Next Article in Journal
Thickness Dependence of Ferroelectric and Optical Properties in Pb(Zr0.53Ti0.47)O3 Thin Films
Previous Article in Journal
Online Feature Selection for Robust Classification of the Microbiological Quality of Traditional Vanilla Cream by Means of Multispectral Imaging
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Comparative Study of Coupling Techniques in Lamb Wave Testing of Metallic and Cementitious Plates

1
Instituto de Telecomunicaciones y Aplicaciones Multimedia (iTEAM), Universitat Politècnica de València, Camino de Vera, s/n, 46022 Valencia, Spain
2
Instituto de Ciencia y Tecnología del Hormigón (ICITECH), Universitat Politècnica de València, Camino de Vera, s/n, 46022 Valencia, Spain
*
Author to whom correspondence should be addressed.
Sensors 2019, 19(19), 4068; https://doi.org/10.3390/s19194068
Submission received: 19 August 2019 / Revised: 12 September 2019 / Accepted: 17 September 2019 / Published: 20 September 2019
(This article belongs to the Section Physical Sensors)

Abstract

:
Lamb waves have emerged as a valuable tool to examine long plate-like structures in a faster way compared to conventional bulk wave techniques, which make them attractive in non-destructive testing. However, they present a multimodal and dispersive nature, which hinders signal identification. Oblique incidence is one of the most known methods to generate and receive Lamb waves and it is applied in different experimental arrangements with different types of sensors. In this work, several setups were conducted and compared to determine the optimal ones to launch and detect ultrasonic Lamb waves, especially in non-homogeneous specimens. The chosen arrangements were contact with angle beam transducers, immersion in a water tank, localised water coupling using conical containers and air coupling. Plates of two different materials were used, stainless steel and Portland cement mortar. Theoretical and experimental dispersion curves were compared to verify the existence of Lamb modes and good correspondence was achieved.

1. Introduction

Conventional ultrasonic inspection techniques used in non-destructive testing (NDT) are based on bulk waves (also called body or volume waves). These procedures are time-consuming since a point-by-point scan is needed to obtain global information from a structure because they only cover the localised area below or adjacent to the transducer [1]. Bulk waves consist of longitudinal (also known as P-waves/primary waves, pressure waves or compressional waves) and transverse waves (also known as S-waves/secondary waves or shear waves). Both kinds of waves occur in solids with dimensions greater than a wavelength in the three dimensions [2,3]. Nevertheless, in non-viscous fluids, such as water or air, only longitudinal waves can propagate [1]. The particle motion of longitudinal waves is parallel to the direction of wave propagation while the particle motion of transverse waves is perpendicular to the direction of wave propagation. The remaining waves, as guided waves, are composed of a combination of longitudinal and transverse wave-particle velocity components [1].
Ultrasonic guided waves have become an important subject in NDT. They are a faster, more sensitive and more economical way of inspecting materials compared with bulk wave techniques. The main advantage is that a whole structure can be inspected globally from a single probe position [4,5]. These waves need boundaries to propagate, unlike bulk waves whose propagation is unaffected by boundaries. Depending on the type of structure with boundaries (waveguide), guided waves receive different names. If the waveguide is a half-space (a thick structure) Rayleigh waves (R-waves) are found, which propagate in the surface of the tested specimen because the wavelength is very small compared to its thickness. Stoneley and Scholte waves travel in solid-solid and liquid-solid interfaces (or half-spaces), respectively. For pipes or cylindrical rods, cylindrical guided waves appear and the guided waves that occur in plates (or multilayered plates [6,7]) are known as Lamb waves [1].
Lamb waves are frequently used in ultrasonic NDT because they can travel long distances with little energy loss [8,9,10]. The main problem with Lamb waves is that they are dispersive (their phase and group velocities depend on frequency) and multimodal (different particle motion). Their velocities and modes can be obtained from characteristic Equations and the graphical solutions of these Equations are known as dispersion curves. As frequency increases, many modes with different velocities exist in the received signal, overlapping and hindering Lamb wave analysis. To improve the interpretation of these dispersive and multimode signals, different recommendations may be followed:
  • A region of the phase/velocity dispersion curves where little dispersion exists (i.e., small variations of phase/group velocity when changing frequency) may be selected to reduce variations in amplitude and the shape of the sensed signal as it propagates along the plate [11,12].
  • Modes with different group velocities may be chosen to separate them in the time domain [11,13,14].
  • Low frequencies may be chosen to minimise the number of modes in the received signals, i.e., “working on the left side” of the dispersion curves. In fact, the best would be to excite only one mode because that would enable better damage identification since the appearance of non-excited modes could indicate the existence of defects, i.e., mode conversion [11,15,16].
The fundamental modes A0 and S0 are frequently used for damage detection [17,18] since, at low frequencies (until the first cut-off frequency), they are the only existing modes. Normally, the A0 mode presents a larger signal magnitude than the S0 mode [6]. To isolate both modes, one can work with low-frequency transducers or inspect plates with the lowest thickness possible. Different types of defects that have been detected by Lamb waves are delamination [16,19], cracks [10,20], notches [21,22] and corrosion [17,23]. Depending on the type of defect, a determined mode may be excited depending on its sensitivity to that defect. Modes that present great displacements in the thickness direction, like the A0 mode, have good sensitivity to surface damage, such as surface cracks [6], while modes with great displacement in the direction of wave propagation, like the S0 mode, mean good sensitivity to any defect in the thickness of the specimen, such as delamination [24]. Some procedures to detect these defects include observing the velocity shifts in the experimental dispersion curves [19], the amplitude of the received signals [3] and peaks of the frequency spectrum [5].
One of the most known methods to excite guided waves, in general, is oblique incidence (also found in the literature as the wedge method or the coincidence principle) [11,14,24,25,26]. It consists of striking a bulk wave (generally a longitudinal wave, since water and air do not support transverse waves) at a determined angle with respect to the normal in a coupling medium (Plexiglas, water, air, etc.) to generate Lamb waves in the specimen by following Snell’s law. In this way, a specific mode can be selectively excited. Traditional ultrasonic probes are usually employed in oblique incidence. Although Lamb waves can be excited and measured by normal beam excitation [9,27,28] this configuration is not as efficient as oblique incidence. However, selective excitation of Lamb wave modes can be achieved by normal beam excitation. For instance, some researchers excite different modes by changing the polarity of piezoceramic lead-zirconate-titanate (PZT) transducers normally attached on both sides of the plate [8,16]. Others modify the excitation frequency and the geometry of piezoelectric wafer active sensors (PWAS, basically a network of PZT sensors) to selectively excite Lamb wave modes (Lamb wave tuning) [10,29,30]. PZT and PWAS are widely used in Lamb wave excitation and damage detection [31,32,33].
Both kind of sensors must be strongly coupled with the inspected structure. For that, different bonded materials have been studied [34,35]. They present several advantages over the typical ultrasonic probes: they are cheaper, lighter, smaller, need less power and can be embedded in structures [31]. Nonetheless, they are fragile, could show non-linearities at certain conditions and usually generate multiple Lamb wave modes [6]. In this work conventional ultrasonic transducers with oblique incidence have been used.
Different techniques can be used to change the inclination of the transducers, such as the contact technique with angle beam wedge transducers [7,36,37], immersion coupling [3,13,38] and air coupling [19,39,40]. Angle beam wedge transducers are basically longitudinal wave transducers placed on wedges with a certain angle. These wedges are usually made with materials, such as Teflon or Plexiglas, with a bulk wave velocity lower than the velocity of the mode that you want to work with. A coupling agent (gel, petroleum jelly) must be used between the transducer-wedge interface and the wedge-plate interface to provide efficient transmission of energy to the specimen [40,41]. Nevertheless, the received signal is sensitive to the thickness of the layer of the coupling gel, the contact pressure between the transducer-wedge and the wedge-plate interfaces [5,28,40] and the reflections produced within the wedges [11]. Non-contact techniques avoid these problems and enable constant coupling and faster, easier scanning [13] of the plate with repeatable results [42] compared to the direct contact case. One non-contact technique [3,13] is immersion coupling, which consists of completely submerging the test object in a fluid tank (usually water). Longitudinal wave immersion transducers are commonly employed in this setup with an acoustic matching layer between the piezoelectric element and water (the same principle is applied to piezoelectric air-coupled transducers). In this case, leaky Lamb waves (LLW), which are formed by the leakage of energy into the surrounding medium (water, air), are measured [1,38,43,44]. Note that if the acoustic impedance (or density) of the surrounding medium is small compared to the acoustic impedance of the plate, the LLW velocities can be approximated by the Lamb wave velocities [40,45,46]. Nonetheless, the received signal is formed with the addition of the generated LLW and other undesirable signals, as the reflections from the water surface, the bottom of the tank, geometrical (or specular) reflection or the direct wave through the water [28,38,44]. These undesirable signals pollute the received waveform, making mode identification difficult. To solve these issues, some researchers place some insulating material in the bottom of the tank and between the transducers [44] or increase the separation between the plate and other surfaces (bottom of the tank, water surface, etc.) [13]. Additionally, the excited mode could suffer great attenuation due to energy leakage from the plate, especially if the separation between transmitter and receiver increases (e.g., the A0 mode) [3,6,11,43,47,48]. Other researchers avoid completely immersing the plate and choose water columns [42], conical water containers [28], water balloons [49] or water wedges [14,50] as the coupling mechanism. However, immersion is not practical for testing large structures, hot structures or specimens whose surface cannot be contaminated [40,51]. In these cases, air coupling is a better solution, since it maintains the advantages from immersion testing, such as constant coupling, provide fast scanning measurements [52] and almost solves the drawbacks from the undesired contributions, except for the direct acoustic wave through the air, which can be attenuated by an acoustic barrier placed between the transducers [53,54]. Two types of air-coupled transducers can be found, capacitive and piezoelectric. The capacitive design generally offers more bandwidth than the piezoelectric ceramic elements [55,56]. Nevertheless, the received waveforms show a low signal-to-noise ratio (SNR) because of the high attenuation present in air [52,53]. Therefore, amplification in transmission/reception and averaging is needed [43,52,57,58]. Furthermore, there are modes which are difficult to excite in air because of its low surface motion, such as the S0 mode [40,43,52,59,60]. On the contrary, A0 is a mode considered suitable for air-coupled ultrasonic testing due to its great out-of-plane displacement [39,43].
Lamb waves have been generated and received in different materials. The most commonly used are metals [3,4,8,14,36,39,43,61], but composites [9,15,24,51,60] and cementitious materials [19,62,63] are also used. However, the excitation and detection of Lamb waves in cementitious plates is still a challenging task.
The goal of this work is to compare and assess different techniques for generating and receiving Lamb waves and determine the optimal technique considering the frequency, material and mode. For this goal, the first specimen tested was a metallic plate (a reference plate in Lamb wave testing) and the contact technique, immersion technique (with two alternatives) and air coupling were analysed. The fundamental Lamb modes were successfully generated and detected. After performing a considerable amount of experiments in the metallic plate, two techniques were chosen to test a plate of a more heterogeneous material, Portland cement mortar. Despite being a more complicated material, Lamb waves were also excited and sensed. This may be due to its small thickness compared to the conventional cementitious specimens, which are often thick.
The paper is organised as follows. In Section 1, a broad introduction to Lamb waves was described. Section 2 shows the mathematical background employed in the theoretical and experimental data. In Section 3, an explanation about the different experimental arrangements is offered along with the results, which consist of matching the experimental and theoretical dispersion curves to verify which modes were excited in every case. Finally, in Section 4, the conclusions are presented.

2. Mathematical Background

Lamb waves propagate in a linear, homogeneous and isotropic elastic plate with stress-free upper and lower surfaces (as if the plate was placed in vacuum [36,62]) and with lateral dimensions (length and width) far greater than the thickness [39]. They are created by the constructive interference of reflections of longitudinal and transverse waves with both plate surfaces as long as the employed wavelength is greater than the plate thickness (a possible relation from [13] is 2 h 3 λ , where 2 h is the total plate thickness and λ the wavelength). Depending on the particle motion with respect to the middle of the plate (see Figure 1), Lamb waves can be classified into antisymmetric and symmetric wave modes (labelled A m and S m , respectively, where m = 0 ,   1 ,   2 ,   3 ,   indicates the “order”). Antisymmetric modes generally present out-of-plane particle displacement (in the transverse direction) while symmetric modes predominantly have in-plane particle displacement (in the longitudinal direction) [6,13,43].

2.1. Theoretical Dispersion Curves

For a linear, homogenous and isotropic elastic plate with stress-free boundaries and lateral dimensions far greater than the thickness (which has a value of 2 h ), the characteristic Equations for guided symmetric Lamb wave motion can be expressed by Equation (1) [1,10,64,65]:
4 k 2 p q sin ( p h ) cos ( q h ) + ( k 2 q 2 ) 2 sin ( q h ) cos ( p h ) = 0 ,
and for antisymmetric Lamb wave motion by Equation (2):
4 k 2 p q sin ( q h ) cos ( p h ) + ( k 2 q 2 ) 2 sin ( p h ) cos ( q h ) = 0 ,
p 2 = ω 2 c L 2 k 2 ,       q 2 = ω 2 c T 2 k 2 ,
where k = ω c p is the wavenumber, ω = 2 π f is the angular frequency, c p is the phase velocity and c L and c T are longitudinal and transverse velocities of the material, respectively. Characteristic Equations (1) and (2) are also known as dispersion Equations for Lamb waves or Rayleigh-Lamb Equations [6].
The solutions of Rayleigh-Lamb Equations for the phase velocity, c p , generally plotted as a frequency depending function, are called dispersion curves. Nonetheless, there is the option to normalize the axis of the dispersion curves with respect to plate properties, e.g., the frequency axis can be multiplied by the plate thickness and the phase velocity axis can be normalized by the transverse velocity of the plate [52,62]. The phase velocity, c p , can be defined as the speed at which the phase of any frequency component of the wave travels [39]. An example of phase velocity dispersion curves where different modes appear is found in Figure 2.
Note that for dispersion curves of symmetric modes, there is a horizontal part where the phase velocity approaches the quasi-longitudinal wave velocity [62]. In fact, the longitudinal wave is the fastest wave that appears in plates [36]. Two modes exist for all frequencies, the fundamental modes A0 and S0. At higher frequencies, both modes approach the Rayleigh wave velocity of the plate while the rest of Lamb wave modes ( A m , S m where m >0) approach the transverse wave velocity of the plate [14]. These higher modes have cut-off frequencies where their phase velocity tends toward infinity [1,61].
When computing the dispersion curves, the real solution of the Rayleigh-Lamb Equations is chosen to represent the propagating modes of the plate. One way to achieve that is to divide Equation (1) by q and Equation (2) by p. Then, the procedure explained in [1,6] can be performed to obtain the phase velocity dispersion curves. Once the phase velocity, c p , is known, other parameters, such as the group velocity, c g , the angle of incidence, θ , or the wavenumber, k , can be determined. Figure 3 shows the dispersion curves of theses specific parameters for a 1.1 mm stainless steel plate.
The group velocity, c g , can be thought of as the speed at which the wavefront of each mode propagates [64] and is given as [18,66,67]:
c g = c p 2 c p f c p f
The incident angle, θ , is governed by Snell’s law [26]:
c p sin θ = c sin θ r
where c is the bulk longitudinal wave velocity of the coupling medium (water, air, Plexiglas, etc.) and θ r is the refraction angle. As θ is the critical angle which selectively excites a Lamb wave mode with a selected phase velocity, c p , θ r must be equal to 90°. Therefore, the optimum angle of incidence, θ , is [14,40,52]:
θ = sin 1 ( c c p )
It is important to choose a coupling material with c < c p . Equation (6) is the basis of the wedge method [14,25] and the coincidence principle [36,44]. Placing the receiver at the same angle as the transmitter in a pitch-catch configuration (both transducers above the plate surface) enables efficient reception (a higher amplitude) of the excited mode, suppressing others [3,11,52,59,68]. One advantage of pitch-catch setups is that they only need access to one side of the structure, in contrast to through-transmission (also called through-thickness) setups which need double-sided access [15].
Therefore, to excite and receive a certain mode, the excitation frequency and the angle of inclination with respect to normal to the transmitter and receiver must be chosen. Incident angle dispersion curves are helpful for this task [52,69]. To identify the generated modes, theoretical and experimental dispersion curves must be overlaid [16,25,37,39,62,70].

2.2. Experimental Dispersion Curves

Experimental phase velocity (or wavenumber) dispersion curves can be mapped out using the two-dimensional fast Fourier transform (2D-FFT) [36]. This method requires the collection of equally spaced waveforms that can be acquired by moving the receiver (or the transmitter) along the specimen [36,39,52]. With this technique, the identification of individual modes is possible [36,65].
Another procedure to measure experimental phase velocity dispersion curves is varying the angle of inclination and locating the peaks or dips [28,38,68,71] from the frequency domain of the received Lamb waves. An automated system is usually required to vary the position of the transmitter/receiver or their angles of inclination [38,39,52].
Experimental group velocity dispersion curves can be constructed from time-frequency representations (TFR) of the received signals, such as the short-time Fourier transform (STFT), the wavelet transform (WT), the scalogram, the Wigner-Ville distribution or the chirplet transform [72,73,74,75]. These techniques require only a single received waveform, which is more practical and less time-consuming [72] than the methods explained to obtain phase velocity dispersion curves (although, group velocities can also be obtained from moving the transmitter/receiver [38,39]). However, TFR presents problems due to the Heisenberg uncertainty principle [72]. In this work, the spectrogram, the graphical display from the STFT analysis that is widely used in Lamb wave studies, was applied to obtain experimental group velocity dispersion curves, which are very useful for identifying guided wave modes [1,16,37,70]. The STFT of a signal s ( t ) is equal to [21,72]:
S ( f , t ) = 1 2 π + s ( τ ) h ( τ t ) e j 2 π f τ d τ ,
where h ( t ) is the window function. The spectrogram is the energy density spectrum of a STFT and it is defined as:
E ( f , t ) = | S ( f , t ) | 2 .
If broadband signal excitation is used [8,16,61], several modes can be excited in a wide frequency range, which can include dispersive regions and non-dispersive regions, and only one spectrogram would be necessary. In this work, different sinusoidal tone burst signals (which produce narrowband excitation) varying in frequency [39,44,47,70,76] were launched in the tested plate and then every received spectrogram was combined into one [37] (hereafter, the combined spectrogram). The mathematical model of the transmitted sinusoidal tone burst signals is shown in Equation (9) [77]:
x n ( t ) = A · sin ( 2 π f n t ) · r e c t ( t N c 2 f n N c f n ) ,
where A is the signal amplitude, f n is the driving frequency, N c the number of cycles and r e c t ( · ) is the rectangular function.
Five [11,36,39,76] and ten cycles [14,58,78] are a reasonable number of cycles for burst signals. Furthermore, windowing the emitted signals can reduce side lobes to avoid exciting other modes [11,40,79] in addition to removing the ringing effects of the transducers [80].
The group velocity, c g , can be obtained from the spectrogram. This velocity is also defined as the quotient of the propagation distance, d , in the plate and the flight time corresponding to ultrasound travel through the plate [60,81]. From the time axis, t , of the spectrogram, the experimental group velocity can be calculated as [3,13,37,70]:
c g = d t t c ,
where t c is the approximate delay suffered in the coupling medium. This delay can be obtained from the following relationships:
t c = t c 1 + t c 2 = d c 1 c + d c 2 c = d c c ,  
where t c 1 and t c 2 are the flight times for the ultrasound travelling in the coupling media from the transmitter to the plate and from the plate to the receiver, respectively, d c 1 and d c 2 are the propagation paths (transmitter-plate and plate-receiver, respectively), d c is the addition of both propagation paths and c is the bulk wave velocity in the coupling material, as was shown before. In this study, the three distances, d , d c 1 and d c 2 , are measured using a ruler. Wherever possible, it is recommended that these distances have a value equal to or greater than the near field distance to work in the far-field region [14,53,68,82,83]. The near field distance, N , is given as [1,53]:
N = D 2 f 4 c ,
where D is the transducer diameter.
This recommendation is due to the near field or Fresnel region, where great ultrasound pressure fluctuations occur, hindering defect detection; while in the far-field or Fraunhofer region, the ultrasound pressure gradually tends toward zero [48,53]. If the transducers are placed close to the plate, undesirable reflections can appear between the transducer face and the specimen [59,84]. However, some researchers ignore the near field effects and adjust the distances according to the received amplitude [28] or the signal fidelity and repeatability of the experiment [3], obtaining good results. One solution to avoid distance calculations is to employ phase velocity measurements [28,36,38].

2.3. System Sensitivity Curves

In every arrangement, in addition to the inspected plate, the transducers and the coupling medium influence the magnitude of the received signals and, consequently, the frequency domain of those signals. By analysing this frequency domain, it is possible to know the frequencies where the excited mode presents a higher signal level (in other words, the frequencies where the mode is more sensitive). To this end, the system sensitivity curves ( S S C ) were computed. A diagram of the calculation of the combined spectrograms and the SSC is represented in Figure 4. The procedure is as follows:
1)
The frequency sweep is performed and tone burst signals of different frequencies ( f 1 , f 2 , …, N f , where N f is the number of driving frequencies) are launched on the plate surface by means of the corresponding actuator.
2)
The generated Lamb waves are detected by the sensor and acquired for post-processing.
3)
Signal processing is applied and the spectrograms and the Fourier transforms (FT) for every sensed signal is obtained.
4)
From every spectrogram, a slice including the driving frequency is extracted and then it is normalised by that driving frequency. The slices are represented by rectangles with white discontinuous lines placed on the two spectrograms, E 1 ( f , t ) and E N f ( f , t ) . The combined spectrograms are constructed with N f slices [37,70].
5)
The SSC are built using the absolute values of the FT of the signals received at every injected frequency: | Y 1 ( f 1 ) | , | Y 2 ( f 2 ) | , …, | Y N f ( f N f ) | . These absolute values are marked by blue circles placed on the two FT, | Y 1 | and | Y N f | . The maximum in the FT should appear in the excitation frequency, although there are cases where this is not fulfilled. These curves act as a filter to highlight which part of the combined spectrogram can be considered in terms of SNR. In the diagram, the SSC presents more signal level between f m i n and f m a x . Therefore, the combined spectrogram is analysed in that frequency range, i.e., between the two vertical black discontinuous lines that connects both combined spectrogram and SSC.
S S C ( f n ) = | Y n ( f n ) | ,         n = 1 ,   2 , , N f ,

3. Experimental

Different setups were exposed for exciting and detecting Lamb waves in two plates of different materials. The transducers were placed in a pitch-catch configuration (see Figure 5 for instance) and they are mounted in holders produced by 3D printer. One transducer acts as an actuator (left) and the other as a sensor (right). To approximate the stress-free boundary conditions of the plates, a piece of insulating material (expanded polystyrene (EPS)) was placed under them [16,66,85]. A testing system was built to control the inclination and distances of the actuator/sensor. That inclination was measured using an angle level composed by an Arduino board (Mega 2560, Arduino, Somerville, MA, USA) and two sensors (MPU-6050, Invensense, San José, CA, USA). These sensors present an accuracy of ± 0.03 ° . For every setup, a detailed explanation of the parameters (equipment, excitation, incidence angles, etc.) was given in the next sections and in Appendix A. Furthermore, the experimental data of every setup can be found as supplementary material. The signal generator and the oscilloscope used in the different setups were managed by a computer with a MATLAB graphical user interface developed by the group [86]. Special care was paid to avoid saturated signals by choosing a proper input voltage. The vertical range of the oscilloscope was adjusted to the maximum amplitude of the received signal to minimise quantification errors. Electrical stray coupling [87,88,89] and other undesired effects were removed in the post-processing stage. The spectrogram parameters remain constant for all cases, a 20 μ s temporal Hamming window size, 75% overlap between windows and zero padding, up to 214 points, was applied for every window. Theoretical group velocity dispersion curves were overlaid on the combined spectrograms (i.e., experimental group velocity dispersion curves). The lower and upper frequencies of the SSC (marked by black discontinuous lines) were superimposed on the figures at 90% of the maximum amplitude of the current SSC.

3.1. Materials

The tested samples were a stainless steel and a mortar plate. The main data of these samples are summarised in Table 1.
Although their widths should be several times larger than their thickness to optimize Lamb wave generation, some articles demonstrate that this condition is not strictly required [16,23,90,91]. The mortar was prepared by mixing a Spanish Portland cement (CEM I-52.5R), sand (quartz, 0.6–1.2 mm particle size) and water in the 1:3:0.35 ratio by mass. Superplasticiser was added in order to achieve appropriate workability. The fresh mortar was put in a mold, which was vibrated in a vibrating table (ToniVIB Model 5533, Toni Technik, Berlin, Germany) in order to homogenise the mortar and to eliminate the air bubbles. The specimens, after 24 h in the mold (temperature 20 °C, relative humidity > 95%), were demoulded and stored under water for 90 days. To determine mortar properties, the same mix was prepared and moulded in 40 × 40 × 160 mm3 specimens (according to UNE EN 196-1:2005 [92]) and cured in the same conditions. The bulk wave velocities of both stainless steel and mortar were measured experimentally using the Ultrasonic Pulse Velocity (UPV) method [93,94].

3.2. Techniques and Results on Stainless Steel

The techniques used in the stainless steel plate were the contact technique with angle beam probes, water coupling (immersion and conical containers) and air-coupled testing.

3.2.1. Direct Contact

Both schematic and photograph of the experimental arrangement for the direct contact technique is shown in Figure 5. A pair of angle beam probes (MUBW 2N, Krautkramer, Huerth, Germany) with a 2 MHz central frequency were employed for Lamb wave excitation and detection. In these transducers, the piezoelectric element (9 mm large × 8 mm width) was embedded in a Plexiglas wedge, which has a longitudinal wave velocity of 2730 m/s (value obtained from the transducers data sheet). The angle can be changed manually from 0° to 60°. Petroleum jelly (Panreac, Darmstadt, Germany) was used as a coupling between the Plexiglas wedges and the plate. The effects of the coupling were ignored in the group velocity calculations. The actuator was excited by a ten-cycle tone burst using a programmable signal generator (33120A, Agilent Technologies, Loveland, CO, USA). The excitation frequencies were swept from 1 MHz to 3 MHz, in 20 kHz steps. The sent signals were amplified 40 dB (5660C, Panametrics, Waltham, MA, USA) and then captured using a digital oscilloscope (DPO3014, Tektronix, Shanghai, China) with a 25 MHz sampling frequency, 10,000 sampling points (a temporal interval of 400 μ s ) and 32 averaging.
Two incidence angles were chosen to excite and “follow” the predominant S0 mode in the frequency range from 1 MHz to 3 MHz: 32° and 50° (see Figure 6). The A0 mode cannot be generated in this arrangement since the maximum angle that can reach these transducers is 60°.
In Figure 7, the combined spectrograms for the two chosen angles along with the theoretical group velocity dispersion curves and system sensitivity curves (SSC) are represented. There is a good correspondence between the theoretical curve of the S0 mode (red discontinuous curve) and the experimentally obtained dispersion curve (yellow) hot spots [1] in the frequency ranges that mark the SSC, from approximately 1.2 to 2.2 MHz in both angles, 32° and 50°. The SSC values for 32° are larger than those for 50° (note that the vertical ranges are set at the maximum SSC value of 32°). This makes sense according to the angle dispersion curves in this case (Figure 6) since 32° is a better angle than 50° to excite the S0 mode (especially at 2 MHz, the centre frequency of the transducer).

3.2.2. Immersion

Both schematic and photograph of the experimental arrangement for the immersion is shown in Figure 8. The plate is supported on small EPS blocks to reduce the effects of the reflections in the bottom of the tank [3,13]. A pair of broadband longitudinal transducers (K2SC, General Electric, Huerth, Germany), with a 2 MHz central frequency and 24 mm diameter were employed for Lamb wave excitation and detection. The longitudinal wave velocity in water is 1490 m/s [28,68]. The actuator was excited by a ten-cycle tone burst using a programmable signal generator (33120A, Agilent Technologies, Loveland, CO, USA). The excitation frequencies went from 1 MHz to 3 MHz, in 20 kHz steps. The sensed signals were amplified by 40 dB (5660C, Panametrics, Waltham, MA, USA) and then captured using a digital oscilloscope (DPO3014, Tektronix, Shanghai, China) with a 50 MHz sampling frequency and 10,000 point length (200 μ s temporal interval) and averaged 32 times. In this case, a higher sampling frequency was used (which implies a lower acquisition time) to eliminate part of the direct wave through the water that appears after the excited modes.
The chosen angles for the water arrangement were 20° and 40° (see Figure 9). It was expected to excite S0 mode with 20° and A0 mode with 40°.
In Figure 10, the combined spectrograms for the two chosen angles along with the theoretical group velocity dispersion curves and SSC are represented. For the 20° spectrogram, the S0 mode is detected only between 1.5 and 1.9 MHz (see that two frequency bands can be established in the 20° SSC, 1.9–2.3 MHz and 2.3–2.9 MHz) and there is close agreement between the A0 mode and the 40° diagram from 1 to 2.8 MHz. In the 20° SSC, other modes appear because of the width of the transmitted ultrasonic beam and the reflections produced. Moreover, it is remarkable the higher values of the 40° SSC compared to those from the 20° SSC (for that reason, the vertical range of the 20° SSC was not set at the maximum value of the 40° SSC). This is attributed to a combination of existing reflections and the large energy leakage of the excited A0 mode (40°) into the surrounding medium compared to the low energy leakage of the S0 mode (20°), since the latter presents an in-plane displacement where almost all the energy is confined inside the plate [6].

3.2.3. Conical Containers

Both schematic and photograph of the experimental arrangement for the setup of water conical containers is presented in Figure 11. These conical volumes were designed following [27,28]. They are bottomless, so the water is in contact with the plate. Putty was used around the conical containers to prevent water leakage. This arrangement helps to focus the ultrasonic beam (which translates in a smaller angular range) and reduces the reflections present in the immersion arrangement. Additionally, the maximum angle is limited by the conical container. A pair of broadband longitudinal transducers (K2SC, General Electric, Huerth, Germany) were employed for the excitation and detection of Lamb waves. The actuator was excited by a ten-cycle tone burst using a programmable signal generator (33120A, Agilent Technologies, Loveland, CO, USA) and amplified by a factor of 50 (WMA-300, Falco Systems, Amsterdam, The Netherlands). The excitation frequencies were varied from 1 MHz to 3 MHz, in 20 kHz steps. The sended signals were amplified 32 dB (AMPLUS-32, Dasel Sistemas, Madrid, Spain) and then captured using a digital oscilloscope (DPO3014, Tektronix, Shanghai, China) with a 25 MHz sampling frequency, 10,000 sampling points (400 μ s temporal interval) and 32 averaging.
The angles chosen, as in the previous case, were 20° and 40°, for the same frequency range (1–3 MHz). It was expected to excite the same modes, S0 with 20° and A0 with 40°, since the coupling medium was not modified and, therefore, the same angle curves from Figure 9 could be used.
In Figure 12 the combined spectrograms for the two chosen angles along with the theoretical group velocity dispersion curves and SSC are presented. There is good agreement with the S0 mode in the 20° diagram (1–2.5 MHz) and with the A0 mode in the 40° diagram (1–1.9 MHz). The 20° combined spectrogram for conical containers differs from the 20° immersion setup. These differences are attributable mainly to the existing reflections mentioned in Section 1 that affect the signals sensed. The trend in water coupling is maintained, the larger the incidence angle, the higher the SSC values. Additionally, the 20° SSC reach higher frequencies (2.5 MHz) than the 40° SSC (1.9 MHz).

3.2.4. Air Coupling

Both schematic and photograph of the experimental arrangement for the air coupling setup is presented in Figure 13. A block of EPS was placed between both actuator and sensor to absorb the direct wave through the air. Two pair of piezoelectric air-coupled transducers with a 32 mm diameter mm and central frequencies of 250 kHz and 500 kHz were employed for the excitation and detection of Lamb waves (information about these transducers can be found in [95,96]). The longitudinal wave velocity in air is 343 m/s [53,58]. The actuator was excited by a five-cycle tone burst using a programmable signal generator (33120A, Agilent Technologies, Loveland, CO, USA) and amplified by a factor of50 (WMA-300, Falco Systems, Amsterdam, The Netherlands). In air coupling, fewer cycles were used to minimise electrical stray coupling effects. The excitation frequencies were swept from 50 kHz to 550 kHz for the 250 kHz transducers and from 200 kHz to 800 kHz for the 500 kHz transducers, using 5 kHz steps in both cases. The sensed signals were amplified 40 dB (5660C, Panametrics, Waltham, MA, USA) and then captured using a digital oscilloscope (RTO 1004, Rohde & Schwartz, München, Germany) with a 25 MHz sampling frequency and 10,000 sampling points (400 μ s temporal interval) and averaged 32 times.
Two incidence angles were chosen to excite and detect the A0 mode with two kinds of transducers, 13.75° with 250 kHz transducers (frequency range from 50 to 550 kHz) and 10° with 500 kHz transducers (from 200 kHz to 800 kHz, see Figure 14). As can be seen, the lower the longitudinal wave velocity of the coupling medium, the lower the angular range in the dispersion curves. Due to the low central frequencies of the transducers (250 kHz and 500 kHz), only the two fundamental modes appear. Small angle increments can be attained since they are easy to manage, and no physical limitation exists, unlike in immersion and conical containers setups.
Several attempts were made to excite S0 mode, but it was not possible as no signal was sensed by the oscilloscope. One reason is that S0 particle motion is predominantly in-plane and it radiates much less energy to the air than the A0 mode (as stated in Section 1) [40,43,59,60].
In Figure 15, the combined spectrograms for the two chosen angles along with the theoretical group velocity dispersion curves and SSC are represented. There is a good correspondence with the A0 mode in the three diagrams for the frequency bands established for the SSC, from 0.17 to 0.33 MHz for the 13.75° diagram and from 0.3 to 0.65 MHz for the 10° diagram. The bandwidth and the SSC values of 10° are greater than those for the SSC of 13.75° (bear in mind that the transducers are different).

3.3. Techniques and Results on Mortar

To test the mortar plate, two techniques were selected, water conical containers and air-coupled transducers. The direct contact technique with the Plexiglas wedges was discarded for two reasons: 1) The central frequency of the wedge transducers (2 MHz) is very high to test an attenuative material like mortar [97] and 2) some modes are unable to propagate in mortar if their velocities are lower than the longitudinal wave velocity in Plexiglas because of the Snell’s law concept [26]. The Plexiglas speed (2730 m/s) is large in comparison to other coupling mediums (1490 m/s in water, 343 m/s in air). Normally, Teflon (1350 m/s) is employed as a wedge material to test cementitious materials [57,83]. Water coupling through conical containers enables the possibility of exciting both fundamental modes while air coupling enables a fast and comfortable measurement. Almost the same equipment was used. However, a lower sampling frequency (10 MHz) was chosen to enhance the temporal interval (1 ms), since the acquired signals in the mortar plate arrive later than those acquired in the metallic plate (i.e., the bulk wave velocities in mortar are lower than in the metallic plate).

3.3.1. Conical Containers

Both schematic and photograph of the experimental arrangement for the setup of water conical containers to measure the mortar plate is shown in Figure 16. A pair of broadband longitudinal transducers (K0,5SC, General Electric) with a 0.5 MHz central frequency and 24 mm diameter were employed for the excitation and detection of Lamb waves.
The actuator was excited by a ten-cycle tone burst using a programmable signal generator (33120A, Agilent Technologies, Loveland, CO, USA). The excitation frequencies were varied from 10 kHz to 1 MHz, in 5 kHz steps. The sensed signals were amplified by 40 dB (5660C, Panametrics, Waltham, MA, USA) and then captured using a digital oscilloscope (DPO3014, Tektronix, Shanghai, China) with a 10 MHz sampling frequency and 10,000 points length (a 1 ms temporal interval) and averaged 32 times.
Two incidence angles, 20° and 40°, were chosen to excite the S0 and the A0 modes (respectively) in the frequency range from 0.01 to 1 MHz (see Figure 17).
In Figure 18, an acceptable coincidence exists between the theoretical and experimental curves. In a), the S0 mode is detected between approximately 0.35 and 0.4 MHz (the bandwidth that shows the SSC goes from 0.25 to 0.55 MHz); in b) the A0 mode is detected from approximately 0.1 to 0.3 MHz (the bandwidth goes from 0.05 to 0.45 MHz). As can be seen, new modes appear in these curves in contrast to the stainless steel dispersion curves (A2, S2), the larger the thickness, the larger the number of Lamb modes. In the 20° SSC, higher amplitude and lower bandwidth are observed. Also, the 20° SSC decays at a higher frequency (0.55 MHz) than the 40° SSC (0.45 MHz), as in Section 3.2.3 (conical containers setup to test the stainless steel plate), although no comparisons should be drawn since the inspected material and transducers are different.

3.3.2. Air Coupling

Both schematic and photograph of the experimental arrangement for the air coupling setup is shown in Figure 19. The pair of piezoelectric air-coupled transducers with a 250 kHz central frequency used to measure the metallic plate were employed for the excitation and detection of Lamb waves since 250 kHz attenuates less than 500 kHz.
The actuator was excited by a five-cycle tone burst using a programmable signal generator (33120A, Agilent Technologies) and amplified by a factor of 50 (WMA-300, Falco Systems, Amsterdam, The Netherlands). The excitation frequencies were swept from 50 kHz to 450 kHz, in 2 kHz steps (201 frequencies). The sensed signals were amplified 40 dB (5660C, Panametrics, Waltham, MA, USA) and then captured using a digital oscilloscope (RTO 1004, Rohde & Schwartz, München, Germany) with a 10 MHz sampling frequency and 10,000 points length (a 1 ms temporal interval) and averaged 32 times. A 10° angle of inclination was chosen to excite the A0 mode in the mortar plate in the 50 to 450 kHz frequency range (see Figure 20).
In Figure 21, there is a good match between the theoretical and experimental curves. The A0 mode is excited in the frequency range that shows the SSC, from 0.15 to 0.3 MHz.

4. Conclusions

The main conclusions are:
1)
This contribution is a research work where different coupling techniques have been compared and analysed: the contact technique with angle beam probes, pure immersion and alternatives with water wedges, and air-coupled ultrasonic testing. First, a metallic plate was used to perform different experimentals and to achieve a solid theoretical basis. With this theoretical knowledge, a more complicated and heterogeneous material as mortar was inspected.
2)
Good matching between theoretical and experimental group velocity dispersion curves was done to determine which modes were generated in the metallic and mortar plates. Close agreement was achieved between theoretical and experimental data, which means that Lamb waves were excited and received successfully in every setup by choosing the same inclination angle in transmission and reception to enhance a particular Lamb mode.
3)
If the A0 mode is needed, air-coupled ultrasonic testing is recommended as the first option. On the other hand, if S0 mode is sought, water coupling using conical containers is suitable as a first alternative. Plexiglas wedges are also an option to excite the S0 mode if the bulk wave velocities of the tested material are higher than the bulk wave velocity of Plexiglas.
4)
System sensitivity curves (SSC), a signal processing tool that represents the bandwidth of the whole “Lamb wave” system has been of great help to analyse the experimental dispersion curves in the proper frequencies.
As future directions, improvements in the conical containers design (as increasing the maximum incidence angles) and testing of more complicated cementitious materials with new configurations (as laser interferometry) will be analysed. With the acquired theoretical knowledge, practical applications of Lamb waves in cementitious materials are studied, as carbonation and thermal damage.

Supplementary Materials

The following are available online at https://www.mdpi.com/1424-8220/19/19/4068/s1, Data file S1: Experimental data from the direct contact setup to measure the stainless steel plate (32°), Data file S2: Experimental data from the direct contact setup to measure the stainless steel plate (50°), Data file S3: Experimental data from the immersion setup to measure the stainless steel plate (20°), Data file S4: Experimental data from the immersion setup to measure the stainless steel plate (40°), Data file S5: Experimental data from the conical containers setup to measure the stainless steel plate (20°), Data file S6: Experimental data from the conical containers setup to measure the stainless steel plate (40°), Data file S7: Experimental data from the air coupling setup to measure the stainless steel plate (13.75°), Data file S8: Experimental data from the air coupling setup to measure the stainless steel plate (10°), Data file S9: Experimental data from the conical containers setup to measure the mortar plate (20°), Data file S10: Experimental data from the conical containers setup to measure the mortar plate (40°), Data file S11: Experimental data from the air coupling setup to measure the mortar plate (10°).

Author Contributions

Conceptualization, all authors; Methodology, S.V., J.G. and I.B.; Software, S.V., J.G., A.C. and C.G.; Validation, S.V., J.G. and I.B.; Formal analysis, S.V.; Investigation, S.V.; Resources, J.G., A.C. and J.P.; Data curation, S.V.; Writing—original draft preparation, S.V.; writing—review and editing, all authors; visualization, S.V.; supervision, J.G. and I.B.; project administration, J.G., I.B. and J.P.; funding acquisition, J.G, I.B. and J.P.

Funding

This research was funded by the Spanish Administration, grants number BES-2015-071958 and BIA2017-87573-C2.

Acknowledgments

The authors thank John S. Popovics, Tribikram Kundu and Tomás Gómez Álvarez-Arenas for their valuable advice on this study.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A

The Table A1 brings together all the parameters of every setup used in this study.
Table A1. Experimental parameters.
Table A1. Experimental parameters.
MaterialTechniqueTransducersSignal
Generator
Signal (tone burst)TX Amplitude (V)TX AmplificationFrequency RangeFrequency Step (kHz)Number of FrequenciesRX Amplification (dB)Acquisition EquipmentSamplesSampling
Frequency (MHz)
Time
Interval
Averaging
Stainless
steel
ContactMUBW 2N
(2 MHz)
Agilent
33120A
10 cycle10x1(1 MHz,
3MHz)
2010040DPO30141000025 400   μ s 32
ImmersionK2SC
(2 MHz)
Agilent
33120A
10 cycle10x1(1 MHz,
3MHz)
2010040DPO30141000050 200   μ s 32
Conical
containers
K2SC
(2 MHz)
Agilent
33120A
10 cycle 3   ( 20 ° ),
6   ( 40 ° )
x50(1 MHz,
3 MHz)
2010032DPO30141000025 400   μ s 32
Air coupling250 kHzAgilent
33120A
5 cycle6x50(50 kHz,
550 kHz)
510040Rohde & Schwartz
RTO 1004
1000025 400   μ s 32
500 kHzAgilent
33120A
5 cycle6x50(200 kHz,
800 kHz)
512040Rohde & Schwartz
RTO 1004
1000025 400   μ s 32
MortarConical
containers
K0,5SC
(0.5 MHz)
Agilent
33120A
10 cycle3No(10 kHz,
1 MHz)
520040DPO301410000101 ms32
Air coupling250 kHzAgilent
33120A
5 cycle6x50(50 kHz,
450 kHz)
220040Rohde & Schwartz
RTO 1004
10000101 ms32

References

  1. Rose, J.L. Ultrasonic Guided Waves in Solid Media, 1st ed.; Cambridge University Press: New York, NY, USA, 2014. [Google Scholar]
  2. Rossing, T.D.; Fletcher, N.H. Principles of Vibration and Sound, 2nd ed.; Springer-Verlag: New York, NY, USA, 2004. [Google Scholar]
  3. Sharma, S.; Mukherjee, A. Damage detection in submerged plates using ultrasonic guided waves. Sadhana 2014, 39, 1009–1034. [Google Scholar] [CrossRef]
  4. Luo, W.; Rose, J.L. Lamb Wave Thickness Measurement Potential with Angle Beam and Normal Beam Excitation. Mater. Eval. 2004, 62, 860–866. [Google Scholar]
  5. Yung, Y.C.; Kundu, T.; Ehsani, M.R. Internal Discontinuity Detection in Concrete by Lamb Waves. Mater. Eval. 2001, 59, 418–423. [Google Scholar]
  6. Su, Z.; Ye, L. Identification of Damage Using Lamb Waves: From Fundamentals to Applications, 1st ed.; Springer-Verlag: London, UK, 2009. [Google Scholar]
  7. Birgani, P.T.; Tahan, K.N.; Sodagar, S.; Shishesaz, M. Theoretical modeling of low-attenuation lamb wave modes generation in three-layer adhesive joints using angle beam transducer. Lat. Am. J. Solids Stru. 2015, 12, 461–476. [Google Scholar] [CrossRef] [Green Version]
  8. Gómez Ullate, Y.; Reynolds, P.; Mould, J. Selective excitation of Lamb Wave Modes in Thin Aluminium Plates using Bonded Piezoceramics: FEM Modelling and Measurements. In Proceedings of the 9th European Conference on NDT, Berlin, Germany, 25–29 September 2006. [Google Scholar]
  9. Seale, M.D.; Smith, B.T.; Prosser, W.H. Lamb wave assessment of fatigue and thermal damage in composites. J. Acoust. Soc. Am. 1998, 103, 2416–2424. [Google Scholar] [CrossRef]
  10. Giurgiutiu, V. Lamb Wave Generation with Piezoelectric Wafer Active Sensors for Structural Health Monitoring. In Proceedings of the SPIE’s 10th Annual International Symposium on Smart Structures and Materials, San Diego, CA, USA, 2–6 March 2002. [Google Scholar]
  11. Alleyne, D.N.; Cawley, P. Optimization of Lamb wave inspection techniques. NDT E Int. 1992, 25, 11–22. [Google Scholar] [CrossRef]
  12. Cawley, P.; Alleyne, D. The use of Lamb waves for the long range inspection of large structures. Ultrasonics 1996, 34, 287–290. [Google Scholar] [CrossRef]
  13. Sharma, S.; Mukherjee, A. A Non-Contact Technique for Damage Monitoring in Submerged Plates Using Guided Waves. J. Test. Eval. 2015, 43, 776–791. [Google Scholar] [CrossRef]
  14. Jia, X. Modal analysis of Lamb wave generation in elastic plates by liquid wedge transducers. J. Acoust. Soc. Am. 1997, 101, 834–842. [Google Scholar] [CrossRef]
  15. Pitch-Catch Air-Coupled Ultrasonic Technique for Detection of Barely Visible Impact Damages in Composite Laminates. In Proceedings of the 7th European Workshop on Structural Health Monitoring, Nantes, France, 8–11 July 2014.
  16. Shelke, A.; Kundu, T.; Amjad, U.; Hahn, K.; Grill, W. Mode-Selective Excitation and Detection of Ultrasonic Guided Waves for Delamination Detection in Laminated Aluminum Plates. IEEE T. Ultrason. Ferr. 2011, 58, 567–577. [Google Scholar] [CrossRef]
  17. Chen, J.; Su, Z.; Cheng, L. Identification of corrosion damage in submerged structures using fundamental anti-symmetric Lamb waves. Smart Mater. Struct. 2010, 19, 015004. [Google Scholar] [CrossRef]
  18. Masurkar, F.A.; Yelve, N.P. Locating a Damage in an Aluminium Plate using Lamb Waves. In Proceedings of the Indian National Seminar & Exhibition on Non-Destructive Evaluation, Hyderabad, India, 26–28 November 2015. [Google Scholar]
  19. Ham, S.; Popovics, J.S. Fast Lamb Wave Energy Shift Approach Using Fully Contactless Ultrasonic System to Characterize Concrete Structures. In Proceedings of the 41st Annual Review of Progress in Quantitative Nondestructive Evaluation, Boise, ID, USA, 20–25 July 2015. [Google Scholar]
  20. Rajana, K.M.; Cho, Y.; Rose, J.L. Utility of Lamb Waves for Near Surface Crack Detection. In Review of Progress in Quantitative Nondestructive Evaluation; Thompson, D.O., Chimenti, D.E., Eds.; Springer: Boston, MA, USA, 1996; pp. 247–252. [Google Scholar] [Green Version]
  21. Benz, R.; Niethammer, M.; Hurlebaus, S.; Jacobs, L.J. Localization of notches with Lamb waves. J. Acoust. Soc. Am. 2003, 114, 677–685. [Google Scholar] [CrossRef] [PubMed]
  22. Beadle, B.M.; Hurlebaus, S.; Jacobs, L.J.; Gaul, L. Detection and Localization of Small Notches in Plates Using Lamb Waves. In Proceedings of the 23rd Conference and Exposition on Structural Dynamics (IMAC–XXIII), Orlando, FL, USA, 31 January–3 February 2005. [Google Scholar]
  23. Singh, R.; Sharma, D.S.; Sharma, M.S. Guided Waves for Damage Monitoring in Plates with Corrosion Defects. Int. J. Innov. Res. Sci. Eng. Technol. 2014, 3, 10072–10083. [Google Scholar]
  24. Guo, N.; Cawley, P. The interaction of Lamb waves with delaminations in composite laminates. J. Acoust. Soc. Am. 1993, 94, 2240–2246. [Google Scholar] [CrossRef]
  25. Xu, K.; Ta, D.; Moilanen, P.; Wang, W. Mode separation of Lamb waves based on dispersion compensation method. J. Acoust. Soc. Am. 2012, 131, 2714–2722. [Google Scholar] [CrossRef]
  26. Viktorov, I.A. Rayleigh and Lamb Waves: Physical Theory and Applications, 1st ed.; Springer: Boston, MA, USA, 1967. [Google Scholar]
  27. Guo, D.; Kundu, T. A new transducer holder mechanism for pipe inspection. J. Acoust. Soc. Am. 2001, 110, 303–309. [Google Scholar] [CrossRef]
  28. Ghosh, T.; Kundu, T.; Karpur, P. Efficient use of Lamb modes for detecting defects in large plates. Ultrasonics 1998, 36, 791–801. [Google Scholar] [CrossRef]
  29. Giurgiutiu, V.; Cuc, A. Embedded Non-destructive Evaluation for Structural Health Monitoring, Damage Detection, and Failure Prevention. Shock Vib. Dig. 2005, 37, 83–105. [Google Scholar] [CrossRef]
  30. Giurgiutiu, V. Tuned Lamb Wave Excitation and Detection with Piezoelectric Wafer Active Sensors for Structural Health Monitoring. J. Intell. Mater. Syst. Struct. 2005, 16, 291–305. [Google Scholar] [CrossRef]
  31. Giurgiutiu, V. Structural Health Monitoring with Piezoelectric Wafer Active Sensors, 2nd ed.; Academic Press: Waltham, MA, USA, 2014. [Google Scholar]
  32. Rguiti, M.; Grondel, S.; El youbi, F.; Courtois, C.; Lippert, M.; Leriche, A. Optimized piezoelectric sensor for a specific application: Detection of Lamb waves. Sens. Actuators A Phys. 2006, 126, 362–368. [Google Scholar] [CrossRef]
  33. Kessler, S.S.; Spearing, S.M.; Soutis, C. Damage detection in composite materials using Lamb wave methods. Smart Mater. Struct. 2002, 11, 269–278. [Google Scholar] [CrossRef] [Green Version]
  34. Mustapha, S.; Ye, L. Bonding Piezoelectric Wafers for Application in Structural Health Monitoring–Adhesive Selection. Res. Nondestruct. Eval. 2015, 26, 23–42. [Google Scholar] [CrossRef]
  35. Sun, H.; Wang, Y.; Qing, X.; Wu, Z. High Strain Survivability of Piezoceramics by Optimal Bonding Adhesive Design. Sensors 2018, 18, 2554. [Google Scholar] [CrossRef] [PubMed]
  36. Alleyne, D.; Cawley, P. A two-dimensional Fourier transform method for the measurement of propagating multimode signals. J. Acoust. Soc. Am. 1991, 89, 1159–1168. [Google Scholar] [CrossRef]
  37. Bezdek, M.; Joseph, K.; Tittmann, B.R. Low Attenuation Waveguide for Structural Health Monitoring with Leaky Surface Waves. J. Korean Soc. Nondestruct. Test. 2012, 32, 241–262. [Google Scholar] [CrossRef] [Green Version]
  38. Kim, Y.H.; Song, S.-J.; Kwon, S.-D. Measurement of phase velocity dispersion curves and group velocities in a plate using leaky Lamb waves. In Proceedings of the National Seminar of ISNT, Chennai, Tamil Nadu, India, 5–7 December 2002. [Google Scholar]
  39. Harb, M.S.; Yuan, F.G. A rapid, fully non-contact, hybrid system for generating Lamb wave dispersion curves. Ultrasonics 2015, 61, 62–70. [Google Scholar] [CrossRef] [PubMed]
  40. Castaings, M.; Cawley, P. The generation, propagation, and detection of Lamb waves in plates using air-coupled ultrasonic transducers. J. Acoust. Soc. Am. 1996, 100, 3070–3077. [Google Scholar] [CrossRef]
  41. Theobald, P.; Zeqiri, B.; Avison, J. Couplants and their Influence on AE sensor sensitivity. J. Acoust. Emiss. 2008, 26, 91–97. [Google Scholar]
  42. Alleyne, D.N.; Cawley, P. The Interaction of Lamb Waves with Defects. IEEE Trans. Ultrason. Ferroelectr. 1992, 39, 381–397. [Google Scholar] [CrossRef]
  43. Mažeika, L.; Raišutis, R.; Maciulevičius, A.; Žukauskas, E.; Kazys, R.; Seniūnas, G.; Vladišauskas, A. Comparison of several techniques of ultrasonic Lamb waves velocities measurements. Ultragarsas (Ultrasound) 2009, 64, 11–17. [Google Scholar]
  44. Lee, K.I.; Yoon, S.W. Feasibility of bone assessment with leaky Lamb waves in bone phantoms and a bovine tibia. J. Acoust. Soc. Am. 2004, 115, 3210–3217. [Google Scholar] [CrossRef] [PubMed]
  45. Alleyne, D.N. The nondestructive testing of plates using ultrasonic Lamb waves. PhD Thesis, Imperial College of Science, Technology and Medicine - University of London, London, UK, 1991. [Google Scholar]
  46. Lee, Y.C.; Cheng, S.W. Measuring Lamb Wave Dispersion Curves of a Bi-Layered Plate and Its Application on Material Characterization of Coating. IEEE Trans. Ultrason. Ferroelectr. 2001, 48, 830–837. [Google Scholar]
  47. Takiy, A.E.; Granja, S.C.G.; Higuti, R.T.; Kitano, C.; Elvira, L.; Martínez-Graullera, O.F. Low attenuation frequency bands for Lamb waves immersed in viscous fluids: theoretical analysis and experimental validation. In Proceedings of the International Congress on Ultrasonics (ICU), Singapore, 2–5 May 2013. [Google Scholar]
  48. Karpur, P.; Benson, D.M.; Matikas, T.E.; Kundu, T.; Nicolaou, P.D. An Approach to Determine the Experimental Transmitter-Receiver Geometry for the Reception of Leaky Lamb Waves. Mater. Eval. 1995, 53, 1348–1352. [Google Scholar]
  49. Goueygou, M.; Abraham, O.; Lataste, J.-F. A comparative study of two non-destructive testing methods to assess near-surface mechanical damage in concrete structures. NDT E Int. 2008, 41, 448–456. [Google Scholar] [CrossRef]
  50. Briers, R.; Leroy, O.; Shkerdin, G. A liquid wedge as generating technique for Lamb and Rayleigh waves. J. Acoust. Soc. Am. 1997, 102, 2117–2124. [Google Scholar] [CrossRef]
  51. Padiyar, M.J.; Hariharan, N.; Maheswaran, M.; Balasubramaniam, K.; Bose, A.C.; Giridharan, N.V. Increase in efficiency of air-coupled L-scan (Lamb wave scan) and its application in inspection of composite structures. In Proceedings of the National Seminar & Exhibition on Non-Destructive Evaluation, Chennai, India, 8–10 December 2011. [Google Scholar]
  52. Hayashi, T.; Kawashima, K. Single Mode Extraction from Multiple Modes of Lamb Wave and Its Application to Defect Detection. JSME Int. J. Ser. A 2003, 46, 620–626. [Google Scholar] [CrossRef] [Green Version]
  53. Gosálbez, J.; Wright, W.M.D.; Jiang, W.; Carrión, A.; Genovés, V.; Bosch, I. Airborne non-contact and contact broadband ultrasounds for frequency attenuation profile estimation of cementitious materials. Ultrasonics 2018, 88, 148–156. [Google Scholar] [CrossRef]
  54. Solodov, I.; Döring, D.; Busse, G. Air-Coupled Lamb and Rayleigh Waves for Remote NDE of Defects and Material Elastic Properties. J. Mech. Eng. 2010, 56, 557–564. [Google Scholar]
  55. Schindel, D.W.; Hutchins, D.A.; Grandia, W.A. Capacitive and piezoelectric air-coupled transducers for resonant ultrasonic inspection. Ultrasonics 1996, 34, 621–627. [Google Scholar] [CrossRef]
  56. Schindel, D.W.; Hutchins, D.A. Through-thickness characterization of solids by wideband air-coupled ultrasound. Ultrasonics 1995, 33, 11–17. [Google Scholar] [CrossRef]
  57. In, C.-W.; Kim, J.-Y.; Kurtis, K.E.; Jacobs, L.J. Characterization of ultrasonic Rayleigh surface waves in asphaltic concrete. NDT E Int. 2009, 42, 610–617. [Google Scholar] [CrossRef]
  58. Fan, Z.; Jiang, W.; Wright, W.M.D. Non-contact ultrasonic gas flow metering using air-coupled leaky Lamb waves. Ultrasonics 2018, 89, 74–83. [Google Scholar] [CrossRef] [PubMed]
  59. Fan, Z.; Jiang, W.; Cai, M.; Wright, W.M.D. The effects of air gap reflections during air-coupled leaky Lamb wave inspection of thin plates. Ultrasonics 2016, 65, 282–295. [Google Scholar] [CrossRef] [PubMed]
  60. Demčenko, A.; Žukauskas, A.D.E.; Kazys, R.; Voleisis, A. Investigation of interaction of the Lamb wave with delamination type defect in GLARE composite using air-coupled ultrasonic technique. In Proceedings of the Forum Acusticum Budapest: 4th European Congress on Acoustics, Budapest, Hungary, 29 August–2 September 2005. [Google Scholar]
  61. Zhao, M.; Zeng, L.; Lin, J.; Wu, W. Mode identification and extraction of broadband ultrasonic guided waves. Meas. Sci. Technol. 2014, 25, 115005. [Google Scholar] [CrossRef]
  62. Ryden, N.; Park, C.B.; Ulriksen, P.; Miller, R.D. Lamb wave analysis for non-destructive testing of concrete plate structures. In Proceedings of the 9th European Meeting of Environmental and Engineering Geophysics (EEGS-ES / EAGE Near Surface Geoscience Division annual meeting), Prague, Czech Republic, 31 August–4 September 2003; pp. 782–793. [Google Scholar]
  63. Lee, Y.H.; Oh, T. The Simple Lamb Wave Analysis to Characterize Concrete Wide Beams by the Practical MASW Test. Materials 2016, 9, 437. [Google Scholar] [CrossRef] [PubMed]
  64. Karpenko, O. Signal Analysis in Guided Wave Structural Health Monitoring. Master’s Thesis, Michigan State University, East Lansing, MI, USA, 2013. [Google Scholar]
  65. Moll, J.; Fritzen, C.-P. Advanced Aspects of Mode-Selective Excitation of Ultrasonic Guided Waves. In Proceedings of the ISMA 2010: International Conference on Noise and Vibration Engineering including USD2010, Leuven, Belgium, 20-22 September 2010. [Google Scholar]
  66. Ullate, Y.G. Estudio de Sistemas Ultrasónicos Basados en Multitransductores para la Detección de Defectos en Estructuras Tipo Placa. Ph.D. Thesis, Universidad Politécnica de Madrid, Madrid, Spain, 2007. [Google Scholar]
  67. Schaal, C.; Samajder, H.; Baid, H.; Mal, A. Rayleigh to Lamb wave conversion at a delamination-like crack. J. Sound Vib. 2015, 353, 150–163. [Google Scholar] [CrossRef]
  68. Zhang, R.; Wan, M.; Cao, W. Parameter Measurement of Thin Elastic Layers Using Low-Frequency Multi-Mode Ultrasonic Lamb Waves. IEEE Trans. Instrum. Meas. 2001, 50, 1397–1403. [Google Scholar] [CrossRef]
  69. Zeroug, S.; Froelich, B. Ultrasonic leaky-Lamb wave imaging through a highly contrasting layer. In Proceedings of the IEEE Symposium on Ultrasonics, Honolulu, HI, USA, 5–8 October 2003; pp. 794–798. [Google Scholar]
  70. Bezdek, M.; Tittmann, B.R.; Joseph, K.L. Fluid-Coupled Waveguide for Leaky Guided Waves. In Proceedings of the 2006 IEEE Ultrasonics Symposium, Vancouver, BC, Canada, 3–6 October 2006. [Google Scholar]
  71. Balasubramaniam, K.; Rose, J.L. Physically Based Dispersion Curve Feature Analysis in the NDE of Composites. Res. Nondestruct. Eval. 1991, 3, 41–67. [Google Scholar] [CrossRef]
  72. Niethammer, M.; Jacobs, L.J.; Qu, J.; Jarzynski, J. Time-frequency representations of Lamb waves. J. Acoust. Soc. Am. 2001, 109, 1841–1847. [Google Scholar] [CrossRef]
  73. Kuttig, H.; Niethammer, M.; Hurlebaus, S.; Jacobs, L.J. Model-based analysis of dispersion curves using chirplets. J. Acoust. Soc. Am. 2006, 119, 2122–2130. [Google Scholar] [CrossRef]
  74. Prosser, W.H.; Seale, M.D.; Smith, B.T. Time-frequency analysis of the dispersion of Lamb modes. J. Acoust. Soc. Am. 1999, 105, 2669–2676. [Google Scholar] [CrossRef] [PubMed]
  75. Zima, B.; Rucka, M. Application of wavelet transform in analysis of guided wave propagation signals for damage detection in a steel plate. Diagnostyka 2015, 16, 43–48. [Google Scholar]
  76. Ray, P.; Manogharan, P.; Srinivasan, B.; Balasubramaniam, K.; Rajagopal, P. Novel Method of Defect Identification in Bent Structures through Feature-Guided Wave Detection using Fiber Bragg Grating Sensors. In Proceedings of the International Conference on Optical Fibre Sensors (OFS24), Curitiba, Brazil, 28 September–2 October 2015; Kalinowski, H.J., Fabris, J.L., Bock, W.J., Eds.; [Google Scholar]
  77. Genovés, V.; Gosálbez, J.; Carrión, A.; Miralles, R.; Payá, J. Optimized ultrasonic attenuation measures for non-homogeneous materials. Ultrasonics 2016, 65, 345–352. [Google Scholar] [CrossRef] [PubMed]
  78. Alleyne, D.N.; Cawley, P. The Excitation of Lamb Waves in Pipes Using Dry-Coupled Piezoelectric Transducers. J. Nondestruct. Eval. 1996, 15, 11–20. [Google Scholar] [CrossRef]
  79. Hora, P.; Cervená, O. Determination of Lamb wave dispersion curves by means of Fourier transform. Appl. Comput. Mech. 2012, 6, 5–16. [Google Scholar]
  80. Ruiz, A.; Ortiz, N.; Medina, A.; Kim, J.-Y.; Jacobs, L.J. Application of ultrasonic methods for early detection of thermal damage in 2205 duplex stainless steel. NDT E Int. 2013, 54, 19–26. [Google Scholar] [CrossRef]
  81. Grondel, S.; Assaad, J.; Delebarre, C.; Blanquet, P.; Moulin, E. The propagation of Lamb waves in multilayered plates: phase-velocity measurement. Meas. Sci. Technol. 1999, 10, 348–353. [Google Scholar] [CrossRef]
  82. Goueygou, M.; Piwakowski, B. Radiation of surface waves into concrete by means of a wedge transducer: design and optimization. In Proceedings of the 16th World Conference on NDT (WCNDT), Montreal, QC, Canada, 30 August–3 September 2004. [Google Scholar]
  83. Goueygou, M.; Piwakowski, B.; Fnine, A.; Kaczmarek, M.; Buyle-Bodin, F. NDE of two-layered mortar samples using high-frequency Rayleigh waves. Ultrasonics 2004, 42, 889–895. [Google Scholar] [CrossRef]
  84. Kim, G.; Kim, J.-Y.; Kurtis, K.E.; Jacobs, L.J.; Le Pape, Y.; Guimaraes, M. Quantitative evaluation of carbonation in concrete using nonlinear ultrasound. Mater Struct 2016, 49, 399–409. [Google Scholar] [CrossRef]
  85. Deán, J.L.; Trillo, C.; López-Vázquez, J.C.; Doval, Á.F.; Fernández, J.L. Determination of the frequency spectrum of Lamb waves from a sequence of maps of the instantaneous acoustic displacement obtained with TV holography. In Proceedings of the SPIR 7389, Optical Measurement Systems for Industrial Inspection VI, Munich, Germany, 15–17 June 2009; Lehmann, P.H., Ed.; [Google Scholar]
  86. Carrión, A.; Genovés, V.; Pérez, G.; Payá, J.; Gosálbez, J. Flipped Accumulative Non-Linear Single Impact Resonance Acoustic Spectroscopy (FANSIRAS): A novel feature extraction algorithm for global damage assessment. J. Sound Vib. 2018, 432, 454–469. [Google Scholar] [CrossRef]
  87. Jain, A.; Greve, D.W.; Oppenheim, I.J. A MEMS Ultrasonic Transducer for Monitoring of Steel Structures. In Proceedings of the SPIE’s 9th Annual International Symposium on Smart Structures and Materials, San Diego, CA, USA, July 2002. [Google Scholar]
  88. Greve, D.W.; Oppenheim, I.J.; Zheng, P. Lamb Waves and Nearly-Longitudinal Waves in Thick Plates. In Proceedings of the Sensors and Smart Structures Technologies for Civil, Mechanical, and Aerospace Systems, San Diego, CA, USA, 8 April 2008; Tomizuka, M., Ed.; SPIE: Bellingham, WA, USA, 2008. [Google Scholar]
  89. Greve, D.W.; Zheng, P.; Oppenheim, I.J. Transition from Lamb waves to longitudinal waves in plates. Smart Mater. Struct. 2008, 17, 11. [Google Scholar] [CrossRef]
  90. Li, F.; Murayama, H.; Kageyama, K.; Shirai, T. Guided Wave and Damage Detection in Composite Laminates Using Different Fiber Optic Sensors. Sensors 2009, 9, 4005–4021. [Google Scholar] [CrossRef] [PubMed]
  91. Yan-Xun, X.; Fu-Zhen, X.; Ming-Xi, D. Evaluation of Thermal Degradation Induced Material Damage Using Nonlinear Lamb Waves. Chin. Phys. Lett. 2010, 27, 016202. [Google Scholar] [CrossRef]
  92. UNE-EN 196-1:2005. Métodos de Ensayo de Cementos; Parte 1: Determinación de resistencias mecánicas; Normalización Española: Madrid, Spain, 2005. [Google Scholar]
  93. Carrión, A.; Genovés, V.; Gosálbez, J.; Miralles, R.; Payá, J. Ultrasonic signal modality: A novel approach for concrete damage evaluation. Cem. Concr. Res. 2017, 101, 25–32. [Google Scholar] [CrossRef] [Green Version]
  94. Goueygou, M.; Lafhaj, Z.; Soltani, F. Assessment of porosity of mortar using ultrasonic Rayleigh waves. NDT E Int. 2009, 42, 353–360. [Google Scholar] [CrossRef]
  95. Sancho-Knapik, D.; Calás, H.; Peguero-Pina, J.J.; Ramos Fernandez, A.; Gil-Pelegrín, E.; Álvarez-Arenas, T.E.G. Air-coupled Ultrasonic Resonant Spectroscopy for the Study of the Relationship Between Plant Leaves’ Elasticity and Their Water Content. IEEE Trans. Ultrason. Ferroelectr. 2012, 59, 319–325. [Google Scholar] [CrossRef]
  96. Álvarez-Arenas, T.E.G. Acoustic Impedance Matching of Piezoelectric Transducers to the Air. IEEE Trans. Ultrason. Ferroelectr. 2004, 51, 624–633. [Google Scholar] [CrossRef]
  97. Neuenschwander, J.; Schmidt, Th.; Lüthi, Th.; Romer, M. Leaky Rayleigh wave investigation on mortar samples. Ultrasonics 2006, 45, 50–55. [Google Scholar] [CrossRef]
Figure 1. Particle displacement of (a) antisymmetric and (b) symmetric Lamb wave modes. The discontinuous line represents the middle of the plate.
Figure 1. Particle displacement of (a) antisymmetric and (b) symmetric Lamb wave modes. The discontinuous line represents the middle of the plate.
Sensors 19 04068 g001
Figure 2. An example of Lamb wave phase velocity dispersion curves for a mortar plate ( c L = 4779 m/s, c T   = 2872 m/s, 2 h = 13 mm). The three non-dispersive velocities (longitudinal (1), shear (2) and Rayleigh (3) wave velocities, continuous lines) and the first cut-off frequency (4) (discontinuous line) are highlighted. Antisymmetric modes are represented by solid lines and symmetric modes are represented by dotted lines.
Figure 2. An example of Lamb wave phase velocity dispersion curves for a mortar plate ( c L = 4779 m/s, c T   = 2872 m/s, 2 h = 13 mm). The three non-dispersive velocities (longitudinal (1), shear (2) and Rayleigh (3) wave velocities, continuous lines) and the first cut-off frequency (4) (discontinuous line) are highlighted. Antisymmetric modes are represented by solid lines and symmetric modes are represented by dotted lines.
Sensors 19 04068 g002
Figure 3. Dispersion curves for a stainless steel plate ( c L = 5851 m/s, c T   = 3056 m/s, 2 h = 1.1 mm): (a) Phase velocity, c p ; (b) Group velocity, c g ; (c) Incident angle, θ , for air (343 m/s); (d) Wavenumber, k .
Figure 3. Dispersion curves for a stainless steel plate ( c L = 5851 m/s, c T   = 3056 m/s, 2 h = 1.1 mm): (a) Phase velocity, c p ; (b) Group velocity, c g ; (c) Incident angle, θ , for air (343 m/s); (d) Wavenumber, k .
Sensors 19 04068 g003
Figure 4. Diagram of the system sensitivity curve (SSC) and the combined spectrogram generation. The transmission and reception stages include part of the employed equipment (amplifiers, signal generator, oscilloscope, etc).
Figure 4. Diagram of the system sensitivity curve (SSC) and the combined spectrogram generation. The transmission and reception stages include part of the employed equipment (amplifiers, signal generator, oscilloscope, etc).
Sensors 19 04068 g004
Figure 5. Direct contact setup with variable angle beam transducers to measure the stainless steel plate: (a) Schematic; (b) Photograph.
Figure 5. Direct contact setup with variable angle beam transducers to measure the stainless steel plate: (a) Schematic; (b) Photograph.
Sensors 19 04068 g005aSensors 19 04068 g005b
Figure 6. Incidence angle dispersion curves for a stainless steel plate and Plexiglas as the coupling medium ( c =   2730 m/s). The chosen angles (32°, 50°) are represented by black discontinuous curves.
Figure 6. Incidence angle dispersion curves for a stainless steel plate and Plexiglas as the coupling medium ( c =   2730 m/s). The chosen angles (32°, 50°) are represented by black discontinuous curves.
Sensors 19 04068 g006
Figure 7. Combined spectrograms and system sensitivity curves (SSC) from the contact technique setup to measure the stainless steel plate: (a) 32°; (b) 50°.
Figure 7. Combined spectrograms and system sensitivity curves (SSC) from the contact technique setup to measure the stainless steel plate: (a) 32°; (b) 50°.
Sensors 19 04068 g007
Figure 8. Immersion setup to measure the stainless-steel plate: (a) Schematic; (b) Photograph.
Figure 8. Immersion setup to measure the stainless-steel plate: (a) Schematic; (b) Photograph.
Sensors 19 04068 g008
Figure 9. Incidence angle dispersion curves for a stainless-steel plate and water as the coupling medium ( c =   1490 m/s). The chosen angles (20°, 40°) are represented by black discontinuous curves.
Figure 9. Incidence angle dispersion curves for a stainless-steel plate and water as the coupling medium ( c =   1490 m/s). The chosen angles (20°, 40°) are represented by black discontinuous curves.
Sensors 19 04068 g009
Figure 10. Combined spectrograms and sensitivity system curves (SSC) from the immersion setup to measure the stainless-steel plate: (a) 20°; (b) 40°.
Figure 10. Combined spectrograms and sensitivity system curves (SSC) from the immersion setup to measure the stainless-steel plate: (a) 20°; (b) 40°.
Sensors 19 04068 g010
Figure 11. Conical containers setup to measure the stainless-steel plate: (a) Schematic; (b) Photograph.
Figure 11. Conical containers setup to measure the stainless-steel plate: (a) Schematic; (b) Photograph.
Sensors 19 04068 g011
Figure 12. Combined spectrograms and system sensitivity curves (SSC) from the setup of conical containers to measure the stainless plate: (a) 20°; (b) 40°.
Figure 12. Combined spectrograms and system sensitivity curves (SSC) from the setup of conical containers to measure the stainless plate: (a) 20°; (b) 40°.
Sensors 19 04068 g012
Figure 13. Air coupling setup.to measure the stainless-steel plate: (a) Schematic; (b) Photograph.
Figure 13. Air coupling setup.to measure the stainless-steel plate: (a) Schematic; (b) Photograph.
Sensors 19 04068 g013
Figure 14. Incidence angle dispersion curves for a stainless-steel plate and air as the coupling medium ( c =   343 m/s). The chosen angles (10°, 13.75°) are represented by black discontinuous curves.
Figure 14. Incidence angle dispersion curves for a stainless-steel plate and air as the coupling medium ( c =   343 m/s). The chosen angles (10°, 13.75°) are represented by black discontinuous curves.
Sensors 19 04068 g014
Figure 15. Combined spectrograms and system sensitivity curves (SSC) from the air coupling setup to measure the stainless-steel plate: (a) 13.75°, 250 kHz transducers; (b) 10°, 500 kHz transducers.
Figure 15. Combined spectrograms and system sensitivity curves (SSC) from the air coupling setup to measure the stainless-steel plate: (a) 13.75°, 250 kHz transducers; (b) 10°, 500 kHz transducers.
Sensors 19 04068 g015
Figure 16. Conical containers setup to measure the mortar plate: (a) Schematic; (b) Photograph.
Figure 16. Conical containers setup to measure the mortar plate: (a) Schematic; (b) Photograph.
Sensors 19 04068 g016
Figure 17. Incidence angle dispersion curves for a mortar plate and water as the coupling medium ( c =   1490 m/s). The chosen angles (20°, 40°) are represented by black discontinuous curves.
Figure 17. Incidence angle dispersion curves for a mortar plate and water as the coupling medium ( c =   1490 m/s). The chosen angles (20°, 40°) are represented by black discontinuous curves.
Sensors 19 04068 g017
Figure 18. Combined spectrograms and system sensitivity curves (SSC) from the conical containers setup to measure the mortar plate: (a) 20°; (b) 40°.
Figure 18. Combined spectrograms and system sensitivity curves (SSC) from the conical containers setup to measure the mortar plate: (a) 20°; (b) 40°.
Sensors 19 04068 g018
Figure 19. Air coupling setup to measure the mortar plate: (a) Schematic; (b) Photograph.
Figure 19. Air coupling setup to measure the mortar plate: (a) Schematic; (b) Photograph.
Sensors 19 04068 g019
Figure 20. Incidence angle dispersion curves for a mortar plate and air as the coupling medium ( c =   343 m/s). The chosen angle (10°) is represented by a black discontinuous curve.
Figure 20. Incidence angle dispersion curves for a mortar plate and air as the coupling medium ( c =   343 m/s). The chosen angle (10°) is represented by a black discontinuous curve.
Sensors 19 04068 g020
Figure 21. Combined spectrogram and system sensitivity curves (SSC) from the air coupling setup to measure the mortar plate: 10°.
Figure 21. Combined spectrogram and system sensitivity curves (SSC) from the air coupling setup to measure the mortar plate: 10°.
Sensors 19 04068 g021
Table 1. Specimen data.
Table 1. Specimen data.
ParametersStainless SteelMortar
Dimensions (length × width × thickness) (mm3)530 × 27 × 1.1240 × 60 × 5
Longitudinal wave velocity c L (m/s)58514779
Transverse wave velocity c T (m/s)30562872

Share and Cite

MDPI and ACS Style

Vázquez, S.; Gosálbez, J.; Bosch, I.; Carrión, A.; Gallardo, C.; Payá, J. Comparative Study of Coupling Techniques in Lamb Wave Testing of Metallic and Cementitious Plates. Sensors 2019, 19, 4068. https://doi.org/10.3390/s19194068

AMA Style

Vázquez S, Gosálbez J, Bosch I, Carrión A, Gallardo C, Payá J. Comparative Study of Coupling Techniques in Lamb Wave Testing of Metallic and Cementitious Plates. Sensors. 2019; 19(19):4068. https://doi.org/10.3390/s19194068

Chicago/Turabian Style

Vázquez, Santiago, Jorge Gosálbez, Ignacio Bosch, Alicia Carrión, Carles Gallardo, and Jordi Payá. 2019. "Comparative Study of Coupling Techniques in Lamb Wave Testing of Metallic and Cementitious Plates" Sensors 19, no. 19: 4068. https://doi.org/10.3390/s19194068

APA Style

Vázquez, S., Gosálbez, J., Bosch, I., Carrión, A., Gallardo, C., & Payá, J. (2019). Comparative Study of Coupling Techniques in Lamb Wave Testing of Metallic and Cementitious Plates. Sensors, 19(19), 4068. https://doi.org/10.3390/s19194068

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop