Next Article in Journal
Managing Perceived Loneliness and Social-Isolation Levels for Older Adults: A Survey with Focus on Wearables-Based Solutions
Next Article in Special Issue
Acceleration of Magnetic Resonance Fingerprinting Reconstruction Using Denoising and Self-Attention Pyramidal Convolutional Neural Network
Previous Article in Journal
A Machine-Learning Model for Lung Age Forecasting by Analyzing Exhalations
Previous Article in Special Issue
Effect of Auditory Discrimination Therapy on Attentional Processes of Tinnitus Patients
 
 
Article
Peer-Review Record

Learning a Metric for Multimodal Medical Image Registration without Supervision Based on Cycle Constraints

Sensors 2022, 22(3), 1107; https://doi.org/10.3390/s22031107
by Hanna Siebert *, Lasse Hansen and Mattias P. Heinrich
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Sensors 2022, 22(3), 1107; https://doi.org/10.3390/s22031107
Submission received: 28 December 2021 / Revised: 24 January 2022 / Accepted: 27 January 2022 / Published: 1 February 2022

Round 1

Reviewer 1 Report

In this paper, the authors introduce a deep learning-based method for multimodal 3D image registration without the supervision of labels or handcrafted metrics. The proposed method uses synthetic three-way (triangular) cycles that for each pair of images comprise two multimodal transformations to be estimated and one known synthetic monomodal transform. The network is optimized by minimizing the cycle discrepancy and adapting the synthetic transformation to be close to the real geometric difference of the image pairs. Experiments on 16 paired abdominal CT and MR scans demonstrate the effectiveness of the proposed method.

This paper is well-motivated and the proposed method is technically sound. The idea of cycle consistency has been extensively studied for 2D image registration in both supervised and unsupervised settings. The author introduces this idea to multimodal 3D image registration and formulates an interesting unsupervised 3D registration method. 

The authors should give implementation details in section 3. For example, the authors failed to explain the rotation and translation range of random transformation R_23.    In Fig. 3, "SSD" is used without definition. It’s better to give the definition in the caption.  

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 2 Report

This paper proposes a medical image registration method based on cycle constraints. Experiments demonstrate the effectiveness of the proposed method when compared with several models. Overall, the paper is well presented. I have some comments as follows:

  1. In Figures 1 and 2, it would be better to replace the patterns of triangle and circle with medical images used in the experiments.
  2. As the authors mentioned that the global statistical metrics and local contrast-invariant edge features. However, only mutual information is used in the experiments. How about other local contrast-invariant edge features?
  3. It is recommended to review the self-supervision in other computer vision domains, such as Self-supervised learning of pretext-invariant representations (CVPR-2020), Unsupervised single image deraining with self-supervised constraints (ICIP-2019), etc.
  4. Some symbols lack explanations, such as the Id, Equations (2) and (3). Please check it.

 

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Round 2

Reviewer 2 Report

The authors have addressed my comments.

Back to TopTop