Next Article in Journal
GLASS: A Citizen-Centric Distributed Data-Sharing Model within an e-Governance Architecture
Previous Article in Journal
An Innovative Malaxer Equipped with SCADA Platform for Improving Extra Virgin Olive Oil Quality
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Review

Sensors and Instruments for Brix Measurement: A Review

by
Swapna A. Jaywant
1,
Harshpreet Singh
2 and
Khalid Mahmood Arif
1,*
1
Department of Mechanical and Electrical Engineering, SF&AT, Massey University, Auckland 0632, New Zealand
2
New Zealand Product Accelerator, The University of Auckland, Auckland 1010, New Zealand
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(6), 2290; https://doi.org/10.3390/s22062290
Submission received: 20 November 2021 / Revised: 6 March 2022 / Accepted: 10 March 2022 / Published: 16 March 2022
(This article belongs to the Section Smart Agriculture)

Abstract

:
Quality assessment of fruits, vegetables, or beverages involves classifying the products according to the quality traits such as, appearance, texture, flavor, sugar content. The measurement of sugar content, or Brix, as it is commonly known, is an essential part of the quality analysis of the agricultural products and alcoholic beverages. The Brix monitoring of fruit and vegetables by destructive methods includes sensory assessment involving sensory panels, instruments such as refractometer, hydrometer, and liquid chromatography. However, these techniques are manual, time-consuming, and most importantly, the fruits or vegetables are damaged during testing. On the other hand, the traditional sample-based methods involve manual sample collection of the liquid from the tank in fruit/vegetable juice making and in wineries or breweries. Labour ineffectiveness can be a significant drawback of such methods. This review presents recent developments in different destructive and nondestructive Brix measurement techniques focused on fruits, vegetables, and beverages. It is concluded that while there exist a variety of methods and instruments for Brix measurement, traits such as promptness and low cost of analysis, minimal sample preparation, and environmental friendliness are still among the prime requirements of the industry.

1. Introduction

Fruits and vegetables play a vital role in the human diet as they are rich in carbohydrates, minerals, vitamins, dietary fibres, and antioxidants. The crop quality can be evaluated by various attributes such as appearance, texture, and flavor. The appearance factors are size, shape, and color whereas the factors associated with texture include firmness, crispiness, and juiciness. The flavor is defined by sweetness and aroma. However, harvest maturity is the crucial factor determining the shelf life and final quality of any fruit or vegetable [1,2]. The different stages of harvest maturity are shown in Figure 1. Sugar content is one of the measures used to judge the various stages of maturity [3]. Hence, it is the pivotal factor in agriculture, the food industry, winemaking and brewing. It is a significant indicator in harvest scheduling, postharvest management, and variety selection for farmers. In the food industry, to produce different soft drinks, tomato concentrates, fruit juices, and honey, sweetness control is the prime element. In a winemaking and brewing, the fermentation process converts sugar into ethanol (alcohol) and carbon dioxide with the help of a microfungus called yeast. The measurement of sugar levels in a grape must or a wort allows winemakers and brewers to estimate the alcohol content after the fermentation. This calculation is an essential part of winemaking and brewing that can affect decisions made in the alcohol industry. Dissolved sugar in the fruit juice is measured in terms of Brix, Baume or Oechsle. However, the most widely used unit in the food and beverage industry is Brix. Brix can be measured in the field, on a plant, or at a shop. It indicates the number of dissolved solids in a liquid measured via its specific gravity (SG). One degree Brix is 1 g of sucrose in 100 g of solution (1 Brix = 1% sugar) [4,5,6,7,8].
Since sugar concentration has a linear relationship with SG and density, Brix can be measured via any of these two quantities. The most popular instrument for Brix measurement is either refractometer or a hydrometer, where a refractometer is an optical device, and a hydrometer is a glass device used to measure the specific gravity [4,5,8]. There are various other methods to measure Brix value such as differential measurement, density measurement, osmotic potential measurement, mass flow measurement, optical, ultrasonic, and biosensors. Brix measurement methods used in agriculture and horticulture commodities are generally categorized as destructive and nondestructive types. In the destructive-type method, fruit extract is required to perform the measurement, whereas, in the nondestructive methods, Brix measurement is performed on the fruit without cutting it. Brix measurement methods used for the process industry can be classified as in-line measurement and sample-based measurement methods. The sugar content analysis of fruit and vegetables using destructive methods include sensory assessment involving sensory panels, using instruments such as a refractometer, hydrometer, and liquid chromatograph [9]. However, these techniques are manual, time-consuming, and, most importantly, the crops (fruits and vegetables) are damaged. In the winery, the traditional sample-based methods require sample collection of the liquid from the tank. In a commercial winery, this is a mundane job that involves several man-hours. Therefore, an automatic and continuous Brix measurement can always be an asset to the wine industry. Hence there is a continuous demand for a real-time, rapid, noninvasive, and continuous monitoring approach for the sugar content analysis [9,10].
In this paper, we present different destructive and nondestructive methods for the determination of Brix in the farming industry, along with the advantages and disadvantages of each. Destructive detection covers methods using tools such as a refractometer, hydrometer, electronic tongue, lossy mode resonances, and SPR-based sensors. Nondestructive detection covers visual and near-infrared spectroscopy and magnetic resonance methods. In addition to this, sample-based Brix measurement methods and commercially available in-line measurement techniques used in wineries and breweries are also discussed, highlighting the advantages/disadvantages and limitations.

2. Sample-Based Destructive Methods for Brix Measurement

All the methods discussed below require a juice sample of fruit or vegetable to perform Brix measurement. Hence these are classified as destructive methods in the farming industry. Due to the sampling nature of these techniques, they are also considered to be sample-based techniques, particularly in the process industry.

2.1. Hydrometer

A hydrometer is a traditional instrument that measures Brix in terms of SG. It works on the concept of buoyancy and is usually calibrated at room temperature (20 C) in different units such as Brix, Baume, and SG, as per the requirement [9,11,12]. The working of hydrometer is based on Archimedes’ principle, which states that “the upward buoyant force that is exerted on a body immersed in a fluid, whether fully or partially, is equal to the weight of the fluid that the body displaces”. Hydrometer consists of a wide-bottom (for buoyancy) sealed hollow glass tube, a lead or mercury ballast for stability, and a graduated narrow stem for measurement as shown in Figure 2. Typically, a hydrometer is lowered into the liquid filled graduated cylinder until it floats freely. The point where the surface of the liquid touches the hydrometer’s stem is proportional to the relative density of the liquid.
The use of a hydrometer is one of the most widely used techniques to read the Brix value in the food and farming industry, and wineries and breweries. However, this method is unreliable as it involves a manual operation that can lead to errors such as the incorrect reading of the meniscus. Moreover, temperature compensation is required as the operation is dependent on temperature variation. Sometimes, surface contamination occurs due to the presence of insoluble particles in the liquid and causes a substantial difference in the surface tension, resulting in an additional error [9]. To overcome these inaccuracies, another optical tool known as refractometer is more widely used to measure the sugar level.

2.2. Refractometer

A refractometer is an optical instrument to measure the Brix level in fruits, vegetables, juices, jams, wines and beers. It is a standard technique based on the measurement of the refractive index of the liquid. Refractive index (RI) is a ratio of the speed of light in a vacuum to the speed of light in the given medium. RI depends on the wavelength of the light and the temperature of the sample liquid [13,14]. Different types of refractometers commonly used in fruits and vegetables research are listed in Table 1 to provide a source of quick information for the readers. Most of the refractometers are either based on the principle of refraction or the principle of critical reflection of light. According to [15], critical angle-based RI measurement is more appropriate as it is not influenced by the colour of the sample liquid or the suspended solids in the sample liquid. As a result, a refractometer can be conveniently used to perform quality checks in industries such as food and dairy [13].
Refractometers are available as digital or analog types, in handheld and bench-top versions. Similar to a hydrometer, an analog refractometer needs temperature correction while performing the measurement. However, Ugwu et al. [14] have fabricated a handheld analog refractometer that includes an in-built temperature compensation. This refractometer consists of focus adjustment, calibration screw, daylight plate, eyepiece, rubber grip, and the main prism assembly. A charge-coupled device (CCD) has been used as the sensor to precisely measure the intensity of the reflected light. The refractometer must be calibrated before taking measurements with the help of a calibration screw and distilled water for temperature compensation. The temperature-compensation thermometer is built within the refractometer. The researchers conducted the test using samples of orange, pineapple, and cashew juices at certain levels of pH values. A significant drawback of such a handheld analog refractometer is that a person takes reading at a time that may be subjective.
Digital refractometers perform temperature compensation automatically. However, such refractometers usually display errors if readings are taken too quickly before a temperature equilibrium is reached [9]. The benchtop devices can provide an accurate reading with automatic temperature control from 0 to 100 along with quality control analysis. They are readily available, robust and not influenced by ambient conditions. Compared to benchtop devices, handheld devices are cost-effective and portable. The most common commercially available refractometers are summarised in Table 2, along with their specifications.
Figure 3 pistorially describes the advantages and limitations of different types of refractometers (analog, digital, and benchtop refractometers). Advantages of analog refractometers include portability and cost-effectiveness, whereas the subjective nature of their readings is their major limitation. Digital refractometers are handheld, cost-effective, and more accurate compared to analog refractometers. Benchtop refractometers are more accurate compared to analog and digital refractometers, however, they are more expensive than other types of hydrometers.

2.3. Other Refractive Index (RI)-Based Destructive Methods

Dongare et al. [13] have developed an RI-based optical Brix meter using an equilateral angle prism. The optical configuration of the developed system includes the use of LED (light-emitting diode) as a source and LDR (light-dependent resistor) as an optical detector. The variation in concentration of sugar causes variation in the RI, and, therefore, the critical angle of the total reflection changes. This, in turn, produces a change in the LDR resistance. The range of the Brix meter is from 1 to 35 Brix. The possibility of another RI method to measure the sugar content has been introduced by Seki et al. [15]. They have investigated RI measurement in sucrose solution with the help of a heterocore-structured fibre optic-based SPR sensor. The sensor was developed by depositing a silver thin film on the surface of a heterocore-structured fiber optic. The result obtained by the sensor confirmed an excellent alignment with the conventional refractometer readings. Furthermore, Zubiate et al. [32] fabricated a Brix sensor that can be used in the real-life applications. The working principle of the sensor is based on Lossy-Mode Resonances (LMRs). They deposited thin indium tin oxide (ITO) films onto D-shaped optical fibers. The sensor can monitor from 0 to 50 Brix. The device has a linear response with a linearity factor R2 0.9961 and 11.8 nm/ Brix sensitivity (~0.005 Brix resolution).

2.4. Electronic Tongue

One of the conventional methods for quality measurement of fruit, vegetables, different types of juices, and wines is the use of a sensory panel [33]. In this method, a qualified group of panelists evaluates the product’s taste based on different qualities such as sweetness, sourness, bitterness, and aroma. The study is always influenced by subjective readings, the lack of standard methods to rank the product, problems of stability, and reproducibility. It is a time-consuming method and also involves a high cost [9,33,34]. Nevertheless, this has inspired many researchers to develop an electronic device (commonly known as an electronic tongue) that can be utilized as artificial human taste buds.
An electronic tongue utilizes an array of sensors based on potentiometric, voltammetric, impedance spectroscopy, and piezoelectric working principle in conjunction with different pattern-recognition methods as shown in Figure 4 [35,36,37,38]. However, the potentiometric principle is more commonly used than the other principles. It makes use of ion-selective electrodes to measure the potential change against a reference electrode in zero-current conditions. Low cost, rapid response, and easy-to-handle measurement setup are the main advantages of any electronic tongue.
Toko and coworkers were the pioneers of the electronic tongue [35]. Since then, many research groups have widely used many electronic tongues for quality analysis of fruits, vegetables, fruit juices, soft drinks, tea, coffee, alcohol, and herbal products. Beullens et al. [39] used an electronic tongue to determine the sugar and acid profile of four tomato cultivars. An array of 15 potentiometric sensors were checked against the high-performance liquid chromatography (HPLC) result. To study the potential of the electronic tongue to predict the concentrations, partial least squares (PLS) regression models were built. PLS study was undertaken using the results of the electronic tongue data and the HPLC measurements. However, the sensor was found to be less sensitive during sugar measurement as it only achieved a correlation of 0.69, 0.53, and 0.63 with sucrose, glucose, and fructose, respectively. Bett-Garber et al. [40] demonstrated the possibility of using an electronic tongue to monitor processed juice quality by comparing six different brands of pomegranate juice. In another study [41], an electronic tongue was applied for monitoring the consequences of postharvest techniques on the fruit-ripening process. In the comparison of instrumental analysis and sensory analysis it was found that the instrumental results were more sensitive to differences than the sensory analysis. Furthermore, Campos et al. [42] demonstrated the capability of an electronic tongue to measure the grape ripeness parameters (Total Acidity (TA) and Brix) parameters that are useful to estimate the harvesting date. They used voltammetry along with metallic electrodes. Similarly, Lozano et al. [43] fabricated a similar electronic sensing device to identify different wines’ ageing processes. Pattern recognition has been performed using both linear techniques such as principal component analysis, (PCA) and nonlinear techniques such as probabilistic neural networks (PNN). These systems achieved a classification success rate of 97% with PCA and 84% with PNN during the detection of different ageing processes. All these researched RI-based sensors are summarized in Table 3.

3. Nondestructive Methods for Brix Measurement in Fruits and Vegetables

3.1. vis/NIR Spectroscopy

In the farming industry, visible and near-infrared (vis/NIR) spectroscopy demonstrate significant potential for sorting and grading fresh produce as per their internal attributes [10,44,45]. Over the last 2 decades, across the world, a large number of studies have been conducted to predict the sugar content in various fruits and vegetables with the help of vis/NIR techniques [44,46,47,48,49,50,51,52,53,54,55,56,57,58]. The method uses visible and NIR range of the electromagnetic spectrum, covering approximately 300–2500 nm to measure the sugar content of different fruits such as apples, bananas, pomegranates, kiwifruit, mandarins, mangoes, oranges, pineapple, sugarcane, and tomatoes (Table 4). The NIR spectral acquisition from the sample is performed with the help of three standard modes—transmission, reflectance, and interactance [59]. In the transmission mode, while infrared light is passing through a sample, some part of the light can be absorbed by the sample and the remaining fraction is measured on the other side of the sample. Alternatively, light can be reflected from the sample and the absorption properties can be obtained from the reflected light in the reflectance mode. In the transmission mode, a light source and a light detector are placed on either side of the sample, whereas in reflectance mode the source and the detector are placed on the same side of the sample. The interactance mode is similar to the reflectance mode, however, the detector is separated from the illuminated sample by the light seal [60]. In the case of reflectance mode, the spectral acquisition is easier with a relatively high-intensity level, but the shortcoming of this mode is that it is prone to the variations in shallow subsurface or on-surface properties of the fruit; on the other hand, the transmission mode may be less sensitive to surface properties, but it has a reduced signal-to-noise ratio as the amount of light penetrating the flesh is very small [60]. Hence it becomes difficult to pick up accurate transmission readings at grading line speeds, especially where high surrounding light is present. Interactance mode can sort out the problems related to reflection and transmission modes. Hence it can be the most effective type of analysis. However, this mode requires proper light sealing, which is challenging with the high-speed conveyors used in current fruit-sorting systems [44].
To evaluate the Brix value of a variety of apple fruits Angra et al. [61] used a filter-based NIR technique where the in-house developed probe recorded the reflectance wavelengths. A standard error of prediction for Brix values of all varieties of apples was between 0.7322 to 1.7809. Sripaurya et al. [46] developed a portable 6-digital-channel NIR device to predict the quality of Gros Michel bananas and grade them by maturity stage. They also proposed a novel automatic classification of banana-ripening classification results, they where classified up to seven ripeness levels with nearly 100% accuracy. Liu et al. [49] explored the viability of the CCD-NIRS (charge-coupled device—near infrared spectroscopy) technique combined with interval partial least squares (iPLS) algorithm for measuring the soluble solid content (SSC) and TA of Nanfeng mandarin fruits. The author observed the improved predictive ability of the SSC model with r of 0.92 and RMSEP (root mean square errors of prediction) of 0.65 Brix using effective wavelengths of 681.36–740.51 nm, 798.60–836.19 nm, and 945.52–962.75 nm. Golic et al. [62] projected glucose, fructose, and sucrose in citrus fruit through reflectance vis/NIRS technique. The predictability of glucose and fructose was much more than sucrose in this analysis. The authors thought that the difference in the molecular weights of sucrose (MW = 342.30 g/mol) when compared to glucose and fructose (MW = 180.16 g/mol) may be a possible reason for the lower predictability of sucrose. The number of sucrose molecules is 1.89 times fewer than glucose and fructose molecules in the same weight of the sample.
Though a nondestructive NIR spectroscopic approach looks promising for the quality analysis of fruits, it requires complex and expensive instrumentation [50]. A few groups attempted successfully to create an in-house portable NIR system to counter the cost of the commercially available systems [50,51,52,53,54,55,56]. Saranwong et al. [55] compared the performance of a commercially-available, portable NIR instrument “FT20” with a research-type spectrophotometer for fruit-quality evaluation using mango fruits and found similar accuracy in both.
Even if the spectroscopic technique is considered the most viable method for nondestructive quality analysis of fruits and vegetables, it still has some limitations. The major disadvantage is that the method is based on reference values that may be susceptible to errors. It requires a new calibration model for every fruit species. Regarding changes in season and location, it is desirable to upgrade calibration models according to these [63]. Still, with the recent developments in vis/NIRS techniques, the technology has great potential in the industry for sorting and grading the produce.

3.2. Magnetic Resonance Methods

Magnetic resonance imaging (MRI) is a noninvasive method based on nuclear magnetic resonance (NMR) principles. It can provide highly resolved spatial information related to the distribution and environment of water in soft tissues [64]. The molecular elements of fruits consist of magnetic resonance (MR) active nuclei that can be utilized to sense the different quality attributes of the fruits, such as texture, taste, internal structure, Brix number [65]. MR attributes such as proton density, relaxation times, chemical shifts, and diffusion constants have the potential for quality analysis of fruits. The relaxation times can be utilized to assess the fruit-ripening stage, fruit firmness, as those are sensitive to mobility. A time-domain analysis is used in the study of relaxation and diffusion data [66]. Table 5 summarises the use of relaxation time for different applications reported in various research.
Andaur et al. [71] studied the growth and ripening of grape berries for three varieties using MRI. Their study confirms that it is possible to visualize the internal features of berries with such a nondestructive technique and to acquire the information related to volume and Brix distribution within a cluster. Zhang and McCarthy [67,68] used the MRI technique to investigate the quantitative analysis of pomegranate quality and tomato maturity. They could relate the soluble solids content of pomegranate with Spin–spin relaxation time, T2, measured using a low magnetic field (0.04 T). Similarly, a detailed study of biophysical, histological, and metabolic traits of fruit development was reported by Musse et al. [69] with a dependable explanation of the changes in relaxation parameters. Hence, MRI can be used to provide information on cell expansion in a noninvasive manner.
A few portable, dedicated MRI equipment are available for various applications in the food and farming industry. Additional attempts at miniaturization are in progress by using microfluidic technology [72] and integrating all the required RF electronics into a single board. Such efforts have been demonstrated by Sun et al. [73]. Windt and Blumler [74] have developed a portable NMR sensor to measure the dynamic changes in fruits.
Even though NMR spectroscopy and MRI have been proved to be an effective nondestructive method for the quality evaluation of fruits, they have found only limited applications in modern industrial and commercial applications. The size, weight, cost, and stability of magnets and the throughput are the major constraints for the wider application of MR-based technology [66]. The specifications of nondestructive Brix sensors are summarized in Table 6.

4. Brix Measurement in Winemaking and Brewing

In wineries and breweries, optimised process control is the prime factor for enhancing wine or beer quality. This includes the quality assessment of physical properties of raw materials (e.g., grape juice/malt), process streams (e.g., ferments), and end products (e.g., wine/beer) [75]. The fermentation process control is crucial during the production of wine or beer to monitor the conversion from raw juice to alcohol [76]. It is important to observe the Brix level at the various stages of beer and wine production (Figure 5). In this section, different types of Brix monitoring methods that are used during the production of wine and beer are discussed.

4.1. Spectroscopic Methods

Over the last few years, it has been found that vis-NIR spectroscopy can be a potential method to observe the conversion process and anticipate the concentration during wine fermentation in real time [77,78]. Currently, in the winery, the principal application of this method is the calculation of alcohol and Brix content with the help of modern filter instruments such as monochromator and diode array spectrophotometers together with advanced algorithms and software for wine analysis [78]. It has been reported that many big Australian wine companies are utilizing NIR spectroscopy to perform composition analysis of red wine [78]. Investigation trials during the wine fermentation process also revealed that these methods could predict the concentration and monitor the extraction and evolution of phenolic compounds during red wine fermentation. Analysis of different wines using vis-NIR spectroscopy by various authors is summarised in Table 7. Similarly, Grassi et al. [79] demonstrated the possibility of FT-NIR spectroscopy combined with multivariate data analysis to monitor and assess process parameters in beer fermentation at different operative conditions.

4.2. Ultrasonic Methods

For the past few years, ultrasonic-based instruments have been used to control a variety of industrial processes. These days, ultrasonic technology is considered to be a relevant method to monitor quality control in food processes due to its qualities such as nondestructiveness and easy automation [85]. Becker et al. [86] used ultrasonic methodology in developing a density-measuring device for monitoring beer fermentation. The possibility of obtaining a theoretical relation between density and ultrasonic propagation velocity was rejected by the authors, being aware of the lack of a representative model for liquid mixtures and solutions [87]. Using an ultrasonic velocity measurement technique, Resa et al. [87] studied the alcoholic fermentation of several carbohydrate aqueous solution. From their study, they derived that during glucose (or fructose) fermentation, the density and sound velocity measurements can relate both the sound velocity and the density to the concentration of the main components. Lamberti et al. [85] also presented a similar work for monitoring the alcoholic wine fermentation. They observed a steady decrease in propagation velocity as the fermentation process progressed. They confirmed that the variation of the propagation velocity is exclusively due to the decreasing of the sucrose concentration.

4.3. Biosensors

Electrochemical biosensor devices have become more prevalent in the food processing industry in the last 20 years due to their attractive performances. The electrochemical biosensing methods can provide an accurate, highly selective, sensitive, and potentially automatic analysis. It is a widely known fact that the traditional legacy-based analytical techniques are time-consuming and often require specific expensive [88] equipment, hence electrochemical biosensor devices have emerged as a promising alternative. Miniaturization and the development of hand-held devices for real-time monitoring is easily achievable with this technology. Additionally, full automation is being custom developed for continuous online or periodic monitoring of various processes in the food industry [88].
Most of the time, electrochemical biosensor devices are based on the amperometry principle, where electrochemical oxidation or reduction of an electroactive species produces the current and the resulting current is proportional to the concentration of the analyte. The basic working principle and the key enzyme reactions used in biosensors to determine the analytes are shown in Figure 6. The measurement system generally includes three electrodes (working, reference, and auxiliary electrodes). The working electrode is either metal (Pt and Au) or carbon-based material (glassy carbon, graphite, screen-printed, or carbon paste electrode) and is usually combined with immobilized enzymes [88].
For the online detection of ethanol, glucose, and glycerol in wines, Niculescu et al. [89] proposed dehydrogenase-based biosensors. However, the fabrication process of these devices required isolation and purification of PQQ-dependent dehydrogenase and the synthesis of the electrochemical mediator (PVI13dmeOs), which was a painstaking process. Esti et al. [90] developed conventional amperometric biosensors inserted into flow injection analysis (FIA) systems for detection of glucose, fructose, and ethanol. The system worked effectively for real-time monitoring of alcoholic fermentation in industrial-scale red wine-making. Still, overall, it was somewhat inconvenient to use. To overcome all these limitations, Goriushkina et al. [91] developed amperometric biosensors to detect glucose and ethanol in wine in a more convenient way. The used specific enzymes such as alcohol and glucose oxidase (AO and GOx) in the biosensor. The working potential was chosen to counter high biosensor sensitivity and low current signal due to electrochemical interference [92].
Platinum or carbon (doped with cobalt phthalocyanine) screen-printed electrodes and alcohol oxidase were used to develop other biosensor devices to detect the ethanol in wine and beer [92,93]. These devices are required to accommodate the different strategies to avoid interference that was caused due to the application of high potential. Piermarini et al. [92] used graphite screen-printed sensors modified with Prussian Blue (PB) to develop the biosensors for the detection of both sugars as well as ethanol to overcome the limitation. In addition, an amperometric ferrocene-mediated glucose biosensor with a Nafion protective film operating at 0.25 V (vs. Ag/AgCl, 3 M KCl) was developed to determine the glucose content in wine [94]. Likewise, neutral red [95] and osmium polymers [88] were used as mediators for glucose analysis in wines and alcoholic beverages.

4.4. Commercially Available In-Line Brix Sensors

In-line Brix sensing can be effectively used to monitor the progress of the wine-fermentation process. These sensors allow more frequent measurements that can provide data more effectively to winemakers. Based on this data, winemakers can make informed process decisions more frequently and in a much shorter time frame, which improves the productivity and reduces the risk of sluggish ferments. A few such commercially available in-line Brix sensors are discussed in the following paragraphs and summarised in Table 8 and the specifications of discussed in-line Brix sensors are listed in the Table 9.
Liquiphant M vibrating fork: This sensor is based on the density measurement principle. It includes a tuning fork that is excited to the resonance frequency. The tuning fork’s oscillating frequency changes according to the density of the liquid in which it is immersed. The shift in frequency is converted into a proportional output signal with the help of a density computing unit. Simultaneously up to four different sensors can be connected to the computing unit [96].
Fermetrol probe: A semipermeable water-responsive polymer is used in this sensor to evaluate the osmotic potential caused by the concentration of sugar in the must. As fermentation progresses, alcohol concentration increases. The polymer responds to the increased concentration of alcohol by expansive pressure in the polymer. The expansive pressure of the polymer is converted to an electrical signal with the help of a highly stable ceramic membrane. The output signal can be displayed locally on the unit or fed into a control system [96].
Micro-LDS: This sensor determines mass-flow utilizing the Coriolis effect. It consists of a small hollow silicon tube as the sensing element. Vibrating at very high frequencies, the sensor detects variations in fluid density as a shift in tube frequency. The sensor has an integrated temperature sensor to provide the fluid temperature reading. The measured density is converted to Baume and can be displayed locally or fed into a control unit. The sensor also has a solenoid valve and filter mechanism to collect the fermented sample physically [96].
VS-3000: This is a solid-state type of sensor based on the absorbance principle. It passes a mid-infrared signal into the fluid and measures the absorbance of the liquid medium. VS-300 has an integrated predefined calibration software to convert the measured signal into a corresponding concentration. The sensor can measure alcohol level, CO2, specific gravity, and Baume [96].
Tilt-hydrometer: This is a free-floating digital hydrometer specifically designed for home brewing. This instrument allows the reading of the specific gravity and temperature on various Bluetooth devices such as Apple iPhone/iPad or Android smartphone/tablet, or Tilt Pi by downloading the free app on the device. It comes with a calibrated and ready-to-use preinstalled battery. The battery life is approximately 12–14 months, depending on the usage. It is also possible to log the data to the cloud using the Google Sheets template. The accuracy is ±0.002 within the Tilt’s range of 0.990 to 1.120. The thermometer is accurate ±1 F (±0.5 C) [101].

5. Discussion and Outlook

In this modern era, awareness of health has gained importance in communities across the world. People are more interested in consuming good-quality food products. Hence, grading and sorting of the product have become an essential task in the food industry. Various quality assessment procedures have been implemented with the help of modern technology. In this review paper, different Brix sensors and methods used for sugar measurement in agriculture, vineyards, wineries and breweries have been discussed in detail.
The methods used to detect the sugar concentration in fruit and vegetables are broadly classified as destructive and non-destructive methods (Figure 7). All the conventional methods such as hydrometry, refractometry, and chromatography come under the umbrella of destructive methods. A few modern technologies such as electronic tongue and SPR-based sensors are also considered to be destructive methods. Though the hydrometer is a widely used, cost-effective, and handheld instrument for Brix measurement, it includes a manual operation that may lead to errors such as the incorrect reading of the meniscus. As its operation is temperature-dependent, temperature compensation has to be incorporated in the final reading. Sometimes, considerable changes in the surface tension due to the surface contamination could lead to additional errors. These limitations are countered by an optical device called a refractometer. This device can be further classified between analog, digital, and benchtop types. An Analog refractometer is a low-cost handheld device that provides a subjective reading and may include some inaccuracies. These drawbacks are covered in digital- and benchtop-type refractometers by compromising on the cost. Electronic tongue is another cost-effective and portable solution for a quick Brix analysis, however, its low selectivity is a potential drawback. Even though most of these destructive techniques are widely used due to their advantages, such as cost-effectiveness, portability, and quick response times, the significant disadvantage of these methods is that the analysis only provides the attributes of the specific product being examined that makes the quality measurement a sample-based evaluation. Another drawback is that the method requires an extract from a fruit or a vegetable for the analysis, resulting in the crushing of the fruits or vegetables. Hence, the use of a nondestructive method becomes a potential requirement in the fruit and vegetable analysis.
Methods such as vis/NIR spectroscopy and magnetic resonance are considered to be widely researched nondestructive techniques. The spectroscopic method is considered to be a candidate method for the nondestructive quality analysis of fruits and vegetables. The recent developments and research in vis/NIR techniques have made it a portable device that can provide rapid and reliable measurements. Hence it is considered to be the best method for nondestructive quality analysis of fruits and vegetables. However, the technique comes with some limitations. It is based on reference values that may be susceptible to errors. Additionally, it requires a new calibration model for every fruit species. The up-gradation of calibration models according to the change in season and location is required as well. On the other hand, though NMR spectroscopy and MRI have been proved effective non-destructive methods for quality evaluation of fruits, they have found only limited use in modern industrial and commercial applications. Currently, the size, weight, cost, and stability of magnets and the throughput are significant constraints for the wider application of MR-based technology.
In wineries and breweries, the fermentation process control plays an important role during the production of wine or beer. Hence, it is necessary to observe the Brix level during beer and wine production stages. Methods for Brix monitoring in the winery and brewery can be categorized as sample-based and in-line monitoring (Figure 8). A hydrometer and a refractometer are traditional sample-based Brix measurement methods used in the winery and brewery. Various research group have developed many electronic methods. The main disadvantage of these methods is that they require a manual sample collection of the liquid from the tank. In a large commercial winery, this tedious job causes a delay in the results and produces chemical waste. Thus, in-line measurement for continuous Brix monitoring is the prime requirement at the winery and brewery.
Currently, in the research community, biosensors are popular to achieve in-line continuous Brix monitoring. The main reason for their popularity is that biosensors do not require extensive sample preparation. A suitable dilution is usually sufficient; the high specificity and wide linear dynamic range allow direct quantification of the target analyte. However, parameters such as the accuracy of measurements, sensor-to-sensor reproducibility, and operational lifetime are substantially influenced by enzyme stability. That means enzyme immobilization is the key factor in the development of efficient biosensors. Sometimes, the adoption of different strategies to eliminate electrochemical interference due to high applied potential is required. Many commercially available in-line measurement methods have been discussed in the paper. Most of them are turned to be an expensive solution for larger wineries. Some of them are complex to install. Keeping the sensor clean during and after fermentation is another problem. Overall, a rapid, sensitive, nondestructive, and in-line Brix monitoring method is currently needed in the industry.

6. Conclusions

Brix measurement is an essential part of the quality analysis of agricultural products and alcoholic beverages. Different Brix measurement methods used in the farming industry, winemaking, and brewing, such as destructive, nondestructive, and sample-based, are reviewed here, along with their advantages and disadvantages. Additionally, commercially available in-line measurement Brix measurement techniques are also discussed. Existing methods for measuring the composition and quality of fruit, vegetables, and alcoholic beverages are not adequate for the demands of production in a global market, because of their cost and slow turnaround time. Factors such as promptness and low cost of analysis, minimal sample preparation, and environmentally friendly methods for Brix measurement are being striven towards by the food, winemaking, and brewing industries.

Author Contributions

Conceptualization, methodology, investigation, S.A.J., H.S. and K.M.A.; writing—original draft preparation, S.A.J.; writing—review and editing, S.A.J., H.S. and K.M.A.; visualization, S.A.J. and H.S.; supervision, project administration, K.M.A. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by the New Zealand Product Accelerator (Grant No. RM23459).

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Abbreviations

The following abbreviations are used in this manuscript:
DMDry matter
HPLCHigh-performance liquid chromatography
LMRLossy mode resonance
PCAPrincipal component analysis
RIRefractive index
SGSpecific gravity
SSCSoluble Solid Content
TATotal Acidity
vis/NIRSVisible/near-infrared spectroscopy

References

  1. Prasad, K.; Jacob, S.; Siddiqui, M.W. Fruit maturity, harvesting, and quality standards. In Preharvest Modulation of Postharvest Fruit and Vegetable Quality; Academic Press: Cambridge, MA, USA, 2018; pp. 41–69. [Google Scholar]
  2. Costa, M.V.; Fontes, C.H.; Carvalho, G.; Júnior, E.C.d.M. UltraBrix: A Device for Measuring the Soluble Solids Content in Sugarcane. Sustainability 2021, 13, 1227. [Google Scholar] [CrossRef]
  3. Wilson, D. Chemical Sensors for Farm-to-Table Monitoring of Fruit Quality. Sensors 2021, 21, 1634. [Google Scholar] [CrossRef] [PubMed]
  4. Son, H.; Hong, Y.; Park, W.; Yu, M.; Lee, C. A Novel Approach for Estimating Sugar and Alcohol Concentrations in Wines Using Refractometer and Hydrometer; Wiley: Hoboken, NJ, USA, 2009; Volume 74, pp. C106–C111. [Google Scholar]
  5. Puangngernmak, N.; Srisurat, T. The Experiment of The Simple High-Frequency Sensor to Detect The Degree of Brix in Sucrose Solutions. In Proceedings of the 2019 16th International Conference on Electrical Engineering/Electronics, Computer, Telecommunications and Information Technology (ECTI-CON), Pattaya, Thailand, 10–13 July 2019; pp. 760–762. [Google Scholar]
  6. Jacobson, J.L. Introduction to Wine Laboratory Practices and Procedures; Springer Science & Business Media: Berlin, Germany, 2006. [Google Scholar]
  7. Silva, S.d.A.; De Queiroz, D.; de AC Pinto, F.; Santos, N. Coffee quality and its relationship with Brix degree and colorimetric information of coffee cherries. Precis. Agric. 2014, 15, 543–554. [Google Scholar] [CrossRef]
  8. Cao-Paz, A.M.; Acevedo, J.M.; Gandoy, J.D.; del Rio Vazquez, A.; Freire, C.M.P.; Soria, M.L. Plastic optical fiber sensor for real time density measurements in wine fermentation. In Proceedings of the 2007 IEEE Instrumentation & Measurement Technology Conference IMTC 2007, Warsaw, Poland, 1–3 May 2007; pp. 1–5. [Google Scholar]
  9. Magwaza, L.S.; Opara, U.L. Analytical methods for determination of sugars and sweetness of horticultural products—A review. Sci. Hortic. 2015, 184, 179–192. [Google Scholar] [CrossRef]
  10. Magwaza, L.S.; Opara, U.L.; Nieuwoudt, H.; Cronje, P.J.; Saeys, W.; Nicolaï, B. NIR spectroscopy applications for internal and external quality analysis of citrus fruit—A review. Food Bioprocess Technol. 2012, 5, 425–444. [Google Scholar] [CrossRef]
  11. Peña-Perez, L.M.; Pedraza-Ortega, J.C.; Ramos-Arreguin, J.M.; Tovar-Arriaga, S.; Aceves-Fernandez, M.A.; Becerra, L.O.; Gorrostieta-Hurtado, E.; Vargas-Soto, J.E. Alignment of the Measurement Scale Mark during Immersion Hydrometer Calibration Using an Image Processing System. Sensors 2013, 13, 14367–14397. [Google Scholar] [CrossRef] [Green Version]
  12. Lorefice, S.; Malengo, A. Calibration of hydrometers. Meas. Sci. Technol. 2006, 17, 2560. [Google Scholar] [CrossRef]
  13. Dongare, M.; Buchade, P.; Shaligram, A. Refractive index based optical Brix measurement technique with equilateral angle prism for sugar and Allied Industries. Optik 2015, 126, 2383–2385. [Google Scholar] [CrossRef]
  14. Ugwu, K.; Mbajiorgu, C.; Okonkwo, W.; Ani, A. Design, fabrication and performance evaluation of a portable hand-held refractometer. Niger. J. Technol. 2018, 37, 537–542. [Google Scholar] [CrossRef]
  15. Seki, A.; Narita, K.; Watanabe, K. Refractive index measurement in sucrose solution and beverage using surface plasmon resonance sensor based on hetero-core structured fiber optic. Procedia Chem. 2016, 20, 115–117. [Google Scholar] [CrossRef] [Green Version]
  16. Khanom, M.M.; Ueda, Y.; Ishimaru, M. Relationship between volatiles and other factors indicating quality of melon (cucumis melo l. cv. prince melon) during fruit development and storage. Sci. Rep.-Grad. Sch. Agric. Biol. Sci. Osaka Prefect. Univ. 2003, 55, 7–14. [Google Scholar]
  17. Tosun, I.; Ustun, N.S.; Tekguler, B. Physical and chemical changes during ripening of blackberry fruits. Sci. Agric. 2008, 65, 87–90. [Google Scholar] [CrossRef]
  18. Migliorini, M.; Cherubini, C.; Mugelli, M.; Gianni, G.; Trapani, S.; Zanoni, B. Relationship between the oil and sugar content in olive oil fruits from Moraiolo and Leccino cultivars during ripening. Sci. Hortic. 2011, 129, 919–921. [Google Scholar] [CrossRef]
  19. Bureau, S.; Ruiz, D.; Reich, M.; Gouble, B.; Bertrand, D.; Audergon, J.M.; Renard, C.M. Application of ATR-FTIR for a rapid and simultaneous determination of sugars and organic acids in apricot fruit. Food Chem. 2009, 115, 1133–1140. [Google Scholar] [CrossRef]
  20. Drogoudi, P.D.; Vemmos, S.; Pantelidis, G.; Petri, E.; Tzoutzoukou, C.; Karayiannis, I. Physical characters and antioxidant, sugar, and mineral nutrient contents in fruit from 29 apricot (Prunus armeniaca L.) cultivars and hybrids. J. Agric. Food Chem. 2008, 56, 10754–10760. [Google Scholar] [CrossRef] [PubMed]
  21. Álvarez-Fernández, A.; Paniagua, P.; Abadía, J.; Abadía, A. Effects of Fe deficiency chlorosis on yield and fruit quality in peach (Prunus persica L. Batsch). J. Agric. Food Chem. 2003, 51, 5738–5744. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  22. Bell Technology Ltd. HENAN Refractometers. Available online: http://www.belltechnology.co.nz/Portable/refractometershe.html (accessed on 23 June 2021).
  23. Test Equip. Available online: https://www.testequip.com.au/starr-brm-brix-sugar-refractometer (accessed on 23 June 2021).
  24. Bell Technology Ltd. KEM Portable Brix Meter BX-1. Available online: http://www.belltechnology.co.nz/Portable/refractometerska.html (accessed on 23 June 2021).
  25. Test Equip. Available online: https://www.testequip.com.au/atago-pal-grape-must-brix-wine-refractometer-cat3880 (accessed on 23 June 2021).
  26. Mettler Toledo. Available online: https://www.mt.com/nz/en/home/products/Laboratory_Analytics_Browse/refractometer/portable-refractometer/refracto-30gs.html (accessed on 23 June 2021).
  27. Mettler Toledo. Available online: https://www.mt.com/nz/en/home/products/Laboratory_Analytics_Browse/refractometer/refractometers/R4.html (accessed on 23 June 2021).
  28. Anton-Paar. Available online: https://www.anton-paar.com/corp-en/products/details/abbemat-heavy-duty-line/ (accessed on 23 June 2021).
  29. Fisher Scientific. Available online: https://www.fishersci.com/shop/products/laxco-rbd-6000-series-digital-abbe-benchtop-brixri-refractometer/RBD6000 (accessed on 23 June 2021).
  30. Fisher Scientific. Available online: https://www.fishersci.com/shop/products/bench-dig-refract-brix-ri-0-60/rhdb105 (accessed on 23 June 2021).
  31. Anton-Paar. Available online: https://www.fishersci.com/shop/products/abbemat-3100/NC1818326 (accessed on 23 June 2021).
  32. Zubiate, P.; Zamarreño, C.; Matias, I.; Arregui, F. Optical fiber brix sensor based on lossy mode resonances (LMRs). In Proceedings of the SENSORS, Valencia, Spain, 2–5 November 2014; pp. 36–38. [Google Scholar]
  33. O’Mahony, M. Taste perception, food quality and consumer acceptance. J. Food Qual. 1991, 14, 9–31. [Google Scholar] [CrossRef]
  34. Aleixandre, M.; Santos, J.P.; Sayago, I.; Cabellos, J.M.; Arroyo, T.; Horrillo, M.C. A wireless and portable electronic nose to differentiate musts of different ripeness degree and grape varieties. Sensors 2015, 15, 8429–8443. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  35. Ciosek, P.; Wróblewski, W. Sensor arrays for liquid sensing–electronic tongue systems. Analyst 2007, 132, 963–978. [Google Scholar] [CrossRef]
  36. Gallardo, J.; Alegret, S.; Del Valle, M. Application of a potentiometric electronic tongue as a classification tool in food analysis. Talanta 2005, 66, 1303–1309. [Google Scholar] [CrossRef]
  37. Németh, D.; Balázs, G.; Daood, H.G.; Kovács, Z.; Bodor, Z.; Zinia Zaukuu, J.L.; Szentpéteri, V.; Kókai, Z.; Kappel, N. Standard analytical methods, sensory evaluation, NIRS and electronic tongue for sensing taste attributes of different melon varieties. Sensors 2019, 19, 5010. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  38. Jiang, H.; Zhang, M.; Bhandari, B.; Adhikari, B. Application of electronic tongue for fresh foods quality evaluation: A review. Food Rev. Int. 2018, 34, 746–769. [Google Scholar] [CrossRef]
  39. Beullens, K.; Kirsanov, D.; Irudayaraj, J.; Rudnitskaya, A.; Legin, A.; Nicolaï, B.M.; Lammertyn, J. The electronic tongue and ATR–FTIR for rapid detection of sugars and acids in tomatoes. Sens. Actuators B Chem. 2006, 116, 107–115. [Google Scholar] [CrossRef]
  40. Bett-Garber, K.L.; Watson, M.A.; Lea, J.M.; Bai, J.; Baldwin, E.; Raithore, S. Efficacy of monitoring the sensory taste characteristics in pomegranate juice with electronic tongue and chemical measurements. J. Food Qual. 2014, 37, 383–394. [Google Scholar] [CrossRef]
  41. Kantor, D.B.; Hitka, G.; Fekete, A.; Balla, C. Electronic tongue for sensing taste changes with apricots during storage. Sens. Actuators B Chem. 2008, 131, 43–47. [Google Scholar] [CrossRef]
  42. Campos, I.; Bataller, R.; Armero, R.; Gandia, J.M.; Soto, J.; Martínez-Máñez, R.; Gil-Sánchez, L. Monitoring grape ripeness using a voltammetric electronic tongue. Food Res. Int. 2013, 54, 1369–1375. [Google Scholar] [CrossRef]
  43. Lozano, J.; Arroyo, T.; Santos, J.; Cabellos, J.; Horrillo, M. Electronic nose for wine ageing detection. Sens. Actuators B Chem. 2008, 133, 180–186. [Google Scholar] [CrossRef]
  44. Khodabakhshian, R.; Emadi, B.; Khojastehpour, M.; Golzarian, M.R. A comparative study of reflectance and transmittance modes of Vis/NIR spectroscopy used in determining internal quality attributes in pomegranate fruits. J. Food Meas. Charact. 2019, 13, 3130–3139. [Google Scholar] [CrossRef]
  45. Mishra, P.; Roger, J.M.; Rutledge, D.N.; Woltering, E. SPORT pre-processing can improve near-infrared quality prediction models for fresh fruits and agro-materials. Postharvest Biol. Technol. 2020, 168, 111271. [Google Scholar] [CrossRef]
  46. Sripaurya, T.; Sengchuai, K.; Booranawong, A.; Chetpattananondh, K. Gros Michel banana soluble solids content evaluation and maturity classification using a developed portable 6 channel NIR device measurement. Measurement 2021, 173, 108615. [Google Scholar] [CrossRef]
  47. Pissard, A.; Fernández Pierna, J.A.; Baeten, V.; Sinnaeve, G.; Lognay, G.; Mouteau, A.; Dupont, P.; Rondia, A.; Lateur, M. Non-destructive measurement of vitamin C, total polyphenol and sugar content in apples using near-infrared spectroscopy. J. Sci. Food Agric. 2013, 93, 238–244. [Google Scholar] [CrossRef] [PubMed]
  48. Nawi, N.M.; Jensen, T.; Chen, G. The application of spectroscopic methods to predict sugarcane quality based on stalk cross-sectional scanning. Am. Soc. Sugar Cane Technol. J. 2012, 32, 16–27. [Google Scholar]
  49. Liu, Y.; Sun, X.; Zhang, H.; Aiguo, O. Nondestructive measurement of internal quality of Nanfeng mandarin fruit by charge coupled device near infrared spectroscopy. Comput. Electron. Agric. 2010, 71, S10–S14. [Google Scholar] [CrossRef]
  50. Chia, K.S.; Jam, M.N.H.; Gan, Z.; Ismail, N. Pre-dispersive near-infrared light sensing in non-destructively classifying the brix of intact pineapples. J. Food Sci. Technol. 2020, 57, 4533–4540. [Google Scholar] [CrossRef] [PubMed]
  51. Jha, S.N.; Matsuoka, T. Non-destructive determination of acid–brix ratio of tomato juice using near infrared spectroscopy. Int. J. Food Sci. Technol. 2004, 39, 425–430. [Google Scholar] [CrossRef]
  52. Larraín, M.; Guesalaga, A.R.; Agosín, E. A multipurpose portable instrument for determining ripeness in wine grapes using NIR spectroscopy. IEEE Trans. Instrum. Meas. 2008, 57, 294–302. [Google Scholar] [CrossRef]
  53. Liu, Y.; Gao, R.; Hao, Y.; Sun, X.; Ouyang, A. Improvement of near-infrared spectral calibration models for brix prediction in ‘Gannan’navel oranges by a portable near-infrared device. Food Bioprocess Technol. 2012, 5, 1106–1112. [Google Scholar] [CrossRef]
  54. Nawi, N.M.; Chen, G.; Jensen, T. Visible and shortwave near infrared spectroscopy for predicting sugar content of sugarcane based on a cross-sectional scanning method. J. Near Infrared Spectrosc. 2013, 21, 289–297. [Google Scholar] [CrossRef] [Green Version]
  55. Saranwong, I.; Sornsrivichai, J.; Kawano, S. Performance of a portable near infrared instrument for Brix value determination of intact mango fruit. J. Near Infrared Spectrosc. 2003, 11, 175–181. [Google Scholar] [CrossRef]
  56. Reita, G.; Peano, C.; Saranwong, S.; Kawano, S. An evaluating technique for variety compatibility of fruit applied to a near infrared Brix calibration system: A case study using Brix calibration for nectarines. J. Near Infrared Spectrosc. 2008, 16, 83–89. [Google Scholar] [CrossRef]
  57. Corrêdo, L.d.P.; Maldaner, L.F.; Bazame, H.C.; Molin, J.P. Evaluation of Minimum Preparation Sampling Strategies for Sugarcane Quality Prediction by vis-NIR Spectroscopy. Sensors 2021, 21, 2195. [Google Scholar] [CrossRef] [PubMed]
  58. Zhang, D.; Xu, Y.; Huang, W.; Tian, X.; Xia, Y.; Xu, L.; Fan, S. Nondestructive measurement of soluble solids content in apple using near infrared hyperspectral imaging coupled with wavelength selection algorithm. Infrared Phys. Technol. 2019, 98, 297–304. [Google Scholar] [CrossRef]
  59. Tran, N.T.; Fukuzawa, M. A portable spectrometric system for quantitative prediction of the soluble solids content of apples with a pre-calibrated multispectral sensor chipset. Sensors 2020, 20, 5883. [Google Scholar] [CrossRef] [PubMed]
  60. Schaare, P.; Fraser, D. Comparison of reflectance, interactance and transmission modes of visible-near infrared spectroscopy for measuring internal properties of kiwifruit (Actinidia chinensis). Postharvest Biol. Technol. 2000, 20, 175–184. [Google Scholar] [CrossRef]
  61. Angra, S.; Dimri, A.; Kapur, P. Nondestructive brix evaluation of apples of different origin using near infrared (nir) filter based reflectance spectroscopy. Instrum. Sci. Technol. 2009, 37, 241–253. [Google Scholar] [CrossRef]
  62. Golic, M.; Walsh, K.; Lawson, P. Short-wavelength near-infrared spectra of sucrose, glucose, and fructose with respect to sugar concentration and temperature. Appl. Spectrosc. 2003, 57, 139–145. [Google Scholar] [CrossRef] [PubMed]
  63. Mehrotra, R.; Siesler, H.W. Application of mid infrared/near infrared spectroscopy in sugar industry. Appl. Spectrosc. Rev. 2003, 38, 307–354. [Google Scholar] [CrossRef]
  64. Hancock, J.; Callow, P.; Serçe, S.; Hanson, E.; Beaudry, R. Effect of cultivar, controlled atmosphere storage, and fruit ripeness on the long-term storage of highbush blueberries. HortTechnology 2008, 18, 199–205. [Google Scholar] [CrossRef]
  65. Colnago, L.A.; Andrade, F.D.; Souza, A.A.; Azeredo, R.B.; Lima, A.A.; Cerioni, L.M.; Osán, T.M.; Pusiol, D.J. Why is inline NMR rarely used as industrial sensor? Challenges and opportunities. Chem. Eng. Technol. 2014, 37, 191–203. [Google Scholar] [CrossRef] [Green Version]
  66. Srivastava, R.; Talluri, S.; Beebi, S.K.; Kumar, B.R. Magnetic resonance imaging for quality evaluation of fruits: A review. Food Anal. Methods 2018, 11, 2943–2960. [Google Scholar] [CrossRef]
  67. Zhang, L.; McCarthy, M.J. Assessment of pomegranate postharvest quality using nuclear magnetic resonance. Postharvest Biol. Technol. 2013, 77, 59–66. [Google Scholar] [CrossRef]
  68. Zhang, L.; McCarthy, M.J. Measurement and evaluation of tomato maturity using magnetic resonance imaging. Postharvest Biol. Technol. 2012, 67, 37–43. [Google Scholar] [CrossRef]
  69. Musse, M.; Bidault, K.; Quellec, S.; Brunel, B.; Collewet, G.; Cambert, M.; Bertin, N. Spatial and temporal evolution of quantitative magnetic resonance imaging parameters of peach and apple fruit–relationship with biophysical and metabolic traits. Plant J. 2021, 105, 62–78. [Google Scholar] [CrossRef] [PubMed]
  70. Geya, Y.; Kimura, T.; Fujisaki, H.; Terada, Y.; Kose, K.; Haishi, T.; Gemma, H.; Sekozawa, Y. Longitudinal NMR parameter measurements of Japanese pear fruit during the growing process using a mobile magnetic resonance imaging system. J. Magn. Reson. 2013, 226, 45–51. [Google Scholar] [CrossRef] [PubMed]
  71. Andaur, J.E.; Guesalaga, A.R.; Agosin, E.E.; Guarini, M.W.; Irarrázaval, P. Magnetic resonance imaging for nondestructive analysis of wine grapes. J. Agric. Food Chem. 2004, 52, 165–170. [Google Scholar] [CrossRef] [PubMed]
  72. Jaywant, S.A.; Arif, K.M. A comprehensive review of microfluidic water quality monitoring sensors. Sensors 2019, 19, 4781. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  73. Sun, N.; Yoon, T.J.; Lee, H.; Andress, W.; Weissleder, R.; Ham, D. Palm NMR and 1-Chip NMR. IEEE J. Solid-State Circuits 2011, 46, 342–352. [Google Scholar] [CrossRef]
  74. Windt, C.W.; Blümler, P. A portable NMR sensor to measure dynamic changes in the amount of water in living stems or fruit and its potential to measure sap flow. Tree Physiol. 2015, 35, 366–375. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  75. Cozzolino, D.; Cynkar, W.; Shah, N.; Smith, P. Technical solutions for analysis of grape juice, must, and wine: The role of infrared spectroscopy and chemometrics. Anal. Bioanal. Chem. 2011, 401, 1475–1484. [Google Scholar] [CrossRef] [PubMed]
  76. de Oliveira, V.A.; Barbero, A.P.L.; Sphaier, L.A.; dos Santos, A.B.; Peixoto, F.C.; Silva, V.N. Real-time fermentation monitoring of synthetic beer wort using etched fiber Bragg grating. IEEE Trans. Instrum. Meas. 2021, 70, 1–7. [Google Scholar] [CrossRef]
  77. Cozzolino, D.; Parker, M.; Dambergs, R.G.; Herderich, M.; Gishen, M. Chemometrics and visible-near infrared spectroscopic monitoring of red wine fermentation in a pilot scale. Biotechnol. Bioeng. 2006, 95, 1101–1107. [Google Scholar] [CrossRef] [PubMed]
  78. Cozzolino, D.; Dambergs, R.; Janik, L.; Cynkar, W.; Gishen, M. Analysis of grapes and wine by near infrared spectroscopy. J. Near Infrared Spectrosc. 2006, 14, 279–289. [Google Scholar] [CrossRef]
  79. Grassi, S.; Amigo, J.M.; Lyndgaard, C.B.; Foschino, R.; Casiraghi, E. Beer fermentation: Monitoring of process parameters by FT-NIR and multivariate data analysis. Food Chem. 2014, 155, 279–286. [Google Scholar] [CrossRef] [PubMed]
  80. Urbano-Cuadrado, M.; De Castro, M.L.; Pérez-Juan, P.; García-Olmo, J.; Gómez-Nieto, M. Near infrared reflectance spectroscopy and multivariate analysis in enology: Determination or screening of fifteen parameters in different types of wines. Anal. Chim. Acta 2004, 527, 81–88. [Google Scholar] [CrossRef]
  81. Manley, M.; Van Zyl, A.; Wolf, E. The evaluation of the applicability of Fourier transform near-infrared (FT-NIR) spectroscopy in the measurement of analytical parameters in must and wine. S. Afr. J. Enol. Vitic. 2001, 22, 328–339. [Google Scholar] [CrossRef] [Green Version]
  82. Martelo-Vidal, M.J.; Vazquez, M. Evaluation of ultraviolet, visible, and near infrared spectroscopy for the analysis of wine compounds. Czech J. Food Sci. 2014, 32, 37–47. [Google Scholar] [CrossRef] [Green Version]
  83. Liu, F.; He, Y.; Wang, L.; Pan, H. Feasibility of the use of visible and near infrared spectroscopy to assess soluble solids content and pH of rice wines. J. Food Eng. 2007, 83, 430–435. [Google Scholar] [CrossRef]
  84. Páscoa, R.N.; Porto, P.A.; Cerdeira, A.L.; Lopes, J.A. The application of near infrared spectroscopy to wine analysis: An innovative approach using lyophilization to remove water bands interference. Talanta 2020, 214, 120852. [Google Scholar] [CrossRef] [PubMed]
  85. Lamberti, N.; Ardia, L.; Albanese, D.; Di Matteo, M. An ultrasound technique for monitoring the alcoholic wine fermentation. Ultrasonics 2009, 49, 94–97. [Google Scholar] [CrossRef]
  86. Becker, T.; Mitzscherling, M.; Delgado, A. Ultrasonic velocity–a noninvasive method for the determination of density during beer fermentation. Eng. Life Sci. 2001, 1, 61–67. [Google Scholar] [CrossRef]
  87. Resa, P.; Elvira, L.; de Espinosa, F.M. Concentration control in alcoholic fermentation processes from ultrasonic velocity measurements. Food Res. Int. 2004, 37, 587–594. [Google Scholar] [CrossRef]
  88. Dodevska, T.; Lazarova, Y.; Shterev, I. Amperometric Biosensors for Glucose and Lactate with Applications in Food Analysis: A Brief Review. Acta Chim. Slov. 2019, 66, 762–776. [Google Scholar] [CrossRef] [PubMed]
  89. Niculescu, M.; Mieliauskiene, R.; Laurinavicius, V.; Csöregi, E. Simultaneous detection of ethanol, glucose and glycerol in wines using pyrroloquinoline quinone-dependent dehydrogenases based biosensors. Food Chem. 2003, 82, 481–489. [Google Scholar] [CrossRef]
  90. Esti, M.; Volpe, G.; Compagnone, D.; Mariotti, G.; Moscone, D.; Palleschi, G. Monitoring alcoholic fermentation of red wine by electrochemical biosensors. Am. J. Enol. Vitic. 2003, 54, 39–45. [Google Scholar]
  91. Goriushkina, T.B.; Soldatkin, A.P.; Dzyadevych, S.V. Application of amperometric biosensors for analysis of ethanol, glucose, and lactate in wine. J. Agric. Food Chem. 2009, 57, 6528–6535. [Google Scholar] [CrossRef] [PubMed]
  92. Piermarini, S.; Volpe, G.; Esti, M.; Simonetti, M.; Palleschi, G. Real time monitoring of alcoholic fermentation with low-cost amperometric biosensors. Food Chem. 2011, 127, 749–754. [Google Scholar] [CrossRef] [Green Version]
  93. Patel, N.; Meier, S.; Cammann, K.; Chemnitius, G.C. Screen-printed biosensors using different alcohol oxidases. Sens. Actuators B Chem. 2001, 75, 101–110. [Google Scholar] [CrossRef]
  94. Ghica, M.; Brett, C. Development of a carbon film electrode ferrocene-mediated glucose biosensor. Anal. Lett. 2005, 38, 907–920. [Google Scholar] [CrossRef]
  95. Ghica, M.E.; Brett, C.M. Development of novel glucose and pyruvate biosensors at poly (neutral red) modified carbon film electrodes. Application to natural samples. Electroanal. Int. J. Devoted Fundam. Pract. Asp. Electroanal. 2006, 18, 748–756. [Google Scholar] [CrossRef] [Green Version]
  96. Final Report to Australian Grape and Wine Authority. Available online: https://www.wineaustralia.com/getmedia/d0e6cb2f-102e-4783-88f1-d291acbb169a/AWR-1401-Final-Report1 (accessed on 22 October 2019).
  97. Endress & Hauser. Available online: http://fotonika.kiev.ua/pdf/FMB70.pdf (accessed on 23 June 2021).
  98. Smar Technology Company. Available online: https://www.smar.com/en/product/dt301-4-to-20-ma-hart-density-transmitter (accessed on 6 March 2022).
  99. SensoTech. Available online: https://www.sensotech.com/en/products/systems/liquisonic-plato-brix (accessed on 6 March 2022).
  100. SCHMIDT HAENSCH. Available online: https://schmidt-haensch.com/wp-content/uploads/2015/01/SHiPR-B3ENG180628.pdf (accessed on 6 March 2022).
  101. Tilt Hydrometer. Available online: https://tilthydrometer.com/products/tilt-pro-wireless-hydrometer-and-thermometer (accessed on 6 March 2022).
  102. VitalSensors Brewing Analyzer. Available online: https://www.hach.com/beer-analyzer/vitalsensors-brewing-analyzer/family?productCategoryId=35547408052 (accessed on 6 March 2022).
  103. Mobile Density Meter. Available online: https://metersolution.com/analytical-laboratory/ (accessed on 6 March 2022).
  104. Inline Brix Refractometer. Available online: https://schmidt-haensch.com/product/ipr-basic3-brix-refractometer/ (accessed on 6 March 2022).
Figure 1. Different stages of fruit maturity.
Figure 1. Different stages of fruit maturity.
Sensors 22 02290 g001
Figure 2. Brix measurement with hydrometer in a graduated cylindrical container.
Figure 2. Brix measurement with hydrometer in a graduated cylindrical container.
Sensors 22 02290 g002
Figure 3. Features of analog, digital, and benchtop refractometers.
Figure 3. Features of analog, digital, and benchtop refractometers.
Sensors 22 02290 g003
Figure 4. Constructional block diagram of an electronic tongue.
Figure 4. Constructional block diagram of an electronic tongue.
Sensors 22 02290 g004
Figure 5. Brix monitoring at different stages of wine and beer production.
Figure 5. Brix monitoring at different stages of wine and beer production.
Sensors 22 02290 g005
Figure 6. General scheme of amperomertic biosensors for detecting sugar and alcohol levels in process industry.
Figure 6. General scheme of amperomertic biosensors for detecting sugar and alcohol levels in process industry.
Sensors 22 02290 g006
Figure 7. Brix measurement methods for fruits and vegetables.
Figure 7. Brix measurement methods for fruits and vegetables.
Sensors 22 02290 g007
Figure 8. Brix measurement methods for wine and beer.
Figure 8. Brix measurement methods for wine and beer.
Sensors 22 02290 g008
Table 1. Summary of refractometers used in various research of horticulture and agriculture products.
Table 1. Summary of refractometers used in various research of horticulture and agriculture products.
ProductApplicationInstrumentModelRef.
MelonSSC (%Brix)Hand RefractometerNot specified,[16]
Atago, Tokyo, Japan
BlackberrySSC (%Brix)Hand RefractometerAbbe refractometer,[17]
Atago, Tokyo, Japan
OliveSugar content ( Brix)Hand RefractometerPR-32 α ,[18]
Atago, Bellevue, WA, USA
ApricotSSC (%Brix)Digital RefractometerPR-101,[19]
Atago, Nor-folk, VA, USA
ApricotSSC (%Brix)Digital RefractometerPR-1,[20]
Atago, Tokyo, Japan
PeachSoluble solids ( Brix)Digital RefractometerPR-101,[21]
Atago, Tokyo, Japan
SSC: Soluble solid content.
Table 2. An overview of commercial refractometers available in the market.
Table 2. An overview of commercial refractometers available in the market.
InstrumentModelMeasuring RangeAccuracyCost (Approx)Ref.
Handheld devices
Analog refractometerHENAN refractometersNot specifiedNot specified [22]
Analog refractometerBRM-100 STARR INSTRUMENTS0 to 32% Brix±0.2 Brix%AUD 99[23]
Digital refractometerBrix Meter-BX-1,0.0 to 85.0% Brix±0.2 Brix% [24]
Bell Technology Ltd.
Digital refractometerATAGO PAL-38800 to 33% BrixNot specifiedAUD 395[25]
Digital refractometer30GS, Mettler toledo0 to 100% Brix±0.2 Brix%Not specified[26]
Benchtop devices
Digital refractometerExcellence R4, Mettler Toledo0 to 100% Brix±0.05 Brix%USD 5140[27]
Digital refractometerAbbemat 450, Anton-Paar0 to 100% Brix±0.05 Brix%Not specified[28]
Digital refractometerLaxco™ RBD60000 to 100% Brix±0.05 Brix%USD 9300[29]
Digital refractometerLaxco™ RHDB1050 to 59.9% Brix±0.2 Brix%USD 1526[30]
Digital refractometerAnton Paar 1734520 to 100% Brix±0.05 Brix%USD 6960[31]
Table 3. Sample-based destructive Brix sensors reported in the literature.
Table 3. Sample-based destructive Brix sensors reported in the literature.
SensorRangeAccuracyResolutionRef.
RI-based optical meter0–85 ( Brix)Not specifiedNot specified[13]
RI-based SPR sensor0–25 ( Brix)Not specifiedNot specified,[15]
LMR-based sensor0–50 ( Brix)Not specified 0.005 Brix[32]
Electronic tongueNot specified90%Not specified[39]
Table 4. Summary of NIR methods reported in the literature.
Table 4. Summary of NIR methods reported in the literature.
ProductInstrumentModeSpectral RangeDetectorNo. of ScanAccuracyRef.
Non-portable methods
ApplesIn-house developedReflectance800 to 1600 nmNot specified5 to 6SEP = 0.7322 to 1.7809[61]
BananaIn-house developedReflectance600 to 870 nmPhotodiode bankNot specified97%[46]
Pomegranatedual-channel spectrometer,Reflectance400 to 1100 nmNot specifiedNot specifiedR = 0.94[44]
AvaSpec-2048TEC SEP = 0.24 Brix
ApplesScanning MonochromatorReflectance400 to 2500 nmSiliconNot specifiedR = 0.82[47]
Model 6500, NRISystems SEP = 0.61 Brix
SugarcaneFR, SpectroradiometerReflectance350 to 2500 nmNot specified25R2 = 0.76[48]
MandarinCCD spectrometerTransmittance350 to 1040 nmPhotodiode arrayNot specifiedR = 0.92[49]
USB4000 SEP = 0.65 Brix
Portable Methods
PineappleIn house developedReflectance780, 850, 870, 910, and 940 nmPhotodiodeNot specified80.56%[50]
TomatoIn house developedReflectance703 to 1124 nmPhotodiode arrayNot specifiedR = 0.92[51]
GrapesIn house developedReflectance640 to 1300 nmNot specifiedNot specifiedR2 = 0.874 to 0.930[52]
OrangesIn house developedReflectance200 to 1100 nmCCDNot specifiedR2 = 0.918[53]
SEP = 0.321 Brix
Sugarcanevis/SW-NIRS,Reflectance325 to 1075 nmNot specified20R = 0.87[54]
FieldSpec HandHeld SEP = 1.45 Brix
MangoFruit Tester 20Reflectance600 to 1000 nmSilicon diode array30R = 0.98[55]
SEP = 0.40 Brix
NectarinesNIR-Gun, FANTECReflectance600 to 1100 nmSilicon diode arrayNot specifiedR = 0.90[56]
SEP = 0.50 Brix
SEP: Square errors of prediction.
Table 5. Summary of MRI methods reported in various research.
Table 5. Summary of MRI methods reported in various research.
ProductMeasured FeaturesApplicationRef.
PomegrantateT2Measurement of SSC[67]
TomatoT1 & T2Study of tomato ripening stages[68]
Peach & AppleT2Measurement of total sugar content[69]
PearT1 & T2Study of fruit growing stages[70]
Measurement of sugar content
GrapesT1 & T2Study of ripening stages of grape berries[71]
SSC: Soluble solid content.
Table 6. Specifications of nondestructive Brix sensors reported in the literature.
Table 6. Specifications of nondestructive Brix sensors reported in the literature.
SensorRangeAccuracyResolutionRef
NIR-based nonportable sensorNot specified97%Not specified[46]
NIR-based portable sensorNot specified80.56%Not specified[50]
MRI-based sensorNot specifiedNot specifiedNot specified[67,68,69]
Table 7. Summary of sugar content in different wines using vis-NIR spectroscopy as reported in various research articles.
Table 7. Summary of sugar content in different wines using vis-NIR spectroscopy as reported in various research articles.
ParametersType of WineWavelength Range nmPredictionsRef.
Reducing sugars (g L−1)Red, rose and white wines400–2500R2 = 0.705[80]
Sugar ( Brix)White wine1000–2500r = 0.99[81]
Glucose (g L−1)Red wine400–2500R2 = 0.987[82]
soluble solids content ( Brix)Rice wine325–1075r = 0.95[83]
Total sugars (g/L)Red, rose and white wines1000–2500R2 = 0.94[84]
Table 8. Summary of commercially available in-line Brix sensors.
Table 8. Summary of commercially available in-line Brix sensors.
SensorSupplierPrincipleMeasurementTemp SensorApproximately CostRef.
Liquiphant MEndress &DensityIn tankExtra Required$2–3K[96]
vibrating forkHauser
Deltapoint SEndress &HydrostaticIn tankNot [97]
Hauserpressure Required$3K
MicropilotEndress &VolumeTank headerNotcombined[96]
Hauser(Level) Required
FermetrolPsitecOsmotic potentialIn tankIntegrated$2K[96]
MicroLDSISSYSMass flow (liquid)By-pass loopIntegrated$1K[96]
VS-3000Vital SensorsAbsorbanceBy-pass loopIntegrated$3K[96]
BiosensorOptiEnzEnzyme responseBy-pass loopNot Required$3K[96]
DT301Smar technologyHydrostatic pressureIn tankIntegratedNot available[98]
Complete analyzerSensotechUltrasonic pressureIn tankIntegratedNot available[99]
iPR B3SCHMIDTAbsorbanceIn tankIntegratedNZD18K[100]
HAENSCH
TiltTilt HydrometerHydrostatic pressureHome brewingIntegrated$250[101]
Table 9. Specifications of in-line Brix sensors.
Table 9. Specifications of in-line Brix sensors.
SensorRangeAccuracyResolutionRef.
DT3010 to 10 g/cm3±0.0004 g/cm3Not specified[98]
VS-30000–20 °Plato±0.1 Brix0.01 Brix[102]
MicroLDS0.6–1.3 g/cc0.0005 g/cc0.0001 g/cc[103]
iPR B30–90 Brix±0.1 Brix0.01 Brix[104]
Tilt0.990 to 1.120 of SG±0.002 of SGNot specified[101]
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Jaywant, S.A.; Singh, H.; Arif, K.M. Sensors and Instruments for Brix Measurement: A Review. Sensors 2022, 22, 2290. https://doi.org/10.3390/s22062290

AMA Style

Jaywant SA, Singh H, Arif KM. Sensors and Instruments for Brix Measurement: A Review. Sensors. 2022; 22(6):2290. https://doi.org/10.3390/s22062290

Chicago/Turabian Style

Jaywant, Swapna A., Harshpreet Singh, and Khalid Mahmood Arif. 2022. "Sensors and Instruments for Brix Measurement: A Review" Sensors 22, no. 6: 2290. https://doi.org/10.3390/s22062290

APA Style

Jaywant, S. A., Singh, H., & Arif, K. M. (2022). Sensors and Instruments for Brix Measurement: A Review. Sensors, 22(6), 2290. https://doi.org/10.3390/s22062290

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop