Peer-Review Record

Validation of Inertial-Measurement-Unit-Based Ex Vivo Knee Kinematics during a Loaded Squat before and after Reference-Frame-Orientation Optimisation

Sensors 2024, 24(11), 3324; https://doi.org/10.3390/s24113324
by Svenja Sagasser 1, Adrian Sauer 1,2,*, Christoph Thorwächter 2, Jana G. Weber 1, Allan Maas 1,2, Matthias Woiczinski 2,3, Thomas M. Grupp 1,2 and Ariana Ortigas-Vásquez 1,2
Reviewer 1:
Reviewer 2: Anonymous
Reviewer 3:
Sensors 2024, 24(11), 3324; https://doi.org/10.3390/s24113324
Submission received: 30 April 2024 / Revised: 14 May 2024 / Accepted: 21 May 2024 / Published: 23 May 2024
(This article belongs to the Special Issue Biosensors for Gait Measurements and Patient Rehabilitation)

Round 1

Reviewer 1 Report

Comments and Suggestions for Authors

The authors have established a need for this study. Background sufficient, explicit purpose, methods mostly complete, results and discussion well-presented, and conclusion relates to primary purpose and most important results. I have a few minor comments to address. See pdf

Comments for author File: Comments.pdf

Comments on the Quality of English Language

No major issues.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 2 Report

Comments and Suggestions for Authors

Dear Authors,

the paper is interesting and well written. I enclose the .pdf file with my minor comments inside.

Best regards

Comments for author File: Comments.pdf

Comments on the Quality of English Language

Minor editing of English language required

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 3 Report

Comments and Suggestions for Authors

This article presents a comparative analysis of joint motion capture effects between an inertial measurement unit and an optical motion capture system, based on knee joint cadaver specimens in specific environments. The study considers the significant influence of knee ligaments on joint motion. Test results demonstrate that local coordinate optimization significantly impacts the differences in results between different motion capture systems, with implications for further enhancing the accuracy of inertial dynamic capture systems.

In order to better improve the manuscript, the following suggestions are listed for reference:

1.The direction of applying a constant 20 N muscle force described  is not given in the text. Please add reference information.

2.The symbol  REFRAMEIMU-GOM should explain its subscript meaning when it first appears.

3.Please further analyze and explain the reason for the significant difference in IMU and GOM lines between the original knee joint angles of ex/internal rotation in Fig 1.

4.Additionally, frame transformations consisting of rotations around the femoral x-axis during optimisation were restricted to prevent non-physiological frame orientations. The limitations of the direction of the frame require further clarification.

5.Notably, the magnitude of errors affecting abduction/adduction angles were seemingly associated with joint flexion, possibly indicating the presence of crosstalk artefact. The relationship between crosstalk artefact and error should be further elaborated. Does it indicate that there may be problems with optimization based on optical measurements?

6. The reason for the significant difference between the IMU and GOM lines of the raw knee joint angle ab/adduction is due to crosstalk artefacts. Could the author further explain the data that affects the raw knee joint angle ab/adduction.

7.The content of Fig 5b is difficult to read and needs to be redrawn.

8. The manuscript takes into account the influence of soft tissues such as ligaments and uses a segment model with articular ligaments in cadaveric trials, but does not further discuss the influence of ligaments on motion capture. Is it the role of ligaments that causes complex motion patterns of the knee joint, resulting in measurement errors?

Author Response

Please see the attachment.

Author Response File: Author Response.pdf