Optical Sensing of Tissue Freezing Depth by Sapphire Cryo-Applicator and Steady-State Diffuse Reflectance Analysis

by
Arsen K. Zotov
1,2,
Aleksandr V. Pushkarev
3,4,
Anna I. Alekseeva
5,
Kirill I. Zaytsev
2,
Sergey S. Ryabikin
3,4,
Dmitry I. Tsiganov
3,4,
Dmitriy A. Zhidkov
3,4,
Ivan A. Burkov
3,
Vladimir N. Kurlov
1 and
Irina N. Dolganova
1,*
1
Osipyan Institute of Solid State Physics of the Russian Academy of Sciences, Chernogolovka 142432, Russia
2
Prokhorov General Physics Institute of the Russian Academy of Sciences, Moscow 119991, Russia
3
Bauman Moscow State Technical University, Moscow 105005, Russia
4
Federal State Budgetary Educational Institution of Further Professional Education “Russian Medical Academy of Continuous Professional Education”, Ministry of Healthcare of the Russian Federation, Moscow 125993, Russia
5
Avtsyn Research Institute of Human Morphology, FSBSI “Petrovsky National Research Centre of Surgery”, Moscow 117418, Russia
*
Author to whom correspondence should be addressed.
Sensors 2024, 24(11), 3655; https://doi.org/10.3390/s24113655
Submission received: 12 April 2024 / Revised: 31 May 2024 / Accepted: 2 June 2024 / Published: 5 June 2024
(This article belongs to the Section Biomedical Sensors)

Abstract

:
This work describes a sapphire cryo-applicator with the ability to sense tissue freezing depth during cryosurgery by illumination of tissue and analyzing diffuse optical signals in a steady-state regime. The applicator was manufactured by the crystal growth technique and has several spatially resolved internal channels for accommodating optical fibers. The method of reconstructing freezing depth proposed in this work requires one illumination and two detection channels. The analysis of the detected intensities yields the estimation of the time evolution of the effective attenuation coefficient, which is compared with the theoretically calculated values obtained for a number of combinations of tissue parameters. The experimental test of the proposed applicator and approach for freezing depth reconstruction was performed using gelatin-based tissue phantom and rat liver tissue in vivo. It revealed the ability to estimate depth up to 8 mm. The in vivo study confirmed the feasibility of the applicator to sense the freezing depth of living tissues despite the possible diversity of their optical parameters. The results justify the potential of the described design of a sapphire instrument for cryosurgery.

1. Introduction

Cryosurgery and cryotherapy are associated with the application of low temperatures to living tissues and cold-induced effects, such as necrosis, osmotic injury, tissue hypoxia, cell apoptosis, as well as hemostatic and immunological effects [1,2,3,4]. Cryosurgery finds its applications in oncology, dermatology, gynecology, urology and other fields for the surface or interstitial non-surgical removal of various lesions and neoplasms [5,6,7,8,9,10,11]. The advantages of cryosurgery also include minimal invasiveness, relative painlessness and short recovery time for patients [12,13].
Since cryosurgery is aimed at cold-induced tissue ablation during freezing, it is connected with the risk of damaging the surrounding healthy tissues as well as the incomplete freezing of the targeted tissue volume [14]. Several approaches have been applied to addressing this problem. The first one deals with the accurate planning of the procedure, numerical simulations and preliminary estimations [15,16,17,18]. The evident drawback of this approach is the absence of real-time monitoring of tissue freezing. Therefore, it is desirable to use other techniques for monitoring, such as ultrasound (US) [19,20] and magnetic resonance imaging (MRI) [21,22,23], computed tomography (CT) [24,25] and temperature measurements [26]. Visualization by US, CT and MRI requires additional sensors or even expensive systems that can not be combined with the cryosurgical instrument. Moreover, MRI needs non-magnetic cryosurgical equipment and CT images can suffer from the artifacts caused by the usage of common metal cryo-applicators, which hinder the visualization of the so-called ice ball formation in tissue. Temperature measurements enable estimation of tissue condition only at several points or on its surface. Among other approaches, we should also mention the application of cryo-protective agents, which help to manage, to some extent, the freezing volume and protect, for example, adjacent nerves [27].
To solve the stated problem of monitoring the cryosurgical process, it is attractive to use optical sensors and methods, such as analysis of the opto-acoustic signal [28,29,30], terahertz monitoring [31], and diffuse reflectance analysis [32,33]. The latter can be implemented with optical fibers and, thus, be considered as a part of a cryosurgical instrument. We have recently suggested the concept of a sapphire applicator for cryosurgery with the ability to detect tissue freezing depth using the analysis of spatially resolved diffuse reflected intensities [34]. Sapphire provides biocompatibility and chemical inertness together with high performance of tissue freezing, since it features high thermal conductivity at cryogenic temperatures [35,36]. It has been demonstrated that a submerged sapphire applicator enables faster tissue freezing and lower temperatures in comparison with metal ones [37]. Moreover, the application of crystal growth techniques [38,39,40] allows for the manufacturing of sapphire shaped crystals with low amounts of surface and volume defects, which, together with the high transparency of sapphire in visible and near-infrared (NIR) ranges, gives the opportunity to deliver light to tissues through the contact part of the applicator. Thus, optical methods of tissue monitoring during cryosurgery can be implemented directly in the sapphire instrument.
The principle of freezing-depth sensing by a sapphire applicator was described in our previous work [34]. Using spatially resolved optical fibers accommodated inside the crystal and connected to the light source and the detector, the diffuse reflected intensities are analyzed during the formation of an ice ball inside the tissue. The reconstruction of the ice ball height, i.e., the freezing depth, is implemented based on the principles of light diffusion in a turbid medium and frequency modulation of the light source [41,42,43,44,45]. However, this applicator needs rather complicated optical equipment and analysis of signals in the frequency domain. Therefore, in the present work, we propose another implementation of the mentioned optical sensing approach, and we studied the possibility of using steady-state analysis of diffuse reflected intensities for freezing-depth reconstruction. We used a multi-channeled sapphire applicator manufactured by the crystal growth technique. One channel serves for the illumination of the sample, while two other channels serve for detection of the diffuse signal. The experimental approbation of the suggested approach was performed with a tissue-mimicking phantom and in vivo rat liver tissues. The results show the advantages and drawbacks of this applicator and approach. Thus, this work demonstrates the feasibility of the optical sensing of tissue freezing by means of a sapphire cryo-applicator and the analysis of spatially resolved optical responses in steady-state regimes.

2. Materials and Methods

2.1. Multi-Channeled Sapphire Applicator

A sapphire applicator with an outer diameter of 12 mm and a length of 149 mm is shown in Figure 1. It was manufactured by the edge-defined film-fed growth (EFG) technique [39] and has internal channels with a diameter of 2 mm for accommodating optical fibers inside the applicator. EFG and related techniques help to obtain shaped crystals with as-grown quality of channel surface without additional mechanical processing. The manufacturing of a crystal by EFG technique requires a growth chamber with a high-purity Ar atmosphere as an ambient under a pressure of 1.1–1.3 atm, a molybdenum crucible with a 22 kHz induction-heated graphite susceptor and a molybdenum die with capillary channels. The die determines the crystal’s cross-section. To maintain the growth process, the applied set-up is equipped by an in-house automated weight control system. The considered crystal shape and the channel size, which is larger than the common fiber diameter, allow for avoiding the appearance of crystal defects and gas inclusions during the growth process. On the other hand, the decreasing of the channel diameter complicates the efforts for preventing the channel collapse by maintaining the heating zone parameters.
To perform optical monitoring of the sample by the sapphire applicator, optical fibers (with diameter of 200 μm, NA = 0.22 ) are accommodated and fixed by epoxy adhesive inside the three channels of the crystal. The first fiber is used for tissue illumination; it is connected to a laser diode (M530F2, Thorlabs, Inc., Newton, NJ, USA) with a central wavelength of 530 nm and minimal output power of 6.8 mW. For this study, a green light source was chosen for implementing the feasibility test of sensing the tissue phantom, which features low absorption, and approving the tested applicator configuration by using several tissues in vivo. However, for further studies, tissue absorption properties should be taken into account and a red/NIR light source is preferable. Nevertheless, the applied source wavelength does not change the method of reconstructing tissue freezing depth, described below. The other two fibers are used for the detection of the diffused optical response; each one being connected to a separate spectrometer (CCS100/M, Thorlabs, Inc., Newton, NJ, USA). The signals are registered within a 3 s period during the application. To protect the fibers from direct contact with the tissue, a 2 mm thick sapphire window is bounded to the bottom end of the applicator by an adhesive layer with a thickness of 200 μm. The remaining channels are assumed to be used for possible additional monitoring or exposure, but they are not involved in the present implementation of optical monitoring. The sapphire window also helps to homogenize the cooling profile over the application time, which is slightly changed by the presence of the channels.
The sapphire applicator is mounted in a thermally insulated tank with an outer diameter of 100 mm (see Figure 1d). The thickness of the thermal insulation is 15 mm. In the bottom, the tank has a threaded fastener for fixation of the applicator. In cases of using liquid samples, they are stored in the container with 90 × 90 × 110 mm dimensions.

2.2. Reconstruction of a Freezing Depth

During the sample freezing, diffuse reflected intensity I ( R , L ) is recorded at each moment t; here, R is the distance between the illumination and detection channels (or the source–detector separation) and L is the freezing depth (see Figure 1b). Using the applicator described above, the distance R is equal to 4 and 8 mm for two detection channels. The signals are then smoothed by moving average filtering.
For a semi-infinite turbid medium, according to the diffuse approximation of the radiation transfer equation [46,47], the diffuse reflectance can be estimated as
I ( R ) = C 1 R m exp ( C 2 R ) ,
where C 1 and C 2 are empirically determined coefficients depending on optical parameters of the turbid medium and parameter m depends on the distance R. For the considered design of the applicator, we set m = 1.5 .
An effective attenuation coefficient μ eff of the medium describes its absorption and scattering properties:
μ eff = [ 3 μ a ( 3 μ a + μ s ) ] 1 / 2 ,
where μ a and μ s are the absorption and reduced scattering coefficients, respectively. From Equation (1), μ eff can be estimated as μ eff C 2 . Consequently, using two detection channels and measuring I ( R 1 ) and I ( R 2 ) , the effective attenuation coefficient is calculated from Equation (1):
μ eff ln [ I ( R 1 ) R 1 m ] ln [ I ( R 2 ) R 2 m ] R 2 R 1 .
In order to evaluate correctly μ eff from the measured signals, the applicator needs calibration using a sample with known optical parameters. The calibration should be performed with a preliminary cooled applicator, and only the initial values of I calib ( R , L = 0 ) —which are obtained before the freezing layer begins to appear—should be applied in the calculations. We compare the slope ln ( I R m ) obtained for the actual measurement of the calibration phantom with the slope of a fit line [ ln C 1 C 2 R ] , where C 2 is estimated from Equation (2). As a result, the required corrections for the measurements are determined.
For reconstruction of freezing depth, we assume the plane-parallel structure of the tissue-mimicking phantom, with an upper layer of frozen medium of thickness L and a bottom semi-infinite layer of unfrozen medium. Theoretically, the diffuse reflectivity from the two-layered turbid medium is described by the following formulas, assuming the extrapolated boundary condition [44]:
I th ( R , L ) = D 1 z Φ ^ ( R , z , L ) | z = 0
Φ ^ ( R , z , L ) = 1 π a 2 n = 1 G ( s n , z , L ) J 0 ( s n R ) J 1 2 ( a s n ) ,
G ( s n , z , L ) = exp ( α 1 | z z 0 | ) exp [ α 1 ( z + z 0 + 2 z b ) ] 2 D 1 α 1 + sinh [ α 1 ( z 0 + z b ) ] sinh [ α 1 ( z + z b ) ] D 1 α 1 exp [ α 1 ( L 1 + z b ) ] × D 1 α 1 D 2 α 2 D 1 α 1 cosh [ α 1 ( L 1 z b ) ] + D 2 α 2 sinh [ α 1 ( L 1 z b ) ] ,
α k = μ a , k / D k + s n 2 ,
where Φ ^ ( R , z , L ) is the photon fluence in the upper layer at an arbitrary position R relative to the source position, z is the sample depth direction, L is the thickness of the upper frozen layer, D k = 1 / ( 3 μ s , k ) is the diffusion coefficient of the k-layer (1, 2 stand for upper frozen and bottom unfrozen layers, respectively), μ s , k and μ a , k are the reduced scattering and absorption coefficients, respectively, of the k-layer, G ( s n , z , L ) is the Green’s function, J p is the Bessel function of the first kind and order p, a = a + z b , a is the characteristic radius of the sample cylinder, z b describes the position of the extrapolated boundary and depends on the refractive indices and scattering coefficients of two layers [41], z 0 = 1 / μ s , 1 and s n is the positive root of the equation J 0 ( a s n ) = 0 . Thus, after numerical calculation of the photon fluence in the upper layer, finding its first derivative on the sample surface, the diffuse reflectivity I th ( R , L ) can be obtained for R 1 and R 2 , which yields the theoretical estimation of the effective attenuation μ eff , th by means of Equation (3) for arbitrary thickness L. In these calculations, we assume that the sample optical parameters may vary in some ranges. Therefore, we obtain I th ( R , L ) and μ eff , th = μ eff , th ( L , n 1 , n 2 , μ a , 1 , μ a , 2 , μ s , 1 , μ s , 2 ) for this set of parameters. The μ eff , th is then compared with μ eff , meas ( t ) obtained from the actually measured I meas ( R , t ) . Minimization of error at every particular moment t gives the freezing-depth dependence L ( t ) :
L ( t ) = argmin L μ eff , meas ( t ) μ eff , th ( L , n 1 , n 2 , μ a , 1 , μ a , 2 , μ s , 1 , μ s , 2 ) .
The condition L i + 1 > L i for the t i + 1 > t i moment is applied.

3. Results

3.1. Calibration

For the calibration of the applicator, we used an aqueous 3% solution of lipid emulsion (Lipoplus 20, B. Brown, Melsungen, Germany). It has thermal [48,49] and optical [50,51,52] properties similar to biotissues. The phantom temperature was ambient (24 °C). The results of the calibration are shown in Figure 2. Here, panel (a) demonstrates the measured intensities. By means of Equations (1) and (2) we calculated the fitted slope (orange line in Figure 2b) and by means of Equations (4)–(7) we calculated I th ( R ) for the calibration phantom, assuming that it consisted of a single unfrozen layer. In Figure 2b, we show the corresponding slope ln [ I th ( R ) R m ] . The measurement, fit and theoretical slopes were vertically shifted to zero at R 1 . It is evident that ln [ I th ( R ) R m ] almost matches the fit line, which justifies the algorithm for theoretical calculation. On the other hand, the noticed significant difference between the actually measured and the theoretical slopes can be explained by the uneven sensitivity of the spectrometers, fiber losses and possible reflections of the light beam in the sapphire window. The calibration of the probe helped to eliminate the measurement error caused by these reasons.

3.2. Sensing of Tissue Phantom Freezing

For the experimental approbation of the sapphire applicator, the tissue-mimicking phantom was applied. It contained gelatin with 95 ± 1 % moisture content and Lipoplus 20 of 1% volume concentration. The phantom was made directly in the sample container (Figure 1d), which eliminated the presence of air bubbles. Its initial temperature was ambient (24 °C). The preliminary cooled applicator touched the surface of the phantom to start the formation of an ice ball. During this process, the optical signals were stored. They are shown in Figure 3a. After freezing, the phantom was thawed at room temperature. The whole experiment was repeated three times.
For measuring the ice ball dimensions, the growth process was recorded by a digital camera (Sony a7 mark II with lens Art 24-70 f/2.8 FE). We should note that to visualize the ice ball growth, the phantom should be transparent in the visible range; therefore, it should not contain lipid emulsion for this measurement. Thus, the dimensions were recorded for a pure gelatin medium. We assumed that the ice ball formation was mainly governed by the gel properties; thus, the absence of emulsion caused negligible errors in measuring the ice ball size in the phantom.
Considering the dispersion of gelatin properties, as well as a slight sedimentation of lipid inclusions in warm gelatin during solidification, we set the variance of the phantom parameters, in order to calculate I th ( R , L ) . Since the scattering and absorption properties of the phantom were mostly determined by the lipid emulsion, which has optical properties similar to those of intralipid [53,54], we assumed the following parameters for the unfrozen layer: refractive index n 2 = [ 1.40 1.42 ] , reduced scattering coefficient μ s , 2 = [ 0.50 0.85 ] mm 1 , absorption coefficient μ a , 2 = [ 0.03 0.05 ] mm 1 . These data were in accordance with the previously measured properties of such phantoms [52,53]. Then, we assumed that the frozen layer was characterized by the slight alteration of the refractive index n 1 = n 2 Δ n , Δ n = [ 0.02 0.04 ] , and that the absorption coefficient μ a , 1 = μ a , 2 Δ μ a , Δ μ a = [ 0.008 0.01 ] mm 1 . At the same time, the reduced scattering coefficient increased as μ s , 1 = μ s , 2 + Δ μ s , where Δ μ s = [ 0.5 1.0 ] mm 1 . The introduced ranges for the properties of the frozen phantom were adjusted numerically. The expanding of these boundaries yielded the same results of depth reconstruction, but increased the computation time. Figure 3b demonstrates the intensity I th ( R , L ) obtained for certain sample properties from the described ranges. In contrast to the behavior of I meas , we can notice uneven alteration of I th at small freezing depth L. The obtained I th decreases to a certain value before starting to increase, which is explained by the usage of extrapolated boundary conditions. While the measured intensities yielded the reduction of μ eff , meas during the ice ball formation (the normalized μ eff , meas is shown in Figure 3c), the reduction of μ eff , th began after a certain depth L (Figure 3d). Thus, we should begin to compare the measured and calculated effective attenuation coefficient and the reconstructed L by Equation (8), assuming some delay from the beginning of freezing (see Figure 3f). This delay is indicated by a blue region in panel (f), which corresponds to a certain depth range, shown by the same color in panels (b) and (d).
Figure 4 shows the reconstructed freezing depth and its comparison with the visualized ice ball height. The proposed approach ensured rather good agreement between two values within a time range from 10 to 50 s. The error of the depth estimation was <1.5 mm. It increased dramatically after 50 s and L = 8 mm. This meant that the size and curvature of the ice ball did not allow for monitoring of the freezing depth by the applicator with the considered positions of the channels. To increase the upper limit of depth reconstruction, one should choose a larger distance between the illumination and detection channels along with the account of the ice ball curvature instead of using plane-parallel approximation, which will lie within the scope of our further studies. However, a large source-detector separation is appropriate only for superficial applicators. Otherwise, the illumination of tissue should be performed by a separate tool, which will complicate the overall instrumentation. At the same time, for this particular sample medium, reconstruction of the freezing depth is impossible for t < 10 s, due to the reason described above.
In Figure 4b, the results of the performed depth sensing are compared with those obtained by the use of frequency-domain solution and four detection channels [34]. Despite applying only two detection channels and a steady-state solution, the errors of the two approaches are almost comparable, but slightly increased for the present applicator. Moreover, the range of freezing depth that can be reconstructed is the same.
We should note that possible errors of reconstruction can be caused by the freezing and thawing of the phantom, which contains lipid emulsion. In order to fix the optical parameters of the phantom medium during several experiments, the measurements were performed with the same phantom volume. Therefore, freezing and thawing of lipid emulsion may still have caused change in the phantom properties, similar to the changes of real tissue properties. Indirectly, this was confirmed by a rather high variance of the measured I ( t ) (Figure 3a). This fact leads to rather high variance of depth reconstruction and imposes the necessity of using a rather wide range of sample initial parameters for theoretical calculations. To overcome this problem, it is possible to use additional methods of measuring initial sample parameters, which would help to strength the results of reconstruction and reduce the calculation efforts.

3.3. Sensing of In Vivo Rat Liver Freezing

To continue the experimental approbation of the proposed sapphire cryo-applicator, we studied the cryodestruction of in vivo liver tissue. For this aim, four Wistar rats (females) weighing 280 380 g were used. The animals were placed under deep anesthesia using Zoletil and xylazine (Interchemie, Püünsi, Estonia)/propofol (Fresenius Kabi, Graz, Austria). After treating the wound surface with an antiseptic, an incision about 4–5 cm long was made to the right of the white midline of the abdomen in the hypochondrium. The skin and muscle wall were incised. Next, several lobes of the liver were isolated (raised to the surface) for immediate cryo-application (Figure 5a). The applicator was cooled down before the experiment. The application was carried out for 5–20 s, namely 5 s (sample 1), 15 s (samples 2 and 3) and 20 s (sample 4). Due to the rather small thickness of the liver lobes, longer exposure could induce total freezing. After thawing, the liver lobe was carefully removed back into the abdominal cavity and sutured with silk threads in layers.
The rats were removed from the experiment in the CO 2 camera 24 h after the cryo-application. After measuring the dimensions of the damaged tissue area, liver specimens were taken for characterization to confirm the destruction of the tissue. They were fixed in formaldehyde solution (4% paraformaldehyde solution, pH = 7.4 ) for 48 h, immersed in ethanol–xylene solutions and embedded in paraffin blocks. Then, sections with a thickness of 4–7 μm were made and were stained with hematoxylin and eosin (H&E). Histological images of the specimens, were obtained at magnification ×200. Figure 5b shows a representative H&E-stained image of the rat liver tissue after cold-induced damage. The border between healthy (on the left) and damaged (on the right) tissue regions was smooth; the damaged region was characterized by either moderate or severe vasodilation and edema. As one can notice, the damaged tissue may feature uneven consistency. Consequently, the parameters of the probed tissue region may change during the application, as the thickness of the frozen layer increases. In order to take this variation into account during obtaining μ eff , th , we assumed that the equivalent homogeneous damaged tissue could be considered, which was characterized by the averaged parameters within the specific ranges.
Figure 6 demonstrates the results of this experiment. In Figure 6a, we show a representative time evolution of I meas for one sample. However, the measured intensities varied significantly from sample to sample. This is explained by the different thicknesses of the liver lobes as well as by the variance of optical properties for different animals. As described in Section 2.2, from I meas ( R , t ) , we could estimate the effective attenuation coefficient (Figure 6c). Since the optical signals were measured within a period of 3 s, μ eff , meas ( t ) was characterized by only a few points.
To obtain I th ( R , L ) , the following tissue parameters were used. The unfrozen layer was characterized by n 2 = [ 1.37 1.40 ] , μ s , 2 = [ 2.70 2.8 ] mm 1 and μ a , 2 = [ 1.00 1.15 ] mm 1 , while for the frozen layer we assumed n 1 = n 2 Δ n , Δ n = [ 0.02 0.1 ]   μ s , 1 = μ s , 2 + Δ μ s , Δ μ s = [ 0.2 1.0 ] mm 1 and μ a , 1 = μ a , 2 Δ μ a , Δ μ a = [ 0.008 0.05 ] mm 1 . Expanding the boundaries of the selected ranges brought no benefits in the results, but increased the computation time. Figure 6b demonstrates the calculated intensities for particular sample parameters. It is clear that the region of uncertainty was much smaller for the liver tissue than for the gelatin-based phantom, and that it was below 0.5 mm. This allowed for freezing-depth sensing during a rather short cryo-application. In Figure 6c, the red circles indicate the estimated μ eff , th ( t ) for sample 4, demonstrating good agreement with the measured values.
The reconstructed freezing depth is shown in Figure 6d. We can see a rather different character of changing L for all the samples. This was associated with a number of factors. The main reason for such a strong variation was the different sizes of the liver lobes in the rats, which impacted on the heat transfer in their tissue. The differences of perfusion, initial optical properties and the state of a rat under anesthesia may also change the freezing rate. The observed results highlight the need for in situ monitoring of the damaged region. However, the final values of the estimated L are in good agreement with the visualized values. Note that due to the applied set-up, the measurement of I within a period of 3 s did not allow us to obtain signals exactly at 5 and 20 s. The results confirm the feasibility of the sapphire applicator to be applied for the sensing of freezing depth in vivo, despite the alteration of the sample properties.

4. Discussion

In this work, we demonstrate the concept of the sapphire cryo-applicator, which provides the ability to estimate tissue freezing depth using optical monitoring based on the steady-state analysis of diffuse reflected intensities. This allows for the monitoring of tissue freezing by means of rather simple optical equipment and three fiber channels in contrast to the previously proposed instrumentation that uses signal analysis in the frequency domain. The limits of possible depth reconstruction depend on the sample properties and separation between the applicator channels R. Our experimental approbation with tissue phantom demonstrated the upper limit of depth reconstruction, which is ∼8 mm, and the error of depth estimation < 1.5 mm. This error can be slightly improved by using the red or NIR light source and consequently increasing the penetration depth. It is also a significant drawback that it is difficult to estimate L at the beginning of sample freezing. However, the in vivo studies of the proposed applicator show the ability to reduce this blind area when scattering and absorption coefficients significantly increase. At the same time, from the theoretically estimated I th ( R , L ) , one can notice that this area depends on R. Thus, using smaller separation between the illumination and detection channels in the applicator, the bottom limit of the depth reconstruction can be also changed. It is clear that by increasing R one can broaden the upper limit for the L estimation. Therefore, adjusting the positions of the applicator’s channels, it is possible to choose the depth range that should be monitored. However, at high L, it is important to account for ice ball curvature. Thus, finding the trade-off between the depth and the reconstruction accuracy requires a separate study.
It is straightforward to roughly estimate the freezing depth using the behavior of μ eff ; in particular, its rate of change μ ˙ eff . From Figure 3e, we can see that it saturates with time. Therefore, the saturation of μ ˙ eff can be applied as the marker of reaching a certain depth by the freezing front. At the same time, it is possible to involve the first derivative in the reconstruction procedure in order to increase the accuracy of Equation (8). We leave this for our further studies, in order to solve the task of improving L reconstruction.
Despite the cost of growing a sapphire crystal being higher than the cost of manufacturing a metal applicator, the higher freezing rate, the absence of tissue sticking to the contact surface and ability to withstand applications without surface degradation [37] make the usage of sapphire for cryosurgery cost-effective. We should highlight the use of the EFG technique for sapphire applicator manufacturing. It gives opportunities for choosing the appropriate crystal shape, changing the number of channels and their positions. It is important for transferring the proposed concept to clinical usage, assuming a significant variety of localizations and sizes of tissue lesions, as well as their possible heterogeneity. However, to study the performance of the sapphire applicator in these various conditions, experiments with more complex phantoms and tissue samples should be conducted.
The proposed concept of the sapphire applicator demonstrates its prospects for cryosurgery; however, it could find applications in other tasks, such as monitoring of tissue state and condition, for example, and detection of hematoma or tissue ishemia.

5. Conclusions

This work is aimed at solving the problem of monitoring tissue freezing depth during cryosurgery. For this, the concept of the sapphire cryo-applicator, which enables the analysis of a diffuse reflected optical response in a steady-state regime, is proposed. The advantages of this sapphire applicator are the minimal number of internal channels used for the illumination of a sample and the detection of optical radiation and the absence of frequency modulation of the light source. Our feasibility test of this applicator, made with a tissue-mimicking phantom, demonstrated its ability to sense tissue freezing depth up to 8 mm. The in vivo approbation of the sapphire applicator demonstrated the ability to detect the freezing depth despite the significant alteration of tissue properties. The results of this work, being a continuation of our previous studies, pave the way for further investigations of the sapphire cryo-applicator along with the development of other sapphire sensors for biophotonics and medicine.

Author Contributions

Conceptualization, I.N.D.; methodology, I.N.D. and A.V.P.; software, A.K.Z. and S.S.R.; validation, A.I.A., I.A.B. and K.I.Z.; investigation, I.N.D.; writing—original draft preparation, A.K.Z., A.V.P. and I.N.D.; writing—reviewing and editing, V.N.K.; visualization, D.A.Z. and A.K.Z.; supervision, V.N.K. and D.I.T. All authors have read and agreed to the published version of the manuscript.

Funding

This work was supported by a grant from the Russian Science Foundation (RSF), Project # 19-79-10212.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki, and approved by the Bioethical Commission of the FSBI Avtsyn Research Institute of Human Morphology (protocol No. 27(3) from 11 October 2021).

Informed Consent Statement

Not applicable.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request, due to privacy.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Baust, J.G.; Gage, A.A. The molecular basis of cryosurgery. BJU Int. 2005, 95, 1187–1191. [Google Scholar] [CrossRef]
  2. Gage, A.A.; Baust, J. Mechanisms of Tissue Injury in Cryosurgery. Cryobiology 1998, 37, 171–186. [Google Scholar] [CrossRef]
  3. Gage, A.; Baust, J.; Baust, J. Experimental cryosurgery investigations in vivo. Cryobiology 2009, 59, 229–243. [Google Scholar] [CrossRef]
  4. Korpan, N.N. (Ed.) Basics of Cryosurgery; Springer: Vienna, Austria, 2001. [Google Scholar]
  5. Mokbel, K.; Kodresko, A.; Ghazal, H.; Mokbel, R.; Trembley, J.; Jouhara, H. The Evolving Role of Cryosurgery in Breast Cancer Management: A Comprehensive Review. Cancers 2023, 15, 4272. [Google Scholar] [CrossRef]
  6. Backman, E.; Polesie, S.; Berglund, S.; Gillstedt, M.; Sjöholm, A.; Modin, M.; Paoli, J. Curettage vs. cryosurgery for superficial basal cell carcinoma: A prospective, randomised and controlled trial. J. Eur. Acad. Dermatol. Venereol. 2022, 36, 1758–1765. [Google Scholar] [CrossRef]
  7. Pustinsky, I.; Dvornikov, A.; Kiva, E.; Chulkova, S.; Egorova, A.; Gladilina, I.; Peterson, S.; Lepkova, N.; Grishchenko, N.; Galaeva, Z.; et al. Cryosurgery for Basal Cell Skin Cancer of the Head: 15 Years of Experience. Life 2023, 13, 2231. [Google Scholar] [CrossRef]
  8. Kwong, A.; Co, M.; Fukuma, E. Prospective Clinical Trial on Expanding Indications for Cryosurgery for Early Breast Cancers. Clin. Breast Cancer 2023, 23, 363–368. [Google Scholar] [CrossRef]
  9. Zheng, W.; Liu, Y.; Tseng, Y.; Zhang, J.; Li, W.; Wang, B.; Pan, Y.; Zhu, J.; Luo, Z.; Luo, F.; et al. A novel endoscopic visible full-thickness cryoablation device on stomach. Sci. Rep. 2020, 10, 5203. [Google Scholar] [CrossRef]
  10. Tan, W.P.; Chang, A.; Sze, C.; Polascik, T.J. Oncological and Functional Outcomes of Patients Undergoing Individualized Partial Gland Cryoablation of the Prostate: A Single-Institution Experience. J. Endourol. 2021, 35, 1290–1299. [Google Scholar] [CrossRef]
  11. Cohen, J.K.; Miller, R.J.; Ahmed, S.; Lotz, M.J.; Baust, J. Ten-Year Biochemical Disease Control for Patients with Prostate Cancer Treated with Cryosurgery as Primary Therapy. Urology 2008, 71, 515–518. [Google Scholar] [CrossRef]
  12. Vansonnenberg, E.; McMullen, W.; Solbiati, L.; Livraghi, T.; Müeller, P.; Silverman, S. Tumor Ablation: Principles and Practice; Springer: New York, NY, USA, 2005. [Google Scholar] [CrossRef]
  13. Chu, K.F.; Dupuy, D.E. Thermal ablation of tumours: Biological mechanisms and advances in therapy. Nat. Rev. Cancer 2014, 14, 199–208. [Google Scholar] [CrossRef]
  14. Shurrab, M.; Wang, H.; Kubo, N.; Fukunaga, T.; Kurata, K.; Takamatsu, H. The cooling performance of a cryoprobe: Establishing guidelines for the safety margins in cryosurgery. Int. J. Refrig. 2016, 67, 308–318. [Google Scholar] [CrossRef]
  15. Burkov, I.; Pushkarev, A.; Ryabikin, S.; Shakurov, A.; Tsiganov, D.; Zherdev, A. Numerical simulation of controlled precision cryosurgery using argon Joule–Thomson and liquid nitrogen evaporation cryoprobes. Int. J. Refrig. 2022, 133, 30–40. [Google Scholar] [CrossRef]
  16. Burkov, I.; Pushkarev, A.; Shakurov, A.; Tsiganov, D.; Zherdev, A. Numerical simulation of multiprobe cryoablation synergy using heat source boundary. Int. J. Heat Mass Transf. 2020, 147, 118946. [Google Scholar] [CrossRef]
  17. Pushkarev, A.; Ryabikin, S.; Saakyan, N.; Tsiganov, D.; Burkov, I.; Vasilev, A. A study of prostate multiprobe cryoablation near urethra for precision treatment planning. Cryobiology 2022, 109, 10–19. [Google Scholar] [CrossRef]
  18. Hossain, S.; Hossain, S.M.C.; Ferdows, M.; Bangalee, M.Z.I.; Alam, M.S. Computer simulation based analysis of transient two-phase thermal evolutions during multiprobe cryosurgery of hepatic tissues embedded without or with blood vessels. Int. J. Thermofluids 2024, 21, 100587. [Google Scholar] [CrossRef]
  19. Ahmed, M.; Weinstein, J.; Hussain, J.; Sarwar, A.; Anderson, M.; Dillon, B. Percutaneous Ultrasound-Guided Cryoablation for Symptomatic Plantar Fibromas. CardioVascular Interv. Radiol. 2018, 41, 298–304. [Google Scholar] [CrossRef]
  20. Ward, R.C.; Lourenco, A.P.; Mainiero, M.B. Ultrasound-Guided Breast Cancer Cryoablation. Am. J. Roentgenol. 2019, 213, 716–722. [Google Scholar] [CrossRef]
  21. de Jager, N.S.; van Oostenbrugge, T.J.; Pätz, T.; Jenniskens, S.F.M.; Fütterer, J.J.; Langenhuijsen, J.F.; Overduin, C.G. Intraoperative MRI-derived volumetric ablation margins and initial correlation with local outcome after MRI-guided cryoablation of renal tumors. Cancer Imaging 2023, 23, 31. [Google Scholar] [CrossRef]
  22. Tokuda, J.; Chauvin, L.; Ninni, B.; Kato, T.; King, F.; Tuncali, K.; Hata, N. Motion compensation for MRI-compatible patient-mounted needle guide device: Estimation of targeting accuracy in MRI-guided kidney cryoablations. Phys. Med. Biol. 2018, 63, 085010. [Google Scholar] [CrossRef]
  23. Zhou, K.; Zhang, Z.; Figini, M.; Shangguan, J.; Wang, Y.; Shi, H.; Li, Y.; Jin, Z.; Pan, J. Combined magnetic resonance imaging and optical surgical navigation system guidance of percutaneous liver cryoablation in a porcine model. Am. J. Transl. Res. 2018, 10, 1531–1538. [Google Scholar]
  24. Yang, Y.; Li, Y.; Wu, Y.; Qiu, S.; Liu, C.; Wang, Q.; Hong, Y.; Lyu, J.; Zhang, Y.; Du, D. Retrospective analysis of CT-guided percutaneous cryoablation for treatment of painful osteolytic bone metastasis. Cryobiology 2020, 92, 203–207. [Google Scholar] [CrossRef]
  25. Jankovic, I.; Poulsen, F.R.; Pedersen, C.B.; Kristensen, B.W.; Schytte, T.; Andersen, T.L.; Langhorn, L.; Graumann, O.; Krone, W.; Høilund-Carlsen, P.F.; et al. Preclinical cerebral cryoablation in non-tumor bearing pigs. Sci. Rep. 2022, 12, 1977. [Google Scholar] [CrossRef]
  26. Tokiwa, T.; Zimin, L.; Ishiguro, H.; Inoue, T.; Kajigaya, H.; Nomura, S.; Suzuki, M.; Yamakawa, T. A Palm-Sized Cryoprobe System With a Built-In Thermocouple and Its Application in an Animal Model of Epilepsy. IEEE Trans. Biomed. Eng. 2019, 66, 3168–3175. [Google Scholar] [CrossRef]
  27. Ranjbartehrani, P.; Ramirez, D.A.; Schmidlin, F.; Etheridge, M.; Iaizzo, P.A.; Shao, Q.; Bischof, J.C. Nerve Protection During Prostate Cryosurgery. Ann. Biomed. Eng. 2023, 51, 538–549. [Google Scholar] [CrossRef]
  28. Larin, K.; Larina, I.; Motamedi, M.; Esenaliev, R. Optoacoustic laser monitoring of cooling and freezing of tissues. Quantum Electron. 2002, 32, 953–958. [Google Scholar] [CrossRef]
  29. Esenaliev, R.O. Optoacoustic diagnostic modality: From idea to clinical studies with highly compact laser diode-based systems. J. Biomed. Opt. 2017, 22, 091512. [Google Scholar] [CrossRef]
  30. Petrova, E.V.; Brecht, H.P.; Motamedi, M.; Oraevsky, A.A.; Ermilov, S.A. In vivo optoacoustic temperature imaging for image-guided cryotherapy of prostate cancer. Phys. Med. Biol. 2018, 63, 064002. [Google Scholar] [CrossRef]
  31. Zotov, A.K.; Gavdush, A.A.; Katyba, G.M.; Safonova, L.P.; Chernomyrdin, N.V.; Dolganova, I.N. In situ terahertz monitoring of an ice ball formation during tissue cryosurgery: A feasibility test. J. Biomed. Opt. 2021, 26, 043003. [Google Scholar] [CrossRef]
  32. Ikiades, A.; Bassukas, I.D.; Kourkoumelis, N. A Fiber Optic Sensor for Monitoring the Spectral Alterations and Depth in Ex Vivo and In Vivo Cryosurgery. Sensors 2023, 23, 2690. [Google Scholar] [CrossRef]
  33. Spasopoulos, D.; Rattas, G.; Kaisas, A.; Dalagiannis, T.; Bassukas, I.D.; Kourkoumelis, N.; Ikiades, A. Fiber Optic Sensor for Real-time Monitoring of Freezing–Thawing Cycle in Cryosurgery. Appl. Sci. 2020, 10, 1053. [Google Scholar] [CrossRef]
  34. Dolganova, I.N.; Zotov, A.K.; Safonova, L.P.; Aleksandrova, P.V.; Reshetov, I.V.; Zaytsev, K.I.; Tuchin, V.V.; Kurlov, V.N. Feasibility test of a sapphire cryoprobe with optical monitoring of tissue freezing. J. Biophotonics 2023, 16, e202200288. [Google Scholar] [CrossRef]
  35. Katyba, G.; Zaytsev, K.; Dolganova, I.; Shikunova, I.; Chernomyrdin, N.; Yurchenko, S.; Komandin, G.; Reshetov, I.; Nesvizhevsky, V.; Kurlov, V. Sapphire shaped crystals for waveguiding, sensing and exposure applications. Prog. Cryst. Growth Charact. Mater. 2018, 64, 133–151. [Google Scholar] [CrossRef]
  36. Kurlov, V. Sapphire: Properties, Growth, and Applications. In Encyclopedia of Materials: Science and Technology; Buschow, K.J., Cahn, R.W., Flemings, M.C., Ilschner, B., Kramer, E.J., Mahajan, S., Veyssière, P., Eds.; Elsevier: Oxford, UK, 2001; pp. 8259–8264. [Google Scholar] [CrossRef]
  37. Pushkarev, A.V.; Ryabikin, S.S.; Tsiganov, D.I.; Zotov, A.K.; Kurlov, V.N.; Dolganova, I.N. Comparison of Probe Materials for Tissue Cryoablation: Operational Properties of Metal and Sapphire Cryoprobes. J. Biomed. Photonics Eng. 2022, 8, 040501. [Google Scholar] [CrossRef]
  38. LaBelle, H. EFG, the invention and application to sapphire growth. J. Cryst. Growth 1980, 50, 8–17. [Google Scholar] [CrossRef]
  39. Kurlov, V.N.; Rossolenko, S.N.; Abrosimov, N.V.; Lebbou, K. Shaped Crystal Growth. In Crystal Growth Processes Based on Capillarity; John Wiley & Sons, Ltd.: Hoboken, NJ, USA, 2010; Chapter 5; pp. 277–354. [Google Scholar] [CrossRef]
  40. Kurlov, V.; Rossolenko, S. Growth of shaped sapphire crystals using automated weight control. J. Cryst. Growth 1997, 173, 417–426. [Google Scholar] [CrossRef]
  41. Fantini, S.; Franceschini, M.A.; Gratton, E. Semi-infinite-geometry boundary problem for light migration in highly scattering media: A frequency-domain study in the diffusion approximation. J. Opt. Soc. Am. B 1994, 11, 2128–2138. [Google Scholar] [CrossRef]
  42. Liemert, A.; Kienle, A. Light diffusion in a turbid cylinder. I. Homogeneous case. Opt. Express 2010, 18, 9456–9473. [Google Scholar] [CrossRef]
  43. Franceschini, M.A.; Fantini, S.; Paunescu, L.A.; Maier, J.S.; Gratton, E. Influence of a superficial layer in the quantitative spectroscopic study of strongly scattering media. Appl. Opt. 1998, 37, 7447–7458. [Google Scholar] [CrossRef]
  44. Liemert, A.; Kienle, A. Light diffusion in a turbid cylinder. II. Layered case. Opt. Express 2010, 18, 9266–9279. [Google Scholar] [CrossRef]
  45. Hallacoglu, B.; Sassaroli, A.; Fantini, S. Optical Characterization of Two-Layered Turbid Media for Non-Invasive, Absolute Oximetry in Cerebral and Extracerebral Tissue. PLoS ONE 2013, 8, e64095. [Google Scholar] [CrossRef]
  46. Ishimaru, A. Diffusion of light in turbid material. Appl. Opt. 1989, 28, 2210–2215. [Google Scholar] [CrossRef]
  47. Farrell, T.J.; Patterson, M.S.; Wilson, B. A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo. Med. Phys. 1992, 19, 879–888. [Google Scholar] [CrossRef]
  48. Agafonkina, I.V.; Belozerov, A.G.; Berezovsky, Y.M.; Korolev, I.A.; Pushkarev, A.V.; Tsiganov, D.I.; Shakurov, A.V.; Zherdev, A.A. Thermal Properties of Biological Tissue Gel-Phantoms in a Wide Low-Temperature Range. J. Eng. Phys. Thermophys. 2021, 94, 790–803. [Google Scholar] [CrossRef]
  49. Agafonkina, I.V.; Belozerov, A.G.; Vasilyev, A.O.; Pushkarev, A.V.; Tsiganov, D.I.; Shakurov, A.V.; Zherdev, A.A. Thermal Properties of Human Soft Tissue and Its Equivalents in a Wide Low-Temperature Range. J. Eng. Phys. Thermophys. 2021, 94, 233–246. [Google Scholar] [CrossRef]
  50. Wagnières, G.; Cheng, S.; Zellweger, M.; Utke, N.; Braichotte, D.; Ballini, J.P.; van den Bergh, H. An optical phantom with tissue-like properties in the visible for use in PDT and fluorescence spectroscopy. Phys. Med. Biol. 1997, 42, 1415. [Google Scholar] [CrossRef]
  51. Tuchin, V.V. Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnostics, 3rd ed.; SPIE Press: Bellingham, WA, USA, 2015. [Google Scholar]
  52. Lai, P.; Xu, X.; Wang, L.V. Dependence of optical scattering from Intralipid in gelatin-gel based tissue-mimicking phantoms on mixing temperature and time. J. Biomed. Opt. 2014, 19, 035002. [Google Scholar] [CrossRef]
  53. Assadi, H.; Karshafian, R.; Douplik, A. Optical Scattering Properties of Intralipid Phantom in Presence of Encapsulated Microbubbles. Int. J. Photoenergy 2014, 2014, 471764. [Google Scholar] [CrossRef]
  54. Aernouts, B.; Beers, R.V.; Watté, R.; Lammertyn, J.; Saeys, W. Dependent scattering in Intralipid® phantoms in the 600–1850 nm range. Opt. Express 2014, 22, 6086–6098. [Google Scholar] [CrossRef]
Figure 1. A sapphire applicator: (a) Schematic of a multi-channeled 12 mm diameter applicator with optical fibers accommodated inside three channels. (b) Schematic of beam path of the detected optical diffuse signals, R 1 = 4 mm, R 2 = 8 mm. (c) A photo of the sapphire applicator. (d) The experiment of an ice ball formation in a tissue phantom by the sapphire applicator.
Figure 1. A sapphire applicator: (a) Schematic of a multi-channeled 12 mm diameter applicator with optical fibers accommodated inside three channels. (b) Schematic of beam path of the detected optical diffuse signals, R 1 = 4 mm, R 2 = 8 mm. (c) A photo of the sapphire applicator. (d) The experiment of an ice ball formation in a tissue phantom by the sapphire applicator.
Sensors 24 03655 g001
Figure 2. Calibration of the applicator: (a) The measured intensities at two channels with R 1 = 4 and R 2 = 8 mm. An aqueous 3% solution of Lipoplus 20 was used as a sample. (b) Comparison of the actually measured signal slope, its theoretical estimation and the fit line [ ln C 1 C 2 R ] , where C 2 μ eff and μ eff were determined from Equation (2). All lines were shifted to zero at R 1 for simplicity.
Figure 2. Calibration of the applicator: (a) The measured intensities at two channels with R 1 = 4 and R 2 = 8 mm. An aqueous 3% solution of Lipoplus 20 was used as a sample. (b) Comparison of the actually measured signal slope, its theoretical estimation and the fit line [ ln C 1 C 2 R ] , where C 2 μ eff and μ eff were determined from Equation (2). All lines were shifted to zero at R 1 for simplicity.
Sensors 24 03655 g002
Figure 3. Signal analysis during the tissue phantom freezing: (a,b) Time and thickness evolution of normalized diffuse reflected intensities I meas and I th , respectively, detected at two channels separated from the illumination channel by distance R; I meas is shown after smoothing filtering. (c,d) Normalized effective attenuation coefficient obtained from the measured signals and calculated intensities, respectively. (e) The rate of change of the measured attenuation. (f) Comparison of μ meas and μ th , assuming a certain time delay indicated by the blue region, aimed at L reconstruction by Equation (8). Gray regions denote measurement errors.
Figure 3. Signal analysis during the tissue phantom freezing: (a,b) Time and thickness evolution of normalized diffuse reflected intensities I meas and I th , respectively, detected at two channels separated from the illumination channel by distance R; I meas is shown after smoothing filtering. (c,d) Normalized effective attenuation coefficient obtained from the measured signals and calculated intensities, respectively. (e) The rate of change of the measured attenuation. (f) Comparison of μ meas and μ th , assuming a certain time delay indicated by the blue region, aimed at L reconstruction by Equation (8). Gray regions denote measurement errors.
Sensors 24 03655 g003
Figure 4. Reconstruction of freezing depth in tissue phantom performed by the sapphire cryo-applicator: (a) Time evolution of the reconstructed depth and the visualized ice ball height in gelatin phantom. (b) Comparison of the reconstructed and visualized freezing depth for frequency-domain solution (data from Ref. [34]) and steady-state solution. The region of error bars represent ± σ .
Figure 4. Reconstruction of freezing depth in tissue phantom performed by the sapphire cryo-applicator: (a) Time evolution of the reconstructed depth and the visualized ice ball height in gelatin phantom. (b) Comparison of the reconstructed and visualized freezing depth for frequency-domain solution (data from Ref. [34]) and steady-state solution. The region of error bars represent ± σ .
Sensors 24 03655 g004
Figure 5. The freezing of rat liver by the sapphire cryo-applicator: (a) An image of the freezing process. (b) A representative H&E histological image of the liver tissue after cold-induced damage; the damaged tissue is shown on the right.
Figure 5. The freezing of rat liver by the sapphire cryo-applicator: (a) An image of the freezing process. (b) A representative H&E histological image of the liver tissue after cold-induced damage; the damaged tissue is shown on the right.
Sensors 24 03655 g005
Figure 6. The results of monitoring the cryodestruction of rat liver tissue: (a,b) Time and thickness evolution of normalized diffuse reflected intensities I meas and I th , respectively, detected at two channels separated from the illumination channel by distance R; I meas is shown after smoothing filtering. (c) Normalized effective attenuation coefficient obtained from the measured signals and calculated intensities. Data are shown for sample 4. (d) Time evolution of the reconstructed freezing depth (crosses) compared with the visualized depth (circles) at the end of exposure.
Figure 6. The results of monitoring the cryodestruction of rat liver tissue: (a,b) Time and thickness evolution of normalized diffuse reflected intensities I meas and I th , respectively, detected at two channels separated from the illumination channel by distance R; I meas is shown after smoothing filtering. (c) Normalized effective attenuation coefficient obtained from the measured signals and calculated intensities. Data are shown for sample 4. (d) Time evolution of the reconstructed freezing depth (crosses) compared with the visualized depth (circles) at the end of exposure.
Sensors 24 03655 g006
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Zotov, A.K.; Pushkarev, A.V.; Alekseeva, A.I.; Zaytsev, K.I.; Ryabikin, S.S.; Tsiganov, D.I.; Zhidkov, D.A.; Burkov, I.A.; Kurlov, V.N.; Dolganova, I.N. Optical Sensing of Tissue Freezing Depth by Sapphire Cryo-Applicator and Steady-State Diffuse Reflectance Analysis. Sensors 2024, 24, 3655. https://doi.org/10.3390/s24113655

AMA Style

Zotov AK, Pushkarev AV, Alekseeva AI, Zaytsev KI, Ryabikin SS, Tsiganov DI, Zhidkov DA, Burkov IA, Kurlov VN, Dolganova IN. Optical Sensing of Tissue Freezing Depth by Sapphire Cryo-Applicator and Steady-State Diffuse Reflectance Analysis. Sensors. 2024; 24(11):3655. https://doi.org/10.3390/s24113655

Chicago/Turabian Style

Zotov, Arsen K., Aleksandr V. Pushkarev, Anna I. Alekseeva, Kirill I. Zaytsev, Sergey S. Ryabikin, Dmitry I. Tsiganov, Dmitriy A. Zhidkov, Ivan A. Burkov, Vladimir N. Kurlov, and Irina N. Dolganova. 2024. "Optical Sensing of Tissue Freezing Depth by Sapphire Cryo-Applicator and Steady-State Diffuse Reflectance Analysis" Sensors 24, no. 11: 3655. https://doi.org/10.3390/s24113655

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop