Plasticizer Effects in the PVC Membrane of the Dibasic Phosphate Selective Electrode
Abstract
:1. Introduction
2. Experimental Section
2.1. Reagents
2.2. Sensor Fabrication
2.3. Calculation of Lipophilicity
2.4. Determination of Sensor Sensitivity
# | Added PO4 µL | Added PO4 mmol/L | Tot Vol µL | TOT PO4 mmol/L | {H2PO4−} mol/L | {HPO4−2} mol/L | log{H2PO4−} | log{HPO42−} |
---|---|---|---|---|---|---|---|---|
1 | 0 | 0 | 100,000 | 1.00E−09 | 6.62E−13 | 6.77E−13 | −12.179 | −12.169 |
2 | 5 | 100 | 100,005 | 5.00E−03 | 2.45E−06 | 2.51E−06 | −5.611 | −5.601 |
3 | 5 | 100 | 100,010 | 1.00E−02 | 4.88E−06 | 4.99E−06 | −5.312 | −5.302 |
4 | 10 | 100 | 100,020 | 2.00E−02 | 9.70E−06 | 9.93E−06 | −5.013 | −5.003 |
5 | 20 | 100 | 100,040 | 4.00E−02 | 1.92E−05 | 1.97E−05 | −4.716 | −4.706 |
6 | 40 | 100 | 100,080 | 7.99E−02 | 3.81E−05 | 3.89E−05 | −4.420 | −4.410 |
7 | 80 | 100 | 100,160 | 1.60E−01 | 7.49E−05 | 7.66E−05 | −4.126 | −4.116 |
8 | 16 | 1000 | 100,176 | 3.19E−01 | 1.47E−04 | 1.50E−04 | −3.834 | −3.824 |
9 | 32 | 1000 | 100,208 | 6.39E−01 | 2.84E−04 | 2.91E−04 | −3.546 | −3.536 |
10 | 64 | 1000 | 100,272 | 1.28E+00 | 5.45E−04 | 5.57E−04 | −3.264 | −3.254 |
11 | 128 | 1000 | 100,400 | 2.55E+00 | 1.03E−03 | 1.05E−03 | −2.989 | −2.979 |
12 | 256 | 1000 | 100,656 | 5.09E+00 | 1.89E−03 | 1.93E−03 | −2.724 | −2.714 |
13 | 512 | 1000 | 101,168 | 1.01E+01 | 3.38E−03 | 3.46E−03 | −2.470 | −2.460 |
14 | 1000 | 1000 | 102,168 | 1.98E+01 | 5.82E−03 | 5.95E−03 | −2.235 | −2.225 |
15 | 2500 | 1000 | 104,668 | 4.32E+01 | 1.07E−02 | 1.09E−02 | −1.973 | −1.963 |
2.5. Determination of Selectivity
2.6. Useable Sensor Life Span (L50) Evaluation
3. Results and Discussion
3.1. Sensor Sensitivity
Plasticizer | logKlipo | Limit of Detection −log{HPO42−} mol/L | Linear Sensitivity Conc. Decades | Linear Slope mV/Conc. Decade | |||
---|---|---|---|---|---|---|---|
Average | Std. Dev | Average | Std. Dev | Average | Std. Dev | ||
DMF | −0.6 | −5.13 | 0.38 | 2.60 | 0.00 | −7.4 | 0.9 |
NPOE | 2.3 | −5.05 | 0.20 | 2.34 | 0.33 | −8.8 | 4.9 |
DBP | 3.6 | −7.01 | 0.18 | 4.85 | 0.21 | −23.8 | 0.6 |
TEHP | 4.4 | −5.56 | 0.92 | 2.70 | 0.46 | −6.7 | 11.6 |
DBS | 4.6 | −5.00 | 0.51 | 2.78 | 0.42 | −9.3 | 2.2 |
BEHS | 6.9 | −5.05 | 0.26 | 3.18 | 1.57 | −16.0 | 4.4 |
3.2. Sensor Selectivity
Plast. | logKLipo | Cl− | NO3− | I− | F− | Cit3− | SO42− | SCN− | Ace− | Lac− | BTP− | B-T− | Br− | HEPES |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
DMF | −0.6 | 1.88 | 1.08 | 1.16 | 0.55 | −0.75 | 0.07 | 0.59 | 0.87 | 0.84 | 0.35 | 0.98 | 0.46 | 1.11 |
NPOE | 2.3 | 2.46 | 1.63 | 2.30 | 1.17 | −0.55 | 0.56 | 1.19 | 1.63 | 1.65 | 0.98 | 1.65 | 1.19 | 1.76 |
DBP | 3.6 | 3.41 | 2.42 | 2.00 | 2.87 | −0.54 | 0.93 | 2.00 | 3.41 | 3.19 | 1.11 | 3.35 | 0.33 | 2.95 |
TEHP | 4.4 | 2.74 | 2.33 | 2.24 | 1.12 | −0.58 | 0.51 | 1.26 | 1.70 | 1.79 | 1.35 | 2.09 | 0.98 | 1.65 |
DBS | 4.6 | 2.46 | 1.77 | 2.01 | 1.05 | −0.58 | 1.07 | 1.48 | 1.98 | 2.03 | 1.45 | 2.17 | 1.23 | 1.97 |
BEHS | 6.9 | 2.49 | 2.31 | 2.38 | 1.13 | −0.57 | 0.52 | 1.40 | 1.91 | 1.96 | 1.41 | 2.16 | 1.17 | 1.74 |
3.3. Useable Sensor Life Span (L50)
4. Conclusions
Acknowledgments
Conflicts of Interest
References
- Carey, C.M.; Riggan, W.B. Cyclic Polyamine Ionophore for Use in a Dibasic Phosphate Selective Electrode. Anal. Chem. 1994, 66, 3587–3591. [Google Scholar] [CrossRef] [PubMed]
- Perez, M.A.A.; Marin, L.P.; Quintana, J.C.; Yazdani-Pedram, M. Influence of different plasticizers on the response of chemical sensors based on polymeric membranes for nitrate ion determination. Sens. Act. 2003, 89B, 262–268. [Google Scholar] [CrossRef]
- Bedlechowicz, I.; Maj-Zurawska, M.; Sokalski, T.; Hulanicki, A. Effect of a plasticizer on the detection limit of calcium-selective electrodes. J. Electroanal. Chem. 2002, 537, 111–118. [Google Scholar] [CrossRef]
- Zhang, W.; Jenny, L.; Spichiger, U.E. A comparison of neutral Mg2+-selective ionophores in solvent polymeric membranes: Complex stoichiometry and lipophilicity. Anal. Sci. 2000, 16, 11–18. [Google Scholar] [CrossRef]
- Eugster, R.; Rosatzin, T.; Rusterholz, B.; Aebersold, B.; Pedrazza, U.; Ruegg, D.; Schmid, A.; Spichiger, U.E.; Simon, W. Plasticizers for liquid polymeric membranes of ion-selective chemical sensors. Anal. Chim. Acta 1994, 289, 1–13. [Google Scholar] [CrossRef]
- Dinten, O.; Spichiger, U.E.; Chaniotakis, N.; Gehrig, P.; Rusterholz, B.; Morf, W.E.; Simon, W. Lifetime of neutral-carrier-based liquid membranes in aqueous samples and blood and the lipophilicity of membrane components. Anal. Chem. 1991, 63, 596–603. [Google Scholar] [CrossRef] [PubMed]
- Van de Waterbeemd, H.; Smith, D.A.; Jones, B.C. Lipophilicity in PK design: Methyl, ethyl, futile. J. Comput.-Aided Mol. Des. 2001, 15, 273–286. [Google Scholar] [CrossRef] [PubMed]
- Buck, R.P.; Lindner, E. Recommendations for nomenclature of ion-selective electrodes (IUPAC Recommendations 1994). Pure Appl. Chem. 1994, 66, 2527–2536. [Google Scholar] [CrossRef]
- Davies, C.W. Ion Association; Butterworth’s: Washington, DC, USA, 1962; Chapter 3; p. 41. [Google Scholar]
- Attiyat, A.S.; Christian, G.D. Effect of interfering ion concentration and method of calculation on potentiometric selectivity coefficients of crown ether based liquid membrane ion selective electrodes. Anal. Sci. 1988, 4, 13–16. [Google Scholar] [CrossRef]
© 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/4.0/).
Share and Cite
Carey, C. Plasticizer Effects in the PVC Membrane of the Dibasic Phosphate Selective Electrode. Chemosensors 2015, 3, 284-294. https://doi.org/10.3390/chemosensors3040284
Carey C. Plasticizer Effects in the PVC Membrane of the Dibasic Phosphate Selective Electrode. Chemosensors. 2015; 3(4):284-294. https://doi.org/10.3390/chemosensors3040284
Chicago/Turabian StyleCarey, Clifton. 2015. "Plasticizer Effects in the PVC Membrane of the Dibasic Phosphate Selective Electrode" Chemosensors 3, no. 4: 284-294. https://doi.org/10.3390/chemosensors3040284