Article

Miniaturizable Chemiluminescence System for ATP Detection in Water

by
Giuseppe E. Capuano
1,
Domenico Corso
1,*,
Roberta Farina
1,2,
Gianni Pezzotti Escobar
3,
Giuseppe A. Screpis
4,
Maria Anna Coniglio
1,4,* and
Sebania Libertino
1
1
Istituto per la Microeletttronica e Microsistemi—Consiglio Nazionale delle Ricerche, VIII Strada Z.I., 5, 95121 Catania, Italy
2
Department of Chemical Sciences, University of Catania, Viale Andrea Doria 6, 95125 Catania, Italy
3
URT “LabSens of Beyond Nano” of the Department of Physical Sciences and Technologies of Matter, National Research Council (CNR-DSFTM-ME), Viale Ferdinando Stagno D’Alcontres 31, 98166 Messina, Italy
4
Department of Medical, Surgical Sciences and Advanced Technologies, University of Catania, Via S. Sofia 78, 95123 Catania, Italy
*
Authors to whom correspondence should be addressed.
Sensors 2024, 24(12), 3921; https://doi.org/10.3390/s24123921
Submission received: 23 May 2024 / Revised: 10 June 2024 / Accepted: 12 June 2024 / Published: 17 June 2024

Abstract

:
We present the design, fabrication, and testing of a low-cost, miniaturized detection system that utilizes chemiluminescence to measure the presence of adenosine triphosphate (ATP), the energy unit in biological systems, in water samples. The ATP–luciferin chemiluminescent solution was faced to a silicon photomultiplier (SiPM) for highly sensitive real-time detection. This system can detect ATP concentrations as low as 0.2 nM, with a sensitivity of 79.5 A/M. Additionally, it offers rapid response times and can measure the characteristic time required for reactant diffusion and mixing within the reaction volume, determined to be 0.3 ± 0.1 s. This corresponds to a diffusion velocity of approximately 44 ± 14 mm2/s.

1. Introduction

The presence of water on earth is essential for the development and sustenance of life, and it is largely used for anthropic activities. Water quality is an important condition for preserving human health. Many diseases can occur when water is contaminated by bacterial strains. Generally, microbial contents are measured to evaluate drinking water quality, food safety [1,2,3], and clinical analysis [4,5,6,7]. Today, more than ever, it is essential to monitor the microbiological quality of water in networks intended for human consumption to ensure public safety and health. This control not only allows for assessing the effectiveness of disinfection systems already in use in Water Safety Plans [8,9] but also for testing and developing new antibiofilm methods that can further contribute to improving water supply safety [10,11]. Several isolation and identification methods are now employed, such as conventional culturing methods, immunological protocols, molecular methods, and spectroscopic techniques. Culturing methods, also coupled with spectroscopy and mass spectrometry techniques, are usually limited to bench-top equipment requiring skilled technicians and are laborious and time-consuming. Immunological and molecular techniques for detecting bacterial pathogens are much faster than the traditional culture methods [12]. They are based on the selective interaction between a specific antibody (Ab) and a specific antigen (Ag) of the searched bacterial strain. The most used immunological techniques are the enzyme-linked immunosorbent assay (ELISA), radiometric methods (RIAs), fluorescence methods (FIAs), and chemiluminescent methods (CLIAs) [13].
However, in most cases, these methods suffer from poor specificity, sometimes returning false positive results and, then, contributing to issues of low sensitivity [14,15].
In the last decades, the polymerase chain reaction (PCR) has been considered a novel technique for bacterial strain detection. It is based on the amplification of bacterial DNA using the polymerase enzyme and consists of cycles in which DNA denaturation, primer annealing, and extension processes are involved [16,17]. Quantitative information concerning the initial amount of DNA contained in a given sample can be obtained by employing fluorescence to monitor bacterial DNA amplification in real time (“quantitative PCR” or “qPCR”) [18,19]. However, this technique requires sample preparation and skilled personnel not only in sample processing but also in interpreting the results.
Recently, researchers developed a novel molecular technique to mitigate the shortcomings of the PCR called the “isothermal technique”. The latter equally employs the nucleic-acid-based method of identification principle and works at a fixed temperature for nucleic acid amplification [20]. Amongst the different isothermal techniques employed in pathogen detection, the loop-mediated isothermal amplification (LAMP) technique has gained popularity due to its specificity, stability, and sensitivity [21,22]. Despite its widespread application, the LAMP technique encounters some limitations during usage [23]. The extraction, amplification, and identification of microbial DNA in various environmental water and food samples are quite difficult, especially when the amount of the pathogen is low. Moreover, all of these techniques require some knowledge of the DNA sequence of the specific gene that must be amplified to properly design some primers to be employed for the correct application of the methods.
It is worth knowing that the use of nanoparticles, graphene quantum dots, and electrospun nanofibers showing pathogens’ selectivity has been recently considered to improve bacterial detection techniques [24].
Currently, adenosine triphosphate (ATP)–luciferin-driven chemiluminescence is one of the most sensitive methods for bacterial charge detection [25]. ATP bioluminescent assays are reliable techniques to measure the number of viable cells [26,27,28] or biomolecules [29] in samples, widely used to indicate microbial contamination. In particular, the ATP–luciferin chemiluminescence reaction is catalyzed by the enzyme luciferase to obtain an intermediate complex (luciferyl adenylate) that combines with oxygen to produce oxyluciferin and a photon at 560 nm. The fully described reaction is as follows [30]:
l u c i f e r i n + A T P + O 2 M g 2 + l u c i f e r a s e o x y l u c i f e r i n + A M P + C O 2 + p y r o p h o s p h a t e + h ν
Santangelo et al. [31] developed the prototype of a low-cost, miniaturized, and highly sensitive ATP detection device employing the ATP–luciferin chemiluminescent reaction and based on 3D printed lab-on-chip microfluidics (LoC) [32,33,34] coupled with a non-commercial 0.4 × 0.4 mm2 silicon photomultiplier (SiPM, fabricated by STMicroelectronics) for highly sensitive, continuous, and real-time ATP detection, reaching a limit of detection (LoD) equal to 8 nM and sensitivity of 1.82 × 10−2 A/M; the authors also performed static measurements by replacing the 3D chip with a glass cuvette, reaching a LoD equal to 3.7 nM and a sensitivity of 9.32 × 10−2 A/M (ampere/moles). The SiPM is a solid-state photodetector consisting of a bi-dimensional array of microcells (pixels), each one composed of a series of quenching resistors and a single photon avalanche diode (SPAD), operating in Geiger mode [35,36,37,38]. It is characterized by high quantum efficiency, high gain, high speed, low operating voltage, simple output electronics, and single photon sensitivity, useful for the detection of very low photon fluxes.
Hu et al. [39] developed a low-cost and portable device based on the ATP–luciferin-driven chemiluminescence reaction taking place in test tubes for bacterial charge detection, where the main components of the acquisition system are a low-noise photoelectric signal detection and a processing circuit based on an SiPM, a power management module, and a high-performance embedded microcontroller subsystem with peripheral circuits. The authors adopted the balanced chopper modulation and lock-in amplification techniques to improve the signal-to-noise ratio and performed a zero-adjustment to eliminate the dark current of the SiPM and increase the dynamic range of the system, detecting a minimum ATP concentration of 3.6 × 10−11 M and obtaining a sensitivity of 0.387 mV/M. However, the proposed system requires complex electronic circuitry and the material to be tested must be sampled and inserted in the instrument by a human operator.
Some commercial devices developed for cellular ATP detection reach low LoD. For example, the BioThema ATP Kit SL [40] allows the monitoring of the ATP concentration ranging from 10−12 to 10−6 M; water samples are treated and inserted into cuvettes or microplates along with the relative reactants, and the measurements are performed through an appropriate luminometer. The Luminultra Quench-Gone Aqueous (QGA) Test Kit [41] is optimized to measure down to 0.1 ng/L (which corresponds to about 2 × 10−13 M of ATP, considering its molecular weight AATP = 507.181 g/mol [42]). In this case, the ATP sample is extracted through filtering, diluted, and inserted into microtubes employed for measurement with a luminometer. The Hygiena handy type luminometer [43] allows the monitoring of the bacterial charge in surfaces to be sanitized, returning information about the cleaning status. In particular, each sample is obtained by using a swab, where the latter is treated and inserted into the luminometer; the surface is considered clean if the system returns concentration values below 20 relative light units (RLU). For all systems, the operator has to prepare the sample to be measured (either by filtering and mixing water samples with reactants or by swabbing a surface and dipping the swab into buffer solutions). It means they depend on the operator’s expertise and the measurement requires a long time for sample preparation.
In this work we propose a low-cost and very high-sensitivity bacterial detection system based on the ATP–luciferin-driven chemiluminescence reaction taking place in a static regime, aiming at reaching a low LoD and a high sensitivity. The system is mainly composed of a 5 × 5 mm2 base-squared quartz cuvette with five optical windows (the lateral ones are appropriately covered with aluminum to avoid the dispersion of the emitted photons) where the ATP–luciferin-driven reaction takes place, and of a 6.07 × 6.07 mm2 commercial SiPM posed under the quartz cuvette, which detects the photons emitted during the reaction. The output signal is a current proportional to the intensity of the emitted light and, consequently, to the ATP concentration. The device possesses simple electronic circuitry and does not need the presence of human operators for sampling and sample treatment, allowing it to perform in situ, remote-sensing, and real-time measurements.

2. Materials and Methods

2.1. SiPM Details

The SiPM utilized in this study is a commercial device manufactured by OnSemiconductor [44] (Phoenix, AZ, USA, MICROFJ-60035-TSV-TR1). It possesses dimensions of 6.07 × 6.07 mm2, housing a total of 22,292 pixels, each consisting of a Single Photon Avalanche Diode (SPAD) operating in Geiger mode [38] and linked to a quenching resistor. The SiPM exhibits a measured breakdown voltage of −24.5 V [44], falling within the voltage range specified by the manufacturer (reverse voltage of 24.2 V–24.7 V).
The operating voltage (Vop) was determined by considering an overvoltage of −2.5 V, chosen to achieve a high gain (2.9 × 10−6 in this instance). For the measurements conducted in this study, the SiPM was biased at Vop = −27 V. At this operating voltage, the detector exhibits a Dark Count (DC) of 5 × 104 Hz/mm2, corresponding to a dark current of 9.0 × 10−7 A, with an 8% crosstalk probability [44]. The measured dark current of 1.06 × 10−6 A aligns closely with the value provided in the datasheet.
Measurements were carried out by biasing the SiPM using a Source Meter Multi-Channel I-V, KEITHLEY 2636 (Tektronix, Beaverton, OR, USA 2600 Series SourceMeter®), employing a Lab-VIEW® (National Instruments, Austin, TX, USA, Version 2021) routine to control the bias voltage and measurement parameters (such as time and integration time), to acquire the current signal, and to automatically record data. Subsequently, the collected data were analyzed offline using a Matlab® (Mathworks, Natick, MA, USA, Version 2023) routine (see Figure 1 for a schematic of the measurement setup).

2.2. Setup Description

The proposed detection system is formed by the SiPM, already described in Section 2.1, and a 5 × 5 mm2 base-squared quartz cuvette with 5 optical windows coupled together (Hellma, Müllheim, Germany, Micro-cuvette 111-057-40 QS). The cuvette sides were appropriately covered with aluminum to collect all the produced light, hence maximizing the electric signal related to the chemiluminescent reaction. The cuvette was filled with 100 μL ATP (see after for preparation details) at different concentrations; then, the standard reaction solution (SRS) containing the luciferin was added through a hand-made pumping system, composed of a syringe linked to a silicon tube. The former was positioned out of the dark chamber containing the detection system (not reached by the environmental light), while the silicon tube was long enough to enter the chamber through a light-shielded entrance. A schematic representation of the experimental set-up is shown in Figure 1. The syringe–tube system delivered 100 μL SRS into the cuvette. In this way, not only can the mixing and the reaction take place while the chamber is closed, but also the syringe–tube device allows one to pump the SRS into the cuvette after the data acquisition is started, obtaining detailed information on the characteristic reaction time required by the reactants to mix in the reaction volume (200 μL), since the reaction occurs instantaneously at room temperature [45].

2.3. Standard Preparation and ATP Bioluminescence Measurements

ATP Determination Kit (A22066) from Molecular Probes [46] was used as a reference for the quantitative determination of ATP concentration with recombinant firefly luciferase and its substrate D-luciferin. It is based on the reaction reported in Equation (1). The assay involves the enzyme luciferase to catalyze luciferin oxidation, where magnesium ions (Mg2+) are considered enzyme cofactors. The ATP concentration is proportional to the amount of emitted light (560 nm) released along with carbon dioxide (CO2), adenosine monophosphate (AMP), and pyrophosphate. The SRS was prepared using MilliQ water from Simplicity UV (Millipore, by Merk, Darmstadt, Germany), 20× Reaction Buffer, dithiothreitol (DTT) to a concentration of 0.1 M, D-luciferin to a concentration of 10 mM, and firefly luciferase (5 mg/mL). The ATP concentrations considered for the chemiluminescence measurements were 500 nM, 250 nM, 125 nM, 62.5 nM, 31.25 nM, 15.62 nM, 7.81 nM, 3.90 nM, 1.95 nM, 0.98 nM, 0.50 nM, and 0.25 nM, starting from the concentration of 5 mM provided by the ATP Determination Kit. Finally, three sets of ATP concentration measurements were taken between 0.25 nM and 500 nM to obtain a satisfying statistic at each concentration.

3. Results

Before proceeding with the ATP detection measurements, the system’s dark conditions and the absence of optical sources interfering with the measurements were verified. Figure 2 reports the comparison between the dark count noise signal and the one obtained by filling the cuvette with only SRS to highlight the absence of optical interferents during the measurements. Therefore, the static performances of the ATP detection system were tested by employing different ATP concentrations ranging between 0.25 nM and 500 nM (see Section 2 for details).
Figure 3a summarizes one of three sets of current measurements in the static configuration as a function of time for the different ATP concentrations. The data show an increase in the optical signal as the ATP concentration in the sample increases. Figure 3b shows a zoom-in of the measurements at low concentrations and allows the verification that even at the lowest measured value the signal is above the dark current. The signal is quite stable, decreasing slowly with time, as already observed in ref. [31]. In particular, the current signal reduces as a function of time more rapidly at higher ATP concentrations.
To study the current reduction as a function of time, the range from 15 s to 35 s was defined. The ratio between the current values at 15 s and 35 s was calculated for each ATP concentration and considering all three sets of measurements. The data, expressed as percentages, are summarized in Figure 4.
This range has been chosen to guarantee that the current is reducing. In fact, for the first few seconds, the signal is stationary, while for longer times the reduction rate changes, going asymptotically toward the dark current. The signal decreases similarly for concentrations lower than 15.62 nM, with a reduction of about 3%. For concentrations higher than 15.62 nM, the current reduction percentage increases with ATP concentration, following the trend shown in Figure 4. At 500 nM, the current reduction is 9% of the starting value, about three times higher than the reduction measured at low ATP concentrations. It is worth noting that current reduction values have been estimated by performing polynomial fits over each current curve in a decreasing phase relative to each set of measurements for each ATP concentration (such as those reported in Figure 3). The observed trend can be explained by considering a faster consumption of the SRS reactant at higher ATP concentrations.
To correctly analyze the data, a study was performed on the first few seconds from the SRS injection. In fact, after inserting the SRS into the cuvette containing the ATP solution, the current signal I(t) increases from the reference (dark noise) to the maximum current value Imax corresponding to each ATP concentration by following the trend reported in Equation (2):
I t = I max   1 e t τ  
where t is the time variable, and τ is the characteristic increasing time in correspondence to which the current intensity is equal to ( 1 1 / e ) · I max     0.63   I max   (see Figure 5).
The data analysis of all three sets of measurements demonstrates that τ is almost the same for all of the concentrations, obtaining a value of 0.3 ± 0.1 s (see Figure 6). It is strictly related to the diffusive mechanism characterizing the SRS mixing with ATP solution into the reaction volume. If the diffusion of the SRS reactant starts from a single point toward the surrounding space into the reaction volume under a concentration gradient, and approximating the reaction volume V to a sphere with radius r = 3 V / 4 π 3 = 3.6   m m , the SRS diffusion velocity D can be estimated using the following expression:
D = r 2 τ 44 ± 14   m m 2 s 1
If the firefly luciferase molecule drives the diffusion mechanism (considering its remarkable weight, i.e., 62 kDa [47]), it is possible to verify the relationship between the experimental diffusion velocity and the firefly luciferase dimensions in a single-point diffusion scenario. If the presence of turbulent mechanisms is neglected, the energy dissipated through the viscous friction force of the diffusing molecules can be equaled to their thermal energy, obtaining the following:
6 π η R D = K B T
where KB is the Boltzmann constant, T = 293 K is the temperature at which the diffusion took place, η is the solution viscosity (assumed equal to the water viscosity, that is, 1.0016 × 10−3 Pa s at 20 °C [48]), and R is the radius of the diffusing molecule (approximated to a sphere). Considering the molecular weight M of the firefly luciferase, which is 62 kDa, the radius of the diffusing molecule can be estimated by considering the following expressions (reported in Equations (2.1) and (2.2) of ref. [49]):
V M = 0.73   c m 3 g 1 × 10 21   n m 3 c m 3 6.023 × 10 23   D a / g × M   D a = 1.212 × 10 3 × M
R = 3 V M / 4 π   3 = 0.066 × M 3 = 2.6   n m
where VM is the volume of the molecule and 0.73 cm3/g is the reciprocal of the mean density of all proteins (about 1.37 g/cm3). The expected diffusion velocity can be inferred through the following expression (also referred to as the “Stokes–Einstein” equation) [50]:
D = K B T 6 π η R
obtaining a velocity of 8.3 × 10−5 mm2/s; this value is five orders of magnitudes lower than that estimated experimentally, implying a diffusion time of 1.6 × 105 s (about 44 h).
The large difference between the expected and measured diffusion velocity values is due to the assumptions considered in the model used. The reaction volume is not spherical (as assumed previously) and the mixing of the SRS with the ATP solution is helped by the direct SRS injection into the cuvette (with a starting velocity different from zero), which acquires a not-negligible kinetic energy before the mixing, generating turbulences and then facilitating its diffusion into the ATP solution. This process remarkably reduces the diffusion times concerning those associated with the condition of single-point diffusion assumed in the model. By dividing the radius associated with the reaction volume by the experimental diffusion time, the mean velocity along one direction can be estimated as follows:
v = r τ = 12 ± 4 m m s
which is a reasonable value for the SRS velocity during the injection process.
For each set of measurements, we collected the maximum current Imax as the average current acquired in a time range going from 2 to 5 s. Following this procedure, we are confident the solution is already in the “fully mixed” regime, before the current reduction range, and the time interval used to average the signal is enough to have a reasonable number of measurement points. For each curve, the maximum current value has been associated with the relative ATP concentration. The minimum value measured with the detection system is 0.25 nM (see Figure 7), more than an order of magnitude lower than the LoD inferred by Santangelo et al. [31] in the static regime (LoD = 3.7 nM) and about an order of magnitude higher than the minimum ATP concentration detected by Hu et al. [38] (3.6 × 10−2 nM). In particular, the LoD value inferred with the system proposed in this work is 0.2 nM, estimated by using the following equation [51]:
L o D = 3 × S D s l o p e
where SD is the standard deviation of the reference signal, and the slope (sensitivity) refers to the linear fit of the calibration curve (that associates each maximum current value with the corresponding ATP concentration). In this case, the measured SD is equal to 5.5 × 10−9 A, while the sensitivity is 79.5 A/M, much higher than the values reported in the literature [31,39]. The sensitivity is also higher than typical sensitivities of high-cost and bulky bench-scale commercial laboratory luminescence devices [40,41,43], although these show very low LoD thanks to the preliminary application of filtering processes.
By knowing the mean amount of ATP characterizing a living cell, the bacterial concentration in a water sample can be inferred by detecting the ATP extracted from bacterial cells through appropriate procedures. For this purpose, it is possible to define the Microbial Equivalent (ME) unit [52], where it is assumed that each bacterial cell holds 1 fg of ATP, which is the average amount for a typical Escherichia coli cell. Therefore, by knowing the ATP molecular weight AATP = 507.181 g/mol [42], a single living cell contains a mean ATP amount of 2 × 10−18 mol. This means that the minimum ATP concentration detectable with the proposed system (LoD = 0.2 nM) corresponds to a minimum detectable bacterial concentration of 108 cells/L.
For bacterial load determination in water samples, pretreatment is carried out in commercial applications [40,41,43]. The water sample is filtered, and the filter (containing the bacteria) is attached to a syringe barrel to start the lysis process. This is achieved by adding 1 mL of lysis buffer to the barrel; it allows the bacterial degradation for both gram-negative and gram-positive bacteria to pass gently through the filter as a consequence of lysozyme activity toward the cell wall. This step is swift; hence, to increase sensitivity, often additional samples are filtered and the lysis process is repeated. Once all bacteria have been disrupted, soluble proteins and nucleic acids are released and ATP is dissolved in a tube containing an ATP dilution reagent and ready to be tested. For our application, we aim to avoid the sample concentration. The final goal of our approach is to avoid, or at least reduce, the amount of water to be filtered.

4. Conclusions

In this work, we describe the design, fabrication, and testing of a low-cost and miniaturized chemiluminescent sensing system for highly sensitive real-time ATP detection. The system is mainly composed of a five-optical-windows quartz squared-based cuvette with the sides appropriately covered with aluminum, coupled with an SiPM as a photodetector. The process of optical cuvette wrapping allows the collection of all the light produced during the reaction, while the use of an SiPM provides a high-quality output signal compared to traditional systems due to its enhanced characteristics. The limit of detection of the proposed system is 0.2 nM, while the sensitivity is equal to 79.5 A/M, much higher than the typical sensitivities of bench-scale commercial laboratory luminescence devices. Moreover, the developed system allows one to obtain a direct measurement of the characteristic time required by the reactants to diffuse and mix after injection in the reaction volume (200 μL), equal to 0.3 ± 0.1 s, through which a diffusion velocity of about 44 ± 14 mm2/s has been inferred. Commercial bench devices require the presence of human operators and longer times to perform the same measurements as the proposed system. It can be implemented in situ to monitor water’s total bacterial charge due to its capacity to perform highly sensitive, in situ, remote-sensing, and real-time measurements, with a minimum detectable total bacteria concentration of about 108 cells/L. Future work that has to be considered of primary importance is the development of early warning systems, thanks to the device characteristics such as miniaturization, portability, and low cost, making it ideal for a network of integrated sensors in the surrounding environment to obtain a qualitative pre-screening of water.

Author Contributions

Conceptualization, G.E.C., D.C., S.L. and M.A.C.; methodology, G.E.C., D.C., S.L., M.A.C., G.A.S., R.F. and G.P.E.; software, D.C.; validation, G.E.C., S.L. and D.C.; formal analysis, G.E.C.; investigation, G.E.C., S.L., D.C., M.A.C., G.A.S., R.F. and G.P.E.; resources, S.L. and M.A.C.; data curation, G.E.C., D.C., G.A.S. and R.F.; writing—original draft preparation, G.E.C.; writing—review and editing, G.E.C., S.L., D.C., M.A.C., G.A.S., R.F. and G.P.E.; visualization, G.E.C., D.C., G.A.S. and R.F.; supervision, S.L., D.C. and M.A.C.; project administration, S.L. and M.A.C.; funding acquisition, S.L. and M.A.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the European Union (NextGeneration EU), through the MUR-PNRR project SAMOTHRACE [GA ECS00000022]; the European project TELEGRAM of the European Union’s Horizon 2020 Research and Innovation Program [GA 101006941]; and Italian PON ARS01_00333_TETI funded by Italian University and Research Minister.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The raw data supporting the conclusions of this article are available in the figures. Ascii files will be made available by the authors upon request.

Acknowledgments

The authors would like to thank Giuseppina Faro for the precious help in sample preparation.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Champiat, D.; Matas, N.; Monfort, B.; Fraass, H. Applications of biochemiluminescence to HACCP. Luminescence 2001, 16, 193–198. [Google Scholar] [CrossRef]
  2. Hameed, S.; Xie, L.; Ying, Y. Conventional and emerging detection techniques for pathogenic bacteria in food science: A review. Trends Food Sci. Technol. 2018, 81, 61–73. [Google Scholar] [CrossRef]
  3. Ferone, M.; Gowen, A.; Fanning, S.; Scannell, A.G.M. Microbial detection and identification methods: Bench top assays to omics approaches. Compr. Rev. Food Sci. Food Saf. 2020, 19, 3106–3129. [Google Scholar] [CrossRef]
  4. Conn, R.B.; Charache, P.; Chappelle, W.E. Limits of applicability of the firefly luminescence ATP assay for the detection of bacteria in clinical specimens. Am. J. Clin. Pathol. 1975, 63, 493–501. [Google Scholar] [CrossRef]
  5. Turner, D.E.; Daugherity, E.K.; Altier, C.; Maurer, K.J. Efficacy and limitations of an ATP-based monitoring system. J. Am. Assoc. Lab. Anim. Sci. 2010, 49, 190–195. [Google Scholar]
  6. Jagrosse, D.; Bommarito, M.; Stahl, J.B. Monitoring the Cleaning of Surgical Instruments with an ATP Detection System. AJIC 2012, 40, 90–91. [Google Scholar] [CrossRef]
  7. Veiga-Malta, I. Preventing Healthcare-Associated Infections by Monitoring the Cleanliness of Medical Devices and Other Critical Points in a Sterilization Service. BI&T 2016, 50, 45–52. [Google Scholar] [CrossRef]
  8. Coniglio, M.A.; Ferrante, M.; Yassin, M.H. Preventing Health care Associated Legionellosis: Results after 3 Years of Continuous Disinfection of Hot Water with Monochloramine and an Effective Water Safety Plan. Int. J. Environ. Res. Public Health 2018, 15, 1594. [Google Scholar] [CrossRef]
  9. Sciuto, E.L.; Laganà, P.; Filice, S.; Scalese, S.; Libertino, S.; Corso, D.; Faro, G.; Coniglio, M.A. Environmental management of Legionella in domestic water systems: Consolidated and innovative approaches for disinfection methods and risk assessment. Microorganisms 2021, 9, 577. [Google Scholar] [CrossRef]
  10. Sciuto, E.L.; Filice, S.; Coniglio, M.A.; Faro, G.; Gradon, L.; Galati, C.; Spinella, N.; Libertino, S.; Scalese, S. Antimicrobials s-PBC coatings for innovative multifunctional water filters. Molecules 2020, 25, 5196. [Google Scholar] [CrossRef] [PubMed]
  11. Filice, S.; Sciuto, E.L.; Scalese, S.; Faro, G.; Libertino, S.; Corso, D.; Timpanaro, R.M.; Laganà, P.; Coniglio, M.A. Innovative Antibiofilm Smart Surface against Legionella for Water Systems. Microorganims 2022, 10, 870. [Google Scholar] [CrossRef]
  12. Zeng, L.; Wang, L.; Hu, J. Current and Emerging Technologies for Rapid Detection of Pathogens. In Biosensing Technologies for the Detection of Pathogens—A Prospective Way for Rapid Analysis; InTech: London, UK, 2018. [Google Scholar] [CrossRef]
  13. Vasilescu, M.N.; Medvedovici, A.V. Pesticides. In Encyclopedia of Analytical Science, 2nd ed.; Worsfold, P., Townshend, A., Poole, C., Eds.; Elsevier: Amsterdam, The Netherlands, 2005; pp. 55–71. [Google Scholar] [CrossRef]
  14. Dincer, C.; Bruch, R.; Costa-Rama, E.; Fernández-Abedul, M.T.; Merkoçi, A.; Manz, A.; Urban, G.A.; Güder, F. Disposable Sensors in Diagnostics, Food, and Environmental Monitoring. Adv. Mater. 2019, 31, 1806739. [Google Scholar] [CrossRef]
  15. Ramírez-Castillo, F.; Loera-Muro, A.; Jacques, M.; Garneau, P.; Avelar-González, F.; Harel, J.; Guerrero-Barrera, A. Waterborne Pathogens: Detection Methods and Challenges. Pathogens 2015, 4, 307–334. [Google Scholar] [CrossRef]
  16. Saiki, R.K.; Scharf, S.; Faloona, F.; Mullis, K.B.; Horn, G.T.; Erlich, H.A.; Arnheim, N. Enzymatic amplification of beta-globin genomic sequences and restriction site analysis for diagnosis of sickle cell anemia. Science 1985, 230, 1350–1354. [Google Scholar] [CrossRef]
  17. Saiki, S.K.; Gelfand, D.H.; Stoffel, S.; Scharf, S.J.; Higuchi, R.; Horn, G.T.; Mullis, K.B.; Erlich, H.A. Primer-directed enzymatic amplification of DNA with a thermostable DNA polymerase. Science 1988, 239, 487–491. [Google Scholar] [CrossRef]
  18. Holland, P.M.; Abramson, R.D.; Watson, R.; Gelfand, D.H. Detection of specific polymerase chain reaction product by utilizing the 5′----3′ exonuclease activity of Thermus aquaticus DNA polymerase. Proc. Natl. Acad. Sci. USA 1991, 88, 7276–7280. [Google Scholar] [CrossRef] [PubMed]
  19. Higuchi, R.; Dollinger, G.; Walsh, P.S.; Griffith, R. Simultaneous amplification and detection of specific DNA sequences. Bio/Technology 1992, 10, 413–417. [Google Scholar] [CrossRef]
  20. Notomi, T.; Okayama, H.; Masubuchi, H.; Yonekawa, T.; Watanabe, K.; Amino, N.; Hase, T. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000, 28, 63e. [Google Scholar] [CrossRef]
  21. Alonso-Padilla, J.; Cortés-Serra, N.; Pinazo, M.J.; Bottazzi, M.E.; Abril, M.; Barreira, F.; Sosa-Estani, S.; Hotez, P.J.; Gascón, J. Strategies to enhance access to diagnosis and treatment for Chagas disease patients in Latin America. Expert Rev. Anti-Infect. Ther. 2019, 17, 145–157. [Google Scholar] [CrossRef]
  22. Caruso, G.; Coniglio, M.A.; Laganà, P.; Fasciana, T.; Arcoleo, G.; Arrigo, I.; Di Carlo, P.; Palermo, M.; Giammanco, A. Validation of a Loop-Mediated Isothermal Amplification-Based Kit for the Detection of Legionella pneumophila in Environmental Samples according to ISO/TS 12869:2012. Microorganisms 2024, 12, 961. [Google Scholar] [CrossRef]
  23. Karami, A.; Gill, P.; Motamedi, M.H.K.; Saghafinia, M. A review of the current isothermal amplification techniques: Applications, advantages and disadvantages. J. Glob. Infect. Dis. 2011, 3, 293. [Google Scholar] [CrossRef]
  24. Castillo-Henríquez, L.; Brenes-Acuña, M.; Castro-Rojas, A.; Cordero-Salmerón, R.; Lopretti-Correa, M.; Vega-Baudrit, J.R. Biosensors for the Detection of Bacterial and Viral Clinical Pathogens. Sensors 2020, 20, 6926. [Google Scholar] [CrossRef]
  25. Liu, S.; Zhao, J.; Guo, Y.; Ma, X.; Sun, C.; Cai, M.; Chib, Y.; Xu, K. Application of ATP-based bioluminescence technology in bacterial detection: A review. Analyst 2023, 148, 3452. [Google Scholar] [CrossRef]
  26. Brolin, S.E.; Wettermark, G. Bioluminescence Analysis; VCH: New York, NY, USA, 1992. [Google Scholar]
  27. Ahn, J.M.; Kim, B.C.; Gu, M.B. Characterization of gltA: luxCDABE fusion in Escherichia coli as a toxicity biosensor. Biotechnol. Bioprocess Eng. 2006, 11, 516–521. [Google Scholar] [CrossRef]
  28. Morciano, G.; Sarti, A.C.; Marchi, S.; Missiroli, S.; Falzoni, S.; Raffaghello, L.; Pistoia, V.; Giorgi, C.; Di Virgilio, F.; Pinton, P. Use of luciferase probes to measure ATP in living cells and animals. Nat. Protoc. 2017, 12, 1542–1562. [Google Scholar] [CrossRef]
  29. Min, J. 17β-Estradiol-stimulated eNOS gene transcriptional activation is regulated through the estrogen-responsive element in eNOS promoter. Biotechnol. Bioprocess Eng. 2007, 12, 446–449. [Google Scholar] [CrossRef]
  30. Tanii, T.; Goto, T.; Iida, T.; Koh-Masahara, M.; Ohdomari, I. Fabrication of adenosine triphosphate-molecule recognition chip by means of bioluminous enzyme luciferase. Jpn. J. Appl. Phys. 2001, 40, 1135–1137. [Google Scholar] [CrossRef]
  31. Santangelo, M.F.; Libertino, S.; Turner, A.P.F.; Filippini, D.; Mak, W.C. Integrating printed microfluidics with silicon photomultipliers for miniaturised and highly sensitive ATP bioluminescence detection. Biosens. Bioelectron. 2018, 99, 464–470. [Google Scholar] [CrossRef]
  32. Comina, G.; Suska, A.; Filippini, D. Low-cost lab-on-a-chip prototyping with a consumer grade 3D printer. Lab. Chip. 2014, 14, 2978–2982. [Google Scholar] [CrossRef] [PubMed]
  33. Comina, G.; Suska, A.; Filippini, D. 3D printed unibody lab-on-a-chip: Features survey and check-valves integration. Micromachines 2015, 6, 437–451. [Google Scholar] [CrossRef]
  34. Comina, G.; Suska, A.; Filippini, D. Autonomous chemical sensing interface for universal cell phone readout. Angew. Chem. 2015, 54, 8708–8712. [Google Scholar] [CrossRef]
  35. Sciacca, E.; Giudice, A.C.; Sanfilippo, D.; Zappa, F.; Lombardo, S.; Consentino, R.; Di Franco, C.; Ghioni, M.; Fallica, G.; Bonanno, G.; et al. Silicon planar technology for single-photon optical detectors. IEEE Trans. Electron Devices 2003, 50, 918–925. [Google Scholar] [CrossRef]
  36. Mazzillo, M.; Condorelli, G.; Piazza, A.; Sanfilippo, D.; Valvo, G.; Carbone, B.; Fallica, G.; Billotta, S.; Belluso, M.; Bonanno, G.; et al. Single-photon avalanche photodiodes with integrated quenching resistor. Nucl. Instrum. Methods Phys. Res. Sect. A 2008, 591, 367–373. [Google Scholar] [CrossRef]
  37. Mazzillo, M.; Condorelli, G.; Sanfilippo, D.; Valvo, G.; Carbone, B.; Fallica, G.; Billotta, S.; Belluso, M.; Bonanno, G.; Cosentino, L.; et al. Silicon photomultiplier technology at stmicroelectronics. IEEE Trans. Nucl. Sci. 2009, 56, 2434. [Google Scholar] [CrossRef]
  38. Pagano, R.; Valvo, G.; Sanfilippo, D.; Libertino, S.; Corso, D.; Fallica, P.G.; Lombardo, S. Silicon photomultiplier device architecture with dark current improved to the ultimate physical limit. Appl. Phys. Lett. 2013, 102, 183502. [Google Scholar] [CrossRef]
  39. Hu, Z.; Zhang, D.; Lin, H.; Ni, H.; Li, H.; Guan, Y.; Jin, Q.; Wu, Y.; Guo, Z. Low-cost portable bioluminescence detector based on silicon photomultiplier for on-site colony detection. Anal. Chim. Acta 2021, 1185, 339080. [Google Scholar] [CrossRef]
  40. ATP Kit SL. Available online: https://biothema.com/shop/kits/atp-kit-sl/ (accessed on 20 May 2024).
  41. 2nd Generation ATP® Testing. Available online: https://www.luminultra.com/tech/2nd-generation-atp (accessed on 29 April 2024).
  42. Adenosine-5′-triphosphate. Available online: https://pubchem.ncbi.nlm.nih.gov/compound/Adenosine-5_-triphosphate (accessed on 29 April 2024).
  43. Welcometo Smarter ATP Hygiene Monitoring. Available online: https://www.hygiena.com/documents/64121/implementation-guide-for-ensure-touch.pdf (accessed on 20 May 2024).
  44. Silicon Photomultipliers (SiPM), High PDE and Timing Resolution Sensors in a TSV Package. Available online: https://www.onsemi.com/pdf/datasheet/microj-series-d.pdf (accessed on 29 April 2024).
  45. Lundin, A. Use of firefly luciferase in ATP-related assays of biomass, enzymes, and metabolites. Methods Enzymol. 2000, 305, 346–370. [Google Scholar] [CrossRef]
  46. Molecular Probes. ATP Determination Kit (A22066). Product Information. 2005. Available online: https://assets.thermofisher.com/TFS-Assets%2FLSG%2Fmanuals%2Fmp22066.pdf (accessed on 29 April 2024).
  47. Conti, E.; Franks, N.P.; Brick, P. Crystal structure of firefly luciferase throws light on a superfamily of adenylate-forming enzymes. Structure 1996, 4, 287–298. [Google Scholar] [CrossRef]
  48. Viscosity of Water. Available online: https://wiki.anton-paar.com/be-en/water/ (accessed on 13 May 2024).
  49. Erickson, H.P. Size and Shape of Protein Molecules at the Nanometer Level Determined by Sedimentation, Gel Filtration, and Electron Microscopy. Biol. Proced. Online 2009, 11, 32. [Google Scholar] [CrossRef]
  50. Cruickshank, M.C. The Stokes-Einstein law for diffusion in solution. Proc. R. Soc. London. Ser. A 1924, 106, 724–749. [Google Scholar] [CrossRef]
  51. Armbruster, D.A.; Tillman, M.D.; Hubbs, L.M. Limit of detection (LQD)/limit of quantitation (LOQ): Comparison of the empirical and the statistical methods. Clin. Chem. 1994, 40 Pt 1, 1233–1238. [Google Scholar] [CrossRef] [PubMed]
  52. Test Kit Instructions. Available online: https://biotechnologysolutions.com/wp-content/uploads/2016/11/LuminUltra-Test-Kit-Instructions-QGA.pdf (accessed on 29 April 2024).
Figure 1. The schematic layout of the ATP detection system in a static configuration.
Figure 1. The schematic layout of the ATP detection system in a static configuration.
Sensors 24 03921 g001
Figure 2. The current versus time measurements of the dark current (black curve) and the signal obtained by using only SRS (red curve).
Figure 2. The current versus time measurements of the dark current (black curve) and the signal obtained by using only SRS (red curve).
Sensors 24 03921 g002
Figure 3. (a) The current signal versus time as a function of ATP concentration, for one of three sets of measurements. (b) The zoom-in of the region identified by the black dashed rectangle shown in (a). The bottom red curve is the dark current signal. The legend refers to the measured ATP concentrations, with values ranging from 0.25 nM to 500 nM (upper black line 500 nM, upper yellow line 250 nM, upper green line 125 nM, upper blue line 62.5, cyan line 31.25 nM, magenta line 15.62 nM, red line 7.81 nM, green line 3.90 nM. blue line 1.95 nM, black line 0.98 nM, yellow line 0.50 nM, and blue line 0.25 nM).
Figure 3. (a) The current signal versus time as a function of ATP concentration, for one of three sets of measurements. (b) The zoom-in of the region identified by the black dashed rectangle shown in (a). The bottom red curve is the dark current signal. The legend refers to the measured ATP concentrations, with values ranging from 0.25 nM to 500 nM (upper black line 500 nM, upper yellow line 250 nM, upper green line 125 nM, upper blue line 62.5, cyan line 31.25 nM, magenta line 15.62 nM, red line 7.81 nM, green line 3.90 nM. blue line 1.95 nM, black line 0.98 nM, yellow line 0.50 nM, and blue line 0.25 nM).
Sensors 24 03921 g003
Figure 4. The current reduction percentage calculated in the time range from 15 s to 35 s, as a function of ATP concentration. Each value (*) and its error bar have been obtained by considering the average and the standard deviation over three measurements taken at each ATP concentration, along with the noise associated with each measure. A linear fit has been performed (solid red line); “x” and “y” in the equation shown in the upper left of the figure represent the ATP concentration and the fit values, respectively.
Figure 4. The current reduction percentage calculated in the time range from 15 s to 35 s, as a function of ATP concentration. Each value (*) and its error bar have been obtained by considering the average and the standard deviation over three measurements taken at each ATP concentration, along with the noise associated with each measure. A linear fit has been performed (solid red line); “x” and “y” in the equation shown in the upper left of the figure represent the ATP concentration and the fit values, respectively.
Sensors 24 03921 g004
Figure 5. (a) The current signal as a function of time, with the time axis ranging between 0 s and 1.2 s, for different ATP concentrations. (b) The zoom-in of the region identified by the black dashed rectangle shown in (a) is reported. The bottom red curve in (a,b) indicates the dark noise signal. The fits performed on each curve by following Equation (2) are also reported as solid lines. ATP concentrations range from 0.25 nM to 500 nM. The set of measurements shown in Figure 3 is considered. To identify the curve color–ATP concentration relation, refer to the legends shown in Figure 3.
Figure 5. (a) The current signal as a function of time, with the time axis ranging between 0 s and 1.2 s, for different ATP concentrations. (b) The zoom-in of the region identified by the black dashed rectangle shown in (a) is reported. The bottom red curve in (a,b) indicates the dark noise signal. The fits performed on each curve by following Equation (2) are also reported as solid lines. ATP concentrations range from 0.25 nM to 500 nM. The set of measurements shown in Figure 3 is considered. To identify the curve color–ATP concentration relation, refer to the legends shown in Figure 3.
Sensors 24 03921 g005
Figure 6. Values of the estimated characteristic SRS diffusion time (τ) as a function of ATP concentration. Each value (*) and its error bar have been obtained by considering the average and the standard deviation over three measurements performed at each ATP concentration. The fit performed over the points (solid red curve) is also shown, suggesting a constant trend of the diffusion time with the concentration. The terms “x” and “y” in the equation shown in the upper left of the figure represent the ATP concentration and the fit values, respectively.
Figure 6. Values of the estimated characteristic SRS diffusion time (τ) as a function of ATP concentration. Each value (*) and its error bar have been obtained by considering the average and the standard deviation over three measurements performed at each ATP concentration. The fit performed over the points (solid red curve) is also shown, suggesting a constant trend of the diffusion time with the concentration. The terms “x” and “y” in the equation shown in the upper left of the figure represent the ATP concentration and the fit values, respectively.
Sensors 24 03921 g006
Figure 7. The calibration curve of the detection system described in this work. The vertical axis reports the difference between the measured maximum current signal and the dark noise (ΔI). Considering all three sets of measurements, the mean maximum current value and the respective error bar for each concentration have been estimated.
Figure 7. The calibration curve of the detection system described in this work. The vertical axis reports the difference between the measured maximum current signal and the dark noise (ΔI). Considering all three sets of measurements, the mean maximum current value and the respective error bar for each concentration have been estimated.
Sensors 24 03921 g007
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Capuano, G.E.; Corso, D.; Farina, R.; Pezzotti Escobar, G.; Screpis, G.A.; Coniglio, M.A.; Libertino, S. Miniaturizable Chemiluminescence System for ATP Detection in Water. Sensors 2024, 24, 3921. https://doi.org/10.3390/s24123921

AMA Style

Capuano GE, Corso D, Farina R, Pezzotti Escobar G, Screpis GA, Coniglio MA, Libertino S. Miniaturizable Chemiluminescence System for ATP Detection in Water. Sensors. 2024; 24(12):3921. https://doi.org/10.3390/s24123921

Chicago/Turabian Style

Capuano, Giuseppe E., Domenico Corso, Roberta Farina, Gianni Pezzotti Escobar, Giuseppe A. Screpis, Maria Anna Coniglio, and Sebania Libertino. 2024. "Miniaturizable Chemiluminescence System for ATP Detection in Water" Sensors 24, no. 12: 3921. https://doi.org/10.3390/s24123921

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop