque reduction (RMNPI)
Study 1: 2 weeks
Study 2: Approximately 3 weeks
Study 1: 50 (0) Study 2: 49 (1) Reduction in RMNPI score (mean ± SD)
Study 1:
OR brush = 0.8275 ± 0.07; sonic brush = 0.6323 ± 0.12 (OR brush 21% greater reduction than sonic brush, p < 0.001)
Study 2:
OR brush = 0.758 ± 0.007; sonic brush = 0.662 ± 0.007 (OR brush 14.6% greater reduction than sonic brush, p < 0.001)
Biesbrock, Walters, Bartizek, Goyal and Qaqish, 2008 [25]Canada, adultsRandomised, examiner-blind, two-treatment, four-period, four-sequence crossover AABB, ABBA, BBAA, BAAB treatment sequences Oscillating–rotating brush Oral-B Triumph with FlossAction brush head with MicroPulse bristles (8800 oscillations/40,000 pulsations per minute) Sonic brush Sonicare FlexCare with ProResults brush head (260 Hz) Full-mouth plaque score, RMNPI post-brushing 12 days approximately 48 (3) Post-brushing plaque reduction adjusted mean ± SE:
OR toothbrush: 0.884 ± 0.013
Sonic toothbrush: 0.724 ± 0.013
Putt, Milleman, Jenkins, Schmitt, Master and State, 2008 [26]Independent clinical research organisation, Fort Wayne, Indiana, USA, adultsSingle-use, examiner-masked, randomised, crossover Sonic toothbrush: Sonicare FlexCare with ProResults radial brush head Oscillating–rotating toothbrush: Oral-B Triumph Professional Care 9000 with FlossAction brush head % reduction in full-mouth plaque score, Turesky modified Quigley–Hein plaque index (TMQH), plaque scores by site safety 2 weeks approximately 94 (1) % reduction in TMQH for interproximal sites (mean ± SD): sonic toothbrush: 33.74 ± 14.79; oscillating–rotating toothbrush: 24.21 ± 14.27
Klukowska, M., Grender, J., Goyal, C., Mandl, C. and Biesbrock, A., 2012 [27]AdultsOpen-label, examiner-blind, two-treatment, parallel-group, randomised study OR toothbrush: Oral-B Triumph with SmartGuide (aka Oral-B Professional Care SmartSeries 5000 in US) with EB25 FlossAction brush head Sonic toothbrush: Philips Sonicare DiamondClean with DiamondClean standard brush head MGI, GBI, RMNPI safety, consumer perception assessment 12 weeks130 (0)Adjusted mean reduction from baseline (SE), % reduction for interproximal RMNPI
Week 6:
OR toothbrush = 0.231 (0.0154), 23.3%
Sonic toothbrush = 0.188 (0.0154), 19.1%
Difference between brushes = 22.9%, p = 0.048
Week 12:
OR toothbrush = 0.387 (0.0156), 31.9%
Sonic toothbrush = 0.311 (0.0156), 31.6% Difference between brushes = 24.4%, p = 0.001
Klukowska, Grender, Conde and Goyal, 2013 [28]USA? (unclear), adultsRandomised controlled, two-treatment, parallel-group, examiner-blind Oscillating–rotating brush Oral-B Triumph with FlossAction AB25 brush head Sonic brush Colgate ProClinical A1500 with Triple Clean brush head MGI, whole-mouth RMNPI 12 weeks130 (4)Adjusted mean reduction from baseline (SE)
Week 4:
OR toothbrush: 0.275 (0.014); sonic toothbrush: 0.198 (0.014)
Week 12:
OR toothbrush: 0.407 (0.017); sonic toothbrush: 0.322 (0.016)
Klukowska, M., Grender, J., Conde, E., Ccahuana Vasquez, R. and Goyal, C., 2014 [29]Research centre, adultsRandomised, two-treatment, examiner-blind, parallel-group design OR toothbrush: Oral-B Triumph with SmartGuide with FlossAction brush head, D34/EB25 Sonic toothbrush: Sonicare FlexCare Platinum with InterCare standard brush head MBI, GBI, number of bleeding sites, RMNPI safety 12 weeks130 (3)Interproximal RMNPI reduction (adjusted mean reduction (SE), % change
6 weeks:
OR brush = 0.205 (0.0142), 20.7%; sonic brush = 0.181 (0.0141), 18.3%; 13.3% difference between brushes, p = 0.231
12 weeks:
OR brush = 0.344 (0.0155), 34.7%; sonic brush = 0.309 (0.0153); 31.2% 11.3% difference between brushes, p = 0.112
Klukowska, M., Grender, J., Conde, E., Goyal, C., Qaqish, J. and Schneider, M., 2014 [30]UnclearRandomised, examiner-blind, two-treatment, parallel-group study OR toothbrush: Oral-B Triumph with SmartGuide with Oral-B CrossAction brush head D34.EB50 Sonic toothbrush: Sonicare DiamondClean with standard brush head MGI, GBI, number of bleeding sites, RMNPI safety, user experience 6 weeks130 (2)Six-week adjusted mean reduction from baseline RMNPI interproximal plaque (SE), % change
OR toothbrush = 0.280 (0.0130), 28.4%;
sonic toothbrush = 0.210 (0.0130), 21.3%; 33.3% difference between brushes, p < 0.001
Ccahuana-Vasquez, R., Conde, E., Grender, J., Cunningham, P., Qaqish, J. and Goyal, C., 2015 [31]UnclearRandomised, examiner-blind, two-treatment, parallel-group Oscillating–rotating brush: Oral-B Professional Care 1000 [D16u] with Oral-B CrossAction brush head [EB50] Sonic brush: Sonicare DiamondClean with standard DiamondClean brush head RMNPI, MGI, GBI 8 weeks ± 2 days 150 (2)Adjusted mean reduction (SE), % change (between baseline and week 8):
OR toothbrush = 0.349 (0.0140), 34.9%;
sonic toothbrush = 0.270 (0.0142), 27%; 29.3% difference between groups, p < 0.001
Lv, Guo and Ling, 2018 [32]Guangzhou city, China Randomised, examiner-blind, parallel-group, with two-treatment sub-trials High-frequency sonic toothbrush: GEVILAN GET011 with GEH011 brush head (683 Hz) (Group A) Oscillating–rotating toothbrush: Oral-B PRO 700 with Cross Action brush head (8800 oscillations and 20,000 pulsations per minute) (Group B)
Sonic brush: Philips Sonicare Healthy White HX6712 with ProResults brush head (517 Hz) (Group C) MGI
RMNPI, MGI, GBI safety 6 months120 (1)Adjusted mean reduction from baseline (SE), % change
Month 3:
A: 0.114 (0.017), 11.5%
B: 0.082 (0.017), 8.3%
C: 0.108 (0.016), 11.1%
Month 6:
A: 0.169 (0.013), 17.1%
B: 0.131 (0.013), 13.3%
C: 0.219 (0.019), 22.4%
Adam, Goyal, Qaqish and Grender, 2020 [33]Canada (unclear) Randomised, open-label, parallel-group, examiner-blind Oscillating–rotating brush Oral-B iO and Ultimate Clean brush head M7/OC15 Sonic brush Philips Sonicare DiamondClean Smart Sonic and Premium Plaque control brush head HX9903/11 Modified Gingival Index, Gingival Bleeding Index, RMNPI safety after using the assigned toothbrush for the study period 8 weeks90 (0)Adjusted mean (SE) change from baseline:
Sonic: 0.324 (0.0234)
Oscillating–rotating: 0.458 (0.0234)
Goyal, C., Adam, R., Timm, H., Grender, J. and Qaqish, J., 2021 [34]Germany/research centreSingle-centre, examiner-blind, two-treatment, open-label, parallel-group, randomised study OR brush: Oral-B iO with Ultimate Clean brush head (M7/OR015) in Daily Clean mode Sonic brush: Sonicare DiamondClean with Premium Plaque Control brush head (HX9903/11) Clean mode with an intensity level of 3 (high) MGI, GBI, RMNPI 6 months110 (0)Adjusted mean (SE) change from baseline interproximal RMNPI
Day 1—single brushing:
OR brush = 0.891 (0.0101);
sonic brush = 0.794 (0.0101); 12% difference between groups, p < 0.001. Week 1—OR brush = 0.116 (0.0126);
sonic brush = 0.063 (0.0126); 84.5% difference between groups, p = 0.003. Week 24—OR brush = 0.467 (0.0247);
sonic brush = 0.371 (0.0247); 25.8% difference between groups, p = 0.007
Table 5. Risk of bias based on Cochrane Risk of Bias 2 [37] chart (studies [20,21,22,23,24,25,26,27,28,29,30,31,32,33]).
Table 5. Risk of bias based on Cochrane Risk of Bias 2 [37] chart (studies [20,21,22,23,24,25,26,27,28,29,30,31,32,33]).
StudyRandomisationAssignment to InterventionAdherence to InterventionMissing Outcome DataMeasurement of the OutcomeSelection of the Reported ResultOverall Risk of Bias
Sjögren et al. (2004) [21] Use of own toothbrush for washout periods, unclear which toothbrush this was
Sharma et al. (2005) [22] Used normal toothbrush during washout period
Strate et al. (2005) [23] Use of own toothbrush for washout periods, unclear which toothbrush this was
Biesbrock et al. (2007) [24]
Biesbrock et al. (2008) [25] Use of own toothbrush for washout periods, unclear which toothbrush this was
Putt et al. (2008) [26]
Klukowska et al. (2012) [27] OR group used SmartGuide
Klukowska et al. (2013) [28]
Klukowska et al. (2014a) [29] OR group used SmartGuide
Klukowska et al. (2014b) [30] OR group used SmartGuide
Ccahuana-Vasquez et al. (2015) [31] Two dropouts, both from comparator group; study underpowered?
Lv et al. (2018) [32]
Adam et al. (2020) [33]
Goyal et al. (2021) [34] App use differed between groups
Key: Green = low-risk; orange = some concerns; red = high-risk.
Table 6. Time-wise comparison of OR and sonic toothbrushes. OR, oscillating–rotating; RMNPI, Rustogi Modified Naval Plaque Index; TMQH, Turesky modified Quigley–Hein plaque index.
Table 6. Time-wise comparison of OR and sonic toothbrushes. OR, oscillating–rotating; RMNPI, Rustogi Modified Naval Plaque Index; TMQH, Turesky modified Quigley–Hein plaque index.
Follow-Up TimeInterproximal Plaque ReductionReference
1 day9.1 mg less remaining wet weight interproximal plaque for sonic toothbrush than OR toothbrushSjögren et al. (2004) [21]
0.097 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushGoyal et al. (2021) [34]
4 days0.237 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushStrate et al. (2005) [23]
1 week0.053 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushGoyal et al. (2021) [34]
8 days11.2 mg less remaining wet weight interproximal plaque for sonic toothbrush than OR toothbrushSjögren et al. (2004) [21]
0.09 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrush (at up to 8 days)Sharma et al. (2005) [22]
12 days0.16 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushBiesbrock et al. (2008) [25]
2 weeks0.1952 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushBiesbrock et al. (2007) [24]
9.52% greater TMQH reduction for sonic toothbrush than OR toothbrushPutt et al. (2008) [26]
3 weeks0.096 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushBiesbrock et al. (2007) [24]
4 weeks0.77 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2013) [28]
6 weeks0.043 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2012) [27]
0.024 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2014a) [29]
0.07 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2014b) [30]
8 weeks0.079 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushCcahuana-Vasquez et al. (2015) [31]
0.134 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushAdam et al. (2020) [33]
12 weeks/3 months0.076 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2012) [27]
0.085 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2013) [28]
0.035 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushKlukowska et al. (2014a) [29]
Between 0.026 and 0.032 greater interproximal RMNPI reduction for two different sonic toothbrushes than OR toothbrushLv, Guo and Ling (2018) [32]
6 monthsBetween 0.026 and 0.032 greater interproximal RMNPI reduction for two different sonic toothbrushes than OR toothbrushLv, Guo and Ling (2018) [32]
0.096 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushGoyal et al. (2021) [34]
Table 7. Time-wise meta-analyses of interproximal plaque reduction, comparison of OR and sonic toothbrushes. OR, oscillating–rotating.
Table 7. Time-wise meta-analyses of interproximal plaque reduction, comparison of OR and sonic toothbrushes. OR, oscillating–rotating.
Follow-Up TimeInterproximal Plaque Reductionp Value; Heterogeneity
6 weeks0.005 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushp = 0.0008; I2 = 32%
8 weeks0.09 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushp < 0.00001; I2 = 51%
12 weeks0.04 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushp = 0.0001; I2 = 79%
6 months0.03 greater interproximal RMNPI reduction for OR toothbrush than sonic toothbrushp = 0.09, I2 = 95%
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Lewis, R.D.; Kanagasingam, S.; Cook, N.; Krysmann, M.; Taylor, K.; Pisani, F. The Effect of Different Electric Toothbrush Technologies on Interdental Plaque Removal: A Systematic Review with a Meta-Analysis. Healthcare 2024, 12, 1035. https://doi.org/10.3390/healthcare12101035

AMA Style

Lewis RD, Kanagasingam S, Cook N, Krysmann M, Taylor K, Pisani F. The Effect of Different Electric Toothbrush Technologies on Interdental Plaque Removal: A Systematic Review with a Meta-Analysis. Healthcare. 2024; 12(10):1035. https://doi.org/10.3390/healthcare12101035

Chicago/Turabian Style

Lewis, Robert David, Shalini Kanagasingam, Neil Cook, Marta Krysmann, Kathryn Taylor, and Flavio Pisani. 2024. "The Effect of Different Electric Toothbrush Technologies on Interdental Plaque Removal: A Systematic Review with a Meta-Analysis" Healthcare 12, no. 10: 1035. https://doi.org/10.3390/healthcare12101035