Detecting Patient Position Using Bed-Reaction Forces for Pressure Injury Prevention and Management

1
KITE Research Institute, Toronto Rehabilitation Institute—University Health Network, Toronto, ON M5G 2A2, Canada
2
Institute of Biomedical Engineering, University of Toronto, Toronto, ON M5S 3G9, Canada
3
Department of Physical Therapy, University of Toronto, Toronto, ON M5G 1V7, Canada
4
Rehabilitation Sciences Institute, University of Toronto, Toronto, ON M5G 1V7, Canada
5
Vector Institute, Toronto, ON M5G 1M1, Canada
*
Author to whom correspondence should be addressed.
Sensors 2024, 24(19), 6483; https://doi.org/10.3390/s24196483
Submission received: 24 August 2024 / Revised: 21 September 2024 / Accepted: 1 October 2024 / Published: 9 October 2024
(This article belongs to the Collection Wearable Sensors for Risk Assessment and Injury Prevention)

Abstract

:
A key best practice to prevent and treat pressure injuries (PIs) is to ensure at-risk individuals are repositioned regularly. Our team designed a non-contact position detection system that predicts an individual’s position in bed using data from load cells under the bed legs. The system was originally designed to predict the individual’s position as left-side lying, right-side lying, or supine. Our previous work suggested that a higher precision for detecting position (classifying more than three positions) may be needed to determine whether key bony prominences on the pelvis at high risk of PIs have been off-loaded. The objective of this study was to determine the impact of categorizing participant position with higher precision using the system prediction F1 score. Data from 18 participants was collected from four load cells placed under the bed legs and a pelvis-mounted inertial measurement unit while the participants assumed 21 positions. The data was used to train classifiers to predict the participants’ transverse pelvic angle using three different position bin sizes (45°, ~30°, and 15°). A leave-one-participant-out cross validation approach was used to evaluate classifier performance for each bin size. Results indicated that our prediction F1 score dropped as the position category precision was increased.

1. Introduction

Pressure injuries (PIs), also known as bed sores or pressure ulcers, are defined as “localized damage to the skin and underlying soft tissue usually over a bony prominence or related to a medical or other device” [1]. PIs are largely preventable, yet they affect one in four Canadians across all healthcare settings [2]. The current practice for preventing PIs is to minimize the risk of prolonged tissue deformation by repositioning patients regularly to allow the compressed tissues to return to their normal state [1,3,4,5,6,7,8]. Unfortunately, evidence suggests that adherence rates to repositioning schedules are poor in clinical environments [6,7,9,10,11,12]. We expect that patients in the home are repositioned even less than in clinical environments due to lack of evidence-based educational materials and tools directed at unpaid home caregivers [13,14,15]. Furthermore, the unpredictability in the home environment may also pose a barrier to adherence to proper protocols and device use compared with a more controlled hospital environment [14]. One study showed that slightly less than one-third of patients were correctly using pressure-redistributing devices in the home environment [14].
The cost of PIs to the healthcare system is enormous, with the estimated cost currently exceeding USD 26.8 billion in the United States [16]. In the UK, approximately USD 4.5–5.1 billion is spent on wound management annually, and two-thirds of this cost is spent on treating wounds in the community setting [17]. Extrapolating from studies conducted in the UK, the average cost for treating PIs not requiring hospitalization is approximately USD 3000 per month in Canada, and the cost increases by an order of magnitude to approximately USD 30,000 per month if hospitalization is required [18,19,20,21,22].
A recent study from the United States found that the prevalence of all at-risk patients presenting to the hospital with a PI was 7.4% [23]. This value is higher than the 2.3% to 5.8% admission rates of two previous studies using electronic databases and greater than the 6.6% prevalence from a study using surveillance methodology [24,25,26]. In another study from the UK, it was found that 81% of patients presenting to the hospital with community-acquired PIs were presenting from a private residence [27].
Therefore, there is a strong need for a study of repositioning in the home environment for the prevention and management of PIs. However, existing methods for tracking repositioning are limited. Leveraging artificial intelligence (AI) could provide an easy solution to gather data and accurately track repositioning. AI has already been used in the clinical healthcare environment to enhance diagnostic accuracy, risk stratification, and treatment efficiency, as well as in the home environment to enhance precision monitoring for personalized medicine [28,29,30,31]. The use of load cells and AI, such as machine learning, may be able to provide a novel non-contact, unobtrusive measurement tool for monitoring patient position.
There are two commercially available technologies that are available for tracking patient positioning in bed in hospitals and nursing homes for PI prevention: pressure mats and body-worn inertial sensors. Pressure mats (e.g., Curiato, XSENSOR) can provide visualizations of the pressure distribution under a patient, which can be used to provide alerts to healthcare providers [6]. Single-use inertial motion sensors (e.g., Leaf Healthcare) attach to a patient’s chest to wirelessly monitor their position and provide an interface to indicate when a patient needs to be repositioned. Each of these technologies has its own limitations. Pressure mats can counteract the benefits of specialized mattresses and have been found to increase the risk of PI by 20% [32]. Pressure mats can also increase the risk of spreading bacteria from infected wounds [33]. Similarly, inertial sensors require attachment to the patient’s skin or clothing using adhesives or some other attachment method, which can be problematic. If adhesives are too strong, they can damage a frail patient’s skin. If the attachment method is not perfectly secure, a sensor can become detached and end up under the patient, potentially causing a PI.
In addition, there is a relatively new type of flexible electronics designed to track positioning during sleep that may be useful for pressure injury prevention [34,35,36]. These devices may have some of the same limitations as pressure mats, but may be beneficial if they can be made to be thin and easy to clean.

The Use of Loadcells to Detect and Classify Patient Movements

There has been some previous work using load cells and machine learning to classify patient movements in bed. Duvall et al. found that a K-nearest neighbor classification system was able to classify four movements in bed (rolls, turns in place, extremity movements, and caregiver-assisted turns) with 94.2% accuracy [37]. Similarly, Adami et al. classified movements into the three categories based on movements of the torso and lower extremities and achieved a 97% classification accuracy [38]. Alaziz et al. developed a load-cell-based system to detect and classify big and small movements using log-peak values and a threshold-based algorithm that achieved error rates of 6.3% and 4.2%, respectively [39]. This team classified nine classes of movements using a multi-level binary decision tree paired with a Support Vector Machine (SVM) at each decision step to achieve an accuracy of 90% and they applied SVM, random forest, and XGBoost techniques separately to classify the movements and then combined the techniques to achieve a 91.5% accuracy [40,41]. Minteer et al. also designed a system that can detect if a patient movement has occurred with 85% reliability; however, the system did not provide any additional information such as type of movement or what position the patient was in [42]. An important benefit of this study was that the system was tested on a real clinical population that was immobile and needed assistance in repositioning.
Overall, detecting movement in bed may be clinically useful, as there needs to be a detected movement for an individual’s position to change. However, these systems are limited as they have only demonstrated their ability to correctly classify a handful of predetermined movements. They may leave out multitudes of other movements that the system will need to contend with (e.g., dressing, changing incontinence briefs, etc.), which may cause false positive classifications of repositioning events. Furthermore, detecting a movement does not mean that an offloading event has occurred, as an individual could return to the same position and load the same tissues. Capturing an offloading event would ensure that the compressed tissue is allowed to return to its nature state, which decreases the risk for developing a PI.
Reflecting on the above, our team has developed a system that uses machine learning to detect the position of a patient based on data from load cells under each bed leg. We believe our approach has the potential to be more robust than previous movement-based approaches because it categorizes patient position when steady-state values are detected between movements (i.e., when the individual is at rest). This ensures that we capture offloading events because we are able to compare the current position to a previous position and decide if there has been enough of a change to offload the area of interest. Further, it is important to note that some positions are more likely to put patients at a higher risk of PIs than others. For instance, an individual that is lying on their side at 90° to the surface of the bed is likely to increase the forces acting on their greater trochanter compared to a position where the individual is lying on their side at a lower angle (e.g., 30–60° from supine).
Our proof-of-concept work was able to detect healthy participant position (supine, left-side, or right-side) with 94.2% accuracy (n = 20) [43]. However, when we tested this model on the data collected from nine older adults sleeping in their own beds at home, we noted there was a large variation in sleeping positions that were adopted at home that were not seen in the laboratory environment [44]. In particular, the larger variability in sleeping positions raised the question of whether it may be necessary to predict an individual’s position in bed with higher precision than only three categories (supine, left, or right). A companion paper will compare the results of this study with the ranges of positions for which key bony areas of the pelvis are off-loaded to determine the optimal position category precision that is needed [45]. Therefore, the objective of this study was to determine the impact of categorizing participant position with higher precision (i.e., more than three position categories) on the system prediction F1 score. We hypothesize that there will be a drop in prediction accuracy when an individual’s position is predicted with a higher precision than the three categories of left-side lying, right-side lying, and supine.

2. Materials and Methods

This study was approved by the Research Ethics Boards of University Health Network (Protocol# 17-5140).

2.1. Participants

A convenience sample of 20 healthy participants (10 males, 10 females) was recruited for this study. Two participants were excluded from the analysis due to an IMU malfunction, leaving data from a total of eighteen participants for data analysis. All participants provided informed consent.

2.2. System Setup

The instrumentation was set up in a similar manner to Wong et al. [43] and Gabison et al. [44]. Data was collected in a simulated patient care environment (CareLab) located at the KITE Research Institute, Toronto Rehabilitation Institute—University Health Network. Single-axis load cells comprised of four load sensors (model DLC902-30KG-HB, Hunan Detail Sensing Technology, Changsha, Hunan, China) arranged in a full Wheatstone bridge circuit were placed under each of the four wheels of a hospital bed (Carroll Hospital Group, Kalamazoo, MI, USA). Figure 1 and Figure 2 are visual representations of the system setup. The load cell signals were amplified, filtered, and converted from analog to digital using a signal conditioner (GEN 5, AMTI, Watertown, MA, USA) configured for 5.0 VDC excitation and a gain of 500 for each channel. NetForce software (version 3.5.2, AMTI, Watertown, MA, USA) running on a laptop PC (Thinkpad T520, Lenovo, Hong Kong, China, 2.5 GHz Intel Core i5 CPU, 4 GB of RAM) was used to collect the load cell data at 50 Hz with 16-bit resolution. An overhead video camera was also positioned above the bed to capture ground truth video of the participant positions.
Participants were fitted with two inertial measurement units (IMU) (Shimmer3, Shimmer Sensing, Dublin, Ireland) at the pelvis. One of these IMUs was used to log the ground truth pelvic angles and the other was used to view the angle in real-time to ensure the participant was positioned appropriately. Two sensors were required because our IMUs did not allow for logging and real-time viewing simultaneously. The IMU was connected to a laptop (Aspire 5, Acer, Xizhi, New Taipei, Taiwan, 1.8 GHz Intel Core i7 CPU, 12 GB of RAM) via Bluetooth to ConsensysPRO (version 1.6.0, Consensys, Dublin, Ireland) to collect data at 256 Hz.

2.3. Defining Participant Position While in Bed

The participant’s position in bed was defined by the angle of the pelvis in the transverse plane using an imaginary reference line perpendicular to the line connecting the left and right anterior superior iliac spines in the frontal plane. The angle between this reference line and a line perpendicular to the surface of the bed was defined as the transverse pelvic angle (TPA). This angle was defined such that the TPA would be 0° when participants were supine, a negative value when participants rolled onto their left side, and a positive value when they rolled onto their right side. Figure 3 shows a visual representation of the TPA.

2.4. Defining Bin Sizes

A total of three different bin sizes were used to test the effects of increased precision on performance (Figure 4). The 45° bin size is seen in Figure 4a, the 30° bin size is seen in Figure 4b, and the 15° bin size is seen in Figure 4c. For all data, the supine position was defined as any angle between −22.5° and +22.5°.

2.5. Data Collection

The data was collected in two phases. The position of participants was monitored using the real-time angle values from one of the IMUs attached to the participant’s pelvis. One of the researchers would gently reposition participants if they deviated more than ±5° from the target angle.
In Phase One, participants were instructed to cycle through a series of 11 unique TPA positions at 0°, ±15°, ±30°, ±45°, ±60°, and ±90°. In Phase Two, participants were instructed to assume any 10 random positions they chose from −180° to +180° to reflect the wide range of positions that can be adopted in bed.
All positions were held for three minutes. Figure 5 shows the order of positions participants were asked to assume, including the intermediate positions (when participants returned to supine) that were held for one minute. The intermediate positions served as a method of standardizing the insertion and removal of pillows for the primary phase.

2.6. Data Supplementation

The resulting dataset consisted of 2458 observations. Additional data was incorporated into the training set to increase the size of the data set. This additional data was only used to train machine and deep learning classifiers for the detection of supine, left, or right positions. Supplemental dataset A (n = 4910) was from the data collected by Wong et al. in a lab environment [43]. Supplemental dataset B (n = 13,913) was from data collected in the home environment by Gabison et al. [44]. The total dataset for Phase One included 21,281 observations.
In our Phase Two dataset, we include only the correct predictions from Phase One (n = 2002), and classify them into smaller Phase Two left (n = 763), supine (n = 622), and right (n = 617) datasets. In Phase Two, we focus only on the left and right datasets. Table 1 illustrates which datasets were used for Phase One and Phase Two classifications.

2.7. Data Processing

2.7.1. Load Cell Data

Load cell data was processed in the same manner as Wong et al. [43] and Gabison et al. [44]. Load cell data was exported from Netforce and processed using MATLAB 2020a (MathWorks, Natick, MA, USA). The data was manually sorted into trials by removing segments of data where positional changes occurred. The center of mass (CoM) of the bed and patient was calculated using Equations (1) and (2), where CoM_x and CoM_y reflect the CoM parallel to the width and length of the bed, respectively. In Equations (1) and (2), LH and RH represent the vertical forces experienced by the left and right sensors at the head of the bed, respectively; LF and RF represent the vertical forces by the left and right sensors at the foot of the bed, respectively; and l and w refer to the distance between the load cells along the length and width of the bed, respectively.
C o M _ x = w 2 × L H + L F R H R F L H + L F + R H + R F
C o M _ y = l 2 × L H + R H L F R F L H + L F + R H + R F
The CoM_x and CoM_y signals were low-pass-filtered with personalized Chebyshev Type II filters in order to isolate the changes in CoM signals related to respiration. The MATLAB filtfilt function was applied, which ensures a zero-phase shift, to obtain CoM_resp_x and CoM_resp_y. The maxima (tmax) and minima (tmin) in the CoM_resp_x and CoM_resp_y signals were identified by finding the zero crossings for the first derivative of each signal, where the maxima and minima coincide with the end of each exhalation and inhalation, respectively [46]. Equation (3) was used to calculate the angle of the principal axis of the ellipsoid traced by the resultant CoM_resp signal relative to the positive x axis for each tmax and subsequent tmin.
C o M _ r e s p _ A N G = a r c t a n C o M _ r e s p _ y ( t m a x ) C o M _ r e s p _ y ( t m i n ) C o M _ r e s p _ x ( t m a x ) C o M _ r e s p _ x ( t m i n )
Every data point used for the training and testing of the machine learning and deep learning classifiers was obtained by taking the average of a 45 s window, containing 2250 observations. New values were computed by shifting the window by 15 s. Approximately 10 data points were calculated from each pose for each participant. Missing data from one participant was interpolated based on values from the same window.

2.7.2. IMU Data

The ground truth data from the IMUs were classified into one of three positions (right-side lying, left-side lying, or supine) based on their Euler angle values. The IMU data were further classified four more times using the generated Euler angles, once for each of the different bin sizes, to allow for more precise TPA detection.

2.7.3. Supplemental Data

The supplemental data were previously classified into one of the three positions (right-side lying, left-side lying, or supine) by reviewing video data collected from supplemental dataset A and by three independent and blinded reviewers using time-lapsed images from supplemental dataset B.

2.8. Feature Selection

Table 2 shows the features extracted from load cell data that paralleled those used by Wong et al. [43] and Gabison et al. [44]. Feature importance was reassessed, showing that ratio_rmsCoM_resp, CoM_resp_ANG, and stdCoM_y were the three most important features. This finding is in line with our previous work that also demonstrated that the most important part of the signal from our load cells had to do with the small changes in the centre of mass of the participant that occurred with each breath. See Appendix A for a graphical representation that shows the relative importance of all 12 features.

2.9. Two-Phase Machine Learning Approach

In order to predict positions more precisely, a two-phase hierarchical classification approach was used, similar to the one used by Liang et al. [47]. In Phase One, we classified participant’s position as left-lying, supine, or right-lying, as previously defined in Section 2.3. Phase One was trained on the total data set and tested only on the PI dataset. Any incorrect classifications from Phase One were marked as incorrect and removed from the dataset. The remaining output of correct classifications from Phase One was divided into smaller datasets for left, supine, and right, as explained in Section 2.4, with the left and right datasets serving as the input for Phase Two.
We chose to remove the incorrect classifications from Phase Two as this is a pilot study that had limited data, so we prioritized the ability to predict position with more precision. Furthermore, as we defined the supine dataset to be a constant range, it could not be analyzed with more precision and was thus not used in Phase Two.
In Phase Two, the algorithm was used to predict the participant’s position more precisely. The side-lying positions were classified with increasing precision by dividing the range between ±22.5° and ±112.5° into two, three, and six bins, which correspond to bin sizes of 45°, 30°, and 15°, respectively. This can be seen in Figure 4a–c. The classification was repeated three times, once for each bin size. The range of positions was limited to positions between ±112.5° because there was not enough data to reliably predict positions outside of this range.
By splitting the classification into two phases, it allowed additional data to be used as Phase One classification did not require associated IMU data. In this way, the model was able to be trained on more data to narrow down a portion of its classification task.
Please see Appendix A for a detailed example of how Phase One and Phase Two classification are integrated.

2.9.1. Leave-One-Participant-Out

A leave-one-participant-out cross validation approach was used to evaluate the accuracy of the classifier while maximizing the number of training observations. Using this method, a classifier was trained on the new data set that incorporated 17 participants and tested on the 1 excluded participant. This procedure was repeated 18 times, once for each participant. The overall performance measures were estimated from the averaged errors for each individual test sample.

2.9.2. Incremental Learning

Incremental learning was used to evaluate the potential of the classifier to adapt to the left-out participant. The classifier was trained using different percentages of the left-out participant’s data (c = 0%, 10%, 20%, 30%). To maintain a uniform test set, the left-out participant’s data was split into a 70% test set and a 30% incremental learning set, from which different c values were added to the training set. The best performing incremental learning level (ILL) was used to create the smaller left and right datasets for the more precise step two classification.

2.9.3. Machine Learning Classifiers

Table 3 shows a list of models used in both phase one and two classifications. Appendix B contains tables that describe the architectures of all the created models.

2.10. Statistical Analysis

A Shapiro–Wilk test and a Levene test were used in all cases to assess the normality and the homogeneity of the data, respectively. Both were not satisfied, resulting in non-parametric tests being used.

2.10.1. Phase One Classification

A Friedman’s ANOVA was used to identify any significant differences in the mean F1 score and accuracy based on the best performing ILL. Post hoc Wilcoxon Rank Sum tests with Bonferroni corrections for three comparisons (p < 0.0167) were used to compare the top three performing models.
A Friedman’s ANOVA was also used to compare the ILLs for each of the top models to determine their impact on performance. Post hoc Wilcoxon Rank Sum tests with Bonferroni corrections for three comparisons (significance needed p < 0.0167) were used to compare each incremental learning level to its adjacent value(s) (i.e., 0% to 10%, 10% to 20%, and 20% to 30%).

2.10.2. Phase Two Classification

ANOVAs were used to compare the classification results to determine the effect of bin size on model performance for left and right classification. The best classifier from each bin size was compared to its adjacent bin size(s) (i.e., 45° to 30°, and 30° to 15°). Post hoc Wilcoxon Rank Sum tests with Bonferroni corrections for two comparisons (p < 0.025), were used to compare the top three performing models.

3. Results

3.1. Participants

Descriptive statistics of all the participants recruited for this study are shown in Table 4.

3.2. Phase One Classification

The best-performing model was the LGB classifier with a F1 score of 0.8343 ± 0.1362 at an ILL of 30%. All other model scores and associated graphs are found in Appendix C.

3.2.1. Comparing Machine Learning Models

The data was confirmed to be non-parametric with a significant difference between the mean F1 score, χ2(8) = 96.398, p < 2.2 × 10−16. The top three performing classifiers were LGB, GBC, and MLP 2. The post hoc comparisons with the LGB model were found to be statistically significant: LGB vs. MLP 2: V = 27, p = 0.0115; LGB vs. GBC: V = 156, p = 0.0023; GBC vs. MLP 2: V = 69, p = 0.740.

3.2.2. Comparing Incremental Learning Levels

As the LGB model was shown to be the best model, the ILLs were compared. The LGB data was confirmed to be non-parametric with a significant difference between the mean F1 score, χ2(3) = 28.787, p < 2.483 × 10−6. Appendix C contains a graphical comparison between the mean F1 score of the different ILLs for the LGB model. Two out of three post hoc comparisons were found to be statistically significant, 0% vs. 10%: V = 10, p = 0.00109; 10% vs. 20%: V = 40, p = 0.155; 20% vs. 30%: V = 13, p = 0.00172. Therefore, the LGB classifier at an ILL of 30% was the best-performing model. It was used to create the left and right data sets for Phase Two.

3.3. Phase Two Classification

Table 5 shows the overall mean F1 score, accuracy, and standard deviation values for the best model, LGBoost, across all 18 participants from Phase Two left and right classification using an ILL of 30%.

3.3.1. Comparing Left Bin Sizes

The data for Phase Two left classification was confirmed to be non-parametric with significant differences between the F1 scores of different bin sizes, χ2(1) = 54, p < 2.00 × 10−13. Figure 6 shows a visual comparison between the different bin sizes for left-sided lying.
Post hoc comparisons were found to be statistically significant for 45° vs. 30° bin size, V = 147, p = 0.00789, but not for 30° vs. 15° bin size, V = 108, p = 0.142.

3.3.2. Comparing Right Bin Sizes

The data for Phase Two right classification was confirmed to be non-parametric with significant differences between the F1 scores of different bin sizes, χ2(1) = 54, p < 2.00 × 10−13. Figure 7 shows a visual comparison between the bin sizes.
Post hoc comparisons for mean F1 was found to be statistically significant for 45° vs. 30° bin size, V = 118, p = 0.0105, but not for 30° vs. 15° bin size, V = 111, p = 0.0280.

4. Discussion

This study aimed to improve our non-contact position detection system for preventing PIs by increasing the precision of position classification. By analyzing data from load cells and IMUs, we assessed how classifying individual positions with finer granularity (45°, ~30°, and 15°) impacts the system’s predictive accuracy. The study found that higher precision in detecting positions led to a decrease in the F1 score of the system. This highlights a critical trade-off between detailed position tracking and predictive reliability, contributing valuable insights into optimizing PI prevention technologies.

4.1. Phase One Classification

The best performing model was the LGBoost. This F1 score is an improvement over the F1 score of 0.6841 that Wong et al.’s previous model was able to achieve on this data [43]. It is likely that the F1 score represents an underestimate of the true F1 score of our new system as the training set includes data without the IMU ground truth data for Phase One. Additionally, we combined data from different settings, which may have negatively impacted the performance of our model as there was more variability to be accounted for by the model, such as the type of bed, size of bed, and carpeted floors. This likely explains why Wong et al.’s model was able to perform well on their lab-controlled dataset but had its performance drop significantly in this mixed dataset.
The Leaf Healthcare system, which attaches an IMU to the patient’s body, will consistently record a real-time angle for patient position. However, this comes at the cost of having an adhesive placed on the skin that could cause problems itself. Further, Leaf Healthcare’s system is advertised as placing the sensor on a patient’s chest, which may not accurately represent position changes and offloading as the hips can rotate independently of the chest. By focusing on predicting the TPA, we believe that our system would more accurately reflect the true position needed to be monitored for offloading.

Comparing Incremental Learning Levels

Our results suggest that collecting data for incremental learning has the potential to better personalize the model to individuals at high risk of developing PIs, with the potential to improve their care. It is important to note that the test set for each participant in the study was limited, ranging from 81 to 206 observations (mean 136.6 observations), meaning that a maximum of 61 observations at 30% incremental learning were added to a data pool of ~20,000 observations for incremental learning. Considering that the incremental learning, with so few observations, was able to improve the overall F1 score by slightly over 0.07, for the LGB model, it would be important to further investigate the impacts of incremental learning with a larger data set. Additionally, it would be important to investigate whether the statistical significance of incremental learning translates to clinical significance and an improved prevention of PIs.

4.2. Phase Two Classification

The LGB model performed significantly better than other models for all test cases in Phase Two. As we removed the incorrect classifications from Phase One before feeding the data into Phase Two, it is important to interpret the F1 score and accuracy as an overestimation. To get a hypothetical maximum F1 score, we would need to multiply the F1 score from Phase One by the F1 scores from Phase Two. This results in a hypothetical maximum F1 score of 0.767 for left using 45° bins and our hypothetical minimum of 0.617 for right using 15° bins. It is difficult to compare Phase Two to other position-based studies as they focused only on predicting left, right, or supine positions, whereas our study increases the precision. As hypothesized, this inherently decreased our F1 score and accuracy compared to other studies. We also identified a trend of left-side positions being predicted correctly slightly more often than right-side positions. The reason for this finding is currently unclear.

Comparing Bin Sizes

We noted that with increasing number of bin sizes, the accuracy of prediction decreased. However, this was only significant in the Level 2 left and right mean F1 score between 45° vs. 30° bin sizes. It was also significant for left mean accuracy between 45° vs. 30° and for right mean accuracy between 30° vs. 15° bin sizes. These results are important because they indicate that there is likely a trade-off between accuracy and precision when making predictions on estimating an individual’s position in bed. It will be important to optimize the bin size for this system to ensure it is recording and classifying movements of interest. The bin size should be optimized based on information gathered from offloading data and clinical expertise to decide what is the smallest positional change that needs to be captured to offload areas on the pelvis that are at high risk of developing PIs. Our companion paper uses offloading data to draw conclusions about the optimal bin size [45].

4.3. Study Limitations and Future Work

This study included some technical limitations. The load cell data was filtered using customized parameters for each participant, which may have led to an increased performance compared to generic parameters. Most training data did not contain IMU ground truth, which may have negatively impacted accuracy. We found that our sample size of 18 and the number of data points (~2500) was too small to perform meaningful one-shot learning or regression analysis. Additionally, the small sample size resulted in removing incorrect predictions from Phase One before feeding the data into Phase Two as we did not have enough data to learn from the incorrect predictions. There was also an imbalance of positions recorded, resulting in fewer positions that were >90° and <−90° for Phase Two classification, which could have negatively impacted performance for predicting positions within that range. Lastly, the hyperparameter tuned architecture was overly complicated, with too many layers for our dataset, and may have overfit the model.
Most of the above limitations can be addressed by collecting additional data. As such, our future work will include collecting longer-duration overnight data with participants in their own homes. After collecting additional data, we plan to treat the detection of patient position as a regression task instead of a classification task and compare their respective performances to identify which approach is better. A regression analysis would strengthen our ability to predict the actual angle rather than an angle bin, which may be exaggerating the errors from our system. Moreover, with additional data, we would be able to learn from incorrect predictions to improve the algorithm’s performance, allowing us to determine the true maximum F1 score attainable. Finally, we plan on further developing this system to allow for real-time prompting so that the device can be implemented in the home setting to monitor offloading.

5. Conclusions

The results of this study showed that when trying to predict position more precisely than left side-lying, right side-lying, or supine, the prediction mean F1 score decreased from 0.919 and 0.900 for the left and right ~45° bin size (91.5% and 89.4% accuracy), respectively, to 0.786 and 0.740 for the 15° bins (81.5% and 75.4% accuracy), respectively.
These results demonstrate that our system can track and predict an individual’s positioning in bed and needs to be further tested in a home care setting to evaluate its ability in preventing PIs. This paper should be interpreted along with its companion paper that expands on these results from a wound management perspective [45].

Author Contributions

Conceptualization, S.G., G.F. and T.D.; methodology, N.P., S.G., G.F. and T.D.; software, N.P., G.E. and E.D.; validation, N.P.; formal analysis, N.P.; investigation, N.P.; resources, S.G., G.E., G.F. and T.D.; data curation, N.P. and G.E.; writing—original draft preparation, N.P., S.G. and T.D.; writing—review and editing, N.P., S.G., G.E., G.F., E.D. and T.D.; visualization, N.P., S.G., G.E. and T.D.; supervision, T.D.; project administration, G.F., S.G. and T.D.; funding acquisition, N.P., S.G., G.F., E.D. and T.D. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Hallisey Family, and the Canadian Institutes of Health Research (Canadian Graduate Scholarship held by N.P. and Project Grant #202309PJT-508073-BME-ADHD-249077 held by T.D and S.G.).

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Research Ethics Board of University Health Network (study protocol 17-5140 approved on 2 February 2018).

Informed Consent Statement

Written informed consent was obtained from all participants involved in the study.

Data Availability Statement

Data is unavailable due to the restrictions of our research ethics approval.

Acknowledgments

The authors thank Gordon Wong for his contributions to earlier versions of the data collection hardware and Davood Dadkhah for his contributions to developing the hardware and software used in the study that made this work possible.

Conflicts of Interest

The authors declare no conflicts of interest.

Appendix A

Appendix A shows a graph of the relative feature importance.
Figure A1. Graph of relative feature importance for features used to predict patient position.
Figure A1. Graph of relative feature importance for features used to predict patient position.
Sensors 24 06483 g0a1

Appendix B

Appendix A illustrates how Phase One and Phase Two classification are integrated. For example, the 45° bin size has five possible bins to classify a position. By splitting the new data set into smaller left and right datasets, the number of possible bins a given position can be classified as has been decreased by over 50% for both left and right prediction. As seen in Figure A1, for the 45° bin size, the algorithm must now predict between two possible classifications for both left-side and right-side lying, rather than having to use the limited IMU data available to directly predict which of the five bins a position belongs to.
Figure A2. Flow diagram of how the two steps of hierarchical classification work together, where the second level shows the example of a bin size of 45° [43].
Figure A2. Flow diagram of how the two steps of hierarchical classification work together, where the second level shows the example of a bin size of 45° [43].
Sensors 24 06483 g0a2

Appendix C

Table A1, Table A2 and Table A3 describe the different model architectures for the MLP models. Three different MLP models were constructed to test Phase One classification, where MLP 1 was the original model used by Wong et al. [43]; MLP 2 was a large, hyperparameter tuned model using keras.tuner; and MLP 3 was a simple MLP model. Table A4 and Table A5 describe the architecture of the CNN and RNN, respectively.
Table A1. Architecture of MLP 1 (the original MLP model used by Wong et al.). The dropout was set to 0.1 for all instances, the batch size was not specified, and the learning rate was 0.001.
Table A1. Architecture of MLP 1 (the original MLP model used by Wong et al.). The dropout was set to 0.1 for all instances, the batch size was not specified, and the learning rate was 0.001.
LayersNumber of NodesActivation Function
Input12ReLu
Fully Connected 164ReLu
Dropout 1N/AN/A
Batch Normalization 1N/AN/A
Fully Connected 2100ReLu
Dropout 2N/AN/A
Batch Normalization 2N/AN/A
Output3Softmax
Table A2. Architecture of MLP 2, the hyperparameter tuned model. The dropout was set to 0.1 for all instances, the batch size was 128, and the learning rate was 0.000417.
Table A2. Architecture of MLP 2, the hyperparameter tuned model. The dropout was set to 0.1 for all instances, the batch size was 128, and the learning rate was 0.000417.
LayersNumber of NodesActivation Function
Input12ReLu
Fully Connected 1128ReLu
Dropout 1 N/AN/A
Batch Normalization 1N/AN/A
Fully Connected 2256ReLu
Dropout 2N/AN/A
Batch Normalization 2N/AN/A
Fully Connected 3160ReLu
Dropout 3N/AN/A
Batch Normalization 3N/AN/A
Fully Connected 432ReLu
Dropout 4N/AN/A
Batch Normalization 4N/AN/A
Output3Softmax
Table A3. Architecture of MLP 3, the simplified tuned model. The dropout was set to 0.1 for all instances, the batch size was 32, and the learning rate was 0.001.
Table A3. Architecture of MLP 3, the simplified tuned model. The dropout was set to 0.1 for all instances, the batch size was 32, and the learning rate was 0.001.
LayersNumber of NodesActivation Function
Input12ReLu
Fully Connected 112ReLu
Dropout 1N/AN/A
Batch Normalization 1N/AN/A
Fully Connected 26ReLu
Dropout 2N/AN/A
Batch Normalization 2N/AN/A
Fully Connected 34ReLu
Dropout 3N/AN/A
Batch Normalization 3N/AN/A
Output3Softmax
Table A4. Architecture of the CNN model. The dropout was set to 0.2, 0.0, 0.05, and 0.45 for the four layers, respectively; the batch size was 16, and the learning rate was 0.00188.
Table A4. Architecture of the CNN model. The dropout was set to 0.2, 0.0, 0.05, and 0.45 for the four layers, respectively; the batch size was 16, and the learning rate was 0.00188.
LayerNumber of FiltersSize/StridePaddingActivation Function
Input13N/AN/AReLu
Convolutional 1321/2SameReLu
Dropout 1N/AN/AN/AN/A
Batch Normalization 1N/AN/AN/AN/A
Convolutional 2327/4SameReLu
Dropout 2N/AN/AN/AN/A
Batch Normalization 2N/AN/AN/AN/A
Convolutional 3325/2SameReLu
Dropout 3N/AN/AN/AN/A
Batch Normalization 3N/AN/AN/AN/A
Max Pooling 1N/A4/3SameN/A
Flatten 1N/AN/AN/AN/A
Fully Connected 1224N/AN/AReLu
Dropout 4N/AN/AN/AN/A
Batch Normalization 4N/AN/AN/AN/A
OutputN/AN/AN/ASoftmax
Table A5. Architecture of the RNN model. The dropout was set to 0.35 for all instances, the batch size was 512, and the learning rate was 0.000933.
Table A5. Architecture of the RNN model. The dropout was set to 0.35 for all instances, the batch size was 512, and the learning rate was 0.000933.
LayerNumber of NodesActivation Function
Input13ReLu
LSTM 196ReLu
Dropout 1N/AN/A
Batch Normalization 1N/AN/A
LSTM 2128ReLu
Dropout 2N/AN/A
Batch Normalization 2N/AN/A
LSTM 3256ReLu
Dropout 3N/AN/A
Batch Normalization 3N/AN/A
LSTM 4224ReLu
Dropout 4N/AN/A
Batch Normalization 4N/AN/A
Fully Connected 132ReLu
Dropout 5N/AN/A
Batch Normalization 5N/AN/A
Output3Softmax

Appendix D

Appendix D contains information related to Phase One of classification. Table A6 describes the mean F1 scores of all models in Phase One. Figure A3 visually depicts the mean F1 scores for an ILL of 30% across all models. Figure A4 visually shows the difference between ILLs for the best-performing model, LGB.
Table A6. Combined mean F1 score and standard deviations of the tested models for Phase One of the classifications of all 18 participants, which classifies positions as supine, left-side lying, or right-side lying. The bolded value indicates the best-performing model.
Table A6. Combined mean F1 score and standard deviations of the tested models for Phase One of the classifications of all 18 participants, which classifies positions as supine, left-side lying, or right-side lying. The bolded value indicates the best-performing model.
Modelc = 0%c = 10%c = 20%c = 30%
ADA Mean0.741 ± 0.1660.743 ± 0.1600.740 ± 0.1640.742 ± 0.168
GBC Mean0.770 ± 0.1670.775 ± 0.1690.773 ± 0.1720.781 ± 0.165
LGB Mean0.762 ± 0.1620.791 ± 0.1470.801 ± 0.1490.834 ± 0.132
XGB Mean0.767 ± 0.1730.770 ± 0.1710.772 ± 0.1690.775 ± 0.172
MLP 1 Mean0.774 ± 0.1590.754 ± 0.1720.762 ± 0.1540.780 ± 0.158
MLP 2 Mean0.747 ± 0.1650.769 ± 0.1530.784 ± 0.1560.782 ± 0.159
MLP 3 Mean0.764 ± 0.1960.762 ± 0.1720.761 ± 0.1790.753 ± 0.203
LSTM Mean0.137 ± 0.03350.143 ± 0.03400.151 ± 0.05250.128 ± 0.0209
CNN Mean0.221 ± 0.05110.203 ± 0.04230.200 ± 0.04110.189 ± 0.0299
Figure A3. Phase One classification mean F1 scores for an ILL of 30%. The error bars represent standard deviation.
Figure A3. Phase One classification mean F1 scores for an ILL of 30%. The error bars represent standard deviation.
Sensors 24 06483 g0a3
Figure A4. Phase One classification mean F1 scores for the top three models across the different ILLs. The error bars represent standard deviation.
Figure A4. Phase One classification mean F1 scores for the top three models across the different ILLs. The error bars represent standard deviation.
Sensors 24 06483 g0a4

References

  1. Edsberg, L.E.; Black, J.M.; Goldberg, M.; McNichol, L.; Moore, L.; Sieggreen, M. Revised National Pressure Ulcer Advisory Panel Pressure Injury Staging System: Revised Pressure Injury Staging System. J. Wound Ostomy Cont. Nurs. 2016, 43, 585. [Google Scholar] [CrossRef] [PubMed]
  2. Perez, E. Pressure Ulcers: Updated Guidelines for Treatment and Prevention. Geriatrics 1993, 48, 39–44. [Google Scholar] [PubMed]
  3. Gefen, A. The Future of Pressure Ulcer Prevention Is Here: Detecting and Targeting Inflammation Early. EWMA J. 2018, 19, 7–13. [Google Scholar]
  4. Lustig, M.; Wiggermann, N.; Gefen, A. How Patient Migration in Bed Affects the Sacral Soft Tissue Loading and Thereby the Risk for a Hospital-Acquired Pressure Injury. Int. Wound J. 2020, 17, 631–640. [Google Scholar] [CrossRef]
  5. Gefen, A.; Soppi, E. What Is New in Our Understanding of Pressure Injuries: The Inextricable Association between Sustained Tissue Deformations and Pain and the Role of the Support Surface. Wound Pract. Res. 2020, 28, 58–66. [Google Scholar] [CrossRef]
  6. Pompeo, M.Q. Pressure Map Technology for Pressure Ulcer Patients: Can We Handle the Truth? Wounds Compend. Clin. Res. Pract. 2013, 25, 34–40. [Google Scholar]
  7. Stockton, L.; Parker, D. Pressure Relief Behaviour and the Prevention of Pressure Ulcers in Wheelchair Users in the Community. J. Tissue Viability 2014, 12, 84–99. [Google Scholar] [CrossRef]
  8. Ontario, H.Q. Pressure Injuries. Care for Patients in All Settings. Qual. Stand. 2017, 51. [Google Scholar]
  9. Lyder, C.H.; Preston, J.; Grady, J.N.; Scinto, J.; Allman, R.; Bergstrom, N.; Rodeheaver, G. Quality of Care for Hospitalized Medicare Patients at Risk for Pressure Ulcers. Arch. Intern. Med. 2001, 161, 1549–1554. [Google Scholar] [CrossRef] [PubMed]
  10. Krishnagopalan, S.; Johnson, E.W.; Low, L.L.; Kaufman, L.J. Body Positioning of Intensive Care Patients: Clinical Practice versus Standards. Crit. Care Med. 2002, 30, 2588–2592. [Google Scholar] [CrossRef]
  11. Renganathan, B.S.; Preejith, S.P.; Nagaiyan, S.; Joseph, J.; Sivaprakasam, M. A Novel System to Tackle Hospital Acquired Pressure Ulcers. In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Orlando, FL, USA, 15–19 July 2016; EMBS: Piscataway, NJ, USA, 2016; Volume 2016-Octob, pp. 4780–4783. [Google Scholar]
  12. Takashini, M.; Black, J.; Dealey, C.; Gefen, A. Pressure in Context. In International review: Pressure ulcer precention: Pressure, shear, friction and microclimate in context. A consensus document. Wounds Int. 2010. [Google Scholar]
  13. Kwiczala-Szydłowska, S.; Skalska, A.; Grodzicki, T. Pressure Ulver Prevention-Evaluation of Awarness in Families of Patients at Risk. Prz. Lek. 2005, 62, 1393–1397. [Google Scholar]
  14. Jackson, D.; Durrant, L.; Bishop, E.; Walthall, H.; Betteridge, R.; Gardner, S.; Coulton, W.; Hutchinson, M.; Neville, S.; Davidson, P.M.; et al. Health Service Provision and the Use of Pressure-Redistributing Devices: Mixed Methods Study of Community Dwelling Individuals with Pressure Injuries. Contemp. Nurse 2017, 53, 378–389. [Google Scholar] [CrossRef] [PubMed]
  15. Jankowski, I.M.; Nadzam, D.M. Identifying Gaps, Barriers, and Solutions in Implementing Pressure Ulcer Prevention Programs. Jt. Comm. J. Qual. Patient Saf. 2011, 37, 253–264. [Google Scholar] [CrossRef] [PubMed]
  16. Padula, W.V.; Delarmente, B.A. The National Cost of Hospital-Acquired Pressure Injuries in the United States. Int. Wound J. 2019, 16, 634–640. [Google Scholar] [CrossRef]
  17. Guest, J.F.; Ayoub, N.; McIlwraith, T.; Uchegbu, I.; Gerrish, A.; Weidlich, D.; Vowden, K.; Vowden, P. Health Economic Burden That Different Wound Types Impose on the UK’s National Health Service. Int. Wound J. 2017, 14, 322–330. [Google Scholar] [CrossRef] [PubMed]
  18. Bennett, G.; Dealey, C.; Posnett, J. The Cost of Pressure Ulcers in the UK. Age Ageing 2004, 33, 230–235. [Google Scholar] [CrossRef]
  19. Guest, J.F.; Fuller, G.W.; Vowden, P.; Vowden, K.R. Cohort Study Evaluating Pressure Ulcer Management in Clinical Practice in the UK Following Initial Presentation in the Community: Costs and Outcomes. BMJ Open 2018, 8, e021769. [Google Scholar] [CrossRef]
  20. Dealey, C.; Posnett, J.; Walker, A. The Cost of Pressure Ulcers in the United Kingdom. J. Wound Care 2012, 21, 261–266. [Google Scholar] [CrossRef]
  21. Norton, L.; Coutts, P.; Sibbald, R.G. Beds: Practical Pressure Management for Surfaces/Mattresses. Adv. Skin Wound Care 2011, 24, 324–332. [Google Scholar] [CrossRef]
  22. Chan, B.C.; Nanwa, N.; Mittmann, N.; Bryant, D.; Coyte, P.C.; Houghton, P.E. The Average Cost of Pressure Ulcer Management in a Community Dwelling Spinal Cord Injury Population. Int. Wound J. 2013, 10, 431–440. [Google Scholar] [CrossRef] [PubMed]
  23. Corbett, L.Q.; Funk, M.; Fortunato, G.; O’Sullivan, D.M. Pressure Injury in a Community Population: A Descriptive Study. J. Wound. Ostomy Cont. Nurs. 2017, 44, 221–227. [Google Scholar] [CrossRef]
  24. Lyder, C.H.; Wang, Y.; Metersky, M.; Curry, M.; Kliman, R.; Verzier, N.R.; Hunt, D.R. Hospital-Acquired Pressure Ulcers: Results from the National Medicare Patient Safety Monitoring System Study. J. Am. Geriatr. Soc. 2012, 60, 1603–1608. [Google Scholar] [CrossRef] [PubMed]
  25. Meddings, J.; Reichert, H.; Rogers, M.A.M.; Hofer, T.P.; McMahon, L.F.; Grazier, K.L. Under Pressure: Financial Effect of the Hospital-Acquired Conditions Initiative-a Statewide Analysis of Pressure Ulcer Development and Payment. J. Am. Geriatr. Soc. 2015, 63, 1407–1412. [Google Scholar] [CrossRef]
  26. Keelaghan, E.; Margolis, D.; Zhan, M.; Baumgarten, M. Prevalence of Pressure Ulcers on Hospital Admission among Nursing Home Residents Transferred to the Hospital. Wound Repair Regen. 2008, 16, 331–336. [Google Scholar] [CrossRef] [PubMed]
  27. Worsley, P.R.; Smith, G.; Schoonhoven, L.; Bader, D.L. Characteristics of Patients Who Are Admitted with or Acquire Pressure Ulcers in a District General Hospital; a 3 Year Retrospective Analysis. Nurs. Open 2016, 3, 152–158. [Google Scholar] [CrossRef]
  28. Topol, E.J. High-Performance Medicine: The Convergence of Human and Artificial Intelligence. Nat. Med. 2019, 25, 44–56. [Google Scholar] [CrossRef]
  29. Jiang, F.; Jiang, Y.; Zhi, H.; Dong, Y.; Li, H.; Ma, S.; Wang, Y.; Dong, Q.; Shen, H.; Wang, Y. Artificial Intelligence in Healthcare: Past, Present and Future. Stroke Vasc. Neurol. 2017, 2, 230. [Google Scholar] [CrossRef]
  30. Karalis, V.D. The Integration of Artificial Intelligence into Clinical Practice. Appl. Biosci. 2024, 3, 14–44. [Google Scholar] [CrossRef]
  31. Ma, Z.; Hua, H.; You, C.; Ma, Z.; Guo, W.; Yang, X.; Qiu, S.; Zhao, N.; Zhang, Y.; Ho, D.; et al. FlexiPulse: A Machine-Learning-Enabled Flexible Pulse Sensor for Cardiovascular Disease Diagnostics. Cell Rep. Phys. Sci. 2023, 4, 101690. [Google Scholar] [CrossRef]
  32. Kayser, S.A.; VanGilder, C.A.; Lachenbruch, C. Predictors of Superficial and Severe Hospital-Acquired Pressure Injuries: A Cross-Sectional Study Using the International Pressure Ulcer PrevalenceTM Survey. Int. J. Nurs. Stud. 2019, 89, 46–52. [Google Scholar] [CrossRef] [PubMed]
  33. Muder, R.R.; Brennen, C.; Wagener, M.M. Bacteremia in a Long-Term-Care Facility: A Five-Year Prospective Study of 163 Consecutive Episodes. Clin. Infect. Dis. 1992, 14, 647–654. [Google Scholar] [CrossRef] [PubMed]
  34. Jiang, Z.; Lee, Y.S.; Wang, Y.; John, H.; Fang, L.; Tang, Y. Advancements in Flexible Sensors for Monitoring Body Movements during Sleep: A Review. Sensors 2024, 24, 5091. [Google Scholar] [CrossRef] [PubMed]
  35. Zhang, Y.; Xiao, A.; Zheng, T.; Xiao, H.; Huang, R. The Relationship between Sleeping Position and Sleep Quality: A Flexible Sensor-Based Study. Sensors 2022, 22, 6220. [Google Scholar] [CrossRef]
  36. Li, Y.; Zhang, P.; Wang, S.; Su, T.; Zhang, K.; Huang, L.; Wu, S. Sleep State Monitoring System Based on Flexible Pressure Sensor Arrays for Evaluating Sleep Quality. IEEE Sens. J. 2024, 1. [Google Scholar] [CrossRef]
  37. Duvall, J.; Karg, P.; Brienza, D.; Pearlman, J. Detection and Classification Methodology for Movements in the Bed That Supports Continuous Pressure Injury Risk Assessment and Repositioning Compliance. J. Tissue Viability 2019, 28, 7–13. [Google Scholar] [CrossRef]
  38. Adami, A.M.; Adami, A.G.; Hayes, T.L.; Beattie, Z.T. A System for Assessment of Limb Movements in Sleep. In Proceedings of the 2013 IEEE 15th International Conference on e-Health Networking, Applications and Services (Healthcom 2013), Lisbon, Portugal, 9–12 October 2013; pp. 419–423. [Google Scholar] [CrossRef]
  39. Alaziz, M.; Jia, Z.; Liu, J.; Howard, R.; Chen, Y.; Zhang, Y. Motion Scale: A Body Motion Monitoring System Using Bed-Mounted Wireless Load Cells. In Proceedings of the 2016 IEEE First International Conference on Connected Health: Applications, Systems and Engineering Technologies (CHASE), Washington, DC, USA, 27–29 June 2016; pp. 183–192. [Google Scholar] [CrossRef]
  40. Alaziz, M.; Jia, Z.; Howard, R.; Lin, X.; Zhang, Y. MotionTree: A Tree-Based In-Bed Body Motion Classification System Using Load-Cells. In Proceedings of the 2017 IEEE/ACM International Conference on Connected Health: Applications, Systems and Engineering Technologies (CHASE), Philadelphia, PA, USA, 17–19 June 2017; pp. 127–136. [Google Scholar] [CrossRef]
  41. Alaziz, M.; Jia, Z.; Howard, R.; Lin, X.; Zhang, Y. In-Bed Body Motion Detection and Classification System. ACM Trans. Sens. Netw. 2020, 16, 1–26. [Google Scholar] [CrossRef]
  42. Minteer, D.M.; Simon, P.; Taylor, D.P.; Jia, W.; Li, Y.; Sun, M.; Rubin, J.P. Pressure Ulcer Monitoring Platform—A Prospective, Human Subject Clinical Study to Validate Patient Repositioning Monitoring Device to Prevent Pressure Ulcers. Adv. Wound Care 2020, 9, 28–33. [Google Scholar] [CrossRef]
  43. Wong, G.; Gabison, S.; Dolatabadi, E.; Evans, G.; Kajaks, T.; Holliday, P.; Alshaer, H.; Fernie, G.; Dutta, T. Toward Mitigating Pressure Injuries: Detecting Patient Orientation from Vertical Bed Reaction Forces. J. Rehabil. Assist. Technol. Eng. 2020, 7, 205566832091216. [Google Scholar] [CrossRef] [PubMed]
  44. Gabison, S.; Pupic, N.; Evans, G.; Dolatabadi, E.; Dutta, T. Measuring Repositioning in Home Care for Pressure Injury Prevention and Management. Sensors 2022, 22, 7013. [Google Scholar] [CrossRef]
  45. Pupic, N.; Gabison, S.; Evans, G.; Fernie, G.; Dolatabadi, E.; Dutta, T. Detecting Patient Position Using Bed-Reaction Forces and Monitoring Skin-Bed Interface Forces for Pressure Injury Prevention and Management. medRxiv 2022. [Google Scholar] [CrossRef]
  46. Beattie, Z.T.; Hagen, C.C.; Hayes, T.L. Classification of Lying Position Using Load Cells under the Bed. In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Boston, MA, USA, 30 August–3 September 2011; pp. 474–477. [Google Scholar] [CrossRef]
  47. Liang, H.; Tsui, B.Y.; Ni, H.; Valentim, C.C.S.; Baxter, S.L.; Liu, G.; Cai, W.; Kermany, D.S.; Sun, X.; Chen, J.; et al. Evaluation and Accurate Diagnoses of Pediatric Diseases Using Artificial Intelligence. Nat. Med. 2019, 25, 433–438. [Google Scholar] [CrossRef] [PubMed]
Figure 1. The data collection setup. Four load cells were placed under the legs of the bed. Signals from the load cell were amplified, filtered, and converted from analog to digital using a signal conditioner. Data was saved on a laptop (PC2). IMU data was collected to PC1 via Bluetooth.
Figure 1. The data collection setup. Four load cells were placed under the legs of the bed. Signals from the load cell were amplified, filtered, and converted from analog to digital using a signal conditioner. Data was saved on a laptop (PC2). IMU data was collected to PC1 via Bluetooth.
Sensors 24 06483 g001
Figure 2. Close-up image of the bed wheel on a load cell.
Figure 2. Close-up image of the bed wheel on a load cell.
Sensors 24 06483 g002
Figure 3. The solid-colored black line represents an individual’s position on the bed in the transverse plane. This line is perpendicular to the line connecting the left and right anterior superior iliac spines on the individual in the frontal plane. The dotted red line represents our true supine reference point at 0° and is defined as the line perpendicular to the surface of the bed. Therefore, the TPA, represented by θ, is defined as the angle between the 0° reference line and the line perpendicular to the frontal plane of the pelvis.
Figure 3. The solid-colored black line represents an individual’s position on the bed in the transverse plane. This line is perpendicular to the line connecting the left and right anterior superior iliac spines on the individual in the frontal plane. The dotted red line represents our true supine reference point at 0° and is defined as the line perpendicular to the surface of the bed. Therefore, the TPA, represented by θ, is defined as the angle between the 0° reference line and the line perpendicular to the frontal plane of the pelvis.
Sensors 24 06483 g003
Figure 4. Three different bin sizes were used to test precision on performance: 45° bins (a); 30° bins (b); 15° bins (c).
Figure 4. Three different bin sizes were used to test precision on performance: 45° bins (a); 30° bins (b); 15° bins (c).
Sensors 24 06483 g004
Figure 5. The 11 different positions adopted by patients during primary testing and the order of positions, including intermediate holds at supine (S), right-side-lying at 90° (R90°), and left-side-lying at 90° (L90°).
Figure 5. The 11 different positions adopted by patients during primary testing and the order of positions, including intermediate holds at supine (S), right-side-lying at 90° (R90°), and left-side-lying at 90° (L90°).
Sensors 24 06483 g005
Figure 6. Mean F1 scores for the top models for Phase Two left classification for the three different bin sizes.
Figure 6. Mean F1 scores for the top models for Phase Two left classification for the three different bin sizes.
Sensors 24 06483 g006
Figure 7. Mean F1 scores for the top models for Phase Two right classification for the three different bin sizes.
Figure 7. Mean F1 scores for the top models for Phase Two right classification for the three different bin sizes.
Sensors 24 06483 g007
Table 1. Table describing the different data sources and how they were used.
Table 1. Table describing the different data sources and how they were used.
Data SourceData SpecificationsData Set Size (Samples)IMUData UsePhase
Supplemental Dataset A20 healthy participants4910NoTrainingOne
Supplemental Dataset B9 healthy participants collected at home and 1 healthy participant collected in a sleep lab13,913NoTrainingOne
PI Dataset 18 healthy participants2458YesTraining and TestingOne
Phase Two Left18 healthy participants763YesTraining and TestingTwo
Phase Two Right18 healthy participants 617YesTraining and TestingTwo
Phase Two Supine 18 healthy participants622YesN/AN/A
Table 2. Features extracted by Wong et al. [43] and Gabison et al. [44].
Table 2. Features extracted by Wong et al. [43] and Gabison et al. [44].
FeatureDescription
meanCoM_xThe mean of CoM_x
meanCoM_yThe mean of CoM_y
ratio_meanCoMThe quotient of meanCoM_y divided by meanCoM_x
stdCoM_xThe standard deviation of CoM_x
stdCoM_y The standard deviation of CoM_y
ratio_stdCoMThe quotient of stdCoM_y divided by stdCoM_x
CoM_resp_ANGCOM angle during inhalation phase only, averaged for all occurrences
stdCoM_resp_ANGStandard deviation of CoM_resp_ANG
rmsCoM_resp_xThe root mean square of the x-component of CoM_resp during both inhale and exhale phases, normalized to the 97th percentile
rmsCoM_resp_yThe root mean square of the y-component of CoM_resp during both inhale and exhalation phases, normalized to the 97th percentile
ratio_rmsCoM_respThe quotient of rmsCoM_resp_y divided by rmsCoM_resp_x
rmsPulseThe root mean square of the load cell signals filtered to capture changes resulting from the cardiac cycle
Table 3. Table describing the models created, which level they were used for, and the features they used.
Table 3. Table describing the models created, which level they were used for, and the features they used.
ModelPhase One PredictionPhase Two PredictionFeatures Used
Logistic Regression X12 features from Wong et al.
Support Vector Machine X12 features from Wong et al.
Gradient Boosting ClassifierXXSome of the 12 features from Wong et al.
AdaBoost ClassifierXX12 features from Wong et al.
XGBoost ClassifierXX12 features from Wong et al.
Light Gradient Boosting Machine ClassifierXX12 features from Wong et al.
Multilayer Perceptron (x3)X 12 features from Wong et al.
Recurrent Neural Network: Long Short-Term MemoryX 12 features from Wong et al.
and
Automatic feature selection
Convolutional Neural Network: 1-DimensionalX 12 features from Wong et al.
and
Automatic feature selection
Table 4. Table showing the participant demographics. Age ranges are included to protect privacy, but the mean and standard deviation were computed with exact values.
Table 4. Table showing the participant demographics. Age ranges are included to protect privacy, but the mean and standard deviation were computed with exact values.
ParticipantSex (M/F)Age Range (Years)Height (cm)Weight (kg)BMI
1F26–30170.092.031.8
2M71–75172.775.225.1
3M31–35181.088.727.1
4M41–45181.099.830.4
5F51–55157.587.835.4
6F21–25157.550.320.3
7M21–25175.073.824.1
8M26–30165.065.524.1
10M31–35198.0120.330.7
11F21–25161.362.624.1
12F21–25165.149.218.0
13F21–25175.057.418.8
14F21–25175.060.819.8
15F21–25170.063.722.1
17M21–25188.0107.630.4
18M21–25179.081.825.5
19M16–20185.478.022.7
20M21–25183.088.526.4
Mean ± (SD) 29.8 ± 13.0174.4 ± 10.677.9 ± 19.225.4 ± 4.7
Table 5. Mean F1 scores and accuracy with standard deviation for the Phase Two left and right classifications of different bin sizes for the best performing model, LGBoost.
Table 5. Mean F1 scores and accuracy with standard deviation for the Phase Two left and right classifications of different bin sizes for the best performing model, LGBoost.
Left
ModelBin = 45°Bin = 30°Bin = 15°
F10.919 ± 0.07360.830 ± 0.1560.786 ± 0.174
Accuracy0.915 ± 0.08020.847 ± 0.1450.815 ± 0.153
Right
ModelBin = 45°Bin = 30°Bin = 15°
F10.900 ± 0.1330.812 ± 0.1570.740 ± 0.188
Accuracy0.894 ± 0.1220.823 ± 0.1420.754 ± 0.177
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Pupic, N.; Gabison, S.; Evans, G.; Fernie, G.; Dolatabadi, E.; Dutta, T. Detecting Patient Position Using Bed-Reaction Forces for Pressure Injury Prevention and Management. Sensors 2024, 24, 6483. https://doi.org/10.3390/s24196483

AMA Style

Pupic N, Gabison S, Evans G, Fernie G, Dolatabadi E, Dutta T. Detecting Patient Position Using Bed-Reaction Forces for Pressure Injury Prevention and Management. Sensors. 2024; 24(19):6483. https://doi.org/10.3390/s24196483

Chicago/Turabian Style

Pupic, Nikola, Sharon Gabison, Gary Evans, Geoff Fernie, Elham Dolatabadi, and Tilak Dutta. 2024. "Detecting Patient Position Using Bed-Reaction Forces for Pressure Injury Prevention and Management" Sensors 24, no. 19: 6483. https://doi.org/10.3390/s24196483

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop