1. Introduction
Salinity is a significant property of natural water and industrial wastewaters. Salinity detection and measurement has widespread applications and a great realistic significance in the protection of the marine environment, oceanic, mariculture, water source quality, and concrete structures [
1]. In concrete structures, Portland cement plays a key function in maintaining the health of the structure, however, it is an alkaline material. Ideally, it is the preferred choice to place the concrete structure at, or near, an alkaline environment to preserve and maintain the integrity of the concrete structure; however, it is challenging in the scenario where concrete structures are built in a marine environment. Due to the pores and micro-cracks in the concrete, the saline solution can infiltrate into the concrete structures. The infiltrated saline solution is a mild acid that lowers the pH of the concrete and compromises the adhesive ability of the cement that holds the concrete structures together. Furthermore, the saline solution also attacks the concrete paste and increases the pore size, which allows more infiltration of the saline solution and other chemical substances into the deeper parts of the concrete structures, thus exacerbating the freeze/thaw cycle damage. When the salt reaches the steel reinforcement, the thin protective layer of iron oxide surrounding the reinforcement has limited ability against the corrosive attack. Eventually, the protective layer will fail and the corrosion will aggravate. Generally, the created corrosion products absorb water and their volume can be 3–6 times larger than their original size; the resulting pressure from the growing corrosion can cause the surrounding concrete to crack and break [
2]. Considering the connection between the solution salinity, the pH of the solution, and the strength of the concrete paste, the salinity concentration can serve as a good indicator for diagnosis and structural monitoring of the concrete under corrosion.
Electrical conductivity is a commonly used parameter for liquid salinity measurement, however, its relationship to the mass of dissolved salts are subject to several other variables and uncertainties, such as the ionic substances in the solution may have different conductivity from the sodium chloride salt solution, the amount of undissolved salts in a saturated saline solution, and the local temperature of the solution where the measurement is made. The operation of the electrical conductivity meter for measurement of the salinity requires ensuring the calibration solution is fresh. If the conductivity meter measures a wide range of conductivities, the calibration solution should have a similar conductivity range [
3].
Abbe refractometers are among the most commonly used system for refractive index (RI) measurement, which is based on determination of the critical angle of an analyte on an optical prism. Generally, small test samples are required and consumed for every measurement and it may not be suitable for an application where in situ measurement and monitoring are required.
Owing to the inexpensive and efficient delivery of a signal, optical fiber devices are commonly used in industrial sensing applications [
4,
5,
6,
7,
8]. The fiber optic sensors inherit many advantages, such as immunity to electromagnetic interference, compact size, high resolution, safety in hazardous environments, high sensitivity, low noise, high durability against seawater corrosion, the possibility of processing the signal at a long distance from the sensor with little degradation, etc.
An optical fiber sensor for salinity measurement was proposed by Zhao et al. [
9]. In the measurement, the deviation of the refracted beam caused by the change in the liquid refractive index was used as an indicator for the change in liquid salinity (PSD). An advanced fabrication technique, such as focused ion beam (FIB) milling, has been employed for manufacturing a three-wave Fabry-Pérot interferometer (FPI) in a single-mode fiber (SMF). The sensitivity matrix method was used for the simultaneous measurement of temperature and salinity of seawater [
10]. Fiber Bragg grating (FBG) biosensors and chemical sensors [
11,
12,
13,
14] are important types of fiber sensors that have received much attention in biochemistry applications. In general, FBGs are functionalized with an RI-sensing ability through the removal of parts or the entire cladding of the fiber by means of post-treatment techniques, such as chemical etching and mechanical polishing, to enhance the evanescent field interaction with the analyte. Liang et al. reported a high-sensitivity RI sensor based on a 1.5 µm diameter etched eroded fiber FPI [
15] that can detect an index variation as small as 1.4 × 10
−5. Long-period grating coupling the core modes into the cladding modes results in the increased intensity of the evanescent field in the fiber cladding. Therefore, it can sense the ambient RI without etching the cladding or a complex structure design. Microfiber is another type of carrier frequently used to design optical sensors due to its very large exposed evanescent field. Many different structures with enhanced sensitivities have been developed [
16]. Tapered plastic fiber optic sensors based on the intensity modulation technique have been implemented for salinity and cement paste curing temperature measurement [
17]. Different sensor’s head designs, such as U-shaped and spiral-shaped, have been investigated and compared. U-shaped sensors have proven to be more efficient in terms of simplicity, sensitivity, and reversibility [
18].
In this paper, a simple and low-cost salinity sensor based on a bundled plastic optical fiber (POF) and a concave mirror is demonstrated. The bundled POF comprises of a POF which serves as a transmitting fiber surrounded by ten smaller POFs, which serve as receiving fibers. A differential detection method is used to eliminate the fluctuation of the light source intensity and the photodetector’s (PD) transfer function. A compact multiplexing system with four integrated sensors is proposed. The multiplexing system comprises of a light-emitting diode (LED) as the light source and the high-speed photologic photodetectors. A He-Ne laser and silicon detector were used in the work of [
19]. The cost of the proposed sensor system is reduced by approximately 70% compared with the early work reported in [
19]. Experimental results indicate that the salinity detection sensitivity of the proposed sensor is 1279 times higher than the previously reported method in [
19].
2. Analytical Results and Discussions
The geometrical configuration of the proposed sensor is shown in
Figure 1a. The longitudinal axis of the transmitting fiber core is coaxial with the normal axis of the concave mirror. The original laser source emitting from point
O is situated inside the transmitting fiber at a distance of
za from the fiber end with an emitting angle
θ1. The emitted light incident on the liquid surface is refracted by an angle
θ2. The liquid height is
x. Light reflected by the air-liquid interface is very weak and its contribution to the received power is not significant. Based on the refracted ray geometry in the liquid, the virtual point source
O1 is located at a point
z1 along the longitudinal axis above the original laser source
O in the air. The refracted light is then reflected by the concave mirror immersed in the liquid. The light converges at a point
O2 and becomes the second virtual emitting point source for the receiving fiber. The location of the new virtual emitting point source
O2 is influenced by the RI of liquid and the height of liquid
x. The new virtual point source
O2 emits at an angle
θ4 with the longitudinal axis in the opposite direction.
Figure 1b illustrates the configuration of the sensor probe. It is a bundled POF that comprises of a transmitting fiber (T) and 10 receiving fibers. The light from the LED is guided by the transmitting fibers to the concave mirror, whereas the receiving fibers are used to collect the reflected light. Since they are made of the same material, all of the fibers are assumed to have the same transmission loss
K. The receiving fibers are divided into two groups in which five of them (R
1) are connected to PD
1, whereas the other five (R
2) are connected to PD
2. As such, the power detected by PD
1 and PD
2 can be given by
, where
Pr1 is the light power detected by PD
1,
Pr2 is the light power detected by PD
2, and
Pre is the light power collected by each receiving fiber.
Pre can be expressed as a function of liquid RI
n and the height of liquid level
x,
where
r and
z are the radius and longitudinal coordinates along the receiving fiber, respectively;
D is the diameter of the concave mirror; and the light power emitted from the transmitting fiber is
PE. The expressions for
v1 and
θ4 are provided in [
19]. It is worth noting that the receiving fibers,
R1, have a larger core area diameter than that of
R2. Both fiber groups will produce different characteristic curves. The intention of having two groups of receiving fibers with different core diameters will become clearer in the latter part of this section.
From the Beer-Lambert Law, we know that the light power in saline has an absorbance for
N attenuating species in the solution,
, where
εi is the attenuation coefficient of the attenuating species
i in the solution,
l is the path length of the beam of light through the solution, and
ci is the concentration of the attenuating species
i in the solution. It is assumed that
PD1 and
PD2 have the same transfer function
F, which converts the detected light powers of
Pr1 and
Pr2 to voltages
V1 and
V2, respectively. The output voltages of
PD1 and
PD2 can be expressed as:
If both receiving fibers are identical, then K
1 = K
2, F
1 = F
2. If the NaCl solution is uniform, it will induce the light absorbance inside the solution to only vary with the liquid level
x. Hence, the light emitted from the second virtual emitting point source has a uniform absorbance, which means the absorbance
A1 for detected
V1 and
A2 for detected
V2 are identical. From the outputs of PD
1 and PD
2, a new characteristic output curve,
V can be produced:
For an intensity modulation-based fiber sensor, the sensor performance is heavily affected by the light source power intensity. However, the sensor’s output V in Equation (4) indicates that the sensor is immune to the variation of K, F, PE, and liquid absorption. According to Equation (4), we can see that the sensor output depends on the variables RI n, liquid level x, and sensor probe position u. For a fixed liquid level x, a small change of RI n or position u excites the variation of the sensor’s responses.
The analytical results of the proposed sensor are presented in
Figure 2, of which,
Figure 2a shows the sensor outputs of two PDs for RI = 1.333 and
u is in the range of 3–25 mm. The near displacement region (0–3.0 mm) is referred to as the blind region for RI sensing because the same power variations are observed in the output responses for different RI. The difference can only be observed when the displacement goes beyond 3.0 mm. One dip located in between the two peaks is observed in each output curve. The dip is located at the position in which the displacement is twice the focal length of the concave mirror in the air medium (RI = 1). However, the dip position is closer to the mirror when the air medium is replaced with a liquid with a higher RI. At these points, the virtual point source
O2 is located at the end facet of the transmitting fiber in which the incident light cone becomes a point at the center of the transmitting fiber core. This result is zero, or minimum reflective light, collected by the receiving fiber.
From Equation (1), it can be seen that the light power detected by the PDs are related to the radius of the receiving fiber
r. Two groups of receiving fibers with different core radii will induces the PDs to have slightly different output curves, which are shown in
Figure 2a. The two output curves in
Figure 2a indicate that a differential detection method (Equation (4)) can be employed to improve the sensors’ sensitivity.
Figure 2b shows the output of the sensor for three different RIs using the differential detection method. Each curve in this figure has one dip and one peak, in which the dip is close to the zero point when the RI is increasing. The results in this figure also show that the proposed sensor has a very broad RI detection range. From the curves in
Figure 2a, it can be seen that a slight change of displacement at ~15 mm will excite the PDs’ output to vary slowly. However, the curve (RI = 1.333) in
Figure 2b shows that the slight change of
u at ~15 mm will induce the output
V to vary sharply. As such, it can be concluded that the proposed sensor’s sensitivity is improved significantly by employing two groups of receiving fibers with different core radii, combined with the differential detection method.
3. Experimental and Results Discussion
The experimental setup of the proposed fiber sensor is shown in
Figure 3a for simultaneous detection of the salinity in four cuvettes. In this system, four sensor probes are multiplexed into one system which integrates four LEDs and eight PDs to emit and detect the light, respectively, as shown in
Figure 3a. Four concave mirrors with the same focal length of 10 mm and diameter of 24 mm are located at the bottom of the four cuvettes, respectively. Four sensor probes, as shown in
Figure 1b, are formed with the POF with the same parameters, with that the radius of the transmitting fibers is 0.25 mm and the radii of the receiving fibers are 0.1 mm and 0.05 mm. The POFs have core radii of 0.09 mm and 0.045 mm, and refractive indices of 1.492 and 1.450 for the core and cladding, respectively. A circuit is designed to provide a stable current to drive the LEDs (red light, centered at 660 nm). The LEDs used in this setup show an excellent stability in the output light intensity.
Figure 3b illustrates one of the sensor structures for salinity detection in a cuvette, in which the light is delivered through the sensor probe and collected by two high-speed photologic PDs. The PD converts the light power into a voltage signal. The circuit also equips a simple function to process the output voltages of the PDs to achieve the equation
V = (
V1 −
V2)/(
V1 +
V2). Of which,
V1 and
V2 are the output voltages of PD
1 and PD
2, respectively. A TDS-530 data logger (Tokyo Sokki Kenkyuio Co. Ltd., Tokyo, Japan) is used to collect the processed output voltage signal
V. The TDS-530 data acquisition system is set at a 640 mV detection range. The data logger is also set for the acquisition of one point each second. Four sensor probes are fastened on a multi-axial stage that provides a precise displacement control. The stage has a displacement resolution of 0.1 µm. The control of displacement in the y-axis direction is used to move the sensor probes in the vertical direction while the other two axes ensure accurate alignment between the longitudinal axis of the fiber probes and the normal axis of the center of the concave mirror. The experiment was conducted in a laboratory where the room temperature was 25 °C and the experimental setup includes four cuvettes, the multi-axial stage, and four sensor probes are placed on an optical table, a vibration-free platform for the experiment. To accurately control the liquid surface flatness, a wait time of 5 min is imposed after each movement of the liquid level or sensor probes to reduce the surface corrugations. Before the experiment, both NaCl salt and distilled water are stored at room temperature (25 °C) for two days to ensure the accuracy of experimental results. The solutions with concentrations of 1 wt % to 5 wt % with 1 wt % intervals were prepared to test the proposed sensor. The system drift is studied by employing the measurement system to detect the received light intensity. The room temperature is also monitored by using a thermocouple. The test is performed for 48 h, and a maximum drift of ~2% and 2.0 °C is found for the light intensity and room temperature, respectively. This indicates that the experimental work occurs under well-controlled ambient conditions.
From Equation (4), it is worth noting that the liquid level
x also affects the sensor response. However, in the practical application the liquid forms a convex surface at
x > 3 mm when it is poured on the concave mirror. The liquid convex surface reflects and diverges the incident light from the emitting fiber. This results in a reduction in the light power collected by the receiving fiber. A minimum liquid height of 3 mm is suggested to avoid the occurrence of the convex liquid surface. The theoretical and experimental curves in
Figure 4 are in good agreement with each other, except for the positions of the dip and the peak. It is shown that the dip and peak of the experimental curve is closer to the zero point than that of the theoretical case. This can be attributed to the concave liquid surface, while a flat surface is assumed in the theoretical approximation. The concave liquid surface yields a lower position of the virtual point source
O2 than that of a flat surface, which contributes the discrepancy between the theory and experiment. It is worth noting the liquid surface depends on the liquid level,
x. However, its influence on the sensor output is minimal and it can be disregarded.
The saline solution is slowly filled into the cuvette to make sure the level is as close to 3 mm as possible. The sensor probes are moved at a step size of 0.1 mm, between 3 mm and 25 mm, twice. After that the collected data are averaged. The measured RIs of the saline solutions by a prism coupler at 632.8 nm are 1.3394, 1.3454, 1.3529, 1.3581, and 1.3654, which increase with the higher salinity concentration. The calibrated experimental results are only plotted in the displacement of 14 mm and 16 mm, which includes five peaks and five dips, as shown in
Figure 5a. The results in
Figure 5a clearly show that the peak of the curve is moving closer to the zero point with the increasing RI, as enclosed in the horizontal dotted ellipse. This can be attributed to the lower virtual point source
O2 produced from the solution with a higher RI. Furthermore, the readings at the displacement of 14.7 mm (refer to the vertical dotted ellipse in
Figure 5a) also varies with the increase of the RI. As such, the points in these two regions can be used as the RI indicators which provide two RI detection approaches for the proposed sensor: (Method 1) a fixed liquid level
x with varying sensor probe position,
u, as indicated by the horizontal dotted ellipse; and (Method 2) a fixed sensor probe position,
u, but varying the liquid level
x, as indicated by the vertical dotted ellipse. Accordingly, the sensitivity of the proposed sensor is shown in
Figure 5b,c for Methods 1 and 2, respectively. A linear curve fitting can be applied to the data points in
Figure 5b for Method 1, which show that the proposed sensor has a sensitivity of 15.419 mm/RIU by using this method. From
Figure 5c, it also can be found that the proposed sensor exhibits a maximum sensitivity of 14,847.486/RIU at RI = 1.3525 by using Method 2. The experimental setup in
Figure 3a shows that the proposed sensor has an ability for the measurement of four samples simultaneously. The experiment was also performed with saline solutions with the same concentration to ensure the consistency of the test. The output data from the four sensors have a standard deviation of
σ = 0.02.
To ensure the repeatability and accuracy of the measurements, re-calibration is performed prior the experiment. The sensor probe is reset to the same position in which distance u is the same in every measurement. This step can be treated as the zero setting. Then, the solution can be slowly poured into the cuvettes to increase the liquid level in steps from 0 mm to 3 mm during the re-calibration work. After each test, the sensor probe and the used cuvettes are cleaned with DI water and dried before proceeding to the next test.