Assessing the Impact of IOS Scanning Accuracy on Additively Manufactured Occlusal Splints

by
Eduardo Anitua
1,2,*,†,
Asier Lazcano
2,3,†,
Asier Eguia
1,4 and
Mohammad Hamdan Alkhraisat
1,2,5
1
University Institute for Regenerative Medicine and Oral Implantology—UIRMI (UPV/EHU-Fundación Eduardo Anitua), 01007 Vitoria, Spain
2
BTI Biotechnology Institute, 01005 Vitoria, Spain
3
Biomedical Research, Universidad del País Vasco/Euskal Herriko Unibertsitaea (UPV/EHU), 48940 Leioa, Spain
4
Faculty of Medicine and Nursing, Universidad del País Vasco/Euskal Herriko Unibertsitaea (UPV/EHU), 48940 Leioa, Spain
5
Faculty of Dentistry, University of Jordan, Amman 11942, Jordan
*
Author to whom correspondence should be addressed.
These authors contributed equally to this work.
Dent. J. 2024, 12(10), 298; https://doi.org/10.3390/dj12100298
Submission received: 23 July 2024 / Revised: 6 September 2024 / Accepted: 18 September 2024 / Published: 24 September 2024
(This article belongs to the Special Issue Feature Papers in Digital Dentistry)

Abstract

:
Introduction: Digital workflow and intraoral scanners (IOSs) are used to clinically obtain data for a wide range of applications in restorative dentistry. The study aimed to compare two different IOSs with inexperienced users in the digital workflow of oral split manufacturing. Material and Methods: Anonymous stone models of upper and lower dentate patients were used. Both models were scanned with a desktop 3D scanner 3Shape D2000 to obtain the reference models (STLR). Ten inexperienced operators scanned each model three times with each IOS system (3Shape TRIOS 3 and Carestream CS 3800). Finally, 20 intraoral scanners were randomly chosen from the obtained dataset (10 per IOS system) to design and manufacture 20 nightguards. All the nightguards were scanned. Trueness and precision were calculated and compared between the two IOS systems. Results: All the mean errors both for trueness and precision were below 40 µm, more than acceptable for the design and manufacturing of intraoral devices such as nightguards. All the mean errors (except one) for trueness between the inner part of the nightguards and the upper control model were below 100 µm, less than a printed layer height. For inexperienced operators, both IOSs are suitable for a digital workflow of manufacturing occlusal splints.

1. Introduction

The occlusal splint is a removable intraoral appliance that clinicians use to treat a wide variety of conditions, including bruxism, temporomandibular joint disorders, and masticatory system disorders [1,2]. The occlusal splint, due to its increased surface area, decreases the stress on teeth and restorations, thus preventing mechanical-induced damage to these structures. There are different factors that influence the clinical efficacy of these types of appliances. The primary factor that determines the occlusal splint’s function is its design. Nowadays, there are different kinds and designs that have been tailored by dictating the degree of teeth coverage (partial or complete), elasticity (hard or soft), optical properties (transparency), and stability [1,2]. The material choice would additionally affect the polishability and cleansibility of the occlusal splint. Patient comfort and adherence to treatment are factors that need to be considered to ensure an effective treatment [3].
Additive manufacturing (AM) has been suggested and implemented to produce occlusal splints. This manufacturing method is based on a three-dimensional digital design of the occlusal splint (CAD) and is then printed layer by layer (CAM) [3]. The biggest advantage of AM is the reduction of the labor load by scaling up the production capacity and reproducibility. Furthermore, it allows for the incorporation of artificial intelligence in the design process to tailor it to clinical efficacy and safety. Salmi et al. have reported reduced chair-side time due to the use of the occlusal splint that was fabricated by AM [3].
The first step in the AM of the occlusal splint is the digitalization of the teeth, gingival tissue, and maxilla–mandible relation. Nowadays, this step is widely performed by extraoral scanning of cast models or the incorporation of intraoral scanners. Intraoral scanners are classified by their accuracy and ergonomics. The more accurate an intraoral scanner is, the better results we obtained in AM solutions. Accuracy consists of two parameters: trueness and precision. Trueness is how closely the data conform to reality (or the best approximation of reality), while precision is how closely the data conform to each other, in a continuous measurement of data [4]. Therefore, for high accuracy, a scanner must have high trueness values (decreasing systematic errors) and high precision values (decreasing random errors). The combination of both leads to a high accuracy (decreasing uncertainty) [5]. Although different studies have compared different types of intraoral scanners, there is a need to assess the impact of these differences on the production of occlusal splints [6]. Patient anatomy, CAD design, manufacturing technology, and the type of material are factors that may directly impact the result of the product.
The purpose of this study has been to analyse the behaviour of 3Shape (Copenhagen, Denmark) TRIOS 3, and Carestream CS 3800 (Atlanta, GA, USA) by inexperienced operators in a full arch dentate model that was going to be used for nightguard (intraoral devices) design and manufacturing. The null hypotheses were that no significant difference would be found between 3Shape TRIOS 3 and CS Carestream CS 3800 accuracy (trueness and precision), and that no significant deviation would be added in occlusal splint manufacturing and post-processing.

2. Materials and Methods

2.1. Reference Model and Scanning Procedure

Anonymous stone models of upper and lower dentate patients were used for the study. Both maxilla and mandible did not present any dental restoration and had ideal occlusal plane. It was a Class I dentate patient, resulting in models that are easy to scan due to their low complexity. Then, reference file was generated to be used as a control for the comparison with the intraoral scanners. Two types of intraoral scanners were used 3Shape TRIOS 3 and Carestream CS 3800. Reference model: The working casts were digitized to serve as references or control models in the study by means of a Blue LED Multi-line structured light (D2000; 3Shape) and four cameras of 5MP. This device projects a series of parallel patterns onto the scan target. When light patterns are projected onto the plaster model, the patterns become distorted. The four cameras capture these visualizations from different angles and process them to find the best possible point cloud reconstruction, obtaining the digitization of the plaster models as STL files named control or reference (STLR). The D2000 3Shape scanner had been previously calibrated following the manufacturer’s manual, reporting an accuracy of 5 µm (ISO 12836). The calibration process is done by inserting a calibration object without touching the top surface. Scanned data were reported to be sufficiently accurate for use as a reference model.
Intraoral scan (IOS) models: Ten unexperienced odontologists performed digital scans of the cast models (maxilla, mandible), three times each for each model with each IOS. The odontologists were identified by numbers (from 122 to 131) to maintain their anonymity. Everyone had at least 3 years of experience in the industry but less than a month in the digital workflow. The number of times was named as follows: A = first-time scanning, B = second-time scanning, and C = third-time scanning (Figure 1). The IOS models were calibrated following the manufacturer’s calibration protocol by inserting the IOS inside a calibration tip. Each operator scanned following the manufacturer’s scanning protocols. Further, 3Shape recommendation is as follows: Start the scanner while it rests occlusal on the molar and wait three-to-five clicks. Move towards the centrals, capturing the occlusal surface. Continue slowly during the centrals and again continue along the occlusal surface until you reach the last molar. Turn buccal slowly by rotating the scanner 60–90 degrees at the last molar and complete the buccal swipe, taking care of areas where soft tissue may interfere with the scan. Go along the buccal side until the last molar on the opposite side is reached. Finally, roll to the palate side and complete the swipe (Figure 2a). Carestream recommendation is as follows: Start the scanner while it rests occlusal on the molar and move along the arch scanning the occlusal surface until you reach the last molar. Turn buccal slowly by rotating the scanner 60–90 degrees at the last molar and complete the buccal swipe, taking care of areas where soft tissue may interfere with the scan. Finally, turn to the palate side and complete the arch scanning (Figure 2b).

2.2. Data Analysis and Assessment of Trueness and Precision

The STLR files were used as a reference digitized model to compare the deviation with the 120 files obtained by the scanning of the operators. In the study, trueness was defined as the mean distance (mm) between the reference model and each scanned model, and it was represented in a colour spectrum. Precision was defined as the mean distance deviation between scan pairs of the same operator. A 3D mesh processing software (Geomagic Studio 12) was used for 3D mesh processing. All the obtained meshes were pre-processed by inspecting the obtained tessellated geometry, cleaning and defining the region of interest (ROI) for the deviation analysis (green line):
To maintain the same ROI for all the datasets, they were first translated into the same coordinate system, and then a spline was defined between the boundary of gum and teeth (Figure 3). All the scanners were between the range of 210,000–220,000 triangles.
Regarding trueness, 60 obtained meshes from the operators were translated to same coordinate system, applying a transformation matrix 4 × 4. This matrix was obtained by means of mesh-to-mesh point-based registrations. It consists of the re-orientation of two or more objects (point or polygon) that comprise a single object so that identical regions of different objects are made to overlap. N-point registration mode type has been used, where objects were aligned on the selection of three-to-nine points on the overlapping section of the objects.
Once the objects were overlapped and located at the same coordinate system, a fine registration was applied to reduce the deviation between the two meshes. This fine registration was made by applying an algorithm known as iterative closest-point (ICP) algorithm [7]. This algorithm minimizes the difference between two sets of points. Once both meshes have been overlapped at their maximum, the difference between two surfaces (control mesh and new measured mesh) was calculated by taking the greatest of all the distances from a point in one set to the closest point in the other set, known as Hausdorff distance. The Hausdorff distance is the maximum distance from any point on either surface to the nearest point on the other. Hausdorff distance, named after the German mathematician Felix Hausdorff, measures the dissimilarity between two sets. Specifically, it measures how far two subsets of a metric space are from each other. This concept was applied in 3D meshes, which are collections of vertices and their connections (usually edges and faces) in three-dimensional space. The concept remains the same as in the general definition but is adapted to work with these mesh structures. With two 3D meshes M1 and M2, the Hausdorff distance [8,9] between them can be defined as follows:
H a u s d o r f f M 1 , M 2 = m a x ( max v 1   M 1 min v 2     M 2 v 1 v 2 ,     m a x v 2   M 2   m i n v 1   M 1 v 2 v 1   )
For each vertex v1 in M1, the minimum distance to any vertex v2 in M2 was found, and the maximum of all these minimum distances was taken, where || v1v2 || denotes the Euclidean distance between vertices v1 and v2. Similarly, for each vertex v 2   in M 2 , the minimum distance to any vertex v1 in M 1 was found, and the maximum of all these minimum distances was taken, where || v2v1 || denotes the Euclidean distance between vertices v2 and v1. The overall Hausdorff distance between the two 3D meshes is then the maximum of the two computed values.
Once the deviation values were obtained, they were plotted as a colour-coded surface mapping of the dental scan, indicating the positive and negative dimensional differences. A positive value in dimensional difference was denoted by a red–yellow spectrum on the color-coded surface map. If there was no significant dimensional difference, it was defined by green, and a negative value in dimensional difference was indicated by a blue spectrum on the colour-coded surface map. For this representation, the vertices of the mesh were labelled with the corresponding RGB colours [10,11,12,13,14]. A histogram was plotted per each analysis to see the distribution of all the cloud points (Figure 4).
Regarding precision, each 3-scan data obtained from each operator were compared in pairs, obtaining 3 sets of analysis results. The 3D deviation analysis was performed using the Hausdorff distance, which was calculated between two surfaces: AB, AC, and BC. The same process as trueness was applied, but instead of comparing the obtained STL to the control mesh, the obtained meshes were compared between them. The normality of the data distribution for the quantitative variables was tested by Shapiro–Wilk test. The standard error and the 95% confidence interval were calculated. The comparison between the study groups was performed with two-way ANOVA test. The statistical analysis was performed in IBM SPSS Statistics (Armonk, NY, USA). The threshold of statistical significance was at p-value < 0.05.

2.3. Manufacturing and Post-Processing of Nightguards for Trueness Evaluation

Two hundred forty analyses were made in total with the 120 obtained digital models: 120 for trueness and 120 for precision. From the 120 models, 20 random meshes were chosen for design and manufacturing of their corresponding nightguards. The randomicity consisted of choosing one of the three scanners made from each operator and each system, naming them with the operator number and a letter T for TRIOS 3 and C for CS 3800. These nightguards were designed by the software BTI 3D INTRAORAL DEVICES version 2.4.10, developed internally for design and manufacturing of nightguards and intraoral device APNia.
All the nightguards were manufactured by DLS technology (with a 75 µm pixel size) and biocompatible class IIa resin material (printed at 100 µm layer height), and they were labelled by their mesh stl name. The manufactured parts were post-processed under inner conditions (N2) following the material’s manufacturer protocol to avoid the creation of an oxygen inhibition layer. The inner part of the 20 splints was scanned using the reference scanner 3Shape D2000. Moreover, the obtained digitized splint models were compared to the control model STLR to see the trueness between the inner part of the nightguard and the control model. To make this possible, the 20 scanned parts were translated to the same coordinate system, and an ROI was applied to all of them (the inner part of the splint). The ROI was defined by a splint, and it was extracted from the STL to obtain a new object to work on it. The obtained 20 ROIs were translated to the 3D coordinate system of the control mesh, and then ICP algorithm was applied for a best fit between control mesh and the inner part of the splint. Finally, the Haussdorf distance was applied to calculate the deviation between objects. These results showed deviation in the manufacturing process, and they were supposed to have higher deviation than the scanned models due to the deviations that may occur in the manufacturing and post-processing processes.

3. Results

Two hundred forty analyses were conducted for trueness and precision. In all groups, the standard deviation was higher than the mean errors from the control mesh. All of the mean errors, both for trueness and precision, were below 40 µm, indicating that the average distance was anywhere between 210,000–220,000 triangles of each scanner, and the control model and each scanner, as well as the following ones (A, B, and C) done by each operator, was below 40 microns, more than acceptable for the design and manufacturing of intraoral devices such as nightguards [13].
Regarding the trueness of the upper model scanned by TRIOS 3 (Table 1), mean values mostly ranged between −0.05 mm and 0.05 mm. Slight deviations were observed, but most data points were concentrated around 0 mm. Standard deviations were relatively consistent across times A, B, and C, with a general range from 0 to 0.1 mm. Maximum positive deviations showed some outliers, but most values were clustered below 1.5 mm. Maximum negative deviations had a few outliers extending below −1 mm, with most values above −1.5 mm.
Regarding the trueness of the lower model scanned by TRIOS 3 (Table 2), mean values were tightly grouped around 0 mm, with occasional slight positive and negative deviations. Standard deviations ranged similarly to the upper model, generally between 0 and 0.1 mm. Maximum positive deviations showed a similar pattern with most values below 1 mm. Maximum negative deviations had more outliers, with values occasionally dropping below −1 mm.
With respect to the trueness of the upper model scanned by CS 3800 (Table 3), negative mean deviations were more common, with most values within −0.1 to 0.05 mm. Standard deviations were slightly higher, reaching up to 0.15 mm. Maximum positive deviations were generally below 1.5 mm but showed some significant outliers. Maximum negative deviations extended significantly, with some values going below −2 mm.
Concerning the trueness of the lower model scanned by CS 3800 (Table 4), mean deviations showed more variation, both positive and negative, with a similar range as the upper model. Standard deviations were higher on average, with more outliers. Maximum positive deviations reached higher values with several significant outliers. Maximum negative deviations showed a broader range with more extreme outliers.
Relating to the precision of the upper model scanned by TRIOS 3 (Table 5), mean deviations were relatively small, mostly within ±0.05 mm. Standard deviations were consistent and low, mostly under 0.05 mm. Maximum positive deviations were mostly under 0.5 mm. Maximum negative deviations were generally under −1 mm, with some outliers.
About the precision of the lower model scanned by TRIOS 3 (Table 6), mean deviations were small and centered around 0 mm. Standard deviations were consistently low. Maximum positive deviations remained low, under 0.5 mm. Maximum negative deviations showed a few outliers but were generally under −1 mm.
With respect to the precision of the upper model scanned by CS 3800 (Table 7), mean deviations were slightly larger, with more variation. Standard deviations were higher and showed more outliers. Maximum positive deviations had higher values, with significant outliers. Maximum negative deviations showed more variation, with values extending further.
Regarding the precision of the lower model scanned by CS 3800 (Table 8), mean deviations were small and around 0 mm but showed more negative values. Standard deviations were low but had more variation. Maximum positive deviations showed higher values and more outliers. Maximum negative deviations had significant outliers and broader ranges.
All results showed that the operators’ experience had little impact on the trueness and precision of the study. In the trueness of TRIOS 3 (Table 9), mean deviations across operators were relatively small and centered around 0 mm, indicating that most operators were able to achieve high trueness. Standard deviations were consistently low across operators, suggesting good repeatability and precision in measurements. Max deviations showed some outliers, but these were not significantly different across operators, indicating that extreme values were likely due to measurement challenges rather than operator inconsistency. In the trueness of Carestream CS 3800 (Table 10), mean deviations varied more among operators, indicating that some operators might have had difficulties achieving high trueness consistently. Standard deviations were higher on average and showed more variability among operators, suggesting that some operators were less consistent in their measurements. Max deviations had more significant outliers, which could be linked to specific operators struggling with certain measurements.
Regarding the precision of TRIOS 3 (Table 11), mean deviations were small and consistent across operators, indicating high trueness between them. Standard deviations were low and consistent, suggesting high precision across operators. Max deviations showed fewer outliers, indicating fewer instances of significant measurement errors among operators. With respect to the precision of CS3800 (Table 12), mean deviations varied more significantly among operators, suggesting differences in measurement techniques or challenges with the equipment. Standard deviations were higher and more variable among operators, indicating inconsistencies in measurements. Max deviations had more significant outliers, pointing to occasional large errors that could be operator-dependent.
Regarding the precision of the IOS, in the mean deviation of lower arch models, there were statistically significant differences, as the p-value was 0.032. However, in the mean deviation of upper arch models, there were no statistically significant differences, as the p-value was 0.921. Regarding trueness, in the mean deviation of lower arch models, there were statistically significant differences as the p-value was 0.000 and in the mean of upper models.
The manufacturing and post-processing processes of polymers also had a great impact on the deviation of the final product in the digital workflow. Standard deviations for trueness between the inner part of the nightguards and the upper control model were higher than the mean errors from the control mesh. All the mean errors for trueness were below 100 microns, except one that was 244 microns. This last one could be affected by the holder of the scanning parts. All the nightguards were placed in a holder inside the scanner, and due to the low Young modulus of the polymer (1063 MPa at environmental temperature), it could suffer some deformation when scanning. However, due to the low stiffness of the material, the deviation suffered in the manufacturing and post-processing processes could be corrected when placing the nightguard in a patient’s mouth. This was a good point because it was possible to quantify all of the possible errors in the digital workflow of the design and manufacturing of nightguards with DLS printing technology and the biocompatible Class ii material that was used in the study (Table 13).
The mean deviations for different splints were relatively small, with values centered around 0 mm. This indicated good trueness overall, as the deviations from the true value are minimal. Also, 122C and 123T splints showed slightly larger mean deviations compared to others, suggesting some variability in the measurements. The other splints generally showed minimal mean deviations, indicating high trueness. The standard deviations were low across all splints, indicating consistent measurements with minimal spread around the mean. Further, 122C and 123T showed slightly higher standard deviations, indicating more variability in the measurements for these splints. The other splints consistently obtained low standard deviations, indicating precise and repeatable measurements. Positive maximum deviations were generally below 2 mm, with a few outliers extending beyond this range. Also, 122C showed the highest positive maximum deviation, indicating occasional significant positive errors. The most positive deviations of the other splints were clustered below 1.5 mm, indicating fewer significant positive errors. Negative maximum deviations showed more variation, with some values extending below −2 mm. Also, 122C again showed the highest negative maximum deviation, indicating occasional significant negative errors. The other negative deviations of splints were generally less extreme, clustered above −1.5 mm (Figure A1).

4. Discussion

4.1. Impact of Digital Workflow and Scanner Accuracy in Clinical Dentistry

Digital workflow is growing exponentially in the dental sector. Although it is supposed to be a more precise technology than conventional workflow, we must consider the deviations that may occur along the workflow. In clinical dentistry, intraoral scanners (IOSs) play a vital role in obtaining digital data of the patient’s mouth for a wide range of applications such as diagnosis or restorative dentistry. Ensuring the accuracy of intraoral scanning is crucial and is determined by two key parameters: precision and trueness. Precision shows how similar repeated measurements are, in other words, the reproducibility of the scanning results of the same anatomical structure, while trueness shows how it is a similar measurement to the value of the measured anatomy. Both 3Shape TRIOS 3 and Carestream CS 3800 are suitable for intraoral devices such as nightguards. The manufacturing and post-processing methods add deviation in the digital workflow. However, due to the low stiffness of the polymer, these small deviations are not critical because it adapts perfectly when the application is placed in the mouth. The flexibility that the polymer offers can lead to the correction of these small deviations.
Other previous studies evaluated the influence of the ambient light scanning conditions on the accuracy of different IOS that have not been considered in this study [15]. The results of this study show the accuracy of just one patient model that did not present any dental restorations, so all the teeth were perfectly aligned among the arch. The number of analyses demonstrates the effectiveness of this study because 240 deviation analyses were performed, more than what had been analysed in other previous studies.

4.2. Comparison of Operating Technologies: Precision, Trueness, and Software Performance

In reference to the operating technologies across all models and metrics, there was a consistent trend of small mean deviations centered around 0 mm, with occasional outliers. The standard deviations were relatively low, indicating precise measurements, though a few models showed higher variation. Both positive and negative maximum deviations showed more significant outliers, indicating occasional larger discrepancies. In general, the comparative performance of the operating technologies was as follows: TRIOS 3 models generally had lower deviations and fewer outliers compared to the CS 3800 models, indicating higher precision and trueness. The results of the mean deviation of models were statistically significantly different in general, as the obtained p-values were as follows: 0.032/0.921 for precision and 0.000/0.000 for trueness. So, the differences were considered significant. Consequently, the first null hypothesis was rejected. In general, 3Shape TRIOS 3 results were slightly better than CS Carestream 3800, as the precision obtained between the different scanning times is very high. Regarding the tessellation of the obtained meshes, the ones obtained by TRIOS 3 were more uniform and equitable compared to the ones obtained by CS 3800. Also, small artifacts were only found in the meshes obtained by CS 3800. This could indicate that 3Shape has a greater capacity to process the data obtained from the point cloud, suggesting that they are superior in software development.

4.3. Operator Consistency

Regarding operator consistency in the study, operators who consistently achieve low mean deviations and low standard deviations contribute positively to the trueness of the measurements. Variability among operators suggests that training, technique, or experience may play a role. The equipment sensitivity also plays a role. Carestream showed more variability among operators, suggesting that the equipment might be more sensitive to operator technique or that certain operators find it more challenging to use consistently. Moreover, improving trueness may involve standardizing measurement techniques and providing additional training for operators who show higher variability in their measurements. The importance of following the manufacturer’s scanning protocol is huge, as a variety of scanning methods may vary the accuracy of results. By focusing on these aspects, it is possible to improve the overall trueness of the measurements by addressing operator-related variability. However, an operator can feel confident even without experience, as their data acquisition is valid for this type of application, as has been demonstrated.

4.4. Impact of Manufacturing and Post-Processing

The manufacturing and post-processing also affect the deviation of the final product. The results of this study demonstrated that the mean values obtained with IOS are lower than the ones that are produced during the manufacturing process. The accuracy values reported are questionable if the manufacturing technology [15], post-processing method, or the material is changed [16,17,18,19]. However, all the mean errors for trueness were below 100 microns (less than the manufactured layer height). So, the second null hypothesis was retained due to the insignificant deviation in the manufacturing and post-processing steps [20].

4.5. Study Limitations and Future Research

The results may vary depending on the tessellation geometry of the reference and test object. If decimation or refining algorithms are applied to the analysed meshes, the Haussdorf distances are modified so they can be manipulated to obtain better or worse results. In this study, the tessellation was not modified. It was used in the “raw” mesh exported from the IOS directly to avoid these possible modifications.
The limitations of the study are as follows: It was an in vitro study conducted with only one patient without prosthetic restorations, with an ideal arch and occlusal planes. The version of the intraoral scanners is not the latest in the market.
It would be interesting to analyse environmental variables such as ambient light and other models with prosthetic restorations or deteriorated teeth. Also, the study evaluates two different intraoral scanners that have different hardware components. However, it would be interesting to analyse the different optical behaviours [21].

5. Conclusions

  • Both 3Shape TRIOS 3 and Carestream CS 3800 are suitable for creating intraoral devices like nightguards.
  • Operators do not need prior experience with any IOS system to achieve successful results.
  • Manufacturing and post-processing methods introduce deviations in the digital workflow.
  • These small deviations are not critical due to the low stiffness of the polymer used.
  • The application can be deformed in the mouth to meet the patient’s anatomical requirements.

Author Contributions

All authors have made substantial contributions to the conception or design of the work or the acquisition, analysis, or interpretation of data for the work; AND drafting the work or revising it critically for important intellectual content; AND have given final approval of the version to be published; AND agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The datasets used and analyzed during the current study are available from the corresponding author upon reasonable request.

Acknowledgments

I would like to express my heartfelt gratitude to all those who have contributed to this study. Special thanks to the operators, whose tireless efforts in repeatedly scanning the models made this labor-intensive work possible. I am especially grateful to Mohammad Hamdan Alkhraisat from BTI ImasD for his insightful comments and suggestions, which were instrumental in shaping the final version of this paper. Additionally, I would like to thank BTI Biotechnology Institute for providing me the opportunity to research new technologies in Industry 4.0.

Conflicts of Interest

E.A. is the Scientific Director of BTI Biotechnology Institute, a dental implant company that investigates the fields of oral implantology and PRGF-Endoret technology. A.E. is a researcher at the Eduardo Anitua Foundation. A.L. and M.H.A. are researchers at BTI Biotechnology Institute.

Appendix A

Figure A1. Mesh-to-mesh point-based registration, ICP algorithm, and colour-coded surface mapping of the manufactured splints and the STLR model.
Figure A1. Mesh-to-mesh point-based registration, ICP algorithm, and colour-coded surface mapping of the manufactured splints and the STLR model.
Dentistry 12 00298 g0a1

References

  1. Nassif, M.; Haddad, C.; Habli, L.; Zoghby, A. Materials and manufacturing techniques for occlusal splints: A literature review. J. Oral Rehabil. 2023, 50, 1348–1354. [Google Scholar] [CrossRef] [PubMed]
  2. Albagieh, H.; Alomran, I.; Binakresh, A.; Alhatarisha, N.; Almeteb, M.; Khalaf, Y.; Alqublan, A.; Alqahatany, M. Occlusal splints-types and effectiveness in temporomandibular disorder management. Saudi Dent. J. 2023, 35, 70–79. [Google Scholar] [CrossRef] [PubMed]
  3. Salmi, M.; Paloheimo, K.S.; Tuomi, J.; Ingman, T.; Mäkitie, A. A digital process for additive manufacturing of occlusal splints: A clinical pilot study. J. R. Soc. Interface 2013, 10, 20130203. [Google Scholar] [CrossRef] [PubMed]
  4. Vag, J.; Stevens, C.D.; Badahman, M.H.; Ludlow, M.; Sharp, M.; Brenes, C.; Renne, W. Trueness and precision of complete arch dentate digital models produced by intraoral and desktop scanners: An ex-vivo study. J. Dent. 2023, 139, 104764. [Google Scholar] [CrossRef] [PubMed]
  5. Osnes, C.A.; Wu, J.H.; Venezia, P.; Ferrari, M.; Keeling, A.J. Full arch precision of six intraoral scanners in vitro. J. Prosthodont. Res. 2020, 64, 6–11. [Google Scholar] [CrossRef] [PubMed]
  6. Reich, S.; Berndt, S.; Kühne, C.; Herstell, H. Accuracy of 3D-printed occlusal devices of different volumes using a digital light processing printer. Appl. Sci. 2022, 12, 1576. [Google Scholar] [CrossRef]
  7. Hu, Z. ICP algorithm for 3D surface registration. Highlights Sci. Eng. Technol. 2022, 24, 94–98. [Google Scholar] [CrossRef]
  8. Atallah, M.J. A linear time algorithm for the Hausdorff distance between convex polygons. Inf. Process. Lett. 1983, 17, 207–209. [Google Scholar] [CrossRef]
  9. Bartoň, M.; Hanniel, I.; Elber, G.; Kim, M.S. Precise Hausdorff distance computation between polygonal meshes. Comput. Aided Geom. Des. 2010, 27, 580–591. [Google Scholar] [CrossRef]
  10. Gao, H.; Liu, X.; Liu, M.; Yang, X.; Tan, J. Accuracy of three digital scanning methods for complete-arch tooth preparation: An in vitro comparison. J. Prosthet. Dent. 2022, 128, 1001–1008. [Google Scholar] [CrossRef] [PubMed]
  11. Aspert, N.; Santa-Cruz, D.; Ebrahimi, T. MESH: Measuring errors between surfaces using the Hausdorff distance. In Proceedings of the IEEE International Conference on Multimedia and Expo, Lausanne, Switzerland, 26–29 August 2002; pp. 705–708. [Google Scholar]
  12. Gilbert, E.; Johnson, D.; Keerthi, S. A fast procedure for computing the distance between complex objects in three-dimensional space. IEEE Trans. Robot. Autom. 1988, 4, 193–203. [Google Scholar] [CrossRef]
  13. Kernen, F.; Schlager, S.; Alvarez, V.S.; Mehrhof, J.; Vach, K.; Kohal, R.; Flügge, T. Accuracy of intraoral scans: An in vivo study of different scanning devices. J. Prosthet. Dent. 2022, 128, 1303–1309. [Google Scholar] [CrossRef] [PubMed]
  14. Revilla-León, M.; Subramanian, S.G.; Özcan, M.; Krishnamurthy, V.R. Clinical Study of the Influence of Ambient Light Scanning Conditions on the Accuracy (Trueness and Precision) of an Intraoral Scanner. J. Prosthodont. 2020, 29, 107–113. [Google Scholar] [CrossRef] [PubMed]
  15. Piedra-Cascón, W.; Krishnamurthy, V.R.; Att, W.; Revilla-León, M. 3D printing parameters, supporting structures, slicing, and post-processing procedures of vat-polymerization additive manufacturing technologies: A narrative review. J. Dent. 2021, 109, 103630. [Google Scholar] [CrossRef] [PubMed]
  16. Vasilescu, M.D. Influence of technological parameters on the structure and deformation of flat or round parts generated by DLP 3D printing. Nonconv. Technol. Rev. 2019, 23, 56. [Google Scholar]
  17. Mayer, J.; Reymus, M.; Mayinger, F.; Edelhoff, D.; Hickel, R.; Stawarczyk, B. Temporary 3D-printed fixed dental prosthesis materials: Impact of postprinting cleaning methods on degree of conversion and surface and mechanical properties. Int. J. Prosthodont. 2021, 34, 784. [Google Scholar] [CrossRef] [PubMed]
  18. Wada, J.; Wada, K.; Gibreel, M.; Wakabayashi, N.; Iwamoto, T.; Vallittu, P.K.; Lassila, L. Effect of nitrogen gas post-curing and printer type on the mechanical properties of 3D-printed hard occlusal splint material. Polymers 2022, 14, 3971. [Google Scholar] [CrossRef] [PubMed]
  19. Benfaida, S.; Hamza, M.; Benabdelaziz, A.; Jouhadi, E.M. The impact of post-processing procedures on the characteristics of the additively manufactured resins: A systematic review. Open Access Libr. J. 2024, 11, 1–16. [Google Scholar] [CrossRef]
  20. Johansson, C.; Dibes, J.; Rodriguez, L.E.L.; Papia, E. Accuracy of 3D printed polymers intended for models and surgical guides printed with two different 3D printers. Dent. Mater. J. 2021, 40, 339–347. [Google Scholar] [CrossRef] [PubMed]
  21. Pilecco, R.O.; Dapieve, K.S.; Baldi, A.; Valandro, L.F.; Scotti, N.; Pereira, G.K.R. Comparing the accuracy of distinct scanning systems and their impact on marginal/internal adaptation of tooth-supported indirect restorations: A scoping review. J. Mech. Behav. Biomed. Mater. 2023, 144, 105975. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Schematic representation of the study design.
Figure 1. Schematic representation of the study design.
Dentistry 12 00298 g001
Figure 2. (a) Scanning protocol for the maxillary complete-arch digital scans performed followed 3Shape recommendation; (b) scanning protocol for the maxillary right quadrant digital scans performed followed Carestream recommendations.
Figure 2. (a) Scanning protocol for the maxillary complete-arch digital scans performed followed 3Shape recommendation; (b) scanning protocol for the maxillary right quadrant digital scans performed followed Carestream recommendations.
Dentistry 12 00298 g002
Figure 3. Selected ROI (in green) defined by a spline in Geomagic Studio 12.
Figure 3. Selected ROI (in green) defined by a spline in Geomagic Studio 12.
Dentistry 12 00298 g003
Figure 4. Mesh-to-mesh point-based registration, ICP algorithm, and colour-coded surface mapping.
Figure 4. Mesh-to-mesh point-based registration, ICP algorithm, and colour-coded surface mapping.
Dentistry 12 00298 g004
Table 1. Statistical aggregates for trueness of the upper maxilla scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
Table 1. Statistical aggregates for trueness of the upper maxilla scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122A0.0030.0430.565−0.976
122B0.0180.0591.101−0.926
122C0.0070.0410.298−1.106
123A0.0110.0590.473−1.442
123B0.0170.0740.389−1.417
123C0.0060.0570.4−1.488
124A0.0080.0470.314−1.333
124B0.0090.0440.352−1
124C0.010.0420.312−1.056
125A0.0070.0460.393−1.116
125B0.0110.0650.451−1.525
125C0.0140.0660.362−1.395
126A0.0060.0390.45−0.901
126B0.0060.0470.314−1
126C0.0040.0440.288−1.094
127A0.0110.0390.595−0.728
127B−0.0030.0430.372−1.004
127C0.0010.0420.381−1.077
128A0.0080.0570.324−1.412
128B0.0090.0480.314−1.131
128C0.0090.0540.298−1.367
129A0.0080.0410.308−1.068
129B0.0020.0530.318−1.285
129C0.0120.0440.326−0.885
130A0.0020.0440.319−1.343
130B0.0050.0370.291−0.738
130C0.0040.0370.29−0.739
131A0.0120.0490.357−1.041
131B0.0070.0480.33−1.248
131C00.0540.335−0.98
Table 2. Statistical aggregates for trueness of the lower mandible scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
Table 2. Statistical aggregates for trueness of the lower mandible scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122A0.0120.0430.46−0.633
122B0.020.060.724−0.621
122C0.0140.0471.268−0.559
123A0.010.0550.872−1.207
123B0.010.0570.745−1.247
123C0.010.0570.8−1.376
124A0.010.0530.651−1.089
124B0.0140.0590.499−0.934
124C0.0130.0520.771−0.947
125A0.0090.0610.675−1.036
125B0.0120.0541.101−1.031
125C0.0150.0571.056−1.197
126A0.0150.061.059−1.432
126B0.0130.0690.542−0.705
126C0.010.0520.473−0.763
127A0.0110.0490.702−1.134
127B0.0090.0450.572−0.664
127C0.0070.0450.446−0.804
128A0.0090.0570.494−0.967
128B0.010.0551.093−1.136
128C0.0140.0561.047−0.947
129A0.0150.0570.724−1.072
129B0.0140.0521.098−1.223
129C0.0190.0560.624−0.918
130A0.010.0470.688−0.69
130B0.0090.0490.608−1.12
130C0.0140.051.154−0.912
131A0.0120.0520.735−1.778
131B0.020.0531.069−1.109
131C0.0160.0561.025−1.12
Table 3. Statistical aggregates for trueness of the upper maxilla scanned by Carestream CS 3800. All values are provided in millimeters (mm).
Table 3. Statistical aggregates for trueness of the upper maxilla scanned by Carestream CS 3800. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122A−0.0320.0750.412−0.857
122B0.0070.0560.419−1.069
122C−0.0160.0650.572−1.204
123A−0.0170.0681.12−0.912
123B−0.0080.060.41−0.824
123C−0.0190.0960.449−0.897
124A−0.010.070.486−1.154
124B−0.0140.0670.681−0.909
124C−0.0320.0990.475−0.954
125A−0.0060.0530.468−0.978
125B0.0080.0490.363−0.972
125C0.0060.050.391−0.938
126A−0.0020.0570.408−0.924
126B−0.0010.0630.401−0.944
126C−0.0080.0630.559−1.016
127A−0.020.0590.413−0.896
127B−0.020.0590.407−1.29
127C−0.0210.0660.48−1.035
128A−0.0020.0590.455−0.859
128B0.0040.0620.441−1.026
128C0.0080.0660.439−0.949
129A00.0510.382−1.051
129B0.0050.0570.483−1.062
129C0.0070.0850.431−1.232
130A0.0080.0620.783−0.962
130B−0.0040.0570.494−1.206
130C−0.0050.0550.475−0.902
131A0.0090.0530.328−0.784
131B0.0020.0560.479−0.939
131C0.0020.0530.427−1.097
Table 4. Statistical aggregates for trueness of the lower mandible scanned by Carestream CS 3800. All values are provided in millimeters (mm).
Table 4. Statistical aggregates for trueness of the lower mandible scanned by Carestream CS 3800. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122A−0.0080.0861.039−3.053
122B−0.0120.0830.913−1.027
122C−0.0130.0990.859−1.267
123A0.0110.090.755−0.97
123B−0.020.0891.2−0.907
123C0.0090.0880.688−0.757
124A−0.0130.0720.804−1.105
124B−0.0030.0611.168−1.219
124C−0.0050.0750.772−0.904
125A−0.0030.0681.036−1.097
125B−0.0020.0570.553−0.926
125C−0.0030.0750.851−0.995
126A0.0010.0650.847−0.756
126B00.0670.608−0.783
126C−0.0030.0710.404−1.277
127A−0.0080.0660.477−0.849
127B−0.0010.0720.899−1.159
127C−0.0180.0680.456−0.935
128A0.010.0570.49−0.965
128B0.010.0551.094−1.137
128C0.0130.0561.044−0.948
129A0.0150.0570.724−1.072
129B00.081.124−0.878
129C−0.0160.0751.122−1.179
130A−0.0020.0710.947−1.03
130B−0.0070.0740.895−0.992
130C−0.0010.0651.084−0.84
131A0.0080.0720.88−1.138
131B00.071.091−1.088
131C0.0050.0651.036−1.04
Table 5. Statistical aggregates for precision of the upper maxilla scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
Table 5. Statistical aggregates for precision of the upper maxilla scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122AB0.0140.0551.261−1.002
122AC0.0040.0290.177−0.564
122BC−0.0120.0640.348−1.522
123AB0.0040.0330.119−0.199
123AC−0.0050.0140.081−0.203
123BC−0.010.0380.29−0.178
124AB0.0010.0150.244−0.127
124AC0.0020.0170.484−0.217
124BC0.0010.0160.357−0.252
125AB0.0040.0360.129−1.327
125AC0.0050.0420.309−1.207
125BC0.0010.0160.349−0.434
126AB00.0320.463−1.06
126AC−0.0020.0280.134−1.097
126BC−0.0020.0220.12−0.524
127AB−0.0140.0520.434−1.314
127AC−0.0090.0470.376−1.41
127BC0.0040.0290.155−0.536
128AB00.0220.506−0.272
128AC0.0010.0190.282−0.217
128BC00.0190.181−0.486
129AB−0.0050.0360.147−1.403
129AC0.0030.0240.901−0.737
129BC0.0070.0261.049−0.189
130AB0.0020.0160.661−0.418
130AC0.0020.0210.613−0.415
130BC−0.0010.0230.62−0.348
131AB−0.0050.0310.353−0.238
131AC−0.0110.0410.368−0.236
131BC−0.0060.0260.431−0.4
Table 6. Statistical aggregates for precision of the lower mandible scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
Table 6. Statistical aggregates for precision of the lower mandible scanned by 3Shape TRIOS 3. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122AB0.0060.040.703−0.448
122AC0.0010.0250.827−0.445
122BC−0.0070.0440.738−1.215
123AB0.0010.0150.136−0.319
123AC0.0010.0140.185−0.816
123BC0.0010.0090.477−0.572
124AB0.0050.0480.189−0.281
124AC0.0040.0310.165−0.409
124BC−0.0010.0230.323−0.338
125AB0.0050.0250.229−0.642
125AC0.0060.050.234−0.569
125BC0.0010.0290.192−0.401
126AB−0.0030.0250.534−0.384
126AC−0.0040.0580.539−0.201
126BC00.0760.459−0.344
127AB−0.0020.0240.691−0.911
127AC−0.0040.0240.415−0.774
127BC−0.0030.0330.599−0.555
128AB00.0150.223−0.255
128AC0.0040.0460.227−0.251
128BC0.0030.0380.227−0.427
129AB00.0190.425−0.212
129AC0.0040.0310.353−0.163
129BC0.0040.0290.361−0.302
130AB−0.0010.0150.331−0.446
130AC0.0040.0180.223−0.436
130BC0.0040.0230.302−0.342
131AB0.0070.0140.232−0.185
131AC0.0030.0230.149−0.208
131BC−0.0040.020.119−0.212
Table 7. Statistical aggregates for precision of the upper maxilla scanned by Carestream CS 3800. All values are provided in millimeters (mm).
Table 7. Statistical aggregates for precision of the upper maxilla scanned by Carestream CS 3800. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122AB0.0350.0780.452−0.643
122AC0.0120.0640.76−0.465
122BC−0.0220.0460.615−0.969
123AB0.010.0320.214−0.418
123AC−0.0010.10.323−0.337
123BC−0.0130.1010.922−1.49
124AB−0.0060.0350.315−0.493
124AC−0.0220.10.462−0.304
124BC−0.0180.0860.459−0.306
125AB0.0120.0210.418−0.19
125AC0.0090.0460.646−1.063
125BC−0.010.1031.287−1.412
126AB00.0280.749−0.536
126AC−0.0120.1241.985−2.794
126BC−0.0050.0470.72−0.811
127AB0.0010.0160.45−0.643
127AC00.0160.385−0.385
127BC−0.0010.0211.366−0.808
128AB0.0070.0270.259−0.387
128AC0.010.0340.274−0.359
128BC0.0020.0210.181−0.226
129AB0.0070.0340.354−1.05
129AC0.0090.0760.276−0.352
129BC0.0010.0690.47−0.24
130AB−0.0120.0260.185−0.478
130AC−0.0150.040.468−0.442
130BC−0.0020.0941.31−1.599
131AB−0.0020.0240.447−0.423
131AC−0.0070.0360.163−0.553
131BC−0.0010.0180.358−0.319
Table 8. Statistical aggregates for precision of the lower mandible scanned by CS Carestream 3800. All values are provided in millimeters (mm).
Table 8. Statistical aggregates for precision of the lower mandible scanned by CS Carestream 3800. All values are provided in millimeters (mm).
SplintTimeMeanSDMax +Min −
122AB−0.0050.0340.517−0.84
122AC−0.0040.0370.511−0.415
122BC0.0010.0320.743−0.358
123AB−0.0310.0720.332−0.463
123AC−0.0040.0410.593−0.317
123BC0.0270.0630.597−0.319
124AB0.0090.0380.409−0.357
124AC0.0070.040.518−0.244
124BC−0.0020.0560.454−0.566
125AB0.0010.0330.192−0.312
125AC00.0710.231−1.276
125BC−0.0010.0530.383−0.868
126AB−0.0030.0230.295−0.63
126AC−0.0040.0350.571−0.641
126BC−0.0020.0310.322−0.714
127AB0.0060.0270.341−0.475
127AC−0.0120.0290.266−0.449
127BC−0.0180.0330.186−0.475
128AB−0.0130.0380.373−0.264
128AC−0.0140.0480.558−0.707
128BC−0.0030.0330.266−0.366
129AB−0.0030.0741.114−0.74
129AC−0.0170.0470.391−0.264
129BC−0.0140.080.398−0.41
130AB−0.0040.0170.291−0.218
130AC0.0020.0250.323−0.235
130BC0.0050.0270.415−0.347
131AB−0.0080.0410.664−0.811
131AC−0.0030.0360.221−0.259
131BC0.0030.0330.205−0.225
Table 9. Trueness 3Shape TRIOS 3.
Table 9. Trueness 3Shape TRIOS 3.
ArchMean (mm)Mean SD (mm)Mean Max +Mean Min −
Mandible0.0070.0480.387−1.127
Maxilla0.0120.0530.792−1.012
Table 10. Trueness Carestream CS 3800.
Table 10. Trueness Carestream CS 3800.
ArchMean (mm)Mean SD (mm)Mean Max +Mean Min −
Mandible−0.0050.0630.484−0.994
Maxilla−0.0010.0710.862−1.076
Table 11. Precision 3Shape TRIOS 3.
Table 11. Precision 3Shape TRIOS 3.
ArchMean (mm)Mean SD (mm)Mean Max +Mean Min −
Mandible00.0290.398−0.617
Maxilla0.0010.0290.36−0.435
Table 12. Precision CS Carestream 3800.
Table 12. Precision CS Carestream 3800.
ArchMean (mm)Mean SD (mm)Mean Max +Mean Min −
Mandible−0.0010.0520.575−0.683
Maxilla−0.0030.0410.422−0.485
Table 13. Statistical aggregates for trueness of the nightguard. All values are provided in millimeters (mm).
Table 13. Statistical aggregates for trueness of the nightguard. All values are provided in millimeters (mm).
SplintMean (mm)SD (mm)Max +Min −
122C−0.0460.7162.444−2.939
122T00.4321.613−1.72
123C−0.0130.2971.318−1.176
123T0.0050.4741.534−1.613
124C−0.0450.2891.004−1.199
124T−0.0740.5962.029−2.266
125C−0.090.3331.274−1.601
125T−0.0120.4821.822−1.897
126C00.2321.095−1.221
126T−0.0590.5741.986−2.248
127C−0.0910.8562.933−2.94
127T−0.0610.4361.345−1.754
128C00.291.3−0.951
128T−0.2440.7062.936−2.94
129C−0.0870.3281.116−1.681
129T−0.0360.4861.526−1.902
130C00.170.672−0.501
130T−0.0850.3751.331−1.647
131C−0.010.2320.623−0.729
131T00.1340.736−0.607
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Anitua, E.; Lazcano, A.; Eguia, A.; Alkhraisat, M.H. Assessing the Impact of IOS Scanning Accuracy on Additively Manufactured Occlusal Splints. Dent. J. 2024, 12, 298. https://doi.org/10.3390/dj12100298

AMA Style

Anitua E, Lazcano A, Eguia A, Alkhraisat MH. Assessing the Impact of IOS Scanning Accuracy on Additively Manufactured Occlusal Splints. Dentistry Journal. 2024; 12(10):298. https://doi.org/10.3390/dj12100298

Chicago/Turabian Style

Anitua, Eduardo, Asier Lazcano, Asier Eguia, and Mohammad Hamdan Alkhraisat. 2024. "Assessing the Impact of IOS Scanning Accuracy on Additively Manufactured Occlusal Splints" Dentistry Journal 12, no. 10: 298. https://doi.org/10.3390/dj12100298

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop