Article

Design of a Selective and Sensitive PVC-Membrane Potentiometric Sensor for Strontium Ion Based on 1,10-Diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dioneas a Neutral Ionophore

by
Mojtaba Shamsipur
1,*,
Sayed Yahya Kazemi
2 and
Hashem Sharghi
3
1
Department of Chemistry, Razi University, Kermanshah, Iran
2
Department of Science, College of Agriculture, Mazandaran University, Sari, Iran
3
Department of Chemistry, Shiraz University, Shiraz, Iran
*
Author to whom correspondence should be addressed.
Sensors 2007, 7(4), 438-447; https://doi.org/10.3390/s7040438
Submission received: 6 February 2007 / Accepted: 16 March 2007 / Published: 12 April 2007

Abstract

:
A novel PVC membrane sensor for the Sr2+ ion based on 1,10-diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dione has been prepared. The sensor possesses a Nernstian slope of 30.0 ± 0.6 mV decade-1 over a wide linear concentration range of 1.6 × 10-6-3.0 ×10-3 M with a detection limit of 6.3 ×10-7 M. It has a fast response time of <15 s and can be used for at least two months without any considerable divergence in potential. The potentiometric response is independent of the pH of test solution in the pH range 4.3-9.4. The proposed electrode shows good selectivities over a variety of alkali, alkaline earth, and transition metal ions.

1. Introduction

Strontium is widely distributed in many biological systems and throughout the environment, at various concentration levels, and plays a fundamental role in ecological systems [1]. It is deposited is preferentially in the bones and teeth of the human body. Some strontium compound such as strontium oxide, strontium hydroxide and strontium carbonate have a strongly irritant effect on skin and mucous membranes especially eyes [2,3]. A radioactive half-life of 29 years makes strontium-90 one of the most hazardous products of radioactive fallout [3]. The activity measurement of Sr2+ in aqueous systems is thus of vital importance from an important environmental point of view.
Ionophore-based PVC membrane electrodes are well-established analytical tools routinely used for the measurement of the activity of a wide variety of different ions selectively and directly in complex biological and environmental samples [4-6]. The key ingredient of such plasticized PVC membrane is an incorporated ionophore, which defines the selectivity of the electrodes via selective complex formation with the cation of interest.
Because macrocyclic ligands can form selective and stable complexes with metal ions of compatible dimensions [7], and can potentially be applied to their selective separation and determination [6,8,9], continuous interest has been focused on the design and synthesis of new functionalized macrocycles for specific applications [10]. Several macrocyclic diamides, prepared from salicylaldehade derivatives and appropriate diamino compounds [11], have been reported to have successful use as ion carriers in the construction of PVC-based ion-selective electrodes for Cu2+ [12], Zn2+ [13], Hg2+ [14], Sr2+ [15], Ca2+ [16], Cs+ [17], Co2+ [18], Ag+ [19], and Be2+ ions [20].
In this paper, the suitable cavity size of macrocyclic diamide derivative 1,10-diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dione (L) as well as its water insolubility led us to examine its ability as a potential ion carrier in the preparation a new PVC-membrane electrode for Sr2+ ions. It should be noted that, despite urgent need for new potentionmetric sensors for the selective determination of low level amounts of strontium in different samples, there are only limited reports on the Sr2+ ion-selective electrodes in the literature [1,15,21-23].
Sensors 07 00438u1

2. Results and Discussion

Ligands for use as ionophores in a strontium ion-selective electrode should fulfill certain conditions including (1) increased selectivity for Sr2+ over other metal ions, (2) rapid exchange kinetics of the resulting complex and (3) sufficient lipophilicity to prevent leaching of the ligand into the surrounding aqueous solution. The selectivity of ordinary crown ethers for alkaline earth metal ions is much lower than that for alkali cations [6]. However, the introduction of an amide linkage in the polyether ring has been clearly shown to modify the binding properties of the crown compounds in favor of an alkaline earth with respect to alkali metal ions [15,16,20]. Moreover, amide substitution in the cavity of benzocrown ethers may not only contribute to their cation selectivity but also allow the macrocycles to have properties more closely resembling those on the naturally occurring ionophores [26].
To investigate the potential response of the macrocyclic diamide L for different ions, it was used as a neutral ionophore to prepare PVC-membranes for a wide variety of cations. The potential responses of the most selective electrodes in the concentration range of 10-7-10-1 M are shown in Figure 1. As seen, except for the Sr2+ ion-selective electrode, in all other cases the slope of the corresponding potential-pM plots is much lower than the expected Nenstian slope of 29.5 mV decade-1 for bivalent cations, although over a limited concentration range. As is obvious from the results obtained, Sr2+ ion with the most sensitive response over a wider concentration range seems to be suitably determined with the membrane electrode based on the ligand L. This is due to the highly selective behavior of the ionophore for Sr2+ over other metal ions. The observed selectivity of ionophore L toward Sr2+ ion is most probably due to the rigid structure of the macrocycle as well as its convenient cavity size (16-membered cavity) for the bivalent strontium ion [7,15].
Some important features of the PVC membranes such as the nature and amount of ionophore, the properties of the plasticizer, the plasticizer/PVC ratio, and specially, the nature of additives used are reported to significantly influence the sensitivity and selectivity of the ion selective electrodes [12-20,27]. Thus, the effects of membrane composition, nature and amount of plasticizer, and amount of additive salt on the potential response of the Sr2+ sensor were investigated, and the results are summarized in Table 1.
As with many ionophore-based membrane electrodes, the potential response of the electrode depends on the concentration of lipophilic anionic salts incorporated in the membranes (Nos. 2-4). As seen, addition of STB will increase the sensitivity of the electrode response considerably. Use of 1.7% STB resulted in a Nernstian behavior (NO.4). The presence of such lipophilic anionic species as tetraphenylborate in cation-selective membrane electrodes is proved to have a beneficial influence on different sensor characteristics. It will not only reduce the ohmic resistance [33] and improve the response behavior and selectivity [27] but also, in cases where the extraction capability of the ionophore is poor, enhance the sensitivity of the membrane electrode [28]. Moreover, the additives may catalyze the exchange kinetics at the sample-membrane interface [2].
Since the nature of plasticizer influences the dielectric constant of the membrane phase, the mobility of the ionophore molecules and the state of ligands [29], it is expected to play an important role in determining the ion-selective characteristics. Thus, five membrane of similar composition (i.e., 30% PVC, 66.3% plasticizer, 2% ionophore and 1.7% STB) but with five different plasticizers having dielectric constant over the range 4-24, namely DMS (ε = 4.0), DBP (ε = 6.4), DOP (ε = 7.0), NPOE (ε = 24.0) and AP (ε = 17.3) were prepared and tested (Table 1, Nos 4-8). It is apparent from the performance characteristics that the membrane plasticized with AP had the highest sensitivity with a nice Nernstian slope, widest linear range and lowest detection limit in this series. The slopes of the corresponding potential responses of membranes plasticized with DMS, DBP, DOP and NPOE were 12.8, 14.7, 16.8 and 23.5 mV decade-1, respectively.
These results indicate that the polar ketone and ether type solvent mediators give a strontium ion sensor with more favorable potentiometric characteristics than nonpolar ester type solvents. This is in agreement with the previous reports showing that the response characteristics and selectivities of multivalent cations of high lipophilicities are favored by the use of high dielectric constant plasticizers [12-20]. As it is obvious from Table 1, the membrane obtained with the PVC/AP/L/STB ratio of 33.0%/66.3%/2.0%/1.7% (no. 4) shows the best response characteristics for selective determination of strontium ion.
Optimum conditioning time for the membrane sensor in a 1.0 ×10-3 M strontium nitrate solution was investigated from 5 to 24 h and found to be 20 h; it then generates stable potentials when placed in contact with Sr2+ solutions.
The average time required for the Sr2+ ion-selective electrode to reach a potential within ± 1 mV of the final equilibrium value was measured after successive immersion of a series of strontium ion solution, each 10-fold different in concentration. The static response time thus obtained was < 15 s for strontium ion concentrations ≤10-3 M, and the potential stayed constant for more than 5 min, after which only a very slow divergence with in the resolution of the pH meter (±0.1 mV) was recorded.
The PVC-membrane electrode based on L prepared under optimized experimental conditions resulted in a potential-pM concentration response with a Nernstian slope of 30.5 ± 0.6 mV/decade (n = 7) over a wide linear concentration range of 1.6 ×10-6-3.0 ×10-3 M (Figure 2). The detection limit, as evaluated from the intersection of the two extrapolated segments of the calibration graph, was 6.3 × 10-7 M.
The membrane sensor prepared could be used for at least two months without any measurable divergence. After two months, the slope of the calibration curve decreased to 27.0 mV decade-1 without any measurable change in the linear range.
The effect of pH of test solution on the response of the membrane electrode for a 1.0 × 10-4 M solution of Sr2+ was studied over the pH range 3.5-11.5, the results are given in Figure 3. It is immediately obvious that the potential stays constant from pH 4.3 to 9.4, beyond which sharp increases in potential are observed. The observed sharp change in potential response below pH 4.3 might be due to the simultaneous response of the electrode to Sr2+ ion and H+ and at pH higher than 9.4, it could be because of relatively strong competition between Na+ (from NaOH) and Sr2+ ions.
The selectivity behavior is obviously one of the most important characteristics of an ion-selective electrode, determining whether a reliable measurement in the electrode proposed can be obtained. Thus, its potential response was investigated in the presence of various interfering foreign cations using the matched potential method (MPM) [30]. The MPM is a recently recommended procedure which gets ride of the limitations of the corresponding methods based on the Nicolski-Eisenman equation for the determination of potentiometric selectivity coefficients. These limitations include non-Nernstian behaviors of interfering ions and inequality of charges of any primary interfering ion.
According to the MPM, the selectivity coefficient is defined as the activity ratio of the primary ion and the interfering ion that gives the same potential change in a reference solution. Thus, one should measure the change potential upon changing the primary ion activity, and then the interfering ion would be added to an identical reference solution until the same potential change is obtained. The selectivity coefficient is determined as: KA,BMPM = Δa′A/aB, where ΔaA = aA′ – aA, aA is the initial primary ion activity and aA′ is the activity of A in the presence of interfering ion, B, and aB is the interfering ion activity. Thus, according to this method, a specific activity (concentration) of the primary ion A (1.0 × 10-6-1.0 × 10-5 M Sr2+) is added to a reference solution (5.0 × 10-7 M Sr2+) and the potential is measured. In a separate experiment, the interfering ions B are successively added to an identical reference solution (from 1.0 × 10-5 to 1.0 × 10-3 M of Mn+) until the measured potential matched that obtained before by adding the primary ions. The KA,BMPM value is then calculated as Δa′A/aB. The resulting values are listed in Table 2. As it is evident from Table 2, most of the interfering ions show low values of selectivity coefficient, indicating negligible interference in the performance of the membrane sensor assembly.
In Table 3, the response characteristics of the proposed strontium-sensor are compared with those of the best Sr2+ ion-selective electrodes reported before [2,20,26,27,29]. From the data given in Table 3 it is immediately obvious that the limit of detection of the proposed sensor is superior to those reported before, while the other response characteristics and selectivity coefficients obtained in this work are within the same ranges in comparison with those previously reported for the case of the best strontium ion-selective electrodes.
The proposed membrane sensor can not only be used for the direct determination of Sr2+ ions in real samples (e.g., quantitative recovery of traces of Sr2+ ion from tap water), but was also found useful as an electrode in potentiometric titration procedures involved strontium ion. As an example, it was successfully applied as an indicator electrode for the titration of 40 ml of 1.0 × 10 –3 M concentration of Sr2+ ions, at pH 5.5, with 0.100 M EDTA solution and results are shown in Figure 4. As seen, the amount of Sr2+ ions in solution can be accurately determined from the resulting titration curve providing a sharp end point.

3. Experimental Section

3.1. Reagents

Reagent-grade 2-nitrophenyl octyl ether (NPOE), dibutyl phthalate (DBP), dioctyl phthalate (DOP), acetophenone (AP), dimethylsebacate (DMS), tetrahydrofuran (THF), sodium tetraphenylborate (STB) and high relative molecular weight PVC were purchased from Fluka Chemical Company and used as received. Reagent-grade nitrate salts of all cations used (all from Merck) were of the highest purity available and used without any further purification except for vacuum drying. The macrocyclic diamide L was synthesized and purified in as described elsewhere [24]. Triply distilled and deionized water was used throughout.

3.2. Electrode Preparation

The general procedure to prepare the PVC membrane was to mix thoroughly 66.3 mg of plasticizer AP, 30 mg of powdered PVC and 2 mg of ionophore and 1.7 mg of STB in 0.5 ml of methanol and 3.5 ml of THF. The resulting mixture was transferred into a glass dish of 2-cm diameter. The solvent was evaporated slowly until an oily concentrated mixture was obtained. A Pyrex tube (4-5 mm diameter on the top) was dipped into the oily mixture for about 10 s so that a non-transparent membrane of about 0.3 mm thickness was formed. The tube was then pulled out from the mixture and at room temperature for about 1 h. The tube was then filled with internal filling solution (1.0 × 10- M strontium nitrate). The electrode was finally conditioned for 20 h by soaking in a 1.0 × 10-3 M solution of Sr(NO3)2. A silver/silver chloride electrode was used as an internal reference electrode.

3.3. EMF Measurements

All emf measurements were carried out the following assembly:
Ag-AgCl (satd) | 3 M KCl | internal solution, 1.0 ×10-3 M Sr(NO3)2 | PVC membrane | test solution | Ag-AgCl, 0.1 M KCl |
A Metrohm ion analyzer pH/mV meter was used for the potential measurements at 25.0 ± 0.1 °C.
Activities were calculated according to Debye-Hückel procedure [25].

Acknowledgments

The support of this work by a research grant (no. 84123/17) from Iran National Science Foundation (INSF) is gratefully acknowledged.

References and Notes

  1. Jain, A.K.; Gupta, V.K.; Raisoni, J.R. Strontium(II)-selective potentiometric sensor based on ester derivative of 4-tert-butylcalix[8]arene in PVC matrix. Sensors 2004, 4, 115–124. [Google Scholar]
  2. Elvers, B.; Hawkins, S.; Russey, W. (Eds.) Ulmann'sEncyclopedia of Industrial Chemistry, 5th Ed.; Vol. A25, 1995; VCH Publisher: New York; p. 321.
  3. Othmer, K. Encyclopedia of Chemical Technology, 4th Ed. ed; Vol. 22, 1992; John Wiley & Sons: New York; p. 947. [Google Scholar]
  4. Umezawa, Y. (Ed.) CRC Handbook of Ion-Selective Electrodes: Selectivity Coefficients; 1990; CRC Press: Boca Raton, FL.
  5. Janata, J.; Jasowicz, M.; DeVaney, D.M. Chemical sensors. Anal. Chem. 1998, 70, 179R–208R. [Google Scholar]
  6. Bühlmann, P.; Pretsch, E.; Bakker, E. Carrier based ion-selective electrodes and the bulk optodes. 2. Ionophoresand for potentiometric and optical sensors. Chem. Rev. 1998, 98, 1593–1678. [Google Scholar]
  7. Izatt, R.M.; Pawlak, K.; Bradshaw, J.; Bruening, R.L. Thermodynamic and kinetic data for macrocycle interaction with cations and anions. Chem. Rev. 1991, 91, 1721–2085. [Google Scholar]
  8. Shih, J.-S. Applications of macrocyclic polyethers. J. Chin. Chem. Soc. 1992, 39, 551–559. [Google Scholar]
  9. Lippolis, V.; Shamsipur, M. Synthesis, coordination properties, and analytical applicationsof mixed donor macrocycles containing the1,10-phenanthroline sub-unit. J. Iran. Chem. Soc. 2006, 3, 105–127. [Google Scholar]
  10. An, H.Y.; Bradshaw, J.S.; Izatt, R.M.; Yan, Z. Bis and oligo(benzocrown ethers). Chem. Rev. 1994, 94, 939–991. [Google Scholar]
  11. Sharghi, H.; Eshghi, H. Efficient synthesis of macrocyclic diamides. Tetrahedron 1995, 51, 913–922. [Google Scholar]
  12. Shamsipur, M.; Rouhani, S.; Ganjali, M. R.; Eshghi, H.; Sharghi, H. Cu(II)-selective membrane electrode based on a recently synthesized macrocyclic diamide. Microchem. J. 1999, 63, 202–210. [Google Scholar]
  13. Shamsipur, M.; Rouhani, S.; Ganjali, M.R.; Sharghi, H.; Eshghi, H. Zinc-selective membrane potentiometric sensor based on a recently synthesized benzo-substituted macrocyclic diamide. Sens. Actuators B 1999, 59, 30–34. [Google Scholar]
  14. Javanbakht, M.; Ganjali, M.R.; Eshghi, H.; Sharghi, H.; Shamsipur, M. Mercury(II) ion-selective electrode based on dibenzo-diazathia-18-crown-6-dione. Electroanalysis 1999, 11, 81–84. [Google Scholar]
  15. Shamsipur, M.; Rouhani, S.; Sharghi, H.; Ganjali, M.R.; Eshghi, H. Strontium-selective membrane electrode based on some recently synthesized benzo-substituted macrocyclic diamides. Anal. Chem. 1999, 71, 4938–4943. [Google Scholar]
  16. Shamsipur, M.; Khayatian, G.; Kazemi, S.Y.; Niknam, K.; Sharghi, H. The synthesis of 1,4-diaza-2,3;8,9-dibenzo-7,10-dioxacyclododecane-5,12-dione and its use in calcium-selective carbon paste electrodes. J. Incl. Phenom. 2001, 40, 303–307. [Google Scholar]
  17. Shamsipur, M.; Kazemi, S.Y.; Sharghi, H.; Niknam, K. Cesium-selective membrane electrode based on a recently synthesized 16-membered macrocyclic diamide. Fresenius J. Anal. Chem. 2001, 371, 1104–1108. [Google Scholar]
  18. Shamsipur, M.; Rouhani, S.; Poursaberi, T.; Ganjali, M.R.; Sharghi, H.; Niknam, K. Cobalt(II)-selective coated graphite PVC-membrane electrode based on a recently synthesized dibenzopyridino-substituted macrocyclicdiamide. Electroanalysis 2002, 14, 729. [Google Scholar]
  19. Shamsipur, M.; Kazemi, S.Y.; Niknam, K.; Sharghi, H. A new pVC-membrane electrode based on a thia-substituted macrocyclic diamide in selective potentiometric determination of silver ion. Bull. Korean Chem. Soc. 2002, 23, 53–58. [Google Scholar]
  20. Shamsipur, M.; Soleymanpour, A.; Akhond, M.; Sharghi, H. New macrocyclic diamides as neutral ionophores for highly selective and sensitive PVC-membrane electrodes for Be2+ ion. Electroanalysis 2004, 16, 282–288. [Google Scholar]
  21. Akmal, N.; Zimmer, H.; Mark, H.B. Strontium ion-selective electrode based on a conducting poly(dibenzo-18-crown-6) film. Anal. Lett. 1991, 24, 1431–1443. [Google Scholar]
  22. Qian, G.; Wu, B.B.; Wu, G.; Huang, S.; Yan, Y.; Tian, B. Strontium-selective electrodes based on diamides with pyridine ring as ionophores. Talanta 1998, 47, 1149–1155. [Google Scholar]
  23. Gupta, V.K.; Ludwig, R.; Agarwal, S. Strontium(II) sensor based on a modified calyx[6]arene in PVC matrix. Anal. Sci. 2005, 21, 293–296. [Google Scholar]
  24. Sharghi, H.; Zare, A. Efficient synthesis of some novel macrocyclic diamides using fast addition method. Synthesis 2006, 999–1004. [Google Scholar]
  25. Kamata, S.; Bhale, A.; Fukunaga, Y.; Murata, A. Copper(II)-selective electrode using thiuram disulfate neutral carriers. Anal. Chem. 1988, 60, 2464–2467. [Google Scholar]
  26. Kimura, K.; Kumami, K.; Kitazawa, S.; Shono, T. Calcium-selective polymeric membrane electrodes based on bicyclic poly(ether amide). Anal. Chem. 1984, 56, 2369–2372. [Google Scholar]
  27. Bakker, E.; Bühlmann, P.; Bakker, E. Carrier-based ion selective electrodes and bulk optodes. Part 1. General characteristics. Chem. Rev. 1997, 97, 3083–3112. [Google Scholar]
  28. Eugster, R.; Gehring, P.M.; Morf, W.E.; Spichiger, U.E.; Simon, W. Selectivity-modifying influence of anionic sites in neutral carrier-based membrane electrodes. Anal. Chem. 1991, 63, 2285–2289. [Google Scholar]
  29. Gehrig, P.M.; Morf, W.E.; Welti, M.; Pretsch, E.; Simon, W. Catalysis of ion transfer by tetraphenylborates in neutral carrier-based ion-selective electrodes. Helv. Chim. Acta 1990, 73, 203–212. [Google Scholar]
  30. Umezawa, Y.; Umezawa, K.; Sato, H. Selectivity coefficients for ion-selective electrodes: recommended methods for reporting Kpot values. Pure Appl. Chem. 1995, 67, 507–518. [Google Scholar]
Figure 1. Potential response of various ion-selective membranes based on ionophore L. Conditions: membrane ingredients, 2% L, 30% PVC, 1.7% STP and 66.3% AP, internal solution, 1.0 × 10-3 M of the corresponding cations for each ion-selective sensor.
Figure 1. Potential response of various ion-selective membranes based on ionophore L. Conditions: membrane ingredients, 2% L, 30% PVC, 1.7% STP and 66.3% AP, internal solution, 1.0 × 10-3 M of the corresponding cations for each ion-selective sensor.
Sensors 07 00438f1
Figure 2. Calibration graph for the strontium ion-selective electrode.
Figure 2. Calibration graph for the strontium ion-selective electrode.
Sensors 07 00438f2
Figure 3. Effect of pH on the strontium test solution.
Figure 3. Effect of pH on the strontium test solution.
Sensors 07 00438f3
Figure 4. Potentiometric titration curve of 40 ml of 0.001 M Sr(NO3)2 with 0.1 M EDTA at pH 6.5, using the proposed sensor as indicator electrode.
Figure 4. Potentiometric titration curve of 40 ml of 0.001 M Sr(NO3)2 with 0.1 M EDTA at pH 6.5, using the proposed sensor as indicator electrode.
Sensors 07 00438f4
Table 1. Optimization of membrane ingredients.
Table 1. Optimization of membrane ingredients.
NoComposition (%)Slope (mV decade-1)Linear rangea (M)

PVCPlasticizerSTBL
132.5AP, 66.00.50.013.54.6 × 10-5-1.3 × 10-3
231.0AP, 67.00.02.022.01.7 × 10-5-6.3 × 10-3
331.0AP, 66.01.02.025.03.6 × 10-6-6.3 × 10-3
430.0AP, 66.31.72.030.51.6 × 10-6-3.0 × 10-3
530.0NPOE, 66.31.72.023.53.6 × 10-6-1.2 × 10-3
630.0DBP, 66.31.72.016.83.6 × 10-6-1.0 × 10-3
730.0DOP, 66.31.72.014.73.6 × 10-6-6.3 × 10-3
830.0DMS, 66.31.72.012.83.6 × 10-6-6.2 × 10-3
aIn all cases, the correlation coefficient of the linear calibration plot was at least 0.99.
Table 2. Selectivity coefficients of various interfering ions.
Table 2. Selectivity coefficients of various interfering ions.
Interfering ionKA,BMPMInterfering ionKA,BMPM
Na+1.3×10-2Ba2+9.2×10-2
K+3.3×100-2Co2+8×10-2
Cs+3.3×10-2Ni2+5.4×10-2
Rb+3.1×10-2Cu2+9.2×10-2
Mg2+6.3×10-2Pb2+7.2×10-2
Ca2+8.9×10-3
Table 3. Response characteristics and selectivity coefficients for different PVC-membrane Sr2+ ion-selective electrodes.
Table 3. Response characteristics and selectivity coefficients for different PVC-membrane Sr2+ ion-selective electrodes.
Sensors 07 00438u2

Share and Cite

MDPI and ACS Style

Shamsipur, M.; Kazemi, S.Y.; Sharghi, H. Design of a Selective and Sensitive PVC-Membrane Potentiometric Sensor for Strontium Ion Based on 1,10-Diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dioneas a Neutral Ionophore. Sensors 2007, 7, 438-447. https://doi.org/10.3390/s7040438

AMA Style

Shamsipur M, Kazemi SY, Sharghi H. Design of a Selective and Sensitive PVC-Membrane Potentiometric Sensor for Strontium Ion Based on 1,10-Diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dioneas a Neutral Ionophore. Sensors. 2007; 7(4):438-447. https://doi.org/10.3390/s7040438

Chicago/Turabian Style

Shamsipur, Mojtaba, Sayed Yahya Kazemi, and Hashem Sharghi. 2007. "Design of a Selective and Sensitive PVC-Membrane Potentiometric Sensor for Strontium Ion Based on 1,10-Diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dioneas a Neutral Ionophore" Sensors 7, no. 4: 438-447. https://doi.org/10.3390/s7040438

Article Metrics

Back to TopTop