1. Introduction
As a radiation diagnostic procedure, posteroanterior (PA) chest radiography is the most commonly applied methodology and it is increasingly important in implications for delivering excessive radiation to patients to provide diagnostically acceptable images. In practice, it has been noticed that radiological technologists have tended to prefer high exposure in X-ray exams because it provides high-quality images. Although it is known that the individual patient dose in PA chest radiography is comparatively low, it can significantly affect the cumulative dose with the frequent use of radiography [
1]. While patients are exposed to excessive radiation, their exposure dose has been highlighted globally. Furthermore, apprehension about possible harm to people caused by radiation exposure has risen as the clinical use of diagnostic procedures increases. The radiological technologists have been demanding dose information [
2,
3].
The entrance surface dose (ESD reported in mGy) value is specified as a useful descriptor to evaluate the absorbed dose of the patient in diagnostic radiology and it is generally defined as the absorbed dose detected at the point of intersection of the central X-ray beam axis with the patient or phantom surface, including backscattered radiation [
2,
3]. In practice, however, the absorbed dose is delivered unevenly in a beam irradiation field because the intensity distribution in an X-ray beam is not uniform due to the influence of the heel effect. Hence, this may lead to the possibility that the ESD value measured at the center of the beam irradiation field can be underestimated because the higher absorbed doses are practically delivered at the other points. It means that an excessive radiation dose can be delivered to a part of the patient's body even though the measured ESD value is under the diagnostic reference level (DRL).
In recent years, a multitude of studies have been conducted to measure the radiation dose in order to optimize radiographic procedures to achieve as reasonably low a radiation dose as possible, and the results have indicated possibilities to reduce the dose without detriment to image quality [
4]. For successful and efficient medical effects, safety and quality are imperative. In addition, precision and sensitivity of the operating characteristics of the imaging system are related to benefits for patient health [
5]. In diagnostic radiology, the correlation of image quality with patient absorbed dose must be set to provide acceptable image quality at the lowest practicable dose level; this balance is usually defined by the exposure parameters of the X-ray imaging system. The periodic assessment of exposures in the comprehensive quality assurance (QA) of an X-ray imaging system is an important part to identify problems and to ensure that diagnostic quality is maintained without over-exposure [
6].
For
in situ and real-time measurements of the absorbed dose at the measuring points, various kinds of clinical dosimeters have been developed, including ionization chambers (ICs) and semiconductor dosimeters (SCDs). However, in the cases of ICs and SCDs, they have a large sensing volume and their sensing probes are composed of high-atomic-number materials. Accordingly, most conventional ICs and SCDs can generate distracting image artifacts and interfere with the examination of X-ray imaging [
7]. To overcome the problems of existing dosimeters, various scintillating fiber-optic dosimeters (SFODs) based on plastic scintillating fibers (PSFs) and plastic optical fibers (POFs) have been investigated as promising candidates because SFODs have many useful dosimetric qualities, such as a small sensing probe with low-atomic-number materials (
i.e., near tissue-equivalent materials), a light weight, adequate flexibility, immunity to electromagnetic interference (EMI), remote operation capabilities, and real-time measurement [
3,
7,
8,
9]. Several studies on the performance evaluation of a fiber-optic-coupled dosimeter (FOC) for use in radiation diagnosis have been reported by Hintenlang
et al. [
9,
10,
11]. However, the design of the phantom dosimeter proposed in the present study is significantly different from that of the existing FOC system fabricated by Hintenlang
et al. Recently, various experimental studies on the real-time measurement of ESD using a SFOD during radiation diagnosis have been carried out by our research team [
3,
7,
8]. In previous experiments, we developed a SFOD with two sensing probes to infer the ESD value at the center of a beam irradiation field without the impediment of X-ray imaging [
8] and demonstrated that a SFOD can be used for the measurement of real-time ESD [
7]. However, it is difficult to rapidly measure the two-dimensional (2D) dose distribution of the heel effect affecting the X-ray beam used in radiation diagnosis with existing SFOD systems.
In the present study, to measure the planar dose distribution and to apply it to the QA of a diagnostic X-ray imaging system, a multichannel all-in-one phantom dosimeter system was developed using nine sensing probes, a chest phantom, an image intensifier, and a complementary metal-oxide semiconductor (CMOS) image sensor. If the beam quality of the X-ray exceeds the standard values, the dose measurement may be invalid and the patient can be exposed to an unnecessary dose [
12]. Accordingly, the phantom dosimeter system was exploited to measure the beam profile of the X-ray used in radiation diagnosis to acquire information on the beam quality. Using a beam profile, we obtained the dose distribution that can be applied to check the X-ray beam quality. Furthermore, we selected a suitable measuring point according to the relationship between each sensing probe for calculating the ESD values.
2. Materials and Instruments
2.1. Fabrication of Nine Sensing Probes
A PSF (BCF-12, Saint-Gobain Ceramic & Plastics, Hiram, OH, USA), which has a core/single-clad structure with a diameter of 1 mm, was employed to generate scintillation signals as a sensitive material of the sensing probes. The core is a synthetic material of polystyrene and fluorescent dopants and the cladding material is polymethylmethacrylate (PMMA). The refractive index of the core is 1.60 and the numerical aperture (NA) is approximately 0.58. The decay time and the emission peak are 2.7 ns and 435 nm, respectively. To transmit the scintillation signals generated from the PSF in the sensing probe, a POF (GH 4001, Mitsubishi Rayon, Tokyo, Japan) with a step-index and multimode characteristics was used. This POF has a core diameter of 0.98 mm and an outer diameter of 1 mm. The core material is PMMA resin and the cladding is made of fluorinated polymer. The POF is covered by a black jacket based on polyethylene to shield extraneous light. The refractive indices of the core and the cladding are 1.49 and 1.402, respectively; therefore, the NA is approximately 0.50. The POF length is about 6 m, which is a sufficient length considering the distance between the examination room and the control room.
The inner structure of the sensing probe is described in
Figure 1. A 10 mm long PSF was polished and coupled to a 6 m long POF. The PSF was then covered by a reflective paint (BC-620, Saint-Gobain Ceramic & Plastics, Hiram, OH, USA) based on titanium dioxide (TiO
2) to improve the light collection efficiency. In addition, the sensing probe was wrapped with black tape to shield ambient light. Before the experiments, we first selected nine promising probes which have almost identical light output by using a multi-pixel photon counter (MPPC: S10362-11-100U, Hamamatsu Photonics, Hamamatsu, Japan). We also obtained the calibration factor by measuring the light intensities of all probes at the same position.
Figure 1.
Inner structure of the sensing probe and the arrangement of nine sensing probes at each center of nine equal parts on the top of the ANSI sensitometry chest phantom.
Figure 1.
Inner structure of the sensing probe and the arrangement of nine sensing probes at each center of nine equal parts on the top of the ANSI sensitometry chest phantom.
As a chest phantom, an American National Standards Institute (ANSI) sensitometry chest phantom was fabricated [
3]. The description and use of the ANSI sensitometry chest phantom is proposed in report No. 31 which is published for the American Association of Physicists in Medicine (AAPM) by the American Institute of Physics (AIP) [
13]. As can be seen in
Figure 1, the top of the chest phantom was divided into nine equal parts virtually and we made nine holes with a diameter of 2.5 mm at each center of the nine equal parts on the top of the chest phantom. The distal ends of nine sensing probes with an identical structure were installed at each center of the nine equal parts as measuring points (
i.e., channel 1–9) to measure the planar dose distribution. The transmitting POFs of nine sensing probes were taken in the form of a coherent bundle with a 3 × 3 square array by using a plastic holder and then connected to an image intensifier (MCP-PROXIFIER
®, ProxiVision GmbH, Bensheim, Germany) with a fiber-optic window and a useful diameter of 40 mm to increase the light intensity of the scintillation signals transmitted via POFs.
2.2. Multichannel All-In-One Phantom Dosimeter System for Measuring Planar Dose Distribution of X-ray Beam
Figure 2 illustrates the experimental setup to obtain the planar dose distribution on a chest phantom using the multichannel all-in-one phantom dosimeter system. The performance evaluation of the proposed phantom dosimeter was carried out using an X-ray tube (Rotanode
TM E7252X, Toshiba Electron Tubes & Devices, Saitama, Japan) of a digital radiography (DR) system (Clear Vision DR 7000F, JPI Healthcare, Osong, Korea) with 0.6/1.2 mm dual focal spots, a 12° rhenium-tungsten-faced molybdenum target, and 0.9 mm Al filtration at 75 kVp [
7]. When the sensing probes installed at the top of the chest phantom are irradiated by an X-ray beam, the scintillation signals are transmitted through the POFs and their light intensities are intensified by the image intensifier. To intensify the scintillation signals, it performs the following three main processes. First, a bialkali photocathode with a composition of K
2SbCs on the distal end of an image intensifier converts the entered photons into electrons. The electrons are then accelerated through the microchannel plates (MCPs). Finally, the phosphor screen with a composition of Gd
2O
2S:Tb on the proximal end of the image intensifier converts electrons back into photons.
Figure 2.
Experimental setup using a multichannel all-in-one phantom dosimeter and a DR system.
Figure 2.
Experimental setup using a multichannel all-in-one phantom dosimeter and a DR system.
The real-time scintillation images (RSIs) including the intensified nine scintillation signals are taken by a CMOS image sensor (acA3800-14uc, BASLER, Ahrensburg, Germany) with a single lens optical system. The RSIs are converted into red-green-blue (RGB) image files by the MATLAB program (MATLAB®, MathWorks, Natick, MA, USA) and subsequently transformed to gray scale. The gray-scale image format is corrected by multiplying the calibration factor acquired from a reference parameter.
Using nine sensing probes of the phantom dosimeter system that are installed in the upper part of the chest phantom, the RSIs were measured to obtain the planar dose distribution, which changed with the variation of the X-ray exposure parameters. We then compared the results with the absorbed doses measured by using a conventional SCD (ALPHA plus, Pehamed, Sulzbach, Germany) with a finger-shaped dose sensor. In this study, the values of the exposure parameters such as the beam field size, focus-to-image receptor distance (FID), tube potential, and current-time product were fixed at 30 × 30 cm2, 100 cm, 100 kVp, and 5 mAs, respectively.
4. Discussion and Conclusions
In this study, we encouraged the apprehension of over-exposure due to the heel effect in part of the patient's body during chest radiography. To obtain an X-ray beam profile, we measured the light intensities of scintillation signals due to the X-ray beam and demonstrated that the planar dose distribution including the ESD can be easily and quickly measured by using the proposed multichannel all-in-one phantom dosimeter system. From the experimental results, we confirmed that the beam profile of an X-ray used in diagnostic radiology is affected predominantly by the heel effect and the absorbed dose which is higher than the ESD values can be delivered to some other points, not the center of the beam irradiation field.
In conclusion, it is expected that the proposed phantom dosimeter will be a useful dosimeter to measure the real-time dose distribution and the beam quality of a DR system and to select the adequate measuring point for application of the modified direct measurement before chest radiography. Further studies will be conducted to develop a newly designed multichannel all-in-one phantom dosimeter system to accurately measure the planar and depth dose distribution with enhanced spatial resolution.