Next Article in Journal
Diagnostic-Quality Guided Wave Signals Synthesized Using Generative Adversarial Neural Networks
Next Article in Special Issue
A New Robotic Endoscope Holder for Ear and Sinus Surgery with an Integrated Safety Device
Previous Article in Journal
Reliable Sarcoidosis Detection Using Chest X-rays with EfficientNets and Stain-Normalization Techniques
Previous Article in Special Issue
Emotion Recognizing by a Robotic Solution Initiative (EMOTIVE Project)
 
 
Article
Peer-Review Record

Validation of an RF Image System for Real-Time Tracking Neurosurgical Tools

Sensors 2022, 22(10), 3845; https://doi.org/10.3390/s22103845
by Carolina Blanco-Angulo, Andrea Martínez-Lozano, Carlos G. Juan, Roberto Gutiérrez-Mazón, Julia Arias-Rodríguez, Ernesto Ávila-Navarro and José M. Sabater-Navarro *
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Sensors 2022, 22(10), 3845; https://doi.org/10.3390/s22103845
Submission received: 8 April 2022 / Revised: 12 May 2022 / Accepted: 17 May 2022 / Published: 19 May 2022
(This article belongs to the Special Issue Assistive Robots for Healthcare and Human-Robot Interaction)

Round 1

Reviewer 1 Report

The paper is very interesting and well presented. However, I have a major concern which authors should try to address.

What happens if you use more realistic phantoms involving human-mimicking tissues? Is the position accuracy affected by the unknown dielectric properties of the scenario? and how? how could you mitigate?

It will be good to add results on this, or at least a detailed discussion.   

  

 

Author Response

Please, see attachment

Author Response File: Author Response.pdf

Reviewer 2 Report

The paper proposed a very interesting subject that can be solved by microwave imaging.

However, there are several concerns regarding imaging and results 

 

1- There are cross talk between antennas that can be presented as Sij such as S12 or erc

however the Authers just provided the S11 and patterns at situation that antenna not used practically

 

2-Such cross talk are really important because this will be source distortions and blurring of imaging, authors required to study it and propose it in imaging equations.

 

3- How Authors find out number of antennas used for imaging are optimum and should not be more or less (Adding spatial radiation elements for improvement of imaging resolution)

 

4-detection error at Figure 17 for soft tissues like human brain seems so high. Please provide more accurate figure and label what’s minimum error and why it’s acceptable.

5- Flow chart or Algorithm of imaging used is required.

 

Author Response

Please see attachment

Author Response File: Author Response.pdf

Round 2

Reviewer 2 Report

The authors addressed my concerns.