Measurement, Validation and Uncertainty of an Experimental Procedure to Characterize the Size-of-Source Effect of Radiation Thermometers, in the Framework of an Industrial Calibration Laboratory
Abstract
:1. Introduction
2. Materials and Methods
2.1. Basic Model for Calculation of SSE Corrections in RT Calibration and Use
2.2. Description of the Experimental Facility
2.2.1. Reference BB
2.2.2. System of Interchangeable Apertures
2.3. Metrological Traceability
2.4. Measurement Process, Model for Analysis of the Results and Uncertainty Calculation
3. Results
4. Discussion
Effecive Apertures in Calibration and Comparison of RT
5. Conclusions
Funding
Institutional Review Board Statement
Informed Consent Statement
Data Availability Statement
Conflicts of Interest
References
- Bureau International des Poids et Mesures (BIPM). The International System of Units (SI), 9th ed.; Bureau International des Poids et Mesures (BIPM): Sèvres, France, 2019. [Google Scholar]
- BIPM; IEC; IFCC; ILAC; ISO; IUPAC; IUPAP; OIML; BIPM Joint Committee for Guides in Metrology. International Vocabulary of Metrology—Basic and General Concepts and Associated Terms (VIM), JCGM 200:2012, 3rd ed.; 2008 version with minor corrections; BIPM: Sèvres, France, 2012. [Google Scholar]
- ISO/IEC 17025:2017; ISO/CASCO Committee on Conformity Assessment, General Requirements for the Competence of Testing and Calibration Laboratories. September 2017 Corrected May 2018; International Organization for Standardization: Geneva, Switzerland, 2018.
- Saunders, P.; Edgar, H. On the characterization and correction of the size-of-source effect in radiation thermometers. Metrologia 2009, 46, 62–74. [Google Scholar] [CrossRef]
- Saunders, P.; Fischer, J.; Sadli, M.; Battuello, M.; Park, C.W.; Yuan, Z.; Yoon, H.; Li, W.; van der Ham, E.; Sakuma, F.; et al. Uncertainty Budgets for Calibration of Radiation Thermometers below the Silver Point. Int. J. Thermophys. 2008, 29, 1066–1083. [Google Scholar] [CrossRef]
- Pušnik, I.; Grgić, G.; Drnovšek, J. System for the determination of the size-of-source effect of radiation thermometers with the direct reading of temperature. Meas. Sci. Technol. 2006, 17, 1330–1336. [Google Scholar] [CrossRef]
- Nutter, G.D. Radiation Thermometers: Design Principles and Operating Characteristics. In Theory and Practice of Radiation Thermometry; Dewitt, D.P., Nutter, G.D., Eds.; Wiley Interscience: New York, NY, USA, 1988; pp. 231–337. [Google Scholar]
- Cywiak, D.; Cárdenas-García, D.; Rodriguez-Arteaga, H. Influence of Size of Source Effect on Accuracy of LWIR Radiation Thermometers. Metrol. Meas. Syst. 2016, 23, 661–667. [Google Scholar] [CrossRef] [Green Version]
- Saunders, P. Calibration and use of low-temperature direct-reading radiation thermometers. Meas. Sci. Technol. 2009, 20, 025104. [Google Scholar] [CrossRef]
- Sakuma, F.; Hattori, S. Establishing a practical temperature standard by using a narrow-band radiation thermometer with a silicon detector. In Temperature: Its Measurement and Control in Science and Industry; Schooley, J.F., Ed.; AIP Conference Proceedings: New York, NY, USA, 1982; Volume 5, pp. 421–427. [Google Scholar]
- Saunders, P. General interpolation equations for the calibration of radiation thermometers. Metrologia 1997, 34, 201–210. [Google Scholar] [CrossRef]
- Saunders, P.; White, D.R. Physical basis of interpolation equations for radiation thermometry. Metrologia 2003, 40, 195–203. [Google Scholar] [CrossRef]
- Bloembergen, P.; Duan, Y.; Bosma, R.; Yuan, Z. The characterization of radiation thermometers subject to the size-of-source effect. In Proceedings of the TEMPMEKO 1996, 6th International Symposium on Temperature and Thermal Measurements in Industry and Science, Torino, Italy, 10–12 September 1996. [Google Scholar]
- Jimeno-Largo, P.; Yamada, Y.; Bloembergen, P.; Villamañan, M.A.; Machin, G. Numerical analysis of the temperature drop across the cavity wall of high temperature fixed points for radiation thermometry. In Proceedings of the TEMPMEKO 2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science, Zagreb, Croatia, 21–25 June 2004. [Google Scholar]
- De Lucas, J. A Simple Geometrical Model for Calculation of the Effective Emissivity in Blackbody Cylindrical Cavities. Int. J. Thermophys. 2015, 36, 267–282. [Google Scholar] [CrossRef]
- BIPM; IEC; IFCC; ISO; IUPAC; IUPAP; OIML; BIPM Joint Committee for Guides in Metrology. Evaluation of Measurement Data Guide to the Expression of Uncertainty in Measurement JCGM 100:2008 (GUM 1995 with Minor Corrections), 1st ed.; BIPM: Sèvres, France, 2008. [Google Scholar]
- Entidad Nacional de Acreditación (ENAC). Accreditation No.16/LC10.007, Rev. 17, Date 15/01/2021. Available online: http://www.enac.es (accessed on 20 July 2022).
- De Lucas, J.; Segovia, J. Uncertainty calculation of the effective emissivity of cylinder-conical blackbody cavities. Metrologia 2016, 53, 61–75. [Google Scholar] [CrossRef]
- ISO 13528:2015; ISO, Technical Committee ISO/TC 69/SC 6 Measurement Methods and Results, Statistical Methods for Use in Proficiency Testing by Interlaboratory Comparison. Aug/2015 Corrected Oct/2016; International Organization for Standardization: Geneva, Switzerland, 2016.
- McMillan, J.L.; Whittam, A.; Rokosz, M.; Simpson, R.C. Towards quantitative small-scale thermal imaging. Measurement 2008, 117, 429–434. [Google Scholar] [CrossRef]
- Saunders, P.; Edgar, H. Size-of-source effect correction for a thermal imaging radiation thermometer. High Temp.-High Press. 1999, 31, 283–292. [Google Scholar] [CrossRef]
- Chrzanowski, K.; Fischer, J.; Matyszkiel, R. Testing and evaluation of thermal cameras for absolute temperature measurement. Opt. Eng. 2000, 39, 2535–2544. [Google Scholar] [CrossRef]
IRCON, UX40P | LAND, C300 | FLUKE, 574CF | |
---|---|---|---|
Focus distance, mm | 700 (focusable) | 500 (focusable) | 300 (fixed) |
εinstr | 1.00 | 1.00 | 1.00 |
dt, mm | 17.5 | 9 | 6 |
Δλ, μm | (8 a 14) | (8 a 13) | (8 a 14) |
Resolution r, °C | 0.1 | 0.1 | 0.1 |
Apertures (Diameter), mm | CSSE, °C | USSE, °C | ΔRT, °C | URT, °C | En |
---|---|---|---|---|---|
300 °C | |||||
30 → 20 | −1.0 | 0.2 | −1.1 | 0.4 | 0.1 |
30 → 40 | 0.4 | 0.1 | 0.4 | 0.1 | |
30 → 50 | 0.6 | 0.1 | 0.6 | 0.1 | |
500 °C | |||||
30 → 20 | −1.8 | 0.3 | −1.9 | 0.7 | 0.2 |
30 → 40 | 0.7 | 0.2 | 0.6 | 0.1 | |
30 → 50 | 1.0 | 0.2 | 0.8 | 0.3 | |
700 °C | |||||
30 → 20 | −2.5 | 0.4 | −2.4 | 1.1 | 0.1 |
30 → 40 | 1.0 | 0.4 | 0.9 | 0.1 | |
30 → 50 | 1.5 | 0.4 | 1.3 | 0.2 | |
900 °C | |||||
30 → 20 | −3.4 | 0.6 | −4.4 | 1.8 | 0.5 |
30 → 40 | 1.4 | 0.5 | 1.6 | 0.1 | |
30 → 50 | 2.0 | 0.5 | 1.9 | 0.1 |
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations. |
© 2022 by the author. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
Share and Cite
de Lucas, J. Measurement, Validation and Uncertainty of an Experimental Procedure to Characterize the Size-of-Source Effect of Radiation Thermometers, in the Framework of an Industrial Calibration Laboratory. Sensors 2022, 22, 8284. https://doi.org/10.3390/s22218284
de Lucas J. Measurement, Validation and Uncertainty of an Experimental Procedure to Characterize the Size-of-Source Effect of Radiation Thermometers, in the Framework of an Industrial Calibration Laboratory. Sensors. 2022; 22(21):8284. https://doi.org/10.3390/s22218284
Chicago/Turabian Stylede Lucas, Javier. 2022. "Measurement, Validation and Uncertainty of an Experimental Procedure to Characterize the Size-of-Source Effect of Radiation Thermometers, in the Framework of an Industrial Calibration Laboratory" Sensors 22, no. 21: 8284. https://doi.org/10.3390/s22218284
APA Stylede Lucas, J. (2022). Measurement, Validation and Uncertainty of an Experimental Procedure to Characterize the Size-of-Source Effect of Radiation Thermometers, in the Framework of an Industrial Calibration Laboratory. Sensors, 22(21), 8284. https://doi.org/10.3390/s22218284