The Design of a Low Noise, Multi-Channel Recording System for Use in Implanted Peripheral Nerve Interfaces

1
Department of Electronic and Electrical Engineering, University of Bath, Bath BA2 7AY, UK
2
Department of Medical Physics and Bioengineering, University College London, London WC1E 6BT, UK
3
Department of Clinical Science and Services, The Royal Veterinary College, Hawkshead Lane, Brookmans Park, Hatfield AL9 7TA, UK
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(9), 3450; https://doi.org/10.3390/s22093450
Submission received: 1 April 2022 / Revised: 18 April 2022 / Accepted: 23 April 2022 / Published: 30 April 2022
(This article belongs to the Section Biomedical Sensors)

Abstract

:
In the development of implantable neural interfaces, the recording of signals from the peripheral nerves is a major challenge. Since the interference from outside the body, other biopotentials, and even random noise can be orders of magnitude larger than the neural signals, a filter network to attenuate the noise and interference is necessary. However, these networks may drastically affect the system performance, especially in recording systems with multiple electrode cuffs (MECs), where a higher number of electrodes leads to complicated circuits. This paper introduces formal analyses of the performance of two commonly used filter networks. To achieve a manageable set of design equations, the state equations of the complete system are simplified. The derived equations help the designer in the task of creating an interface network for specific applications. The noise, crosstalk and common-mode rejection ratio (CMRR) of the recording system are computed as a function of electrode impedance, filter component values and amplifier specifications. The effect of electrode mismatches as an inherent part of any multi-electrode system is also discussed, using measured data taken from a MEC implanted in a sheep. The accuracy of these analyses is then verified by simulations of the complete system. The results indicate good agreement between analytic equations and simulations. This work highlights the critical importance of understanding the effect of interface circuits on the performance of neural recording systems.

1. Introduction

1.1. Neural Recording

The recording of signals from the peripheral nervous system (PNS) (the electroneurogram—ENG) using chronically implanted electrodes is one of the major challenges in current neuroprosthetic research. Several types of implantable interfaces have been proposed, but very few have been validated with long-term chronic studies. One of the most well-established types, both for stimulation and recording, is the extraneural nerve cuff [1]. Cuffs are widely used for the electrical stimulation of the peripheral nerves, including in commercial devices, such as the LivaNova vagus nerve stimulator for intractable epilepsy [2,3]. Likewise, there have been several demonstrations of recording from the PNS using cuffs, albeit predominantly in acute studies [4]. Tripolar stimulation cuffs, in which a current is driven between the centre electrode and the outer pair of electrodes (which are usually connected together) are also frequently employed [5]. For recording purposes, a monopole, dipole or tripole electrode structure is connected to a differential amplifier or a double-differential amplifier. The dipole recording configuration is most common, although it has been shown that the tripole arrangement can reduce interference by improving common-mode (CM) rejection [6].
The neural signals within the PNS may be classified as afferent or efferent, corresponding to the direction of propagation. Most peripheral nerves (especially when interfaced using an extraneural approach) are mixed and contain many afferent and efferent axons. In order to record selectively from specific axons, it is necessary to perform some form of signal processing to separate the propagating signals. There are many approaches to this problem, such as the use of spatio-temporal filters [7], source localisation [8], electrical impedance tomography [9], or discrimination based on conduction velocity [10]. Velocity Selective Recording (VSR) uses multi-electrode cuffs (MECs) to detect and classify neural signals based on the different velocities present, exploiting the correspondence between conduction velocity and fibre diameter, at least for myelinated nerves [11,12].

1.2. Recording Challenges of MECs

To implement VSR, the electrodes of the MEC must be connected to a bank of differential amplifiers [11] and this increases the complexity of the amplifier design task. As for the dipoles and tripoles, the amplitudes of spontaneous (i.e., naturally occurring) neural signals recorded are very small (generally less than 1 µV [13]). Therefore, the differential voltage gain must be high (typically 60–100 dB) with an adequate signal-to-noise ratio (SNR), requiring very low noise front-end amplifiers, i.e., the noise floor of the amplifier must be less than a few nV/√Hz [14]. The outline schematic of an example recording system is shown in Figure 1, where N (typically about 10) electrodes are fitted to an insulating nerve cuff forming a MEC. The electrodes are shown connected directly to an amplifier array as dipoles.
For the MEC, in addition to the exacting differential specification required of each amplifier, in order to mitigate the effects of undesired interfering signal sources, CM gain and crosstalk between the individual channels must be minimised. In the recording of spontaneous neural signals, interference comes from nearby muscles, AC mains and radio frequency (RF) pick-up [15]. Singly or in combination, such interference can saturate high-gain amplifiers. Therefore, a bandpass filter interface network, placed between the recording electrodes and the front-end amplifiers, is essential to limit the effects of high- and low-frequency interfering signals. Unfortunately, the presence of such a filter network conflicts with the noise and CM rejection behaviour of the amplifiers and so the interface between the two requires careful design. Moreover, the impedance mismatch between channels in a multi-electrode recording system is inevitable and will also affect the system’s performance. In this paper, two interface circuits suitable for a multi-electrode recording system are examined for cuff electrodes. Analyses of noise, common-mode rejection ratio (CMRR) and crosstalk are presented for two input networks (‘Type 1’ and ‘Type 2’) (these correspond to the networks called ‘Type 1’ and ‘Type 2’ in [15]). The two networks were originally introduced and analysed for the case of a single tripole (one differential amplifier) in [16], and it was found that the ‘Type 2’ arrangement was superior to the ‘Type 1’ network in terms of both CMRR and noise performance. In this paper, we extend and enlarge the results of [16] and demonstrate that the single-amplifier results do not extend to a multi-electrode system. The two networks are shown in Figure 2.
The overall intention of this work is to aid and inform the designer of such a system and provide the theory that simulation alone cannot. Having said that, the validation study presented is based mostly on simulation. The justification for this approach is discussed in Section 4 and relates to the inherently low sensitivity of the circuits employed to component tolerances (including the electrode impedances) and the reliability of simulation models in the bandwidth under consideration. Equations are derived symbolically using MATLAB and simplified for practical component values, reflecting how the various parameters affect the design criteria. Although the equations are derived for a 10-channel recording system, they can be extended and generalised for an N electrode system. SPICE simulations and some measured data are used to validate the accuracy of the procedure and indicate good agreement between the detailed analytic equations and the simplified versions.
The outline of this paper is as follows. CMRR, crosstalk and noise performance of a 10-channel multi-electrode system with two types of filters are discussed, and corresponding approximate equations are derived in Section 2. Section 4 investigates the accuracy of the approximate equations derived in Section 2 using SPICE simulation. Section 5 discusses how these analyses can help designers to achieve the desired specifications in a multi-electrode recording system and, finally, conclusions are drawn in Section 6.

2. Interface Circuits

Figure 2a,b show the two candidate front-end circuits that are referred to as ‘Type 1’ and ‘Type 2’ [16]. Both circuits place an RC bandpass filter between the cuff electrodes and the input terminals of the amplifiers, realised by first-order high pass and low pass sections connected in cascade, the time constants being realised by the series and parallel capacitors (Cs, Cp), respectively, and associated resistances, including the two impedances Rd and Re, (see below). The cut-off frequencies are chosen to be about 100 Hz and 100 kHz [14], respectively, attenuating both low- and high-frequency interfering signals while satisfying the bandwidth requirements of a 10-channel VSR system for neural recording [7]. Note, also, that the sensitivities of the circuit transfer functions to the various component values (including those associated with the electrodes) will be low.
In all cases, the electrodes, which are grouped as dipoles, are represented by an equivalent circuit consisting of a voltage source (Vd) in series with an impedance (Rd) representing the axial component of the section of the tissue inside the cuff. This combination of voltage generator and impedance appears to be a generic Thévenin source, but actually, it is more complicated than that. The source will be the superposition of voltages from all the separate axons, and each axon produces action potentials at the dipole electrodes which is the product of Rd and its action current—the action current is being determined by the travelling transmembrane (TMAP) potential and the high axon resistance [11]. The action currents are therefore a property of the nerve and only Rd can be altered by the design of the cuff. It follows that Vd and Rd are not independent but, in fact, Vd is proportional to Rd. Rd is resistive and has a typical value of about 0.5 kΩ [9]. By contrast, the impedances of the electrodes connecting the tissue to the front-end amplifier circuits (Re) are complex, with typical values (moduli) of about 1 kΩ at a frequency of 1 kHz. These element values and assumptions have been employed in assessing the performances of the two candidate circuits discussed in the paper and to simplify the design equations.
The only difference between the two circuits in Figure 2 is that in the Type 1 circuit of Figure 2a the bias current path to ground for each amplifier input is provided by individual resistors (Ra) while in the Type 2 configuration of Figure 2b, a centre tapped pair of resistors (R1) is employed [8]. The two resistors’ Rcm represents the impedances between the reference electrode (normally placed far away from the cuff) and the end electrodes of the cuff. Vcm is the CM input voltage (i.e., the interfering source in this case). At passband (mid-band) frequencies, it is assumed that the shunt capacitors become open circuits and the series capacitors short circuit. In the following sub-sections, expressions for CMRR, crosstalk and noise are derived for both versions of the circuit, for passband operation in all cases.
Finally, it is worth noting that all the circuits employed are based on simple first-order RC circuits and, since no overall feedback is employed, the poles of the transfer functions lie on the negative real axis of the complex plane. This has the advantage that Q values are low in the frequency domain and hence sensitivity to component tolerances is also low.

2.1. Common-Mode Rejection Ratio (CMRR)

Using an amplifier with high CMRR does not guarantee that the entire system will operate with high CMRR. A front-end filter network of the type shown in Figure 2 (either version) can generate differential signals from common-mode inputs, reducing the overall CMRR of the system. It can be shown that the CMRR of the entire system is related to the CMRR of the individual amplifiers by the following expression [8]:
1 CMRR 1 CMRR a + V a i V c m
where CMRR a is the amplifier’s common-mode rejection ratio and Vai/Vcm represents the differential gain from the common-mode input (see Figure 2—both forms) to the i′th differential input of the amplifier array. Therefore, an inappropriately designed front-end filter can degrade the performance of the entire system, irrespective of the quality of the amplifiers. To investigate the effect of the front-end network on the overall CMRR, the common-mode gain at the differential input of each amplifier Vai was calculated for both candidate structures, for N = 10.

2.1.1. ‘Type 1’ Biasing Structure

To calculate the effect of a common-mode signal at the inputs of the amplifiers, all differential input voltage sources Vdi are set to zero, and the KCL equations for nodes V1, …, V10, V′1, …, V′10 are written in the form:
A × [ V 1 : V 10 V 1 : V 10 ] = B × V c m
where A is the (20 × 20) connection matrix shown in Table 1 and B is a (20 × 1) column vector. In B, all elements are zero except for the 11th and 20th, which are Gcm1 and Gcm2, respectively.
B = [ 0 0 : 0 G cm 1 0 0 : G cm 2 ]
In the analysis, the conductances Gcm, Gd, etc., are the inverses of the resistances Rcm, Rd in Figure 2. Assuming that the amplifiers are ideal, V1, …, V10, V′1, …, V′10 can be calculated as:
[ V 1 : V 10 V 1 : V 10 ] = A 1 × B × V c m
The nodal equations for V1–V10 were solved symbolically using MATLAB (see Appendix A for an example equation). The input impedance (Rai) should be larger than the electrode impedances in order to maximise the voltage drop across the input. Therefore, making the reasonable assumptions that:
  • Rai >> Rei;
  • Rai >> Rcmi and
  • Rai >> Rdi; all i
and all the Ra are equal, all the Re are equal, and the Rcm are equal, the additional CM gain in the i’th channel due to the filter network can be expressed as follows:
V a i V c m ( N 2 i ) R d R a   ; i = 1 , 2 ,   , N / 2
where Vai is the differential input voltage to the i’th amplifier and N, the number of electrodes is assumed to be even (similar arguments apply for N odd) and with a value of 10 in this case. Equation (5) shows that the outer channels have higher common-mode gain and therefore lower CMRR than those near the centre of the array, where the overall CMRR approaches that of the amplifiers. Note that the additional CM gain can be reduced by reducing the ratio Rd/Ra as far as possible.

2.1.2. ‘Type 2’ Biasing Structure

Figure 2b shows the front-end network employing the ‘Type 2’ biasing circuit. As in the case of the ‘Type 1’ structure, KCL equations can be derived and the voltages V1V10 calculated. Proceeding as in the ‘Type 1’ case, assuming this time that for all i, R2i >> Rcmi, Rdi and Rei and that, in addition, Rei << R1i, the CM gain at the differential input of the amplifier is approximated as:
V a i V c m ( N 2 i ) ( R d | | 2 R 1 ) R 2 ;   i = 1 ,   2 , , N / 2
Equation (6) shows that increasing R2 and reducing R1 improves the CMRR. However, note that reducing R1 also decreases the differential gain of the channels in addition to possibly reducing the accuracy of the approximation, which requires, as already noted, that Rei << R1i. Therefore, some design trade-offs may be required in this case.

2.2. Crosstalk between the Channels

The differential gain from each input source Vdi to the amplifier inputs Vaj can be discussed in two parts. Firstly, the gain from each source to its corresponding amplifier input should be as close to unity as possible and, secondly, the gain from each source to the inputs of the other amplifiers, which, for the purposes of this paper, we refer to as crosstalk, should ideally be zero. In this section, the differential gain at each amplifier input is calculated for both candidate structures.

2.2.1. ‘Type 1’ Biasing Structure

To analyse the differential gain at each amplifier input Vaj, all the differential inputs Vdi except one are set to zero (the CM input voltage is also set to zero). For this calculation, it is convenient to use the Norton form of the equivalent circuit for each source, where:
I d i = V d i R d i
The KCL equations for V1, …, V10, V′1, …, V′10 are as follows:
A × [ V 1 : V 10 V 1 : V 10 ] = C i × I d i  
where A is defined in Table 1 and the column vector Ci is:
  C i = [ c 1 : c k : c 10 ]       ,   c k = { 1 k = i 1 k = i + 1 0 e l s e  
The voltages V1V10 follow from:
[ V 1 : V 10 ] = A 1 × C i × I d i  
and the differential gain from each source to each amplifier input is given by Equation (11):
V a j V d i = { 1 R d 2 R c m + 9 R d i f   i = j R d 2 R c m + 9 R d i f   i j  
Equation (11) shows that crosstalk is reduced by reducing the ratio Rd/Rcm.

2.2.2. ‘Type 2’ Biasing Structure

The analysis is very similar to that employed for the ‘Type 1’ structure and the differential gain in this case is:
V a j V d i = { R d | | 2 R 1 R d ( 1 R d | | 2 R 1 2 R c m + 9 ( R d | | 2 R 1 ) ) i f   i = j R d | | 2 R 1 R d ( R d | | 2 R 1 2 R c m + 9 ( R d | | 2 R 1 ) ) i f   i j  
As in the case of the ‘Type 1’ structure, crosstalk in the ‘Type 2’ biasing arrangement is reduced by decreasing the ratio of Rd/Rcm. Note, in addition, that in this case, decreasing R1 reduces both crosstalk and differential gain, which may be unacceptable in some applications.

2.3. Noise Analysis

In this section, the main sources of noise and their effect on the behaviour of the circuit, including the front-end filter networks (as shown in Figure 2) are described. The subject of noise in multi-channel VSR systems was dealt with comprehensively in [12]. In this paper, we present a simplified approach based on the earlier work, however, we intend to inform the designer of the main issues involved and their possible mitigation. The contributions of thermal noise, amplifier voltage noise and amplifier current noise are considered separately, and the individual sources are assumed to be Gaussian and uncorrelated in all cases.
In both the Type 1 and Type 2 arrangements, the axial components of the electrode/tissue combination are denoted Rd, while the Re is the electrode impedances (see Figure 2). The two resistances Rcm are the common-mode elements connecting the ends of the cuff system to the reference (shown as ground in the figure). In the Type 1 circuit, the Ra are bias resistors, essentially used to define the DC bias point of the amplifier inputs, whereas, in the Type 2 circuit, a balanced-T network is used for this purpose. As already noted, one consequence of this arrangement is that a potential divider exists involving Re and R1, that influences both the signal and noise amplitudes.

2.3.1. Thermal Noise

For noise modelling, as in other sections of the paper, all the resistors are assumed to be real (i.e., ohmic) except for Re, which are complex impedances. In the passband of the filter, the capacitors are chosen, so that those connected in series are short circuits while those in shunt are open circuits. The following assumptions amongst the impedances are made:
Rai >>Rei;
Rai >> Rcmi,
R2i >> R1i and
Rai >> Rdi; all i
For noise calculations, it is helpful to redraw the filter circuit in the passband as a ladder network, where the impedances of each type are assumed to be equal. For the Type 1 circuit, it is as shown in Figure 3a.
Noting the inequalities above, the Ra appear in parallel with much smaller resistances and so can be omitted from the noise calculations. The simplified circuit is redrawn, as shown in Figure 3b.
If each Rd is removed from the circuit in turn, the Thévenin equivalent circuit of residual impedance at the port where the resistor was removed is [(N − 1)Rd + 2Rcm]. For N = 10 (i.e., large) and, given that Rcm is not negligible, this residue will be in excess of 10Rd. This resistance appears in parallel with the Rd removed and so it is reasonable to assume that the only significant contribution to thermal noise density at the input to an amplifier is given by one Rd and the associated pair of impedances Re. The thermal noise density contribution appearing at the input to this amplifier, expressed as an rms voltage, is, therefore:
V a 1 ( r m s ) = 4 k T ( R d + 2 r e ( R e )
where re(Re) represents the real part of Re.
Similar expansion and simplification procedures can be applied to the Type 2 circuit, resulting in the equivalent circuit shown in Figure 4. As already noted in Section 2.1.2 of the paper, the presence of the resistors R1 places a potential divider (involving the impedances Re as the other element) in each signal path. Due to the small size of the recorded signals, it is highly desirable to make the gain of this divider as close to unity as possible and so we can add a further inequality to the previous list:
R1i >> ∣Rei∣, all i
Figure 4. Simplified passband equivalent circuit for thermal noise analysis for a Type 2 circuit. This is a ladder expansion of Figure 2b, to which the same simplifying procedures have been applied as were used in Figure 3a. In addition, as already noted, for maximum signal gain we require that R1 >> ∣Re∣ and so the circuit reduces to that of Figure 3b.
Figure 4. Simplified passband equivalent circuit for thermal noise analysis for a Type 2 circuit. This is a ladder expansion of Figure 2b, to which the same simplifying procedures have been applied as were used in Figure 3a. In addition, as already noted, for maximum signal gain we require that R1 >> ∣Re∣ and so the circuit reduces to that of Figure 3b.
Sensors 22 03450 g004
A consequence of this is that the R1 will contribute little to the overall noise calculation since they also appear in parallel with many small resistances. The thermal noise equivalent circuits for both Type 1 and Type 2 circuits are therefore the same (Figure 3b) and the thermal noise density appearing at the input to each amplifier is given by Equation (1).

2.3.2. Amplifier Noise

Figure 5 is the noise equivalent circuit of a single-ended output, differential voltage amplifier, where vn and in are rms noise sources (densities), respectively. Assuming the input resistance of the amplifiers is so large as to be effectively infinite, the voltage noise contribution of each amplifier appears only at the corresponding output, since no current flows into the other parts of the circuit as a result of the presence of vn.
The equivalent circuit for the amplifier current noise is shown in Figure 6, where the simplified form of the circuit developed above has been used. It is clear from this circuit that each amplifier will cause noise current to flow around a primary loop consisting of one Rd and two Re and also around a secondary loop consisting of (N − 1) resistors Rd and two Rcm. Using the same arguments as for thermal noise, the current in the secondary loop will be significantly less than in the primary loop and can be ignored for practical purposes. Using the principle of superposition (i.e., taking each source individually, the others being removed from the circuit) and treating the various resistances and impedances as noiseless (since the noise contributions of these components have been considered separately), the contribution to the rms voltage Va1 due to ia1 is:
Va11 = ia1|Rd1 + Re1 + Re2|
Since Va1 appears across one of the pairs of electrodes at the end of the array, considering only currents flowing in the primary loops, the only other contribution comes from ia2:
Va12 = ia2|Re2|
Assuming that all the Rd are equal and all the Re are equal and that ia1 and ia2 are uncorrelated sources of the same statistics and of equal amplitude ia, the total rms value of Va1 is given by summing the two contributions, recalling that Re is a complex impedance:
V a n = i a | R d + 2 R e | 2 + | R e | 2
where n = 1 or N − 1, and, similarly, for the other inputs (i.e., those not placed at the ends of the array):
V a m = i a | R d + 2 R e | 2 + 2 | R e | 2
In this section, in summary, expressions have been derived for the thermal noise and amplifier voltage and current noise appearing at the input of one of the amplifiers in the N channel system shown in Figure 2. The analysis is a simplified form of that given previously in [12] and, given certain assumptions, ensures the accuracy of the thermal and amplifier current noise calculations is better than approximately 1/N, where N is typically 10 (the amplifier voltage noise calculation is exact). Since all three sources are assumed to be white and uncorrelated, an expression for the total input-referred noise density for one amplifier can be written using superposition. This is valid for both Type 1 and Type 2 arrangements:
v a ( total ) = 4 k T [ R d + 2 r e ( R e ) ] + v n 2 + i a 2 ( | R d + 2 R e | 2 + p | R e | 2 )
where p = 1 at the end amplifiers, and p = 2 for the others.
Table 2 summarizes the equations derived in this section.

3. Electrode Impedance Mismatch

In Section 2, the calculation of CMRR, crosstalk and noise performance for the case of matched electrode impedances was discussed. However, in practice, impedance variation of the individual electrodes, as well as the mismatch between the electrodes, is inevitable. Typical reasons for this are fabrication process non-idealities and inconsistencies and the growth of encapsulation tissue around the electrodes. In addition, changes in electrode size and separation would affect the impedances of the electrodes (Rd and Re), and hence, both the gain and the upper cut-off frequency of the bandpass filter to some extent. However, the inherently low sensitivity design of the filter circuits ensures that the effect on the system parameters would be small.
As an illustration of typical impedance variations, a 10 electrode MEC with stainless steel ring electrodes (diameter 1 mm, electrode pitch 1.5 mm) was fabricated and implanted on the second sacral spinal nerve root S2 (left) of a female sheep. Impedance measurements were taken using a two-wire configuration at 1 kHz with a 100 mV source voltage. The measurements are shown in Table 3 and indicate that the modulus of the impedance mismatch between pairs of electrodes may be as high as 200%, and therefore not negligible (note that the impedance values in Table 3 are illustrative only. This is because these values include contributions from both Re and Rd. At present we do not have estimates of these impedances separately). Clearly, then, it is critically important that the impedance mismatch is included in the design and analysis of the recording instrumentation.
To illustrate the effect of impedance mismatch, the effect on CMRR and crosstalk are discussed for both types of interfaces. As in the case of the matched interface networks, analysed in Section 2, KCL equations are derived, and the CM gain and crosstalk are calculated for both interfaces with mismatched electrode impedances. Note, also, that the same assumptions are applied as in Section 2. As a result, only the axial component of the electrode impedance (Rd) appears in the equations in Table 4 and Table 5 and only variations of Rd and Rcm need to be considered.
The effect of mismatch between two reference impedances (Rcm) on the gain is shown in Table 4, where Rcm and ΔRcm represent the average value of the reference impedance and its mismatch. For both circuits, the first term of the CM gain equations is the same as the gain of a network without mismatch. The second term is added to reflect the Rcm mismatch and therefore degrades the CMRR. However, it does not affect the differential gain and so crosstalk is not affected by Rcm mismatch.
The effect of axial impedance mismatch on common-mode and differential gain is summarised in Table 4. In these equations, Rdi represents the axial impedance of the ith electrode. Therefore, the mismatch of each electrode affects the CMRR and crosstalk of its corresponding channel only in both types of structure.

4. Validation by Simulation

4.1. Accuracy of the Approximate Equations

To verify the accuracy of the approximate equations derived in Section 2, a 10-channel interface was simulated (using SPICE) for both the ‘Type 1’ and ‘Type 2’ structures. The amplifier is assumed to have a high input impedance and not to interact with the interface, except for the amplifier’s current noise source. As already noted, this causes currents to flow in the interface circuit, resulting in differential voltages at the amplifier inputs. The effect of the amplifier current noise source is therefore expressed as an impedance. The other simulations considered are CMRR, crosstalk and thermal noise (note that the input-referred rms voltage noise does not depend on the design of the interface circuit and so is not considered). The comparison between the two sets of results for CMRR is shown in Figure 7 for different cases where some parameters are kept constant while others are swept over a practical range of interest (note that for compactness, only the curves describing the CMRR performance are shown). The variables chosen for this validation process are as shown in Table 6.
In Figure 7, the x-axes show the swept parameters, the others being held at their nominal values. The CMRR variation as a function of Rd and Ra (‘Type 1’)/R2 (‘Type 2’) is shown in Figure 7a,b. The inaccuracies caused by the various approximations are less than 0.05 dB compared with the SPICE simulations. Similarly, for crosstalk, where Rd and Rcm are varied, the inaccuracies caused by the approximations are less than 0.1 dB. The errors in the calculation of thermal noise and amplifier current noise gain are less than 3.6% and 0.08%, respectively.

4.2. Simulation of the Complete System

As an example of the characterisation of a complete recording system, a 10-channel ‘Type 1’ interface network (Figure 2a operating in the passband), connected to a low noise CMOS amplifier array, was simulated using SPICE. In this example, all the components in the interface network were matched (Ra = 10 MΩ, Rcm = Rd = 1 kΩ). The specifications of the ENG amplifier are summarised in Table 7.
The simulated and calculated results are shown in Table 8. It should be noted that for the calculated CMRR value, the measured amplifier CMRR (77.5 dB) is included (Equation (1)), whereas neglecting the effect of the amplifier results in a CMRR of 67.96 dB. For the noise density, Equation (14) calculates the total noise density referred to as the input of the amplifier, which results in 8.5   nV / Hz and, in order to refer the noise to the voltage source Vdi, it should be divided by the corresponding gain from (Equations (11) and (12)). For both crosstalk and noise parameters, the calculated values and simulation results are in good agreement, as illustrated in Table 8. Unfortunately, as explained above, it is not possible at present to incorporate the measured results given in Table 4 into these simulations, as separate values for Rd and Re are not available.

5. Discussion

5.1. Validity of Assumptions

5.1.1. Use of a Simulation-Based Analytical Study

As noted above, with few exceptions, the study presented in this paper is based on simulation. The justification for this is that: (a) the circuits employed to achieve a bandpass characteristic (first-order RC combinations) have inherently low sensitivity to parameter values; (b) that the bandwidth (centred around 1 kHz) is very low; (c) the circuits are fee-forward, i.e., no overall feedback is applied. Taken together, these aspects favour the use of simulation because (i) the use of RC circuits for frequency selection and the absence of overall feedback means that the transfer functions of the circuits have low Q in the frequency domain and, therefore, have low sensitivity to component tolerance. Furthermore, (ii) in the bandwidth of the systems discussed, the component models employed in commercial circuit simulators provide an adequate level of precision for most practical purposes.

5.1.2. The Use of Gaussian Noise Models

This paper has presented a detailed analysis of the noise performance (in terms of thermal noise, amplifier voltage noise and amplifier current noise) of the Type 1 and Type 2 circuits, with a focus on the design guidelines that aim to maximise the SNR. Throughout this analysis, the noise sources have been modelled using independent Gaussian processes—a common approach for noise modelling. Interference has been assumed to arise predominantly from high-frequency radio transmissions (RF) and would be dealt with via the appropriate selection of the corner frequencies of the bandpass filter. This is made possible because the RF sources are significantly out-of-band when compared to ENG. However, other biological sources within the body will also contribute to the interference, and it is likely that some of these will produce impulsive noise that may be better represented by coloured noise (e.g., Brownian noise). These sources will most likely be in-band, and thus not readily removed by simple filtering in the frequency domain. These sources are not considered in this paper; however, they would most likely appear at the recording interface as correlated sources akin to the ECG artefact. Thus, the analysis of the common-mode rejection ratio considered in Section 2.1 would apply in this case.

5.2. Optimisation of Component Values

A set of analytical design equations were derived in Section 2, linking the main system parameters (CMRR, crosstalk and noise) for two types (Types 1 and 2) of practical multi-channel interface networks to the underlying component values. These equations were validated (by simulation) in Section 3 and found to be accurate in all cases, at least when reasonable assumptions were made. In fact, it was noted that if R1 >> Rd, and R2 ~ Ra, the same design equations can be used for both types of interface networks, suggesting a unified graphical presentation to aid the designer. This is important, since the design of a multi-channel interface system, optimised to a particular specification, presents the designer with a very large set of design choices and possibilities.
The four plots in Figure 8 show the design equations for CMRR, crosstalk, thermal noise and amplifier current noise derived in Section 2 for both interface structures in a compact, visually accessible form. Figure 8a (based on Equation (5)) shows that CMRR is increased by increasing the values of the grounded resistors (Ra) and/or reducing the cuff impedances (Rd), always recognising that the upper limit of CMRR is determined by the amplifier itself. Note that crosstalk (Figure 8b and Equation (11)) can be reduced by reducing the ratio of Rd to Rcm, but for practical values of these elements, crosstalk is fairly constant.
In Figure 8c,d, the effects of thermal noise and gain of amplifier current noise at preamplifier input are plotted at constant temperature (37 °C) in terms of Rd and Rcm, assuming that Re is negligible in comparison with other impedance values. These plots demonstrate the effect of the interface network impedances on the components of the total input-referred noise that depends on them, i.e., neglecting the voltage noise contribution of the amplifier. Both these parameters can be reduced by decreasing the electrode impedance (Rd) but are both almost independent of the common-mode resistance, Rcm. Note, however, that since the total input-referred noise is the superposition of all three components (see Equation (14)), reducing any or all of them will reduce the total noise.
In summary, ∣Re∣ should be small because this will reduce both thermal and amplifier current noise (Equations (13) and (14)) and avoid attenuation of the neural signal. Equations (13) and (14) are valid for the noise, however, they need to be considered carefully when optimising the design. The noise is minimised by reducing both Re and Rd, however, because the signal amplitude is proportional to Rd [11,17], in practice, to maximise signal/noise, the cuff should be designed to maximise Rd and the effect that has on noise is unavoidable. Rcm should be as large as possible in order to minimise crosstalk (Equations (11) and (12)), which will only have a small effect on noise. Increasing Ra (Type 1) or R2 (Type 2) reduces CMRR (Equations (5) and (6)); the analysis in Section 3 does not show any difference between the two biasing networks. Finally, note that the maximum value of Ra is limited by the input impedance of the amplifier (possible potential divider effects reducing the signal amplitude) and the effect on input filter design (time constants, etc.).
The analysis in this paper is for networks operated in the passband. However, for a complete filter design, the capacitor sizes should be determined based on the resistor values chosen using the criteria discussed in this paper. The required lower and upper cut-off frequencies determine the Cs and Cp capacitor sizes, respectively. It should be noted that to have equal cut-off frequencies, all the Cs should be equal, and all the Cp should be equal as well.

6. Conclusions

Bandpass filter networks are essential for interfacing arrays of electrodes (such as nerve cuff electrodes) to amplifier arrays for applications, such as velocity selective recording (VSR). However, the inclusion of these networks influences the CMRR, crosstalk, and noise performance of such systems. In this paper, an analysis has been provided relating the performance criteria of two commonly used front-end biasing networks to component values. The effect of electrode mismatches is also discussed for both networks. It is shown that the ‘Type 1’ and ‘Type 2’ biasing networks result in a similar performance, provided that certain assumptions are satisfied.
The approach taken in the paper is to generalise the design problem as far as possible in order to aid the designer in the complex task of designing an interface network for a particular application. Using reasonable assumptions, the complicated state equations of the complete system are simplified into a manageable set of design equations, which are then verified for accuracy by comparison with the SPICE simulations of the complete system. Finally, a set of three-dimensional plots has been derived, which show how the main parameters of interest (CMRR, crosstalk and noise) depend in a very general way on the circuit component values. The knowledge thus provided to the designer will help in the construction of a first, probably approximate design, that can subsequently be refined by simulation.

Author Contributions

Conceptualization, S.S., B.M., N.D., N.G., J.P. and J.T.; Formal analysis, S.S. and J.T.; Funding acquisition, J.T.; Methodology, S.S., B.M. and J.T.; Supervision, J.T.; Validation, S.S.; Writing—original draft, S.S.; Writing—review & editing, B.M., N.D., N.G., J.P. and J.T. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by EPSRC grant number EP/P018947/1, and the APC is not applicable.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A

Example of the Symbolic Calculation of Common-Mode (CM) Gain Using Matlab

The gain at the input of the first amplifier to the common-mode input voltage Vcm, for the ‘Type 1’ structure, is given by the ratio of two polynomials a and b:
V a 1 V c m = a b  
where a and b are:
a =(A2)
36Ga9Gcm2Gd7Ge2 + 210Ga9Gcm2Gd6Ge3 + 462Ga9Gcm2Gd5Ge4 + 495Ga9Gcm2Gd4Ge5 + 286Ga9Gcm2Gd3Ge6 + 91Ga9Gcm2Gd2Ge7 +
15Ga9Gcm2GdGe8 + Ga9Gcm2Ge9 + 8Ga9GcmGd8Ge2 + 120Ga9GcmGd7Ge3 + 462Ga9GcmGd6Ge4 + 792Ga9GcmGd5Ge5 + 715Ga9GcmGd4Ge6
+ 364Ga9GcmGd3Ge7 + 105Ga9GcmGd2Ge8 + 16Ga9GcmGdGe9 + Ga9GcmGe10 + 288Ga8Gcm2Gd7Ge3 + 1470Ga8Gcm2Gd6Ge4 +
2772Ga8Gcm2Gd5Ge5 + 2475Ga8Gcm2Gd4Ge6 + 1144Ga8Gcm2Gd3Ge7 + 273Ga8Gcm2Gd2Ge8 + 30Ga8Gcm2GdGe9 + Ga8Gcm2Ge10 +
64Ga8GcmGd8Ge3 + 840Ga8GcmGd7Ge4 + 2772Ga8GcmGd6Ge5 + 3960Ga8GcmGd5Ge6 + 2860Ga8GcmGd4Ge7 + 1092Ga8GcmGd3Ge8 +
210Ga8GcmGd2Ge9 + 16Ga8GcmGdGe10 + 1008Ga7Gcm2Gd7Ge4 + 4410Ga7Gcm2Gd6Ge5 + 6930Ga7Gcm2Gd5Ge6 + 4950Ga7Gcm2Gd4Ge7 +
1716Ga7Gcm2Gd3Ge8 + 273Ga7Gcm2Gd2Ge9 + 15Ga7Gcm2GdGe10 + 224Ga7GcmGd8Ge4 + 2520Ga7GcmGd7Ge5 + 6930Ga7GcmGd6Ge6 +
7920Ga7GcmGd5Ge7 + 4290Ga7GcmGd4Ge8 + 1092Ga7GcmGd3Ge9 + 105Ga7GcmGd2Ge10 + 2016Ga6Gcm2Gd7Ge5 + 7350Ga6Gcm2Gd6Ge6 +
9240Ga6Gcm2Gd5Ge7 + 4950Ga6Gcm2Gd4Ge8 + 1144Ga6Gcm2Gd3Ge9 + 91Ga6Gcm2Gd2Ge10 + 448Ga6GcmGd8Ge5 + 4200Ga6GcmGd7Ge6 +
9240Ga6GcmGd6Ge7 + 7920Ga6GcmGd5Ge8 + 2860Ga6GcmGd4Ge9 + 364Ga6GcmGd3Ge10 + 2520Ga5Gcm2Gd7Ge6 + 7350Ga5Gcm2Gd6Ge7 +
6930Ga5Gcm2Gd5Ge8 + 2475Ga5Gcm2Gd4Ge9 + 286Ga5Gcm2Gd3Ge10 + 560Ga5GcmGd8Ge6 + 4200Ga5GcmGd7Ge7 + 6930Ga5GcmGd6Ge8 +
3960Ga5GcmGd5Ge9 + 715Ga5GcmGd4Ge10 + 2016Ga4Gcm2Gd7Ge7 + 4410Ga4Gcm2Gd6Ge8 + 2772Ga4Gcm2Gd5Ge9 + 495Ga4Gcm2Gd4Ge10
+ 448Ga4GcmGd8Ge7 + 2520Ga4GcmGd7Ge8 + 2772Ga4GcmGd6Ge9 + 792Ga4GcmGd5Ge10 + 1008Ga3Gcm2Gd7Ge8
+ 1470Ga3Gcm2Gd6Ge9 + 462Ga3Gcm2Gd5Ge10 + 224Ga3GcmGd8Ge8 + 840Ga3GcmGd7Ge9 + 462Ga3GcmGd6Ge10 + 288Ga2Gcm2Gd7Ge9 +
210Ga2Gcm2Gd6Ge10 + 64Ga2GcmGd8Ge9 + 120Ga2GcmGd7Ge10 + 36GaGcm2Gd7Ge10 + 8GaGcmGd8Ge10
b =(A3)
9Ga10Gcm2Gd8 + 120Ga10Gcm2Gd7Ge + 462Ga10Gcm2Gd6Ge2 + 792Ga10Gcm2Gd5Ge3 + 715Ga10Gcm2Gd4Ge4 + 364Ga10Gcm2Gd3Ge5 +
105Ga10Gcm2Gd2Ge6 + 16Ga10Gcm2GdGe7 + Ga10Gcm2Ge8 + 2Ga10GcmGd9 + 90Ga10GcmGd8Ge + 660Ga10GcmGd7Ge2 +
1848Ga10GcmGd6Ge3 + 2574Ga10GcmGd5Ge4 + 2002Ga10GcmGd4Ge5 + 910Ga10GcmGd3Ge6 + 240Ga10GcmGd2Ge7 + 34Ga10GcmGdGe8 +
2Ga10GcmGe9 + 10Ga10Gd9Ge + 165Ga10Gd8Ge2 + 792Ga10Gd7Ge3 + 1716Ga10Gd6Ge4 + 2002Ga10Gd5Ge5 + 1365Ga10Gd4Ge6 +
560Ga10Gd3Ge7 + 136Ga10Gd2Ge8 + 18Ga10GdGe9 + Ga10Ge10 + 90Ga9Gcm2Gd8Ge + 1080Ga9Gcm2Gd7Ge2 + 3696Ga9Gcm2Gd6Ge3 +
5544Ga9Gcm2Gd5Ge4 + 4290Ga9Gcm2Gd4Ge5 + 1820Ga9Gcm2Gd3Ge6 + 420Ga9Gcm2Gd2Ge7 + 48Ga9Gcm2GdGe8 + 2Ga9Gcm2Ge9 +
20Ga9GcmGd9Ge + 810Ga9GcmGd8Ge2 + 5280Ga9GcmGd7Ge3 + 12,936Ga9GcmGd6Ge4 + 15,444Ga9GcmGd5Ge5 + 10,010Ga9GcmGd4Ge6 +
3640Ga9GcmGd3Ge7 + 720Ga9GcmGd2Ge8 + 68Ga9GcmGdGe9 + 2Ga9GcmGe10 + 90Ga9Gd9Ge2 + 1320Ga9Gd8Ge3 + 5544Ga9Gd7Ge4 +
10,296Ga9Gd6Ge5 + 10,010Ga9Gd5Ge6 + 5460Ga9Gd4Ge7 + 1680Ga9Gd3Ge8 + 272Ga9Gd2Ge9 + 18Ga9GdGe10 + 405Ga8Gcm2Gd8Ge2 +
4320Ga8Gcm2Gd7Ge3 + 12,936Ga8Gcm2Gd6Ge4 + 16,632Ga8Gcm2Gd5Ge5 + 10,725Ga8Gcm2Gd4Ge6 + 3640Ga8Gcm2Gd3Ge7 +
630Ga8Gcm2Gd2Ge8 + 48Ga8Gcm2GdGe9 + Ga8Gcm2Ge10 + 90Ga8GcmGd9Ge2 + 3240Ga8GcmGd8Ge3 + 18,480Ga8GcmGd7Ge4 +
38,808Ga8GcmGd6Ge5 + 38,610Ga8GcmGd5Ge6 + 20,020Ga8GcmGd4Ge7 + 5460Ga8GcmGd3Ge8 + 720Ga8GcmGd2Ge9 + 34Ga8GcmGdGe10
+ 360Ga8Gd9Ge3 + 4620Ga8Gd8Ge4 + 16,632Ga8Gd7Ge5 + 25,740Ga8Gd6Ge6 + 20,020Ga8Gd5Ge7 + 8190Ga8Gd4Ge8 + 1680Ga8Gd3Ge9
+ 136Ga8Gd2Ge10 + 1080Ga7Gcm2Gd8Ge3 + 10,080Ga7Gcm2Gd7Ge4 + 25,872Ga7Gcm2Gd6Ge5 + 27,720Ga7Gcm2Gd5Ge6 +
14,300Ga7Gcm2Gd4Ge7 + 3640Ga7Gcm2Gd3Ge8 + 420Ga7Gcm2Gd2Ge9 + 16Ga7Gcm2GdGe10 + 240Ga7GcmGd9Ge3 + 7560Ga7GcmGd8Ge4 +
36,960Ga7GcmGd7Ge5 + 64,680Ga7GcmGd6Ge6 + 51,480Ga7GcmGd5Ge7 + 20,020Ga7GcmGd4Ge8 + 3640Ga7GcmGd3Ge9 +
240Ga7GcmGd2Ge10 + 840Ga7Gd9Ge4 + 9240Ga7Gd8Ge5 + 27,720Ga7Gd7Ge6 + 34,320Ga7Gd6Ge7 + 20,020Ga7Gd5Ge8 + 5460Ga7Gd4Ge9
+ 560Ga7Gd3Ge10 + 1890Ga6Gcm2Gd8Ge4 + 15,120Ga6Gcm2Gd7Ge5 + 32,340Ga6Gcm2Gd6Ge6 + 27,720Ga6Gcm2Gd5Ge7 +
10,725Ga6Gcm2Gd4Ge8 + 1820Ga6Gcm2Gd3Ge9 + 105Ga6Gcm2Gd2Ge10 + 420Ga6GcmGd9Ge4 + 11,340Ga6GcmGd8Ge5 +
46,200Ga6GcmGd7Ge6 + 64,680Ga6GcmGd6Ge7 + 38,610Ga6GcmGd5Ge8 + 10,010Ga6GcmGd4Ge9 + 910Ga6GcmGd3Ge10 + 1260Ga6Gd9Ge5
+ 11,550Ga6Gd8Ge6 + 27,720Ga6Gd7Ge7 + 25,740Ga6Gd6Ge8 + 10,010Ga6Gd5Ge9 + 1365Ga6Gd4Ge10 + 2268Ga5Gcm2Gd8Ge5 +
15,120Ga5Gcm2Gd7Ge6 + 25,872Ga5Gcm2Gd6Ge7 + 16,632Ga5Gcm2Gd5Ge8 + 4290Ga5Gcm2Gd4Ge9 + 364Ga5Gcm2Gd3Ge10 +
504Ga5GcmGd9Ge5 + 11,340Ga5GcmGd8Ge6 + 36,960Ga5GcmGd7Ge7 + 38,808Ga5GcmGd6Ge8 + 15,444Ga5GcmGd5Ge9 +
2002Ga5GcmGd4Ge10 + 1260Ga5Gd9Ge6 + 9240Ga5Gd8Ge7 + 16,632Ga5Gd7Ge8 + 10,296Ga5Gd6Ge9 + 2002Ga5Gd5Ge10 +
1890Ga4Gcm2Gd8Ge6 + 10,080Ga4Gcm2Gd7Ge7 + 12,936Ga4Gcm2Gd6Ge8 + 5544Ga4Gcm2Gd5Ge9 + 715Ga4Gcm2Gd4Ge10 +
420Ga4GcmGd9Ge6 + 7560Ga4GcmGd8Ge7 + 18,480Ga4GcmGd7Ge8 + 12,936Ga4GcmGd6Ge9 + 2574Ga4GcmGd5Ge10 + 840Ga4Gd9Ge7 +
4620Ga4Gd8Ge8 + 5544Ga4Gd7Ge9 + 1716Ga4Gd6Ge10 + 1080Ga3Gcm2Gd8Ge7 + 4320Ga3Gcm2Gd7Ge8 + 3696Ga3Gcm2Gd6Ge9 +
792Ga3Gcm2Gd5Ge10 + 240Ga3GcmGd9Ge7 + 3240Ga3GcmGd8Ge8 + 5280Ga3GcmGd7Ge9 + 1848Ga3GcmGd6Ge10 + 360Ga3Gd9Ge8 +
1320Ga3Gd8Ge9 + 792Ga3Gd7Ge10 + 405Ga2Gcm2Gd8Ge8 + 1080Ga2Gcm2Gd7Ge9 + 462Ga2Gcm2Gd6Ge10 + 90Ga2GcmGd9Ge8 +
810Ga2GcmGd8Ge9 + 660Ga2GcmGd7Ge10 + 90Ga2Gd9Ge9 + 165Ga2Gd8Ge10
+ 90GaGcm2Gd8Ge9 + 120GaGcm2Gd7Ge10 + 20GaGcmGd9Ge9 + 90GaGcmGd8Ge10 + 10GaGd9Ge10 + 9Gcm2Gd8Ge10 + 2GcmGd9Ge10
where
Gx = 1/Rx
Due to their complexity, these equations do not provide useful insight into the effects of the various parameters. Consider the effect of the following (reasonable) assumptions:
Rai >> Rcmi and
Rai >> Rei;
Rai >> Rdi; all i
a 36 G a G c m 2 G d 7 G e 10 + 8 G a G c m G d 8 8 G e 10  
b 9 G c m 2 G d 8 G e 10 + 2 G c m G d 9 G e 10  
Hence, the CM gain for the first channel can be expressed as follows:
V a 1 V c m 36 G a G c m 2 G d 7 G e 10 + 8 G a G c m G d 8 8 G e 10 9 G c m 2 G d 8 G e 10 + 2 G c m G d 9 G e 10 4 G a G d = 4 R d R a  
In general, for N = 10, this can be written as:
V a i V c m ( 5 i ) ( R d | | 2 R 1 ) R 2  

References

  1. Haugland, M. A flexible method for fabrication of nerve cuff electrodes. In Proceedings of the 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Amsterdam, The Netherlands, 31 October–3 November 1996; Volume 1, pp. 359–360. [Google Scholar]
  2. Kawai, K.; Tanaka, T.; Baba, H.; Bunker, M.; Ikeda, A.; Inoue, Y.; Kameyama, S.; Kaneko, S.; Kato, A.; Nozawa, T. Outcome of vagus nerve stimulation for drug-resistant epilepsy: The first three years of a prospective Japanese registry. Epileptic Disord. 2017, 19, 327–338. [Google Scholar] [CrossRef] [PubMed]
  3. Purser, M.F.; Mladsi, D.M.; Beckman, A.; Barion, F.; Forsey, J. Expected budget impact and health outcomes of expanded use of vagus nerve stimulation therapy for drug-resistant epilepsy. Adv. Ther. 2018, 35, 1686–1696. [Google Scholar] [CrossRef]
  4. Metcalfe, B.W.; Nielsen, T.N.; de N Donaldson, N.; Hunter, A.J.; Taylor, J. First demonstration of velocity selective recording from the pig vagus using a nerve cuff shows respiration afferents. Biomed. Eng. Lett. 2018, 8, 127–136. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Metcalfe, B.W.; Nielsen, T.; Taylor, J. Velocity selective recording: A demonstration of effectiveness on the vagus nerve in pig. In Proceedings of the 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Honolulu, HI, USA, 18–21 July 2018; pp. 1–4. [Google Scholar]
  6. Pflaum, C.; Riso, R.R.; Wiesspeiner, G. Performance of alternative amplifier configurations for tripolar nerve cuff recorded ENG. In Proceedings of the 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Amsterdam, The Netherlands, 31 October–3 November 1996; Volume 1, pp. 375–376. [Google Scholar]
  7. Koh, R.G.; Nachman, A.I.; Zariffa, J. Use of spatiotemporal templates for pathway discrimination in peripheral nerve recordings: A simulation study. J. Neural Eng. 2016, 14, 016013. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  8. Silveira, C.; Brunton, E.; Spendiff, S.; Nazarpour, K. Influence of nerve cuff channel count and implantation site on the separability of afferent ENG. J. Neural Eng. 2018, 15, 046004. [Google Scholar] [CrossRef] [PubMed]
  9. Aristovich, K.; Donegá, M.; Blochet, C.; Avery, J.; Hannan, S.; Chew, D.J.; Holder, D. Imaging fast neural traffic at fascicular level with electrical impedance tomography: Proof of principle in rat sciatic nerve. J. Neural Eng. 2018, 15, 056025. [Google Scholar] [CrossRef] [PubMed]
  10. Yoshida, K.; Kurstjens, G.; Hennings, K. Experimental validation of the nerve conduction velocity selective recording technique using a multi-contact cuff electrode. Med. Eng. Phys. 2009, 31, 1261–1270. [Google Scholar] [CrossRef] [PubMed]
  11. Taylor, J.; Donaldson, N.; Winter, J. Multiple-electrode nerve cuffs for low-velocity and velocity-selective neural recording. Med. Biol. Eng. Comput. 2004, 42, 634–643. [Google Scholar] [CrossRef] [PubMed]
  12. Donaldson, N.; Rieger, R.; Schuettler, M.; Taylor, J. Noise and selectivity of velocity-selective multi-electrode nerve cuffs. Med. Biol. Eng. Comput. 2008, 46, 1005–1018. [Google Scholar] [CrossRef] [PubMed]
  13. Metcalfe, B.W.; Chew, D.J.; Clarke, C.T.; Donaldson, N.D.N.; Taylor, J.T. A new method for spike extraction using velocity selective recording demonstrated with physiological ENG in Rat. J. Neurosci. Methods 2015, 251, 47–55. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  14. Rieger, R.; Taylor, J.; Demosthenous, A.; Donaldson, N.; Langlois, P.J. Design of a low-noise preamplifier for nerve cuff electrode recording. IEEE J. Solid-State Circuits 2003, 38, 1373–1379. [Google Scholar] [CrossRef]
  15. Rahal, M.; Winter, J.; Taylor, J.; Donaldson, N. An improved configuration for the reduction of EMG in electrode cuff recordings: A theoretical approach. IEEE Trans. Biomed. Eng. 2000, 47, 1281–1284. [Google Scholar] [CrossRef] [PubMed]
  16. Casas, O.; Pallas-Areny, R. Optimal bias circuit for instrumentation amplifiers. In Proceedings of the 14th Imeko World Congress, Tampere, Finland, 2–6 June 1997; pp. 143–148. [Google Scholar]
  17. Stein, R.; Pearson, K. Predicted amplitude and form of action potentials recorded from unmyelinated nerve fibres. J. Theor. Biol. 1971, 32, 539–558. [Google Scholar] [CrossRef]
Figure 1. Simplified schematic of a typical recording configuration using a MEC fitted to a peripheral nerve. There may be one or more ranks of amplification after the front-end network.
Figure 1. Simplified schematic of a typical recording configuration using a MEC fitted to a peripheral nerve. There may be one or more ranks of amplification after the front-end network.
Sensors 22 03450 g001
Figure 2. Front-end network with (a) ‘Type 1’ biasing arrangement and (b) ‘Type 2’ biasing arrangement.
Figure 2. Front-end network with (a) ‘Type 1’ biasing arrangement and (b) ‘Type 2’ biasing arrangement.
Sensors 22 03450 g002aSensors 22 03450 g002b
Figure 3. (a) Passband equivalent circuit for thermal noise analysis. This is a ladder expansion of the circuit of Figure 2a, i.e., for a Type 1 circuit. (b) The circuit of (a) with the resistors Ra is removed to reflect the fact that they appear in parallel with much smaller resistors and therefore contribute little to the thermal noise calculation.
Figure 3. (a) Passband equivalent circuit for thermal noise analysis. This is a ladder expansion of the circuit of Figure 2a, i.e., for a Type 1 circuit. (b) The circuit of (a) with the resistors Ra is removed to reflect the fact that they appear in parallel with much smaller resistors and therefore contribute little to the thermal noise calculation.
Sensors 22 03450 g003
Figure 5. Noise equivalent circuit of the amplifiers. The input impedance of the amplifiers is assumed to be infinite.
Figure 5. Noise equivalent circuit of the amplifiers. The input impedance of the amplifiers is assumed to be infinite.
Sensors 22 03450 g005
Figure 6. Simplified passband equivalent circuit for amplifier current noise analysis applicable to both Type 1 and Type 2 circuits.
Figure 6. Simplified passband equivalent circuit for amplifier current noise analysis applicable to both Type 1 and Type 2 circuits.
Sensors 22 03450 g006
Figure 7. Examples comparing the approximate equations and simulation results for the ‘Type 1’ and ‘Type 2’ interface networks. Minimum CMRR versus (a) Rd and (b) Ra (for Type 1)/R2 (for Type 2). The nominal component values are: R2 = Ra = 10 MΩ, R1 = 10 kΩ, Rcm = Rd = 1 kΩ. Note that the comparisons between calculation and simulation for the other parameters (i.e., crosstalk, thermal noise and amplifier current noise are not plotted, but the limiting values are quoted in the text).
Figure 7. Examples comparing the approximate equations and simulation results for the ‘Type 1’ and ‘Type 2’ interface networks. Minimum CMRR versus (a) Rd and (b) Ra (for Type 1)/R2 (for Type 2). The nominal component values are: R2 = Ra = 10 MΩ, R1 = 10 kΩ, Rcm = Rd = 1 kΩ. Note that the comparisons between calculation and simulation for the other parameters (i.e., crosstalk, thermal noise and amplifier current noise are not plotted, but the limiting values are quoted in the text).
Sensors 22 03450 g007
Figure 8. The effect of design parameters on (a) CMRR; (b) crosstalk; (c) thermal noise and (d) current noise gain. This 3D presentation of the material from Section 2 and Section 4 is discussed in Section 5.
Figure 8. The effect of design parameters on (a) CMRR; (b) crosstalk; (c) thermal noise and (d) current noise gain. This 3D presentation of the material from Section 2 and Section 4 is discussed in Section 5.
Sensors 22 03450 g008
Table 1. Connection matrix A for ‘Type 1’ front-end network.
Table 1. Connection matrix A for ‘Type 1’ front-end network.
12101112131420
1Ge1 + Ga100−Ge10000
20Ge2 + Ga200−Ge2000
: : : : : : : : : :
1000Ge10 + Ga100000−Ge10
11−Ge100Gcm1 + Ge1 + Gd1−Gd1000
120−Ge20−Gd1Ge2 + Gd1 + Gd2−Gd200
130000−Gd2Ge3 + Gd2 + Gd3−Gd30
: : : : : : : : : :
2000−Ge100000Gcm2 + Ge10 + Gd9
Table 2. Summary of the equations derived in this section (all Re, Ra, Rd are equal).
Table 2. Summary of the equations derived in this section (all Re, Ra, Rd are equal).
Description‘Type 1’ Circuit‘Type 2’ CircuitEquations No.
Common-Mode Gain V a i V c m ( 5 i ) R d R a V a i V c m ( 5 i ) ( R d | | 2 R 1 ) R 2 (5), (6)
Crosstalk Between Channels V a j V d i = { 1 R d 2 R c m + 9 R d i f   i = j R d 2 R c m + 9 R d i f   i j V a j V d i = { R d | | 2 R 1 R d ( 1 R d | | 2 R 1 2 R c m + 9 ( R d | | 2 R 1 ) ) i f   i = j R d | | 2 R 1 R d ( R d | | 2 R 1 2 R c m + 9 ( R d | | 2 R 1 ) ) i f   i j (11), (12)
Total Thermal Noise Density V a 1 ( r m s ) = 4 k T ( R d + 2 r e ( R e ) (13)
Total Input-Referred rms Noise Density v a ( total ) = 4 k T [ R d + 2 r e ( R e ) ] + v n 2 + i a 2 ( | R d + 2 R e | 2 + p | R e | 2 )   (14)
Table 3. Two-wire impedance measurements of the electrodes as dipoles for an implanted cuff in sheep.
Table 3. Two-wire impedance measurements of the electrodes as dipoles for an implanted cuff in sheep.
Electrodes2—Wire Impedance Measurements (100 mV, 1 kHz)
1–22.4 kΩ/−59°
2–32.0 kΩ/−58°
3–42.6 kΩ/−59°
4–53.3 kΩ/−60°
5–63.9 kΩ/−51°
6–72.5 kΩ/−47°
7–81.7 kΩ/−61°
8–91.4 kΩ/−59°
9–101.3 kΩ/−60°
Reference-11.1 kΩ/−48°
Reference-101.1 kΩ/−48°
Table 4. Common-mode and differential gain for mismatched common-mode impedances (only Rcms not equal).
Table 4. Common-mode and differential gain for mismatched common-mode impedances (only Rcms not equal).
StructureCommon-Mode GainDifferential Mode Gain
‘Type 1’ V a i V c m = ( 5 i ) R d R a + ( 2.5 R d R a ( 1 + 4.5 R d R c m ) ) ( Δ R c m R c m ) V a i V d j = { 1 R d 2 R c m + 10 R d i f   i = j R d 2 R c m + 10 R d i f   i j
‘Type 2’ V a i V c m = ( 5 i ) R d R 2 + R d | | 2 R 1 ( 4 R 2 9 ) ( 1 + 4.5 R d | | 2 R 1 R c m ) ( Δ R c m R c m ) V a j V d i = { R d | | 2 R 1 R d ( 1 R d | | 2 R 1 2 R c m + 9 ( R d | | 2 R 1 ) ) i f   i = j R d | | 2 R 1 R d ( R d | | 2 R 1 2 R c m + 9 ( R d | | 2 R 1 ) ) i f   i j
Table 5. Common-mode and differential gain for mismatched electrode impedances (Rds are not equal but Ras, Rcms are equal).
Table 5. Common-mode and differential gain for mismatched electrode impedances (Rds are not equal but Ras, Rcms are equal).
StructureCommon-Mode GainDifferential Mode Gain
‘Type 1’ V a i V c m = ( 5 i ) R d i R a V a i V d j = { 1 R d i 2 R c m + x = 1 10 R d x i f   i = j R d i 2 R c m + x = 1 10 R d x i f   i j
‘Type 2’ V a i V c m = ( 5 i ) R d i R 2
Table 6. Variables chosen for validation.
Table 6. Variables chosen for validation.
ParameterVariables Held ConstantSwept Variables
both circuits
CMRRR2 = Ra = 10 MΩRd, Ra (or R2 for Type 2)
CrosstalkR1 = 10 kΩRd, Rcm
Thermal noiseRcm = Rd = 1 kΩRd
Amplifier current noise Rd, Rcm
Table 7. Summary of amplifier specifications [14].
Table 7. Summary of amplifier specifications [14].
ParameterSpecifications
Technology0.35 µm 4-metal 2-poly CMOS
Power supply±1.5 V
Midband Gain79.7 dB
−3 dB frequencies
Lower258 Hz
Upper24.1 kHz
CMRR (@3 KHz)77.5 dB (AV,CM = 1.29)
PSRR(@3KHz)
VDD50.57 dB
VSS40.2 dB
Adjacent channel interference(crosstalk)<−100.1 dB
Total input-referred voltage noise density@3 kHz 7.5   nV / Hz
Total input-referred current noise density@3 kHz 0.55   pA / Hz
Total input-referred rms voltage noise 1 Hz–31 kHz for a source resistance of 1 kΩ 1.82   uV
Table 8. Summary of a complete system with Type 1 interface (Ra = 10 MΩ, Rcm = Rd = 1 kΩ, Re = 0).
Table 8. Summary of a complete system with Type 1 interface (Ra = 10 MΩ, Rcm = Rd = 1 kΩ, Re = 0).
ParameterCalculationSimulation
Min(CMRR) (@3 KHz)65.29 dB65.08 dB
Adjacent channel interference (crosstalk)−20.83 dB−20.81 dB
Total input-referred voltage noise density@3 kHz 9.35   nV / Hz 9.32   nV / Hz
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Sadrafshari, S.; Metcalfe, B.; Donaldson, N.; Granger, N.; Prager, J.; Taylor, J. The Design of a Low Noise, Multi-Channel Recording System for Use in Implanted Peripheral Nerve Interfaces. Sensors 2022, 22, 3450. https://doi.org/10.3390/s22093450

AMA Style

Sadrafshari S, Metcalfe B, Donaldson N, Granger N, Prager J, Taylor J. The Design of a Low Noise, Multi-Channel Recording System for Use in Implanted Peripheral Nerve Interfaces. Sensors. 2022; 22(9):3450. https://doi.org/10.3390/s22093450

Chicago/Turabian Style

Sadrafshari, Shamin, Benjamin Metcalfe, Nick Donaldson, Nicolas Granger, Jon Prager, and John Taylor. 2022. "The Design of a Low Noise, Multi-Channel Recording System for Use in Implanted Peripheral Nerve Interfaces" Sensors 22, no. 9: 3450. https://doi.org/10.3390/s22093450

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.