Next Article in Journal
Advances in Wearable Sensors: Signalling the Provenance of Garments Using Radio Frequency Watermarks
Next Article in Special Issue
Wearable and Nearable Biosensors and Systems for Healthcare
Previous Article in Journal
Quantitative X-ray Channel-Cut Crystal Diffraction Wavefront Metrology Using the Speckle Scanning Technique
Previous Article in Special Issue
Machine-Learning Analysis of Voice Samples Recorded through Smartphones: The Combined Effect of Ageing and Gender
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Thin-Film Flexible Wireless Pressure Sensor for Continuous Pressure Monitoring in Medical Applications

1
Smart Sensors Lab, School of Medicine, National University of Ireland Galway, Galway H91 TK33, Ireland
2
School of Physics, National University of Ireland Galway, Galway H91 TK33, Ireland
3
Lambe Institute of Translational Research, National University of Ireland Galway, Galway H91 TK33, Ireland
*
Author to whom correspondence should be addressed.
Sensors 2020, 20(22), 6653; https://doi.org/10.3390/s20226653
Submission received: 26 October 2020 / Revised: 15 November 2020 / Accepted: 17 November 2020 / Published: 20 November 2020
(This article belongs to the Special Issue Wearable and Nearable Biosensors and Systems for Healthcare)

Abstract

:
Physiological pressure measurement is one of the most common applications of sensors in healthcare. Particularly, continuous pressure monitoring provides key information for early diagnosis, patient-specific treatment, and preventive healthcare. This paper presents a thin-film flexible wireless pressure sensor for continuous pressure measurement in a wide range of medical applications but mainly focused on interface pressure monitoring during compression therapy to treat venous insufficiency. The sensor is based on a pressure-dependent capacitor ( C ) and printed inductive coil ( L ) that form an inductor-capacitor (LC) resonant circuit. A matched reader coil provides an excellent coupling at the fundamental resonance frequency of the sensor. Considering varying requirements of venous ulceration, two versions of the sensor, with different sizes, were finalized after design parameter optimization and fabricated using a cost-effective and simple etching method. A test setup consisting of a glass pressure chamber and a vacuum pump was developed to test and characterize the response of the sensors. Both sensors were tested for a narrow range (0–100 mmHg) and a wide range (0–300 mmHg) to cover most of the physiological pressure measurement applications. Both sensors showed good linearity with high sensitivity in the lower pressure range <100 mmHg, providing a wireless monitoring platform for compression therapy in venous ulceration.

1. Introduction

Physiological pressure, including intraocular, intracranial, and cardiovascular pressure, is a key parameter for the assessment of human health and provides opportunities for early diagnosis, personalized therapy, and preventive healthcare [1]. Pressure monitoring has been used in diagnosing lower limb problems, muscle rehabilitation, and wound monitoring [1,2,3,4]. A common medical application of non-invasive pressure sensing is the monitoring of compression therapy to treat venous leg ulcers. Venous insufficiency occurs when blood is unable to return to the heart and accumulates in the lower limbs. Chronic venous insufficiency (CVI) may cause swelling, pain, edema, and ulcerations [5,6]. The most effective treatment for CVI is compression therapy, in which a compression bandage is used to apply gradual pressure between the ankle and knee to improve the circulation of blood in the lower limb [7,8]. The typical pressure range for compression therapy is between 10 and 50 mmHg, where the bandage pressure has direct impact on the healing of ulcer [4,9]. To improve the healing process of venous ulcers, continuous monitoring of applied pressure is essential and has become the focus of current research and commercial solutions. Clinical evidence suggests that compression therapy becomes more effective with a feedback sensing system. This feedback system is achieved by using a pressure sensor. Existing solutions that are commonly used in clinical practice are accurate and robust, but they are mostly tethered, rigid, bulky, and require an additional power supply [2,10,11].
The need for wireless, small scale, lightweight, and mobile sensing solutions has led current research to focus on miniaturized thin-film and microelectromechanical (MEMS) pressure sensing devices [12,13]. Current pressure monitoring technologies are generally based on either pneumatic, fluid-filled, piezoelectric, resistive, or capacitive working principles [14,15]. In a pneumatic pressure sensing system, the force of compression bandage is transferred to air pressure and later this air pressure is converted into an electrical signal for further processing [16]. Pneumatic sensors are cheap, flexible, and thin but they are not suitable for dynamic pressure applications and are prone to temperature drift and hysteresis [7,15]. Fluid-filled pressure sensors are similar to pneumatic pressure sensors, where water or oil is used instead of air [17]. The main drawbacks of a fluid-filled sensing system are the air bubbles in the fluid, leakage risk, and bulkiness [16]. In the piezoelectric sensing technology, when pressure is applied on a piezoelectric material, it gets polarized and generates a voltage differential across the device. The piezoelectric effect is proportional to the applied pressure on the device. Thin-film piezoelectric pressure sensors are used for arterial pulse monitoring, respiratory rate, and integrated with a catheter for intravascular pressure measurements, and biomedical implants [18,19,20]. Piezoelectric pressure sensors are self-powered, low cost, and good for dynamic pressure applications, but they are not suitable for static pressure measurements due to current leakages [21,22]. In resistive pressure sensing technology, the contact area between the active thin-film resistive layer and the electrodes changes with the applied pressure so the effective resistance of the sensor changes [23,24]. Resistive pressure sensors are easier to fabricate, faster in response, and less expensive than piezoelectric pressure sensors; however, an active power source with additional adapting circuitry is required to enable pressure sensing and they are very sensitive to temperature [25]. In capacitive pressure sensing technology, the distance between capacitor electrodes is a function of the applied pressure. A capacitive sensor can be either an active sensing device where the applied pressure can be measured by the changes in capacitance or more often a passive wireless sensing device by combining it with an inductor coil [8]. The combination of the sensing capacitor and inductor coil makes an inductor-capacitor (LC) resonant tank circuit, which makes it suitable for wireless sensing via inductive coupling with an external antenna. The pressure is measured from relative changes in the resonance frequency of the LC resonant tank [26,27,28]. Due to this wireless communication between sensor and reader coil, capacitive pressure sensors are more practical for wearable and implantable applications as compared to resistive and piezoelectric sensing technologies that demand wired connection to communicate. Capacitive pressure sensing technology is generally used in MEMS and thin-film pressure sensors. MEMS-based sensors are accurate, miniaturized, wirelessly powered, and are widely used in wearable and implantable applications [29]. However, MEMS sensors are generally rigid and have a complex fabrication process that requires specialized equipment. On the other hand, thin-film based capacitive pressure sensors are flexible, less expensive, and simple to fabricate [30].
In past decades, many commercial solutions have been developed for pressure monitoring during compression therapy with growing research focused on lightweight, flexible, and wireless sensing systems. PicoPress (Microlab Electronica, Ponte S. Nicolo, Italy), air-pack type analyzer (AMI Techno, Tokyo, Japan), Kikuhime pneumatic transducer (Advancis Medical, Nottinghamshire, UK), Medical stocking tester (MST, Salzmann AG (SAG), St. Gallen, Switzerland), SIGaT tester (Ganzoni-Sigvaris, St. Gallen, Switzerland), and Oxford pressure monitor MK II (Talley Ltd., Romsey, UK) are available pneumatic sensor-based solutions to monitor the pressure during compression therapy [31,32,33,34]. A comparative study has confirmed that PicoPress and Kikuhime are more accurate compared to SIGaT [31]. PicoPress, Kikuhime, MST, and SIGaT are the most common medical devices focused on clinical applications, with relatively higher costs compared to stand-alone sensors [4,8]. Because of the pneumatic sensing principle, these systems are not appropriate for continuous dynamic pressure measurements [9,34].
On the other hand, Quantum tunneling composite (QTC, Peratech, Richmond, UK), ThruMode Force Sensing Resistor (FSR, Sensitronics, Bow, WA, USA), Interlink FSR (Interlink Electronics Inc., Camarillo, CA, USA), F-Scan (Tekscan, Inc., Boston, MA, USA) and Tactilus (Sensor Products Inc., Madison, NJ, USA) are commercially available piezoresistive pressure sensors being widely used to measure interface pressure during compression therapy [35]. Although these sensors are low-cost, thin, and flexible, they require a wired connection and additional electronics to work which makes the system bulky and impractical for real-time pressure measurements [35,36].
In addition to the commercially available compression therapy monitoring solutions, several research studies on pressure sensors and systems have been reported in the literature. Raj et al. [37] used water-filled polymerizing vinyl chloride (PVC) envelopes connected to an electrical pressure transducer to measure the interface pressure at four positions and reported that only within 6–8 h of daily routine applied pressure falls significantly. Hafner et al. [38] reported a silicone oil-filled pressure sensing system to train healthcare staff for an optimal compression in venous ulcer patient management; however, no details about the effect of temperature, hysteresis, and dynamic pressure are reported. Barbenel et al. [16] demonstrated a pressure sensing system for interface pressure using PVC probes filled with vegetable oil and was only limited to a pressure range of 0–37.5 mmHg. Burke et al. [4] developed an interface pressure monitoring system using four commercially available force sensors after integrated with a microcontroller and was capable to work in a range of 0–96 mmHg. However, there was observed a large hysteresis and lack of repeatability. Mehmood et al. [39] reported a telemetric mobile-based sub-bandage for monitoring the pressure and moisture of wounds but because of improper integration of commercial sensors, the system size was big. Casey et al. [8] reported a wearable capacitive flexible pressure sensing technology for sub-acute compression therapy monitoring. This flexible sensor array is built on active capacitor-based pressure sensing. Therefore, it requires a connected power supply and control unit. Farooqui et al. [40] reported a low-cost inkjet-printed wireless sensing system for chronic wound monitoring by measuring the pH level and physical pressure at the wound site. Rahimi et al. [41] has proposed an LC wireless strain sensor for wound monitoring by directly printing the conductive traces on the wound dressing but linearity was limited to 35% strain and no details reported about the repeatability and reliability of the system. Deng et al. [42] fabricated an LC wireless sensor for wound monitoring with a sensitivity of 270 kHz/mmHg in the range between 0 and 200 mmHg.
The majority of implantable or wearable sensors are based in LC systems due to the wireless communication between sensor and reader coil. Fonseca et al. [43] presented a very flexible wireless LC pressure sensor that was rollable and foldable to a compact shape for catheter-based delivery. This sensor was tested acutely in vivo for greater than 30 days in canine models simulating abdominal aortic aneurysms (AAA). Li et al. [44] reported a low power flexible sensor for intracranial pressure (ICP) monitoring, with a dual-mode operation in piezoelectric and capacitive modes, accuracy and reliability can be improved using dual-mode capability. Chen et al. [45] presented a wireless pressure sensor for continuous intraocular pressure monitoring of glaucoma patients with a long sensing distance and small physical form factor. Lei et al. [3] reported a flexible capacitive pressure sensor for plantar pressure measurements, different ratios of polydimethylsiloxane (PDMS) prepolymer and curing agent were mixed to improve the linearity by tweaking the stiffness factor.
The work presented here shows a flexible thin-film capacitive pressure sensor that can be fabricated using a simple and cost-effective etching process. The proposed sensor can be used in a wide range of medical applications, including intra and extracranial pressure, wound healing, and muscle rehabilitation monitoring; although, in this instance it has been designed mainly for interface pressure monitoring during compression therapy.
Considering varying ulcer sizes and lower limb curvatures, as well as different positions, two versions of the sensor with different sizes were fabricated, after optimization of their design parameters for best quality factor and resonance frequencies. Nevertheless, both sensors are LC resonant tank circuits and work on a capacitive sensing mechanism. The optimization of such parameters is reported as analytical results. In the experimental work, the performance of these sensors was evaluated over a pressure range of 0–100 mmHg. In addition, both sensors were also tested for a wider pressure range of 0–300 mmHg, as to suit a varying range of medical applications.
The rest of the paper is organized as follows: Section 2 describes the methodology, including the design, fabrication, and validation of the sensor; Section 3 presents the results obtained (analytical and experimental); Section 4 and Section 5 provide the final discussion and conclusions, respectively.

2. Materials and Methods

The proposed sensor is based on an LC resonance circuit, where the resonance frequency of the LC circuit is proportional to the applied pressure. The schematic diagram of the wireless sensing system is presented in Figure 1a. By placing multiple sensors under compression bandage as shown in Figure 1b, an array of wireless sensors can be formed to help in delivering a more controlled personalized compression therapy for the fast recovery of venous ulcers. A wearable readout band can keep records of pressure profiles during the daily routine.

2.1. Sensor Design

The LC sensor is designed as a disc capacitor made of two parallel circular plates, and the inductor is a planar circular spiral coil located around one of the capacitor electrodes suited for a flexible design for a bandage–skin interface. A geometrical representation of the sensor and reader are shown in Figure 2a,b respectively. The resonance frequency ( f o ) of the proposed LC sensor depends on the inductance ( L s ) and capacitance ( C s ) of the sensor, as given in Equation (1):
f o = 1 2 π L s C s
The capacitance of the sensor can be calculated as in Equation (2):
C s = ϵ o ϵ r π r 2 d
where ϵ o is the permittivity of free space, ϵ r is the relative permittivity of dielectric material in the capacitor, and r is the radius of the disk capacitor. The inductance of the planar spiral inductor is calculated using its current sheet expression [46], which depends on the inner diameter d i n , outer diameter d o u t and number of turns N , as given in Equation (3):
L s = μ o N 2 d a v g C 1 2 ( ln ( C 2 / τ ) + C 3 τ + C 4 τ 2 )
where μ o is the permeability of free space, N is the number of turns, d a v g = ( d i n + d o u t ) 2 , τ = ( d o u t d i n ) ( d o u t + d i n ) , and C 1 , C 2 , C 3 and C 4 are the coefficients for the current sheet expression, which are 1, 2.46, 0, and 0.2 for a circular design [46].

Parasitic Components

The inductive part of the sensor, consisting of circular spirals, can be modeled accurately using lumped elements. Its elements are an inductor ( L s ) , a parasitic resistance ( R t o t ) , and parasitic capacitance ( C p ), where L s and R t o t are in series in parallel to C p as shown in Figure 3a.
One of the major parasitic effects that play a major role in the quality factor of the inductor is the series resistance, which is modeled as R t o t in this paper. A large R t o t will result in a poor quality factor of the inductor in the sensor, as well as in the reader coil. This R t o t can be represented by Equation (4), which includes direct current resistance ( R d c ) and alternating current resistance ( R a c ) .
R t o t = R d c + R a c
R d c can be calculated according to Equation (5), where ρ is the resistivity of the conductor, l is the length of the spiral conductor, w is the trace width and t is the trace thickness.
R d c = ρ l w t
For a spiral inductor with N number of turns, outer and inner diameters d o u t and d i n , the length of the conductive traces can be calculated using Equation (6).
l = π N ( d i n + d o u t ) 2
The component R a c in Equation (4) is affected by the values of R s k i n and R p r o x , which correspond to the skin effect and proximity effect, respectively:
R a c = R s k i n + R p r o x
The skin effect occurs at higher frequencies when current does not flow through the complete cross-sectional area of the conductor, and it starts flowing only through its surface as shown in Figure 3b, which increases the effective resistance. In Figure 3b, the red color represents the skin depth ( δ ) for current flow and the blue color shows the area without electric current. This effect is represented by the skin depth δ . The mathematical expression to compute R s k i n is given in Equation (8) [47]. Here μ o is the permeability constant and μ r is the relative permeability of the conductor and f is the operational frequency.
R s k i n = ρ l w δ ( 1 e t δ ) ( 1 + t w ) , where   δ = ρ π μ o μ r f
The proximity effect is another major contributor to R a c that becomes significant above a frequency specific to the design, known as crowding frequency, f c r i t . In the signal frequencies above f c r i t , magnetic forces surrounding the conductor become significant and result in a nonuniform current flow through the conductor. This redistribution of the current causes an increase in effective resistance and can be calculated through Equation (9) [48].
R p r o x = R d c 10 ( f f c r i t ) 2 , where   f c r i t = 3.1 ( w + s ) ρ 2 π μ o w 2 t
The parasitic capacitance between the nearby turns can be computed from Equation (10) [49,50], where α and β are 0.9 and 0.1, respectively, and represent the parasitic contribution due to the air gap between the coil turns, and the gap between the metallic tracks and the substrate, as shown in Figure 3c. ϵ r c and ϵ r s are the relative permittivity of air and substrate material respectively.
C p = C p c + C p s = l t ϵ o s ( α ϵ r c + β ϵ r s )
The value of the self-resonance frequency f S R F of an inductor is critical, as above this frequency the parasitic capacitance of the inductor becomes dominant. The f S R F can be calculated using Equation (11) [50].
f S R F = 1 2 π L s C p
Finally, the quality factor of the LC sensor is given by Equation (12) [51].
Q F = 1 R t o t L s C s

2.2. Device Fabrication

After the optimization of design parameters that is discussed in Section 3.1, a wet etching process was used to fabricate the two different sensors and their reader antennas. Figure 4 shows the stages in the fabrication process. In step I, as shown in Figure 4a, the mask of the sensor was directly printed on a 50 μm thick copper-coated polyimide film (Flexible isolating circuit 50 μm-coppered 35 μm-1 side, CIF, Buc, France) with a LaserJet printer (HP M553, HP Technology, Dublin, Ireland). In step II, the printed copper sheets were immersed in an etchant solution (CIF, Boosted ferric chloride solution). After manual stirring for 15 min at room temperature, all the unwanted copper was removed as shown in Figure 4b, and the patterned sheet was washed with hot water. Acetone was used to remove the ink particles from the copper surface after the etching process. In the next step, a polydimethylsiloxane (PDMS) layer (Ultra-thin film, 30° shore A hardness, Silex Ltd., Bordon, UK) of 200 μm thickness was cut into a circular shape equal to the diameter of the capacitor electrodes and was placed on the bottom electrode as shown in Figure 4c. PDMS is widely used as a dielectric layer in capacitive pressure sensors due to its low Young’s modulus and compressibility. An adhesive layer composed of polypropylene and synthetic rubber of 90 μm thickness (Tesa64621, Tesa, Norderstedt, Germany) was placed around the PDMS layer as shown in Figure 4d. In the final step, the top layer of the sensor was folded onto the PDMS layer for the final assembly of the sensor. Figure 4e,f shows the top and bottom views of the fabricated sensor. The reader antenna was also fabricated by the same etching procedure, and flexible multithread wires were soldered to connect with a Sub-Miniature version A (SMA) connector.

2.3. Device Validation

To test the fabricated system (sensor with reader coils), a bench-test model was developed using a vector network analyzer (VNA E5063, Keysight Technologies Inc., Santa Rosa, CA, USA), a high-pressure glass bottle (Pressure+ 1000, Duran, Mainz, Germany), and a digital pressure gauge (Traceable 3462, Fisher Scientific Ltd., Loughborough, UK), as shown in Figure 5. The sensor was placed inside the pressure chamber and its response recorded using the reader antenna, which was placed outside the wall of the chamber. The pressure was varied using a vacuum pump (FB70155 Pump, Fisher Scientific Ltd., Loughborough, UK) to produce positive pressure inside the chamber, which was measured as well by the digital pressure gauge. The input impedance of the VNA was 50 Ω. A frequency sweep was generated from the VNA to observe the variation in resonance frequency against the varying pressure, and the S parameters of the sensor were recorded simultaneously.

3. Results

The results presented in this paper comprise of the outcomes of two types of investigation: analytical investigations (Section 3.1) and experimental investigations (Section 3.2). The analytical investigations are performed for optimization of design parameters ( d o u t , N , w , s ) to achieve the best quality factor ( Q F ) , and lower resonance frequencies ( f o ) . The experimental investigations are performed to test and characterize the performance of the two fabricated prototype sensors on suitable testbeds.

3.1. Analytical Results: Numerical Estimation of Sensor Parameters

Sensor optimization was done in two steps. In the first step, the outer diameter ( d o u t ) and the number of turns ( N ) of the inductor were optimized while keeping the trace width ( w ) and trace separation ( s ) constant. In the second step, after selecting the optimal values of d o u t and N , both s and w were adjusted to achieve the best quality factor ( Q F ) with a low resonance frequency ( f o ) .

3.1.1. Optimization of Outer Diameter ( d o u t ) and Number of Turns ( N )

Before the fabrication stage of the sensor, MATLAB numerical modeling was performed to achieve the best quality factor ( Q F ) within low resonance frequency ( f o ) range to achieve a better signal to noise ratio (SNR). The two different designs of the sensor, sensor 1 (S1) and sensor 2 (S2), were characterized according to their individual parameters. S1 was modeled for different d o u t values, between 36 and 45 mm, and a varying N from 1 to 10, while keeping s = w = 500 μm. As can be seen from the data point shown in Figure 6, the best Q F was 106.4 , with a correspondent resonance frequency of 17.147 MHz, when d o u t and N were 45 mm and 10 respectively. However, to keep the sensor size small, we selected d o u t = 40 mm and N = 10 for the fabrication as there was no significant loss in Q F (97.46), and f o was also low (19.188 MHz).
A similar model was computed for S2 as shown in Figure 7. In this case, the objective was to design a relatively small sensor; therefore, d o u t was varied between 10 and 14 mm and N between 1 and 5 turns, while s and w were kept constant at 500 and 200 μm respectively. For S2, the highest Q F was ~32, with a f o of 222.4 MHz for d o u t = 14 mm and N = 5; however, we selected d o u t = 12 mm and N = 5 to achieve an optimal set of Q F (23.93) and f o (259.44 MHz) against the size of the sensor.

3.1.2. Optimization of Trace width ( w ) and Trace separation ( s )

Trace width and trace separation also affect the Q F and resonance frequency; therefore, complete numerical modeling was performed for the selection of the trace geometry. Q F and f o were analyzed for different values of s and w , while the number of turns and d o u t were fixed this time. Both trace width ( w ) and trace gap ( s ) were varied within the maximum allowable range to fit within the limits of given sensor size and number of turns. For S1, values of s and w were modeled between 200 and 600 μm and d o u t and N were 40 mm and 10 , respectively. As shown in Figure 8, the highest Q F (103.5) was observed for s = 325 μm and w = 400 μm, with a resonance frequency of 16.82 MHz. For an equally distributed pattern with a trace width ( w ) and trace gap ( s ) of 500 μm, a very small loss in Q F (~5%) was observed, therefore, w = s = 500 μm were chosen for the design of S1.
S2 was modeled by varying s between 200 and 300 μm and w from 200 and 500 μm, while keeping d o u t = 12 mm and N = 5 fixed, as shown in Figure 9. Maximum Q F was 23.93 with a resonance frequency of 259.44 MHz for a combination of s = 500 μm and w = 200 μm. As both the Q F and the resonance frequency of S2 were very sensitive to trace width and separation, the combination of s and w that produced the best Q F were chosen for S2.

3.2. Experimental Prototype and Results

After selecting the optimized design parameters ( d o u t , N , s , and   w ) , two sensors, of outer diameters 40 and 12 mm (shown in Figure 10), were fabricated and tested using the test-bench described in Section 2.3. The key design parameters, results, and operating frequencies for both sensors and respective reader coils are listed in Table 1.
As discussed in Section 1, since bandage pressure varies between 10 and 60 mmHg during compression therapy, both fabricated sensors were tested for a pressure range of 0 to 100 mmHg. The reader coil connected with the network analyzer was magnetically coupled with the sensor, and the response of the sensor over varying pressure was measured. The measurements from VNA were triggered at an interval of 5 mmHg for a narrow range of 0–100 mmHg. These measurements are the reflection coefficients (S11 parameter) and are shown in Figure 11 and Figure 12 for the sensors S1 and S2, respectively.
In addition to the compression therapy monitoring, the proposed sensors could be used for other medical applications, including physiological pressure measurement. Therefore, both sensors were also tested over a wider range of 0 to 300 mmHg that covers almost the entire physiological pressure range. The measurements from VNA were triggered at an interval of 25 mmHg for a wide range of 0–300 mmHg. Figure 13 and Figure 14 show the measured reflection coefficients (S11 parameter) of S1 and S2 over this broad pressure range.
As the response of both the sensors was linear within the targeted pressure range of 0 to 100 mmHg, a first-order polynomial was fitted over the measured response of the sensors. The coefficients of the linear fitted model are given in Table 2. The measure sensor response (dotted) and fitted curve (solid) for both sensors are shown in Figure 15.
In the sensor response over a wide range of pressure up to 300 mmHg, a nonlinearity, associated with compression saturation of the dielectric layer, was observed at higher pressures as shown in Figure 16. Therefore, a second-order polynomial function was fitted to the measured response to obtain a model relating the resonance frequency to the pressure. The values of R-square (goodness of fit) and the model coefficients are listed in Table 3.
To assess the repeatability of pressure measurement with both sensors, the response of the sensors for six different pressure points between 0 and 100 mmHg was measured repeatedly for 10 cycles. Figure 17 and Figure 18 show the repeatability of S1 and S2, respectively. The mean values of the frequency response against applied pressure ( f u ) and standard deviation ( σ ) of 10 repeated measurements at 6 pressure points (100, 80, 60, 40, 20, 0) mmHg are given in Table 4.

4. Discussion

An LC pressure sensing system is developed to measure the pressure in compression therapy due to wireless communication between sensor and reader coil. Optimization of the sensors is essential to achieve the best quality factor and resonance frequency while keeping the sensor size limited. Optimized values of outer diameter ( d o u t ) and the number of turns ( N ) , trace width ( w ) , and trace separation ( s ) are listed in Table 1. The parasitic components of the sensor which are parasitic capacitance and parasitic resistance at resonance frequency were analyzed through numerical modeling and their values are reported in Table 1. The reported sensors were fabricated using a wet etching process, which is cost-effective and very simple but comes at the cost of less control on trace widths. In these circumstances, the thinnest trace width achieved was 200 μm. Both sensors were characterized using a bench test setup that was developed during this research work. Both sensors showed good linearity and repeatability for a pressure <100 mmHg.
As shown in Figure 15, the response of both designed sensors was linear over a pressure range of 0–100 mmHg, with a sensitivity of 8 kHz/mmHg for S1 and 65 kHz/mmHg for S2. The sensor response was observed as nonlinear at the higher pressure range of 0 to 300 mmHg, as shown in Figure 16. This is due to the nonlinear effect of the compression saturation of the dielectric layer of the capacitor in the sensor. Up to 100 mmHg, the sensitivity of S1 was 8.11 kHz/mmHg, which was reduced at higher pressure due to the dielectric layer saturation. Similar behavior was noticed for S2, where sensitivity was 65.48 kHz/mmHg up to 100 mmHg, and was reduced when the sensor was loaded with higher values of pressure.
Both sensors offered good repeatability as shown in Figure 17 and Figure 18, for a pressure range <80 mmHg; however, variability in measurements started growing in the sensor response for higher applied pressures (>80 mmHg), due to the already mentioned hysteresis of the dielectric layer. As it can be noticed from Table 4 that the average repeatability for both the sensors over the pressure range of 0–100 mmHg is slightly larger than the sensitivity per mmHg, the measurement uncertainty is estimated as less than ±1 mmHg.
From Table 1, it can be noticed that Q F of S1 was better than S2, which is due to the exponential increase of the ac resistance at higher frequencies for S2 caused by the skin effect. In addition, by comparing the amplitude of S parameters of both sensors in Figure 13 and Figure 14, it is quite clear that S1 has a better signal to noise ratio ( SNR ) compared to S2.
There was noticed a difference between the calculated and measured resonance frequencies of both sensors (S1 and S2) was due to numerous possible reasons. The first possible reason might be the value of the PDMS dielectric constant ( ϵ r _ PDMS ) , which reported between 2.3 and 2.8 in literature [52]; however, for this research ϵ r _ PDMS was selected 2.65 as stated in Table 1. The second reason for this difference might be due to the roughness of conductive traces caused by an over-etching effect during the fabrication process. This difference was greater for S1 due to the uneven distribution of the dielectric layer and air gaps between the capacitor plates, which were relatively bigger as compared to S2. In the future, a more controlled fabrication process can be used to improve the etching process and dielectric layer deposition to overcome the mismatch between analytical and real values of sensor parameters.
A comparison of the developed sensors with previously reported systems is given in Table 5. It includes sensors developed explicitly for wound compression therapy, and as an extension, implantable sensors that measure bodily pressures in different locations. Although not designed specifically for the application targeted in this work, these implantable sensors are based on the same sensing concept of LC systems and operate in similar pressure ranges (as shown in Table 5). From the observation of the values listed in the table, it is noticeable that the sensitivity of the S2 sensor, 65.48 kHz/mmHg, is comparable with the prototypes reported in the literature. This is, in the author’s view, a noteworthy achievement, considering the fact that the sensor proposed here is based on a very simple and non-expensive fabrication method. By contrast, most states of the art sensors are based on microfabrication techniques, which are very expensive and laborious.

5. Conclusions

This work presented the design of a wireless capacitive pressure sensor of low-cost fabrication for medical applications. In particular, the sensor is designed to be used for monitoring of compression therapy in venous leg ulcers. The sensor design was optimized to achieve an optimal quality factor and resonance frequency by numerical modeling of the design parameters. The proposed thin-film flexible wireless pressure sensor was fabricated using a simple and cost-effective fabrication method. Two versions of the sensors, with 40 and 12 mm outer diameters respectively, were developed and characterized between 0–100 and 0–300 mmHg to cover the pressure range of compression therapy and the nominal range of all other physiological applications. A bench test setup was also developed for sensor validation using a glass pressure bottle, pressure pump, and a network analyzer. Both sensors showed good sensitivity, linearity, and repeatability for the lower pressure regime (0–100 mmHg). A MATLAB curve-fitting tool was used to model the relationship between the shift in resonance frequency and the change in pressure.
The focus of this research work was on the early prototype development of the sensor, which is characterized by the benchtop model. However, in the future, improved and miniaturized prototypes will be fabricated by a more controlled fabrication process, and an extensive study will be performed on human subjects to validate the effectiveness. The miniaturization and replacement of the dielectric material used in the proposed sensors with other elastomeric polymers, can improve the linearity, sensitivity, and repeatability of the sensor and will make it more suitable for numerous medical applications.

Author Contributions

Conceptualization, M.F. and A.S.; methodology, M.F., N.F., and S.T.; software, M.F.; validation, M.F., T.I., and A.S.; formal analysis, M.F.; investigation, M.F., and A.S.; resources, A.S. and W.W.; data curation, M.F. and P.V.; writing—original draft preparation, M.F.; writing—review and editing, M.F., A.S., and W.W., P.V.; visualization, M.F.; supervision, A.S. and W.W.; project administration, M.F.; funding acquisition, W.W. All authors have read and agreed to the published version of the manuscript.

Funding

The research leading to this publication was funded by the Science Foundation Ireland Research Professorship Award (grant no. 15/RP/2765), the Government of Ireland Disruptive Technology Innovation Fund (grant no. DT20180031A), and Enterprise Ireland (grant no. CF-2019-1125).

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Trung, T.Q.; Lee, N.E. Flexible and Stretchable Physical Sensor Integrated Platforms for Wearable Human-Activity Monitoringand Personal Healthcare. Adv. Mater. 2016, 28, 4338–4372. [Google Scholar] [CrossRef] [PubMed]
  2. Liu, Y.; Pharr, M.; Salvatore, G.A. Lab-on-Skin: A Review of Flexible and Stretchable Electronics for Wearable Health Monitoring. ACS Nano 2017, 11, 9614–9635. [Google Scholar] [CrossRef] [PubMed]
  3. Lei, K.F.; Lee, K.F.; Lee, M.Y. Development of a flexible PDMS capacitive tactile sensor with adjustable measurement range for plantar pressure measurement. Microsyst. Technol. 2014, 20, 1351–1358. [Google Scholar] [CrossRef]
  4. Burke, M.; Murphy, B.; Geraghty, D. Measurement of sub-bandage pressure during venous compression therapy using flexible force sensors. In Proceedings of the IEEE SENSORS Conference, Valencia, Spain, 2–5 November 2014; pp. 1623–1626. [Google Scholar] [CrossRef]
  5. Ghosh, S.; Mukhopadhyay, A.; Sikka, M.; Nagla, K.S. Pressure mapping and performance of the compression bandage/garment for venous leg ulcer treatment. J. Tissue Viability 2008, 17, 82–94. [Google Scholar] [CrossRef]
  6. Valencia, I.C.; Falabella, A.; Kirsner, R.S.; Eaglstein, W.H. Chronic venous insufficiency and venous leg ulceration. J. Am. Acad. Dermatol. 2001, 44, 401–424. [Google Scholar] [CrossRef]
  7. Partsch, H.; Clark, M.; Bassez, S.; Benigni, J.; Becker, F.; Blazek, V.; Caprini, J.; Cornu-Thénard, A.; Hafner, J.; Flour, M.; et al. Measurement of lower leg compression in vivo: Recommendations for the performance of measurements of interface pressure and stiffness. Dermatol. Surg. 2006, 32, 224–233. [Google Scholar] [CrossRef]
  8. Casey, V.; McAree, B.; Moloney, M.C.; Grace, P. Wearable sub-bandage pressure measurement system. In Proceedings of the 2010 IEEE Sensors Applications Symposium (SAS), Cancun, Mexico, 9–11 November 2016; pp. 41–45. [Google Scholar] [CrossRef]
  9. Flaud, P.; Bassez, S.; Counord, J.L. Comparative in vitro study of three interface pressure sensors used to evaluate medical compression hosiery. Dermatol. Surg. 2010, 36, 1930–1940. [Google Scholar] [CrossRef]
  10. Hegarty, M.S.; Grant, E.; Reid, L. An overview of technologies related to care for venous leg ulcers. IEEE Trans. Inf. Technol. Biomed. 2010, 14, 387–393. [Google Scholar] [CrossRef]
  11. Rahimi, M.; Blaber, A.P.; Menon, C. Towards the evaluation of force-sensing resistors for in situ measurement of interface pressure during leg compression therapy. In Proceedings of the 2016 IEEE Healthcare Innovation Point-Of-Care Technologies Conference (HI-POCT), Cancun, Mexico, 9–11 November 2016; pp. 25–28. [Google Scholar] [CrossRef]
  12. Kaisti, M.; Panula, T.; Leppänen, J.; Punkkinen, R.; Tadi, M.J.; Vasankari, T.; Jaakkola, S.; Kiviniemi, T.; Airaksinen, J.; Kostiainen, P.; et al. Clinical assessment of a non-invasive wearable MEMS pressure sensor array for monitoring of arterial pulse waveform, heart rate and detection of atrial fibrillation. NPJ Digit. Med. 2019, 2, 1–10. [Google Scholar] [CrossRef] [Green Version]
  13. Brown, M.S.; Ashley, B.; Koh, A. Wearable technology for chronic wound monitoring: Current dressings, advancements, and future prospects. Front. Bioeng. Biotechnol. 2018, 6, 1–21. [Google Scholar] [CrossRef]
  14. Zang, Y.; Zhang, F.; Di, C.A.; Zhu, D. Advances of flexible pressure sensors toward artificial intelligence and health care applications. Mater. Horiz. 2015, 2, 140–156. [Google Scholar] [CrossRef]
  15. Fraulini, K.E. Interface pressure. Decubitis 2014, 3, 1–12. [Google Scholar] [CrossRef]
  16. Barbenel, J.C.; Sockalingham, S. Device for measuring soft tissue interface pressures. J. Biomed. Eng. 1990, 12, 519–522. [Google Scholar] [CrossRef]
  17. Van Pijkeren, T.; Naeff, M.; Kwee, H.H. A new method for the measurement of normal pressure between amputation residual limb and socket. Bull. Prosthet. Res. 1980, 10, 31–34. [Google Scholar]
  18. Bu, N.; Ueno, N.; Fukuda, O. Monitoring of respiration and heartbeat during sleep using a flexible piezoelectric film sensor and empirical mode decomposition. In Proceedings of the 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Lyon, France, 22–26 August 2007; pp. 1362–1366. [Google Scholar] [CrossRef]
  19. Han, M.; Wang, H.; Yang, Y.; Bai, W.; Yan, Z.; Li, H.; Xue, Y.; Wang, X.; Akar, B.; Zhao, H.; et al. Three-dimensional piezoelectric polymer microsystems for vibrational energy harvesting, robotic interfaces and biomedical implants. Nat. Electron. 2019, 2, 26–35. [Google Scholar] [CrossRef]
  20. Sharma, T.; Je, S.S.; Gill, B.; Zhang, J.X.J. Patterning piezoelectric thin film PVDF-TrFE based pressure sensor for catheter application. Sens. Actuators A Phys. 2012, 177, 87–92. [Google Scholar] [CrossRef]
  21. Gautschi, G. Piezoelectric Sensorics; Springer: Berlin/Heidelberg, Gemany, 2002; Volume 3. [Google Scholar]
  22. Chen, X.; Shao, J.; An, N.; Li, X.; Tian, H.; Xua, C.; Ding, Y. Self-powered flexible pressure sensors with vertically well-aligned piezoelectric nanowire arrays for monitoring vital signs. J. Mater. Chem. C 2015, 3, 11806–11814. [Google Scholar] [CrossRef]
  23. Gong, S.; Schwalb, W.H.; Wang, Y.; Chen, Y.; Tang, Y.; Si, K.J.; Shirinzadeh, B.; Cheng, W. A wearable and highly sensitive pressure sensor with ultrathin gold nanowires. Nat. Commun. 2014, 5, 1–8. [Google Scholar] [CrossRef] [Green Version]
  24. Ng, K.C.-K.; Leung, Y.-F.; Lee, Y.-L. An ultra-sensitive resistive pressure sensor based on hollow-sphere microstructure induced elasticity in conducting polymer fil. Hand Surg. 2014, 19, 423–425. [Google Scholar] [CrossRef]
  25. Kenry, J.; Yeo, C.; Lim, C.T. Emerging flexible and wearable physical sensing platforms for healthcare and biomedical applications. Microsyst. Nanoeng. 2016, 2. [Google Scholar] [CrossRef]
  26. Chitnis, G.; Maleki, T.; Samuels, B.; Cantor, L.B.; Ziaie, B. A minimally invasive implantable wireless pressure sensor for continuous IOP monitoring. IEEE Trans. Biomed. Eng. 2013, 60, 250–256. [Google Scholar] [CrossRef] [PubMed]
  27. Boutry, C.M.; Beker, L.; Kaizawa, Y.; Vassos, C.; Tran, H.; Hinckley, A.C.; Pfattner, R.; Niu, S.; Li, J.; Claverie, J.; et al. Biodegradable and flexible arterial-pulse sensor for the wireless monitoring of blood flow. Nat. Biomed. Eng. 2019, 3, 47–57. [Google Scholar] [CrossRef] [PubMed]
  28. Takahata, K.; Dehennis, A.; Wise, K.D.; Gianchandani, Y.B. A Wireless Microsensor For Monitoring Flow and Pressure in a Blood Vessel Utilizing a Dual-Inductor Antenna Stent and Two Pressure Sensors. In Proceedings of the 17th IEEE International Conference on Micro Electro Mechanical Systems, Maastricht, The Netherlands, 25–29 January 2004; pp. 5–8. [Google Scholar]
  29. Park, J.; Kim, J.K.; Patil, S.J.; Park, J.K.; Park, S.A.; Lee, D.W. A wireless pressure sensor integrated with a biodegradable polymer stent for biomedical applications. Sensors 2016, 16, 809. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  30. Chhetry, A.; Yoon, H.; Park, J.Y. A flexible and highly sensitive capacitive pressure sensor based on conductive fibers with a microporous dielectric for wearable electronics. J. Mater. Chem. C 2017, 5, 10068–10076. [Google Scholar] [CrossRef]
  31. Al Khaburi, J.A.J. Pressure Mapping of Medical Compression Bandages Used for Venous Leg Ulcer Treatment. 2010. Available online: http://etheses.whiterose.ac.uk/3386/1/Thesis_Corrected.pdf (accessed on 10 September 2020).
  32. Partsch, H. The Static Stiffness Index: A Simple Method to Assess the Elastic Property of Compression Material In Vivo. Dermatol. Surg. 2006, 31, 625–630. [Google Scholar] [CrossRef]
  33. Van den Kerckhove, E.; Fieuws, S.; Massagé, P.; Hierner, R.; Boeckx, W.; Deleuze, J.-P.; Laperre, J.; Anthonissen, M. Reproducibility of repeated measurements with the Kikuhime pressure sensor under pressure garments in burn scar treatment. Burns 2007, 33, 572–578. [Google Scholar] [CrossRef]
  34. Finnie, A. Interface pressure measurements in leg ulcer management. Br. J. Nurs. 2000, 9 (Suppl. 1), S8–S18. [Google Scholar] [CrossRef]
  35. Parmar, S.; Khodasevych, I.; Troynikov, O. Evaluation of flexible force sensors for pressure monitoring in treatment of chronic venous disorders. Sensors 2017, 17, 1923. [Google Scholar] [CrossRef]
  36. Khodasevych, I.; Parmar, S.; Troynikov, O. Flexible sensors for pressure therapy: Effect of substrate curvature and stiffness on sensor performance. Sensors 2017, 17, 2399. [Google Scholar] [CrossRef]
  37. Rai, T.B.; Goddard, M.; Makin, G.S. How long do compression bandages maintain their pressure during ambulatory treatment of varicose veins. Br. J. Surg. 1980, 67, 122–124. [Google Scholar]
  38. Hafner, J.; Lüthi, W.; Hänssle, H.; Kammerlander, G.; Burg, G. Instruction of compression therapy by means of interface pressure measurement. Dermatol. Surg. 2000, 26, 481–488. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  39. Mehmood, N.; Hariz, A.; Templeton, S.; Voelcker, N.H. An improved flexible telemetry system to autonomously monitor sub-bandage pressure and wound moisture. Sensors 2014, 14, 21770–21790. [Google Scholar] [CrossRef] [PubMed]
  40. Farooqui, M.F.; Shamim, A. Low cost inkjet printed smart bandage for wireless monitoring of chronic wounds. Sci. Rep. 2016, 6, 28949. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  41. Rahimi, R.; Ochoa, M.; Zieger, M.; Sood, R.; Ziaie, B. A wireless strain sensor for wound monitoring with direct laser-defined patterning on a commercial dressing. In Proceedings of the 2016 IEEE 29th International Conference on Micro Electro Mechanical Systems (MEMS), Shanghai, China, 24–28 January 2016; pp. 481–484. [Google Scholar] [CrossRef]
  42. Deng, W.J.; Wang, L.F.; Dong, L.; Huang, Q.A. LC Wireless Sensitive Pressure Sensors with Microstructured PDMS Dielectric Layers for Wound Monitoring. IEEE Sens. J. 2018, 18, 4886–4892. [Google Scholar] [CrossRef]
  43. Fonseca, M.A.; Allen, M.G.; Kroh, J.; White, J. Flexible Wireless Passive Pressure Sensors for Biomedical Applicatons. In Proceedings of the Solid-State Sensors, Actuators, and Microsystems Workshop, Hilton Head Island, SC, USA, 4–8 June 2006; pp. 37–42. [Google Scholar]
  44. Li, C.; Wu, P.M.; Shutter, L.A.; Narayan, R.K. Dual-mode operation of flexible piezoelectric polymer diaphragm for intracranial pressure measurement. Appl. Phys. Lett. 2010, 96, 2008–2011. [Google Scholar] [CrossRef]
  45. Chen, P.-J.; Saati, S.; Varma, R.; Humayun, M.S.; Tai, Y.-C. Wireless Intraocular Pressure Sensing Using Microfabricated Minimally Invasive Flexible-Coiled LC Sensor Implant. J. Microelectromech. Syst. 2010, 19, 721–734. [Google Scholar] [CrossRef]
  46. Mohan, S.S.; Hershenson, M.D.M.; Boyd, S.P.; Lee, T.H. Simple accurate expressions for planar spiral inductances. IEEE J. Solid-State Circuits 1999, 34, 1419–1420. [Google Scholar] [CrossRef] [Green Version]
  47. Schormans, M.; Valente, V.; Demosthenous, A. Practical Inductive Link Design for Biomedical Wireless Power Transfer: A Tutorial. IEEE Trans. Biomed. Circuits Syst. 2018, 1–19. [Google Scholar] [CrossRef] [Green Version]
  48. Kuhn, W.B.; Ibrahim, N.M. Analysis of current crowding effects in multiturn spiral inductors. IEEE Trans. Microw. Theory Tech. 2001, 49, 31–38. [Google Scholar] [CrossRef] [Green Version]
  49. Ammouri, A.; Salah, T.B.; Morel, H. A spiral planar inductor: An experimentally verified physically based model for frequency and time domains. Int. J. Numer. Model. Electron. Netw. Devices Fields 2018, 31, 1–13. [Google Scholar] [CrossRef]
  50. Islam, A.B.; Islam, S.K.; Tulip, F.S. Design and Optimization of Printed Circuit Board Inductors for Wireless Power Transfer System. Circuits Syst. 2013, 4, 237–244. [Google Scholar] [CrossRef] [Green Version]
  51. Huang, Q.A.; Dong, L.; Wang, L.F. LC Passive Wireless Sensors Toward a Wireless Sensing Platform: Status, Prospects, and Challenges. J. Microelectromech. Syst. 2016, 25, 822–841. [Google Scholar] [CrossRef]
  52. MIT.edu. Material: PDMS (Polydimethylsiloxane). Available online: http://www.mit.edu/~6.777/matprops/pdms.htm (accessed on 24 April 2019).
Figure 1. (a) Schematic diagram of wireless LC sensing system showing sensor and reader coil connected with a vector network analyzer (b) An application demonstration using flexible pressure sensors under the compression bandage.
Figure 1. (a) Schematic diagram of wireless LC sensing system showing sensor and reader coil connected with a vector network analyzer (b) An application demonstration using flexible pressure sensors under the compression bandage.
Sensors 20 06653 g001
Figure 2. Geometrical representation of the proposed LC system: (a) LC sensor with a capacitor of the radius ( r ) and planar inductor with an inner diameter ( d i n ) shown with a solid line, outer diameter ( d o u t ) shown with a dotted line, trace separation ( s ) and trace width ( w ); (b) Reader antenna with the same design parameters ( d i n , d o u t , s , w ).
Figure 2. Geometrical representation of the proposed LC system: (a) LC sensor with a capacitor of the radius ( r ) and planar inductor with an inner diameter ( d i n ) shown with a solid line, outer diameter ( d o u t ) shown with a dotted line, trace separation ( s ) and trace width ( w ); (b) Reader antenna with the same design parameters ( d i n , d o u t , s , w ).
Sensors 20 06653 g002
Figure 3. Parasitic effects: (a) Lumped model of the spiral inductor showing the inductor ( L s ), parasitic resistance and capacitance; (b) Skin effect on a rectangular conductor, with current flowing only in the red area; (c) The parasitic capacitance is due to the air gap between coil turns ( C p c ) and the substrate material ( C p s ) .
Figure 3. Parasitic effects: (a) Lumped model of the spiral inductor showing the inductor ( L s ), parasitic resistance and capacitance; (b) Skin effect on a rectangular conductor, with current flowing only in the red area; (c) The parasitic capacitance is due to the air gap between coil turns ( C p c ) and the substrate material ( C p s ) .
Sensors 20 06653 g003
Figure 4. Fabrication process: (a) Copper-coated polyimide film with ink printed mask. (b) Etched pattern showing capacitor electrodes and planar inductor spirals. (c) Dielectric layer of PDMS elastomer. (d) Adhesive layer placement around the PDMS dielectric. (e) Top view and (f) bottom view of the LC sensor.
Figure 4. Fabrication process: (a) Copper-coated polyimide film with ink printed mask. (b) Etched pattern showing capacitor electrodes and planar inductor spirals. (c) Dielectric layer of PDMS elastomer. (d) Adhesive layer placement around the PDMS dielectric. (e) Top view and (f) bottom view of the LC sensor.
Sensors 20 06653 g004
Figure 5. Bench test setup for sensor validation where reader coil is connected to vector network analyzer and sensor is kept inside the pressure chamber and pressure is varied using pressure pump.
Figure 5. Bench test setup for sensor validation where reader coil is connected to vector network analyzer and sensor is kept inside the pressure chamber and pressure is varied using pressure pump.
Sensors 20 06653 g005
Figure 6. Analysis of S1 quality factor and resonance frequency for different number of turns and outer diameters, when trace separation and width were kept constant at 500 μm.
Figure 6. Analysis of S1 quality factor and resonance frequency for different number of turns and outer diameters, when trace separation and width were kept constant at 500 μm.
Sensors 20 06653 g006
Figure 7. Analysis of S2 quality factor and resonance frequency for different number of turns and outer diameters, when trace separation was 500 μm and trace width was 200 μm.
Figure 7. Analysis of S2 quality factor and resonance frequency for different number of turns and outer diameters, when trace separation was 500 μm and trace width was 200 μm.
Sensors 20 06653 g007
Figure 8. Analysis of S1 quality factor and resonance frequency for different trace separation and trace width, when the number of turns and outer diameter were 10 and 40 mm, respectively.
Figure 8. Analysis of S1 quality factor and resonance frequency for different trace separation and trace width, when the number of turns and outer diameter were 10 and 40 mm, respectively.
Sensors 20 06653 g008
Figure 9. S2 quality factor and resonance frequency analysis for different trace separation and trace width when the number of turns and outer diameter were 5 and 12 mm, respectively.
Figure 9. S2 quality factor and resonance frequency analysis for different trace separation and trace width when the number of turns and outer diameter were 5 and 12 mm, respectively.
Sensors 20 06653 g009
Figure 10. Fully labeled images of final prototypes, (a) Fabricated wireless LC resonance sensors: left, S1 (40 mm in diameter), and right, S2 (12 mm diameter); (b) Reader coils for S1 (left) and S2 (right).
Figure 10. Fully labeled images of final prototypes, (a) Fabricated wireless LC resonance sensors: left, S1 (40 mm in diameter), and right, S2 (12 mm diameter); (b) Reader coils for S1 (left) and S2 (right).
Sensors 20 06653 g010
Figure 11. Reflection coefficients (S11 parameter) of S1 for a pressure range of 0 to 100 mmHg.
Figure 11. Reflection coefficients (S11 parameter) of S1 for a pressure range of 0 to 100 mmHg.
Sensors 20 06653 g011
Figure 12. Reflection coefficients (S11 parameter) of S2 for a pressure range of 0 to 100 mmHg.
Figure 12. Reflection coefficients (S11 parameter) of S2 for a pressure range of 0 to 100 mmHg.
Sensors 20 06653 g012
Figure 13. Reflection coefficients (S11 parameter) of S1 for a pressure range between 0 to 300 mmHg.
Figure 13. Reflection coefficients (S11 parameter) of S1 for a pressure range between 0 to 300 mmHg.
Sensors 20 06653 g013
Figure 14. Reflection coefficients (S11 parameter) of S2 for a pressure range between 0 to 300 mmHg.
Figure 14. Reflection coefficients (S11 parameter) of S2 for a pressure range between 0 to 300 mmHg.
Sensors 20 06653 g014
Figure 15. Linear model fitting for a pressure range of 0 to 100 mmHg: (a) Measured response (dotted) and linear fit (solid) of S1; (b) Measured response (dotted) and linear fit (solid) of S2.
Figure 15. Linear model fitting for a pressure range of 0 to 100 mmHg: (a) Measured response (dotted) and linear fit (solid) of S1; (b) Measured response (dotted) and linear fit (solid) of S2.
Sensors 20 06653 g015
Figure 16. Nonlinear model fitting for a pressure range of 0 to 300 mmHg: (a) Measured response (dotted) and linear fit (solid) of S1; (b) Measured response (dotted) and linear fit (solid) of S2.
Figure 16. Nonlinear model fitting for a pressure range of 0 to 300 mmHg: (a) Measured response (dotted) and linear fit (solid) of S1; (b) Measured response (dotted) and linear fit (solid) of S2.
Sensors 20 06653 g016
Figure 17. Repeatability of measurements with S1 over 10 cycles.
Figure 17. Repeatability of measurements with S1 over 10 cycles.
Sensors 20 06653 g017
Figure 18. Repeatability of measurements with S2 over 10 cycles.
Figure 18. Repeatability of measurements with S2 over 10 cycles.
Sensors 20 06653 g018
Table 1. Key design parameters and results for both sensors (S1, S2) and their readers (R1, R2).
Table 1. Key design parameters and results for both sensors (S1, S2) and their readers (R1, R2).
Parameter/ResultS1/R1 (40 mm)S2/R2 (12 mm)
Outer diameter, d o u t (mm)4012
Inner diameter, d i n (mm)205.3
Trace width, w (mm)0.50.2
Trace separation, s (mm)0.50.5
Number of turns, N 105
Capacitor electrode radius, r (mm)72
Dielectric layer thickness, d (μm)200200
Spiral length, l (mm)942.5135.9
Skin depth, δ (μm)15.154.12
DC resistance, R d c (Ω)0.93710.3377
AC resistance, R a c (Ω)3.775517.0536
Relative permittivity of PDMS [3], ϵ r _ PDMS 2.652.65
Calculated resonance frequency, f o _ cal (MHz)19.188259.44
Measured resonance frequency, f o _ meas (MHz)30.843750274.10
Sensitivity, m (kHz/mmHg)−8.110−65.48
Capacitance, C s (pF)18.061.4743
Parasitic capacitance, C p (pF)0.724340.10442
Inductance, L s (μH)3.80950.25572
Self-resonance frequency, f S R F (MHz)95.81974.80
Quality factor, Q F at f o 97.458423.92
Table 2. Coefficients of the polynomial equation ( f ( P ) = m × P + β ; where P is pressure) curve fitting between measured resonance frequencies and applied pressure.
Table 2. Coefficients of the polynomial equation ( f ( P ) = m × P + β ; where P is pressure) curve fitting between measured resonance frequencies and applied pressure.
ParametersS1 (40 mm)S2 (12 mm)
m (sensitivty)−8.11 × 103−65.48 × 103
β ( f o )3.083 × 1072.74 × 108
R 2 (goodness of fit)0.99770.9989
Table 3. Coefficients of 2nd order polynomial ( f ( P ) = a × P 2 + b × P + β ; where P is pressure) curve fitting between measured resonance frequencies and applied pressure.
Table 3. Coefficients of 2nd order polynomial ( f ( P ) = a × P 2 + b × P + β ; where P is pressure) curve fitting between measured resonance frequencies and applied pressure.
ParametersS1 (40 mm)S2 (12 mm)
a 15.2777.12
b −9.797 × 103−8.439 × 104
β ( f o )3.084 × 1072.75 × 108
R 2 (goodness of fit)0.99910.993
Table 4. Mean value and standard deviation of measured resonance frequencies when both sensors were tested under different pressures for 10 cycles.
Table 4. Mean value and standard deviation of measured resonance frequencies when both sensors were tested under different pressures for 10 cycles.
SensorParameterPressure (mmHg)
100806040200
S1Mean, f u (MHz)30.0530.1830.3230.4930.6630.86
Standard deviation, σ (kHz)22.487.485.274.933.013.01
S2Mean, f u (MHz)267.64268.69270.00271.29272.62274.12
Standard deviation, σ (kHz)26.8180.9659.8645.9442.4967.49
Table 5. Comparative analysis of this study with the literature.
Table 5. Comparative analysis of this study with the literature.
StudySensing MechanismMethodologyLinear Operational RangeSensitivityFabrication CostApplication
This studyCapacitiveLaserJet printing, Copper etching, Sandwiching of PDMS layer0–100 mmHg8.11 kHz/mmHg
65.48 kHz/mmHg
LowInterface pressure monitoring during compression therapy
Deng et al. [42]CapacitiveSi wafer moulding, PDMS casting, Conductive printing, Packaging0–200 mmHg270 kHz/mmHgHighWound monitoring
Casey et al. [8]CapacitiveMicro-machining, Si wafer moulding, Electrodes patterning, Component mounting10–80 mmHgN/AHighSub-bandage pressure measurements
Farooqui et al. [40]CapacitiveScreen printing, Conductive printing, Component mounting, Packaging5–100 mmHg0.0523 pF/mmHgMediumSmart bandage for chronic wounds
Fonseca et al. [43]CapacitiveStandard lithography, Wet-chemical etching, Laser-cutting70–120 mmHg5.76 kHz/mmHgHighImplantable pressure sensing
Chen et al. [45]CapacitiveOxide patterning, Coating, and patterning, Metal deposition, Deep Si etching0–100 mmHg160 kHz/mmHgHighIntraocular pressure monitoring
Li et al. [44]CapacitiveMicrofabrication0–50 mmHg0.419 kHz/mmHgHighIntracranial blood pressure monitoring
Rahimi et al. [41]InductiveCorona treatments, Laser patterning of mask, Screen printing of electrodes, Temperature curing0–35%150 kHz% strainHighWound monitoring
Mehmood et al. [39]Resistive (FSR)Off the shelf sensors integration with electronics, Biocompatible coating0–60 mmHgN/AMediumSub-bandage pressure and wound moisture
Burke et al. [4]Resistive (FSR)Off the shelf sensors integration with electronics0–96 mmHg31.27 mV/mmHgMediumSub-bandage pressure during venous compression therapy
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Farooq, M.; Iqbal, T.; Vazquez, P.; Farid, N.; Thampi, S.; Wijns, W.; Shahzad, A. Thin-Film Flexible Wireless Pressure Sensor for Continuous Pressure Monitoring in Medical Applications. Sensors 2020, 20, 6653. https://doi.org/10.3390/s20226653

AMA Style

Farooq M, Iqbal T, Vazquez P, Farid N, Thampi S, Wijns W, Shahzad A. Thin-Film Flexible Wireless Pressure Sensor for Continuous Pressure Monitoring in Medical Applications. Sensors. 2020; 20(22):6653. https://doi.org/10.3390/s20226653

Chicago/Turabian Style

Farooq, Muhammad, Talha Iqbal, Patricia Vazquez, Nazar Farid, Sudhin Thampi, William Wijns, and Atif Shahzad. 2020. "Thin-Film Flexible Wireless Pressure Sensor for Continuous Pressure Monitoring in Medical Applications" Sensors 20, no. 22: 6653. https://doi.org/10.3390/s20226653

APA Style

Farooq, M., Iqbal, T., Vazquez, P., Farid, N., Thampi, S., Wijns, W., & Shahzad, A. (2020). Thin-Film Flexible Wireless Pressure Sensor for Continuous Pressure Monitoring in Medical Applications. Sensors, 20(22), 6653. https://doi.org/10.3390/s20226653

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop