Validations of Blood Pressure Measuring Devices Using Recognized Protocols

by
Victoria Mazoteras-Pardo
1,
Sagrario Gómez-Cantarino
1,*,
Miguel Ramírez-Jiménez
2,
Emmanuel Navarro-Flores
3 and
María Idoia Ugarte-Gurrutxaga
1
1
Research Group “ENDOCU”, Department of Nursing, Physiotherapy and Occupational Therapy, Faculty of Physiotherapy and Nursing of Toledo, University of Castilla-La Mancha, 45071 Toledo, Spain
2
Faculty of Education REDAFLED, University of Valladolid, 42004 Soria, Spain
3
Frailty Research Organized Group (FROG), Department of Nursing, Faculty of Nursing and Podiatry, University of Valencia, 46001 Valencia, Spain
*
Author to whom correspondence should be addressed.
J. Pers. Med. 2023, 13(1), 9; https://doi.org/10.3390/jpm13010009
Submission received: 30 October 2022 / Revised: 13 December 2022 / Accepted: 19 December 2022 / Published: 21 December 2022

Abstract

:
Preventing, diagnosing, and controlling high blood pressure is a global health priority. The self-measurement of blood pressure is therefore fundamental and should be done with devices validated by recognized protocols, although most are not. The most widely used and current protocols are the 2010 European Society of Hypertension (ESH) revision and the 2018 Association for the Advancement of Medical Instrumentation (AAMI)/ ESH/ the International Organization for Standardization (ISO) universal standard, respectively. The aim of this study was to find out which blood pressure measuring devices have been adequately validated by the above protocols. A narrative review of blood pressure device validations was conducted by searching the PubMed database. From 52 records identified, 37 studies were included. Most validations follow the 2010 revision and only six follow the 2018 protocol, which is more demanding. Almost all validated sphygmomanometers are automated oscillometric sphygmomanometers in the general population. Wrist devices and devices combining new technologies are also validated, as well as in specific populations, such as the obese, pregnant women, or children. There is sufficient evidence to confirm that the universal AAMI/ ESH/ISO standard is considered the protocol of the century. However, it is necessary to increase the number of validations following it and, above all, validations of the new technologies that are invading the current market.

1. Introduction

After obesity, arterial hypertension (HBP) is the second most common cause of cardiovascular disease (CVD) [1]. If we add that CVD is the leading cause of morbidity and mortality [2,3], we can conclude that AHT carries a high risk of CV morbidity and mortality [4,5,6,7].
Therefore, one of the main objectives of health systems is to identify people with AHT and ensure that they have good control of their blood pressure (BP), since the higher the blood pressure levels, the greater the risk and morbidity and mortality from CV events [8,9,10]. Thus, preventing, diagnosing, treating, and controlling hypertension is a global health priority [4,5,7,9,10,11,12,13,14].
The detection and diagnosis of hypertension must be done by measuring BP. BP measurement is considered as one of the most frequently performed procedures at the clinical level, in primary or specialized care [15].
The first time BP was recorded in a consultation was in 1896, and since the end of the last century, out-of-office measurement has been established in two versions: ambulatory BP monitoring (ABPM) and home blood pressure monitoring (HBPM). The major advantage of these two methods is that by allowing BP to be measured on multiple occasions outside the healthcare environment, it provides a more reliable BP reading and has a higher prognostic value than doing it in the office [16,17,18].
If we focus on HBPM, it has become a very beneficial and increasingly used simple procedure, for highly consistent reasons [19,20,21]. The subject’s self-measurements are very useful for the monitoring of HBP, since it is estimated that about 80–90% of the doubts in the diagnosis and control of this pathology can be solved with this procedure, but, logically, these benefits can be obtained if the HBPM is done properly and with validated devices [15,17,22,23].
With regard to blood pressure validations, in the last 30 years, there have been several protocols for this purpose, such as the British Society of Hypertension (BSH) [24] protocol, the Association for the Advancement of Medical Instrumentation (AAMI) protocol, [25] and the international protocol published by the European Society of Hypertension (ESH) [26] and its review [27]. The latter [26,27] were the most current and widely used, but given the need to increase the validity of these devices, in 2018, experts from the AAMI, ESH, and the International Organization for Standardization (ISO) agreed to develop a universal standard for their validation. Today, it is considered as the single universal standard and replaces all other previous standards/protocols [25,28,29].
Given the relevance of HBPM and the need for increased monitoring of BP devices, according to the literature, the results of the present study will mainly serve to find out which devices can be validly used by subjects to measure and monitor their blood pressure, according to the most used and/or current protocols [27,29].

2. Materials and Methods

2.1. Study Design

A narrative review was carried out following the applicable recommendations of the Scale for the Assessment of Narrative Review Articles (SANRA) [30]. This scale contains six items to assess the quality of narrative review articles. It can be found in the Supplementary Material. The aim of this review was to update the data of the devices available on the market that are valid for measuring blood pressure since there are no recent previous systematic reviews on the issue.

2.2. Search Strategy

The database search was carried out during November–January 2021. Pubmed was the database used for this process.
The advanced search strategy was as follows, combining the terms with the Boolean and grouping operators that follow: (European Society of Hypertension [Title]) OR AAMI/ESH/ISO[Title]) AND validation [Title].
As for the search filters, only one of the publication dates is applied: Last 5 years.
The search strategy has been filtered by title because, by regulation, all validations following the “the revised 2010 European Society of Hypertension international protocol” and “the 2018 AAMI/ESH/ISO universal standard” should be named in a standardized way and, consequently, the titles of these publications should also be named in the same way.

2.3. Selection Criteria

Inclusion criteria comprised blood pressure device validation studies that followed the 2010 European Society of Hypertension international protocol review and/or the 2018 AAMI/ESH/ISO universal standard, conducted within the last 5 years.
Exclusion criteria included studies with a publication date prior to 2017, and those that were not device validations or did not follow the revised 2010 European Society of Hypertension international protocol and/or the 2018 AAMI/ESH/ISO universal standard.

2.4. Data Extraction

The data obtained were divided according to the validation protocol used. The common methodology governing the validation conditions was recorded, such as sample size, blood pressure range, and other variables, in accordance with the protocol used.
Study characteristics were also recorded, such as citation year and place of validation, types of devices used, population characteristics and origin, and validation conclusions.
Additional information was provided where necessary.

2.5. Flow Diagram

From 52 records identified through the searching process, 15 were removed due to exclusion, and finally, 37 studies were included in narrative synthesis (Figure 1).
15 studies were excluded for the following reasons:
  • − 7 articles discussed the validation standard itself and/or revisions of the validation standard
  • − 2 studies commented with the proper use of validation protocols
  • − 4 publications were corrections of previously included articles.
  • − 1 paper was about validation but only followed the ESH 2002 protocol.
  • − 1 study did not analyze validity but reproducibility.

3. Results

3.1. Characteristics of the Validation Protocols

The studies of the last five years on blood pressure measuring devices mainly apply two validation protocols for BP devices, which are the review of the international ESH protocol [27] and the international universal standard AAMI/ESH/ISO 81160-2:2018 [29], the latter being the most recent.
Thus, 31 articles following the ESH review [27] and six based on the AAMI/ESH/ISO universal standard [29] are found in the literature. The methodology of these studies has common characteristics, depending on the validation protocol they follow (Table 1). The characteristics of the sample, the sample recruitment criteria, the blood pressure measurement method, or the analyses required to validate the device in question will differ.
Both protocols use a similar sequential BP measurement procedure by alternating two devices (reference and test). Validated devices are already used as a reference, but the 2010 ESH prefers these to be two mercury sphygmomanometers with stethoscopes and the AAMI/ESH/ISO standard indicates that the reference can be as stated above, but can also be non-mercury sphygmomanometers, aneroid manometers, or other types. The measurement conditions are similar, where the subject being measured must remain relaxed and calm in a certain position. Also, the human validation equipment is almost identical, and it has a similar tolerable BP measurement error. In both protocols, Bland–Altman charts are required.
On the other hand, they vary substantially in the range of BP recruitment, sample characteristics and size, and validation criteria.
For a device to pass the ESH [27], it must pass two phases. To pass the first phase, two conditions must be met: a minimum of 65, 81, and 93 comparisons falling within 5, 10, and 15 mm/Hg, respectively; a minimum of two of the following three requirements: 73, 87, and 96 differences must be within the category of 5, 10, and 15 mm/Hg, respectively. In the second phase, a minimum of 24 subjects are required to have two of their three differences in the 5 mm/Hg category, and a maximum of three individuals with the three differences greater than 5 mm/Hg is allowed.
To pass the universal standard [29], the device must also pass two phases. In the first phase, the mean BP difference must be 5 mm/hg or less, and its standard deviations 8 mm/hg or less for SBP and DBP. In the second phase, the standard deviation of 85 averaged BP differences (test minus reference BP per subject) must be within a threshold defined by the mean test-reference BP difference listed for SBP and DBP.
In terms of specific populations, the standards proposed by the three societies as a whole are more explicit.
Thus, although the ESH [27] protocol improved on previous protocols [24,26,27] by eliminating some validation steps and reducing the sample size, it is simpler to apply than the most recent protocol. The international universal validation protocol (AAMI/ESH/ISO) of 2018 [29] is currently the most complete but more complex than the previous one.

3.2. Characteristics of the Studies

All the studies collected have the same quality in terms of study type, as they all follow the same standards and have the same design: prospective cross-sectional observational studies [31].
Having analyzed the common and different aspects in terms of the validation conditions governing the respective protocols (Table 1), we now turn to the rest of the characteristics, which are summarized in Table 2.

4. Discussion

We found positive results in most studies, exceeding the protocol in question and validating the test devices for their recommended use, as well as in the population analyzed, with the exception of two studies with four test devices, Omron RS6 (hem-6221-E), Microlife WatchBP O3 [33], Yuwell YE680a, and Cofoe KF-65B [67].
The validated devices are different depending on the measurement method, the area of the body where they are applied, their characteristics, or their function: self-measurement, BP monitoring, or professional use, among others. Only one study has validated a device that uses the auscultatory method to measure BP [66]. Typically, oscillometric upper arm sphygmomanometers are validated as the conventional devices for measuring BP and heart rate (HR). These consist of an automatic monitor with an LCD display and a cuff connected by rubber tubes, with some exceptions, which do not have tubes connecting the pump to the cuff; rather, the device itself is embedded in the cuff [42,47,53,55]. They are usually battery operated and some have various cuff sizes.
The cuff used by most authors in their test device is usually the standard or medium size (22–32 cm), except for some who have tested more than one cuff, such as small (18–22 cm) [65], or large (32–42 cm) [65,66].
There are authors who opt for other wrist devices of the same type [33,35,38,44,50,61] and therefore do not have rubber tubes. These are more applicable to obese subjects because they have a larger arm circumference, giving erroneous measurement results if a standard upper cuff is used [33].
Of the above, many are recommended devices for HBPM but some are designed for ABPM [32,36,46,48,53], while there are few validated sphygmomanometers that are used only in a professional manner [58,59,65,66]. According to the ESH protocol, the validation of ABPM devices is performed only in static conditions and does not require testing in ambulatory conditions.
Among those used for HBPM, other validated devices differ more from the above. The Inbody BPBIO320 [49] and the Inbody BPBIO750 [62] are for public use as well as being automatic right upper arm oscillometric devices, but in this case, it is a kiosk type with a fixed hole for the user to insert their arm to measure their BP. Other devices have also been validated that measure BP oscillometrically and can be connected wirelessly to electronic devices allowing for better visual recording and monitoring [46,51,63]. Other units have similar behavior but are operated directly from an app connected to the oscillometric cuff via Bluetooth [42,47,55]. To this effect, one study tests an app that measures BP without a cuff through fingerprints [68] but it is difficult to follow the validation protocol with it.
Therefore, of all the validations, only one device without a cuff to measure BP is attempted to be validated [68], and the results are unsuccessful.
It should be noted that the use of new technologies for health promotion and disease prevention is booming and, in particular, BP control using these has gained importance, adding additional advantages to HBPM, which results in a more active subject with better adherence and follow-up [68,69,70,71]. Despite this and its widespread use and downloading by society, only six apps are being validated [42,46,47,51,55,63,68], considering that one of them, Qardioarm, has been validated on several occasions, by different teams and in different populations [42,47,55]. This is not a unique case, since it draws our attention to the fact that the Omron RS6 is tested on obese subjects and is found to be invalid by some authors [33] and valid by others [61]. This may be due to the fact that both studies follow the ESH review and do not have specific criteria for vulnerable populations [27].
As for the parameters assessed in the populations, all studies include recruitment SBP and DBP, age (in years), and circumference of the upper arm [32,34,36,37,39,40,41,42,43,45,46,47,48,49,51,52,53,54,55,56,57,58,59,60,62,63,64,65,66,67], wrist [38,40], or both [25,33,44,61] (in mm), depending on the device to be tested. Only one study does not measure any circumference of its participants [68]. BMI is also assessed by some authors [33,34,38,45,46,47,51,55,57,61], although only some of these [47,51,55,61] reflect the weight and height of their participants.
These measured characteristics are generally similar as they all meet the recruitment criteria of the protocols [27,29], except for special populations.
Most of these studies are conducted in the general population, which are healthy adults over 25 years of age for one protocol [27], and 12 years of age for another [29]. In all of them, the number of women recruited tends to be higher than that of men.
Some studies look at other specific populations prone to HBP, such as the obese [33,47,61], children [36,65], diabetics [42], pregnant women [48], or subjects with chronic kidney disease [55]. Furthermore, the difference between the recruited and tested subjects is often greater in these specific populations, as more failures encountered tend to emerge. Most of the reasons for exclusion are full BP ranges. The above are necessary because the results obtained in the general population cannot be extrapolated to specific populations, as they do not follow the same recruitment conditions, requiring a larger sample in most cases. Hence, the universal standard already applies explicit criteria for such populations [29] where even stricter BP control is necessary.
As for the validation team, it also follows the rules of the protocols [27,29], reflected in Table 1, so it usually corresponds to two observers and a supervisor, except for others that have additional explanations, such as teams formed by three licensed physicians [32,48,58], three nurses [47,55], four physicians [49], or three medical technologists [50,59]. All teams were trained in BP measurement.
Although the universal standard is considered the protocol of the century [29], only seven articles have been seen to follow it [56,58,59,62,65,66,68] compared to the large number of studies that follow the ESH [27], almost all of them with positive results [72,73], because although they share conditions, it is less strict (Table 1).
For example, it was recommended to the journals that, from November 2019 onward, they no longer accept articles following the ESH to validate the devices [74] and only accept the most recent protocol [29], so it is considered necessary to increase the monitoring of this type of device, especially their apps, following the same protocol. In addition, a mercury-free sphygmomanometer can be used as the gold standard in this protocol and is used by the ESH.
To conclude this section, it is worth mentioning that one of the main limitations of our study is that we have only used one database, Pubmed. We were guided by a study that analysed optimal database combinations for literature searches and found that “Sixteen percent of the included references (291 articles) were only found in a single database” [75].

5. Conclusions

Since 2017, 37 devices have been attempted to be validated to measure BP with reputable protocols, with most obtaining positive results. It is necessary to increase the number of studies according to what is considered the protocol of the century, the universal AAMI/ESH/ISO standard (ISO 81160-2:2018), and especially using new technologies, as few validations have been found on them.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/jpm13010009/s1. Scale S1: Scale for the Assessment of Narrative Review Articles (SANRA).

Author Contributions

Conceptualization, V.M.-P. and M.R.-J.; formal analysis, V.M.-P. and M.I.U.-G.; data curation, V.M.-P., S.G.-C., E.N.-F. and M.R.-J.; writing—original draft preparation, V.M.-P. and M.R.-J.; writing—review and editing, V.M.-P., S.G.-C., M.R.-J., E.N.-F. and M.I.U.-G.; supervision, V.M.-P., S.G.-C., M.R.-J., E.N.-F. and M.I.U.-G. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Bolaji, A.A. Simulation of a Real-time Mobile Health Monitoring System Model for Hypertensive Patient in Rural Nigeria. Afr. J. Comput. Ict. 2014, 7, 95–100. [Google Scholar]
  2. Wang, H.; Dwyer-Lindgren, L.; Lofgren, K.T.; Rajaratnam, J.K.; Marcus, J.R.; Levin-Rector, A.; Levitz, C.E.; Lopez, A.D.; Murray, C.J.L. Age-specific and sex-specific mortality in 187 countries, 1970-2010: A systematic analysis for the Global Burden of Disease Study 2010. Lancet 2012, 380, 2071–2094. [Google Scholar] [CrossRef] [PubMed]
  3. Raghu, A.; Praveen, D.; Peiris, D.; Tarassenko, L.; Clifford, G. Engineering a mobile health tool for resource-poor settings to assess and manage cardiovascular disease risk: SMARThealth study. BMC Med. Inf. Decis. 2015, 15, 36. Available online: https://pubmed.ncbi.nlm.nih.gov/25924825/ (accessed on 12 December 2021). [CrossRef] [PubMed] [Green Version]
  4. Mancia, G.; Fagard, R.; Narkiewicz, K.; Redon, J.; Zanchetti, A.; Böhm, M.; Christiaens, T.; Cifkova, R.; De Backer, G.; Dominiczak, A.; et al. 2013 ESH/ESC guidelines for the management of arterial hypertension: The Task Force for the management of arterial hypertension of the European Society of Hypertension (ESH) and of the European Society of Cardiology (ESC). Eur. Heart J. 2013, 34, 2159–2219. [Google Scholar] [PubMed] [Green Version]
  5. Chow, C.K.; Teo, K.K.; Rangarajan, S.; Islam, S.; Gupta, R.; Avezum, A.; Bahonar, A.; Chifamba, J.; Dagenais, G.; Diaz, R.; et al. Prevalence, awareness, treatment, and control of hypertension in rural and urban communities in high-, middle-, and low-income countries. JAMA 2013, 310, 959–968. Available online: https://pubmed.ncbi.nlm.nih.gov/24002282/ (accessed on 12 December 2021). [CrossRef] [Green Version]
  6. Levin, M.G.; Klarin, D.; Walker, V.M.; Gill, D.; Lynch, J.; Hellwege, J.N.; Keaton, J.M.; Lee, K.M.; Assimes, T.L.; Natarajan, P.; et al. Association between genetic variation in blood pressure and increased lifetime risk of peripheral artery disease. Arterioscler. Thromb. Vasc. Biol. 2021, 41, 2027–2034. Available online: https://www.ahajournals.org/doi/abs/10.1161/ATVBAHA.120.315482 (accessed on 12 December 2021). [CrossRef]
  7. Visseren, F.L.J.; Mach, F.; Smulders, Y.M.; Carballo, D.; Koskinas, K.C.; Bäck, M.; Benetos, A.; Biffo, A.; Boavida, J.-M.; Capodanno, D.; et al. 2021 ESC Guidelines on cardiovascular disease prevention in clinical practice. Eur. J. Prev. Cardiol. 2021, 29, 5–115. Available online: https://pubmed.ncbi.nlm.nih.gov/34558602/ (accessed on 12 December 2021). [CrossRef]
  8. BP lowering and cardiovascular risk reduction. Drug Ther. Bull. 2021, 60, 3. Available online: https://pubmed.ncbi.nlm.nih.gov/34893501/ (accessed on 12 December 2021).
  9. Siu, A.L.; Bibbins-Domingo, K.; Grossman, D.; Baumann, L.C.; Davidson, K.W.; Ebell, M. US Preventive Services Task Force. Screening for high blood pressure in adults: U.S. preventive services task force recommendation statement. Ann. Intern. Med. 2015, 163, 778–786. [Google Scholar] [CrossRef] [Green Version]
  10. Casey, D.E., Jr.; Daniel, D.M.; Bhatt, J.; Carey, R.M.; Commodore-Mensah, Y.; Holmes, A.; Smith, A.P.; Wozniak, G.; Wright, J.T., Jr. Controlling High Blood Pressure: An Evidence-Based Blueprint for Change. Am. J. Med. Qual. 2021, 37, 22–31. [Google Scholar] [CrossRef]
  11. Khatib, R.; Schwalm, J.D.; Yusuf, S.; Haynes, R.B.; McKee, M.; Khan, M.; Nieuwlaat, R. Patient and healthcare provider barriers to hypertension awareness, treatment and follow up: A systematic review and meta-analysis of qualitative and quantitative studies. PLoS ONE 2014, 9, e84238. Available online: https://pubmed.ncbi.nlm.nih.gov/24454721/ (accessed on 12 December 2021). [CrossRef] [Green Version]
  12. Rodgers, A.; Chow, C.K.; Jackson, R.T.; Patel, A.; Usherwood, T. Guideline for the diagnosis and management of hypertension in adults—2016. Med. J. Aust. 2017, 206, 141. Available online: https://pubmed.ncbi.nlm.nih.gov/28208048/ (accessed on 13 December 2021). [CrossRef]
  13. Carey, R.M.; Wright, J.T.; Taler, S.J.; Whelton, P.K. Guideline-Driven Management of Hypertension: An Evidence-Based Update. Circ. Res. 2021, 128, 827–846. Available online: https://pubmed.ncbi.nlm.nih.gov/33793326/ (accessed on 13 December 2021).
  14. Unger, T.; Borghi, C.; Charchar, F.; Khan, N.A.; Poulter, N.R.; Prabhakaran, D.; Ramirez, A.; Schlaich, M.; Stergiou, G.S.; Tomaszewski, M.; et al. 2020 International Society of Hypertension Global Hypertension Practice Guidelines. Hypertension 2020, 75, 1334–1357. Available online: http://ahajournals.org (accessed on 13 December 2021). [CrossRef]
  15. Muntner, P.; Shimbo, D.; Carey, R.M.; Charleston, J.B.; Gaillard, T.; Misra, S.; Myers, M.G.; Ogedegbe, G.; Schwawartz, J.; Townsend, R.R.; et al. Measurement of Blood Pressure in Humans: A Scientific Statement From the American Heart Association. Hypertension 2019, 73, e35–e66. Available online: https://www.ahajournals.org/doi/abs/10.1161/HYP.0000000000000087 (accessed on 13 December 2021). [CrossRef]
  16. Parati, G.; Stergiou, G.; O’Brien, E.; Asmar, R.; Beilin, L.; Bilo, G.; Clement, D.; de la Sierra, A.; de Leuuw, P.; Dolan, E.; et al. European Society of Hypertension practice guidelines for ambulatory blood pressure monitoring. J. Hypertens. 2014, 32, 1359–1366. Available online: https://pubmed.ncbi.nlm.nih.gov/24886823/ (accessed on 14 December 2021). [CrossRef] [Green Version]
  17. Parati, G.; Stergiou, G.S.; Bilo, G.; Kollias, A.; Pengo, M.; Ochoa, J.E.; Agarwal, R.; Asayama, K.; Asmar, R.; Burnier, M.; et al. Home blood pressure monitoring: Methodology, clinical relevance and practical application: A 2021 position paper by the Working Group on Blood Pressure Monitoring and Cardiovascular Variability of the European Society of Hypertension. J. Hypertens. 2021, 39, 1742–1767. Available online: https://pubmed.ncbi.nlm.nih.gov/34269334/ (accessed on 14 December 2021). [CrossRef]
  18. O’Brien, E.; Parati, G.; Stergiou, G.; Asmar, R.; Beilin, L.; Bilo, G.; Clement, D.; de la Sierra, A.; de Leeuw, P.; Dolan, E.; et al. European Society of Hypertension position paper on ambulatory blood pressure monitoring. J. Hypertens. 2013, 31, 1731–1768. Available online: https://pubmed.ncbi.nlm.nih.gov/24029863/ (accessed on 14 December 2021). [CrossRef] [Green Version]
  19. Zhang, S.; Zhou, X.; Chen, Y.; Wang, L.; Zhu, B.; Jiang, Y.; Bu, P.; Liu, W.; Li, D.; Li, Y.; et al. Changes in Home Blood Pressure Monitored among Elderly Patients with Hypertension during the COVID-19 Outbreak: A Longitudinal Study in China Leveraging a Smartphone-Based Application. Circ. Cardiovasc. Qual. Outcomes 2021, 14, 605–612. Available online: https://www.ahajournals.org/doi/abs/10.1161/CIRCOUTCOMES.120.007098 (accessed on 14 December 2021). [CrossRef]
  20. Feitosa, F.G.A.M.; Feitosa, A.D.M.; Paiva, A.M.G.; Mota-Gomes, M.A.; Barroso, W.S.; Miranda, R.D.; Barbosa, E.C.D.; Brandão, A.A.; Lima-Filho, J.L.; Sposito, A.C.; et al. Impact of the COVID-19 pandemic on blood pressure control: A nationwide home blood pressure monitoring study. Hypertens. Res. 2021, 45, 364–368. Available online: https://pubmed.ncbi.nlm.nih.gov/34857897/ (accessed on 14 December 2021). [CrossRef]
  21. Citoni, B.; Figliuzzi, I.; Presta, V.; Volpe, M.; Tocci, G. Home Blood Pressure and Telemedicine: A Modern Approach for Managing Hypertension During and After COVID-19 Pandemic. High Blood Press Cardiovasc. Prev. 2021, 29, 1–14. Available online: https://pubmed.ncbi.nlm.nih.gov/34855154/ (accessed on 14 December 2021). [CrossRef]
  22. Narita, K.; Hoshide, S.; Kario, K. Nighttime Home Blood Pressure Is Associated with the Cardiovascular Disease Events Risk in Treatment-Resistant Hypertension. Hypertension 2022, 79, e18–e20. Available online: https://www.ahajournals.org/doi/abs/10.1161/HYPERTENSIONAHA.121.18534 (accessed on 14 December 2021). [CrossRef] [PubMed]
  23. Bryant, K.B.; Green, M.B.; Shimbo, D.; Schwartz, J.E.; Kronish, I.M.; Zhang, Y.; Sheppard, J.P.; McManus, R.J.; Moran, A.E.; Bellows, B.K. Home Blood Pressure Monitoring for Hypertension Diagnosis by Current Recommendations: A Long Way to Go. Hypertension 2021, 79, e15–e17. Available online: https://pubmed.ncbi.nlm.nih.gov/34852639/ (accessed on 15 December 2021). [CrossRef] [PubMed]
  24. O’Brien, E.; Petrie, J.; Littler, W.; De Swiet, M.; Padfield, P.L.; O’Malley, K.; Jamieson, M.; Altman, D.; Bland, M.; Atkins, N. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems. J. Hypertens. 1990, 8, 607–619. Available online: https://pubmed.ncbi.nlm.nih.gov/2168451/ (accessed on 15 December 2021). [CrossRef] [PubMed] [Green Version]
  25. Stergiou, G.S.; Alpert, B.; Mieke, S.; Asmar, R.; Atkins, N.; Eckert, S.; Frick, G.; Friedman, B.; Graßl, T.; Ichikawa, T.; et al. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J. Hypertens. 2018, 71, 368–374. Available online: https://pubmed.ncbi.nlm.nih.gov/29386350/ (accessed on 15 December 2021). [CrossRef] [PubMed]
  26. O’Brien, E.; Pickering, T.; Asmar, R.; Myers, M.; Parati, G.; Staessen, J.; Mengden, T.; Imai, Y.; Waeber, B.; Palatini, P.; et al. Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults. Blood Press. Monit. 2002, 7, 3–17. Available online: https://pubmed.ncbi.nlm.nih.gov/12040236/ (accessed on 15 December 2021). [CrossRef] [PubMed]
  27. O’Brien, E.; Atkins, N.; Stergiou, G.; Karpettas, N.; Parati, G.; Asmar, R.; Imai, Y.; Wang, J.; Mengden, T.; Shennan, A.; et al. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press. Monit. 2010, 5, 23–38. Available online: https://pubmed.ncbi.nlm.nih.gov/20110786/ (accessed on 15 December 2021). [CrossRef] [PubMed]
  28. Stergiou, G.S.; Alpert, B.; Mieke, S.; Asmar, R.; Atkins, N.; Eckert, S.; Frick, G.; Friedman, B.; Graßl, T.; Ichikawa, T.; et al. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J. Hypertens. 2018, 36, 472–478. [Google Scholar] [CrossRef]
  29. Stergiou, G.S.; Palatini, P.; Asmar, R.; Ioannidis, J.P.; Kollias, A.; Lacy, P.; McManus, R.J.; Myers, M.G.; Parati, G.; Shennan, A.; et al. Recommendations and Practical Guidance for performing and reporting validation studies according to the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO). J. Hypertens. 2019, 37, 459–466. Available online: https://pubmed.ncbi.nlm.nih.gov/30702492/ (accessed on 15 December 2021).
  30. Baethge, C.; Goldbeck-Wood, S.; Mertens, S. SANRA—A scale for the quality assessment of narrative review articles. Res. Integr. Peer Rev. 2019, 4, 5. Available online: https://pubmed.ncbi.nlm.nih.gov/30962953/ (accessed on 15 December 2021). [CrossRef] [Green Version]
  31. Vandenbroucke, J.P.; von Elm, E.; Altman, D.G.; Gøtzsche, P.C.; Mulrow, C.D.; Pocock, S.J.; Poole, C.; Schlesselman, J.J.; Egger, M. Strengthening the Reporting of Observational Studies in Epidemiology (STROBE): Explanation and elaboration. Int. J. Surg. 2014, 12, 1500–1524. [Google Scholar] [CrossRef] [Green Version]
  32. Abou-Dakn, M.; Döhmen, C.; Wenzel, S. Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010. J. Hum. Hypertens. 2017, 31, 89–92. Available online: https://pubmed.ncbi.nlm.nih.gov/27411300/ (accessed on 8 January 2022). [CrossRef]
  33. Azaki, A.; Diab, R.; Harb, A.; Asmar, R.; Chahine, M.N. Questionable accuracy of home blood pressure measurements in the obese population—Validation of the Microlife WatchBP O3® and Omron RS6® devices according to the European Society of Hypertension-International Protocol. Vasc. Health Risk Manag. 2017, 13, 61–69. Available online: https://pubmed.ncbi.nlm.nih.gov/28280348/ (accessed on 9 January 2022). [CrossRef] [Green Version]
  34. Chen, Q.; Kang, Y.Y.; Li, Y.; Wang, J.G. Validation of the BPUMP BF1112 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2017, 22, 105–108. Available online: https://pubmed.ncbi.nlm.nih.gov/28072599/ (accessed on 9 January 2022). [CrossRef]
  35. Liu, Z.Y.; Zhang, Q.H.; Ye, X.L.; Liu, D.P.; Cheng, K.; Zhang, C.H.; Wan, Y. Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols. Blood Press. Monit. 2017, 22, 101–104. Available online: https://https://pubmed.ncbi.nlm.nih.gov/28177942/ (accessed on 9 January 2022). [CrossRef]
  36. Beime, B.; Deutsch, C.; Krüger, R.; Wolf, A.; Müller, P.; Hammel, G.; Bramlage, P. Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010. Eur. J. Pediatr. 2017, 176, 573–580. Available online: https://pubmed.ncbi.nlm.nih.gov/28236027/ (accessed on 9 January 2022). [CrossRef]
  37. Fania, C.; Albertini, F.; Palatini, P. Validation of the A&D UM-201 device for office blood pressure measurement according to the European Society of Hypertension International Protocol Revision 2010. Blood Press. Monit. 2017, 22, 234–237. Available online: https://pubmed.ncbi.nlm.nih.gov/28362645/ (accessed on 10 January 2022).
  38. Kang, Y.Y.; Chen, Q.; Liu, C.Y.; Li, Y.; Wang, J.G. Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2017, 22, 230–233. Available online: https://pubmed.ncbi.nlm.nih.gov/28383290/ (accessed on 10 January 2022). [CrossRef]
  39. Fania, C.; Albertini, F.; Palatini, P. Validation of the A&D UM-211 device for office blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2017, 22, 302–305. Available online: https://pubmed.ncbi.nlm.nih.gov/28520592/ (accessed on 10 January 2022).
  40. Chen, Q.; Lei, L.; Li, Y.; Wang, J.G. Validation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2017, 22, 295–297. Available online: https://pubmed.ncbi.nlm.nih.gov/28617717/ (accessed on 10 January 2022). [CrossRef]
  41. Grover-Páez, F.; Cardona-Muñoz, E.G.; Cardona-Müller, D.; Guzmán-Saldívar, V.H.; Rodríguez-De La Cerda, M.; Jiménez-Cázarez, M.B.; Totsuka-Sutto, S.E.; Alanis-Sánchez, G.A.; Ramos-Becerra, C.G. Validation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population. Blood Press. Monit. 2017, 22, 375–378. Available online: https://pubmed.ncbi.nlm.nih.gov/28945665/ (accessed on 15 January 2022).
  42. Chahine, M.N.; Topouchian, J.; Zelveian, P.; Hakobyan, Z.; Melkonyan, A.; Azaki, A.; Diab, R.; Harb, A.; Asmar, R. Validation of BP devices QardioArm® in the general population and Omron M6 Comfort® in type II diabetic patients according to the European Society of Hypertension International Protocol (ESH-IP). Med. Devices 2017, 11, 11–20. Available online: https://pubmed.ncbi.nlm.nih.gov/29343992/ (accessed on 15 January 2022). [CrossRef] [PubMed] [Green Version]
  43. Chen, L.; Li, J.; Wen, J.; Guo, C.; Zhang, J.; Yu, Z. Validation of the Pangao PG-800B26 upper arm blood pressure monitor in the general population according to the European Society of Hypertension and the British Hypertension Society protocols. Blood Press. Monit. 2018, 23, 41–44. Available online: https://pubmed.ncbi.nlm.nih.gov/28902676/ (accessed on 15 January 2022). [CrossRef] [PubMed]
  44. Zhao, H.; Qiao, W.; Zhang, R.; Cui, P.; Hou, F.; Zhang, W. Validation of the Pangao PG-800A36 automatic wrist blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols. Blood Press. Monit. 2018, 23, 37–40. Available online: https://pubmed.ncbi.nlm.nih.gov/28926362/ (accessed on 15 January 2022). [CrossRef] [PubMed]
  45. Kang, Y.Y.; Chen, Q.; Liu, C.Y.; Li, Y.; Wang, J.G. Validation of the AVITA BPM64 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2018, 23, 45–48. Available online: https://pubmed.ncbi.nlm.nih.gov/29049094/ (accessed on 15 January 2022). [CrossRef] [PubMed]
  46. Pereira, T.; Guimarães, J. Validation of the Beneware model ABP-021 ambulatory blood pressure monitor according to the revised 2010 European Society of hypertension international protocol. Blood Press. Monit. 2018, 23, 210–213. Available online: https://pubmed.ncbi.nlm.nih.gov/29894314/ (accessed on 16 January 2022). [CrossRef]
  47. Mazoteras-Pardo, V.; Becerro-De-Bengoa-Vallejo, R.; Losa-Iglesias, M.E.; López-López, D.; Palomo-López, P.; Rodríguez-Sanz, D.; Calvo-Lobo, C. The QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010. JMIR mHealth uHealth. 2018, 6, e11632. Available online: https://pubmed.ncbi.nlm.nih.gov/30361193/ (accessed on 16 January 2022). [CrossRef]
  48. Abou-Dakn, M.; Wenzel, S. Validation of the PHYSIO-PORT UP ambulatory blood pressure monitor in pregnant women according to the European Society of Hypertension International Protocol revision 2010. J. Hum. Hypertens. 2018, 32, 770–774. Available online: https://pubmed.ncbi.nlm.nih.gov/30232401/ (accessed on 16 January 2022). [CrossRef]
  49. Kollias, A.; Stambolliu, E.; Kyriakoulis, K.G.; Papadatos, S.S.; Stergiou, G.S. Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO320 for public use according to the 2010 European Society of Hypertension International Protocol. Blood Press. Monit. 2019, 24, 30–32. Available online: https://pubmed.ncbi.nlm.nih.gov/30531495/ (accessed on 17 January 2022). [CrossRef]
  50. Saito, K.; Hishiki, Y.; Takahashi, H. Validation of two automatic devices, Omron HEM-6232T and HEM-6181, for self-measurement of blood pressure at the wrist according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the European Society of Hypertension International Protocol revision 2010. Vasc. Health Risk Manag. 2019, 15, 47–55. Available online: https://pubmed.ncbi.nlm.nih.gov/30881007/ (accessed on 17 January 2022). [CrossRef] [Green Version]
  51. Mazoteras-Pardo, V.; Becerro-De-Bengoa-Vallejo, R.; Losa-Iglesias, M.E.; López-López, D.; Palomo-López, P.; Rodríguez-Sanz, D.; Calvo-Lobo, C. Validation in the General Population of the iHealth Track Blood Pressure Monitor for Self-Measurement According to the European Society of Hypertension International Protocol Revision 2010: Descriptive Investigation. JMIR mHealth uHealth. 2019, 7, e13137. Available online: https://pubmed.ncbi.nlm.nih.gov/30888331/ (accessed on 17 January 2022). [CrossRef]
  52. Reshetnik, A.; Gohlisch, C.; Abou-Dakn, M.; Tölle, M.; Zidek, W.; Van Der Giet, M. Validation of noninvasive oscillometric blood pressure 2020 up pressure upper arm blood pressure monitoring technology according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2019, 24, 99–101. Available online: https://pubmed.ncbi.nlm.nih.gov/30762596/ (accessed on 17 January 2022). [CrossRef]
  53. Fania, C.; Vezzù, L.; Lazzaretto, I.; Palatini, P. Validation of the Hingmed WBP-02A device for ambulatory blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2019, 24, 151–154. Available online: https://pubmed.ncbi.nlm.nih.gov/30807305/ (accessed on 17 January 2022). [CrossRef]
  54. Liu, Z.; Chen, L. Validation of Transtek TMB-1776 according to European Society of Hypertension International Protocol revision 2010 in adults. Blood Press. Monit. 2019, 24, 319–322. Available online: https://pubmed.ncbi.nlm.nih.gov/31567185/ (accessed on 18 January 2022). [CrossRef]
  55. Mazoteras-Pardo, V.; Becerro-De-Bengoa-Vallejo, R.; Losa-Iglesias, M.E.; López-López, D.; Rodríguez-Sanz, D.; Casado-Hernández, I.; Calvo-Lobo, C.; Palomo-López, P. QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010. J. Med. Internet Res. 2019, 21, e14686. Available online: https://pubmed.ncbi.nlm.nih.gov/31789600/ (accessed on 18 January 2022). [CrossRef]
  56. Kollias, A.; Anagnostopoulos, I.; Gravvani, A.; Stambolliu, E.; Bountzona, I.; Menti, A.; Stergiou, G.S. Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Press. Monit. 2020, 25, 50–52. Available online: https://pubmed.ncbi.nlm.nih.gov/31633518/ (accessed on 18 January 2022).
  57. Zhang, W.; Lei, L.; Li, Y.; Wang, J.G. Validation of the HL868ED upper-arm blood pressure monitor for clinical use and self-measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2020, 25, 53–57. Available online: https://pubmed.ncbi.nlm.nih.gov/31658108/ (accessed on 18 January 2022). [CrossRef]
  58. Kollias, A.; Gravvani, A.; Anagnostopoulos, I.; Kyriakoulis, K.G.; Bountzon, I.; Menti, A.; Stergiou, G.S. Validation of the InBody BPBIO250 oscillometric blood pressure monitor for professional office use in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Press. Monit. 2020, 15, 115–117. Available online: https://pubmed.ncbi.nlm.nih.gov/31913149/ (accessed on 18 January 2022).
  59. Saito, K.; Hishiki, Y.; Takahashi, H. Validation of the Omron HBP-1320 for professional use according to the ANSI/AAMI/ISO 81060-2: 2013 protocol and the 2010 revision of the European Society of Hypertension International Protocol. Blood Press. Monit. 2020, 25, 162–166. Available online: https://pubmed.ncbi.nlm.nih.gov/32118675/ (accessed on 18 January 2022). [CrossRef]
  60. Song, C.; Yu, Y.; Lu, B.C.; Yan, X.L. Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2020, 25, 291–294. Available online: https://pubmed.ncbi.nlm.nih.gov/32898351/ (accessed on 18 January 2022). [CrossRef]
  61. Omar, S.M.; Ali, E.A.; Ibrahim, Y.; Al-Wutayd, O.; Adam, I. Validation of the wrist blood pressure measuring device Omron RS6 (HEM-6221-E) among obese Sudanese patients according to the European Society of Hypertension International Protocol Revision 2010. F1000Resarch 2020, 9, 1284. Available online: https://pubmed.ncbi.nlm.nih.gov/34691394/ (accessed on 20 January 2022).
  62. Ntineri, A.; Prapa, S.; Bountzona, I.; Menti, A.; Stergiou, G.S. Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO750 for public spaces according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard. Blood Press. Monit. 2021, 26, 146–148. Available online: https://pubmed.ncbi.nlm.nih.gov/33323723/ (accessed on 20 January 2022). [PubMed]
  63. Chahine, M.N.; Harb, S.B.; Saad, A.R.; Sarkis, P.; Azaki, A.; Harb, A.; Allouch, A.; Asmar, R. Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2021, 26, 237–241. Available online: https://pubmed.ncbi.nlm.nih.gov/33661139/ (accessed on 20 January 2022). [CrossRef] [PubMed]
  64. Deutsch, C.; Bramlage, C.; Botta, B.; Krüger, R.; Forstner, K.; Bramlage, P.; Beime, B. Validation of the blood pressure measurement device Beurer BM 28 according to the European Society of Hypertension International Protocol revision 2010. Blood Press. Monit. 2021, 26, 292–298. Available online: https://pubmed.ncbi.nlm.nih.gov/33741775/ (accessed on 20 January 2022). [CrossRef] [PubMed]
  65. Zhang, H.-J.; Zhang, J.; Wang, S.-L.; Zhang, J.; Teng, L.-N.; Zhang, S.-J.; Zhou, D.-J.; Long, M.-Z. Validation of the YuWell YE900 oscillometric blood pressure monitor for professional office use in adults and children according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018). Blood Press. Monit. 2021, 26, 396–399. Available online: https://pubmed.ncbi.nlm.nih.gov/34480474/ (accessed on 20 January 2022). [CrossRef] [PubMed]
  66. Ntineri, A.; Menti, A.; Konstantinos, G.; Bountzona, K.J.; Prapa, S.; Kollias, G.S.S. Validation of the InBody BPBIO210 manual auscultatory hybrid device for professional office use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization. Blood Press. Monit. 2021, 27, 135–138. [Google Scholar] [CrossRef]
  67. Zhang, P.; Li, X.; Fang, Z.; Lu, Y.; Cui, J.; Du, X.; Hu, R. Smartphone application-supported validation of three automatic devices for self-measurement of blood pressure according to the European Society of Hypertension International Protocol revision 2010: The Omron HEM-7120, Yuwell YE680A and Cofoe KF-65B. Blood Press. Monit. 2021, 26, 435–440. Available online: https://pubmed.ncbi.nlm.nih.gov/34001755/ (accessed on 20 January 2022). [CrossRef]
  68. Degott, J.; Ghajarzadeh-Wurzner, A.; Hofmann, G.; Proença, M.; Bonnier, G.; Lemkaddem, A.; Lemay, M.; Christen, U.; Knebel, J.-F.; Durgnat, V.; et al. Smartphone based blood pressure measurement: Accuracy of the OptiBP mobile application according to the AAMI/ESH/ISO universal validation protocol. Blood Press. Monit. 2021, 26, 441–448. Available online: https://pubmed.ncbi.nlm.nih.gov/34139747/ (accessed on 20 January 2022). [CrossRef]
  69. Taylor, P.; Berg, C.; Thompson, J.; Dean, K.; Yuan, T.; Nallamshetty, S.; Tong, I. Effective Access to Care in a Crisis Period: Hypertension Control During the COVID-19 Pandemic by Telemedicine. Mayo Clin. Proc. Innov. Qual. Outcomes 2022, 6, 19–26. Available online: https://pubmed.ncbi.nlm.nih.gov/34805763/ (accessed on 23 January 2022). [CrossRef]
  70. Wu, D.; Huyan, X.; She, Y.; Hu, J.; Duan, H.; Deng, N. Exploring and Characterizing Patient Multibehavior Engagement Trails and Patient Behavior Preference Patterns in Pathway-Based mHealth Hypertension Self-Management: Analysis of Use Data. JMIR mHealth uHealth 2022, 10, e33189. Available online: https://pubmed.ncbi.nlm.nih.gov/35113032/ (accessed on 23 January 2022). [CrossRef]
  71. Bricca, A.; Pellegrini, A.; Zangger, G.; Ahler, J.J.; Jäger, M.; Skou, S.T. The Quality of Health Apps and Their Potential to Promote Behavior Change in Patients With a Chronic Condition or Multimorbidity: Systematic Search in App Store and Google Play. JMIR mHealth uHealth 2022, 10, e33168. Available online: https://pubmed.ncbi.nlm.nih.gov/35119367/ (accessed on 23 January 2022). [CrossRef]
  72. Stergiou, G.S.; Asmar, R.; Myers, M.; Palatini, P.; Parati, G.; Shennan, A.; Wang, J.; O’Brien, E.; European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability. Improving the accuracy of blood pressure measurement: The influence of the European Society of Hypertension International Protocol (ESH-IP) for the validation of blood pressure measuring devices and future perspectives. J. Hypertens. 2018, 36, 479–487. Available online: https://pubmed.ncbi.nlm.nih.gov/29384984/ (accessed on 23 January 2022). [CrossRef]
  73. O’Brien, E.; Stergiou, G.; Palatini, P.; Asmar, R.; Ioannidis, J.P.; Kollias, A.; Lacy, P.; McManus, R.J.; Myers, M.G.; Shennan, A.; et al. Validation protocols for blood pressure measuring devices: The impact of the European Society of Hypertension International Protocol and the development of a Universal Standard. Blood Press. Monit. 2019, 24, 163–166. Available online: https://pubmed.ncbi.nlm.nih.gov/31116156/ (accessed on 23 January 2022). [CrossRef]
  74. Alpert, B.S.; Sarkis, J.; Dart, R.A.; Quinn, D.; Friedman, B.; Townsend, R.R.; Shimbo, D. Future use of the European Society of Hypertension International Protocol for validation of automated sphygmomanometers. Blood Press. Monit. 2019, 24, 161–162. Available online: https://pubmed.ncbi.nlm.nih.gov/31116154/ (accessed on 23 January 2022). [CrossRef]
  75. Bramer, W.M.; Rethlefsen, M.L.; Kleijnen, J.; Franco, O.H. Optimal database combinations for literature searches in systematic reviews: A prospective exploratory study. Syst. Rev. 2017, 6, 245. [Google Scholar] [CrossRef]
Figure 1. Flow diagram of the narrative review in the Pubmed database.
Figure 1. Flow diagram of the narrative review in the Pubmed database.
Jpm 13 00009 g001
Table 1. Characteristics of the main validation protocols.
Table 1. Characteristics of the main validation protocols.
ESH Protocol
Review 2010
AAMI/ESH/ISO
Universal Standard
ISO 81160-2:2018
Validation Team1 independent supervisor, 2 observers trained in BP 1 independent supervisor, ≥ 2 observers trained in BP
Sample Men and women ≥ 25 years oldMales and females ≥ 12 years old
Sample Size33 subjects (≥10 men; ≥10 women)85 subjects (≥30% men; ≥30% women)
Recruitment BP
mm/hg
SBP:
10–12 subjects in 3 groups of SBP: 90–129, 130–160, 160–180
DBP:
10–12 subjects in 3 groups of DBP: 40–79,80–100, 100–130
SBP readings:
≥ 5% ≤100; ≥ 5% must be ≥ 160 and ≥ 20% ≥ 140
DBP readings:
≥ 5% ≤ 60; ≥ 5% ≥ 100
and ≥ 20% ≥ 85
Cuff sizesNot applicableFor test devices with a single cuff:
≥40% subjects must have an arm circumference within the (upper half) of the specified range of use of the cuff; ≥40% (lower half); ≥20% (higher quarter; ≥20% (lower quarter); ≥10% (higher octal); ≥10% (lower octal)
Measurement conditionsRoom: Quiet, isolated, comfortable temperature.
Subject: Relaxed ≥ 5 min, empty bladder, silent, sitting, feet resting on the floor without crossing their legs, their arm on a flat surface at the level of the heart, palm of the hand upwards.
Idem
Measurement method
9 sequential measurements (2 initial + 7 validation) alternating two devices
Same sequential arm
-Gold standard: Two standard mercury sphygmomanometers with nonelectronic stethoscopes (observers)
-Test instrument (supervisor)
30–60 s between measurements
-Gold standard: The auscultatory method with dual-head stethoscope (observers).
-Test instrument (supervisor)
60 s between measurements
The opposite arm simultaneous method can also be used.
Analysis method3 comparisons per subject =
99 pairs of measurements.
Each test measurement is compared with two references (the one before and the one after) and the smallest difference remains.
2 validation phases, based on absolute BP differences within 5, 10, and 15 mm/hg
Bland–Altman graphics
3 comparisons per subject = 255 paired BP comparisons
Each of the test device measurements is compared against the average of the previous and succeeding reference BP readings.
Differences are calculated by subtracting the reference from the test device measurement.
2 validation phases:
  • BP differences ≤ 5 and SD ≤8 mm/hg
  • For individual subjects, the SD of 85 averaged BP differences must be within a threshold defined. Only applicable in samples of 85 subjects
Bland–Altman graphics
Exclusion criteriaArrhythmia; Other problems during the validation
Reasons for exclusion reported
Idem
Specialized populationsNo particular conditions are indicated Suggests adapting the conditionsSpecial conditions: Children ≤ 3 years, pregnant; arm circumference ≥ 42 cm
AAMI, Association for Advancement of Medical Instrumentation; BP, blood pressure; cm, centimeters; DBP, diastolic blood pressure; ESH, European Society of Hypertension; ISO, International Organization for Standardization; mm/hg, millimeters of mercury; s, seconds; SBP, systolic blood pressure; SD, standard deviation.
Table 2. Comparative table of validation studies.
Table 2. Comparative table of validation studies.
PUBLICATIONPROTOCOLBP DEVICES POPULATION RESULT
2017. Abou-Dakn et al. [32]ESH-IP 2010Gold standard: ERKAMETER 3000 (ERKA, Bad Tölz, Ger-many), mercury sphygmomanometer
Test device: TONOPORT VI, oscillometric, ABPM
General populationTONOPORT VI passed
2017. Azaki et al. [33]ESH-IP 2010Gold standard:
two mercury sphygmomanometers and stethoscope
2 Test devices:
OMRON RS6 (HEM-6221-E), automatic oscillometric, HBPM, wrist
MICROLIFE WATCHBP O3 (BP 3 MZ1-1), automatic oscillometric office, ABPM, HBBM, arm
Obese populationOMRON RS6, MICROLIFE WATCHBP O3 failed
2017. Chen et al. [34]ESH-IP 2010Gold standard:
a mercury sphygmomanometer
Test device:
BPUMP BF1112, automated electronic digital oscillometric, HBPM, arm
General populationBPUMP BF1112 passed
2017. Liu et al. [35]ESH-IP 2010, BHS protocol; ISO 81060-2:2013 criteriaGold standard: mercury sphygmomanometers (Yuyue; Yuyue
Medical Equipment Co., Ltd., Jiangsu, China) with a double stethoscope
Test device:
G.LAB
MD2200, automated oscillometric, wrist
General populationG.LAB MD2200 passed
2017. Beime et al. [36]ESH-IP 2010 modified according to KIGGS Gold standard: ERKAMETER 3000 (ERKA, Bad Tölz, Ger-many), mercury sphygmomanometer with a double stethoscope (ErkaPhon).
Test device:
CUSTO SCREEN PEDIATRIC, oscillometric, automatic zero balancing, ABPM of children
Child populationCUSTO SCREEN PEDIATRIC passed
2017. Fania et al. [37]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
A&D UM-201 (Tokyo, Japan), automatic oscillometric, office BP, arm
General populationA&D UM-201 passed
2017. Kang et al. [38]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
AVITA BPM17 (AVITA Medical,
Taipei, Taiwan), automatic oscillometric digital, HBPM, wrist
General populationAVITA BPM17 passed
2017. Fania et al. [39]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
A&D UM-211 (Tokyo, Japan), automatic oscillometric, office BP measurement, arm
General populationA&D UM-211 passed
2017. Chen Q et al. [40]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
YUWELL YE690A (Danyang, Jiangsu
Province, China), automatic digital oscillometric, office BP, HBPM, arm
General populationYUWELL YE690A passed
2017. Grover-Páez et al. [41]ESH-IP 2010Gold standard: mercury sphygmomanometer

Test device:
OMRON HEM-7320-LA (Omron Healthcare Co., Ltd., Kyoto, Japan), automatic intelligent wrap, oscillometric, office BP, HBPM, arm
General populationOMRON HEM-7320-LA passed
2017. Chahine et al. [42]ESH-IP 2010Gold standard: mercury sphygmomanometer
2 Test devices:
QARDIOARM (San Francisco, CA, USA), automatic oscillometric, HBPM, upper, connected by bluetooth to mobile phones and tablets through the Qardio app
OMRON M6 IT COMFORT® HEM-7322U-E (Omron Healthcare Co., Ltd.,Kyoto, Japan), automatic digital oscillometric, HBPM, arm
1: General population
2: Non-insulin-dependent type II diabetic
QARDIOARM,
OMRON M6 passed
2018. Chen L. et al. [43]ESH-IP 2010; BHS ProtocolGold standard: a mercury sphygmomanometer with Y-tubing dual-head stethoscope
Test device:
PANGAO PG-800B26 (Shenzhen Pango Electronic Co., Ltd., Shenzhen,
China), automatic electronic oscillometric, HBPM, arm
General populationPANGAO PG-800B26 passed
2018. Zhao et al. [44]ESH-IP 2010; BHS ProtocolGold standard: a mercury sphygmomanometer
Test device:
PANGAO PG-800A36 (Shenzhen Pango Electronic Co., Ltd., Shenzhen, China), automatic electronic oscillometric, HBPM, wrist
General populationPANGAO PG 800A36 passed
2018. Kang et al. [45]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
AVITA BPM64 (AVITA Medical,
Taipei, Taiwan), an automated electronic digital oscillometric, HBPM, arm
General populationAVITA BPM64 passed
2018. Pereira et al. [46]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
BENEWARE MODEL ABP-021 (Suzhou Beneware Medical Equipment Co., Ltd., Hangzhou, China), automatic oscillometric ABPM, arm
This has a built-in USB communication
interface enabling connection with a PC operating the dedicated ABPM analysis software.
General populationBENEWARE MODEL ABP-021 passed
2018. Mazoteras-Pardo et al. [47]ESH-IP 2010Gold standard: mercury sphygmomanometer.
Test device:
QARDIOARM (Qardioarm, Atten Electronic Co., Dongguan, China), automatic oscillometric HBPM, arm connected by bluetooth to mobile phones and tablets through the Qardio app
Obese populationQARDIOARM passed
2018. Abou-Dakn et al. [48]ESH-IP 2010Gold standard: ERKAMETER 3000 (ERKA, Bad Tölz, Germany), mercury sphygmomanometer
Test device:
PHYSIO-PORT UP (AR Medizintechnik GmbH & Co.KG, Berlin, Germany), oscillometric ABPM, arm
PregnantPHYSIO-PORT UP passed
2019. Kollias et al. [49]ESH-IP 2010Gold standard: Two mercury sphygmomanometers (Baumanometer; WA Baum Co., Inc., New York City, NY, USA) with stethoscope (3M Littmann Classic II
SE; 3M, St Paul, MN, USA)
Test device:
INBODY BPBIO320 (InBody, Seoul, Korea), kiosk-type automated oscillometric HBPM, right arm, public spaces. It has a fixed hole to insert the user’s arm, with an implanted single cuff arm.
General populationINBODY BPBIO320 passed
2019. Saito et al. [50]ESH-IP 2010; ANSI/AAMI/ISO 81060-2:2013 protocolGold standard: mercury sphygmomanometer
2 Test devices:
OMRON HEM-6232T (Omron Healthcare
Co., Ltd., Kyoto, Japan), automatic oscillometric with a sensor for the angle of the forearm, wrist
OMRON HEM-6181 (Omron Healthcare Co. Ltd., Kyoto, Japan), automatic oscillometric, wrist
General populationOMRON HEM-6232T, OMRON HEM-6181 passed
2019. Mazoteras-Pardo et al. [51]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
IHEALTH TRACK (KN-550BT; iHealthLabs Europe, Paris, France), automatic oscillometric, arm
General population IHEALTH TRACK passed
2019. Reshetnik et al. [52]ESH-IP 2010Gold standard: ERKAMETER 3000 (ERKA, Bad Tölz, Germany), mercury sphygmomanometer
Test device:
NIBP2020 UP technology (PAR Medizintechnik GmbH and Co. KG, Berlin, Germany), an electronic board, which must be installed into a host system: BP+ (Uscom Ltd., Sydney, Australia) has been used as a host device, automatic oscillometry, arm
General populationNIBP2020 UP technology passed
2019. Fania et al. [53]ESH-IP 2010 Gold standard: mercury sphygmomanometer
Test device:
HINGMED WBP-02A (Hingmed Company Shenzhen, China), automated oscillometric for ABPM 24 h, arm
General populationHINGMED
WBP-02A passed
2019. Liu et al. [54]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
TRANSTEK TMB-1776 (Guangdong Transtek Medical
Electronics Co., Ltd., Zhongshan, China)
an automatic oscillometric, HBPM, arm
General populationTRANSTEK TMB-1776
passed
2019. Mazoteras-Pardo et al. [55]ESH-IP 2010Gold standard: Omron M3 Intellisense (Omron Healthcare)
Test device:
QARDIOARM (Qardioarm, Atten Electronic Co., Dongguan, China), automatic oscillometric, HBPM, upper connected by bluetooth to mobile phones and tablets through the Qardio app
Patients with chronic kidney diseaseQARDIOARM
passed
2020. Kollias et al. [56] AAMI/ESH/ISO universal standard
(ISO 81160-2:2018)
Gold standard: two mercury sphygmomanometers
Test device:
INBODY BP170
an automated oscillometric, HBPM, upper
General populationINBODY BP170
passed
2020. Zhang et al. [57]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
HL868ED, automated electronic digital oscillometric, HBPM, office BP, arm
General populationHL868ED
passed
2020. Kollias et al. [58]AAMI/ESH/ISO universal standard
(ISO 81160-2:2018)
Gold standard: two mercury sphygmomanometers (Baumanometer; WA Baum Co., Inc., New York, NY, USA)
Test device:
INBODY BPBIO250, automated oscillometric, office BP, arm
General populationINBODY BPBIO250
passed
2020. Saito et al. [59]ESH-IP 2010; ANSI/AAMI/ISO 81060-2:2013 protocol
Gold standard: mercury sphygmomanometer
Test device:
OMRON HBP-1320 (Omron Healthcare
Co., Ltd., Kyoto, Japan), upper arm BP device. Both the automatic oscillometric mode and the auscultation mode can be selected.
Oscillometric mode has been used.
General populationOMRON HBP-1320 passed
2020. Song et al. [60]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
GLOBALCARE GCE603 (Globalcare Medical
Technology CO., LTD., Zhongshan, Guangdong, China), an automatic digital oscillometric, HBPM, arm
General populationGLOBALCARE GCE603
passed
2020. Omar et al. [61]ESH-IP 2010Gold standard: Two mercury sphygmomanometers and two non-electronic stethoscopes
Test device:
OMRON RS6 (HEM-6221-E),
automatic digital oscillometric wrist device for HBPM.
Obese populationOMRON RS6 (HEM-6221-E)
passed
2021. Ntineri et al. [62]AAMI/ESH/ISO universal standard
(ISO 81160-2:2018)
Gold standard: mercury sphygmomanometer
Test device:
INBODY BPBIO750 (InBody, Seoul, Korea), kiosk-type automated oscillometric right arm, HBPM, public spaces.
General populationINBODY BPBIO750
passed
2021. Chachine et al. [63]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
PHILIPS DL8760, utomatic oscillometric, HBPM, arm
This unit is integrated into the cuff
General populationPHILIPS DL8760
passed
2021. Deutsch et al. [64]ESH-IP 2010Gold standard: mercury sphygmomanometer
Test device:
BEURER BM 28, automatic digital oscillometric, HBPM, arm
General populationBEURER BM 28
passed
2021. Zhang et al. [65]AAMI/ESH/ISO universal standard
(ISO 81160-2:2018)
Gold standard: mercury sphygmomanometer
Test device:
YUWELL YE900, automatic digital oscillometric, office BP, arm. 3 cuffs
General population and childrenYUWELL YE900 passed
2021. Ntineri et al.
[66]
AAMI/ESH/ISO universal standard
(ISO 81160-2:2018)
Gold standard: mercury sphygmomanometer
Test device:
INBODY BPBIO210 (InBody, Seoul, Korea), a manual auscultatory mercury-free hybrid, office BP, arm, 2 cuffs
General populationINBODY BPBIO210
passed
2021. Zhang et al. [67]ESH-IP 2010Gold standard: mercury sphygmomanometer
3 Test devices:
OMRON HEM-7120 (OMRON Healthcare Co., Ltd., China);
YUWELL YE680A (Jiangsu Yuwell Medical Equipment Co., Ltd.); COFOE KF-65B (Hunan Cofoe Medical Technology Development Co., Ltd.), automatic oscillometric, HBPM, arm
General populationOMRON HEM-7120
passed
YUWELL YE680A and
COFOE KF-65B failed
2021. Degott et al. [68]AAMI/ESH/ISO universal standard
(ISO 81160-2:2018) with modifications imposed by cuffless measurements.
Gold standard: A&D UM-101, mercury-free sphygmomanometer (A&D Company,
Ltd., Toshima Ku, Tokyo, Japan)
Test device:
OPTIBP MOBILE APPLICATION (Biospectal SA, Lausanne, CH) installed on a Samsumg Galaxy S7 (Samsung GEC, 26, Sangil-ro 6-gil, Gagdong-gu, Seoul, Korea), app that estimates BP by applying the fingertip to the mobile camera
General populationOPTIBP MOBILE APPLICATION passed
AAMI, Association for the Advancement of Medical Instrumentation; ABPM, ambulatory blood pressure monitoring; ANSI, American National Standards Institute; BHS, British Hypertension Society; BP, blood pressure; ESH-IP 2010, European Society of Hypertension International Protocol revision 2010; HBPM, home blood pressure monitoring; ISO, International Organization for Standardization; KIGGS, German Health Interview and Examination Survey for Children and Adolescents 2013.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Mazoteras-Pardo, V.; Gómez-Cantarino, S.; Ramírez-Jiménez, M.; Navarro-Flores, E.; Ugarte-Gurrutxaga, M.I. Validations of Blood Pressure Measuring Devices Using Recognized Protocols. J. Pers. Med. 2023, 13, 9. https://doi.org/10.3390/jpm13010009

AMA Style

Mazoteras-Pardo V, Gómez-Cantarino S, Ramírez-Jiménez M, Navarro-Flores E, Ugarte-Gurrutxaga MI. Validations of Blood Pressure Measuring Devices Using Recognized Protocols. Journal of Personalized Medicine. 2023; 13(1):9. https://doi.org/10.3390/jpm13010009

Chicago/Turabian Style

Mazoteras-Pardo, Victoria, Sagrario Gómez-Cantarino, Miguel Ramírez-Jiménez, Emmanuel Navarro-Flores, and María Idoia Ugarte-Gurrutxaga. 2023. "Validations of Blood Pressure Measuring Devices Using Recognized Protocols" Journal of Personalized Medicine 13, no. 1: 9. https://doi.org/10.3390/jpm13010009