4.1. Equivalent Sphere Diameter (ESD)
The
ESD results for the relatively badly soluble air and well-soluble CO
2 bubbles show a very different behavior. An overview of the changes in the
ESD of air bubbles is presented in
Figure 7. Here, and in all following figures, the legend first displays the capillary diameter, then the gas flow rate is referred to as “g”, and the liquid flow rate as “l”. As expected, with larger capillary diameters, larger bubbles were generated at the same gas and liquid volume flow rate (
Figure 7a). The bubble diameters remained almost the same over the column height; just a slight growth of around 0.1 mm could be recognized in the investigated length (1 m). This bubble growth is more visible in
Figure 7b–d due to the changed scales and was induced by the decrease in the hydrostatic pressure inside the column.
With increasing counter-current liquid flow, the bubble size became bigger (
Figure 7b) along the complete column height. It is noticeable that, in the case of stagnant liquid, similar bubble sizes as those recorded at the highest 55.5 l·min
−1 counter-current liquid flow rate were found. This similarity was not observed with the 3.6 mm and 0.13 mm capillaries. This suggests that the bubble-forming mechanism at the nozzles was not the same due to different materials being used (here Teflon, see
Table 1), thus exerting surface tension effects.
The influence of the gas volume flow rate is represented in
Figure 7c,d at two different counter-current liquid flow rates and for two different capillary sizes. With an increasing gas flow rate, the bubble size increased, but the difference between the 6 l·h
−1 and 10 l·h
−1 gas flow rates was negligible for the 0.18 mm capillary. Also, a higher counter-current liquid flow rate led to bigger bubbles (
Figure 7c) for this small capillary size. For the biggest 3.6 mm capillaries, bubble sizes were not affected by the counter-current liquid flow rate in most of the cases, as only an influence of the gas flow rate was observable (
Figure 7d). However, large fluctuations in bubble size were visible at the highest counter-current liquid flow rate, especially in the top half of the investigated area.
In contrast to the air bubbles, at the same gas mass flow rates, the CO
2 bubbles showed a very different behavior (
Figure 8). With an increasing capillary size, the generated bubble size increased also, but, due to the high solubility in water, it decreased continuously along the column height (
Figure 8a). The experiments show that the bubbles reached a minimum size of about 1 mm and then they stopped shrinking further (
Figure 8a). Probably, all CO
2 was dissolved from these bubbles into the liquid and the remaining bubbles consisted of less soluble and back-diffused inert gases, as already known from the literature [
24,
25]. Slightly below this size, the automatic bubble recognition reached its minimum, because the shadows of these small bubbles were almost as bright as the background. Therefore, reliable bubble recognition was not possible for
ESD ≤ 1 mm, and the representation in
Figure 8 is, therefore, interrupted (
Figure 8a, yellow diamonds).
Similarly to the air bubbles, the CO
2 bubbles were influenced by the counter-current flow (
Figure 8b). Here, however, with an increasing counter-current liquid flow rate, the bubble size became smaller over the column height, while starting with nearly the same bubble diameter at the capillary outlets. The bubble shrinkage gradually increased with increasing counter-current liquid flow rates. The bubble residence time increased, further favoring CO
2 dissolution. Also, the liquid renewal on the bubble surfaces became faster with increasing liquid flow rates, enhancing the mass transfer from the bubbles to the liquid.
The influence of the gas flow rate on the generated bubble size is depicted for two different counter-current liquid flows in
Figure 8c. With an increasing gas flow rate, the bubble size increased and gradually decreased to about 1 mm along the height of the column. At higher counter-current liquid flow rates, the bubble shrinkage became more significant, as the bubbles reached much earlier the diameter of about 1 mm. In this plot, and also in
Figure 8a, it is observable that bubble shrinkage occurred in a parallel way for the same counter-current flow conditions. For this reason, it can be assumed that the mass transfer from the bubbles to the liquid does not only depend on the initial bubble size, but also, essentially, on the counter-current flow conditions. With the 3.6 mm capillaries (
Figure 8d), an increasing gas flow rate led to bigger bubble sizes at the bottom of the column. Since bubble shrinkage was less pronounced with the lower liquid flow rate, bubble sizes were bigger at the column top compared to the high liquid counter-current flow for the same gas flow rate.
4.2. Bubble Velocity
In the following paragraphs, the measured bubble velocities will be analyzed.
Figure 9 shows the air bubble velocities under different flow conditions, with
Figure 9a representing the velocities of the bubbles generated with different capillaries. They differed significantly just above the gas outlets, where the larger bubbles had lower velocities than the smaller ones. After about half the measurement section, all bubbles reached the same mean terminal velocity of about 0.27 m·s
−1. With increasing counter-current liquid flow rates, the bubble velocity decreased, but the velocity difference between the two smallest liquid flows and the stagnant condition was negligible (
Figure 9b), corresponding to the bubble terminal velocity given by different well-known correlations [
26,
27]. Also, the influence of the gas flow rate on the bubbles generated by smaller capillaries was negligible at low counter-current liquid flow rates (
Figure 9c), while, at the maximum counter-current liquid flow rate, the velocity of the bubbles in the lower part of the column was bigger for higher gas flow rates, until reaching similar terminal velocities at about half the column height. Due to the higher gas flow rate differences with the 3.6 mm capillaries (
Figure 9d), the velocities of the bubbles differed more for the three flow conditions shown. With increasing gas flow rates, the bubble velocity increased for all counter-current liquid flow rates, but remained smaller for the higher counter-current flow.
At the same gas mass flow rates, the velocities of the CO
2 bubbles showed larger differences (
Figure 10) due to the stronger size change of the bubbles. With increasing capillary sizes, the bubbles starting velocities decreased. Then, according to bubble shrinkage, the bubble velocities decreased differently along the column height (
Figure 10a). When increasing the counter-current liquid flow rate, the bubble velocity decreased along the complete column height (
Figure 10b), while increasing gas flow rate increased the velocity for all counter-current liquid flow rates (
Figure 10c,d). The velocity of rounder bubbles (aspect ratio
E > 0.8, see Figure 12) decreased more strongly (
Figure 10c,d, highest counter-current liquid flow rate), while the velocity of the larger bubbles with a lower aspect ratio remained almost the same along the investigated height (
Figure 10d, orange, green, and purple symbols).
4.3. Aspect Ratio (E)
In the following two figures (
Figure 11 and
Figure 12), the bubble aspect ratio is plotted along the column height under different flow conditions but equivalent gas mass flow rates for air and CO
2, respectively. In the case of air bubbles, the aspect ratio of the smallest bubbles generated by the 0.13 mm capillaries increased from the bottom to the top, while it remained almost the same for the two other capillary sizes (
Figure 11a). After the bubbles had left the capillaries, the hydrostatic pressure gradually decreased and had a visible influence on the aspect ratio of the small bubbles. It is worth pointing out that neither the counter-current liquid flow rate (
Figure 11b) nor the gas flow rate (
Figure 11c) influenced significantly the aspect ratio of the bubbles. Only for the largest bubbles at the highest liquid counter-current flow rate a difference was observable (
Figure 11d, yellow, blue, and red symbols).
In the case of the CO
2 bubbles, the aspect ratio change was dominated by the bubble shrinkage (
Figure 12). With decreasing capillary sizes, the aspect ratio increased earlier, due to the initially smaller bubbles (
Figure 12a). Since the counter-current liquid flow increased bubble shrinkage, the aspect ratio also increased earlier with higher liquid flow rates (
Figure 12b). Also, with increasing gas flow rates, the aspect ratio of the small CO
2 bubbles started decreasing directly at the outlet (
Figure 12c), while the decrease started later for the bigger bubbles (
Figure 12d).
4.4. Liquid Velocity
The general liquid flow behavior in the column is depicted in
Figure 13 by streamlines, where air bubbles were generated with the 0.18 mm capillaries at a 10 l·h
−1 gas flow rate as an example.
In this figure, blue arrows indicate the main liquid flow directions. At the lowest counter-current liquid flow rate, the bubble-generated upward flow dominated almost the complete column height. With increasing counter-current liquid flow rates, the stagnation plane between the bubble-generated upward flow and the downward liquid flow moved more and more downwards. At the highest investigated counter-current liquid flow rate, no upward liquid velocity existed anymore. The extent of this effect depends on the gas flow rate and bubble size: the higher the gas flow rate and bubble size, the higher the upward velocity field that exists.
As for the bubble parameters, the vertical liquid flow velocities were averaged over the column width along the column height and are presented for comparison in the following figures.
Figure 14 shows the mean vertical liquid velocities for air bubbles under different flow conditions. Significant differences can be found in the liquid velocity between the two small (0.13 mm and 0.18 mm) and the large (3.6 mm) capillaries in
Figure 14a. The measured mean vertical liquid velocities for the two smaller capillaries were similar and slightly decreased along the column height, although the bubble size (
Figure 7a) and velocity (
Figure 9a) differed appreciably. Apparently, here, the gas flow rate was the strongest influencing parameter, and it was the same for both capillaries. In the case of the 3.6 mm capillaries, largely different flow behaviors were observed. The much larger bubbles generated an upward liquid velocity that remained almost constant over the whole investigated height. The velocity also remained always positive (upwards) since, due to their larger surface, drag, and buoyancy, the larger bubbles can mobilize more liquid and generate a stronger upward flow. Moreover, larger velocity fluctuations were found in the case of the larger bubbles.
The effect of the counter-current liquid flow on the bubble-generated upward flow is noticeable in
Figure 14b. Depending on the counter-current flow rate, the liquid velocity direction was inverted earlier from positive to negative values. This liquid flow direction change corresponds to that visible in the streamline visualization in
Figure 13. In the top 100 mm of the column, the liquid velocity reached that of a potential single-phase flow. The mean vertical liquid velocity profiles for the stagnant case and the case with the lowest counter-current liquid flow rate (0.6 l·min
−1) were very similar, except for a slight change in the top region. At a 30.8 l·min
−1 counter-current liquid flow rate (blue symbols), the measured vertical liquid velocity became negative all over the investigated height. At the highest counter-current liquid flow rate of 55.5 l·min
−1 (purple symbols), the vertical liquid velocity decreased directly from the gas inlets up to about 600 mm, where the single-phase velocity was reached.
The influence of the gas flow rate is depicted in
Figure 14c,d. While for the smaller capillaries the influence of the gas flow rate was negligible (
Figure 14c), the liquid velocities differed noticeably with the large capillaries (
Figure 14d) under larger gas flow rates. At the lowest counter-current liquid flow rate of 0.6 l·min
−1 (orange, green, and purple symbols), the liquid velocity increased appreciably with increasing gas flow rates. The highest liquid velocities were measured at around 100 mm close to the gas inlet, while they remained almost constant through the rest of the investigated height. At the highest counter-current liquid flow rate of 55.5 l·min
−1 (yellow, blue, and red symbols), the liquid velocity decreased over the complete height of the column with decreasing gas flow rates.
The mean vertical liquid velocities with CO
2 bubbles at equivalent gas mass flow rates are shown in
Figure 15. By comparing the mean vertical liquid velocities for different capillary sizes at the 30.8 l·min
−1 counter-current liquid flow rate (
Figure 15a), it becomes visible that there was no difference between the results of the two smaller capillaries (orange and green symbols), despite the large differences in bubble sizes (
Figure 8a) and bubble velocities (
Figure 10a). Due to the mass transfer from the CO
2 bubbles to the liquid, bubble sizes decreased over the rising path alongside bubble velocities; therefore, the shrinking bubbles generated a less intense upward flow than the air bubbles under the same conditions (
Figure 14a). For this reason, the liquid velocity reached nearly the velocity of the single-phase flow already at around 300 mm, and, thus, much earlier than in the case of the air bubbles. The bubbles generated with the 3.6 diameter capillaries (red symbols) were considerably larger (
Figure 8); their size decreased during their rise as well, but the bubble velocities remained constant (
Figure 10). For this reason, the upward liquid flow was stronger than in the other two cases, resulting in a continuously decreasing vertical liquid velocity profile. However, from around 800 mm, it reached the liquid velocity of the other two cases, corresponding to the single-phase velocity of that liquid flow rate.
When increasing the counter-current liquid flow rate, the liquid velocity decreased (
Figure 15b) as it did in the case of the air bubbles (
Figure 14b). It is also noticeable that the measured liquid velocities remained almost constant above 300 mm, contrary to the results from the air bubbles (
Figure 14b) where a constant velocity was only reached higher up in the column. The source of this difference can be found in
Figure 7,
Figure 8,
Figure 9 and
Figure 10. The size and velocity of the air bubbles remained nearly the same over the whole investigated height, while the CO
2 bubbles shrunk constantly and their velocities decreased during their rise. Due to the decreasing size and velocity, their influence on the liquid flow decreased as well.
The influence of the gas flow rate on the liquid velocity is depicted in
Figure 15c,d. As for the air bubbles, just a minor difference was found in the mean vertical velocities for the two lowest gas flow rates from the small capillary under different counter-current liquid flow rates (
Figure 15c). Despite the decreasing bubble size and velocity and the increasing bubble aspect ratio, the mean vertical liquid velocity decreased only in the very bottom over the gas inlets and then remained almost constant in the investigated section. Due to the bubble shrinkage and, therefore, weaker upward flow, the vertical liquid velocity reached the single-phase velocity much earlier than in the case of air bubbles. On the other hand, in the case of the 3.6 mm capillaries (
Figure 15d), the mean vertical liquid velocity differed appreciably among the different gas flow rates. With increasing gas flow rates, the liquid velocity increased in the lower half of the column before reaching the single-phase flow velocity under the highest counter-current liquid flow rate (yellow, blue, and red symbols). At the lowest counter-current liquid flow rate (orange, green, and purple symbols), the velocity decrease was much smaller and even. At the top, positive, ascending velocities were measured.
4.5. Bubble Slip Velocity
After capturing the bubble and liquid velocities, the relative bubble velocities—also called slip velocities—were calculated by transferring the discrete bubble velocities on a Cartesian coordinate system.
Figure 16 shows exemplary slip velocities (calculated from the PIV (liquid velocity) and shadow (bubble velocity) measurements) for the two applied gases at equivalent gas mass flow rates and equal counter-current liquid flow rates. Due to the very small liquid velocities compared to the bubble velocities, the slip velocities were mainly determined by the bubble velocities. The slip velocity of the air bubbles was rather evenly distributed throughout the whole column due to the rather constant bubble velocities in that case. On the other hand, the slip velocity of the CO
2 bubbles decreased continuously with the column height, since also the bubble velocities were decreasing for the CO
2 bubbles. These slip velocity distribution results were then reduced, in the same manner as that used for the former properties, to mean values over the column width. These are used in the following paragraphs along the column height for further comparisons.
Figure 17a represents the slip velocities of the air bubbles generated with different capillaries at the same liquid and gas flow rates. Close to the gas inlets, larger velocities can be observed according to the bubble and liquid velocity plots (
Figure 9a and
Figure 14a).
With increasing bubble sizes, the slip velocity decreased and remained almost constant from about half of the column height. With increasing counter-current liquid flow rates (
Figure 17b), the bubble velocity, as well as the liquid velocity, decreased; therefore, the calculated slip velocities remained rather constant over the column height. By comparing different gas flow rates for small and large capillaries at different counter-current liquid flow rates (
Figure 17c,d), it can be seen that, in both cases—small and large capillaries—the relative velocity remained nearly the same for one capillary size, despite the different bubble sizes (
Figure 7) measured over the column height. However, with the 3.6 mm capillaries, larger fluctuations in the slip velocities were observed. All in all, these results are theoretically expected and show the consistency of the separately performed gas and liquid velocity measurements: if the bubble size and shape remain the same, the gas and counter-current liquid flow rates have no influence on the slip velocity.
Describing the slip velocities of the CO
2 bubbles (
Figure 18) is more complex since mass transfer from the bubbles to the liquid changes the bubble size and shape during their rise. Accordingly, the slip velocity changed as well along the column height. Due to the gradually decreasing bubble size and velocity, the slip velocity decreased continually for all capillary sizes examined (
Figure 18a). When the bubbles reached a constant size, shape, and velocity, the slip velocity also became constant.
The effect of the counter-current liquid flow is depicted in
Figure 18b. The slip velocity for all counter-current liquid flows was constant up to around 300 mm. From this point, the slip velocity decreased, becoming stronger the higher the counter-current liquid flow rate was. With increasing counter-current liquid flow rates, the shrinkage of the bubbles became faster due to the longer contact time with the liquid. With decreasing bubble sizes, the velocity decreased as well and the aspect ratio increased. For this reason, the slip velocity decreased with increasing counter-current liquid flow rates. After the CO
2 was depleted from the bubbles, the remaining small gas bubbles became spherical, exhibiting about the same size in the upper part of the column. As it has already been described for the air bubbles, bubbles with the same size and aspect ratio have the same slip velocity, as it can be observed here as well.
The influence of the gas flow rate was weak for the smaller capillary (
Figure 18c), but smaller gas flow rates led to a slightly slower slip velocity, especially in the upper half of the column where an influence of the counter-current liquid flow rate became obvious. For the biggest bubbles (
Figure 18d), the slip velocity stayed more or less the same, apart from when under the highest counter-current liquid flow rate, where it decreased in the upper part of the column, as it was the case for the bubble velocity.
4.6. Comparison of Mean Bubble Diameters and Aspect Ratios with Literature Correlations
Mean aspect ratios and bubble diameters were finally determined from all measurement results, as mentioned in
Section 3 and
Figure 6, in order to compare these results to empirical correlations for bubble aspect ratio (
Figure 19 and
Figure 20). The correlations used here are those of Besagni and Deen [
28], obtained from data from different fluid and bubble column combinations, and that of Liu et al. [
29] for air bubble swarms in a pseudo-2D column.
It is clearly visible that, with different capillary sizes, that is, different bubble diameters, different bubble aspect ratios were obtained. Moreover, the measurement results for air and CO
2 bubbles are similar for the different capillaries (see symbols in
Figure 19 and
Figure 20). Generally speaking, with increasing bubble diameters, the aspect ratio decreased. Only one exception exists for bubbles produced with the 0.18 mm nozzles and having diameters between 3 mm and 5 mm. Here, the aspect ratio increased with increasing ESD. This behavior may be attributed to the nozzle material, which was Teflon in that case, that behaves differently from the stainless steel and peek nozzles of the other diameters.
These mean aspect ratios are compared to the aforementioned empirical correlations from the literature (see lines in
Figure 19 and
Figure 20). Globally, both correlations capture well the decreasing trend of the aspect ratio with increasing bubble diameter. However, for both gases used, the correlation of Besagni and Deen [
28] underestimated the aspect ratio of bubble diameters up to about 2 mm, while it overestimated the bubble aspect ratios of nearly all cases bigger than 2 mm. The correlation of Liu et al. [
29] performed better for the bubble aspect ratios of air bubbles generated with the 0.13 mm capillaries and for the CO
2 bubbles generated with the 0.18 mm nozzles. For bubbles bigger than 5 mm, the aspect ratios were mostly underestimated with this correlation.
These discrepancies that cannot be explained by the measurement uncertainty (see
Section 2) nor by the validity range of the correlations, show the necessity for more sophisticated correlations, taking into account the gas and liquid properties as well as the bubble generation conditions. Therefore, the actual bubble data presented in this paper are also further used for the evaluation of correlations incorporating these influencing parameters.