Next Article in Journal
Pre-Chirp-Managed Adiabatic Soliton Compression in Pressure-Gradient Hollow-Core Fibers
Next Article in Special Issue
Segmentation and Quantitative Analysis of Photoacoustic Imaging: A Review
Previous Article in Journal
Evaluation of Three Iterative Algorithms for Phase Modulation Regarding Their Application in Concentrating Light Inside Biological Tissues for Laser Induced Photothermal Therapy
Previous Article in Special Issue
High-Sensitivity Optical-Resolution Photoacoustic Microscopy with an Optical-Acoustic Combiner Based on an Off-Axis Parabolic Acoustic Mirror
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Fast Correction of “Finite Aperture Effect” in Photoacoustic Tomography Based on Spatial Impulse Response

Department of Biomedical Engineering, School of Basic Medical Science, Central South University, Changsha 410083, China
*
Author to whom correspondence should be addressed.
Photonics 2021, 8(9), 356; https://doi.org/10.3390/photonics8090356
Submission received: 4 August 2021 / Revised: 19 August 2021 / Accepted: 24 August 2021 / Published: 27 August 2021
(This article belongs to the Special Issue Photoacoustic Imaging for Biomedical Applications)

Abstract

:
Photoacoustic computed tomography (PACT) is a fast-developing imaging technique, which can provide structural and functional information in biological tissues with high-resolution beyond the depth of the optical diffusion limit. However, the most current PACT reconstruction method generally employs a point detector assumption, whereas in most PAT systems with circular or spherical scanning modes, the transducer is mostly flat and with a finite size. This model mismatch leads to a notable deterioration in the lateral direction in regions far from the rotation center, which is known as the “finite aperture effect”. In this work, we propose to compensate a novel Back-projection (BP) method based on the transducer’s spatial impulse response (SIR) for fast correction of the “finite aperture effect”. The SIR accounts for the waveform change of the transducer for an arbitrary point source due to the geometry of the detection surface. Simulation results showed that the proposed SIR-BP method can effectively improve the lateral resolution and signal to noise ratio (SNR) in the off-center regions. For a target 4.5 mm far from the rotation center, this new method improved the lateral resolution about five times along with a 7 dB increase in the SNR. Experimental results also showed that this SIR-BP method can well restore the image angular blur to recover small structures, as demonstrated by the imaging of leaf veins. This new method offers a valuable alternative to the conventional BP method, and can guide the design of PAT systems based on circular/spherical scan.

1. Introduction

Photoacoustic computed tomography (PACT) combines the merits of high ultrasound penetration and high optical contrast in a single modality, and is capable of high-resolution structural and functional imaging in biological tissue beyond the depth of optical diffusion limit [1,2,3]. To date, it has been widely employed in a variety of biomedical applications [3,4,5], such as brain vascular network visualization [6], small animal whole body imaging [7], breast cancer detection [8], and finger joint imaging [9].
In PACT, wide illumination over the whole sample is used, and ultrasound signals are collected by scanning the transducer around the sample or using transducer arrays, assuming that each transducer element covers the whole image domain. Then, an image reconstruction method is utilized to restore the light absorption distribution of the whole region at once [6,9,10,11].
The scanning geometry of PACT can be linear, circular, and spherical, or the combinations of the above geometries. Compared with linear scanning [11], the circular (for 2D imaging) [9] and spherical (for 3D imaging) [10] scanning can provide better imaging quality for covering a more complete angle of views of the target. Besides the optimization of the scanning geometry, a fast and accurate image reconstruction algorithm is the key to improve the quality of the reconstructed images [12].
Among various PACT reconstruction algorithms, the back-projection (BP) method is mostly employed due to its good stability and low computation cost [13,14,15]. However, most of the current PAT reconstruction algorithms including the BP method simply model the ultrasound transducer as point detectors, which however are usually planar and can reach several millimeters in size in circular or spherical scanning [16,17]. This geometry miss-model of the transducer detection surface in the reconstruction algorithm leads to the “finite aperture effect” [17,18,19]. Due to this effect, in PACT based on circular/spherical scanning, when the target is far away from the rotation center, the lateral resolution will deteriorate rapidly. Theoretical analysis shows that the lateral resolution is proportional to the distance from the rotation center, and equal to the transducer aperture size on the transducer detection surface [19].
To date, several methods have been proposed to improve this elongated lateral resolution. A large number of researchers employ focus transducers in circular-scanning-based PAT with a virtual-point-detector method [20,21], but these methods are reported to be unable to reliably reconstruct the high frequency part of the target [22]. Model-based method have been proposed to incorporate the transducer geometries to overcome the “finite aperture effect”, but the computational cost is too high to be applied on routing practice imaging. Deconvolution as an image post-processing method can also potentially improve the lateral resolution, but as a typical inverse method, it can induce strong image noises [23,24,25,26]. Therefore, new PACT reconstruction algorithms are still needed to overcome the influence of the finite transducer sized in circular/spherical-scanning-based PACT.
Essentially, the decrease of the lateral resolution is due to the spatial averaging of the photoacoustic signal over the transducer aperture, which makes the spatial impulse response (SIR) of a finite-sized transducer different from an ideal point detector [27]. Therefore, in this work we propose a novel BP method that is based on the fast calculation of SIR related parameters. Here, the lateral resolution improvement with this novel SIR based BP (SIR-BP) method was demonstrated with both simulations and phantom experiments in 2D circular-scanning-based PAT. Results show that the computational time of this SIR-BP method was as short as the conventional BP method.

2. Materials and Methods

2.1. Reconstruction Algorithms

Figure 1 shows the schematic of a conventional 2D circular PAT scan. Aerial illumination is given on the target with an infinitely short, pulsed laser. The generated photoacoustic signal p ( i , t ) is collected at a total of N d transducer positions around the target. If the acoustic velocity is assumed to be homogenous, there is analytical solutions to the optical absorption distribution H ( r ) in the image domain, which can be given as the following universal BP equation by Xu and Wang [14],
H ( r ) = i = 1 N d w i r R i × t p ( i , Δ t ) = i = 1 N d w i r R i × t p ( i , r R i / v )
Here, p ( i , t ) is the received acoustic signal by the i -th transducer at R i . The coefficient w i is to weight the contributions from different transducers based on their angular coverage with respect to the target.
The above BP methods models the transducer as an ideal point detector. However, in most current PAT systems the transducer aperture is mostly planar, and the aperture size can be of several millimeters. If considering the effect of the transducer aperture, the actual received photoacoustic pressure p ( R , t ) at the transducer position R should be written as the follows integral over the transducer detection surface S
p ( R , t ) = S p ( r , t ) d S ( r )     = Γ 4 π v t S [ v H ( r ) δ ( t r r / v ) r r d r ] d S ( r )   = Γ 4 π v t v H ( r ) S I R ( r R , t ) d r
Here, r is the position of the photoacoustic source, and r is the position on the transducer. Therefore,
S I R ( r R , t ) = S δ ( t r r / v ) r r d S ( r )
It can be seen that p ( R , t ) is greatly affected by the SIR due to the spatial integral over the transducer detection surface. Thus, the analytical solution in (1) becomes invalid in this case. The result is that the image will be blurred in the regions far from the rotation center due to an elongated lateral resolution, which is also known as the “finite aperture effect”. From a deeper point of view, the SIR determines the flight time from the acoustic point source (or the time delay), as well as the spatial-dependent sensitivity. Because these two factors are different in the SIR of a finite-sized transducer and the SIR of an ideal point detector, they must be reflected in the image reconstruction method.
Therefore, we propose to replace the time delay and the weighting factor in (1) with SIR related expressions, which is the basis of our proposed SIR-BP method in this work. Firstly, in the conventional BP method, the time delay is expressed as r R i / v . In the SIR-BP method, considering the finite transducer aperture, the time delay is a function of the relative distance from the transducer to the acoustic source r R i , which can be termed to be Δ t _ S I R ( r R i ) . Thus, (1) becomes
H ( r ) = i = 1 N d Δ t _ S I R ( r R i ) × t p i ( Δ t _ S I R ( r R i ) )
Here, the weighting factor w i is not considered and we assume signals collected at different transducer positions give equal contributions. We call this kind of time delay compensation-based BP method to be the TDC-BP method.
Furthermore, in the conventional BP method given by (1), the weighting factor w i is set to be constant. However, because the acoustic field or the sensitivity of the finite-sized transducer is spatially dependent, and it is also a function of the vector r R i , we termed the weighting factor to be w ( r R i ) . Then, the SIR-BP method obtained from (4) can be written as follows
H ( r ) = i = 1 N d w ( r R i ) Δ t _ S I R ( r R i ) × t p ( i , Δ t _ S I R ( r R i ) )
Here, the weighting factor is set to be the reciprocal of the amplitude of the SIR waveform I _ S I R ( r ) , as in (6)
w ( r ) = 1 I _ S I R ( r )

2.2. SIR-Related Time-Delay and Weighting Factor Expressions

Next, we have to find the expressions based on the SIR for Δ t _ S I R ( r R i ) and I _ S I R ( r R i ) . As in (3), the SIR is calculated to be the integral over the transducer’s detection surface. Because most planar transducers’ detection surface is circular, the SIR can be obtained analytically in the cylindrical coordination due to the symmetry.
As shown in Figure 2, we set the detection surface of the transducer in the z = 0 plane and take its center to be the origin of the coordinates. The detection surface of the transducer has a radius of a , and the laser pulse width is sufficiently small. Then, for an observation point r ( r , θ , z ) , the SIR is
S I R ( r , θ , z , t ) = S I R ( r , z , t ) = 0 2 π 0 a δ ( t ( R cos φ r ) 2 + ( R sin φ ) 2 + z 2 / v ) 2 π ( R cos φ r ) 2 + ( R sin φ ) 2 + z 2 R d R d φ
There are analytical expressions of the integral of (7), which can be derived in two different cases depending on whether the acoustic point source is within the aperture of the disk in the x-y plane or not [28], as shown in (8)
for   r a S I R ( r , z , t ) = 0 t z v v z v < t < t 1 v π cos 1 ( ( t v ) 2 z 2 + r 2 a 2 2 r ( t v ) 2 z 2 ) , t 1 < t < t 2 0 t > t 2 for   r > a S I R ( r , z , t ) = 0 t t 1 v π cos 1 ( ( t v ) 2 z 2 + r 2 a 2 2 r ( t v ) 2 z 2 ) , t 1 < t t 2 0 t > t 2
Here, t1 and t2 correspond to the flight time from the edges of the disk closest and farthest to the photoacoustic source position r ( r , θ , z ) respectively, so that
t 1 = z 2 + ( a r ) 2 / v t 2 = z 2 + ( a + r ) 2 / v
In this work, we take the peak arrival time of the SIR waveform to be Δ t _ S I R ( r ) . From (8), we can see that the analytical expression of this time delay can also be divided into two cases:
Δ t _ S I R ( r ) = z v r a t 1 r > a
Additionally, the analytical expression for the amplitude of the SIR waveform I _ S I R ( r R ) can be obtained from (10) after dropping the constant of v , so that
I _ S I R ( r ) = 1 r a 1 π cos 1 ( 1 a 2 r 2 ) r > a
The final reconstruction results were given by taking (10) and (11) into (7) in SIR-BP, or just taking (10) into (6) in TDC-BP.

2.3. Numerical Simulations

Point targets imaging simulations were carried out. A planar transducer was employed, which had a bandwidth of 70% and a diameter of 5 mm, and the distance from the transducer to the rotation center was 25 mm. There were 720 detectors in the full scan with an angular interval of 0.5 degree. In these simulations, different transducer central frequencies was tested, which were 3 MHz, 5 MHz, 10 MHz, and 20 MHz, respectively, and the sampling frequency was 100MHz. This was to demonstrate the adaptability of our proposed SIR-BP method to different transducer frequencies. The simulated transducer signal p d ( R , t ) was generated by convoluting the acoustic pressure at the transducer detection surface p ( R , t ) given by (4) with the transducer’s electrical response function he(t), as in (12).
p d ( R , t ) = h e ( t ) p ( R , t ) = Γ 4 π v t = t = + v H ( r ) t [ S I R ( r R , t ) ] h e ( t t ) d r d t = Γ 4 π v v H ( r ) S I R ( r R , t ) h e ( t t ) d r t = t = + + Γ 4 π v t = t = + v H ( r ) S I R ( r R , t ) t h e ( t t ) d r d t = ( t h e ( t ) ) Γ 4 π v v H ( r ) [ S I R ( r R , t ) ] d r
Here, we utilized that he(t) equals to zero at t = and t = + . We define h 0 ( t ) = ( / t ) h e ( t ) to be the impulse function of the system, which can be experimentally obtained by measuring the photoacoustic signal of an infinite small point target. In this work, its expression was given by a Gaussian enveloped sine function as in (13).
h 0 ( t ) = sin ( 2 π f 0 t ) [ sin ( 2 π f 0 t ) e t 2 2 δ 2 ]
Here, f 0 represents the central frequency of the transducer, and δ is the pulse width of this impulse function, which is determined by the bandwidth of the transducer.
For noise assessment, an ensemble of 1000 trials were applied. In each trial, the Box-Muller method was employed to add Gaussian white noise to the data, with the standard deviation to be 5 % of the signal amplitude of the point target at the rotation center. The mean value of the 1000 reconstructed amplitudes for each target was taken as the signal As, and the standard deviation of these reconstructed amplitudes was taken as the noise An, and the signal to noise ratio (SNR) was calculated to be
S N R = 20 log 10 ( A s A n )

2.4. Phantom Experiments

A conventional 2D PAT system was built to test our method, whose schematic is shown in Figure 3. Pulsed 740 nm laser from an optical parametric oscillator (OPO) laser (SpitLight 600 OPO-532 midband, Innolas, München, Germany) was directed to give illumination of the target. The optical intensity on the top of the phantom was about 5 mJ/cm2, and the repetition rate was 20 Hz. The total step number of the 2D scan was 360, with an angular step 1 degree. The employed planar transducer (Zunen Industries, Shanghai, China) had a central frequency of 5 MHz, its bandwidth was 80%, and its active element was 3 mm in diameter. The distance between the rotation center to the transducer detection surface was about 20 mm. The signal was first amplified by the a pulser/receiver (DPR500, Ultrasonics California, San Juan Capistrano, CA, USA), and then digitized with an acquisition card (NI-5124, 12 bit, 100 MHz sampling frequency) in the computer. The whole system was synchronized by the laser, and the data were collected on the hard disk for later processing.
Two phantoms were imaged. The phantoms were made of 2% ager, water and intralipid, which had a diameter of 3 cm and a scattering coefficient of 1 /mm. The first phantom had 7 pencil leads (0.5 mm thick) vertically inserted as point targets, and the second phantom had a piece of leaf vein (about 8 mm in size) embedded on its surface.

3. Results

Figure 4 presents the reconstruction results of the simulated data with 5 MHz central frequency and 5 % noise. There were four point-targets evenly distributed on the x-axis from 0 to 4.5 mm. The image domain was −1 mm < x < 5.5 mm, −1.2 mm < y < 1.2 mm. Here, Figure 4a–c are reconstructed with the conventional BP method (with (3)), the TDC-BP method (with (6)), and the SIR-BP method (with (7)), respectively. Because the transducers were distributed uniformly, the weighting factor w i in (3) was set to be one for all the detectors. The cross sections of the four targets in Figure 4a–c were also extracted and compared in Figure 4d–g. The full width at half height (FWHM) for all these lateral profiles of the point targets are obtained as lateral resolutions, which are listed in Table 1 along with the SNRs of the four targets.
It can be seen from Figure 4 that when the target is away from the center of rotation, the lateral profile of the target becomes wider. For the point target at x = 4.5 mm, it’s lateral resolution was 0.654mm, which was about five times the target of the rotation center (0.136 mm). For the point target at x = 4.5 mm, the lateral resolutions given by the TDC-BP method and the SIR-BP method were 0.204 mm and 0.138 mm, which were about 3.2 and 4.7 times smaller than that given by the conventional BP method.
For SNR comparison and noise assessment, it’s noted that in Figure 4a,b the background noise is evenly distributed in the image domain. However, the background noise in Figure 4c is stronger for regions far from the rotation center than the regions near the rotation center. From Table 1, for all the three off-center targets, both the TDC-BP method and the SIR-BP method give higher SNRs than the conventional BP method. For the target at x = 4.5 mm, the SNR with the conventional BP method is 24.863 dB, but with the TDC-BP method and the SIR-BP method, the SNR improves to 33.311 dB and 27.562 dB, respectively. However, it’s noted that the SNR given by the SIR-BP method is lower than that by the TDC-BP method in the off-center regions. This is due to the fact that the transducer’s sensitivity in the off-center regions is low, so that the background noise in these regions is increased when inverting the SIR amplitude to obtain the weighting factor with (8).
Figure 5 shows the reconstruction results for different central frequencies, as to further explore the adaptability of the proposed method. Here, different rows represent results with different central frequencies, and different columns are reconstructed with different methods. Table 2 lists the SNRs and FWHMs for the most right-side target at x = 4.5 mm. From the first column, it’s clearly noted that the effect of finite transducer aperture is frequency dependent. For lower frequencies such as 3 MHz and 5 MHz, due to the larger divergence angle of the transducer, the lateral blur is not serious for the target at x = 1.5 mm, which however is more notable for higher frequencies. With the time delay compensation by the TDC-BP method, the lateral blur for the off-center targets is well restored for all the frequencies, and the reconstructed amplitudes of these targets are also increased, which is reflected in the improved SNRs, as in Table 2. With the SIR-BP method, the lateral resolutions and the reconstructed amplitudes of the off-center targets are also further increased. However, due to the concurrent increased background noise, the SNRs are not as good as those given by the TDC-BP method, but are still better than those by the conventional BP method. As seen in Table 2, for the target at x = 4.5 mm, the SIR-BP method improves the lateral resolution four to nine times, along with a 3 to 7 dB improvement in the SNR, as compared with the conventional BP method.
The first phantom experiment was to test the improvement of lateral resolution with our proposed method. There were seven pencil leads vertically inserted in the phantom as point targets, as seen in Figure 6a, and the reconstructed images with the conventional BP method, the TDC-BP method, and the SIR-BP method are shown in Figure 6b–d, respectively. Compared with the photo image, it’s seen that all three images can reliably reveal the locations of the seven targets. However, in Figure 6b the five targets located far from the rotation center are distorted in the lateral directions due to the effect of the finite transducer aperture. With the time delay compensation, the lateral resolutions of these targets are well restored, as seen in Figure 6c. When directional sensitivity compensation is applied, the lateral blur of these five targets is further reduced in Figure 6d.
Comparing the obtained FWHMs from the three curves of about 0.70 mm, 0.25 mm and 0.21 mm, it can be seen that the improvement of the FWHM with the TDC-BP and SIR-BP methods were about 2.8 times and 3.3 times, respectively, with respect to the conventional BP method. It’s also noted that the SNRs for Figure 7d–f were 19.09 dB, 22.78 dB and 19.23 dB, respectively. These results are in good agreement with the conclusions we achieved in the simulation studies.
As seen in Table 3, in simulation results for the target at x = 4.5 mm, the SIR-BP method improves the lateral resolution four to nine times as compared with the conventional BP method. In phantom experiment result for the target at about 4 mm away from the rotation center, the SIR-BP method improves the lateral resolution 3.3 times as compared with the conventional BP method.
From Figure 8 we can see the complex reconstruction imaging results of a piece of leaf veins. The size of the leaf veins was about 8 mm. It is noticed that many small structures around the margins of the target were seriously distorted and missed in Figure 8b. One example is the small branch as indicated with arrow 1. After applying the time delay obtained from the SIR of the transducer, it can be well reconstructed, as seen in Figure 8c,d. Furthermore, it can be seen that many veins along the axial direction in the dotted rectangle in Figure 8a were missed in Figure 8b,c, whereas it can be better revealed in Figure 8d with the SIR-BP method. Another example is the small branch pointed with arrow 2, which is less clear in Figure 8c than that in Figure 8d. This proves the advantage of the SIR-BP method in reconstructing complex objects over the conventional BP method and the TDC-BP method. In Figure 8d, there is more background noise, and this was because we were trying to suppress the contributions from the in-axial signals while boosting the contributions from the off-axial signals, which makes the signal amplitude low and noise level higher.
In terms of the computational time, because the three methods shared the same BP structure, the reconstruction time cost was similar. For the reconstruction of Figure 8, which had 601 × 601 pixels and 360 detectors, the time cost with the conventional BP method, the TDC-BP method and the SIR-BP method were 15.2 s, 18.3 s, and 20.2 s, respectively, running on a single CPU (1.80 GHz) on MATLAB.

4. Discussion

The “finite aperture effect” is one of the biggest problem that hinders the acquisition of high-quality images in circular/spherical-scanning-based PACT. The purpose of this paper is to find a fast and stable image reconstruction method for tackling this problem. Because the degraded lateral resolution is largely due to the time delay error and spatial-dependent sensitivity introduced by the SIR of the finite transducer aperture, this work proposes to calculate and compensate for these two factors in the framework of the BP method. Therefore, compared with existing model-based algorithms [18,27,28,29] for this “finite aperture effect” problem, our proposed method inherits the simplicity, stability and low computation cost from the BP method [30,31], so that it is more practical.
The proposed method was validated with both numerical simulations and phantom experiments. As seen in the simulation study, by correcting the time delay in the BP equation with the SIR, both the lateral resolution and SNR in the off-center regions are effectively improved. By further compensating the sensitivity loss in the off-center regions with the SIR-BP method, simulation results indicate that the elongated lateral resolution can almost be completely restored.
It is noted that our proposed method was tested in simulations in the frequency range from 3 MHz to 20 MHz, which covers the common frequency range of existing PAT systems. Results showed that the proposed method showed consistent improvement of the lateral resolution and the SNRs over the conventional BP method, which proves its adaptability and effectiveness. The stability of the proposed method has also been proved with simulated data with 5% noise and the experimental data. In addition, the SIR-BP method is almost as fast as the conventional BP method and the TDC-BP method in terms of run time. All these indicates that the SIR-BP method is a better alternative to the conventional BP method, and can be used for guidance of system designs.
The proposed method has been successfully tested with two phantom experiments. The first experiment validated the improvement in lateral resolution and SNR with the proposed method. The second experiment was to explore the potential of the proposed method for imaging complicated targets. Results showed the proposed method can well recover the blurred structures in the off-center regions, especially for small features along the axial directions. In short, the proposed SIR-BP method can comprehensively improve the image fidelity. In future works, we will accelerate the proposed method with GPU based parallel techniques [32], and extend it to the spherical-scanning-based PAT for 3D imaging.

5. Conclusions

In this work, a novel SIR-BP algorithm was proposed to counter the “finite aperture effect” in 2D circular-scanning-based PAT. By applying accurate time delay and compensating for the spatial-dependent transducer sensitivity in the BP method obtained from the SIR of the transducer, the proposed SIR-BP method can well restore the elongated lateral resolution, giving higher SNRs, and recovering subtitle image features in regions far from the rotation center, as compared with the conventional BP method. The proposed method is of high significance to the guidance of circular/spherical-scanning-based system designs.

Author Contributions

B.W. and X.L. proposed the algorithm and wrote the manuscript. X.L. designed and performed all the experiments, collected raw data, analyzed the data, and prepared the figures for the manuscript. C.W. analyzed the data and prepared the figures for the manuscript. B.W. and J.X. conceived and supervised the project, designed experiments, interpreted data. All authors contributed to the critical reading and writing of the manuscript. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the Fundamental Research Funds for the Central Universities of Central South University under grant 2019zzts324, 2020zzts784, the Emergency Science and Technology Project of Hunan Province under grant 2020YJ004, the Department of Science and Technology of Hunan Province, High-tech Industry Science and Technology Innovation Leading Program 2020SK2003 and the Innovation Driven Program team project of Central South University 2020CX004.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data presented in this study are available on request from the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Wang, L.V.; Hu, S. Photoacoustic tomography: In vivo imaging from organelles to organs. Science 2012, 335, 1458–1462. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Wang, L.V.; Yao, J.J. A practical guide to photoacoustic tomography in the life sciences. Nat. Methods 2016, 13, 627–638. [Google Scholar] [CrossRef] [PubMed]
  3. Zhou, Y.; Liang, S.; Li, M.; Liu, C.; Lai, P.; Wang, L. Optical-resolution photoacoustic microscopy with ultrafast dual-wavelength excitation. J. Biophotonics 2020, 13, e201960229. [Google Scholar] [CrossRef] [PubMed]
  4. Park, E.Y.; Park, S.; Lee, H.; Kang, M.; Kim, C.; Kim, J. Simultaneous Dual-Modal Multispectral Photoacoustic and Ultrasound Macroscopy for Three-Dimensional Whole-Body Imaging of Small Animals. Photonics 2021, 8, 13. [Google Scholar] [CrossRef]
  5. Zhang, X.; Liu, Y.; Tao, C.; Yin, J.; Hu, Z.Z.; Yuan, S.T.; Liu, Q.H.; Liu, X.J. High-Sensitivity Optical-Resolution Photoacoustic Microscopy with an Optical-Acoustic Combiner Based on an Off-Axis Parabolic Acoustic Mirror. Photonics 2021, 8, 127. [Google Scholar] [CrossRef]
  6. Yang, X.M.; Maurudis, A.; Gamelin, J.; Aguirre, A.; Zhu, Q.; Wang, L.V. Photoacoustic tomography of small animal brain with a curved array transducer. J. Biomed. Opt. 2009, 14, 054007. [Google Scholar] [CrossRef] [Green Version]
  7. Xia, J.; Wang, L.V. Small-Animal Whole-Body Photoacoustic Tomography: A Review. IEEE Trans. Bio.-Med. Eng. 2014, 61, 1380–1389. [Google Scholar] [CrossRef] [Green Version]
  8. Menke, J. Photoacoustic breast tomography prototypes with reported human applications. Eur. Radiol. 2015, 25, 2205–2213. [Google Scholar] [CrossRef]
  9. Xiao, J.Y.; Yao, L.; Sun, Y.; Sobel, E.S.; He, J.S.; Jiang, H.B. Quantitative two-dimensional photoacoustic tomography of osteoarthritis in the finger joints. Opt. Express 2010, 18, 14359–14365. [Google Scholar] [CrossRef]
  10. Wang, B.; Xiao, J.Y.; Jiang, H.B. Simultaneous real-time 3D photoacoustic tomography and EEG for neurovascular coupling study in an animal model of epilepsy. J. Neural. Eng. 2014, 11, 046013. [Google Scholar] [CrossRef]
  11. Li, M.C.; Liu, C.B.; Gong, X.J.; Zheng, R.Q.; Bai, Y.Y.; Xing, M.Y.; Du, X.M.; Liu, X.Y.; Zeng, J.; Lin, R.Q.; et al. Linear array-based real-time photoacoustic imaging system with a compact coaxial excitation handheld probe for noninvasive sentinel lymph node mapping. Biomed. Opt. Express 2018, 9, 1408–1422. [Google Scholar] [CrossRef] [Green Version]
  12. Zhou, Y.; Yao, J.J.; Wang, L.H.V. Tutorial on photoacoustic tomography. J. Biomed. Opt. 2016, 21, 061007. [Google Scholar] [CrossRef] [Green Version]
  13. Burgholzer, P.; Bauer-Marschallinger, J.; Grun, H.; Haltmeier, M.; Paltauf, G. Temporal back-projection algorithms for photoacoustic tomography with integrating line detectors. Inverse Probl. 2007, 23, S65–S80. [Google Scholar] [CrossRef]
  14. Xu, M.H.; Wang, L.V. Universal back-projection algorithm for photoacoustic computed tomography. Phys. Rev. E 2005, 71, 059903. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. Huang, H.; Bustamante, G.; Peterson, R.; Ye, J.Y. An adaptive filtered back-projection for photoacoustic image reconstruction. Med. Phys. 2015, 42, 2169–2178. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  16. Burgholzer, P.; Matt, G.J.; Haltmeier, M.; Paltauf, G. Exact and approximative imaging methods for photoacoustic tomography using an arbitrary detection surface. Phys. Rev. E 2007, 75, 046706. [Google Scholar] [CrossRef]
  17. Roitner, H.; Haltmeier, M.; Nuster, R.; O’Leary, D.P.; Berer, T.; Paltauf, G.; Grun, H.; Burgholzer, P. Deblurring algorithms accounting for the finite detector size in photoacoustic tomography. J. Biomed. Opt. 2014, 19, 056011. [Google Scholar] [CrossRef]
  18. Li, M.L.; Tseng, Y.C.; Cheng, C.C. Model-based correction of finite aperture effect in photoacoustic tomography. Opt. Express 2010, 18, 26285–26292. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  19. Xu, M.H.; Wang, L.V. Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction. Phys. Rev. E 2003, 67, 056605. [Google Scholar] [CrossRef] [Green Version]
  20. Pramanik, M.; Ku, G.; Wang, L.V. Tangential resolution improvement in thermoacoustic and photoacoustic tomography using a negative acoustic lens. J. Biomed. Opt. 2009, 14, 024028. [Google Scholar] [CrossRef]
  21. Li, C.H.; Wang, L.V. High-numerical-aperture-based virtual point detectors for photoacoustic tomography. Appl. Phys. Lett. 2008, 93, 033902. [Google Scholar] [CrossRef] [PubMed]
  22. Drozdov, G.; Rosenthal, A. Analysis of Negatively Focused Ultrasound Detectors in Optoacoustic Tomography. IEEE Trans. Med. Imaging 2017, 36, 301–309. [Google Scholar] [CrossRef] [PubMed]
  23. Zhang, C.; Wang, Y.Y. Deconvolution reconstruction of full-view and limited-view photoacoustic tomography: A simulation study. J. Opt. Soc. Am. A 2008, 25, 2436–2443. [Google Scholar] [CrossRef] [PubMed]
  24. Prakash, J.; Raju, A.S.; Shaw, C.B.; Pramanik, M.; Yalavarthy, P.K. Basis pursuit deconvolution for improving model-based reconstructed images in photoacoustic tomography. Biomed. Opt. Express 2014, 5, 1363–1377. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  25. Cai, D.; Li, Z.F.; Li, Y.; Guo, Z.D.; Chen, S.L. Photoacoustic microscopy in vivo using synthetic-aperture focusing technique combined with three-dimensional deconvolution. Opt. Express 2017, 25, 1421–1434. [Google Scholar] [CrossRef]
  26. Chen, J.H.; Lin, R.Q.; Wang, H.N.; Meng, J.; Zheng, H.R.; Song, L. Blind-deconvolution optical-resolution photoacoustic microscopy in vivo. Opt. Express 2013, 21, 7316–7327. [Google Scholar] [CrossRef]
  27. Wang, K.; Ermilov, S.A.; Su, R.; Brecht, H.P.; Oraevsky, A.A.; Anastasio, M.A. An Imaging Model Incorporating Ultrasonic Transducer Properties for Three-Dimensional Optoacoustic Tomography. IEEE Trans. Med. Imaging 2011, 30, 203–214. [Google Scholar] [CrossRef] [Green Version]
  28. Piwakowski, B.; Delannoy, B. Method for computing spatial pulse response: Time-domain approach. J. Acoust. Soc. Am. 1989, 86, 2422–2432. [Google Scholar] [CrossRef]
  29. Rosenthal, A.; Ntziachristos, V.; Razansky, D. Acoustic Inversion in Optoacoustic Tomography: A Review. Curr. Med. Rev. Imaging 2013, 9, 318–336. [Google Scholar] [CrossRef] [Green Version]
  30. Poudel, J.; Lou, Y.; Anastasio, M.A. A survey of computational frameworks for solving the acoustic inverse problem in three-dimensional photoacoustic computed tomography. Phys. Med. Biol. 2019, 64, 14TR01. [Google Scholar] [CrossRef]
  31. Wang, K.; Su, R.; Oraevsky, A.A.; Anastasio, M.A. Investigation of iterative image reconstruction in three-dimensional optoacoustic tomography. Phys. Med. Biol. 2012, 57, 5399–5423. [Google Scholar] [CrossRef] [PubMed]
  32. Wang, K.; Huang, C.; Kao, Y.J.; Chou, C.Y.; Oraevsky, A.A.; Anastasio, M.A. Accelerating image reconstruction in three-dimensional optoacoustic tomography on graphics processing units. Med. Phys. 2013, 40, 023301. [Google Scholar] [CrossRef] [PubMed]
Figure 1. 2D circular-scanning-based PAT system.
Figure 1. 2D circular-scanning-based PAT system.
Photonics 08 00356 g001
Figure 2. The geometry and notations for the SIR calculation of a circular planar transducer.
Figure 2. The geometry and notations for the SIR calculation of a circular planar transducer.
Photonics 08 00356 g002
Figure 3. Schematic of the 2D PAT system.
Figure 3. Schematic of the 2D PAT system.
Photonics 08 00356 g003
Figure 4. Simulation results for point targets imaged with different reconstruction methods. (a) The conventional BP method; (b) the TDC-BP method; (c) the SIR-BP method. The diameter of the employed planar transducer was 5 mm, and the four targets are at 0 mm, 1.5 mm, 3 mm, and 4.5 mm, respectively. (dg) are the lateral profiles for the four targets obtained from (ac). The black, blue and red lines represent the results from the conventional BP method, TDC-BP method, and SIR-BP method, respectively. The simulated data were added with 5 % noise. The amplitudes of these lines are normalized to 1 for comparison.
Figure 4. Simulation results for point targets imaged with different reconstruction methods. (a) The conventional BP method; (b) the TDC-BP method; (c) the SIR-BP method. The diameter of the employed planar transducer was 5 mm, and the four targets are at 0 mm, 1.5 mm, 3 mm, and 4.5 mm, respectively. (dg) are the lateral profiles for the four targets obtained from (ac). The black, blue and red lines represent the results from the conventional BP method, TDC-BP method, and SIR-BP method, respectively. The simulated data were added with 5 % noise. The amplitudes of these lines are normalized to 1 for comparison.
Photonics 08 00356 g004
Figure 5. Simulation results with different reconstruction methods when the transducer central frequency varies. (ad) The results with the conventional BP method when the central frequencies are 3 MHz, 5 MHz, 10 MHz and 20 MHz, respectively. (eh) The corresponding results with the TDC-BP method. (il) The corresponding results with the SIR-BP method. The diameter of the employed planar transducer was 5 mm, and the locations of the four points are 0 mm, 1.5 mm, 3 mm, and 4.5 mm, respectively. The simulated data were added with 5 % noise. The amplitudes of all the images are normalized to 1 for comparison.
Figure 5. Simulation results with different reconstruction methods when the transducer central frequency varies. (ad) The results with the conventional BP method when the central frequencies are 3 MHz, 5 MHz, 10 MHz and 20 MHz, respectively. (eh) The corresponding results with the TDC-BP method. (il) The corresponding results with the SIR-BP method. The diameter of the employed planar transducer was 5 mm, and the locations of the four points are 0 mm, 1.5 mm, 3 mm, and 4.5 mm, respectively. The simulated data were added with 5 % noise. The amplitudes of all the images are normalized to 1 for comparison.
Photonics 08 00356 g005
Figure 6. Experiment results for imaging point targets with different reconstruction algorithms. (a) Photo of the phantom; (b) the conventional BP algorithm; (c) the TDC-BP algorithm; (d) The proposed SIR-BP algorithm. The amplitude of (bd) are all normalized to 1.
Figure 6. Experiment results for imaging point targets with different reconstruction algorithms. (a) Photo of the phantom; (b) the conventional BP algorithm; (c) the TDC-BP algorithm; (d) The proposed SIR-BP algorithm. The amplitude of (bd) are all normalized to 1.
Photonics 08 00356 g006
Figure 7. Enlarged views of the reconstructed images with different BP methods for one representative point target. (a) The conventional BP method; (b) the TDC-BP algorithm; (c) The SIR-BP algorithm. (df) The extracted lateral profiles correspondingly in (ac).
Figure 7. Enlarged views of the reconstructed images with different BP methods for one representative point target. (a) The conventional BP method; (b) the TDC-BP algorithm; (c) The SIR-BP algorithm. (df) The extracted lateral profiles correspondingly in (ac).
Photonics 08 00356 g007
Figure 8. Experiment results for imaging a piece of leaf veins with different reconstruction methods. (a) Photo image of the leaf veins; (bd) are the reconstructed images with the conventional BP algorithm, the TDC-BP algorithm and the SIR-BP algorithm, respectively.
Figure 8. Experiment results for imaging a piece of leaf veins with different reconstruction methods. (a) Photo image of the leaf veins; (bd) are the reconstructed images with the conventional BP algorithm, the TDC-BP algorithm and the SIR-BP algorithm, respectively.
Photonics 08 00356 g008
Table 1. The calculated FWHMs and SNRs of the four targets with different reconstruction methods.
Table 1. The calculated FWHMs and SNRs of the four targets with different reconstruction methods.
Targets LocationParametersBPTDC-BPSIR-BP
0.0 mmFWHM (mm)0.1360.1360.136
SNR (dB)40.90740.90740.907
1.5 mmFWHM (mm)0.1700.1360.136
SNR (dB)37.90739.63739.637
3.0 mmFWHM (mm)0.4080.1620.136
SNR (dB)27.28136.73833.908
4.5 mmFWHM (mm)0.6540.2040.138
SNR (dB)24.86333.31127.562
Table 2. The calculated FWHMs and SNRs for the off-center target at 4.5mm reconstructed with different methods and different transducer frequencies.
Table 2. The calculated FWHMs and SNRs for the off-center target at 4.5mm reconstructed with different methods and different transducer frequencies.
Transducer
Frequency
ParametersEquation (3)Equation (6)Equation (7)
3 MHzFWHM (mm)0.8300.3080.215
SNR (dB)20.86029.80725.324
5 MHzFWHM (mm)0.6540.2040.138
SNR (dB)24.86333.31127.562
10 MHzFWHM (mm)0.5220.1070.073
SNR (dB)28.00939.95033.860
20 MHzFWHM (mm)0.3600.0580.040
SNR (dB)29.95944.06436.930
Table 3. The calculated FWHMs improved for the off-center target at x = 4.5 mm reconstructed with different methods and different transducer frequencies in simulation results and experiment result for the off-center target at about 4mm.
Table 3. The calculated FWHMs improved for the off-center target at x = 4.5 mm reconstructed with different methods and different transducer frequencies in simulation results and experiment result for the off-center target at about 4mm.
ResultsTDC-BPSIR_BP
3 MHz2.73.9
5 MHz3.24.7
10 MHz4.97.2
20 MHz6.29.0
experiment2.83.3
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Luo, X.; Xiao, J.; Wang, C.; Wang, B. Fast Correction of “Finite Aperture Effect” in Photoacoustic Tomography Based on Spatial Impulse Response. Photonics 2021, 8, 356. https://doi.org/10.3390/photonics8090356

AMA Style

Luo X, Xiao J, Wang C, Wang B. Fast Correction of “Finite Aperture Effect” in Photoacoustic Tomography Based on Spatial Impulse Response. Photonics. 2021; 8(9):356. https://doi.org/10.3390/photonics8090356

Chicago/Turabian Style

Luo, Xiaofei, Jiaying Xiao, Congcong Wang, and Bo Wang. 2021. "Fast Correction of “Finite Aperture Effect” in Photoacoustic Tomography Based on Spatial Impulse Response" Photonics 8, no. 9: 356. https://doi.org/10.3390/photonics8090356

APA Style

Luo, X., Xiao, J., Wang, C., & Wang, B. (2021). Fast Correction of “Finite Aperture Effect” in Photoacoustic Tomography Based on Spatial Impulse Response. Photonics, 8(9), 356. https://doi.org/10.3390/photonics8090356

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop