Next Article in Journal
Railway Track Fault Detection Using Selective MFCC Features from Acoustic Data
Previous Article in Journal
Sensor Fusion and Advanced Controller for Connected and Automated Vehicles
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Determining Magnitudes of Forces at Known Locations through a Strain Gauge Force Transducer

Department of Mechanical and Electrical Engineering, Wilkes University, Wilkes-Barre, PA 18766, USA
*
Author to whom correspondence should be addressed.
Sensors 2023, 23(16), 7017; https://doi.org/10.3390/s23167017
Submission received: 26 May 2023 / Revised: 29 July 2023 / Accepted: 2 August 2023 / Published: 8 August 2023
(This article belongs to the Section Physical Sensors)

Abstract

:
A novel strain gauge force transducer was developed to minimize the number of strain gauges needed to determine the magnitudes of loads when the locations are known. This innovative methodology requires only one strain gauge for each force magnitude desired, reducing the complexity and cost associated with traditional approaches. The theory was verified with laboratory experiments. Seven uniaxial strain gauges were attached to the underside of a simply supported, slender, aluminum beam. One or more loads were applied either directly atop strain gauges or in known positions between strain gauges. Experiments were conducted on several different single and double-load configurations to evaluate the extent of the new methodology which yielded average errors under 5% for the cases where loads were direct atop strain gauges and 6.6% for the cases where the loads were between strain gauges. These findings indicate the potential of this novel strain gauge force transducer to revolutionize load measurement in scenarios where load locations are predetermined.

1. Introduction

Strain gauge force transducers are not new in the realm of engineering, as strain gauges are especially useful and prominent tools in structural analysis. This methodology implements the use of strain gauges on a small-scale laboratory experiment with the ability to expand as needed into larger structures. A transducer is an instrument that converts one form of energy into another [1].
Regarding the context of the laboratory experiment, the mechanical energy of the applied forces are converted into an electrical signal via the use of a Wheatstone bridge circuit that is integrated in a Vishay P3 recorder [2]. Strain gauges are constructed on a thin, flexible material with high-gauge wire. These gauges operate such that the resistance of the wire changes under an applied load. The Wheatstone bridge that is used is able to measure very fine changes in the resistance of the wire. Essentially, the circuit is made up of four resistors wired in a diamond pattern, with three of those resistors being highly precise values and the other being the strain gauge in the system to complete the circuit. By assessing the change of resistance of the strain gauge, the strain value is the response to an applied load.Load identification and the determination of forces and/or loads on a structure from measured responses are critical in monitoring structural health. The process as a whole is challenging due to the uncertainty of measurements and the complexity of loads.
As these tools are very effective, they have increasingly become a topic of research for load determination. An alternative method of force determination to strain gauges are Fiber Bragg Grating (FBG) sensors and fiber optics, which are prominently used in structural analysis. One study shows that researchers were able to create a method to identify the load of a moving car on a bridge through the use of long-gauge fiber optic strain responses [3]. Similarly, another study shows that the use of an FBG sensor is able to measure longitudinal strains when a beam is subjected to multiple loads [4]. An additional study looked at using an FGB sensor to measure the force and position of a weight that was attached to a surgical forceps [5]. The practicality of FBG sensors continues as these are also used in measuring dynamic strain in cantilever beams, which can yield strains in linear and nonlinear systems [6]. Ultimately, the decision was made to focus on strain gauges. Strain gauges display remarkable accuracy in the detection of minor changes of resistance which as a result can precisely quantify any applied loads. The methodology found in this study reduces the number of necessary strain gauges and promises a more cost-effective approach which results in a practical application of identifying loads.
Innovative techniques using nonlinear strain gauges are frequently being studied, such as one that highlights a triaxial force transducer and the calibration method [7]. Although rigidly mounted, the strain gauges are still able to observe dynamic forces such as those caused by acceleration and vibration, thus making this method highly desirable in load testing for aircraft and automobiles [7]. As the testing highlighted in this study is static, nonlinear strain gauges are not necessary. Furthermore, the strain present is only in one direction, thus making a triaxial strain gauge excessive.
Previously, a strain gauge force transducer was created to determine both the magnitudes and locations of several applied forces at the same time [8]. To do so, four strain gauges are required for the initial force and two more are necessary for each additional applied force. The methodology focused on in this study aimed to reduce the number of necessary strain gauges for determining the load magnitude by eliminating the unknown locations. The benefit of knowing the location where the force is applied is such that only one strain gauge is needed per force.

2. Theory

Consider a simply supported beam with a given length L subjected to a force F (see Appendix A) at a location A, and that has two reaction forces at each end, R A and R B , as shown in Figure 1. The beam is assumed to be in static equilibrium. Therefore, it is expected that the maximum moment, M max is to occur at the point of the applied force A, as shown in Figure 2. To recreate this graph, a piece-wise function composed of two equations is used: one representing the increasing moment before the point of the applied force and the other representing the decreasing moment after the point of the applied force [9].
M = F x ( 1 A L ) 0 x A F A ( 1 x L ) A x L
The applied force F can be determined if the moment at the location of the force A is known, from Equation (1). However, it is not possible to directly measure the moment, so the relationship between bending stress and moment must be utilized. This relationship is represented in Equation (2), where σ denotes bending stress, y is the distance from the neutral axis of the beam to where the stress is calculated, and I is the area moment of inertia [10].
σ = M y I
Bending stress itself cannot be directly measured, but it can be accurately calculated using the well-established relationship between stress and strain. This relationship is represented in Equation (3), where E denotes Young’s modulus, a material property that quantifies the stiffness of a solid, and ε symbolizes the strain, a dimensionless measure of the deformation experienced by the material. By utilizing this fundamental relationship, it is possible to infer the bending stress from the measured strain values [10].
σ = E ε
Strain, being a physical quantity, can be measured using strain gauges. By combining Equations (2) and (3), a relationship can be established between the bending moment and strain resulting in Equation (4).
M = E I y ε
A calibration factor, β , is used to account for different geometric and material properties. In this experiment, one β was used since a beam that was homogeneous with a constant cross-section was utilized. However, if the beam has variable width or thickness, or is non-homogeneous, β must be calculated separately for each strain gauge. The experimental β can be determined with the use of a known force value and a tool such as Excel Solver. The theoretical β can be represented by Equation (5). Equation (4) was subsequently rewritten to include β and is represented by Equation (6).
β = E I y
M = β ε
By using the relationship established between moment and strain, the magnitude of a force can be calculated using Equations (1) and (6). In situations where multiple forces are present, as shown in Figure 3, it is necessary to introduce multiple strain gauges, with one strain gauge placed at each force location. If superposition is used, which states that the total effect of multiple forces on a system is equal to the sum of the effects of each individual force, Equation (1) can be turned into a summation of all strain gauges to cover all the forces [11]. This is represented in Equation (7), where m represents the location at which the moment is calculated, N represents the total number of forces, and n is the dummy index. L n and L m represent the location of the forces. Lastly, the F n represents the force magnitude for various locations.
M m = n = 1 m F n L n ( 1 L m L ) + n = m + 1 N F n L m ( 1 L n L )
In the experiment being conducted, seven force locations were utilized. If Equation (7) is to be written in its full matrix form to encompass all seven forces, it would result in Equation (8).
M 1 M 2 M 3 M 4 M 5 M 6 M 7 = L 1 ( 1 L 1 L ) L 1 ( 1 L 2 L ) L 1 ( 1 L 3 L ) L 1 ( 1 L 4 L ) L 1 ( 1 L 5 L ) L 1 ( 1 L 6 L ) L 1 ( 1 L 7 L ) L 1 ( 1 L 2 L ) L 2 ( 1 L 2 L ) L 2 ( 1 L 3 L ) L 2 ( 1 L 4 L ) L 2 ( 1 L 5 L ) L 2 ( 1 L 6 L ) L 2 ( 1 L 7 L ) L 1 ( 1 L 3 L ) L 2 ( 1 L 3 L ) L 3 ( 1 L 3 L ) L 3 ( 1 L 4 L ) L 3 ( 1 L 5 L ) L 3 ( 1 L 6 L ) L 3 ( 1 L 7 L ) L 1 ( 1 L 4 L ) L 2 ( 1 L 4 L ) L 3 ( 1 L 4 L ) L 4 ( 1 L 4 L ) L 4 ( 1 L 5 L ) L 4 ( 1 L 6 L ) L 4 ( 1 L 7 L ) L 1 ( 1 L 5 L ) L 2 ( 1 L 5 L ) L 3 ( 1 L 5 L ) L 4 ( 1 L 5 L ) L 5 ( 1 L 5 L ) L 5 ( 1 L 6 L ) L 5 ( 1 L 7 L ) L 1 ( 1 L 6 L ) L 2 ( 1 L 6 L ) L 3 ( 1 L 6 L ) L 4 ( 1 L 6 L ) L 5 ( 1 L 6 L ) L 6 ( 1 L 6 L ) L 6 ( 1 L 7 L ) L 1 ( 1 L 7 L ) L 2 ( 1 L 7 L ) L 3 ( 1 L 7 L ) L 4 ( 1 L 7 L ) L 5 ( 1 L 7 L ) L 6 ( 1 L 7 L ) L 7 ( 1 L 7 L ) F 1 F 2 F 3 F 4 F 5 F 6 F 7
Equation (8) can be written in a condensed form, represented by Equation (9), where [M] is the column vector representing the bending moment at each location, [F] is the column vector representing the magnitude of the applied force at each location, and [L] is the geometric matrix.
M = L F
As matrix [L] is symmetric, it can be inverted to calculate the forces at each of the locations as represented in Equation (10). Additionally, Equation (6) can be used to replace the moment matrix with the strain gauge values from the experiment and multiplied by the calibration factor, as represented in Equation (11). This establishes the final relationship between the strain gauge readings and the forces.
F = L 1 M
F = β L 1 [ ε ]
If the position of a force between two strain gauges is known, the force will be distributed to the two nearby strain gauges as shown in Figure 4. Force F b is the actual force being applied; however, the strain gauges will see that the force is distributed among the two strain gauges nearby as F 1 and F 2 . b represents the position of the force referenced from the left adjacent strain gauge and refers to the spacing between such adjacent strain gauges. In Equations (12) and (13), the forces at the adjacent strain gauges can be calculated by knowing such values.
F 1 = ( 1 b ) F b
F 2 = b F b

3. Experimental Results

3.1. Experimental Setup

The experiments conducted here represent only a few of the many scenarios in which the methodology stated can be used for load identification purposes. The rationale behind selecting these scenarios is rooted in their prevalence within real-world situations. They represent commonly encountered cases rather than extreme or overly simplistic scenarios. As shown in Figure 5, masses were hung from a thin cord on top of a strain gauge or between strain gauges to simulate a point load condition. This was repeated for various placements of one or multiple loads on the beam. With the knowledge of the strain gauge locations, the strain readings, and the locations of the applied loads, one could use the methodology presented to calculate the force magnitudes. In order to conduct the experiment, certain values were recorded such as the strain gauge readings, the location of the strain gauges themselves, and the positions of the applied loads.
The setup for this experiment consisted of an aluminum 6061-T6 simply supported beam with the dimensions of 3.81 cm × 0.476 cm × 61 cm shown in Figure 5. At each end, 5.5 cm of the beam was in contact with the supports, resulting in a gap of 50 cm. Seven Vishay, uniaxial strain gauges of model number CEA-12-240-UZ-120 were attached to the underside of the beam, as seen in Figure 6, at locations shown in Figure 7. These strain gauges were applied to the beam by using Vishay application techniques [12]. Wires were soldered to the leads on the strain gauge to read the strain values by using two Vishay P3 indicators.
To establish the necessary connections, a total of three wires were used for each strain gauge. Referring to Figure 6, a red wire was connected to one of the pads on the strain gauge, while the second pad had two wires connected to it: white and black. Figure 8 illustrates the other end of these wires. The red wire was linked to the terminal labeled P + , the white wire to the terminal labeled S , and finally, the black wire to the terminal labeled D 120 . P is the excitation voltage (typically 10 V) and S is the measurement voltage of the Wheatstone bridge. The connection of the black wire to the D 120 terminal was based on the grid resistance of 120 ohms. This resistance value is determined by the strain gauge manufacturer and the construction of the strain gauge itself.

3.2. Calibration Factor

To effectively carry out the experiment, it was crucial to first determine the calibration factor. To obtain the value, a simultaneous uniform loading was applied. This was done by loading 10.8 N first on the center strain gauge and repeatedly adding 10.8 N adjacent to the existing loads until all gauges had a force applied. Upon collecting the data, the solver function in Excel was employed to attain the optimal fit for the force, ensuring it equaled 10.8 N. The actual calibration factor was determined to be 0.00932, and to verify its accuracy, the theoretical beta factor was calculated using the beam geometry and material properties from Equation (5), resulting in a value of 0.00994. The comparison between the theoretical beta and the calibrated beta value led to a minor difference of 6.3%, as shown in Table 1. By multiplying the obtained strain data by the calibration factor (beta), the moment could be calculated, providing insights into how the beam responds under different loading conditions.

3.3. Single Load Applied on Top of a Strain Gauge

For this condition, a 9.8 N load was applied on top of a single strain gauge and all the strains were read and shown in Table 2. For each test, the load was moved to a different strain gauge. The point location in Table 2 refers to the distance away from the left end at which the beam makes contact with the supports.
The moment graph was created by using Equation (6). By observing the moment graph for the 9.8 N experiment shown in Figure 9, it can be determined how each load applied affects the surrounding area on the beam. Wherever the load is applied on the beam, the moment distribution will peak at that location. This shows the concentration of stress and bending at that particular location where the load was applied.
The forces were calculated by taking the moment matrix and multiplying it by the inverse geometric matrix using Equation (11), and the results are presented in Table 3. Upon examining the table, it can be observed that the force at the strain gauge location with the applied load should be close to 9.8 N, while all other strain gauge locations should theoretically register zero force. This observation is consistent with the expectation that only the applied load should affect the force at the specific strain gauge location.
The force errors in the experiment take into account the differences between expected and calculated forces. In contrast, the zero-force errors refer to the comparison of forces at all other strain gauge locations without the applied force, effectively evaluating the accuracy of the zero-force assumption at these locations. To compute the averages for both force errors and zero force errors, the respective values were summed up and then divided by the total number of forces applied.
By comparing calculated and expected forces, average and zero errors can be computed not only for the 9.8 N experiment but also for the 19.6 N experiment. The results of this comparison can be observed in Table 4. This additional analysis helps to assess the overall accuracy and reliability of the experimental setup and the underlying mathematical models used for force calculation across different load magnitudes.

3.4. Multiple Loads Applied on Top of Strain Gauges

The adjacent load experiment (Figure 10) consisted of two 9.8 N forces applied at the first and second points, and then moving the pair of loads down the beam. Table 5 and Table 6 show the strain readings and calculated forces for this scenario simulating a two-axle vehicle moving along a bridge.
The skip strain gauge experiment (Figure 11) involved the application of two 9.8 N loads on strain gauges placed at specific locations, with an unloaded strain gauge in-between. The first load was placed on strain gauge 1 and the second load on strain gauge 3, intentionally skipping strain gauge 2. This trial skips strain gauge 2 which simulates loads being apart from each other with the strain gauge in the middle. This experiment is done to see how the beam responds to multiple loads that are not directly adjacent to each other. This same procedure is repeated for all the points. Table 7 and Table 8 show the strain readings and the calculated forces from this method.
In both multiple load experiments, the average force and zero error are calculated by comparing the forces with the expected value, and the results are presented in Table 9.

3.5. Single Load Applied In-Between Two Strain Gauges

The data collected from the in-between test used a 19.6 N load which was applied at various different ratios between known locations. Three different ratios were tested, 25%, 50%, and 75%. The load distribution relies solely on the location of the load relative to the adjacent strain gauges. For example, if the load is applied 25% away from strain gauge 2, that means the force is 75% away from strain gauge 3. Inversely, the experienced load at strain gauge 2 will be 75% of the applied load, and strain gauge 3 will experience 25% of the applied load. Table 10 shows the data that were collected for this experiment. From the data collected, the data points start at strain gauge 3, where 3.25 represents the load 25% of the way of the strain gauge, 3.50 represents the load 50% between the strain gauges 3 and 4, and 3.75 represents the load being applied 75% of the way between the strain gauges 3 and 4. The final forces that were calculated are shown in Table 11.
Figure 12 shows the force distribution for the in-between test. It relates to the values from Table 11, solidifying the concept of the method by showing that placing a load 50% away from one strain gauge and 50% away from the other indicates that both of the strain gauges should experience a similar amount of load. Keep in mind that this is not the full magnitude of the force but rather 50% as the force was placed directly in the middle. Figure 13 shows the specific case of the load being placed halfway between strain gauges 4 and 5. Theoretically, the peak of the force should occur directly between strain gauges 4 and 5. As there is no strain gauge available in this scenario, to get the peak force value, it can be observed that the force is distributed into strain gauges 4 and 5 shown by the orange line. Referring back to Figure 4, this is the expected outcome. Where the applied force F b cannot be directly measured, the strain gauges pick up the force that is now distributed by the beam, and the distance is what determines the magnitude of the force that will be seen by the strain gauges, and this is supported by Equations (12) and (13). This can be seen in Figure 13, by drawing a theoretical dotted line that peaks between strain gauges 4 and 5, as represented by the blue dotted line. Using Table 11, the average force error and the zero force error can now be calculated which will determine how accurate this experiment was in determining the magnitude of a force that was not directly atop a strain gauge; the results are shown in Table 12.

3.6. Load Applied on Top of All Strain Gauges

The experiment evaluated the system response under simultaneous loading, exploring varied and uniform weight conditions. The varied case was initially tested by applying distinct loads to each strain gauge to validate the ability of the system to distinguish the loads based on their magnitudes. Table 13 outlines the tested scenarios for this case, beginning with a single gauge under load and incrementally adding loads to the remaining gauges. The recorded strain readings from the experiment are documented in Table 14. A moment diagram was subsequently derived using the above equations, as illustrated in Figure 14.
The calculated moments were then used to estimate the forces. As indicated in Table 15, the forces were largely accurate in identifying correct magnitudes, barring the final scenario where a load was applied at every gauge, including the ends. It is postulated that the errors in force readings arose from the lifting off of the edges of the beam from the surfaces they were resting on, leading to inaccurate strain readings.
The second tested case was the simultaneous uniform case, where identical loads were applied to all strain gauges. Table 16 presents the tested scenarios for this case. Strain values were recorded during the experiment, as shown in Table 17. A moment diagram was subsequently derived, with the moment curve represented in Figure 15. Interestingly, this curve exhibited a single peak despite identical loads being applied at every strain gauge—a result of the superposition principle, where the individual load peaks accumulate into a larger single peak.
The moments were then utilized to estimate the forces, as detailed in Table 18. The forces were plotted to examine their peaks, as illustrated in Figure 16. Despite the moment curve displaying a single peak, the superposition principle allowed the discernment of individual force magnitudes. The final scenario, with an identical load applied at all gauges, yielded a nearly flat line at the top, indicating the uniformly applied force.
Upon calculating the forces, force averages were determined to gauge the accuracy of both cases, as shown in Table 19. Two primary distinctions between the cases and their force average errors emerged. The first potential cause is the calibration factor, originally derived from a case similar to the uniform load scenario, which may account for its lower error. Additionally, the edges of the beam were observed to lift from the surfaces they were resting on during the varied load case, potentially contributing to additional errors. Despite these challenges, both cases were successful overall in accurately identifying the force loads applied to the beam.

4. Conclusions

This new methodology of force determination using a uniaxial strain gauge at each force location yielded results such that the process can be deemed accurate for simply supported, uniform beams. Additionally, this methodology identifies the magnitudes of loads at given locations while successfully calculating zero force loads elsewhere on the beam. The major benefit of this methodology is the minimization of the number of strain gauges required to determine force magnitudes on a beam. A reduction in strain gauges minimizes the number of channels required, thus further lowering the overall cost per test.
Before the experimental data were collected, the goal was to have the error fall within 5%. The overall force errors for cases where the loads were in line with the strain gauge locations peaked at 4.9%. The majority of errors for these data sets could be attributed to slight discrepancies in strain gauge placements and load location, as well as small imperfections in the thickness of the beam. Additionally, movement could have occurred resulting in the designated resting areas possibly shifting.
The data from the load being applied between strain gauges yielded an error of 6.6%. This is slightly higher than the goal, but causes for errors are most likely attributed to the inaccurate placement of the loads between strain gauges. Overall, the methodology and theory both have the ability to be modified for various load applications to suit any need while maintaining accuracy.

Author Contributions

Conceptualization, E.B.III; methodology, E.B.III; validation, J.P., B.H.J. and C.D.; formal analysis, J.P.; investigation, J.P.; resources, Wilkes University; data curation, B.H.J. and C.D.; writing—original draft preparation, J.P., B.H.J. and C.D.; writing—review and editing, E.B.III and A.S.; supervision, A.S.; project administration, E.B.III; funding acquisition, Provided by Wilkes University. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data is contained within the article.

Acknowledgments

Brian Dwyer, Gavin Scofield, Tyler Weatherby, Dan Yenca, and Dillon Yoshitomi.

Conflicts of Interest

The authors declare no conflict of interest.

Appendix A

FApplied Force
xDistance away from the left side of the beam
AApplied Force Location
σ Bending Stress
MBending moment of the beam at the strain gauge location
yDistance from the Neutral Axis
ICross-sectional area moment of inertia of the beam
EYoung’s Modulus
ε Strain
β Calibration Factor

References

  1. The Technical Staff of Measurements Group Inc. Strain Gage Based Transducers: Their Design and Construction; Measurements Group, Inc.: Raleigh, NC, USA, 1988. [Google Scholar]
  2. Vishay Measurements Group Inc. Measurement of Force, Torque, and Other Mechanical Variables with Strain Gages; Measurements Group, Inc.: Raleigh, NC, USA, 2011. [Google Scholar]
  3. Zhang, Q.; Zhao, W.; Zhang, J. Moving load identification with long gauge fiber optic strain sensing. Balt. J. Road Bridge Eng. 2021, 16, 131–158. [Google Scholar] [CrossRef]
  4. Hong, K.; Lee, J.; Choi, S.W.; Kim, Y.; Park, H.S. A strain-based load identification model for beams in building structures. Sensors 2013, 13, 9909–9920. [Google Scholar] [CrossRef] [Green Version]
  5. Liang, Q.; Cheng, L.; Long, J.; Li, Z.; Sun, W.; Zhang, D.; Wang, Y. A force/position measurement method of surgical forceps based on fiber Bragg gratings. IEEE Sens. J. 2023, 23, 363–373. [Google Scholar] [CrossRef]
  6. Song, X.; Zhang, Y.; Liang, D. Load identification for a cantilever beam based on fiber Bragg grating sensors. Sensors 2017, 17, 1733. [Google Scholar] [CrossRef] [Green Version]
  7. Wang, Q.; Xu, F.; Guo, W.; Gao, M. New technique for impact calibration of wide-range triaxial force transducer using Hopkinson Bar. Sensors 2022, 22, 4885. [Google Scholar] [CrossRef] [PubMed]
  8. Bednarz, E.T., III; Zhu, W.D.; Smith, S.A. Identifying the magnitude and location of a load on a slender beam using a strain gauge based force transducer. Strain 2013, 49, 276–285. [Google Scholar] [CrossRef]
  9. Hibbeler, R.C. Structural Analysis, 10th ed.; Pearson: New York, NY, USA, 2018. [Google Scholar]
  10. Hibbeler, R.C. Mechanics of Materials Edition, 10th ed.; Pearson: Boston, MA, USA, 2017. [Google Scholar]
  11. Hugh, D.Y.; Freedman, R.A.; Ford, A.L. Sears and Zemansky’s University Physics, 13th ed.; Addison-Wesley: Boston, MA, USA, 2012. [Google Scholar]
  12. Micro-Measurements. Strain Gage Installation with M-Bond 200 Adhesive. 2018. Available online: https://www.micro-measurements.com/ (accessed on 30 May 2023).
Figure 1. Simply Supported Beam with Single Point Force.
Figure 1. Simply Supported Beam with Single Point Force.
Sensors 23 07017 g001
Figure 2. Bending Moment Curve for Simply Supported Beam with Single Point Force.
Figure 2. Bending Moment Curve for Simply Supported Beam with Single Point Force.
Sensors 23 07017 g002
Figure 3. Strain Gauge Locations and Corresponding Force Labels for Reference.
Figure 3. Strain Gauge Locations and Corresponding Force Labels for Reference.
Sensors 23 07017 g003
Figure 4. Theoretical Force Distribution Between Adjacent Strain Gauges in In-Between Case.
Figure 4. Theoretical Force Distribution Between Adjacent Strain Gauges in In-Between Case.
Sensors 23 07017 g004
Figure 5. Experimental Setup.
Figure 5. Experimental Setup.
Sensors 23 07017 g005
Figure 6. Installation of Strain Gauges on the Beam.
Figure 6. Installation of Strain Gauges on the Beam.
Sensors 23 07017 g006
Figure 7. Beam Layout with Strain Gauge Locations and Dimensions.
Figure 7. Beam Layout with Strain Gauge Locations and Dimensions.
Sensors 23 07017 g007
Figure 8. Vishay P3 Strain Indicator and Recorder.
Figure 8. Vishay P3 Strain Indicator and Recorder.
Sensors 23 07017 g008
Figure 9. Moment Distribution: Single Load 9.8 N Case.
Figure 9. Moment Distribution: Single Load 9.8 N Case.
Sensors 23 07017 g009
Figure 10. Moment Distribution: Adjacent Load Case.
Figure 10. Moment Distribution: Adjacent Load Case.
Sensors 23 07017 g010
Figure 11. Moment Distribution: Skip Strain Gauge Load Case.
Figure 11. Moment Distribution: Skip Strain Gauge Load Case.
Sensors 23 07017 g011
Figure 12. Force Distribution: In-Between Load 44.5 N Case.
Figure 12. Force Distribution: In-Between Load 44.5 N Case.
Sensors 23 07017 g012
Figure 13. Comparison of Force Distribution Curve for 50% Load Between Strain Gauges 4 and 5 and Theoretical Applied Load.
Figure 13. Comparison of Force Distribution Curve for 50% Load Between Strain Gauges 4 and 5 and Theoretical Applied Load.
Sensors 23 07017 g013
Figure 14. Moment Distribution: Simultaneous Varied Load Case.
Figure 14. Moment Distribution: Simultaneous Varied Load Case.
Sensors 23 07017 g014
Figure 15. Moment Distribution: Simultaneous Uniform Load Case.
Figure 15. Moment Distribution: Simultaneous Uniform Load Case.
Sensors 23 07017 g015
Figure 16. Force Distribution: Simultaneous Uniform Load Case.
Figure 16. Force Distribution: Simultaneous Uniform Load Case.
Sensors 23 07017 g016
Table 1. Beta Calculation Values.
Table 1. Beta Calculation Values.
b (m)h (m)I (m4)y (m)E (Pa)Theo. BetaCalibrated Beta% Difference
0.03810.004763.43 × 10−100.0023869 × 1090.009940.009326.3%
Table 2. Strain Readings: Single Load 9.8 N Case.
Table 2. Strain Readings: Single Load 9.8 N Case.
Point (m)Strain ( μ ε )
1234567
0.0763574737302011
0.13541028870553821
0.194587124102805531
0.2537711021351057342
0.312854781031289152
0.37203754729110662
0.4310203039516066
Table 3. Force Values: Single Load 9.8 N Case.
Table 3. Force Values: Single Load 9.8 N Case.
Point (m)Force (N)
1234567
0.079.30.60.0−0.50.5−0.20.1
0.13−0.39.60.6−0.50.30.00.2
0.19−0.50.89.20.00.5−0.20.4
0.25−0.40.5−0.39.80.3−0.20.8
0.31−0.30.3−0.20.09.60.30.9
0.370.00.0−0.2−0.20.69.21.4
0.43−0.20.00.2−0.50.50.59.7
Table 4. Error Values: Single Loading Cases.
Table 4. Error Values: Single Loading Cases.
9.8 N19.6 N
Force Average Error3.3%4.9%
Zero Average Error3.5%3.1%
Table 5. Strain Readings: Adjacent Load Case (2 × 9.8 N).
Table 5. Strain Readings: Adjacent Load Case (2 × 9.8 N).
Point 1 (m)Point 2 (m)Strain ( μ ε )
1234567
0.070.13120158136109845833
0.130.191031872131731349353
0.190.258315722823518412872
0.250.316612418123523016291
0.310.374891133176217196112
0.370.43315986113141166128
Table 6. Force Values: Adjacent Load Case (2 × 9.8 N).
Table 6. Force Values: Adjacent Load Case (2 × 9.8 N).
Point 1 (m)Point 2 (m)Force (N)
1234567
0.070.1310.19.30.8−0.30.2−0.20.5
0.130.190.79.010.2−0.20.3−0.20.8
0.190.25−0.40.59.99.00.80.00.9
0.250.31−0.20.20.59.29.80.51.1
0.310.37−0.30.2−0.20.39.69.81.9
0.370.43−0.20.20.0−0.20.59.811.1
Table 7. Strain Readings: Skip Strain Gauge Load Case (2 × 9.8 N).
Table 7. Strain Readings: Skip Strain Gauge Load Case (2 × 9.8 N).
Point 1 (m)Point 2 (m)Strain ( μ ε )
1234567
0.070.191091441721411107643
0.130.259117119120516011163
0.190.317414020320520614682
0.250.3756108157205195178102
0.310.433775109144179151115
Table 8. Force Values: Skip Strain Gauge Load Case (2 × 9.8 N).
Table 8. Force Values: Skip Strain Gauge Load Case (2 × 9.8 N).
Point 1 (m)Point 2 (m)Force (N)
1234567
0.070.199.11.19.20.00.5−0.20.6
0.130.25−0.39.30.99.20.6−0.20.9
0.190.31−0.40.59.50.29.50.61.0
0.250.37−0.60.50.29.01.19.21.8
0.310.43−1.00.6−0.20.09.81.29.7
Table 9. Average Error Values: Multiple Loading Cases (Adjacent and Skip Strain Gauge).
Table 9. Average Error Values: Multiple Loading Cases (Adjacent and Skip Strain Gauge).
AdjacentSkip Strain Gauge
Force Average Error4.4%4.9%
Zero Average Error4.3%6.1%
Table 10. Strain Readings: In-Between Load 19.6 N Case.
Table 10. Strain Readings: In-Between Load 19.6 N Case.
LocationStrain ( μ ε )
1234567
3.258616423722017011966
3.508015622623818312871
3.757814821625419513675
4.257213119325722015483
4.506612518424123316288
4.756411717222524517093
Table 11. Force Values: In-Between Load 19.6 N Case.
Table 11. Force Values: In-Between Load 19.6 N Case.
LocationForce (N)
1234567
3.25−0.70.814.05.10.20.30.6
3.50−1.20.99.010.40.00.30.6
3.75−0.50.34.715.10.00.30.5
4.250.4−0.5−0.315.74.50.80.0
4.50−0.40.00.310.19.80.50.2
4.750.3−0.30.35.114.80.30.4
Table 12. Average Error Values: In-Between Load 44.5 N Case.
Table 12. Average Error Values: In-Between Load 44.5 N Case.
In-Between Strain Gauges
Force Average Error4.5%
Zero Average Error2.1%
Table 13. Applied Forces to Each Strain Gauge in the ’Simultaneous Varied Load Case’ Experiment.
Table 13. Applied Forces to Each Strain Gauge in the ’Simultaneous Varied Load Case’ Experiment.
ScenariosStrain Gauge
1234567
100019.6 N000
20014.7 N19.6 N14.7 N00
309.8 N14.7 N19.6 N14.7 N9.8 N0
44.9 N9.8 N14.7 N19.6 N14.7 N9.8 N4.9 N
Table 14. Strain Readings: Simultaneous Varied Load Case.
Table 14. Strain Readings: Simultaneous Varied Load Case.
Point (m)Strain ( μ ε )
1234
0.0774187268303
0.13140350490528
0.19207511655693
0.25265570717755
0.31207510657694
0.37144355501535
0.4380192276307
Table 15. Force Values: Simultaneous Varied Load Case.
Table 15. Force Values: Simultaneous Varied Load Case.
Point (m)Force (N)
1234
0.07−0.4−0.41.25.4
0.13−0.20.38.99.3
0.191.415.816.016.0
0.2518.018.518.919.1
0.310.814.814.915.2
0.370.21.210.710.7
0.430.70.21.85.5
Table 16. Applied Forces to Each Strain Gauge in the ’Simultaneous Uniform Load Case’ Experiment.
Table 16. Applied Forces to Each Strain Gauge in the ’Simultaneous Uniform Load Case’ Experiment.
ScenariosStrain Gauge
1234567
10009.8 N000
2009.8 N9.8 N9.8 N00
309.8 N9.8 N9.8 N9.8 N9.8 N0
49.8 N9.8 N9.8 N9.8 N9.8 N9.8 N9.8 N
Table 17. Strain Readings: Simultaneous Uniform Load Case.
Table 17. Strain Readings: Simultaneous Uniform Load Case.
Point (m)Strain ( μ ε )
1234
0.0737110186262
0.1370209347423
0.19103304446521
0.25133335479555
0.31104305448522
0.3772213355424
0.4340116196261
Table 18. Force Values: Simultaneous Uniform Load Case.
Table 18. Force Values: Simultaneous Uniform Load Case.
Point (m)Force (N)
1234
0.07−0.2−0.7−0.29.9
0.130.00.69.69.8
0.190.59.910.29.9
0.259.29.59.910.4
0.310.59.69.610.1
0.370.00.810.210.1
0.430.40.41.49.4
Table 19. Error Values: Simultaneous Load Case.
Table 19. Error Values: Simultaneous Load Case.
VariedUniform
Force Average Error6.5%2.8%
Zero Average Error10.9%4.8%
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Bednarz, E., III; Dietrich, C.; Hepner, B., Jr.; Patel, J.; Sabouni, A. Determining Magnitudes of Forces at Known Locations through a Strain Gauge Force Transducer. Sensors 2023, 23, 7017. https://doi.org/10.3390/s23167017

AMA Style

Bednarz E III, Dietrich C, Hepner B Jr., Patel J, Sabouni A. Determining Magnitudes of Forces at Known Locations through a Strain Gauge Force Transducer. Sensors. 2023; 23(16):7017. https://doi.org/10.3390/s23167017

Chicago/Turabian Style

Bednarz, Edward, III, Christian Dietrich, Brad Hepner, Jr., Jay Patel, and Abas Sabouni. 2023. "Determining Magnitudes of Forces at Known Locations through a Strain Gauge Force Transducer" Sensors 23, no. 16: 7017. https://doi.org/10.3390/s23167017

APA Style

Bednarz, E., III, Dietrich, C., Hepner, B., Jr., Patel, J., & Sabouni, A. (2023). Determining Magnitudes of Forces at Known Locations through a Strain Gauge Force Transducer. Sensors, 23(16), 7017. https://doi.org/10.3390/s23167017

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop