Next Article in Journal
Research on an Autonomous Localization Method for Trains Based on Pulse Observation in a Tunnel Environment
Previous Article in Journal
Development of a Bitterness Sensor Using Partially Dissociated Amine Compounds
Previous Article in Special Issue
Accurate Range Modeling for High-Resolution Spaceborne Synthetic Aperture Radar
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Communication

Dual-Mode Embedded Impulse-Radio Ultra-Wideband Radar System for Biomedical Applications

by
Wei-Ping Hung
1 and
Chia-Hung Chang
2,*
1
Department of Computer Science and Information Engineering, National United University, Miaoli 306301, Taiwan
2
Department of Electrical Engineering, National Yunlin University of Science and Technology, Yunlin County 64002, Taiwan
*
Author to whom correspondence should be addressed.
Sensors 2024, 24(17), 5555; https://doi.org/10.3390/s24175555
Submission received: 30 June 2024 / Revised: 12 August 2024 / Accepted: 26 August 2024 / Published: 28 August 2024
(This article belongs to the Special Issue Radar Receiver Design and Application)

Abstract

:
This paper presents a real-time and non-contact dual-mode embedded impulse-radio (IR) ultra-wideband (UWB) radar system designed for microwave imaging and vital sign applications. The system is fully customized and composed of three main components, an RF front-end transmission block, an analog signal processing (ASP) block, and a digital processing block, which are integrated in an embedded system. The ASP block enables dual-path receiving for image construction and vital sign detection, while the digital part deals with the inverse scattering and direct current (DC) offset issues. The self-calibration technique is also incorporated into the algorithm to adjust the DC level of each antenna for DC offset compensation. The experimental results demonstrate that the IR-UWB radar, based on the proposed algorithm, successfully detected the 2D image profile of the object as confirmed by numerical derivation. In addition, the radar can wirelessly monitor vital sign behavior such as respiration and heartbeat information.

1. Introduction

Recently, wireless and real-time detection using Doppler radar for healthcare applications has gained significant attention. Biomedical or physiological information such as respiration and heartbeat can be captured by microwave sensing to achieve long-term monitoring in hospitals [1]. An impulse-radio (IR) ultra-wideband (UWB) radar with the characteristics of continuous monitoring and low-power consumption is considered a great candidate to supersede conventional medical electrodes [2]. Several studies on the IR-UWB radars for vital sign detection have been reported in [3,4]. In addition to vital sign application, microwave imaging is widely used in non-invasive inspections, remote sensing, or biomedical images for further diagnostics and treatments. The emitted microwave energy is imposed on a dielectric material, accompanied by the transmitted, reflected, and scattering waves. The probing signal experiences a static object, and the corresponding energy is then reflected back to the radar sensor for the following signal processing. However, material sensing with microwave imaging suffers from inverse scattering issues [5,6], which can degrade the quality of the received signal and image. In addition, the direct current (DC) offset caused by the reflected waves or clutter signals is another crucial issue that can degrade receiving sensitivity not only in microwave imaging but also in vital sign monitoring. Various approaches, such as quadrature radar architecture, DC clutter canceller, DC compensation circuit, and servo feedback loop [7,8,9,10,11,12], have been developed to overcome the DC offset for vital sign applications. Nevertheless, the DC offset with time-variant behavior deviates from different types of material objects. Accordingly, a digital-assisted algorithm with a self-calibration technique is proposed to calibrate the received DC level. This study develops an integrated IR-UWB radar sensor that can simultaneously perform biomedical imaging and vital sign monitoring. The main contributions are summarized as follows:
(1) Instead of using a commercial radar module that serves a single purpose in the previous works, a customized front-end circuit featuring a dual-channel analog signal processing circuit (ASP) has been developed for detecting object contours and monitoring life activities. This compact and portable radar device is integrated into an IR-UWB system for both applications.
(2) An inverse scattering equation with the iterated τ is developed to overcome the nonlinear issue for a digital imaging algorithm, which is implemented in an embedded system. A self-calibration mechanism for the DC offset issue is also incorporated into the algorithm to achieve real-time and adaptive calibration. This concept has the capability to identify material properties by using statistical or artificial intelligence techniques.

2. Design of the Embedded IR-UWB System

The system architecture of the proposed IR-UWB radar system with dual applications is shown in Figure 1. The radar module comprises an RF front-end block, a switching antenna array, a dual-path ASP, and the data processing block. The RF front-end circuit uses a monostatic antenna system to transmit a periodic sequence of nanosecond pulses at a pulse repetition frequency (PRF) and subsequently receives the reflected waves. These waves are demodulated to baseband for additional signal processing. The transmission and reception modes are determined by the T/R switch and Ref. switches, which are controlled by the microcontroller (MCU) [10]. In addition, the antenna is a crucial component of the radar, as it is responsible for detecting the signal transmission and echo reception. An antenna array with eight sub-array antenna units forms a square shape to increase the effective sensing area for image detection. Further details on the design concept and implementation are provided in the following sections.

2.1. Hardware Implementation

The IR-UWB radar detects the object or patient’s chest by correlating the phase delay between the probing and reflected pulses. A fixed sensing distance of Ro between the target and the radiated antenna can be obtained by Ro = (Ctd)/2, where C is the speed of the electromagnetic (EM) wave, and td is the delay time of propagation. To achieve this, a series of probing pulses are provided by the RF pulse generator [13] and coupled to the antenna array. In this work, a probing RF pulse sequence with a resonant frequency of 5.5 GHz, a pulse repetition frequency of 2 MHz, and a radio pulse width of 3 ns is designed. The reflected waves carry the EM energy with image and physiological information back to the radar module, which are then down-converted into intermediate frequency (IF) signals for analog and digital processing. Assuming that a sensing distance of Ro between the target and the radiated antenna is fixed, the captured signal Z(t) can be characterized as follows:
Z ( t ) = E e f f cos ( ϕ ( t ) + Δ ϕ ) ,
where Eeff is an effective amplitude of the receiving signal; the Δϕ represents a fixed phase delay or spreading time between the radiated waves to the target and back to the antenna; and time variant ϕ(t) is an instantaneous phase value, which is determined by the sensing object or life activities behavior. It is observed that Z(t) is expressed as a demodulated signal with the target information of Eeff, ϕ(t), and Δϕ.
Accordingly, from the perspective of microwave imaging, the antenna’s performance is critical in enhancing Eeff and thereby determining the sensing quality. Herein, a monostatic antenna system comprising eight sub-array antenna units controlled by an antenna switch is adopted to achieve larger sensing coverage for object contour. Additionally, considering the antenna design, the conventional resonant antenna is unsuitable for IR-UWB radar detection due to the short-time nanosecond pulse, which might produce ringing phenomena and lead to severe waveform distortion [14]. Thus, a square patch antenna with a terminated load is adopted at the sensor output for nano-pulse radar application [15,16]. Figure 2 illustrates the prototype of the antenna unit design with an input impedance matched network, radiated at a wide frequency band from 5.1 GHz to 5.7 GHz to avoid process variation. The feeding point is located at the center of the module for measurement convenience. Subsequently, this prototype is used as a sub-array unit for planning eight switching antennas in a square shape. A one-to-eight power divider is utilized to equally distribute the transmitting power to each antenna unit.
The captured signal, Z(t), is first demodulated to the IF and then passed through the analog circuits consisting of dual channels for image and vital sign applications, as shown in Figure 3. The processing block for vital sign path is composed of a read-out stage, a low-pass filter (LPF) with Chebyshev behavior, a high-pass filter (HPF), and a gain stage. The ASP block with a bandpass frequency response is designed to reject out-of-band interference [10,17,18]. On the other hand, the image channel bypasses the HPF without DC rejection. The DC components remain for recording each DC value from eight antenna units. These DC values are then compensated by the proposed self-calibration algorithm to avoid image distortion. This allows for the real-time sensing of object profile or vital sign information, which can be wirelessly transmitted via the Bluetooth (BT) interface. The data can then be visualized using the graphical programming language, LabVIEW.

2.2. Data Processing Block and Algorithms

The block diagram of the data processing module and graphical user interface (GUI) is illustrated in Figure 4. A BT module is adopted to communicate with others for functional selection and timing sequence of all controlling switches. The system integrates two modes for vital sign acquisition and imaging reconstruction. The sensing data from the IR-UWB radar are digitized at a speed of 64 Hz/s with a 16-bit resolution analog-to digital converter in all modes. Mode 1 serves the purpose of transmitting the captured raw data of the vital signs from the IR-UWB sensor and its corresponding processed signals via digital filter with the HPF/LPF difference equations. The main algorithm of the vital sign is similar to our previous studies [19,20], except for the power spectral density and peak detection. Instead of searching the peak value of the estimated respiration and heartbeat rates, these signals are directly transmitted to a personal computer (PC), and then the real-time vital signs are observed via the BT module.
Mode 2 is responsible for image reconstruction. The sampled data from the radar sensor are fed into a processing sequence, including DC offset calibration and estimated dielectric constant iterations, τ . The DC level compensation from each antenna unit is initially considered to avoid signal distortion. The procedure for the proposed self-calibration algorithm in the first step is to record the DC values of the eight sub-array antenna units, which are switched in turn to detect a 2D object profile [21,22]. Then, a dataset of DC values can be obtained by n times antenna scanning to evaluate the background information, where n can be determined by the user interface. The mean value is adopted to compensate the required DC level for antenna array. Furthermore, an estimated τ from the sensing object is required to depict object contour. Accordingly, the microwave imaging algorithm is developed to deal with the inverse scattering issue. For the sake of brevity, the complicated integral equations and microwave theory are omitted. The basic concept and the corresponding procedure are listed as follows. The electric scattering field is initially established. An inhomogeneous object is assumed to have permeability of μ 0 and permittivity of ε 0 in free space. The total electric field E is the sum of the incident electric field E i and the scattering field E s . The scattering field in the z-direction can be expressed as follows:
E z s ( r ) = s G ( r , r ) k 0 2 ( ε r ( r ) 1 ) E z ( r ) d s ,
where G ( r , r ) is the 2D Green function, and k 0 is a wave factor in the free space. The relative permittivity ε r and the total electric field E z can be expressed as follows:
ε r ( x , y ) = n = 1 N ε r n P n ( x , y ) ,
E z ( x , y ) = n = 1 N E z n P n ( x , y ) ,
where ε r n and E z n are the permittivity and the total electric field with the nth slicing area in the z-direction, respectively; P n ( x , y ) is the basis function. Similarly, the scattering field and incident field can be expressed as follows:
E z s ( x , y ) = n = 1 N E z n s P n ( x , y ) ,
E z i ( x , y ) = n = 1 N E z n i P n ( x , y ) ,
Based on Equations (3)–(6) and the corresponding electric field integral equation, the scattering field can be further derived in a matrix form as follows:
[ E z s ] = [ G 2 ] [ τ z ] ( [ G 1 ] [ τ z ] [ I ] ) 1 ( E z i ) ,
where ( τ z ) n = ε r n ( x , y ) 1 .
In addition, in order to simplify the integration operation of ( G 1 ) and ( G 2 ) , each small area may be replaced with a circle of equal area. Here, ( G 1 ) and ( G 2 ) is
( G 1 ) m n = j π k 0 a n 2 J 1 ( k 0 a n ) H 0 ( 2 ) ( k 0 ρ m n )    m n = j 2 [ π k 0 a n H 1 ( 2 ) ( k 0 a n ) 2 j ]      m = n
( G 2 ) m n = j π k 0 a n 2 J 1 ( k 0 a n ) H 0 ( 2 ) ( k 0 ρ m n )
Here, J 1 is the first-order Bessel Function of the first kind, a n is the equivalent circle radius of the n-th small region, ( x n , x y ) is the center position of the n-th small region, and
ρ m n = ( x m x n ) 2 + ( y m y n ) 2
Thus, to establish the inverse scattering formula, the nonlinear equation of (7) can be solved by an iteration technique to obtain the dielectric constant [ τ z ] , which can be further expressed as
[ G 2 ] ( [ I ] [ τ ] [ G 1 ] ) 1 [ δ τ ] ( [ I ] [ G 1 ] [ τ ] ) 1 [ E i ] = [ δ E s ] ,
It is observed that the factors of [ δ τ ] and [ δ E s ] are linear dependent, where [ δ τ ] is a diagonal matrix. Equation (11) can be expressed as a linear relationship, as shown in Equation (12)
[ P δ ] [ x δ ] = [ Q δ ]
where [ P δ ] is related to [ G 2 ] ( [ I ] [ τ ] [ G 1 ] ) 1 and ( [ I ] [ G 1 ] [ τ ] ) 1 [ E i ] . [ Q δ ] is related to [ δ E s ] .
Consequently, the initial and actual dielectric constants are assumed to be [ τ i n i t i a l ] and [ τ a c t u a l ] , respectively. In the first iteration, an estimated scattered field [ E ^ s ] can be calculated by substituting [ τ i n i t i a l ] into Equation (7). [ δ E s ] is determined by the difference between [ E ^ s ] and [ E s ] . Herein, [ δ τ ] can be obtained from Equation (8), thereby enabling an updated dielectric constant for the next iteration.
Consequently, the vital sign information and image profile can be obtained by the proposed algorithm, where the corresponding mode selection is determined by the user end. It is noteworthy that only a single antenna is required for vital sign detection because the sensing target focuses on the subject’s chest movement. The continuous time domain and frequency spectral of vital signs can be observed on GUI.

3. Experimental Results

Figure 5 displays photographs of the implemented IR-UWB radar sensor, a square-shaped antenna array, and a one-to-eight power divider/combiner, all of which are fabricated on the standard four-layer FR-4 printed circuit board. The physical sizes of the radar module and a 2 × 2 antenna unit are 9.9 × 5.7 cm2 and 8 × 10 cm2, respectively. A housing package is implemented in the RF front-end block to prevent radiation and coupling issues. The use of eight antenna units with digital switching control helps increase the overall sensing area. Table 1 provides the specifications of the main components used in this customized radar system.
The RF probing pulses are radiated and received by the transmitting/receiving antenna unit. The captured waves carry the desired information based on the Doppler effect. The antenna unit’s performance was evaluated by measuring the return loss, which was found to be lower than −10 dB within the frequency range of 5.10–5.69 GHz, indicating a wideband characteristic, as shown in Figure 6. The radiation patterns of the antenna unit were also measured on both the E- and H-planes, as depicted in Figure 7a and Figure 7b, respectively. The gain patterns at frequencies of 5.1 GHz, 5.35 GHz, and 5.69 GHz exhibited uniformity and a wide directional behavior across the entire frequency band. Thus, eight individual antennas forming an antenna set were adopted for the microwave image, while one of the antenna units was employed for vital sign monitoring.
To verify the radar system, two types of experiments were conducted to demonstrate its concept and capabilities. The first experiment involved detecting the object profile at a fixed nominal distance of 15 cm between the antenna module and the object. The sensing distance can be adjusted by delay line control. Due to the low-power nature of the IR-UWB system, short-range detection was employed. The sensing data were observed on a LabVIEW program GUI. In this experiment, a hand was first placed in front of the antenna to observe the variation in the reflected signal, shown in Figure 8. The speed of the switching antenna unit was defined as 0.5 s with a full scan of 4 s. A background data matrix was first measured to obtain the background threshold value, which was in terms of environment. Each testing environment had different antenna threshold values. Subsequently, the radar received the reflected waves based on the switching antennas within a sensing coverage. These waves were further processed by digital signal processing to form a data matrix of the signal strength distribution. Thus, an L-shaped wooden stick was then used to observe the reflected wave based on the signal degradation curve with a 2D profile distribution. Figure 9 depicts a real-time contour plot with great imaging quality on the GUI frame, where the concentrated area represents the object profile based on the proposed algorithm with the iterated τ calculation.
Additionally, to emphasize the effectiveness of the proposed dual-mode embedded IR-UWB radar system in measuring breathing rate and heartbeat, a single antenna unit sensor was first employed for the continuous monitoring of thoracic movements at a fixed distance of 15 cm between the antenna and the subject. Subsequently, a sensing distance of 30 cm was recorded for comparison. Figure 10 illustrates a photograph of the experimental setup to show the actual positions for vital sign detection. The sensing range can be adjusted by the delay time control [23]. However, due to a trade-off between the transmitted power and propagation loss in free space, the maximum detected range is limited to less than 30 cm. The captured vital signs are processed digitally by the MCU using an LPF with a roll-off frequency at 2 Hz. Figure 11 illustrates the recorded time domain vital signs based on radar sensing for a duration of 1 min. The results reveal that the periodical physiological signal behavior can be successfully detected using the proposed sensor. In addition, the reliability of the sensor was evaluated by calculating the error rate percentage, which was determined by comparing the sensor’s breathing and heart rates with reference values. The reference value for respiration was obtained by counting the number of the subject’s breaths during the test, while the reference heartbeat was measured using a commercial product (Guider smart band, 600Z) that utilized a photoplethysmography (PPG) sensing technique. Accordingly, the measured vital sign results can be compared with the reference data to observe the sensing accuracy. Experiments were conducted with three healthy male volunteers (graduate students aged 23 to 25) who were seated directly in front of the radar sensor. The eight sub-array antenna was set to only a single antenna unit for transmitting/receiving vital signs. Measurements were taken at two distinct sensing distances—15 cm and 30 cm—for each participant. Breathing and heartbeat rates were recorded over 30 s to calculate beats per minute and the corresponding percentage error rate. Figure 12 depicts the error rate percentage for vital sign sensing at different distances and subjects. It is noteworthy that the sensor achieved an error rate of less than 2% at a sensing range of 15 cm. As the detection range increased, the sensing accuracy degraded, but the estimated error rates remained below 4%. These results demonstrate the overall reliability of the implemented sensor.

4. Conclusions

This study presents an integrated IR-UWB radar system including hardware and algorithm realizations for both microwave imaging detection and vital sign sensing. The ASP was utilized to deal with out-of-band interference while maintaining the DC component for the subsequent self-calibration algorithm for object profile detection. Additionally, an inverse scattering equation with the iterated τ was established to overcome the nonlinear issue. The proposed concept has potential to detect material characteristics along with statistical or artificial intelligent techniques for biomedical applications.

Author Contributions

Conceptualization, W.-P.H. and C.-H.C.; methodology, C.-H.C.; software, W.-P.H.; validation, W.-P.H.; formal analysis, W.-P.H. and C.-H.C.; investigation, W.-P.H. and C.-H.C.; resources, W.-P.H.; data curation, W.-P.H. and C.-H.C.; writing—original draft preparation, W.-P.H.; writing—review and editing, C.-H.C.; visualization, W.-P.H. and C.-H.C.; supervision, W.-P.H. and C.-H.C.; project administration, W.-P.H.; funding acquisition, W.-P.H. All authors have read and agreed to the published version of the manuscript.

Funding

This research was funded by the National Science and Technology Council, R.O.C, grant numbers NSTC 112-2221-E-224-008 and NSTC 112-2622-E-224-017.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The original contributions presented in the study are included in the article, further inquiries can be directed to the corresponding authors.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Li, H.; Miura, R. Fundamental study on UWB radar for respiration and heartbeat detection. IEICE Trans. Commun. 2014, 97, 594–601. [Google Scholar] [CrossRef]
  2. Schleicher, B.; Nasr, I.; Trasser, A.; Schumacher, H. IR-UWB Radar Demonstrator for Ultra-Fine Movement Detection and Vital-Sign Monitoring. IEEE Trans. Microw. Theory Tech. 2013, 61, 2076–2085. [Google Scholar] [CrossRef]
  3. Dang, X.; Zhang, J.; Hao, Z. A Non-Contact Detection Method for Multi-Person Vital Signs Based on IR-UWB Radar. Sensors 2022, 22, 6116. [Google Scholar] [CrossRef]
  4. Zhang, X.; Yang, X.; Ding, Y.; Wang, Y.; Zhou, J.; Zhang, L. Contactless Simultaneous Breathing and Heart Rate Detections in Physical Activity Using IR-UWB Radars. Sensors 2021, 21, 5503. [Google Scholar] [CrossRef] [PubMed]
  5. Otto, G.P.; Chew, W.C. Microwave inverse scattering/spl minus/ local shape function imaging for improved resolution of strong scatterers. IEEE Trans. Microw. Theory Tech. 1994, 42, 137–141. [Google Scholar] [CrossRef]
  6. Cui, T.J.; Chew, W.C.; Aydiner, A.A.; Chen, S. Inverse scattering of two-dimensional dielectric objects buried in a lossy earth using the distorted Born iterative method. IEEE Trans. Geosci. Remote Sens. 2001, 39, 339–346. [Google Scholar]
  7. Park, B.-K.; Boric-Lubecke, O.; Lubecke, V.M. Arctangent Demodulation with DC Offset Compensation in Quadrature Doppler Radar Receiver Systems. IEEE Trans. Microw. Theory Tech. 2007, 55, 1073–1079. [Google Scholar] [CrossRef]
  8. Kuo, H.; Lin, C.; Yu, C.; Lo, P.; Lyu, J.; Chou, C.; Chung, H. A Fully Integrated 60-GHz CMOS Direct-Conversion Doppler Radar RF Sensor with Clutter Canceller for Single-Antenna Noncontact Human Vital-Signs Detection. IEEE Trans. Microw. Theory Tech. 2016, 64, 1018–1028. [Google Scholar] [CrossRef]
  9. Yang, Z.; Chiang, Y. Vital Signal Radar with Adaptive Compensation Circuits to Effectively Eliminate DC Offsets. IEEE Microw. Wirel. Compon. Lett. 2018, 28, 88–90. [Google Scholar] [CrossRef]
  10. Chang, C.; Chen, W. Vital-Sign Processing Receiver with Clutter Elimination Using Servo Feedback Loop for UWB Pulse Radar System. IEEE Trans. Very Large Scale Integr. Syst. 2020, 28, 292–296. [Google Scholar] [CrossRef]
  11. Ko, G.-H.; Moon, S.-J.; Kim, S.-H.; Kim, J.-G.; Baek, D. Fully Integrated 24-GHz 1TX-2RX Transceiver for Compact FMCW Radar Applications. Sensors 2024, 24, 1460. [Google Scholar] [CrossRef]
  12. Mercuri, M.; Torfs, T.; Rykunov, M.; Laureti, S.; Ricci, M.; Crupi, F. Analysis of Signal Processing Methods to Reject the DC Offset Contribution of Static Reflectors in FMCW Radar-Based Vital Signs Monitoring. Sensors 2022, 22, 9697. [Google Scholar] [CrossRef]
  13. Lyu, H.; Liu, X.; Babakhani, A. A 100-M/s 2.6-pJ/pulse compact UWB impulse transmitter based on antenna-and-pulse-generator codesign. IEICE Electron. Express 2019, 16, 20190672. [Google Scholar] [CrossRef]
  14. Ghavami, M.; Michael, L.B.; Kohno, R. Ultra Wideband Signals and Systems in Communication Engineering, 2nd ed.; John Wiley & Sons: Hoboken, NJ, USA, 2007. [Google Scholar]
  15. Mahdi, R.A.H.; Taha, S.M.R. Miniaturization of rectangular microstrip patch antenna using topology optimized metamaterial. IEICE Electron. Express 2017, 19, 20170787. [Google Scholar] [CrossRef]
  16. Venkatachalam, D.; Govindasamy, M. A miniaturized planar antenna with defective ground structure for UWB applications. IEICE Electron. Express 2019, 16, 20190672. [Google Scholar] [CrossRef]
  17. Chang, C. A programmable analog baseband filter with DC offset reduction using servo loop feedback. Int. J. Circuit Theory Appl. 2018, 46, 1328–1337. [Google Scholar] [CrossRef]
  18. Kai, Y.; Zhao, S.; Huang, X.; Lv, W. Accurate Doppler radar-based heart rate measurement using matched filter. IEICE Electron. Express 2020, 17, 20200062. [Google Scholar]
  19. Chang, C.; Hu, W. Design and Implementation of an Embedded Cardio-Respiratory Monitoring System for Wheelchair Users. IEEE Embedded Syst. Lett. 2021, 13, 150–153. [Google Scholar] [CrossRef]
  20. Hu, W.; Chang, C.; Yang, G.; Li, C. New Paradigm for Contactless Vital-Sign Sensing Using UWB Radar and Hybrid Optical Wireless Communication. IEEE Embed. Syst. Lett. 2023, 15, 121–124. [Google Scholar] [CrossRef]
  21. Henault, S.; Antar, Y.M.M. Accurate evaluation of mutual coupling for array calibration. In Proceedings of the Computational Electromagnetics International Workshop, Izmir, Turkey, 20–23 July 2009; pp. 34–37. [Google Scholar]
  22. Leshem, A.; Wax, M. Array calibration in the presence of multipath. IEEE Trans. Signal Process. 2000, 48, 53–59. [Google Scholar] [CrossRef]
  23. Chang, C.-L.; Chang, C.-H.; Li, D.-H.; Hu, W.-W.; Lo, Y.-L.; Li, C.-P. An impulse radio duty-cycled radar with ultra-wideband VCO using frequency hopping technique. IEICE Electron. Express 2019, 16, 20190318. [Google Scholar] [CrossRef]
Figure 1. System architecture of the IR-UWB radar for hybrid biomedical applications.
Figure 1. System architecture of the IR-UWB radar for hybrid biomedical applications.
Sensors 24 05555 g001
Figure 2. Structure of antenna unit.
Figure 2. Structure of antenna unit.
Sensors 24 05555 g002
Figure 3. Dual-channel ASP for microwave imaging and vital sign data.
Figure 3. Dual-channel ASP for microwave imaging and vital sign data.
Sensors 24 05555 g003
Figure 4. Dual-channel ASP for microwave imaging and physiological data.
Figure 4. Dual-channel ASP for microwave imaging and physiological data.
Sensors 24 05555 g004
Figure 5. (a) Antenna array module with eight 2 × 2 square patch antenna units and (b) backside view of the implemented radar sensor.
Figure 5. (a) Antenna array module with eight 2 × 2 square patch antenna units and (b) backside view of the implemented radar sensor.
Sensors 24 05555 g005
Figure 6. Measured return loss of the antenna unit.
Figure 6. Measured return loss of the antenna unit.
Sensors 24 05555 g006
Figure 7. Measured (a) E-plane and (b) H-plane of the antenna unit.
Figure 7. Measured (a) E-plane and (b) H-plane of the antenna unit.
Sensors 24 05555 g007
Figure 8. Recorded real-time signal with a periodical full scan antenna.
Figure 8. Recorded real-time signal with a periodical full scan antenna.
Sensors 24 05555 g008
Figure 9. Profile plot of the L-shaped object.
Figure 9. Profile plot of the L-shaped object.
Sensors 24 05555 g009
Figure 10. Radar experimental setup for vital sign detection.
Figure 10. Radar experimental setup for vital sign detection.
Sensors 24 05555 g010
Figure 11. Measured vital sign results from the IR-UWB sensor.
Figure 11. Measured vital sign results from the IR-UWB sensor.
Sensors 24 05555 g011
Figure 12. Error rate percentage of vital sign detection with different subjects and distances.
Figure 12. Error rate percentage of vital sign detection with different subjects and distances.
Sensors 24 05555 g012
Table 1. Specifications of components used in radar board.
Table 1. Specifications of components used in radar board.
BlockManufacturerSpecifications
RF switchAnalog DevicesDC—8 GHz, 1.2 dB insertion loss
Driving AmplifierAnalog DevicesDC to 6 GHz, 20 dB gain
LNAAvago1.5–8 GHz, 23 dB gain, 1.6 dB NF
DownconverterAnalog Devices4.5–6 GHz, 30 dB LO-RF isolation, 6.5 dB conversion loss
Low Pass FilterMini-CircuitsDC—400 MHz, low insertion loss
Low-Dropout RegulatorTexas InstrumentsAvailable in 5 V; Output current: 800 mA
OPAAnalog Devices1.8–5 V supply and rail to rail
MicrocontrollerTexas InstrumentsMaximum frequency of 80 MHz
BluetoothITead StudioHC05 Bluetooth to serial port
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Hung, W.-P.; Chang, C.-H. Dual-Mode Embedded Impulse-Radio Ultra-Wideband Radar System for Biomedical Applications. Sensors 2024, 24, 5555. https://doi.org/10.3390/s24175555

AMA Style

Hung W-P, Chang C-H. Dual-Mode Embedded Impulse-Radio Ultra-Wideband Radar System for Biomedical Applications. Sensors. 2024; 24(17):5555. https://doi.org/10.3390/s24175555

Chicago/Turabian Style

Hung, Wei-Ping, and Chia-Hung Chang. 2024. "Dual-Mode Embedded Impulse-Radio Ultra-Wideband Radar System for Biomedical Applications" Sensors 24, no. 17: 5555. https://doi.org/10.3390/s24175555

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop