Next Article in Journal
The Use of Image Analysis to Detect Seed Contamination—A Case Study of Triticale
Next Article in Special Issue
Imbalanced Loss-Integrated Deep-Learning-Based Ultrasound Image Analysis for Diagnosis of Rotator-Cuff Tear
Previous Article in Journal
Combined Use of Vis-NIR and XRF Sensors for Tropical Soil Fertility Analysis: Assessing Different Data Fusion Approaches
Previous Article in Special Issue
FRET-Based Ca2+ Biosensor Single Cell Imaging Interrogated by High-Frequency Ultrasound
 
 
Letter
Peer-Review Record

Design of an Ultrasound Transceiver ASIC with a Switching-Artifact Reduction Technique for 3D Carotid Artery Imaging

Sensors 2021, 21(1), 150; https://doi.org/10.3390/s21010150
by Taehoon Kim 1,*, Fabian Fool 2, Djalma Simoes dos Santos 2, Zu-Yao Chang 1, Emile Noothout 2, Hendrik J. Vos 2,3, Johan G. Bosch 3, Martin D. Verweij 2,3, Nico de Jong 2,3 and Michiel A. P. Pertijs 1
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Sensors 2021, 21(1), 150; https://doi.org/10.3390/s21010150
Submission received: 30 November 2020 / Revised: 24 December 2020 / Accepted: 25 December 2020 / Published: 29 December 2020
(This article belongs to the Special Issue Ultrasonic Systems for Biomedical Sensing)

Round 1

Reviewer 1 Report

This manuscript entitled "Design of an Ultrasound Transceiver ASIC with a Switching-Artifact Reduction Technique for 3-D Carotid Artery Imaging", proposed a new switching algorithm to reduce the switching artifacts which degrade the ultrasound image quality.

In this manuscript, the architecture of the proposed ASIC is well described. However,  the experimental result should be explained in more detail.

Please clarify how much of the impact the 20 dB noise reduction will have on the actual imaging.
In the experiment, the target was a quartz plate and its echo signal was not covered from the switching artifacts. In the actual use, the artifacts will cover the echoes from many reflectors.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 2 Report

Dear authors,

your paper intitled "Design of an Ultrasound Transceiver ASIC 2 with aSwitching-Artifact Reduction Technique 3 for 3-D Carotid Artery Imaging" is really interesting.
The paper is well written, but in my opinion an overview on the cross section of ASIC is mandatory and it is to be added in the manuscript. Please add arepresentative figure, in which you can show layer by layer the construction/manufacturing of the sensor (from bottom to the patient).
Are there matching layers?
And about the backing?
Did you use a filling between the PZT squares?
There is no description of the material involved.

Take care.

Author Response

Please see the attachment.

Author Response File: Author Response.pdf