Article

Perfluorooctanesulfonate Can Cause Negative Bias in Creatinine Measurement in Hemodialysis Patients Using Polysulfone Dialysis Membranes

1
Division of Nephrology, Department of Medicine, Taipei City Hospital Zhongxing Branch, Taipei 103, Taiwan
2
School of Medicine, National Yang Ming Chiao Tung University, Hsinchu 300, Taiwan
3
College of Science and Engineering, Fu Jen Catholic University, New Taipei City 242, Taiwan
4
Institute of Food Safety and Health Risk Assessment, National Yang Ming Chiao Tung University, Hsinchu 300, Taiwan
5
Department of Special Education, University of Taipei, Taipei 100, Taiwan
6
Department of Pediatrics, Taipei Veterans General Hospital, Taipei 112, Taiwan
7
Division of Nephrology, Department of Medicine, Taipei Veterans General Hospital, Taipei 112, Taiwan
8
Kim Forest Enterprise Co., Ltd., New Taipei City 221, Taiwan
9
Department of Child Psychiatry, Chang Gung Memorial Hospital and University, Taoyuan 333, Taiwan
10
College of Medicine, National Taiwan University, Taipei 100, Taiwan
11
Department of Physical Medicine and Rehabilitation, National Taiwan University Hospital Hsin-Chu Branch, Hsinchu 300, Taiwan
*
Authors to whom correspondence should be addressed.
These authors contributed equally to this manuscript.
Membranes 2022, 12(8), 778; https://doi.org/10.3390/membranes12080778
Submission received: 29 June 2022 / Revised: 3 August 2022 / Accepted: 9 August 2022 / Published: 13 August 2022
(This article belongs to the Special Issue Membranes in Renal Replacement Therapy)

Abstract

:
Serum creatinine is an important clinical marker for renal clearance. However, two conventional methods (Jaffe and enzymatic) are prone to interferences with organic compounds as compared to the standard method (isotope dilution–liquid chromatography–mass spectrometry) and can cause a significant negative bias. Perfluorooctanesulfonate (PFOS) and perfluorooctanoic acid (PFOA) are two common perfluorochemicals (PFCs) that can easily be accumulated in humans. We aimed to verify whether this bias is the result of an accumulation of PFCs. The serum creatinine values of 124 hemodialysis patients were analyzed using the three methods. We also aimed to evaluate which biochemical parameters will influence the difference between the conventional methods and the standard method. We found that a significant underestimation occurred when using the conventional methods. Albumin is an independent factor associated with negative bias, but it loses this correlation after dialysis, likely due to the removal of protein-bound uremic toxins. PFOS can cause negative bias when using the enzymatic method. Furthermore, this linear correlation is more significant in patients who used polysulfone-based dialysis membranes, possibly due to the better clearance of other uremic toxins. The serum creatinine of uremic patients can be significantly underestimated when using conventional methods. PFCs, as well the type of dialysis membrane being used, can be influencing factors.

1. Introduction

Inulin clearance is considered to be the most accurate determination of renal function [1,2]. However, it is difficult to perform the procedure on a regular basis. Therefore, serum creatinine and its clearance is generally used as a convenient marker to evaluate renal function [3]. Serum creatinine is the most commonly used method to determine renal function in the clinical setting, but serum creatinine values can be influenced not only by the method of measurement but also by other confounding factors [4]. There are two methods available clinically: the more popular Jaffe method [5] and the enzymatic method. There is also the gold standard method, which is the isotope dilution–liquid chromatography–mass spectrometry (IDLCMS) method [6,7].
The Jaffe method utilizes oxidation to form a chromophore in serum creatinine measurement in the laboratory. However, it can be easily confounded with chromogens such as nitromethane, as well as glucose and protein with positive bias [8,9,10,11]. On the other hand, the enzymatic method is less affected by glucose. It is useful to check creatinine levels in high-glucose-containing peritoneal dialysate. However, the enzymatic method is still influenced by certain organic compounds [6]. Among the three methods, IDLCMS is the most accurate, but it is expensive, which limits its application [12,13,14,15]. In a healthy population, these three methods show a good linear correlation.
Uremic patients whose serum creatinine levels are higher than the normal population suffer from renal failure and are subjected to an accumulation of toxic waste [16,17]. The linear correlation of the three different methods in uremic patients was tested in our previous study in 2012, and it was found that albumin acts as a negative bias in uremic serum instead of a positive bias in normal serum. Positive bias means that the measured value is higher than the real value and negative bias means the measured value is lower [18]. In that study, we proposed a theory that the reason albumin acts as a negative bias may be related to protein-bound uremic toxins. We hypothesized that these toxins which cause negative bias may accumulate in uremic patients and be bounded by albumin, but failed to find proof of this over several years. In another previous study, in 2018, we found that perfluorochemicals (PFCs) are higher in uremic patients and are mainly carried by albumin [19]. The most two common PFCs are perfluorooctanoic acid (PFOA) and perfluorooctanesulfonate (PFOS), which are also persistent organic pollutants. These compounds with long half-lives are difficult to eliminate from the human body and may cause eventual endocrine disturbance [20,21], Furthermore, not all dialysis membranes can provide adequate clearance for PFCs. Results from a previous study showed that polysulfone dialysis membranes provide better clearance for PFCs [21].
Because there is no study at the moment that proves that protein-bound toxins may interfere with serum creatinine measurement in uremic patients, the aim of this study is to determine whether serum PFCs may cause errors in the Jaffe and enzymatic methods in uremic patients. We also aim to investigate how different dialysis membrane properties and the clearance of PFCs may influence the serum creatinine measurement error in the Jaffe and enzymatic methods.

2. Materials and Methods

2.1. Inclusion and Exclusion Criteria

We included patients aged 18 to 90 who had been undergoing 4 h of routine hemodialysis three times per week for more than three months. Blood samples before and after dialysis were collected at the beginning of the month from a teaching hospital in northern Taiwan.
Patients who had undergone peritoneal dialysis and transplantation were excluded. Patients who had intravenous medications, such as lipid nutrition supplement, propofol, dopamine, methotrexate, vancomycin, furosemide, and cyclosporine were also excluded. Those who had blood loss with transfusion or admission were also excluded. A total of 124 patients were included.

2.2. Clinical and Biochemistry Data Collection

Demographic and clinical data such as gender, cause of renal failure (diabetes or chronic glomerulonephritis), age, and duration under hemodialysis were obtained from patients’ medical records. Laboratory parameters were gathered at the beginning of the month prior to hemodialysis.
In this study, the liquid chromatography tandem mass spectrometry (LC-MS/MS) was used with an isotope dilution to quantify the serum creatinine, PFOA, and PFOS. The differences in creatinine, PFOA, and PFOS before and after dialysis were also evaluated. The hemogram auto-analyzer used was SYSMEX XE2100 (Sysmex, Kobe, Japan). The Jaffe and other biochemical parameters were determined by the ADVIA® 1800 Chemistry System (Siemens Healthcare GmbH, Erlangen, Germany). The enzymatic method was determined by the Beckman AU640 (Beckman Coulter Inc., Brea, CA, USA) with amidohydrolase procedure. We used IDLCMS as the gold standard to evaluate the differences between the other two methods. The difference between Jaffe and IDLCMS is termed D-Jaffe and that of the enzymatic method and IDLCMS is termed D-enzymatic.
First, we checked the creatinine values of the three methods. Then, we checked the association of D-Jaffe and D-enzymatic between clinical and biochemical factors by linear regression. Those significant factors were further analyzed with multiple linear regression to find the independent influencing factors for creatinine measurement. The post-dialysis values of independent influencing factors were analyzed again with D-Jaffe and D-enzymatic. Patients with different dialysis membranes were compared with their PFC levels and the association with D-Jaffe and D-enzymatic.
The procedures of measuring creatinine and PFCs is described in the supplement. The study was conducted in accordance with the Declaration of Helsinki and the protocol was approved by the Institutional Review Board of the hospital.

2.3. Statistical Analysis

Continuous data were expressed as mean ± standard deviation. Statistical analysis was performed using IBM SPSS Statistics for Windows, Version 20.0 (IBM Corp., Armonk, NY, USA). The paired t-test was used to compare the difference before and after dialysis. The χ2 test was used for categorical variables and the t-test for continuous variables. Distributions of continuous variables in groups were expressed as mean ± SD. Multivariate forward logistic regression analysis was applied to identify the independent determinant factors. A p value of less than 0.05 was regarded as statistically significant.

3. Results

Table 1 and Figure 1 showed the value of serum creatinine measured in three methods and the difference between Jaffe and IDLCMS (D-Jaffe) and that of the enzymatic method and IDLCMS (D-enzymatic). The underestimation is about 10–15%. The creatinine values of the enzymatic method were also significantly lower than those of the Jaffe method due to the larger negative bias of D-enzymatic (Table 1).
Table 2 showed that gender and DM as comorbidities did not affect the measurement error of serum creatinine. Continuous variables such as age, serum biochemical markers, and complete blood count index were also analyzed with linear regression to the D-Jaffe and D-enzymatic. (Table 2).
The factors which were significantly correlated with D-Jaffe and D-enzymatic were further analyzed with multiple linear regression in order to find the independent influencing factors (Table 3). The independent factors were serum albumin and serum Cr (both Jaffe and enzymatic method). As for PFCs, PFOA was not significantly removed by dialysis and did not affect the D-Jaffe and D-enzymatic values. On the other hand, PFOS only affected the enzymatic method. PFOS fulfilled the aforementioned requirement and was shown to cause negative bias in the enzymatic method pre-hemodialysis (Figure 2).
After dialysis, markers related to serum albumin, creatinine, and PFCs were analyzed in Table 4. Serum albumin lost its ability to associate with negative bias in both the D-Jaffe and D-enzymatic methods (Figure 3). There were lower PFOS levels post-dialysis and these did not affect serum creatinine values in the enzymatic method. However, serum creatinine was still strongly related with negative error for creatinine measurement.
Upon further analysis of dialysis membranes, PFOS causing negative bias in D-enzymatic were mainly seen in patients who used polysulfone-based dialysis membranes and this can be confirmed by the lower PFOS levels seen in patients using polysulfone-based dialysis membranes (Table 5). This may be related to the better clearance of uremic toxins and fewer confounding toxins.

4. Discussion

We found significant negative measurement errors using the Jaffe method (D-Jaffe) and enzymatic method (D-enzymatic), as compared with the gold standard IDLCMS method. The error may cause approximately a 10–15% reduction compared to the real value, which is more significant when serum creatinine is high [18]. This finding explains why serum creatinine is not a sensitive marker to monitor renal function deterioration. The more toxins that are accumulated during advanced chronic kidney disease, the more they provide a negative bias, which offsets the expected increase during serum creatinine measurement [22].
Table 2 showed the factors that are significantly associated with D-Jaffe and D-enzymatic. After multiple linear regression, the two common independent factors related to the negative bias of D-Jaffe and D-enzymatic were serum albumin and the absolute value of Cr, as shown in Table 3. The post-dialysis association of independent factors related to D-Jaffe and D-enzymatic were shown in Table 4.
The fact that serum albumin lost its ability to cause negative bias post-dialysis was compatible with our hypothesis that at least one protein-bound uremic toxin may be removed by dialysis. Albumin is a negatively charged protein and a good carrier which binds with various cations and toxins, including PFCs [19] (Table 4 and Figure 3).
The other factor is creatinine, which is an indicator of glomerular filtration rate and renal function. Various toxins may accumulate in the bloodstream of uremic patients and some of them may not be dialyzable [19]. Hence, they may be still exist even after dialysis and produce negative bias on the serum creatinine value. Serum phosphorus was still significantly associated with the negative bias of serum creatinine measurement post-dialysis. This may be due to serum phosphorus also being associated with renal function decline, and only being able to be removed by half on regular hemodialysis [23]. On the other hand, serum potassium is easily removed by HD. Hence, serum potassium was not related with renal function post-dialysis.
As for PFCs, there is a higher concentration of PFOS than any other PFCs (such as PFOA) because of a longer history of usage [21]. Through multiple linear regression analysis, we found that PFOS was also an independent factor causing negative bias in the enzymatic method but not in the Jaffe method (Table 3). Upon further comparison of the differences between D-Jaffe and D-enzymatic, and analysis with linear regression, we found that PFOS was highly associated with this difference (p = 0.048, R = 0.226) and PFOS may be the reason why serum creatinine was lower in the enzymatic method as compared with the Jaffe method.
PFOS confirmed our previous hypothesis [18] that uremic toxins causing negative bias may be bound with albumin. After dialysis, the toxin is removed and albumin also simultaneously loses its association with the negative bias of serum creatinine.
There are several factors that contribute to the results of this study. First, uremic toxins are composed of many materials, many of which are not yet fully understood. Second, the toxin concentrations may be too low to measure in an ordinary laboratory setting. IDLCMS was required to accurately measure PFCs [24]. The procedure for the measurement of serum creatinine and PFCs is described in the Supplementary Material; it requires appropriate equipment and well-trained staff. Third, the association between the dialysis membrane and clearance was also an important factor. There is a lower PFOS concentration in the serum of patients with polysulfone-based dialysis membranes compared with other materials (Table 5). This may be related to the better clearance of the polysulfone-based membrane for protein-bound toxins [24]. Polysulfone is more hydrophobic than other materials such as cellulose, and it has a better compatibility and clearance for toxins [20]. Therefore, the linear regression of PFOS and D-enzymatic were more obvious in this patient group due to fewer confounding toxins. As more studies showed the advantage of polysulfone-based membranes, it has gradually become the main dialysis membrane material in recent years [25]. In our current study, 75% of patients were using polysulfone-based dialysis membranes.
The limitation of the study is that we found only one toxin (PFOS) which causes a negative bias and is bound with albumin. However, it still confirmed our hypothesis, put forward in 2012, as to why albumin acts differently in the serum creatinine measurement of uremic patients. The strength of the study is that we used IDLCMS to evaluate the true value of serum creatinine, which is quite difficult and requires elaborate procedures. The concentrations of environmental toxins (PFOA and PFOS) were also too low to be measured in the ordinary laboratory setting. Furthermore, without the better clearance provided by polysulfone-based dialysis membranes, it will be impossible to find the linear association between PFOS and the negative bias measured in the enzymatic method because there are too many other confounding toxins found in the serum. More toxins and membrane materials still need to be investigated. There may be other non-dialyzable toxins causing post-dialysis measurement errors. The negative bias associated with post-dialysis serum creatinine indicates the above mentioned possibility.

5. Conclusions

The serum creatinine of uremic patients can be significantly underestimated when using conventional methods. PFOS, a type of PFC, as well the type of dialysis membrane being used, can be influencing factors.

Supplementary Materials

The following supporting information can be downloaded at: www.mdpi.com/article/10.3390/membranes12080778/s1, Figure S1: Molecular structures of PFOA and PFOS, Figure S2: Product ion scans of PFOA and PFOS.

Author Contributions

Conceptualization, W.-S.L., H.-H.T., H.-L.C. and Y.-T.L.; methodology, H.-H.T.; software, C.-H.L.; validation, S.-Y.L., C.-C.L. and T.-Y.L.; formal analysis, W.-S.L.; investigation, W.-S.L.; resources, H.-H.T.; data curation, Y.-T.L.; writing—original draft preparation, W.-S.L.; writing—editing, A.C.T.; visualization, H.-L.C.; supervision, H.-L.C. project administration, Y.-T.L.; funding acquisition, W.-S.L. and H.-L.C. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by Taipei Veterans General Hospital (2011-02-022IC) and Taipei City Hospital (TPCH-108-27).

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki, and approved by the Institutional Review Board of Taipei Veterans General Hospital (2011-02-022IC) and Taipei City Hospital (TCHIRB-10901017-E).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Ocampo, J.H.; Rosales, A.T.; Castellanos, F.R. Comparison of Four Methods for Measuring Glomerular Filtration Rate by Inulin Clearance in Healthy Individuals and Patients with Renal Failure. Nefrologia 2010, 30, 324–330. [Google Scholar] [CrossRef]
  2. Schwartz, G.J.; Furth, S.L. Glomerular filtration rate measurement and estimation in chronic kidney disease. Pediatr. Nephrol. 2007, 22, 1839–1848. [Google Scholar] [CrossRef] [PubMed]
  3. Bjornsson, T.D. Use of serum creatinine concentrations to determine renal function. Clin. Pharmacokinet. 1979, 4, 200–222. [Google Scholar] [CrossRef]
  4. Weber, J.A.; Vanzanten, A.P. Interferences in current methods for measurements of creatinine. Clin. Chem. 1991, 37, 695–700. [Google Scholar] [CrossRef] [PubMed]
  5. Blass, K.G.; Thibert, R.J.; Lam, L.K. Study of Mechanism of Jaffe Reaction. Z. Fur Klin. Chem. Und Klin. Biochem. 1974, 12, 336–343. [Google Scholar] [CrossRef] [PubMed]
  6. Gottschalk, E.M.; Holdt-Lehmann, B.; Bastian, M.; Schmitz, R.A.; Schuff-Werner, P. Automatisation and interference testing of a new creatinine assay using recombinant creatinine deiminase. Clin. Chem. 2004, 50, A27. [Google Scholar]
  7. Tsikas, D.; Wolf, A.; Mitschke, A.; Gutzki, F.-M.; Will, W.; Bader, M. GC-MS determination of creatinine in human biological fluids as pentafluorobenzyl derivative in clinical studies and biomonitoring: Inter-laboratory comparison in urine with Jaffe, HPLC and enzymatic assays. J. Chromatogr. B 2010, 878, 2582–2592. [Google Scholar] [CrossRef]
  8. Booth, C.J.; Naidoo, D.; Rosenberg, A.R.; Kainer, G. Elevated creatinine after ingestion of model aviation fuel: Interference with the Jaffe reaction by nitromethane. J. Paediatr. Child Health 1999, 35, 503–504. [Google Scholar] [CrossRef]
  9. Bruns, D.E. Lactulose interferes in the alkaline picrate assay for creatinine. Clin. Chem. 1988, 34, 2592–2593. [Google Scholar] [CrossRef]
  10. Deleacy, E.A.; Brown, N.N.; Clague, A.E. Nitromethane interferes in assay of creatinine by the Jaffé reaction. Clin. Chem. 1989, 35, 1772–1774. [Google Scholar] [CrossRef]
  11. Feld, L.G.; Langford, D.J.; Schwartz, G.J. The effect of neonatal hyperbilirubinemia on the measurement of plasma creatinine. Clin. Pediatr. 1984, 23, 154–156. [Google Scholar] [CrossRef] [PubMed]
  12. Guy, J.M.; Legg, E.F. An improved cation exchange HPLC method for the measurement of serum creatinine. Ann. Clin. Biochem. 1990, 27, 223–226. [Google Scholar] [CrossRef] [PubMed]
  13. Linnet, K.; Bruunshuus, I. HPLC with enzymatic detection as a candidate reference method for serum creatinine. Clin. Chem. 1991, 37, 1669–1675. [Google Scholar] [CrossRef] [PubMed]
  14. Seronie-Vivien, S.; Galteau, M.M.; Carlier, M.C.; Hadj-Aissa, A.; Hanser, A.M.; Hym, B.; Marchal, A.; Michotey, O.; Pouteil-Noble, C.; Sternberg, M.; et al. Improving the interlaboratory variation for creatinine serum assay. Ann. Biol. Clin. 2004, 62, 165–175. [Google Scholar]
  15. Yuen, P.S.T.; Dunn, S.R.; Miyaji, T.; Yasuda, H.; Sharma, K.; Star, R.A. A simplified method for HPLC determination of creatinine in mouse serum. Am. J. Physiol.-Renal Physiol. 2004, 286, F1116–F1119. [Google Scholar] [CrossRef]
  16. Kikuchi, K.; Itoh, Y.; Tateoka, R.; Ezawa, A.; Murakami, K.; Niwa, T. Metabolomic search for uremic toxins as indicators of the effect of an oral sorbent AST-120 by liquid chromatography/tandem mass spectrometry. J. Chromatogr. B 2010, 878, 2997–3002. [Google Scholar] [CrossRef]
  17. Myers, G.L.; Miller, W.G.; Coresh, J.; Fleming, J.; Greenberg, N.; Greene, T.; Hostetter, T.; Levey, A.S.; Panteghini, M.; Welch, M.; et al. Recommendations for improving serum creatinine measurement: A report from the laboratory working group of the National Kidney Disease Education Program. Clin. Chem. 2006, 52, 5–18. [Google Scholar] [CrossRef]
  18. Liu, W.S.; Chung, Y.T.; Yang, C.Y.; Lin, C.C.; Tsai, K.H.; Yang, W.C.; Chen, T.W.; Lai, Y.T.; Li, S.Y.; Liu, T.Y. Serum creatinine determined by Jaffe, enzymatic method, and isotope dilution-liquid chromatography-mass spectrometry in patients under hemodialysis. J. Clin. Lab. Anal. 2012, 26, 206–214. [Google Scholar] [CrossRef]
  19. Liu, W.S.; Lai, Y.T.; Chan, H.L.; Li, S.Y.; Lin, C.C.; Liu, C.K.; Tsou, H.H.; Liu, T.Y. Associations between perfluorinated chemicals and serum biochemical markers and performance status in uremic patients under hemodialysis. PLoS ONE 2018, 13, e0200271. [Google Scholar] [CrossRef]
  20. Coperchini, F.; Awwad, O.; Rotondi, M.; Santini, F.; Imbriani, M.; Chiovato, L. Thyroid disruption by perfluorooctane sulfonate (PFOS) and perfluorooctanoate (PFOA). J. Endocrinol. Investig. 2017, 40, 105–121. [Google Scholar] [CrossRef]
  21. Liu, W.S.; Chan, H.L.; Lai, Y.T.; Lin, C.C.; Li, S.Y.; Liu, C.K.; Tsou, H.H.; Liu, T.Y. Dialysis Membranes Influence Perfluorochemical Concentrations and Liver Function in Patients on Hemodialysis. Int. J. Environ. Res. Public Health 2018, 15, 2574. [Google Scholar] [CrossRef] [PubMed]
  22. Levey, A.S.; Titan, S.M.; Powe, N.R.; Coresh, J.; Inker, L.A. Kidney Disease, Race, and GFR Estimation. Clin. J. Am. Soc. Nephrol. 2020, 15, 1203–1212. [Google Scholar] [CrossRef] [PubMed]
  23. Zhang, W.; Ye, G.; Bi, Z.; Chen, W.; Qian, J.; Zhang, M.; Ding, D.; Wang, M.; Chen, J. Higher one-year achievement rate of serum phosphate associated with lower cardiovascular mortality in hemodialysis patients. BMC Nephrol. 2021, 22, 398. [Google Scholar] [CrossRef] [PubMed]
  24. Derkits, D.R.; Meggs, W.J.; Parker Cote, J.L. Trending of a falsely elevated serum creatinine after a pediatric nitromethane ingestion: A case report. J. Am. Coll. Emerg. Physicians Open 2022, 3, e12689. [Google Scholar] [CrossRef]
  25. Tsuruoka, S.; Sugimoto, K.I.; Hayasaka, T.; Saito, T.; Fujimura, A. Ranitidine clearance during hemodialysis with high-flux membrane: Comparison of polysulfone and cellulose acetate hemodialyzers. Eur. J. Clin. Pharmacol. 2000, 56, 581–583. [Google Scholar] [CrossRef]
Figure 1. Creatinine measurement using the Jaffe and enzymatic methods as compared with the IDLCMS standard (* p < 0.05). (A) Creatinine measurement in the three methods before dialysis; (B) D-Jaffe and D-enzymatic before dialysis; (C) Creatinine measurement in the three methods after dialysis; (D) D-Jaffe and D-enzymatic after dialysis.
Figure 1. Creatinine measurement using the Jaffe and enzymatic methods as compared with the IDLCMS standard (* p < 0.05). (A) Creatinine measurement in the three methods before dialysis; (B) D-Jaffe and D-enzymatic before dialysis; (C) Creatinine measurement in the three methods after dialysis; (D) D-Jaffe and D-enzymatic after dialysis.
Membranes 12 00778 g001
Figure 2. Linear regression of PFOA and PFOS with Cr difference in Jaffe and enzymatic methods (pre-dialysis). (D-Jaffe, Jaffe Cr−IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr); (A) Linear regression of PFOA with D-Jaffe; (B) Linear regression of PFOA with D-Enzymatic; (C) Linear regression of PFOS with D-Jaffe; (D) Linear regression of PFOS with D-Enzymatic.
Figure 2. Linear regression of PFOA and PFOS with Cr difference in Jaffe and enzymatic methods (pre-dialysis). (D-Jaffe, Jaffe Cr−IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr); (A) Linear regression of PFOA with D-Jaffe; (B) Linear regression of PFOA with D-Enzymatic; (C) Linear regression of PFOS with D-Jaffe; (D) Linear regression of PFOS with D-Enzymatic.
Membranes 12 00778 g002
Figure 3. Linear regression of albumin with Cr difference using the Jaffe and enzymatic methods. (A) Linear regression of albumin with D-Jaffe pre-dialysis; (B) Linear regression of albumin with D-enzymatic pre-dialysis; (C) Linear regression of albumin with D-Jaffe post-dialysis; (D) Linear regression of albumin with D-enzymatic post-dialysis.
Figure 3. Linear regression of albumin with Cr difference using the Jaffe and enzymatic methods. (A) Linear regression of albumin with D-Jaffe pre-dialysis; (B) Linear regression of albumin with D-enzymatic pre-dialysis; (C) Linear regression of albumin with D-Jaffe post-dialysis; (D) Linear regression of albumin with D-enzymatic post-dialysis.
Membranes 12 00778 g003
Table 1. Serum creatinine values of three measuring methods and D-Jaffe and D-enzymatic pre- and post-dialysis.
Table 1. Serum creatinine values of three measuring methods and D-Jaffe and D-enzymatic pre- and post-dialysis.
Serum Cr
n = 124
Pre-HD Mean±SDt-Test
(p)
Post-HD
Mean
±SDt-Test
(p)
Jaffe Cr (mg/dL)9.402.26<0.001 *3.041.04<0.001 *
Enzymatic Cr(mg/dL)9.272.34<0.001 *2.580.94<0.001 *
vs. IDLCMS Cr (mg/dL) as standard11.122.79 3.261.17
D-Jaffe = Jaffe Cr − IDLCMS Cr−1.730.860.023 *−0.310.55<0.001 *
D-enzymatic = enzymatic Cr − IDLCMS Cr−1.850.97 −0.740.47
HD, hemodialysis; SD, standard deviation; Cr, creatinine; D-Jaffe, Jaffe Cr–IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr; IDLCMS, isotope dilution–liquid chromatography–mass spectrometry. * p < 0.05.
Table 2. Demographic, treatment, hemogram, and biochemical data with linear regression to D-Jaffe and D-enzymatic pre-hemodialysis.
Table 2. Demographic, treatment, hemogram, and biochemical data with linear regression to D-Jaffe and D-enzymatic pre-hemodialysis.
N = 124n (%) D-JaffeD-Enzymatic
Mean ± SDpMean ± SDp
Male72 (58.0%) −1.79 ± 0.920.185−1.92 ± 1.060.172
Female52 (42.0%) −1.59 ± 0.72 −1.70 ± 0.77
DM51 (41.1%) −1.59 ± 0.810.126−1.73 ± 0.900.252
Non-DM73 (58.9%) −1.82 ± 0.88 −1.93 ± 1.01
D-Jaffe D-enzymatic
Linear regressionMean±SD(β)(p)(β)(p)
Dialysis duration (month)59.7567.750.0000.9240.0000.913
Dialysis frequency (/week)2.980.120.4230.4830.6790.702
Dialysis time (hours)4.0230.37−0.2440.223−0.3470.124
Age (year) *59.7514.790.0180.001 *0.0170.002 *
Jaffe Cr (mg/dL) *9.402.26−0.2300.000 *−0.1860.000 *
Enzymatic Cr (mg/dL) *9.272.34−0.1920.000 *−0.1200.001 *
IDLCMS Cr (mg/dL) *11.122.79−0.2200.000 *−0.2060.000 *
Ccr (mL/min) *6.032.000.1970.000 *0.1660.000 *
WBC (×1000/μL)6.832.460.0190.5270.0140.686
RBC (×106/μL)3.360.50−0.2420.099−0.2850.084
MCV (fl)91.177.230.0020.856−0.0090.429
Hb (g/dL) *9.891.20−0.1350.027 *−0.2460.000 *
Platelet (×1000/μL)195.7168.31−0.0010.529−0.0010.391
Cholesterol (mg/dL) *154.7035.57−0.0050.011 *−0.0060.010
Glucose (mg/dL)136.8156.860.0010.2800.0020.186
Total protein(gm/dL)6.943.980.0130.4900.0070.744
Albumin (gm/dL) *3.920.37−1.0580.000 *−0.10260.000 *
Globulin2.954.020.0090.6130.0060.799
AST (IU/L)22.7010.450.0100.1520.0170.029 *
ALT (IU/L)18.7710.88−0.0010.891−0.0020.838
Alk-P (IU/L)93.7283.060.0010.1000.0010.409
Total Bilirubin (mg/dL)0.540.15−0.0340.9470.5480.340
Na (mEq/L) *138.922.73−0.0600.027 *−0.0400.200
K (mEq/L) *4.560.66−0.2320.041 *−0.1640.200
Cl (mEq/L)98.835.62−0.0200.144−0.0160.297
Ca (mg/dL) *9.270.86−0.2050.018 *−0.2870.003 *
P (mg/dL) *4.691.33−0.1380.013 *−0.0590.352
BUN (mg/dL)61.6917.09−0.0100.021−0.0110.020 *
Fe (ug/dL)59.3922.15−0.0060.067−0.0100.007 *
TIBC (ug/dL) *248.3848.10−0.0030.029 *0.0000.808
Ferritin(ng/mL)488.95422.600.0000.6030.0000.059
iPTH (pg/mL)118.40(44.17, 239.17)0.0000.4348.91 × 10−50.763
Al (ng/mL)15.169.85−0.0190.229−0.0210.193
PFOA (ng/mL)0.530.27−0.3460.362−0.4940.247
PFOS (ng/mL) *5.50(1.17,24.7)−0.0020.307−0.0050.048 *
D-Jaffe, Jaffe Cr−IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr; SD, standard deviation; DM, diabetes mellitus; IDLCMS, isotope dilution-liquid chromatography-mass spectrometry; Ccr, creatinine clearance; WBC, white blood cell; RBC, red blood cell; Hb, hemoglobin; MCV, mean corpuscular volume; AST, aspartate aminotransferase; ALT, alanine transaminase; Alk-P, alkaline phosphatase; Na, sodium; K, potassium; Cl, chloride; Ca, calcium; P, phosphorus; BUN, blood urea nitrogen; Fe, iron; TIBC, total iron-binding capacity; iPTH, parathyroid hormone; Al, aluminum; PFOA, perfluorooctanoic acid; PFOS, perfluorooctanesulfonate. Pre-dialysis PFOS and PTH variables were not normally distributed and displayed as quartile. * p < 0.05.
Table 3. Multivariate linear regression model of factors associated with D-Jaffe and D-enzymatic post-dialysis.
Table 3. Multivariate linear regression model of factors associated with D-Jaffe and D-enzymatic post-dialysis.
D-Jaffe
(r2 = 0.352, p < 0.001)
B EstimatepD-Enzymatic
(r2 = 0.435, p < 0.001)
B Estimatep
Jaffe Cr (mg/dL) *−0.1530.000 *Albumin (gm/dL) *−0.9880.013 *
Albumin (gm/dL) *−0.5930.011 *PFOS (ng/mL) *−0.0050.038 *
Cholesterol (mg/dL)−0.0030.112Jaffe Cr (mg/dL) *−0.1060.043 *
P (mg/dL)0.0880.156Fe (μg/dL)−0.0070.130
K (meq/L)−0.0870.410AST (IU/L)0.0100.365
Na (meq/L)−0.0100.697Total Ca (mg/dL)0.1230.383
Age0.0020.760Hb (g/dL)−0.0750.427
Hb (g/dL)−0.0050.927Cholesterol (mg/dL)−0.0010.634
Total Ca (mg/dL)0.0070.933BUN (mg/dL)0.0030.660
TIBC (ug/dL)0.0000.950Age0.0010.870
BUN (mg/dL)0.0000.954
D-Jaffe, Jaffe Cr−IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr; P, phosphorus; K, potassium; Na, sodium; Hb, hemoglobin; Ca, calcium; TIBC, total iron-binding capacity; BUN, blood urea nitrogen; PFOS, perfluorooctanesulfonate; Fe, iron; AST, aspartate aminotransferase; BUN, blood urea nitrogen. * p < 0.05.
Table 4. Biochemical markers associated with serum albumin, Cr, and perfluorochemicals (PFCs) with linear regression to D-Jaffe and D-enzymatic post-dialysis.
Table 4. Biochemical markers associated with serum albumin, Cr, and perfluorochemicals (PFCs) with linear regression to D-Jaffe and D-enzymatic post-dialysis.
n = 124
Post-HD
Mean±SDD-JaffeD-Enzymatic
(β)(p)(β)(p)
Albumin (gm/dL)4.120.650.6610.2780.5750.349
Total protein (gm/dL)7.171.270.0060.838−0.0070.811
Jaffe Cr (mg/dL) *3.041.04−0.0960.000 *−0.2060.000 *
Enzymatic Cr (mg/dL) *2.580.94−0.1460.000 *−0.2080.000 *
IDLCMS Cr (mg/dL) *3.261.17−0.1260.002 *−0.0710.000 *
BUN (mg/dL) *15.865.66−0.1070.048 *−0.0290.000 *
P (mg/dL) *1.850.61−0.1090.009−0.2080.000 *
K (mEq/L)3.861.040.0040.909−0.0240.522
PFOA (ng/mL)0.530.340.0120.927−0.1510.202
PFOS (ng/mL)2.001.130.0000.9930.0220.552
SD, standard deviation; D-Jaffe, Jaffe Cr−IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr; Cr, creatinine; IDLCMS, isotope dilution–liquid chromatography-–mass spectrometry; BUN, blood urea nitrogen; P, phosphorus; K, potassium; PFOA, perfluorooctanoic acid; PFOS, perfluorooctanesulfonate. * p < 0.05.
Table 5. Serum PFOS concentration in patients with different dialysis membranes pre- and post-dialysis and linear regression of PFOS to D-Jaffe and D-enzymatic.
Table 5. Serum PFOS concentration in patients with different dialysis membranes pre- and post-dialysis and linear regression of PFOS to D-Jaffe and D-enzymatic.
Dialysis MembranePFOS Concentration
(ng/mL)
Linear Regression
PFOS vs. D-Jaffe
Linear Regression
PFOS vs. D-Enzymatic
β (p)β (p)
Pre-HD All (n = 124)5.50 (1.17,24.7)−0.002 (0.307)−0.005 (0.048) *
PS (n = 93)2.37 (1.17, 15.47)−0.002 (0.247)−0.006 (0.035) *
Non-PS (n = 31)23.89 (5.93,55.53)0.103 (0.485)−0.002 (0.687)
t-testp = 0.026 *
Post-HD All (n = 124)2.00 ± 1.130.000 (0.993)0.022 (0.552)
PS (n = 93)2.08 ± 1.22−0.046 (0.430)0.008 (0.870)
Non-PS (n = 31)
t-test
1.74 ± 0.69
p = 0.055
0.082 (0.476)0.152 (0.180)
PFOS, perfluorooctanesulfonate; D-Jaffe, Jaffe Cr−IDLCMS Cr; D-enzymatic, enzymatic Cr−IDLCMS Cr; HD, hemodialysis; PS: polysulfone. Pre-dialysis PFOS were not normally distributed and displayed as quartile. * p < 0.05.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Liu, W.-S.; Lin, C.-H.; Li, S.-Y.; Lin, C.-C.; Liu, T.-Y.; Tan, A.C.; Tsou, H.-H.; Chan, H.-L.; Lai, Y.-T. Perfluorooctanesulfonate Can Cause Negative Bias in Creatinine Measurement in Hemodialysis Patients Using Polysulfone Dialysis Membranes. Membranes 2022, 12, 778. https://doi.org/10.3390/membranes12080778

AMA Style

Liu W-S, Lin C-H, Li S-Y, Lin C-C, Liu T-Y, Tan AC, Tsou H-H, Chan H-L, Lai Y-T. Perfluorooctanesulfonate Can Cause Negative Bias in Creatinine Measurement in Hemodialysis Patients Using Polysulfone Dialysis Membranes. Membranes. 2022; 12(8):778. https://doi.org/10.3390/membranes12080778

Chicago/Turabian Style

Liu, Wen-Sheng, Chien-Hung Lin, Szu-Yuan Li, Chih-Ching Lin, Tsung-Yun Liu, Ann Charis Tan, Han-Hsing Tsou, Hsiang-Lin Chan, and Yen-Ting Lai. 2022. "Perfluorooctanesulfonate Can Cause Negative Bias in Creatinine Measurement in Hemodialysis Patients Using Polysulfone Dialysis Membranes" Membranes 12, no. 8: 778. https://doi.org/10.3390/membranes12080778

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.