Next Article in Journal
A Novel Deep-Learning Model Compression Based on Filter-Stripe Group Pruning and Its IoT Application
Next Article in Special Issue
C-Axis Textured, 2–3 μm Thick Al0.75Sc0.25N Films Grown on Chemically Formed TiN/Ti Seeding Layers for MEMS Applications
Previous Article in Journal
VOC Detections by p-Type Semiconducting Sensors Using Nano-Sized SmFeO3 Particles
Previous Article in Special Issue
High Electrocaloric Effect in Lead Scandium Tantalate Thin Films with Interdigitated Electrodes
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition

1
Department of Materials Science and Engineering, Penn State University, University Park, PA 16802, USA
2
Department of Biomedical Engineering, Penn State University, University Park, PA 16802, USA
3
Department of Materials Science and Engineering and Materials Research Institute, Penn State University, University Park, PA 16802, USA
*
Author to whom correspondence should be addressed.
Sensors 2022, 22(15), 5614; https://doi.org/10.3390/s22155614
Submission received: 28 June 2022 / Revised: 23 July 2022 / Accepted: 25 July 2022 / Published: 27 July 2022
(This article belongs to the Special Issue Application of Ferroelectric Thin Films in MEMS)

Abstract

:
The receive sensitivity of lead zirconate titanate (PZT) piezoelectric micromachined ultrasound transducers (PMUTs) was improved by applying a DC bias during operation. The PMUT receive sensitivity is governed by the voltage piezoelectric coefficient, h31,f. With applied DC biases (up to 15 V) on a 2 μm PbZr0.52Ti0.48O3 film, e31,f increased 1.6 times, permittivity decreased by a factor of 0.6, and the voltage coefficient increased by ~2.5 times. For released PMUT devices, the ultrasound receive sensitivity improved by 2.5 times and the photoacoustic signal improved 1.9 times with 15 V applied DC bias. B-mode photoacoustic imaging experiments showed that with DC bias, the PMUT received clearer photoacoustic signals from pencil leads at 4.3 cm, compared to 3.7 cm without DC bias.

1. Introduction

Miniaturized ultrasound sensors with low voltage operation and high transmit/receive signals are desirable for many applications. Micromachined ultrasound transducers prepared via MEMS fabrication have been developed to meet this need. Typically, their output acoustic signals are smaller than those from transducers built using bulk piezoelectrics [1]. A variety of methodologies have been utilized to improve the acoustic characteristics of MUTs, such as operating in the collapse mode for capacitive micromachined ultrasound transducers or fabricating dome-shaped membranes for piezoelectric micromachined ultrasound transducers [2,3].
Lead zirconate titanate and aluminum nitride are commonly used materials for piezoelectric micromachined ultrasound transducers. PZT is valued for its high piezoelectric coefficient and thus high transmit pressures at low actuation voltages [4]. The piezoelectric coefficient of aluminum nitride (AlN) is about an order of magnitude lower than PZT, and its dielectric constant is approximately two orders of magnitude lower than that of PZT [5,6]. Although the receive sensitivity of AlN is desirable, higher transmit power is needed for most ultrasonic imaging.
Receive sensitivity is a function of the piezoelectric voltage coefficient, h31,f, which is expressed as:
h31,f = e31,f/εr
where e31,f and ɛr are the piezoelectric coefficient and permittivity, respectively. Thus, to improve receive sensitivity, the piezoelectric coefficient must be high and/or the permittivity must be low.
When large electrical fields are applied to a ferroelectric, such as PZT, the permittivity decreases below the zero field value due to reduced intrinsic and extrinsic contributions to the permittivity [7]. Moreover, the piezoelectric coefficient may also increase with large DC electric fields, as domains more readily remain aligned with the applied electric field [8,9,10,11]. Thus, if a DC electric field is applied while the PMUT functions as a receiver, the receive sensitivity should be higher than at zero applied bias. The trends in piezoelectric coefficient and permittivity have been reported for air-coupled transducers [6] but have not been demonstrated in underwater ultrasonic and photoacoustic applications.
In addition, this paper discusses different PZT compositions and their suitability for ultrasound receive sensitivity. The morphotropic phase boundary (MPB) composition of PZT with a Zr/Ti ratio of 52/48 is often utilized for its high permittivity and piezoelectric coefficients [12]. As the composition moves off the MPB, the piezoelectric coefficient and permittivity decrease [12,13] at different rates. Thus, on changing the composition, the permittivity may decrease faster than that of the piezoelectric coefficient. It is hypothesized that for ultrasound applications moving off the MPB may result in higher receive sensitivity while maintaining relatively high transmit power.

2. Materials and Methods

To test whether the permittivity and piezoelectric coefficient increase with DC bias, unreleased samples with an architecture similar to that of PMUTs were fabricated. A 500 μm Si wafer with 1 μm of SiO2 on both sides (NOVA Electronic Materials, TX, USA) was used as the substrate. The bottom electrode was deposited by sputtering 30 nm of Ti, annealing it at 700 °C for 15 min with 10 sccm of oxygen flow, and then sputtering 100 nm of Pt at a substrate temperature of 500 °C [14]. To facilitate orientation of the piezoelectric film, a PZT seed layer (Mitsubishi Materials Corporation, Hyogo, Japan) was spun, pyrolyzed, and crystallized, as described elsewhere [10,15,16,17]. For the MPB composition, 14 mol% lead excess Pb(Zr0.52Ti0.48)O3 solution doped with 2% Nb (Mitsubishi Materials Corporation, Hyogo, Japan) was spun on at 2750 RPM for 45 s. The film was pyrolyzed at 100 °C for 1 min and 300 °C for 4 min, followed by crystallization in a lead-rich rapid thermal annealer for 1 min at 700 °C. This process was repeated ~20 times until a film thickness of ~2.0 μm was achieved.
In order to investigate various compositions, an inverted mixing order process was used to prepare PZT solutions with different Zr/Ti ratios [18,19,20,21,22,23,24]. 0.4 M solutions of PZT with Zr/Ti ratios of 52/48, 40/60, 30/70, and 20/80 were spun on the hot-sputtered Pt substrates with the Mitsubishi seed layer at 3000 RPM for 30 s. The film was pyrolyzed at 250 °C for 30 s and 400 °C for 1 min, followed by crystallization in a lead-rich rapid thermal annealer for 2 min at 650 °C. The top Pt electrodes were patterned via liftoff and annealed at 600 °C for 1 min. These samples were then diced. Pieces were adhered to the center of a Si wafer, and strain gauges were superglued onto the die. The e31,f was measured via the wafer flexure method [25].
In addition, 14 mol% lead excess Pb(Zr0.52Ti0.48)O3 solution doped with 2% Nb (Mitsubishi Materials Corporation, Hyogo, Japan) was spun on a separate wafer at 2750 rpm for 45 s. The film was pyrolyzed at 100 °C for 1 min and 300 °C for 4 min, followed by crystallization in a lead-rich rapid thermal annealer for 1 min at 700 °C. This process was repeated ~20 times until a total thickness of ~2.0 µm was achieved. The resulting film had a {100} Lotgering factor >97%. The top Pt electrodes were patterned via liftoff and annealed at 600 °C for 1 min. The wafer was then diced and clamped to the center of a Si carrier wafer, and strain gauges were glued onto the die.
Following sample fabrication, the effect of DC bias on e31,f and εr was measured on clamped samples. Prior to measurement, each device was poled at 35 V for 20 min at 150 °C. Pre-poling allowed the DC bias dependence to be deconvoluted from poling, as was apparent in previous studies. The wafer flexure method was used to characterize e31,f.
The design, fabrication process, and characterization for circular 6–8 MHz PMUTs are described elsewhere [10,17]. The PMUTs were wire-bonded to a pin grid array (PGA) or a circuit board and coated with 2 µm of parylene for waterproofing.
The receive sensitivity was obtained with the setup in Figure 1a. The bulk transducer, functioning as a pressure source, was excited using a unipolar 5 Vpp burst sine waveform at its resonance, 3.5 MHz, 8.5 mm distance from the receiving PMUT.
For photoacoustic measurements, a PMUT was submerged in a distilled water tank facing a black card, as seen in Figure 1b. A tunable (680–980 nm) nanosecond laser (Phocus Mobile, Opotek Inc., Carlsbad, CA, USA) delivered 10 ns pulse width laser pulses at a repetition frequency of 10 Hz and 120 mJ/pulse energy through a custom-fabricated fiber optic bundle (Fiberoptic Systems Inc., Simi Valley, CA, USA). The PMUT was housed in a custom printed circuit board and optical fiber bundle apparatus with an additional external gain of 20 dB as described elsewhere [10,17]. Pencil leads 0.3 mm in diameter were suspended at different heights in an agar-gel phantom to test the photoacoustic sensitivity of the PMUT as a function of depth and to see if DC biasing produced improved depth sensitivity. The preparation of the agar-gel phantom is described elsewhere [17]. A single element of the PMUT was linearly scanned across the fixed target, and the photoacoustic A-lines were compiled to form a B-mode photoacoustic image.
For pitch-catch measurements, one PMUT element functioned as the ultrasound transmitter, and an adjacent element on the same PMUT device functioned as a receiver. The ultrasound waves travel from the transmitter and, at a defined distance, reflect at an area of high acoustic impedance mismatch. The waves then travel back to the device, where the signal is acquired by the receiver. In this setup, the PMUT was placed inside a pin-grid array cavity, which was filled with distilled water. A glass slide was placed on top of the cavity, acting as a reflector. One PMUT element was excited with a single cycle of unipolar 5 Vpp in a sinusoidal burst mode. Another PMUT element functioned as a receiver with an external 39 dB gain. A bias tee was used to apply a DC bias to the PMUT during receive.

3. Results

3.1. Structural Measurements

The X-Ray diffraction patterns and FESEM images of various PZT compositions are shown in Figure 2. The Lotgering factors are similar and show highly oriented, phase pure perovskite PZT for each composition. As expected, the degree of tetragonal splitting increases as the film becomes more Ti rich. In addition, the grain sizes were determined using the line intercept method. The grain sizes for 52/48, 40/60, 30/70, and 20/80 were 212 ± 19 nm, 203 ± 113 nm, 197 ± 24 nm, and 232 ± 26 nm, respectively, indicating statistically similar grain sizes for all compositions.

3.2. Electrical Measurements

The baseline dielectric characterization of the films is given in Figure 3. As expected, the permittivity decreased for compositions of the morphotropic phase boundary (MPB). As the composition approaches lead titanate, the remanent polarization increases due to increased tetragonality and possibly, a reduction in the film stress. It should be noted that for the 20/80 composition, a maximum of 400 kV/cm was applied for the hysteresis loop measurement, as the film becomes leaky at greater electric fields, as documented in the literature [12].
Figure 4 shows the permittivity (ɛr) and piezoelectric coefficient (e31,f) as a function of DC bias for the hot-poled PZT 52/48 film. At 15 V (75 kV/cm), the permittivity decreases by a factor of 0.6, and the piezoelectric coefficient increases by a factor of 1.7 relative to the zero bias values. The e31,f starts to saturate at ~10 V, corresponding to the value of the coercive field ~50 kV/cm. Based on the normalized voltage coefficient, the receive sensitivity is expected to improve by ~2.5 times with a DC bias of 15 V.
Figure 5 shows the tunability of the different PZT compositions. The dielectric tunabilities for the 52/48, 40/60, 30/70, and 20/80 compositions were 43%, 30%, 16%, and 11%, respectively, at an electric field of 75 kV/cm. The PZT 52/48 had the largest piezoelectric and permittivity tunabilities, as expected, since the maximum polarizability occurs at the brink of structural instability (i.e., at the MPB). The decrease in the permittivity and piezoelectric coefficient as a function of composition was different. From the MPB to the PZT 20/80 film, the zero bias permittivity and piezoelectric coefficient decreased by a factor of 65% and 43%, respectively.
The resulting voltage coefficient, which controls the receive sensitivity, is given in Figure 6. For operation at 0 DC bias, PZT 20/80 has a higher voltage coefficient than the other compositions. However, when a DC bias of 75 kV/cm is applied, the voltage coefficient of a 52/48 composition can match and potentially surpass the voltage coefficient of the PZT 20/80 composition.

3.3. Underwater Acoustic Measurements

The underwater receive sensitivity and photoacoustic sensitivity as a function of DC bias are shown in Figure 7 for a PMUT made with MPB PZT. The ultrasound receive sensitivity and the photoacoustic receive signal increased by approximately two times, agreeing with the electrical measurements in Figure 6. With no DC bias, the e31,f and the permittivity values were –10 C/m2 and 1300, respectively, with h31,f = −0.0062 C/m2. At 15 V DC bias, the e31,f and the relative permittivity were −13.8 C/m2 and 740, respectively, yielding h31,f = −0.018 C/m2. This is ~18% of the voltage coefficient of AlN, which has a piezoelectric coefficient and permittivity of −1.05 C/m2 and 10.5, respectively [5,6].
It has been reported that the resonant frequencies and bandwidth can increase as much as 33% with an applied DC bias for air-coupled transducers [6]. This may be a consequence of the applied voltage increasing the in-plane stress of the piezoelectric. This would occur in the case when stress dominates over the flexural rigidity, such that increased stress increases the resonant frequency [26]. To examine whether significant changes can be seen in operating frequency and bandwidth as a function of DC bias for these underwater PMUT, the center frequency and bandwidth were extracted from the Fourier transform of the pitch-catch and photoacoustic receive signals, as shown in Figure 7c. In both cases, the center frequency (6.89 MHz) and bandwidth remained relatively constant as a function of DC bias (from 0–15 V). The center frequency had a standard deviation of 0.6%. The photoacoustic bandwidth and pitch catch bandwidths had standard deviations of 2% and 8%, respectively. The photoacoustic center frequencies and bandwidth also showed little variation with DC bias, albeit the photoacoustic bandwidth was higher (69%) than the pitch-catch bandwidth (41%). The higher bandwidth for photoacoustics compared to pitch-catch is due to two factors: First, in pitch-catch ultrasound imaging, spatial resolution is governed by the frequency of the transmitted acoustic wave arriving at the receiver, whereas in photoacoustic imaging, the laser pulses have nanosecond-scale excitation with much wider frequency content [27,28]. Second, the bandwidth for one-direction sound propagation is larger than the two-way bandwidth.
B-mode photoacoustic images were reconstructed from the raw voltage-time data, as shown in Figure 8 and Figure 9. Without DC bias, a clear photoacoustic signal was only obtained from four of the five pencil leads, the fourth being at ~3.7 cm in depth. With 15 V DC bias, the photoacoustic signal was increased for all pencil lead targets. In addition, the PMUT was able to receive a photoacoustic signal from the fifth pencil lead at 4.3 cm depth. Figure 8 shows an ~20 dB improvement in the image signal. The cause of this is that images with and without DC bias were subjected to log compression together in the same data matrix, e.g., the raw data for Figure 8b,c were subjected to one log compression operation for overall image comparison. Thus, when compressing the two images in a single data matrix, the normalization caused the observed signal intensity difference between the two images to exceed two times. When the log compression was done separately for each image, as shown in Figure 9, the photoacoustic signal with DC bias was ~2 times greater than the signals seen without DC bias, agreeing with the electrical measurements. It is observed that the signal-to-noise ratio (SNR) was significantly better with DC bias than without DC bias and applying the log compression of both images in the same dataset caused a larger improvement in the imaging signal and quality. It is not clear why this should be the case.

4. Conclusions

The receive sensitivities increased by ~2 times when 15 V (75 kV/cm) was applied to the MPB PZT PMUT, due to the increase in the e31,f and a decrease in the relative permittivity. Operating at a 15 V DC bias also yielded stronger photoacoustic signals and allowed detection of signals at greater depths compared to without DC bias. When no DC bias is applied, PZT with composition 20/80 has a significantly higher voltage coefficient than other PZT compositions tested. When 15 V DC bias (75 kV/cm) is applied, PZT 52/48 matched the receive sensitivity of PZT 20/80. Thus, operating at higher electric fields in piezoelectric-based thin film devices can improve sensitivities. It is likely that a comparable improvement could be achieved by imprinting the PZT. This would be useful as it would allow higher receive sensitivities for the PMUT without the need for a DC bias on receive.
It is known that higher applied DC electric fields result in a shorter lifetime and faster appearance of electrothermal breakdown events [29,30,31]. In practice, the times for which the DC voltage would need to be applied are quite short (microsecond time scales per pulse), so it is unlikely that the PMUT lifetime will be limited by the DC bias based on previously reported highly accelerated lifetime data. However, it is also known that the lifetime and field-induced deflection can further be reduced by combinations of humidity and AC voltage [32]. Thus, the time-dependence of the receive sensitivity at different DC biases in the waterproofed PMUT should be explored while the device is submerged in future work.

Author Contributions

Conceptualization, C.C. and S.T.-M.; methodology, C.C., T.P. and S.T.-M.; software, A.D., S.A. and H.C.; validation, C.C., A.D., S.A. and H.C.; formal analysis, C.C.; investigation, C.C.; resources, S.-R.K.; data curation, S.A. and C.C.; writing—original draft preparation, C.C.; writing—review and editing, S.T.-M. and S.-R.K.; visualization, C.C.; supervision, S.T.-M. and S.-R.K.; project administration, S.T.-M.; funding acquisition, S.T.-M. and S.-R.K. All authors have read and agreed to the published version of the manuscript.

Funding

This manuscript is based on work supported by four funding sources: the National Science Foundation Material Research Science and Engineering Center (MRSEC) supported by the National Science Foundation under grant DMR-1420620, the National Defense Science and Engineering Graduate Fellowship, and 500000019275 ASSIST Year 10 NSF Funding and R00EB017729-04 (SRK) grant.

Data Availability Statement

The corresponding author will provide data available on reasonable request due to restrictions. The data are not publicly available due to funding agent policies.

Conflicts of Interest

The authors declare no conflict of interest. The funders had no role in the design of the study; in the collection, analyses, or interpretation of data; in the writing of the manuscript, or in the decision to publish the results.

References

  1. Mattiat, O.E. Ultrasonic Transducer Materials; Plenum Press: New York, NY, USA, 1971. [Google Scholar]
  2. Park, K.K.; Oralkan, O.; Khuri-Yakub, B.T. A Comparison Between Conventional and Collapse-Mode Capacitive Micromachined Ultrasonic Transducers in 10-MHz 1-D Arrays. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2013, 60, 1245–1255. [Google Scholar] [CrossRef] [PubMed]
  3. Hajati, A.; Latev, D.; Gardner, D.; Hajati, A.; Imai, D.; Torrey, M.; Schoeppler, M. Three-Dimensional Micro Electromechanical System Piezoelectric Ultrasound Transducer. Appl. Phys. Lett. 2012, 101, 253101. [Google Scholar] [CrossRef] [Green Version]
  4. Kim, K.; Choi, H. High-Efficiency High-Voltage Class F Amplifier for High-Frequency Wireless Ultrasound Systems. PLoS ONE 2021, 16, e0249034. [Google Scholar] [CrossRef]
  5. Wang, Q.; Lu, Y.; Mishin, S.; Oshmyansky, Y.; Horsley, D.A. Design, Fabrication, and Characterization of Scandium Aluminum Nitride-Based Piezoelectric Micromachined Ultrasonic Transducers. J. Microelectromech. Syst. 2017, 26, 1132–1139. [Google Scholar] [CrossRef]
  6. Kusano, Y.; Wang, Q.; Luo, G.L.; Lu, Y.; Rudy, R.Q.; Polcawich, R.G.; Horsley, D.A. Effects of DC Bias Tuning on Air-Coupled PZT Piezoelectric Micromachined Ultrasonic Transducers. J. Microelectromech. Syst. 2018, 27, 296–304. [Google Scholar] [CrossRef]
  7. Bassiri-Gharb, N.; Fujii, I.; Hong, E.; Trolier-McKinstry, S.; Taylor, D.; Damjanovic, D. Domain Wall Contributions to the Properties of Piezoelectric Thin Films. J. Electroceram. 2007, 19, 49–67. [Google Scholar] [CrossRef]
  8. Muralt, P.; Ledermann, N.; Baborowski, J.; Barzegar, A.; Gentil, S.; Belgacem, B.; Petitgrand, S.; Bosseboeuf, A.; Setter, N. Piezoelectric Micromachined Ultrasonic Transducers Based on PZT Thin Films. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2005, 52, 2276–2288. [Google Scholar] [CrossRef] [PubMed]
  9. Masuda, Y.; Baba, A.D.C. Bias and Frequency Dependence of the Dielectric Constant PZT Family Ferroelectric Ceramics. Jpn. J. Appl. Phys. 1985, 24, 113–116. [Google Scholar] [CrossRef] [Green Version]
  10. Cheng, C.Y.; Dangi, A.; Ren, L.; Tiwari, S.; Benoit, R.R.; Qiu, Y.; Lay, H.S.; Agrawal, S.; Pratap, R.; Kothapalli, S.R.; et al. Thin Film PZT-Based PMUT Arrays for Deterministic Particle Manipulation. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2019, 66, 1606–1615. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  11. Dausch, D.E.; Gilchrist, K.H.; Carlson, J.R.; Castelucci, J.B.; Chou, D.R.; von Ramm, O.T. Improved Pulse-Echo Imaging Performance for Flexure-Mode pMUT Arrays. In Proceedings of the 2010 IEEE International Ultrasonics Symposium, San Diego, CA, USA, 11–14 October 2010; pp. 451–454. [Google Scholar]
  12. Hiboux, S.; Muralt, P.; Maeder, T. Domain and Lattice Contributions to Dielectric and Piezoelectric Properties of Pb(Zrx,Ti1-x)O3 Thin Films as a Function of Composition. J. Mater. Res. 1999, 14, 4308–4318. [Google Scholar] [CrossRef] [Green Version]
  13. Gerber, P.; Bottger, U.; Waser, R. Composition Influences on the Electrical and Electromechanical Properties of Lead Zirconate Titanate Thin Films. J. Appl. Phys. 2006, 100, 124105. [Google Scholar] [CrossRef] [Green Version]
  14. Fox, A.; Drawl, B.; Fox, G.R.; Gibbons, B.J.; Trolier-McKinstry, S. Control of Crystallographic Texture and Surface Morphology of Pt/TiO2 Templates for Enhanced PZT Thin Film Texture. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2015, 62, 56–61. [Google Scholar] [CrossRef] [PubMed]
  15. Borman, T.; Ko, S.; Mardilovich, P.; Trolier-McKinstry, S. Development of Crystallographic Texture in Chemical Solution Deposited Lead Zirconate Titanate Seed Layers. J. Am. Ceram. Soc. 2017, 100, 4476–4482. [Google Scholar] [CrossRef]
  16. Borman, T.; Zhu, W.; Wang, K.; Ko, S.; Mardilovich, P.; Trolier-McKinstry, S. Effect of Lead Content on the Performance of Niobium-doped {100} Textured Lead Zirconate Titanate Films. J. Am. Ceram. Soc. 2017, 100, 3558–3567. [Google Scholar] [CrossRef]
  17. Dangi, A.; Cheng, C.Y.; Agrawal, S.; Tiwari, S.; Datta, G.R.; Benoit, R.R.; Pratap, R.; Trolier-McKinstry, S.; Kothapalli, S.R. A Photoacoustic Imaging Device Using Piezoelectric Micromachined Ultrasound Transducers (PMUTs). IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2020, 67, 801–809. [Google Scholar] [CrossRef]
  18. Yi, G.; Wu, Z.; Sayer, M. Preparation of Pb(Zr, Ti)O3 Thin Films By Sol Gel Processing: Electrical, Optical, and Electro-Optic Properties. J. Appl. Phys. 1988, 64, 2717. [Google Scholar] [CrossRef]
  19. Assink, R.A.; Schwartz, R.W. Proton and Carbon-13 NMR Investigations of Lead Zirconate Titanate (Pb(Zr, Ti)O3) Thin-Film Precursor Solutions. Chem. Mater. 1993, 5, 511–517. [Google Scholar] [CrossRef]
  20. Olding, T.; Leclerc, B.; Sayer, M. Processing of Multilayer PZT Coatings for Device Purposes. Integr. Ferroelectr. 1999, 26, 225–241. [Google Scholar] [CrossRef]
  21. Schwartz, R.W. Chemical Solution Deposition of Perovskite Thin Films. Chem. Mater. 1997, 9, 2325–2340. [Google Scholar]
  22. Tuttle, B.A.; Schwartz, R.W. Solution Deposition of Ferroelectric Thin Films. MRS Bull. 1996, 21, 49–54. [Google Scholar] [CrossRef]
  23. Schwartz, R.; Boyle, T.; Lockwood, S.; Sinclair, M.B.; Dimos, D.; Buchheit, C.D. Sol-gel Processing of PZT Thin Films: A Review of The State-of-the-Art and Process Optimization Strategies. Integr. Ferroelectr. 1994, 7, 259–277. [Google Scholar] [CrossRef]
  24. Lockwood, S.J.; Schwartz, R.W.; Tuttle, B.A.; Thomas, E.V. Solution Chemistry Optimization of Sol-Gel Processed PZT Thin Films. MRS OPL 1993, 310, 275–280. [Google Scholar] [CrossRef] [Green Version]
  25. Shepard, J.F., Jr.; Moses, P.J.; Trolier-McKinstry, S. The Wafer Flexure Technique for the Determination of the Transverse Piezoelectric Coefficient (d31) of PZT Thin Films. Sens. Actuators A Phys. 1998, 71, 133–138. [Google Scholar] [CrossRef]
  26. Dangi, A.; Pratap, R. System Level Modeling and Design Maps of PMUTs With Residual Stresses. Sens. Actuators A Phys. 2017, 262, 18–28. [Google Scholar] [CrossRef]
  27. Jiang, X.; Lu, Y.; Tang, H.; Tsai, J.; Ng, E.; Daneman, M.; Boser, B.; Horsley, D.A. Monolithic Ultrasound Fingerprint Sensor. Nature 2017, 3, 17059. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  28. Balasingam, J.A.; Swaminathan, S.; Emadi, A. A Low-Frequency Piezoelectric Micromachined Ultrasonic Transducer based on Multi-User MEMS Process with Enhanced Output Pressure. In Proceedings of the 2020 IEEE International Ultrasonics Symposium (IUS), Las Vegas, NV, USA, 7–11 September 2020; pp. 1–4. [Google Scholar]
  29. Lysne, P.C. Dielectric Breakdown of Shock-Loaded PZT 65/35. J. Appl. Phys. 1972, 44, 577–582. [Google Scholar] [CrossRef]
  30. Cheng, C. Piezoelectric Micromachined Ultrasound Transducers Using Lead Zirconate Titanate Films. Ph.D. Thesis, Department of Materials Science and Engineering, The Pennsylvania State University, University Park, PA, USA, 2021. [Google Scholar]
  31. Akkopru-Akgun, B.; Zhu, W.; Randall, C.A.; Lanagan, M.T.; Trolier-McKinstry, S. Polarity Dependent DC Resistance Degradation and Electrical Breakdown in Nb Doped PZT Films. APL Mat. 2019, 7, 120901. [Google Scholar] [CrossRef] [Green Version]
  32. Dahl-Hansen, R.P.; Tyholdt, F.; Gjessing, J.; Vogl, A.; Wittendorp, P.; Vedum, J.; Tybell, T. On the effect of Water-Induced Degradation of Thin-Film Piezoelectric Microelectromechanical Systems. J. Microelectromech. Syst. 2020, 30, 105–115. [Google Scholar] [CrossRef]
Figure 1. Experimental setup for (a) acoustic and (b) photoacoustic measurements. For pitch catch measurements, the bulk transducer source in (a) was replaced with a glass slide.
Figure 1. Experimental setup for (a) acoustic and (b) photoacoustic measurements. For pitch catch measurements, the bulk transducer source in (a) was replaced with a glass slide.
Sensors 22 05614 g001
Figure 2. Structural characterization of fabricated samples of various PZT compositions (given as Zr/Ti ratio). (a) XRD analysis and (be) are surface FESEM images. (* = substrate peaks).
Figure 2. Structural characterization of fabricated samples of various PZT compositions (given as Zr/Ti ratio). (a) XRD analysis and (be) are surface FESEM images. (* = substrate peaks).
Sensors 22 05614 g002
Figure 3. (a) Permittivity vs. frequency and (b) hysteresis loops of various PZT compositions.
Figure 3. (a) Permittivity vs. frequency and (b) hysteresis loops of various PZT compositions.
Sensors 22 05614 g003
Figure 4. (a) Permittivity (b) e31,f piezoelectric coefficient, and (c) normalized h31,f coefficient as a function of DC bias with various PZT compositions.
Figure 4. (a) Permittivity (b) e31,f piezoelectric coefficient, and (c) normalized h31,f coefficient as a function of DC bias with various PZT compositions.
Sensors 22 05614 g004
Figure 5. (a) Permittivity and (b) piezoelectric coefficient, as a function of DC bias for various average PZT film compositions.
Figure 5. (a) Permittivity and (b) piezoelectric coefficient, as a function of DC bias for various average PZT film compositions.
Sensors 22 05614 g005
Figure 6. Voltage coefficient versus DC bias for various PZT compositions.
Figure 6. Voltage coefficient versus DC bias for various PZT compositions.
Sensors 22 05614 g006
Figure 7. (a) Ultrasound receive signal and (b) normalized photoacoustic signal as a function of DC bias based on setups in Figure 1a,b, respectively. (c) indicates the center frequency (black) and bandwidths (red) as a function of DC bias. The device used for these was PZT 52/48 with 2 μm thickness.
Figure 7. (a) Ultrasound receive signal and (b) normalized photoacoustic signal as a function of DC bias based on setups in Figure 1a,b, respectively. (c) indicates the center frequency (black) and bandwidths (red) as a function of DC bias. The device used for these was PZT 52/48 with 2 μm thickness.
Sensors 22 05614 g007
Figure 8. (a) Picture of the light-guide fiber optic assembly housing the PMUT. The PMUT is facing downward at the agar gel phantom with pencil leads of varying heights. The resulting images are with (b) DC bias and (c) 15 V DC bias operation (75 kV/cm) for 52/48 composition. Log compression was done together with (b,c) for comparison.
Figure 8. (a) Picture of the light-guide fiber optic assembly housing the PMUT. The PMUT is facing downward at the agar gel phantom with pencil leads of varying heights. The resulting images are with (b) DC bias and (c) 15 V DC bias operation (75 kV/cm) for 52/48 composition. Log compression was done together with (b,c) for comparison.
Sensors 22 05614 g008
Figure 9. Images of Figure 8 where log compression was done separately for images (a) without DC bias and (b) with DC bias.
Figure 9. Images of Figure 8 where log compression was done separately for images (a) without DC bias and (b) with DC bias.
Sensors 22 05614 g009
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Cheng, C.; Peters, T.; Dangi, A.; Agrawal, S.; Chen, H.; Kothapalli, S.-R.; Trolier-McKinstry, S. Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition. Sensors 2022, 22, 5614. https://doi.org/10.3390/s22155614

AMA Style

Cheng C, Peters T, Dangi A, Agrawal S, Chen H, Kothapalli S-R, Trolier-McKinstry S. Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition. Sensors. 2022; 22(15):5614. https://doi.org/10.3390/s22155614

Chicago/Turabian Style

Cheng, Christopher, Travis Peters, Ajay Dangi, Sumit Agrawal, Haoyang Chen, Sri-Rajasekhar Kothapalli, and Susan Trolier-McKinstry. 2022. "Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition" Sensors 22, no. 15: 5614. https://doi.org/10.3390/s22155614

APA Style

Cheng, C., Peters, T., Dangi, A., Agrawal, S., Chen, H., Kothapalli, S. -R., & Trolier-McKinstry, S. (2022). Improving PMUT Receive Sensitivity via DC Bias and Piezoelectric Composition. Sensors, 22(15), 5614. https://doi.org/10.3390/s22155614

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop