Systematic Review

Sensing and Control Strategies Used in FES Systems Aimed at Assistance and Rehabilitation of Foot Drop: A Systematic Literature Review

by
Estefanía González-Graniel
1,
Jorge A. Mercado-Gutierrez
2,
Saúl Martínez-Díaz
1,*,†,
Iliana Castro-Liera
1,
Israel M. Santillan-Mendez
1,
Oscar Yanez-Suarez
3,
Ivett Quiñones-Uriostegui
2 and
Gerardo Rodríguez-Reyes
2,*,†
1
División de estudios de Posgrado e Investiagación, TecNM-Instituto Tecnológico de la Paz, La Paz 28080, Mexico
2
Instituto Nacional de Rehabilitación Luis Guillermo Ibarra Ibarra, Mexico City 14389, Mexico
3
Electrical Engineering Department, Universidad Autónoma Metropolitana—Unidad Iztapalapa, Mexico City 09340, Mexico
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this work.
J. Pers. Med. 2024, 14(8), 874; https://doi.org/10.3390/jpm14080874
Submission received: 30 June 2024 / Revised: 11 August 2024 / Accepted: 12 August 2024 / Published: 17 August 2024

Abstract

:
Functional electrical stimulation (FES) is a rehabilitation and assistive technique used for stroke survivors. FES systems mainly consist of sensors, a control algorithm, and a stimulation unit. However, there is a critical need to reassess sensing and control techniques in FES systems to enhance their efficiency. This SLR was carried out following the PRISMA 2020 statement. Four databases (PubMed, Scopus, Web of Science, Wiley Online Library) from 2010 to 2024 were searched using terms related to sensing and control strategies in FES systems. A total of 322 articles were chosen in the first stage, while only 60 of them remained after the final filtering stage. This systematic review mainly focused on sensor techniques and control strategies to deliver FES. The most commonly used sensors reported were inertial measurement units (IMUs), 45% (27); biopotential electrodes, 36.7% (22); vision-based systems, 18.3% (11); and switches, 18.3% (11). The control strategy most reported is closed-loop; however, most of the current commercial FES systems employ open-loop strategies due to their simplicity. Three main factors were identified that should be considered when choosing a sensor for gait-oriented FES systems: wearability, accuracy, and affordability. We believe that the combination of computer vision systems with artificial intelligence-based control algorithms can contribute to the development of minimally invasive and personalized FES systems for the gait rehabilitation of patients with FDS.

1. Introduction

The study of targeted rehabilitation has become an important aspect of stroke management. Stroke is a major cause of severe and long-term disability, often resulting in foot drop (FD), an impairment seen in 20% to 30% of stroke survivors. [1,2]. FD can result in gait related deficiencies. Is characterized by weakness or loss of control in the ankle and toe dorsiflexor muscles. Many survivors are permanently disabled and usually need physical therapy to regain their daily living abilities and prevent further loss of their remaining voluntary functions [1,3,4]. FD can arise from a variety of conditions, including muscular, neurological, spinal, autoimmune, and musculoskeletal disorders; these conditions frequently follow traumatic events such as vehicle accidents, sports and recreational activities, and medical procedures such as lumbosacral spine and hip replacement surgeries, among others [5,6].
However, survivors still retain excitable peripheral nerves and muscle tissues that may be re-established through functional electrical stimulation (FES) [7]. This method bypasses the biological lesion and delivers the necessary stimulation to induce muscle contractions [7]. FES is a rehabilitation tool for restoring motor skills in stroke survivors. By delivering electrical impulses through the skin and targeting specific muscles and nerves, it enables movements involved in specific functional tasks. This method is widely employed to help neurologically impaired individuals restore walking ability [8]. The aim of FES intervention is to enhance gait function by either substituting or assisting voluntary movements to achieve the desired motion [8,9]. Nevertheless, significant challenges remain despite ongoing advancements. A key issue lies in the structure of control strategies of neuroprostheses, which need to synchronize muscle movements with sensory feedback integration [8].
FES is a complex and dynamic process influenced by muscle fatigue, patient effort, and spasticity, which can vary unpredictably during treatment. Minor variations in electrode placement from day to day can significantly alter motor responses. To achieve a more robust FES system, it is essential to measure human motion and muscle contraction via lightweight, portable, and real-time-capable measurement systems [7].
Building on this understanding, in the following sections, two key blocks of FES systems are explored, sensor techniques and control strategies, in order to present an overview of which are used today, their limitations, and the advantages.

1.1. Sensing Techniques

FES systems consist mainly of sensors, a control algorithm, and a stimulation unit. Sensors offer crucial feedback to FES systems, allowing the control system to adjust stimulation outputs based on changes in parameters and environmental interactions [6]. Therefore, feedback sensors are a key element in the design of FES systems for the gait rehabilitation of patients with foot drop. Hence, having a quick reference of the state of the art in this topic would be of great help for new and potential researchers in the field. Such a reference would save them plenty of time, work, and resources, before testing different combinations. This systematic literature review (SLR) is focused on these topics, with the main aim of serving as a first reference and guide to FES researchers and designers, before embarking in technology development or clinical trials.
Wearable sensors, such as foot pressure insoles, foot switches, accelerometers, gyroscopes, inertial measurement units (IMUs), and electromyography (EMG) electrodes, have been extensively utilized in FES control strategies [8]. Particularly, IMUs have been used extensively in this field, since they integrate gyro-scopes, accelerometers, and magnetometers to estimate joint angles, gait, and angular velocities through sensor fusion techniques. These combined measures enable detailed motion monitoring in gait-oriented FES applications, enabling access to specific kinematic variables in real-time, like acceleration and orientation of relevant body segments, such as the foot, ankle, and lower and upper leg. The accuracy and precision of wearable sensing systems are influenced by the number and placement of sensors, their alignment with the body’s coordinate system, and the signal processing algorithms employed. Moreover, effective calibration methods in FES systems depend on sensors, such as non-restrictive sensor-to-segment calibration or automatic anatomical calibration, and are crucial for obtaining precise measurements and reliable data [10,11,12].
Muscle activity, specifically the recruitment of motor units, can be monitored using surface electromyography (sEMG). The sEMG signal represents the activity muscle in the form of a voltage that can be measured by surface electrodes and is closely related to the timing and intensity of muscle contraction [10]. sEMG signals are less favored in wearable gait, such as those based on FES, due to their complexity in acquisition and post-processing. Nevertheless, the sEMG signals evoked by electrical stimulation can be employed to predict the resultant joint torque, which provides a necessary prediction of the muscle response before achieving accurate joint torque controlled by FES [8].
Vision-based non-wearable technologies utilizing standard or depth cameras are increasingly used for tracking human motion. It has been reported recently that these cameras can extract the detailed information necessary for biomechanical analyses [11]. Some studies have explored the applicability of Microsoft Kinect as the only sensory input for the CL of an FES system [13,14]. In general, Kinect requires an RGB camera for image capture and an infrared camera for depth estimation to provide the three-dimensional coordinates of objects. Furthermore, using a sequence of images, it is possible to estimate the speed at which objects are moving. It also includes efficient algorithms that allow the detection of the body skeleton in real time. One of the advantages of Microsoft Kinect is that it allows the estimation of the distance and pattern of the moving objects in the scene [13]. Nevertheless, these studies were only tested for upper limb applications (assisted grasping).
Also, ultrasound (US) imaging-derived echogenicity signals have been used for measuring muscle activation [15]. Authors refer to US echogenicity as the ability to reflect US waves in the context of surrounding tissues, and it can be used to estimate its displacement. Echogenicity is obtained by calculating the average intensity of pixels, within a region of interest, in each frame of the generated images. However, due to the high computational cost of processing US images, the sampling rate allowed by this technique is low.
Another approach, which has apparently been less commonly applied for that purpose, is the use of binocular vision systems. A binocular vision system requires only two calibrated RGB cameras to estimate the three-dimensional distance between objects, via the triangulation of corresponding points in both cameras. The hardware required for this approach can be even cheaper than that required by depth cameras.
However, it remains unclear whether vision-based non-wearable technologies have been fully exploited beyond subjective assessment and documentation. The potential use of standard camera footage for motion analysis would require simpler setup requirements and fewer recording constraints, and could be integrated more seamlessly into daily research, clinical, and telemedical settings [11]. These characteristics could potentially result in lower costs, providing non-wearable marker-less tools with significant advantages over traditional gold-standard technologies, which are expensive and challenging to implement in everyday activities where FES systems are used and evaluated.

1.2. Control Strategies

Once human kinematics and muscle activity are measured, this information can be used to adjust the stimulation parameters automatically to the needs of the user in order to delay the onset of fatigue and to target desired movements in an optimal way. These needs can be addressed through the implementation of feedback control [10].
A variety of control strategies are utilized to replicate or support functions carried out by the central nervous system (CNS), enabling the activation of muscles for the execution of natural movements. These strategies can be classified into two modalities: open-loop (OL) or closed-loop (CL) [16].
An OL control strategy in FES is a straightforward yet dependable method for regulating stimulation timing. OL architectures are commonly utilized in medical environments. These systems offer less precise movement control as they depend on manual input to initiate stimulation delivery [8,17].
On the other hand, CL FES control strategies can enhance the stability and robustness of position and force control by adjusting various stimulation parameters. These strategies aim to correct model errors and address internal disturbances (such as muscle fatigue) and external disturbances (such as obstacles) through feedback information [16].

1.3. Objectives and Structure of the Paper

This SLR aims to identify which sensing and control methods are currently used in lower limb FES applications, and to carry out a systematic comparison of key features related to them, such as invasiveness, cost-effectiveness, ease of use, practicality and suitability for the target application.
This SLR can serve as a complement to other recent reviews [18] that explore control strategies and different type of sensors for upper and lower limb applications of FES. It is our goal to help researchers interested in FES by carrying out a process of selecting suitable combinations of sensing and control strategies for lower limb FES applications, taking into consideration technical and clinical parameters. For example making a more informed choice would help to implement more effective FES control systems for lower limbs. Once these variables are chosen, they directly influence the type of sensor. Although several papers have been published on the exploration of sensors and control strategies, many of them do not consider computer vision as a measurement tool for the controlled variable.
The main aim of this work is to perform a SLR of alternative measurement methods for parameters used in lower limb FES control systems, such as foot angle and velocity. A secondary aim is to explore current and potentially novel approaches for control and sensing in lower limb FES systems and applications, beyond classical approaches.
The remainder of this article is organized as follows. First, the Materials and Methods section details the search strategy and how the PRISMA methodology was followed to conduct this SLR. The analysis of each article reviewed is presented in a table format. In Section 3, the results are presented, addressing each of the research questions proposed in Section 2. Additionally, the main findings of the results are highlighted, suggesting a classification based on the analyzed data. Section 4 covers the discussion, where the implications of the findings of the SRL are mentioned and proposals are given for implementation in FES systems. Finally, a conclusion is given in Section 5.

2. Materials and Methods

The PRISMA methodology was followed to conduct the SLR search [19]. A set of four academic and scientific databases were searched, PubMed, Scopus, Wiley Online Library, and Web of Science, from 2010 to 2024. Records were imported to the web application rayyan.ai for filtering and classification. This SLR does not have registration information.

2.1. Search Strategy

The three steps involved in the manual literature search process are summarized in the PRISMA flow diagram in Figure 1. In the first step (Step 1—Identification), the titles of articles reporting sensing and control strategies were identified from electronic databases. It was specified that the search terms should be found in the titles, keywords, and abstracts in databases that allowed such specification, such as Web of Science and Scopus. Data collection was performed independently by two authors to reduce the risk of selection errors and selection bias. The database search concluded in April 2024. Then, data extraction from abstracts and keywords was performed, and duplicate records, unrelated studies, and articles published before 2010 were removed.
The second step was a more detailed review of the full-text articles (according to the inclusion and exclusion criteria), to assess the eligibility of the selected papers (Step 2-Screening). If the abstract did not indicate clearly whether or not the inclusion and exclusion criteria were met, the full-text paper was also read. In the last step (Step 3-Included), studies considered relevant and those on recent advances were selected for further analysis in this SLR. The last step of filtering was applied to papers after reading the full text, taking into consideration whether or not they reported any sensing or control strategy.

2.1.1. Types of Population

This SLR focuses on works where FES systems and interventions are designed or ap-plied for the assistance or rehabilitation of patients with FD. However, FES technology is particularly useful for the treatment of foot drop of central origin, when peripheral structures (nerves, muscles, and joints) are functional. Additionally, this SLR was not limited only to poststroke patients, in order to obtain a more comprehensive overview. Finally, the review included those studies that tested FES systems or interventions in healthy volunteers and that reported on the development or preliminary testing of FES systems for patients with FD.

2.1.2. Types of Intervention

This review included studies that used FES systems for gait assistance or rehabilitation. Only studies where non-invasive, surface electrodes were used for the application of FES were included. Works using any type of invasive stimulation electrodes were excluded. Reports of hybrid systems that combined FES with another type of actuator such as exoskeletons were also included for the same purposes. Technical tests, proof-of concepts, pilot studies, and controlled clinical trials were included.

2.1.3. Types of Comparison

Due to the technical nature of this SLR, no comparison was performed in clinical terms. Only the applications of different control and sensor strategies employed in the different reports were compared.

2.1.4. Types of Outcomes

No outcomes were considered, since the purpose of the paper was to explore the application and combination of different control and sensing strategies in FES systems for FD assistance and rehabilitation.

2.2. Research Questions

As a complement to other reviews that explored control strategies and sensors for FES systems [18], our objective was to focus this research on drop foot FES applications and to expand the investigation to include vision systems as a new sensor. The aims of this SLR were translated into a set of research questions to better explain and summarize the evidence about FES systems for lower limb applications.
RQ1: What types of sensors are used to measure the body’s state and its response to FES?
RQ2: What variable is measured by this sensor?
RQ3: Which control strategies are employed in lower limb FES systems and applications?
RQ4: What stimulation parameters are modulated in CL lower limb FES applications?

2.3. Inclusion and Exclusion Criteria

The keywords used for a Boolean search through the databases were as follows: (“Foot Drop” OR “Drop Foot” OR “Dropped foot” OR “Foot drop syndrome”) AND (“Functional electrical stimulation” OR “Therapeutic Electrical Stimulation” OR “Electrotherapy” OR “Electric stimulation”) AND (“Electromyography” OR “EMG” OR “Electric Myography” OR “Elec-tromyographic techniques” OR “Inertial sensors” OR “Inertial sensing devices” OR “Inertial measurement units” OR “IMUs” OR “Computer vision” OR “Machine vision” OR “Vision systems” OR “Leap Motion Controller” OR “Kinect” OR “Stereoscopic vision” OR “ToF camera” OR “Structured light imaging”) AND (“Open-loop” OR “Open-loop control systems” OR “Open loop” OR “Closed-loop” OR “Feedback sensory” OR “Feedback mechanisms” OR “Closed-loop systems” OR “Closed loop”).
Articles in which the subjects of study were animal models were excluded. Also, articles using invasive electrodes and those that focused on upper limbs were excluded as they were not of interest for this SLR. SLRs and conferences older than 3 years were also excluded. Additionally, book chapters, abstracts and posters, and any articles written in languages different to English were excluded.

2.4. Quality Assessment

To analyze the information, all the 60 included articles were reviewed, to extract from them relevant data based on the questions asked and other information considered important, such as the test subjects, the type of target movement, the purpose of the study, and the type of application (rehabilitation or assistance). It is crucial to note that some articles did not report or focus on some of the parameters mentioned above.

2.5. Data Extraction and Analysis

To analyze the information, all the 60 included articles were reviewed, to extract from them relevant data based on the questions asked and other information considered important. It is crucial to note that some articles do not report or focus on some of the parameters mentioned above. Experimentally informed studies and the data acquisition reported in each article were analyzed. This analysis aimed to evaluate the type of movement, the sensor used, and the variable being measured. It also sought to establish the health condition of the participants in each study. With regard to the sensing strategy, the equipment (sensors) utilized for data acquisition, as well as the specific parameters that were measured during this acquisition, was considered. Subsequently, the control strategy pursued by each article was examined. It should be noted that some articles did not report on the implementation of a control strategy, while others did not provide sufficient detail regarding the strategy’s implementation. It is also important to mention that, in some cases, the control strategy was not applied to the users from whom the data were obtained but was instead implemented in a simulation. Also, due to its relevance to the FES control strategy, an additional analysis was conducted to examine whether or not and how the electrical parameters were modulated.

3. Results

The 60 papers remaining after the filtering stages, out of the original 322, were included as relevant to this SLR and then selected for data extraction and further analysis.
Table 1 shows the 60 articles considered relevant for this SLR.
It is important to mention that in 45% (27) of the articles, tests were conducted with healthy subjects. In total, 35% (21) performed tests with patients, while 16.7% (10) conducted tests with both patients and healthy subjects. Also, it was observed that 31.7% (19) of the tests were conducted with subjects seated, another 38.3% (23) were performed while subjects were walking, and 11.7% (7) performed both types of tests. For 78.3% (47) of the analyzed articles, the goal of the work was rehabilitation, while 11.3% stated that the purpose was assistance. The remaining works considered both rehabilitation and assistance as objectives.
It is worth highlighting that the target population in most articles was stroke patients (19). Other pathologies found in the included articles were spinal cord injury (SCI) (15), neurologically impaired individuals (8), and CNS disorders (3). Some studies did not specify the pathology they focused on (13). The remaining studies focused on pathologies that were mentioned only once.

3.1. RQ1: What Types of Sensors Are Used to Measure the Body’s State and Its Response to FES?

For the analyzed articles, sensors were classified according to their purpose into four categories: switches, electrodes, vision-based, and IMUs. Additionally, it is important to note that most papers utilized more than one type of sensor.
In total, 45% (27) of the articles used IMUs as sensors to measure body/limb posture. About 36.7% (22) used electrodes, encompassing terms such as “EMG sensors”, “sEMG”, “Electrode pads”, and “EEG sensors”. Additionally, the category of vision-based systems was included, which comprised all systems using images generated by any hardware. This category was present in 18.3% (11) of the articles. The systems included in this category are as follows: motion capture systems, 11.7% (7); US imaging, 5% (3); and depth camera, 1.7% (1). Meanwhile, 18.3% (11) were categorized under switches, including terms like “FSRs”, “forefoot switches”, “heel switches”, “force transducers”, “Foot pressure insoles”, and “Foot Plantar Pressure Sensors”. Goniometers were present in 6.7% (4), while dynamometers also were present in 5% (3) of the articles.
As shown in Table 1, there were other types of sensors used for these measurements. However, it is relevant to mention those that are more commonly used. Other sensors were analyzed in the category of others, 28.3% (17).
The results yielded a set of key criteria (Figure 2) useful for selecting sensor strategies in FES systems targeting FD. It is important to note that the inclusion of more than one of these criteria is essential for consideration in an FES system. However, articles were divided according to the main criterion they considered in order to categorize them. It is also important to mention that this categorization is based solely on the sensing method. Those articles that did not focus on the criteria for selecting the sensor were not included in this categorization. The proposed criteria are described in detail below and illustrated in Figure 2.
  • Accuracy
Twenty-three of the articles included in this SLR emphasize the necessity of employing a sensor or sensing strategy that allows for precise measurement. The design of systems for gait rehabilitation is a complex process that requires precise and robust control systems that incorporate sophisticated actuation and sensing capabilities. Precise identification of gait phases and a responsive control strategy are vital for effective assistance and rehabilitation of walking using FES, since they directly affect the precise timing of stimulation. To achieve suitable controllers that deliver stimulation that mirrors normal muscle activation patterns, reliable online detection of gait phases has been shown to be essential [17,33,43]. It is of great importance to acknowledge that the accuracy of a FES system is reliant on both the selection of the sensor and the processing and control algorithm technique employed. However, the focus of this section is the examination of the precision of the sensors utilized.
In [38], the authors address that a FD stimulator should be able to differentiate between walking and other type of activities (exercises) to avoid unnecessary stimulations in non-walking conditions. This allows patients to safely warm up with the device donned and eliminates the need to remove it after warming up.
Another example where accuracy in some aspect of FES systems has been shown to be a factor in the performance of FES systems for gait assistance and rehabilitation is the work reported in [17]. The researchers created a real-time gait phase detection system for an FES system, with optimized sensors and data processing. However, they found that the use of wearable IMU sensors caused a delay in the detection of gait phases, and led to stimulation timing errors, which may have increased the time shift between the intended and actual actions assisted by FES by up to 100%. The origin of those delays can come from wireless IMU data streaming, USB communication protocol timing uncertainties, or even the non-real-time nature of the Windows operating system.
In order to improve accuracy in FES systems, multiple studies, 46.7% (28), included in the SLR used two or more sensors. Previous research [74] indicates that a single foot switch, due to its low detection reliability, is not suitable for triggering FES stimulation sequences for walking assistance/rehabilitation. Therefore, the researchers developed a system comprising a novel sensor combination: three FSRs to measure the load force in a shoe insole and a miniature gyroscope chip to measure the foot’s rotational velocity. This way, the authors achieved a novel gait phase detection sensor that accurately identifies transitions between the stance, heel off, swing, and heel strike phases of walking.
In another work [21], the authors reported that using only motion sensors in FES systems poses challenges to their performance, due to their low signal-to-noise ratio and the need for extensive signal post-processing. Therefore, they report on the combination of FSRs and motion sensors. By placing FSRs under the foot and attaching accelerometers to the shank, feedback signals can be gathered to generate FES sequences for four muscles, guided by a controller with rules derived from human data.
  • Wearability
The authors of 18 of the articles presented in this SLR were motivated to implement wearable sensors, although these studies also sought to achieve high accuracy. The primary considerations influencing the selection of sensors were the practicality of incorporating them into daily activities or settings outside the laboratory, and the ability to allow patients to walk freely. Sensors in FES systems are often employed to measure the state of the body through variables such as muscle force, joint angle, velocity, and acceleration. Although some variables are measured with specific sensors, sometimes they are not practical for the operation conditions of the system. Some projects involving FES for FD are suitable to be used solely in the laboratory, while others are operated outside of it. For ex-ample, since torque is considered an important variable related to bodily response [68] in FES systems, the use of conventional, mechanical torque sensors is desired. However, torque sensors that are available commercially are unsuitable to be used in patients’ daily lives. Similar reasons have motivated researchers to employ different types of sensors to indirectly measure some variables; for example, a number of methods have been proposed to estimate muscle force or joint torque from sEMG signals [46,75].
Another situation related to wearability is when multiple stimulation electrodes and sensors are placed on the body, which is complicated when using traditional wired connections. Wireless sensing and stimulation systems offer a more streamlined and mobile solution to their wired alternatives. However, only a few external wireless FES stimulators currently attempt to connect and coordinate individual channels via a network. In this regard, in [66], the authors presented a wireless distributed FES architecture. The system relies on the collection of potentially diverse distributed stimulation and measurement units overseen by a wearable controller.
  • Affordability
While less predominant, five of the articles that addressed the motivation behind the selection of sensors were driven by the desire to implement sensors that are cost-effective. In some instances, this objective coincides with that of the incorporation of wearable sensors.
The rise in the use of open-source and low-cost prototyping electronics has enabled the popularization of platforms like the Arduino microcontroller unit to be employed in the development of a wide variety of electronic systems. These platforms utilize high-level libraries that allow for the quick and easy integration of various sensors and actuators into a testing platform, making them ideal for developing low-cost, portable gait neuroprostheses. In [22], a flexible, low-cost micro-controller-based platform for the rapid prototyping of FES-based neuroprostheses is presented. The system was designed to reduce computational complexity, the time for development, and production costs. Its small size and weight (225 g with all modules connected) make it perfect to be attached to a below-the-knee cuff or at the hip.
The study conducted by [73] sought to explore the practicality of using a more cost-effective and accessible sensor, such as a depth camera or pressure mat, as an alternative to the force plate.

3.2. RQ2: What Variables Does the Sensor Measure in Gait-Oriented FES Systems?

Due to the dispersion of the variables, we decided to categorize them for the analysis. The category Angles, which includes all variables related to angles of body segments, such as knee joint, ankle joint, foot, etc., was present in 38.3% (23) of the articles.
Another category included was electrical activity, which was present in 18.3% (11) of the articles included in this review.
The next most used variable was angular velocity, which appeared in 13.3% (8) of the articles. Following this, acceleration was present in 11.7% (7) of the works. Another category that was found to be important to group was Torque, which encompassed all methods related to torque measurements from different parts of the body; this category appeared in 10% (6) of the articles. Finally, the category gait phases, which includes all variables related to identifying gait features, such as heel strike and toe off, gait events, etc, was present in 10% (6) of the articles. The remaining variables were reported only in one or two works and can be seen in Table 1. Some of them are commented below.
In [30], the researchers carried out preliminary work on a sensor-based multichannel FES system for post-stroke gait where FES was used to control knee joints. Their research highlights that the application of FES to a stiff knee in hemiplegic individuals reduces spasticity in both the knee flexors and extensors, while enhancing their range of motion. Additionally, they highlight that avoiding hyperextension and allowing slight knee flexion in the affected limb during the stance phase significantly aid in gait recovery. Therefore, accurate control of knee angle is vital for improving gait in people with CNS disorders.
Due to FD, patients often experience ankle dyskinesia. In [45], the human ankle dorsiflexion angle is considered the research object in the authors’ FES system. An accurate ankle model helps explore movement characteristics under different electrical stimulation parameters. The motion characteristics of the ankle angle, including hysteresis, time variance, and nonlinearity, can be summarized based on experiments that measure ankle angle changes induced by electrical stimulation.
In [58], the researchers predict torque output based on evoked electromyography (eEMG) recordings only. Volitional EMG-based joint movement estimation is effective in healthy subjects; however, SCI patients cannot generate volitional EMG due to spinal cord damage. Consequently, performance prediction is more stable in healthy subjects than in SCI patients. On the other hand, evoked signals in SCI patients are reliable due to the absence of volitional contractions, since their weaker muscles result in less progressive and stable recruitment. Nevertheless, the extent of spinal lesions varies widely among SCI patients, requiring more patient-specific adjustments compared to healthy individuals.

3.3. RQ3: Which Control Strategies Are Employed in Lower Limb FES Systems and Applications?

In total, 63.3% (38) of the articles employed a CL strategy, 8.3% (5) employed an OL strategy, and the remaining were categorized as Others. Some articles reported the strategy used to implement the CL, with the most common being finite state machine, 8.3% (5); iterative learning control, 5% (3); and proportional–integral derivative (PID), 1.7% (1); and P controller, 3.3% (2).
According to some works in this SLR, adaptive CL control is ideal for FES systems but is rarely developed [17,43]. Most commercial FES-based gait rehabilitation systems use simpler OL control solutions. Parastep I (Sigmedics, Inc., Fairborn, OH, USA) and RehaStim (Hasomed Inc., Magdeburg, Germany) are two commercially available FES systems that use OL control strategies. These simple solutions demand continuous or repeated user input, usually via a device button, for muscle activation, thus requiring the complete focus of the user, clinical personnel, and/or caregiver [17,43].
Finite-state-controlled FES systems combine the accuracy of CL controllers with the simplicity of OL setups, using minimal sensors, generally single or a few foot switches, or an IMU. In general, they operate with preset stimulation sequences triggered by specific conditions. However, there are works that report offline adjustments between the period of use as a strategy to combat muscle fatigue [17].
Additionally, six articles were found that reported on the use of machine learning techniques. Among them, neural networks were the most used technique, with a focus on model accuracy and personalization to improve clinical outcomes. As examples, three studies are described below.
In [45], ankle angle characteristics are used to train a Hammerstein (H) model based on a neural network, with model parameters identified using a genetic algorithm. This approach effectively predicts ankle angle changes induced by electrical stimulation. Experimental results confirm that the neural-network-based H model can accurately forecast the output changes in the ankle angle due to the electrical stimulation pulse.
Another study [58] introduces a real-time system for estimating FES-induced torque, utilizing a wireless portable stimulator. The system uses a Kalman filter and recurrent neural network (RNN) to predict torque output from eEMG recordings. Experiments with able-bodied subjects and SCI patients demonstrate its promising performance. This system offers personalized muscle response evaluation, beneficial for clinical diagnostics.
Another paper [60] introduces a new machine learning approach to improve lower limb tracking control for individuals with SCI through neuromuscular electrical stimulation/FES. The method uses data-driven models with historical rehabilitation data, applies robust integral of the sign of the error (RISE) as a control technique for stability, and employs an enhanced genetic algorithm for effective controller tuning.
Reinforcement learning (RL) is a machine learning technique that relies on rewards and involves an agent who observes the environment and learns an optimal policy for action selection based on the states [26]. In [26], the researchers introduce an RL algorithm utilizing a decayed epsilon greedy approach to investigate different pulse parameter variations, with the goal of optimizing stimulation patterns during FES cycling sessions.

3.4. RQ4: What Stimulation Parameters Are Modulated in CL Lower Limb FES Applications?

In total, 60% (36) of the articles specify the use of adaptive FES systems. Some of them specify which parameters are modulated, with the most frequently modulated ones being pulse width, 26.7% (16); intensity, 8.3% (5); timing, 8.3% (5); frequency, 3.3% (2); and amplitude, 8.3% (5). Below, three examples of parameter modulation strategies are described.
In [45], an experimental platform was used to study the effects of different electrical stimulation parameters on ankle motion characteristics. The parameters investigated were frequency, amplitude, and pulse width. The findings highlight that lower frequencies are preferred to avoid muscle fatigue, with 25 Hz being optimal. Increased amplitude generally led to a greater ankle angle, with 25 mA being the ideal setting. Adjustments in pulse width affected the ankle angle, with discomfort noted at higher widths. The results provided valuable data for neural network-based model input parameters and demonstrated that electrical stimulation parameters significantly influence ankle joint motion.
In 2016, the authors of [27] proposed a control method for an FES system aiding in walking. It adjusts the quadricep stimulation amplitude based on knee flexion angles increasing it by 10 mA for angles over 10° (knee unlock), maintaining amplitude for 5–10° (knee extension), and gradually reducing it by 5 mA over 2 s for angles below 5° (knee lock). After each increment, a 1 s interval follows before potentially adjusting it again, with flexibility if knee buckling exceeds 10° per second. This approach aims to optimize muscle stimulation for enhanced stability and control during the stance phase of walking, adapting in real-time to knee angle feedback.
In the work reported in [33], the aim was to optimize FES assistance by dynamically adjusting stimulation parameters according to real-time feedback, thereby improving the effectiveness of gait assistance across different walking speeds and conditions. The parameters (τ and PW) were updated dynamically based on data from the previous five gait cycles, ensuring that the assistance provided by the FES system aligned with the current walking conditions and speed.

4. Discussion

4.1. Sensing

Feedback signals based on artificial or natural sensors are required to build CL control systems for FES-based gait assistance.
Although sensors are employed to measure the state of the body, response variables such as muscle force, joint angle, velocity, and acceleration are not the only factors to con-sider when choosing which sensor or sensors to use in the FES system. Even though certain variables may seem more suitable to measure with specific sensors, this is not the sole consideration.
A key consideration is the purpose for which the FES system will be implemented. Some projects are intended to be used solely in the laboratory, while others aim to be utilized outside of it. For example, the implementation of torque sensors is often pursued because torque is considered an important bodily response. However, existing torque sensors are unsuitable to be used in patients’ daily lives. Reasons like this motivate the search for non-evident or conventional sensors to indirectly measure some variables. Other examples include reports of methods that have been proposed to estimate muscle force or joint torque from sEMG signals [46,75].
Another consideration is the choice of the control system. Choosing a sensor digitized with a low sampling rate can be challenging to incorporate into CL FES control. For instance, the sampling frequency of the echogenicity signal derived from US imaging is significantly lower than kinematic measurements obtained from IMUs or angular encoders [15]. This serves as an example that some reports of the implementation of FES systems employing novel sensing strategies are not always aimed at demonstrating improvements over an existing method but rather at showcasing the potential advantages one sensor has over others.
Moreover, the performance assessment of FES systems is a crucial factor for their design. Previous research has concentrated on the torque or angle between intended and actual limb movements. For instance, Sharma et al. explored the nonlinear FES tracking control of human limbs and assessed the system used by monitoring the knee’s trajectory to control the angular deviation from the ideal path. Zhang et al. (2013) employed sEMG-based CL torque control in the FES system and analyzed the discrepancy between the generated and desired torque [32].
Furthermore, it is important to mention that some sensors are more complex and expensive than others; this is another important thing to consider. For example, the use of motion capture data is another common sensing strategy utilized for feedback and the assessment of FES systems’ performance; however, the equipment used for capturing gait using video (photogrammetry) is costly and complicated. A video-based motion capture system typically involves placing reflective markers on the subject’s skin or using specialized clothing to record human movement using cameras and specialized software. Additionally, the captured video image data can be easily influenced by factors such as lighting and range of movement [73].
According to the articles reviewed, IMUs and biopotential electrodes are the most commonly used sensors in research. It is also important to mention that most of the analyzed articles use more than one sensor.
Taking all of this into account, to aid in the development of a more robust FES system, it is important to explore new alternatives that have high measurement precision, are portable, can be integrated into control systems, and are not prohibitively expensive.
To our knowledge, and based on the results of this SLR, few articles have used depth cameras as a method of measuring the body’s response to FES. In fact, only a few articles were found in this review to employ depth cameras, and these were mostly focused on balance training or upper limb applications [73,76].
Using depth cameras as feedback sensors, cheaper, and more portable FES systems can be implemented. In 2024 [73], a depth camera showed higher potential as a low-cost, portable sensor, compared to a pressure mat in an FES balance application. Also, a depth camera is a feasible option as a replacement for the force plate for use in a FES + VFBT (virtual feedback balance training) system.
Furthermore, no works were found that used binocular vision, which can lead to similar results to those of using depth cameras, at a lower cost. This is because infrared sensors used in depth cameras are more expensive than RGB cameras. The above discussion opens up an area of opportunity to explore new techniques and improve existing ones.
Measuring the body’s state and its reactions can provide valuable information about the system’s effectiveness, with each type of measurement offering different insights. For example, FES-restored motion relies on active joint torques and environmental interactions. Joint angle feedback cannot distinguish between motion from stimulation and external forces, making joint torque control superior. Torque control provides essential compliance, especially for environmental interactions and daily human activities [46].
Focusing on a specific joint can be crucial for post-stroke gait recovery. Several studies have indicated that preventing hyperextension and allowing slight knee flexion of the paretic limb during the stance phase can significantly enhance gait recovery. This is the reason why many studies focus on measuring the knee joint [30].

4.2. Control Strategies

When it comes to muscle stimulation, most commercial FES systems operate in OL mode, where the stimulation intensity—comprising pulse width and pulse amplitude—remains constant. The primary drawback of OL FES is its inability to adjust stimulation intensity in response to changes in residual muscle activity and spasticity, which can happen within as few as five strides [34].
In contrast, a CL FES control strategy is considered a better solution for gait rehabilitation applications, as it more accurately replicates the nervous system’s control of gait. This approach uses feedback information to modify stimulation parameters according to the desired joint angle or moment trajectories [43]. The difficulties in CL control stem from the complex and nonlinear nature of electrical muscle stimulation, involving unpredictable mapping from electrical input to muscle force, muscle fatigue, and delayed muscle responses [39]. To address these challenges, various control strategies have been sought that are robust against unpredictable changes in input delay and the uncertain dynamics of muscles.
The key muscles involved in gait include the hamstrings (biceps femoris), quadriceps (vastus medialis), calf muscles (soleus), and tibialis anterior; each have unique functions during the stance and swing phases of the gait cycle, and they are activated in a specific order [77]. In FES-based rehabilitative gait systems, the activity of these muscles must transition in the same sequence. A finite-state machine (FSM) can be employed to achieve this, as it allows for the sequential stimulation of muscles. An FSM model, which sequences ON/OFF actions, can control the timing of muscle stimulations within the different states of the gait cycle [43].
Another strategy used is iterative learning control (ILC), which is suitable for systems performing repeated tasks [4]. ILC aims to minimize trajectory tracking errors through repeated system operation. It continuously adjusts the controlled system’s trajectory towards the desired path, characterized by a strong mathematical foundation and flexibility for systems with uncertain parameters, which are particularly beneficial for nonlinear systems [52].
The functional effectiveness of an FES system for gait heavily depends on the precise timing of the applied stimulation within the gait cycle. The most straightforward way to control this timing is through manual press buttons or foot switches, which constitute a common approach in most commercial products [21]. The success of FES in treating FD patients relies on the control and modulation of stimulation intensity. Two popular modulation methods for FD intervention are the trapezoidal stimulation profile and the EMG-modulated natural stimulation profile [32].
Although the use of machine learning-based control algorithms was reported in some works reviewed in this SLR, deep learning algorithms were not found, although these techniques have been applied in other technologies used for motor rehabilitation that require interaction with the user, like brain–computer interfaces [78]. In the future, it would be interesting to attempt to incorporate such algorithms to improve sensing/control strategies, for example, to create intelligent–adaptive FES systems, that require no or little previous calibration.
The integration of these kinds of intelligent algorithms, trained with large databases comprising multimodal variables from conventional and novel sensors, validated by gold-standard techniques, such as photogrammetry, would contribute greatly to improving the design and performance of FES systems. To this end, multi-center collaboration would be ideal in overcoming the limitations of the sample sizes of most works, especially when treating patients with FD caused by low-incidence medical conditions, like spinal cord injury or cerebral palsy.
One of the main limitations identified in the studies reviewed is that they tested their systems only with healthy volunteers or with a small sample of patients. Hence, larger pilot and controlled clinical trials are required to better assess the benefits and feasibility of novel sensing and control approaches. Another key limitation was that many studies did not explicitly report the type of sensors they employed, which reduced the information available.
To summarize, some key considerations for practical applications derived from this SLR are as follows:
  • Applying a multimodal sensing strategy, comprising more than one sensor, is a highly recommended approach in CL gait-oriented FES systems for FD, with the potential to improve system accuracy and performance.
  • Novel sensing and control strategies, such as US sensors and machine/deep learning-based controllers, should be considered, possibly combined with the multimodal approach mentioned above. To achieve this, fast and reliable digitizing and processing systems are required, to be feasible in real-life applications.
  • The integration of markerless motion analysis tools, employing computer vision and depth sensors, is an alternative approach worthwhile to explore in gait-oriented FES applications, the relevance of which has already been shown in upper limb and stability training applications.

5. Conclusions

This SLR will serve as a global framework of sensing and control strategies employed in FES systems for gait rehabilitation for patients with FD. This topic is of great relevance to researchers and developers, since both elements play a fundamental role in the overall performance of FES systems and applications; knowing beforehand the different existing techniques already implemented and tested by other research groups could be vital at the initial stages of the design process of an effective FES system, according to specific needs and purposes.
Taking into account the criteria proposed for the selection of sensors for FES systems, one conclusion is that the use of depth cameras as a measurement method is feasible, these being non-invasive, portable, and low-cost sensors. Future research will focus on comparing the accuracy of depth cameras and other vision-based methods with other sensors mentioned in this SLR.
In particular, the variable angles was of utmost interest in the articles analyzed, due to its direct relation with the target ankle movement to be induced by the FES system. Furthermore, it is worth noting that measuring the angle variable through depth cameras is highly feasible, in static or dynamic conditions, through signal processing algorithms.
For future research, it is recommended to further explore the implementation of control methods based on machine learning algorithms, especially in combination with multimodal sensing techniques if possible, including novel approaches in the field, such as US or depth cameras/computer vision. Such combinations could allow the development of better, more natural interfaces for FES systems, especially if they are coupled with high-performance computing systems, like Field programmable gate array circuits, real-time microprocessor devices, or even emerging artificial intelligence dedicated processing units.

Author Contributions

Conceptualization, E.G.-G., J.A.M.-G. and S.M.-D.; methodology, E.G.-G., J.A.M.-G. and S.M.-D.; validation, I.C.-L. and I.M.S.-M.; formal analysis, G.R.-R., I.Q.-U. and O.Y.-S.; investigation, E.G.-G., S.M.-D., J.A.M.-G. and G.R.-R.; resources, S.M.-D., J.A.M.-G. and G.R.-R.; writing—original draft preparation, E.G.-G.; writing—review and editing, E.G.-G., J.A.M.-G., S.M.-D., I.C.-L., I.M.S.-M., G.R.-R., I.Q.-U. and O.Y.-S.; supervision, I.C.-L., I.M.S.-M., G.R.-R., I.Q.-U. and O.Y.-S. All authors have read and agreed to the published version of the manuscript.

Funding

This research was supported by Secretaría de Educación, Ciencia, Tecnologia e Innovacion (SECTEI), through grants SECTEI/182/2021 and SECTEI/181/2021. The APC was funded by Tecnológico Nacional de México (TecNM).

Data Availability Statement

No new data were created or analyzed in this study. Data sharing is not applicable to this article.

Acknowledgments

We acknowledge TecNM and CONAHCyT for supporting the M.Sc. studies of the first author.

Conflicts of Interest

The authors declare no conflicts of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript; or in the decision to publish the results.

References

  1. Pilkar, R.; Ramanujam, A.; Nolan, K.J. Alterations in Spectral Attributes of Surface Electromyograms after Utilization of a Foot Drop Stimulator during Post-Stroke Gait. Front. Neurol. 2017, 8, 449. [Google Scholar] [CrossRef] [PubMed]
  2. Peishun, C.; Haiwang, Z.; Taotao, L.; Hongli, G.; Yu, M.; Wanrong, Z. Changes in Gait Characteristics of Stroke Patients with Foot Drop after the Combination Treatment of Foot Drop Stimulator and Moving Treadmill Training. Neural Plast. 2021, 2021, 1–5. [Google Scholar] [CrossRef]
  3. Alouane, M.A.; Rifai, H.; Kim, K.; Amirat, Y.; Mohammed, S. Hybrid impedance control of a knee joint orthosis. Ind. Robot Int. J. Robot. Res. Appl. 2019, 46, 192–201. [Google Scholar] [CrossRef]
  4. Page, A.P.; Freeman, C.T. Point-to-point repetitive control of functional electrical stimulation for drop-foot. Control Eng. Pract. 2020, 96, 104280. [Google Scholar] [CrossRef]
  5. Nori, S.L.; Stretanski, M.F. Foot Drop. In StatPearls; StatPearls Publishing: Treasure Island, FL, USA, 2024. Available online: http://www.ncbi.nlm.nih.gov/books/NBK554393/ (accessed on 28 July 2024).
  6. Nath, R.K.; Somasundaram, C. Incidence, Etiology, Risk Factors Associated with Foot Drop. Eplasty 2023, 23, e16. [Google Scholar]
  7. Carvalho, S.; Correia, A.; Figueiredo, J.; Martins, J.M.; Santos, C.P. Functional Electrical Stimulation System for Drop Foot Correction Using a Dynamic NARX Neural Network. Machines 2021, 9, 253. [Google Scholar] [CrossRef]
  8. Meng, L.; Porr, B.; Gollee, H. Technical developments of functional electrical stimulation to restore gait functions: Sensing, control strategies, and current commercial systems. Chin. J. Sci. Instrum. 2017, 2017, 1319–1334. [Google Scholar]
  9. Khan, M.A.; Fares, H.; Ghayvat, H.; Brunner, I.C.; Puthusserypady, S.; Razavi, B.; Lansberg, M.; Poon, A.; Meador, K.J. A systematic review on functional electrical stimulation based rehabilitation systems for upper limb post-stroke recovery. Front. Neurol. 2023, 14, 1272992. [Google Scholar] [CrossRef] [PubMed]
  10. Schauer, T. Sensing motion and muscle activity for feedback control of functional electrical stimulation: Ten years of experience in Berlin. Annu. Rev. Control 2017, 44, 355–374. [Google Scholar] [CrossRef]
  11. Salchow-Hömmen, C.; Skrobot, M.; Jochner, M.C.E.; Schauer, T.; Kühn, A.A.; Wenger, N. Review-Emerging Portable Technologies for Gait Analysis in Neurological Disorders. Front. Hum. Neurosci. 2022, 16, 768575. [Google Scholar] [CrossRef]
  12. Seel, T.; Raisch, J.; Schauer, T. IMU-Based Joint Angle Measurement for Gait Analysis. Sensors 2014, 14, 6891–6909. [Google Scholar] [CrossRef]
  13. Strbac, M.; Malesevic, N.; Cobeljic, R.; Schwirtlich, L. Feedback control of the forearm movement of tetraplegic patient based on microsoft kinect and multi-pad electrodes. J. Autom. Control 2013, 21, 7–11. [Google Scholar] [CrossRef]
  14. Strbac, M.; Kočović, S.; Marković, M.; Popović, D.B. Microsoft kinect-based artificial perception system for control of functional electrical stimulation assisted grasping. BioMed Res. Int. 2014, 2014, 740469. [Google Scholar] [CrossRef] [PubMed]
  15. Zhang, Q.; Lambeth, K.; Iyer, A.; Sun, Z.; Sharma, N. Ultrasound Imaging-Based Closed-Loop Control of Functional Electrical Stimulation for Drop Foot Correction. IEEE Trans. Control Syst. Technol. 2023, 31, 989–1005. [Google Scholar] [CrossRef]
  16. Gil-Castillo, J.; Alnajjar, F.; Koutsou, A.; Torricelli, D.; Moreno, J.C. Advances in neuroprosthetic management of foot drop: A review. J. Neuroeng. Rehabil. 2020, 17, 46. [Google Scholar] [CrossRef]
  17. Zahradka, N.; Behboodi, A.; Wright, H.; Bodt, B.; Lee, S. Evaluation of Gait Phase Detection Delay Compensation Strategies to Control a Gyroscope-Controlled Functional Electrical Stimulation System During Walking. Sensors 2019, 19, 2471. [Google Scholar] [CrossRef]
  18. Chaikho, L.; Clark, E.; Raison, M. Transcutaneous Functional Electrical Stimulation Controlled by a System of Sensors for the Lower Limbs: A Systematic Review. Sensors 2022, 22, 9812. [Google Scholar] [CrossRef] [PubMed]
  19. Page, M.J.; McKenzie, J.E.; Bossuyt, P.M.; Boutron, I.; Hoffmann, T.C.; Mulrow, C.D.; Shamseer, L.; Tetzlaff, J.M.; Akl, E.A.; Brennan, S.E.; et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ 2021, 372, n71. [Google Scholar] [CrossRef]
  20. Xu, R.; Jiang, N.; Mrachacz-Kersting, N.; Lin, C.; Prieto, G.A.; Moreno, J.C.; Pons, J.L.; Dremstrup, K.; Farina, D. A closed-loop brain-computer interface triggering an active ankle-foot orthosis for inducing cortical neural plasticity. IEEE Trans. Biomed. Eng. 2014, 61, 2092–2101. [Google Scholar] [CrossRef]
  21. Meng, L.; Porr, B.; Macleod, C.A.; Gollee, H. A functional electrical stimulation system for human walking inspired by reflexive control principles. Proc. Inst. Mech. Eng. H 2017, 231, 315–325. [Google Scholar] [CrossRef]
  22. de Melo, P.L.; da Silva, M.T.; Martins, J.; Newman, D. A microcontroller platform for the rapid prototyping of functional electrical stimulation-based gait neuroprostheses. Artif. Organs 2015, 39, E56–E66. [Google Scholar] [CrossRef]
  23. Alibeji, N.; Kirsch, N.; Dicianno, B.E.; Sharma, N. A Modified Dynamic Surface Controller for Delayed Neuromuscular Electrical Stimulation. IEEE ASME Trans. Mechatron. 2017, 22, 1755–1764. [Google Scholar] [CrossRef] [PubMed]
  24. Ahmed, M.; Huq, M.S.; Ibrahim, B.S.K.K. A New Method towards Achieving FES-Induced Movement. Int. J. Integr. Eng. 2018, 10, 30–34. [Google Scholar] [CrossRef]
  25. Le Guillou, R.; Schmoll, M.; Sijobert, B.; Borges, D.L.; Fachin-Martins, E.; Resende, H.; Pissard-Gibollet, R.; Fattal, C.; Coste, C.A. A Novel Framework for Quantifying Accuracy and Precision of Event Detection Algorithms in FES-Cycling. Sensors 2021, 21, 4571. [Google Scholar] [CrossRef]
  26. Coelho-Magalhães, T.; Coste, C.A.; Resende-Martins, H. A Novel Functional Electrical Stimulation-Induced Cycling Controller Using Reinforcement Learning to Optimize Online Muscle Activation Pattern. Sensors 2022, 22, 9126. [Google Scholar] [CrossRef]
  27. Braz, G.P.; Russold, M.F.; Fornusek, C.; Hamzaid, N.A.; Smith, R.M.; Davis, G.M. A novel motion sensor-driven control system for FES-assisted walking after spinal cord injury: A pilot study. Med. Eng. Phys. 2016, 38, 1223–1231. [Google Scholar] [CrossRef] [PubMed]
  28. Perumal, R.; Wexler, A.S.; Kesar, T.M.; Jancosko, A.; Laufer, Y.; Binder-Macleod, S.A. A phenomenological model that predicts forces generated when electrical stimulation is superimposed on submaximal volitional contractions. J. Appl. Physiol. 2010, 108, 1595–1604. [Google Scholar] [CrossRef]
  29. Hobbs, B.; Artemiadis, P. A Review of Robot-Assisted Lower-Limb Stroke Therapy: Unexplored Paths and Future Directions in Gait Rehabilitation. Front. Neurorobot. 2020, 14, 19. [Google Scholar] [CrossRef]
  30. Sijobert, B.; Azevedo, C.; Pontier, J.; Graf, S.; Fattal, C. A Sensor-Based Multichannel FES System to Control Knee Joint and Reduce Stance Phase Asymmetry in Post-Stroke Gait. Sensors 2021, 21, 2134. [Google Scholar] [CrossRef]
  31. Li, Y.; Yang, Q.; Fang, P.; Song, R. Adaptive Stimulation Profiles Adjustment of Functional Electrical Stimulation for Foot Drop Based on Iterative Learning Control. In Proceedings of the 2023 International Conference on Advanced Robotics and Mechatronics (ICARM), Sanya, China, 8–10 July 2023; IEEE: Piscataway, NJ, USA, 2023; pp. 895–899. [Google Scholar] [CrossRef]
  32. Li, Y.; Yang, X.; Zhou, Y.; Chen, J.; Du, M.; Yang, Y. Adaptive Stimulation Profiles Modulation for Foot Drop Correction Using Functional Electrical Stimulation: A Proof of Concept Study. IEEE J. Biomed. Health Inform. 2021, 25, 59–68. [Google Scholar] [CrossRef]
  33. Dong, H.; Hou, J.; Song, Z.; Xu, R.; Meng, L.; Ming, D. An adaptive reflexive control strategy for walking assistance system based on functional electrical stimulation. Front. Neurosci. 2022, 16, 944291. [Google Scholar] [CrossRef]
  34. Anaya-Reyes, F.; Narayan, A.; Aguirre-Ollinger, G.; Cheng, H.-J.; Yu, H. An Omnidirectional Assistive Platform Integrated With Functional Electrical Stimulation for Gait Rehabilitation: A Case Study. IEEE Trans. Neural Syst. Rehabil. Eng. 2020, 28, 710–719. [Google Scholar] [CrossRef]
  35. Zhang, Q.; Meng, Y.; Wu, L.; Xiang, X.; Xiong, C. Artificially induced joint movement control with musculoskeletal model-integrated iterative learning algorithm. Biomed. Signal Process. Control 2020, 59, 101843. [Google Scholar] [CrossRef]
  36. Hosiasson, M.; Rigotti-Thompson, M.; Appelgren-Gonzalez, J.; Covarrubias-Escudero, F.; Urzua, B.; Barría, P.; Aguilar, R. Biomechanical Gait Effects of a Single Intervention with Wearable Closed Loop Control FES System in Chronic Stroke Patients. A Proof-of-Concept Pilot Study. In Proceedings of the 2023 International Conference on Rehabilitation Robotics (ICORR), Singapore, 24–28 September 2023; IEEE: Piscataway, NJ, USA, 2023; pp. 1–6. [Google Scholar] [CrossRef]
  37. Yeom, H.; Chang, Y.-H. Autogenic EMG-controlled functional electrical stimulation for ankle dorsiflexion control. J. Neurosci. Methods 2010, 193, 118–125. [Google Scholar] [CrossRef] [PubMed]
  38. Shaikh, M.F.; Salcic, Z.; Wang, K.I.-K.; Hu, A.P. Bipedal gait model for precise gait recognition and optimal triggering in foot drop stimulator: A proof of concept. Med. Biol. Eng. Comput. 2018, 56, 1731–1746. [Google Scholar] [CrossRef] [PubMed]
  39. Obuz, S.; Duenas, V.H.; Downey, R.J.; Klotz, J.R.; Dixon, W.E. Closed-Loop Neuromuscular Electrical Stimulation Method Provides Robustness to Unknown Time-Varying Input Delay in Muscle Dynamics. IEEE Trans. Control Syst. Technol. 2020, 28, 2482–2489. [Google Scholar] [CrossRef]
  40. Xu, S.; Li, C.; Wei, C.; Kang, X.; Shu, S.; Liu, G.; Xu, Z.; Han, M.; Luo, J.; Tang, W. Closed-Loop Wearable Device Network of Intrinsically-Controlled, Bilateral Coordinated Functional Electrical Stimulation for Stroke. Adv. Sci. 2024, 11, 2304763. [Google Scholar] [CrossRef]
  41. Singh, M.; Sharma, N. Data-driven Model Predictive Control for Drop Foot Correction. In Proceedings of the 2023 American Control Conference (ACC), San Diego, CA, USA, 31 May–2 June 2023; IEEE: Piscataway, NJ, USA, 2023; pp. 2615–2620. [Google Scholar] [CrossRef]
  42. Watanabe, T.; Tadano, T. Design of Closed-Loop Fuzzy FES Controller and Tests in Controlling Knee Extension Movements. IEICE Trans. Inf. Syst. 2017, E100.D, 2261–2264. [Google Scholar] [CrossRef]
  43. Hayami, N.; Williams, H.E.; Shibagaki, K.; Vette, A.H.; Suzuki, Y.; Nakazawa, K.; Nomura, T.; Milosevic, M. Development and Validation of a Closed-Loop Functional Electrical Stimulation-Based Controller for Gait Rehabilitation Using a Finite State Machine Model. IEEE Trans. Neural Syst. Rehabil. Eng. 2022, 30, 1642–1651. [Google Scholar] [CrossRef]
  44. Qian, J.; Rong, K.; Qian, Z.; Wen, C.; Zhang, S. Effects of a multichannel dynamic functional electrical stimulation system on hemiplegic gait and muscle forces. J. Phys. Ther. Sci. 2015, 27, 3541–3544. [Google Scholar] [CrossRef]
  45. Zhou, H.Y.; Huang, L.K.; Gao, Y.M.; Vasic, Z.L.; Cifrek, M.; Du, M. Estimating the Ankle Angle Induced by FES via the Neural Network-Based Hammerstein Model. IEEE Access 2019, 7, 141277–141286. [Google Scholar] [CrossRef]
  46. Zhang, Q.; Hayashibe, M.; Azevedo-Coste, C. Evoked Electromyography-Based Closed-Loop Torque Control in Functional Electrical Stimulation. IEEE Trans. Biomed. Eng. 2013, 60, 2299–2307. [Google Scholar] [CrossRef] [PubMed]
  47. Embrey, D.G.; Holtz, S.L.; Alon, G.; Brandsma, B.A.; McCoy, S.W. Functional Electrical Stimulation to Dorsiflexors and Plantar Flexors During Gait to Improve Walking in Adults With Chronic Hemiplegia. Arch. Phys. Med. Rehabil. 2010, 91, 687–696. [Google Scholar] [CrossRef] [PubMed]
  48. Chen, W.; Li, J.; Zhu, S.; Zhang, X.; Men, Y.; Wu, H. Gait Recognition for Lower Limb Exoskeletons Based on Interactive Information Fusion. Appl. Bionics Biomech. 2022, 2022, 9933018. [Google Scholar] [CrossRef]
  49. Wiesener, C.; Seel, T.; Spieker, L.; Niedeggen, A.; Schauer, T. Inertial-Sensor-Controlled Functional Electrical Stimulation for Swimming in Paraplegics: Enabling a Novel Hybrid Exercise Modality. IEEE Control Syst. 2020, 40, 117–135. [Google Scholar] [CrossRef]
  50. Merad, M.; Downey, R.J.; Obuz, S.; Dixon, W.E. Isometric Torque Control for Neuromuscular Electrical Stimulation with Time-Varying Input Delay. IEEE Trans. Control Syst. Technol. 2016, 24, 971–978. [Google Scholar] [CrossRef]
  51. Seel, T.; Werner, C.; Raisch, J.; Schauer, T. Iterative learning control of a drop foot neuroprosthesis—Generating physiological foot motion in paretic gait by automatic feedback control. Control Eng. Pract. 2016, 48, 87–97. [Google Scholar] [CrossRef]
  52. Guan, W.; Zhou, L.; Cao, Y. Joint Motion Control for Lower Limb Rehabilitation Based on Iterative Learning Control (ILC) Algorithm. Complexity 2021, 2021, 6651495. [Google Scholar] [CrossRef]
  53. Mendes, L.A.; Lima, I.N.; Souza, T.; Nascimento, G.C.D.; Resqueti, V.R.; Fregonezi, G.A. Motor neuroprosthesis for promoting recovery of function after stroke. Cochrane Database Syst. Rev. 2020, 2020, CD012991. [Google Scholar] [CrossRef]
  54. Duenas, V.H.; Cousin, C.A.; Parikh, A.; Freeborn, P.; Fox, E.J.; Dixon, W.E. Motorized and Functional Electrical Stimulation Induced Cycling via Switched Repetitive Learning Control. IEEE Trans. Control Syst. Technol. 2019, 27, 1468–1479. [Google Scholar] [CrossRef]
  55. Beaulieu, L.-D.; Massé-Alarie, H.; Brouwer, B.; Schneider, C. Noninvasive neurostimulation in chronic stroke: A double-blind randomized sham-controlled testing of clinical and corticomotor effects. Top. Stroke Rehabil. 2015, 22, 8–17. [Google Scholar] [CrossRef] [PubMed]
  56. Jochumsen, M.; Niazi, I.K.; Signal, N.; Nedergaard, R.W.; Holt, K.; Haavik, H.; Taylor, D. Pairing Voluntary Movement and Muscle-Located Electrical Stimulation Increases Cortical Excitability. Front. Hum. Neurosci. 2016, 10, 482. [Google Scholar] [CrossRef]
  57. Li, Z.; Guiraud, D.; Andreu, D.; Gelis, A.; Fattal, C.; Hayashibe, M. Real-Time Closed-Loop Functional Electrical Stimulation Control of Muscle Activation with Evoked Electromyography Feedback for Spinal Cord Injured Patients. Int. J. Neural Syst. 2018, 28, 1750063. [Google Scholar] [CrossRef]
  58. Li, Z.; Guiraud, D.; Andreu, D.; Benoussaad, M.; Fattal, C.; Hayashibe, M. Real-time estimation of FES-induced joint torque with evoked EMG: Application to spinal cord injured patients. J. Neuroeng. Rehabil. 2016, 13, 60. [Google Scholar] [CrossRef] [PubMed]
  59. Kojović, J.; Miljković, N.; Janković, M.M.; Popović, D.B. Recovery of motor function after stroke: A polymyography-based analysis. J. Neurosci. Methods 2011, 194, 321–328. [Google Scholar] [CrossRef]
  60. Arcolezi, H.H.; Nunes, W.R.; de Araujo, R.A.; Cerna, S.; Sanches, M.A.; Teixeira, M.C.; de Carvalho, A.A. RISE controller tuning and system identification through machine learning for human lower limb rehabilitation via neuromuscular electrical stimulation. Eng. Appl. Artif. Intell. 2021, 102, 104294. [Google Scholar] [CrossRef]
  61. Zhang, Q.; Iyer, A.; Sun, Z.; Dodson, A.; Sharma, N. Sampled-Data Observer Based Dynamic Surface Control of Delayed Neuromuscular Functional Electrical Stimulation. In Proceedings of the ASME 2020 Dynamic Systems and Control Conference, Virtual, Online, 5–7 October 2020; p. V001T14A003. [Google Scholar] [CrossRef]
  62. Downey, R.J.; Cheng, T.-H.; Bellman, M.J.; Dixon, W.E. Switched Tracking Control of the Lower Limb During Asynchronous Neuromuscular Electrical Stimulation: Theory and Experiments. IEEE Trans. Cybern. 2017, 47, 1251–1262. [Google Scholar] [CrossRef]
  63. Zhang, Q.; Iyer, A.; Lambeth, K.; Kim, K.; Sharma, N. Ultrasound Echogenicity as an Indicator of Muscle Fatigue during Functional Electrical Stimulation. Sensors 2022, 22, 335. [Google Scholar] [CrossRef]
  64. Jung, J.; Lee, D.-W.; Son, Y.; Kim, B.; Gu, J.; Shin, H.C. Volitional EMG Controlled Wearable FES System for Lower Limb Rehabilitation. In Proceedings of the 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC), Mexico City, Mexico, 1–5 November 2021; IEEE: Piscataway, NJ, USA, 2021; pp. 7099–7102. [Google Scholar] [CrossRef]
  65. Jung, J.; Lee, D.-W.; Son, Y.K.; Kim, B.S.; Shin, H.C. Volitional EMG Estimation Method during Functional Electrical Stimulation by Dual-Channel Surface EMGs. Sensors 2021, 21, 8015. [Google Scholar] [CrossRef]
  66. Andreu, D.; Sijobert, B.; Toussaint, M.; Fattal, C.; Azevedo-Coste, C.; Guiraud, D. Wireless Electrical Stimulators and Sensors Network for Closed Loop Control in Rehabilitation. Front. Neurosci. 2020, 14, 117. [Google Scholar] [CrossRef] [PubMed]
  67. Mrachacz-Kersting, N.; Jiang, N.; Stevenson, A.J.T.; Niazi, I.K.; Kostic, V.; Pavlovic, A.; Radovanovic, S.; Djuric-Jovicic, M.; Agosta, F.; Dremstrup, K.; et al. Efficient neuroplasticity induction in chronic stroke patients by an associative brain-computer interface. J. Neurophysiol. 2016, 115, 1410–1421. [Google Scholar] [CrossRef] [PubMed]
  68. Hayashibe, M. Evoked Electromyographically Controlled Electrical Stimulation. Front. Neurosci. 2016, 10, 335. [Google Scholar] [CrossRef] [PubMed]
  69. Benoussaad, M.; Rotella, F.; Chaibi, I. Flatness of musculoskeletal systems under functional electrical stimulation. Med. Biol. Eng. Comput. 2020, 58, 1113–1126. [Google Scholar] [CrossRef]
  70. Rezaee, Z.; Kobravi, H.R. Human Gait Control Using Functional Electrical Stimulation Based on Controlling the Shank Dynamics. Basic Clin. Neurosci. J. 2020, 11, 1–14. [Google Scholar] [CrossRef] [PubMed]
  71. Sun, Z.; Iyer, A.; Lambeth, K.; Cleveland, C.; Sharma, N. Knee extension tracking and fatigue regulation results using a robust MPC approach in a hybrid exoskeleton. Control Eng. Pract. 2023, 141, 105717. [Google Scholar] [CrossRef]
  72. Li, W.; Rovetta, A.; Ding, X.; Chen, L.; Han, Y. Virtual Multi-Interaction for Rehabilitation Robotics. In Proceedings of the 2020 5th International Conference on Advanced Robotics and Mechatronics (ICARM), Shenzhen, China, 18–21 December 2020; IEEE: Piscataway, NJ, USA, 2020; pp. 204–208. [Google Scholar] [CrossRef]
  73. Lim, D.; Pei, W.; Lee, J.W.; Musselman, K.E.; Masani, K. Feasibility of using a depth camera or pressure mat for visual feedback balance training with functional electrical stimulation. Biomed. Eng. Online 2024, 23, 19. [Google Scholar] [CrossRef]
  74. Pappas, I.P.; Popovic, M.R.; Keller, T.; Dietz, V.; Morari, M. A reliable gait phase detection system. IEEE Trans. Neural Syst. Rehabil. Eng. Publ. IEEE Eng. Med. Biol. Soc. 2001, 9, 113–125. [Google Scholar] [CrossRef] [PubMed]
  75. Erfanian, A.; Chizeck, H.J.; Hashemi, R.M. Using evoked EMG as a synthetic force sensor of isometric electrically stimulated muscle. IEEE Trans. Biomed. Eng. 1998, 45, 188–202. [Google Scholar] [CrossRef]
  76. Kutlu, M.; Freeman, C.T.; Hallewell, E.; Hughes, A.-M.; Laila, D.S. Upper-limb stroke rehabilitation using electrode-array based functional electrical stimulation with sensing and control innovations. Med. Eng. Phys. 2016, 38, 366–379. [Google Scholar] [CrossRef]
  77. Whittle, M.W. Clinical gait analysis: A review. Hum. Mov. Sci. 1996, 15, 369–387. [Google Scholar] [CrossRef]
  78. Gutierrez-Martinez, J.; Mercado-Gutierrez, J.A.; Carvajal-Gámez, B.E.; Rosas-Trigueros, J.L.; Contreras-Martinez, A.E. Artificial Intelligence Algorithms in Visual Evoked Potential-Based Brain-Computer Interfaces for Motor Rehabilitation Applications: Systematic Review and Future Directions. Front. Hum. Neurosci. 2021, 15, 772837. [Google Scholar] [CrossRef]
Figure 1. PRISMA flow diagram.
Figure 1. PRISMA flow diagram.
Jpm 14 00874 g001
Figure 2. Key criteria to choose sensing methods in FES systems for FD.
Figure 2. Key criteria to choose sensing methods in FES systems for FD.
Jpm 14 00874 g002
Table 1. Sensing and control strategies in FES systems.
Table 1. Sensing and control strategies in FES systems.
AuthorSubjectsType of MovementPurposeSensing Technology/StrategyMeasured Variable/SignalControl StrategyParameters Modulated
[20]HealthySittingRehabilitationEMG and EEG sensors
Torque transducers
Real-time control interface of the MAFO -
[21]HealthyWalkingAssistanceFSRs
IMUs
Ground reaction force
Angular rate
Acceleration
Reflexive control algorithmTiming and coordination of muscle activations
Stimulation sequences for specific muscles’
stimulation amplitude
[22]HealthyWalkingRehabilitationIMUs
FSRs
Dynamics of the leg and foot
Gait events
Ground reaction forces
CLPulse width
[23]BothSittingRehabilitation and AssistanceLoad cellTorque generated by the muscle contractions
Knee joint angle
CL, PID-
[24]-Sit-to-standAssistance--Combination of sliding mode control and wavelet networks-
[25]PatientsFES cyclingRehabilitationIMUsAngular velocity
Linear acceleration
Adaptive methodTiming of muscle contractions
[26]Healthy and patientWalking and sittingRehabilitationsEMG sensors
Torque sensors
Kinematic sensors
Joint angles
Angular velocity
Acceleration
Linear control schemesStimulation intensity
Pulse width
Frequency
[27]PatientsWalkingRehabilitationIMUsAngular displacements CL, FSMStimulation amplitude
[28]HealthySittingRehabilitation and AssistanceForce transducersForceCLStimulation amplitude
[29]PatientsWalking/sittingRehabilitationEMG/EEG --
[30]PatientsWalkingRehabilitationIMUs
Force sensors
Knee angle
Gait events
CL, Proportional controllerPulse width
[31]HealthyWalking and sittingRehabilitationMotion capture systemAnkle joint angleCL, ILCPulse width
Stimulation intensity
[32]BothWalkingRehabilitationsEMG sensors
Pressure Sensors
Angular velocity
Ground reaction forces
CL, NSStimulation intensity
[33]HealthyWalkingAssistanceFSRs
IMUs
Sagittal shank angle
Angular velocity
Acceleration
CL, FSMTime coefficient parameter (τ) and pulse width
[34]PatientsWalkingRehabilitationIMUsAngular velocity
Acceleration
CL, Proportional controllerPulse width
[35]HealthySittingRehabilitationGoniometerKnee jointMusculoskeletal model-integrated iterative learning control (MMILC)Pulse width
[36]PatientsWalkingRehabilitationIMUsFoot pose
initiation of the subject’s step
CL, NSYes, NS
[37]HealthySittingRehabilitationsEMG sensorsVolitional electrical activity of the musclesCL,
Gram–Schmidt filtering algorithm
Waveforms proportional to the measured vEMG envelope
[38]HealthyWalkingRehabilitationIMUs
FSRs
Heel pressure, shank tilt, and foot rotationsCL, FSMYes, NS
[39] HealthySittingRehabilitationOptical encodersKnee joint angleCLPulse amplitude
[40]PatientsWalking and sittingRehabilitation and AssistancesEMG sensors
IMUs
Angles of the ankles and feet Electrical activity of musclesCL, NSPWM signals
[41]PatientWalking and sittingRehabilitationIMUsFoot pitch
Roll rate state
CL, NS-
[42]HealthySittingRehabilitationIMUsknee joint angleCL
fuzzy controller
Yes, NS
[43]HealthyWalkingRehabilitationsEMG sensors
Ground reaction forces
Motion capture
Electrical activity
Timing of heel-strikes
hip and knee joint
CL, FSMTiming and sequence
[44]Patients
WalkingRehabilitationMotion Analysis systemMuscle Forces
Step length
Ankle angles
OLNo
[45]HealthySittingRehabilitationIMUsankle angleneural network-based H modelPulse amplitude
[17]HealthyWalkingRehabilitationIMUs
Motion Capture System
Angular velocityFSMNo
[46]HealthySittingRehabilitationsEMG sensors joint torqueCL, NSYes, NS
[7]PatientsWalkingAssistanceIMUsFoot angle and angular velocityCL, NSPulse width
[47]PatientsWalkingRehabilitationForefoot switch
accelerometer
Timing and intensity of muscle contractions No
[48]Both RehabilitationVideo-based motion capture/
IMUs/
Force-sensing resistance sensors/
EEG-sEMG sensors
Motion law of joints/
joint acceleration, velocity and angular acceleration/
planta pressure/
bioelectrical information
--
[3]HealthySittingRehabilitationEncoder
sEMG sensors
Knee joint angle, angular velocity, and accelerationOL-
[49]PatientsSwimingAssistanceIMUsTrunk roll angle and rate
Upper-arm inclination angle
Knee joint angle
Torso roll angle
CL and OL-
[50]HealthySittingRehabilitation and AssistanceLeg extension machine Torque produced about the knee jointCL
nonlinear control
Pulse width
[51]PatientsWalking and sittingRehabilitationIMUsAccelerations, Angular rates of the footCL, ILCStimulation intensity and timing
[52]PatientsWalkingRehabilitationangle sensorsangles of the knee jointCL, ILCPulse sequences
[53]PatientsWalkingRehabilitationHeel switches, accelerometers, and tilt sensorsGait symmetry, rhythmicity, and ankle movementsCL, NSDuration and intensity of the electrical stimulation
[54]BothCyclingRehabilitationsEMG
Optical encoder
Electrical activity
Crank position
CL
repetitive learning control
Pulse width
[55]PatientsSittingRehabilitationGoniometers
Dynamometers
sEMG sensors
--
[56]HealthySittingRehabilitationsEMG sensorsMotor-evoked potentials (MEPs)OLNo
[4]PatientsWalkingRehabilitation and AssistanceElectrode padsAngle of ankle dorsiflexion
Position
Electrical activity of muscles
CL, RCPulse width
[57]HealthyWalkingRehabilitationsEMG sensorsM-waveCL
model predictive control (MPC)
Pulse width
[58]BothSittingRehabilitationsEMG sensors
Dynamometer
Evoked electromyography
Ankle joint torque
--
[59]BothSittingAssistanceEMG amplifiers
IMUs
goniometers
Muscle activation patterns, Joint angles--
[60]BothStandingRehabilitationElectrogoniometer
Gyroscope
Accelerometers
Angular position of lower limbCL
RISE control strategy
-
[61]HealthySittingRehabilitationUS imagingDelayed muscle activationCL, NS-
[62]HealthyStandingRehabilitationforce transducers
optical encoders
torque produced about the knee-joint and the knee-joint angleCL, asynchronous stimulation controllerPulse width
Stimulation channels
[63]HealthySittingRehabilitationUSUS echogenicity signalCL, NSPulse width threshold and saturation values
[15]HealthyWalkingRehabilitationUS
Force plates
IMUs
US echogenicity
Ground reaction force
2D motion in the sagittal plane
CL, NSPulse width
[64]HealthySquat/heel lift/walkingRehabilitationsEMG sensors
IMUs
Motion intention
Joint angle
CL, NSYes, NS
[65]HealthySittingRehabilitation and AssistancesEMG sensorsMuscle activity--
[66]PatientsWalkingRehabilitationDistributed Measurement Units (DMU)Knee joint angle Gait phasesCL, NSYes, NS
[44]PatientsWalkingRehabilitationFoot switch
Motion capture
Heel strike and toe off
Muscle forces
Step length
Maximum knee and ankle angles
OL-
[67]PatientWalking and sittingRehabilitationSurface electrodes
IMUs
EEG and EMG
Tapping frequency of finger or foot
CL-
[68]--RehabilitationEMGEvoked EMGCL-
[69]Simulation-Rehabilitation--OL-
[70]HealthyWalkingRehabilitationMotion analyzerShank and thigh movementFuzzy controllerStimulation intensity
[71]BothWalking and sittingAssistedDynamometerTorqueRISE controller-
[72]PatientWalkingRehabilitationIMUs
Force sensor
Posture information of limb movement
Specific movement of limb
OL-
[73]HealthyStanding balance exerciseRehabilitationDepth camera
Pressure mat
Center of mass
Center of pressure
--
EMG—electromyography; EEG—electroencephalogram; IMUs—inertial measurement units; FSRs—force-sensing resistor; CL—closed-loop; FSM—finite state machine; RC—reflexive control; OL—open-loop; NS—non-specified; ILC—iterative learning control; US—ultrasound; MAFO—motorized ankle–foot orthosis; RISE—robust integral of the sign of the error; PID—proportional–integral derivative; vEMG—volitional electromyography; sEMG—surface electromyography; PMW—pulse width modulation.
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

González-Graniel, E.; Mercado-Gutierrez, J.A.; Martínez-Díaz, S.; Castro-Liera, I.; Santillan-Mendez, I.M.; Yanez-Suarez, O.; Quiñones-Uriostegui, I.; Rodríguez-Reyes, G. Sensing and Control Strategies Used in FES Systems Aimed at Assistance and Rehabilitation of Foot Drop: A Systematic Literature Review. J. Pers. Med. 2024, 14, 874. https://doi.org/10.3390/jpm14080874

AMA Style

González-Graniel E, Mercado-Gutierrez JA, Martínez-Díaz S, Castro-Liera I, Santillan-Mendez IM, Yanez-Suarez O, Quiñones-Uriostegui I, Rodríguez-Reyes G. Sensing and Control Strategies Used in FES Systems Aimed at Assistance and Rehabilitation of Foot Drop: A Systematic Literature Review. Journal of Personalized Medicine. 2024; 14(8):874. https://doi.org/10.3390/jpm14080874

Chicago/Turabian Style

González-Graniel, Estefanía, Jorge A. Mercado-Gutierrez, Saúl Martínez-Díaz, Iliana Castro-Liera, Israel M. Santillan-Mendez, Oscar Yanez-Suarez, Ivett Quiñones-Uriostegui, and Gerardo Rodríguez-Reyes. 2024. "Sensing and Control Strategies Used in FES Systems Aimed at Assistance and Rehabilitation of Foot Drop: A Systematic Literature Review" Journal of Personalized Medicine 14, no. 8: 874. https://doi.org/10.3390/jpm14080874

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop