Next Article in Journal
Characterization of Defocused Coherent Imaging Systems with Periodic Objects
Previous Article in Journal
MPE-HRNetL: A Lightweight High-Resolution Network for Multispecies Animal Pose Estimation
Previous Article in Special Issue
Best Practices for Measuring the Modulation Transfer Function of Video Endoscopes
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study

1
Department of Medical Imaging, IMAGINE UR UM 103, Montpellier University, Nimes University Hospital, Bd Prof Robert Debré, CEDEX 9, 30029 Nîmes, France
2
Medical Physics and Radiation Protection Department, APHP Lariboisière University Hospital, 75010 Paris, France
3
Department of Radiology, Georges Pompidou European Hospital, Paris Cité University, APHP, 75015 Paris, France
4
Department of Radiodiagnostics, CHU Angers, 4 Rue Larrey, 49933 Angers, France
*
Author to whom correspondence should be addressed.
Sensors 2024, 24(21), 6883; https://doi.org/10.3390/s24216883
Submission received: 13 September 2024 / Revised: 15 October 2024 / Accepted: 23 October 2024 / Published: 26 October 2024
(This article belongs to the Special Issue Medical Imaging and Sensing Technologies)

Abstract

:
An image-quality CT phantom was scanned with three different 3D X-ray imaging guidance devices in the operating theatre: O-Arm, Loop-X, and Airo TruCT. Default acquisition and reconstruction parameters for lumbar spine procedures were used on each device. The tube current was set to a dose level of around 27 mGy. A task-based image quality assessment was performed by calculating the noise power spectrum (NPS) and task transfer function (TTF). A detectability index (d’) was calculated for three simulated bone lesions. The noise magnitude of the O-Arm was higher than the Airo TruCT, and the Loop-X had higher noise than the Airo TruCT. The highest average NPS frequency was for the O-Arm images, and the lowest was for the Loop-X. The TTFs at 50% values were similar for the Airo TruCT and Loop-X devices. Compared to Airo TruCT, the TTF at 50% value increased with the O-Arm by 53.12% and 41.20% for the Teflon and Delrin inserts, respectively. Compared to Airo TruCT, the d’ value was lower with Loop-X by −26.73%, −27.02%, and −23.95% for lytic lesions, sclerotic lesions, and high-density bone, respectively. Each 3D-imaging spine surgery guidance device has its own strengths and weaknesses in terms of image quality. Cone-beam CT systems apparently offer the best compromise between noise and spatial resolution for spine surgery.

1. Introduction

Three-dimensional (3D) pre- or intra-operative imaging is a fast-growing technique mainly used in neurosurgery and orthopedic/trauma surgery [1,2]. Over the years, improvements in technology have led to the development of 3D image reconstruction algorithms and flat-panel detectors, which have significantly improved image quality. Unlike two-dimensional (2D) projection images, which offer a limited field of view, spatial resolution, and noise, 3D images can better depict challenging anatomical features. The main benefits of intraoperative 3D imaging have been demonstrated in spine surgery to tackle the challenge of joint surface reconstructions [3,4]. Studies have shown that 3D imaging increases the accuracy of pedicle screw or implant placement and reduces the number of surgical revisions and operative time compared to conventional C-arms [5,6]. Similarly, the use of navigation seems to be particularly beneficial for procedures which are difficult to visualize [7,8]. Other applications have also been validated, such as stereotaxic brain surgery [9], treatment of intracranial aneurysms [10], epilepsy [3], or complex joint fractures including the sacroiliac joint, wrist [11], tibia [12], acetabulum [13,14], and syndesmosis [15]. Thus, 3D imaging has become very common in minimally invasive surgery and provides a basis for surgical planning, procedure accuracy, and patient safety.
Many mobile 3D-imaging devices are available in operating rooms. Each system offers a unique technical approach and has its own advantages and pitfalls. These systems can be classified into three categories: C-arm geometry with ~200° rotation, O-ring geometry with 360° rotation, and Computed Tomography (CT) scanners. C-arm and O-ring devices use a cone-shaped beam to generate a 3D CT-like dataset referred to as cone-beam CT (CBCT). The O-arm® (Medtronic, Minneapolis, MN, USA), Airo TruCT® (Stryker, Kalamazoo, MI, USA), and Loop-XTM (Brainlab AG, Munich, Germany) are three mobile systems designed and used for 3D imaging in operating rooms. The O-arm system is based on an X-ray tube coupled with a flat-panel detector, which rotates through 360 degrees to acquire CBCT images. The Airo TruCT system is a 32-sclice CT scanner combining self-propelled mobility with a large inner bore. The Loop-X is a mobile imaging robot equipped with an independent moving imaging source and detector panels to allow for flexible patient positioning, and non-isocentric imaging to focus on the region of interest. It features 2D planar, 3D-CBCT, and fluoroscopic imaging.
Several studies have shown the benefits of the three systems in image-guided surgery to improve patient safety [5,16,17,18,19]. Additionally, a few studies have been concerned about patient and staff exposure to ionizing radiation and subjective image quality [1,19,20,21,22,23,24]. However, to the best of our knowledge, no previous studies have ever assessed the image quality of these three devices objectively with respect to radiation dose using a task-based image quality evaluation.
This work aimed to compare three different mobile 3D intraoperative systems in terms of image quality and possible dose reduction. To complete this study, a task-based image quality assessment was conducted using an image quality phantom for spine applications. We examined image resolution using task transfer function (TTF), noise using the noise power spectrum (NPS), and detectability (d’).

2. Materials and Methods

Figure 1 shows a summary diagram of the study objective and methodological steps.

2.1. Acquisition and Reconstruction Parameters

A Catphan® 600 image quality CT phantom (The Phantom Laboratory, Salem, NY, USA) was scanned with the three 3D X-ray imaging guidance devices used in the operating theatre: the O-Arm (Medtronic, Minneapolis, MN, USA), the Loop-X (Brainlab AG, Germany), and the Airo® TruCT® MobiCT-32 (Stryker, MI, USA). Helical acquisitions were performed with the Airo TruCT scanner with a pitch factor of 1.415. The technical specifications of these devices are given in Table 1. Clinical acquisition and reconstruction parameters for lumbar spine procedures were used for each device (Table 2).
For Airo TruCT, an air calibration was performed before acquisition according to the manufacturers ‘requirements.
Clinical acquisition and reconstruction parameters for lumbar spine procedures were used for each device (Table 2).
The tube current was set to obtain a weighted CT dose index (CTDIw) or volume CT dose index (CTDIvol) of around 27 mGy. This value is the standard national dose for lumbar spine CT scans [25]. Acquisitions were repeated three times on each device to increase the accuracy of the image quality analyses.

2.2. Task-Based Image Quality Assessment

A task-based image quality assessment was performed using iQMetrix-CT software (version 1.1, SFPM) [26] to assess noise magnitude and texture using the NPS and spatial resolution using the TTF [27]. Images from the three acquisitions were pooled per equipment to calculate the NPS and TTF. This increased the number of images used to calculate these metrics and reduced the impact of noise on TTF calculations [28]. Finally, the detectability index (d′) was computed to assess the ability of the radiologist to detect three types of lesions encountered in lumbar spine treatment.
The choice of these metrics was based on their ability to characterize image quality in conditions more similar to clinical use. The NPS allows for the assessment of the average noise in the image, as well as its frequency distribution, thus characterizing image texture, which can also have an impact on image quality through smoothing effects. TTF, on the other hand, enables spatial resolution to be assessed for a contrast and dose level closer to those used in clinical practice. Finally, detectability simulates the surgeon’s ability to perform a given clinical task by integrating into the calculation the impact of noise (NPS), spatial resolution (TTF), the characteristics of the clinical task (contrast, lesion shape, and size), and a visual response function that takes account of image viewing conditions. These metrics have been widely used for CT image quality evaluations and to compare CT and CBCT image quality, as well [29,30].

2.2.1. Noise Power Spectrum

For each device, the NPS was computed on the uniform module (CTP 486) of the Catphan 600 phantom using 30 consecutive slices. Four square regions of interest (ROIs) measuring 92 × 92 pixels with a margin of 30 pixels were placed in this uniform module (Figure 2A), and the NPS was calculated according to the following formula:
N P S 2 D f x , f y = x y L x L y 1 N R O I i = 1 N R O I F F T 2 D R O I i x , y F I T i ( x , y ) 2
where Δx and Δy are the pixel sizes in the x- and y-directions; FFT is the Fast Fourier Transform; Lx and Ly are the lengths of the ROIs in the x- and y-directions; NROI is the number of ROIs; ROIi (x,y) is the mean pixel value measured for a ROI at the position (x, y); and FITi (x,y) is a second-order polynomial fit of ROIi (x,y).
The noise magnitude and the average spatial frequency (fav) were calculated to quantify the noise level and noise texture, respectively. The following formula was used to compute the fav values:
f a v = f . N P S ( f ) d f N P S f d f
where f is the radial spatial frequency and NPS(f) is the radially re-binned/average 1D NPS [28].

2.2.2. Task-Based Transfer Function

The TTF was assessed using Delrin and Teflon inserts on the sensitometry module (CT404) of the Catphan phantom (Figure 2B) according to the methodology reported by Richard et al. [27].
A circular ROI was placed around the inserts, and a circular-edge technique was used to measure the Edge Spread Function (ESF) computed using 15 consecutive slices. The ESF was obtained by calculating the radius of each pixel from the centre of each pixel of the insert. The Line Spread Function (LSF) was calculated by derivation from the ESF. Finally, the TTF was computed from the LSF normalized Fourier transformation.

2.2.3. Detectability Index

A detectability index (d’) was calculated using a non-prewhitening observer model with an eye filter (NPWE):
d N P W E 2 = W ( u , v ) 2 . T T F u , v 2 . E u , v 2 d u d v 2 W u , v 2 . T T F u , v 2 . N P S u , v . E ( u , v ) 4 d u d v
where u and v are spatial frequencies in the x- and y-directions, E is the eye filter that models the human visual system sensitivity to different spatial frequencies [31], and W(u,v) is the task function defined as:
W = F h 1 x , y h 2 x , y
where F is the Fourier transform, and h1 (x,y) and h2 (x,y) correspond to the object present and object absent hypotheses, respectively.
The eye filter was modelled according to the visual response function using a zoom factor of 1.5 and a 500 mm viewing distance.
Two task functions measuring 5 mm in diameter were simulated to represent the three types of bone lesion. The task function was assumed to represent a circular signal, with a contrast value of 100 HU for the lytic bone lesion, 350 HU for the sclerotic lesion, and 900 HU for high-density bone. The TTF results of the Delrin insert were used for the lytic and sclerotic lesions, and the TTF results of the Teflon insert were used for the high-density bone. These parameters had already been defined thanks to a previous study [32].

3. Results

3.1. Noise Power Spectrum

The NPS curves are presented in Figure 3, and the noise magnitude, fav, and fpeak results in Table 3.
The highest noise magnitude value was found for the O-Arm, and the lowest for the Airo TruCT. The O-Arm’s noise magnitude was 3.79 times higher than the Airo TruCT, and the Loop-X had 1.62 times higher noise than the Airo TruCT.
Concerning the fav results, the highest value was observed for the O-Arm images and the lowest for the Loop-X. Compared to the Airo TruCT, the fav was 2.13 times higher with the O-Arm and 1.29 times lower with the Loop-X. Similar results were found for fpeak, with the highest value for the O-Arm and the lowest value for Loop-X.

3.2. Task-Based Transfer Function

The results for TTF are depicted in Figure 4 and Table 4.
For all devices, the TTF50 values were similar for both Teflon and Delrin inserts.
Compared to the Airo TruCT, the TTF50 value increased with the O-Arm by 53.12% and 41.20% for the Teflon and Delrin inserts, respectively. A smaller increase was found with the Loop-X compared to Airo TruCT, with 21.8% and 9.0% for Teflon and Delrin, respectively. TTF50 values were higher with the O-Arm compared to Loop-X for both inserts.

3.3. Detectability Index

The results of the detectability index for the three simulated lesions are shown in Table 5 for the three imaging devices.
For all three devices, the highest d’ values were found with high-density bone and the lowest values were observed with lytic lesions.
Compared to the Airo TruCT, the d’ value decreased with Loop-X by −26.73, −27.02, and −23.95% for lytic lesions, sclerotic lesions, and high-density bone. Higher differences were found with the O-Arm, decreasing the d’ by −76.80%, −76.82%, and −76.18%, respectively.

4. Discussion

In this phantom study, the image quality of three intraoperative 3D X-ray imaging devices used in the operating theatre was compared. For this purpose, a task-based image quality assessment was made, including calculations of the noise power spectrum (NPS), task transfer function (TTF), and detectability of three simulated bone lesions. The results show that the Airo TruCT provided a lower image noise level and the O-Arm device provided a higher level. However, the O-Arm presented a higher average frequency (fav) than the Loop-X, which provided the lowest value for fav. Similar spatial resolution values were found for the Airo TruCT and Loop-X, but higher values were observed for the O-Arm. The detectability index values were highest with the Airo TruCT and lowest with the O-Arm for all of the simulated lesions.
The results show that the lowest noise level was obtained with the Airo TruCT. This is mainly explained by the low contribution of scattered radiation to image formation for this scanner, compared to the other two CBCT devices. In fact, the Airo TruCT beam is a 32 × 1 mm wide fan beam, whereas the beam used on CBCT is a wide beam exceeding 20 cm at the isocenter. This wider beam increases the ratio between the quantity of scattered photons and primary photons reaching the detector, which is around 0.2 for a scanner with 40 mm collimation and 3 for a CBCT [33]. As scattered photons contribute to image quality degradation and increased image noise, reducing them helps to reduce noise. In addition, the results show that the Loop-X system provides less noise than the O-Arm, but has a lower average NPS frequency. Low fav reflects a change in noise texture, sometimes producing very smooth and blotchy images [34,35,36,37]. This difference in texture can be linked mainly to image post-processing and denoising methods. These methods differ between the two devices, resulting in moderate denoising and image texture preservation for the O-Arm and significant denoising with an impact on image texture for the Loop-X.
The images with the best spatial resolution were those obtained with the O-Arm, whereas the lowest resolution was observed with the Airo TruCT. These results may be explained by the intrinsic characteristics of these devices. The main factors affecting spatial resolution are focal spot dimensions, detector types, and detector matrix sizes. As far as focal spot size is concerned, both CBCT devices allow for the use of a smaller size than the Airo TruCT device. A small focal spot improves the spatial resolution by limiting the effect of geometric blurring [38]. The differences in spatial resolution observed between the O-Arm and the Loop-X may be explained by the differences in detection technology. Indeed, the O-Arm uses amorphous silicon with a high matrix size (2048 × 1536), whereas the Loop-X uses a 1024 × 1024 matrix, which may explain the better TTF50 offered by the O-Arm.
For the detectability index, the highest detectability was observed with the Airo TruCT and the lowest with the O-Arm for the three defined clinical tasks. In fact, the d’ outcome is a consequence of the NPS and TTF results. However, as the NPS has a high impact on calculating the detectability index [39], the device with the lowest noise level was the one with the best detectability.
From a clinical viewpoint, the results obtained in this study show that the three X-ray imaging guidance systems for spine surgery each have their own advantages and drawbacks in terms of image quality. To detect low-contrast lesions such as lytic lesions, the Airo TruCT seems to be the preferred system thanks to its low noise level, although it also has the lowest spatial resolution. For high-contrast lesions such as sclerotic or bone lesions, the Airo TruCT system still offers the best detectability thanks to its low noise level, but, given the high contrast, noise in these cases becomes less of a problem, and spatial resolution becomes the main criterion. In this instance, then, the O-Arm seems to give better results in terms of spatial resolution than the Loop-X, but the latter has a lower noise level. It should also be pointed out that, compared to the Airo TruCT device, CBCT systems are more sensitive to artefacts such as scatter artefacts, cone-beam artefacts, and beam-hardening artefacts [40], particularly in the presence of dense materials. In addition to image quality performance, it is important to consider other parameters related to the operating theatre environment, as stated by Rossi et al. (2022) [41]. The device’s ease of mobility, maneuverability, compatibility with navigation software, and availability of optimization tools are all important criteria to take into account. CBCT systems like the Loop-X have the advantage of greater clearance and easier handling than the other two systems. In terms of tools for optimizing patient dose and image quality, the Airo TruCT scanner is equipped with an intensity modulation system and a metal artefact-correction algorithm, depending on the software version. The CBCT systems compensate for the absence of modulation by offering a wide range of acquisition and reconstruction protocols so that operators can choose the one best suited to the procedure. All three devices are suitable for all types of spine surgery, with or without navigation [42,43,44]. Nevertheless, the results show that performance in terms of noise and spatial resolution differed, which may advantage devices with high spatial resolution, such as the AIRO and LoopX, for bone surgery, such as lumbar spine surgery, given the high contrast of bone tissue and the need for good spatial resolution.
This study has certain limitations. First, a non-anthropomorphic phantom with anatomical structures that differ from those of a real patient was used for the evaluation of image quality. The image quality assessment was carried out on an image quality phantom to characterize the intrinsic performance of each piece of equipment under identical conditions using objective metrics. The generalization of these results and their confirmation in clinical practice requires further evaluations of patient images by physicians through subjective assessments of image quality. This might highlight the constraint of inter-physician and inter-patient variability, which could be resolved by a large sample of patients. Moreover, this study needs to be supplemented by evaluations of other anatomical regions, particularly those involved in brain surgery. Finally, the radiation dose levels used are not representative of the levels used clinically for spinal surgery, particularly with CBCT. However, in the absence of data on clinical dose levels with CBCT, the national DRL values for CT scans were chosen to compare the image quality of the three devices at equivalent dose levels.

5. Conclusions

In conclusion, each 3D imaging device for spine surgery guidance has its own strengths and weaknesses in terms of image quality. For an equivalent patient dose, CBCT systems offer the best compromise between noise and spatial resolution for spine surgery.

Author Contributions

Conceptualization, D.D., L.H.-B., and C.S.; methodology, D.D., L.H.-B., and C.S.; software, D.D. and J.G.; validation, all authors.; formal analysis, D.D. and L.H.-B.; investigation, D.D., L.H.-B., and C.S.; resources, D.D., L.H.-B., C.S., I.F., C.V.N.T., P.P., and J.-P.B.; data curation, D.D.; writing—original draft preparation D.D. and L.H.-B.; writing—review and editing, All authors.; visualization, Not concerned.; supervision, D.D., L.H.-B., and J.-P.B.; project administration, D.D. and L.H.-B.; funding acquisition, no funding. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Data are available on request from the corresponding author.

Acknowledgments

We thank Teresa Sawyers, Medical Writer at the BESPIM, Nîmes University Hospital, France, for her help in editing this manuscript.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Tonetti, J.; Boudissa, M.; Kerschbaumer, G.; Seurat, O. Role of 3D intraoperative imaging in orthopedic and trauma surgery. Orthop. Traumatol. Surg. Res. 2020, 106, S19–S25. [Google Scholar] [CrossRef] [PubMed]
  2. Keil, H.; Trapp, O. Fluoroscopic imaging: New advances. Injury 2022, 53 (Suppl. S3), S8–S15. [Google Scholar] [CrossRef] [PubMed]
  3. Kirnaz, S.; Gebhard, H.; Wong, T.; Nangunoori, R.; Schmidt, F.A.; Sato, K.; Härtl, R. Intraoperative image guidance for cervical spine surgery. Ann. Transl. Med. 2021, 9, 93. [Google Scholar] [CrossRef] [PubMed]
  4. Navarro-Ramirez, R.; Lang, G.; Lian, X.; Berlin, C.; Janssen, I.; Jada, A.; Alimi, M.; Härtl, R. Total Navigation in Spine Surgery; A Concise Guide to Eliminate Fluoroscopy Using a Portable Intraoperative Computed Tomography 3-Dimensional Navigation System. World Neurosurg. 2017, 100, 325–335. [Google Scholar] [CrossRef]
  5. Ganguly, R.; Minnema, A.; Singh, V.; Grossbach, A. Retrospective analysis of pedicle screw accuracy for patients undergoing spinal surgery assisted by intraoperative computed tomography (CT) scanner AIRO® and BrainLab© navigation. Clin. Neurol. Neurosurg. 2020, 198, 106113. [Google Scholar] [CrossRef]
  6. Mezger, U.; Jendrewski, C.; Bartels, M. Navigation in surgery. Langenbecks Arch. Surg. 2013, 398, 501–514. [Google Scholar] [CrossRef]
  7. Hussain, I.; Cosar, M.; Kirnaz, S.; Schmidt, F.A.; Wipplinger, C.; Wong, T.; Härtl, R. Evolving Navigation, Robotics, and Augmented Reality in Minimally Invasive Spine Surgery. Glob. Spine J. 2020, 10 (Suppl. S2), 22S–33S. [Google Scholar] [CrossRef]
  8. Hussain, I.; Schmidt, F.A.; Kirnaz, S.; Wipplinger, C.; Schwartz, T.H.; Härtl, R. MIS approaches in the cervical spine. J. Spine Surg. 2019, 5, S74–S83. [Google Scholar] [CrossRef]
  9. Weise, L.; Eibach, S.; Seifert, V.; Setzer, M. Intraoperative 3D fluoroscopy in stereotactic surgery. Acta Neurochir. 2012, 154, 815–821. [Google Scholar] [CrossRef]
  10. Westermaier, T.; Linsenmann, T.; Homola, G.A.; Loehr, M.; Stetter, C.; Willner, N.; Ernestus, R.I.; Solymosi, L.; Vince, G.H. 3D rotational fluoroscopy for intraoperative clip control in patients with intracranial aneurysms—Assessment of feasibility and image quality. BMC Med. Imaging 2016, 16, 30. [Google Scholar] [CrossRef]
  11. Hammerle, D.; Osterhoff, G.; Allemann, F.; Werner, C.M.L. Comparison of intraoperative 2D vs. 3D imaging in open reduction and fixation of distal radius fractures. Eur. J. Trauma Emerg. Surg. 2020, 46, 557–563. [Google Scholar] [CrossRef] [PubMed]
  12. Vetter, S.Y.; Euler, F.; von Recum, J.; Wendl, K.; Grützner, P.A.; Franke, J. Impact of Intraoperative Cone Beam Computed Tomography on Reduction Quality and Implant Position in Treatment of Tibial Plafond Fractures. Foot Ankle Int. 2016, 37, 977–982. [Google Scholar] [CrossRef] [PubMed]
  13. Keil, H.; Beisemann, N.; Schnetzke, M.; Vetter, S.Y.; Grützner, P.A.; Franke, J. First experiences with the Airo mobile intraoperative CT scanner in acetabular surgery—An analysis of 10 cases. Int. J. Med. Robot. Comput. Assist. Surg. 2019, 15, e1986. [Google Scholar] [CrossRef] [PubMed]
  14. Amer, A.; Pomme, J.; Moreau, E.; Camille, R.; Guillaume, R. Intraoperative Cone Beam Tomography and Navigation for Displaced Acetabular Fractures: A Comparative Study. J. Orthop. Trauma 2018, 32, 612–616. Available online: https://journals.lww.com/jorthotrauma/Abstract/2018/12000/Intraoperative_Cone_Beam_Tomography_and_Navigation.4.aspx (accessed on 1 August 2023).
  15. Franke, J.; von Recum, J.; Suda, A.J.; Grützner, P.A.; Wendl, K. Intraoperative three-dimensional imaging in the treatment of acute unstable syndesmotic injuries. J. Bone Jt. Surg. Am. 2012, 94, 1386–1390. [Google Scholar] [CrossRef]
  16. Bichsel, O.; Oertel, M.F.; Stieglitz, L.H. Mobile intraoperative CT-assisted frameless stereotactic biopsies achieved single-millimeter trajectory accuracy for deep-seated brain lesions in a sample of 7 patients. BMC Neurol. 2021, 21, 285. [Google Scholar] [CrossRef]
  17. Jermakowicz, W.J.; Diaz, R.J.; Cass, S.H.; Ivan, M.E.; Komotar, R.J. Use of a Mobile Intraoperative Computed Tomography Scanner for Navigation Registration During Laser Interstitial Thermal Therapy of Brain Tumors. World Neurosurg. 2016, 94, 418–425. [Google Scholar] [CrossRef]
  18. Hecht, N.; Kamphuis, M.; Czabanka, M.; Hamm, B.; König, S.; Woitzik, J.; Synowitz, M.; Vajkoczy, P. Accuracy and workflow of navigated spinal instrumentation with the mobile AIRO® CT scanner. Eur. Spine J. 2016, 25, 716–723. [Google Scholar] [CrossRef]
  19. Wojdyn, M.; Pierzak, O.; Zapałowicz, K.; Radek, M. Use of O-arm with neuronavigation in percutaneous vertebroplasty reduces the surgeon’s exposure to intraoperative radiation. Arch. Med. Sci. 2021, 17, 113–119. [Google Scholar] [CrossRef]
  20. Casiraghi, M.; Scarone, P.; Bellesi, L.; Piliero, M.A.; Pupillo, F.; Gaudino, D.; Fumagalli, G.; Del Grande, F.; Presilla, S. Effective dose and image quality for intraoperative imaging with a cone-beam CT and a mobile multi-slice CT in spinal surgery: A phantom study. Phys. Med. 2021, 81, 9–19. [Google Scholar] [CrossRef]
  21. Foster, N.; Shaffrey, C.; Buchholz, A.; Turner, R.; Yang, L.Z.; Niedzwiecki, D.; Goode, A. Image Quality and Dose Comparison of 3 Mobile Intraoperative Three-Dimensional Imaging Systems in Spine Surgery. World Neurosurg. 2022, 160, e142–e151. [Google Scholar] [CrossRef] [PubMed]
  22. Keil, H.; Luxenhofer, M.; Vetter, S.Y.; Beisemann, N.; Grützner, P.A.; Franke, J. Evaluation of image quality and assessability of a new flat-panel 3D C-arm compared to mobile and fixed computed tomography in posterior spinal fixation. Int. J. Med. Robot. Comput. Assist. Surg. 2021, 17, e2181. [Google Scholar] [CrossRef] [PubMed]
  23. Nachabe, R.; Strauss, K.; Schueler, B.; Bydon, M. Radiation dose and image quality comparison during spine surgery with two different, intraoperative 3D imaging navigation systems. J. Appl. Clin. Med. Phys. 2019, 20, 136–145. [Google Scholar] [CrossRef] [PubMed]
  24. Lange, J.; Karellas, A.; Street, J.; Eck, J.C.; Lapinsky, A.; Connolly, P.J.; DiPaola, C.P. Estimating the effective radiation dose imparted to patients by intraoperative cone-beam computed tomography in thoracolumbar spinal surgery. Spine 2013, 38, E306–E312. [Google Scholar] [CrossRef] [PubMed]
  25. Arrêté du 23 mai 2019 Portant Homologation de la Décision n° 2019-DC-0667 de l’Autorité de Sûreté Nucléaire du 18 Avril 2019 Relative aux Modalités D’évaluation des Doses de Rayonnements Ionisants Délivrées aux Patients lors d’un acte de Radiologie, de Pratiques Interventionnelles Radioguidées ou de Médecine Nucléaire et à la Mise à Jour des Niveaux de Référence Diagnostiques Associés. Available online: https://www.legifrance.gouv.fr/loda/id/JORFTEXT000038529178 (accessed on 23 May 2019).
  26. Greffier, J.; Barbotteau, Y.; Gardavaud, F. iQMetrix-CT: New software for task-based image quality assessment of phantom CT images. Diagn. Interv. Imaging 2022, 103, 555–562. [Google Scholar] [CrossRef]
  27. Richard, S.; Husarik, D.B.; Yadava, G.; Murphy, S.N.; Samei, E. Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms. Med. Phys. 2012, 39, 4115–4122. [Google Scholar] [CrossRef]
  28. Samei, E.; Bakalyar, D.; Boedeker, K.L.; Brady, S.; Fan, J.; Leng, S.; Myers, K.J.; Popescu, L.M.; Ramirez Giraldo, J.C.; Ranallo, F.; et al. Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233. Med. Phys. 2019, 46, e735–e756. [Google Scholar] [CrossRef]
  29. Verdun, F.R.; Racine, D.; Ott, J.G.; Tapiovaara, M.J.; Toroi, P.; Bochud, F.O.; Veldkamp, W.J.H.; Schegerer, A.; Bouwman, R.W.; Giron, I.H.; et al. Image quality in CT: From physical measurements to model observers. Phys. Medica 2015, 31, 823–843. [Google Scholar] [CrossRef]
  30. Vorbau, R.; Hulthén, M.; Omar, A. Task-based image quality assessment of an intraoperative CBCT for spine surgery compared with conventional CT. Phys. Med. 2024, 124, 103426. [Google Scholar] [CrossRef]
  31. Eckstein, M.P.; Bartroff, J.L.; Abbey, C.K.; Whiting, J.S.; Bochud, F.O. Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks. Opt. Express 2003, 11, 460. [Google Scholar] [CrossRef]
  32. Greffier, J.; Frandon, J.; Pereira, F.; Hamard, A.; Beregi, J.P.; Larbi, A.; Omoumi, P. Optimization of radiation dose for CT detection of lytic and sclerotic bone lesions: A phantom study. Eur. Radiol. 2020, 30, 1075–1078. [Google Scholar] [CrossRef] [PubMed]
  33. Gonçalves, O.D.; Boldt, S.; Nadaes, M.; Devito, K.L. Evaluating the scattered radiation intensity in CBCT. Radiat. Phys. Chem. 2018, 144, 159–164. [Google Scholar] [CrossRef]
  34. Willemink, M.J.; de Jong, P.A.; Leiner, T.; de Heer, L.M.; Nievelstein, R.A.; Budde, R.P.; Schilham, A.M. Iterative reconstruction techniques for computed tomography Part 1: Technical principles. Eur. Radiol. 2013, 23, 1623–1631. [Google Scholar] [CrossRef] [PubMed]
  35. Geyer, L.L.; Schoepf, U.J.; Meinel, F.G.; Nance, J.W., Jr.; Bastarrika, G.; Leipsic, J.A.; Paul, N.S.; Rengo, M.; Laghi, A.; De Cecco, C.N. State of the Art: Iterative CT Reconstruction Techniques. Radiology 2015, 276, 339–357. [Google Scholar] [CrossRef] [PubMed]
  36. Greffier, J.; Boccalini, S.; Beregi, J.P.; Vlassenbroek, A.; Vuillod, A.; Dupuis-Girod, S.; Boussel, L.; Douek, P.; Si-Mohamed, S. CT dose optimization for the detection of pulmonary arteriovenous malformation (PAVM): A phantom study. Diagn. Interv. Imaging 2020, 101, 289–297. [Google Scholar] [CrossRef]
  37. Greffier, J.; Frandon, J.; Larbi, A.; Beregi, J.P.; Pereira, F. CT iterative reconstruction algorithms: A task-based image quality assessment. Eur. Radiol. 2020, 30, 487–500. [Google Scholar] [CrossRef]
  38. Chen, L.; Shaw, C.C.; Altunbas, M.C.; Lai, C.J.; Liu, X. Spatial resolution properties in cone beam CT: A simulation study. Med. Phys. 2008, 35, 724–734. [Google Scholar] [CrossRef]
  39. Dolly, S.; Chen, H.C.; Anastasio, M.; Mutic, S.; Li, H. Practical considerations for noise power spectra estimation for clinical CT scanners. J. Appl. Clin. Med. Phys. 2016, 17, 392–407. [Google Scholar] [CrossRef]
  40. Schulze, R.; Heil, U.; Gross, D.; Bruellmann, D.D.; Dranischnikow, E.; Schwanecke, U.; Schoemer, E. Artefacts in CBCT: A review. Dentomaxillofac. Radiol. 2011, 40, 265–273. [Google Scholar] [CrossRef] [PubMed] [PubMed Central]
  41. Rossi, M.; Cerveri, P. Comparison of Supervised and Unsupervised Approaches for the Generation of Synthetic CT from Cone-Beam CT. Diagnostics 2021, 11, 1435. [Google Scholar] [CrossRef]
  42. Battistelli, M.; Valeri, F.; D’Ercole, M.; Izzo, A.; Rapisarda, A.; Polli, F.M.; Montano, N. The use of intraoperative CT-neuronavigation in Wiltse approach. A technical note. Front. Surg. 2024, 11, 1433273. [Google Scholar] [CrossRef] [PubMed]
  43. Baumgart, L.; Ille, S.; Kirschke, J.S.; Meyer, B.; Krieg, S.M. Radiation doses and accuracy of navigated pedicle screw placement in cervical and thoracic spine surgery: A comparison of sliding gantry CT and mobile cone-beam CT in a homogeneous cohort. J. Neurosurg. Spine 2023, 39, 363–369. [Google Scholar] [CrossRef] [PubMed]
  44. Servello, D.; Saleh, C.; Zekaj, E. Intraoperative mobile computed tomography in deep brain stimulation: Comparison between Airo CT and O-arm CT. Surg. Neurol. Int. 2022, 13, 258. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Summary diagram of the study objective and steps of methodology.
Figure 1. Summary diagram of the study objective and steps of methodology.
Sensors 24 06883 g001
Figure 2. Catphan phantom images: (A) a uniform module (CTP 486) with the region of interest for NPS calculation. (B) Sensitometry module (CT404) with the region of interest placed on Delrin and Teflon inserts for TTF calculation.
Figure 2. Catphan phantom images: (A) a uniform module (CTP 486) with the region of interest for NPS calculation. (B) Sensitometry module (CT404) with the region of interest placed on Delrin and Teflon inserts for TTF calculation.
Sensors 24 06883 g002
Figure 3. Noise power spectrum results obtained for the three devices (Airo TruCT, Loop-X, and O-Arm).
Figure 3. Noise power spectrum results obtained for the three devices (Airo TruCT, Loop-X, and O-Arm).
Sensors 24 06883 g003
Figure 4. Task transfer function for the three devices (Airo TruCT, Loop-X, and O-Arm) for two inserts: Teflon and Delrin.
Figure 4. Task transfer function for the three devices (Airo TruCT, Loop-X, and O-Arm) for two inserts: Teflon and Delrin.
Sensors 24 06883 g004
Table 1. Technical specifications of the three intraoperative imaging systems (Airo TruCT, Loop-X, and O-Arm).
Table 1. Technical specifications of the three intraoperative imaging systems (Airo TruCT, Loop-X, and O-Arm).
Airo TruCTLoop-XO-Arm
Free space (cm)107102–12196.5
Detector typeSolid-State Array CdWO4CsI(Tl)a-Si
Matrix size 512 × 5121024 × 10242048 × 1536
FOV (cm)25.6–51.225–60 (2D)21.2–39.7
3–25–48 (3D)
Generator power (kW)3014.432
Nominal kV120120150
kVp range80–12040–12040–140 *
mA range5–2500.2–8 continuous10–100 *
5–120 pulsed
Focus size (mm)1 × 10.3 (small) 0.6 × 0.9 (small)
0.6 (large)1.2 × 1.7 (large)
Total filtration (mmAl)6.8 (at 120 kVp)4.4 (at 75 kVp) 4.7

(at 75 kVp)
Bow-Tie filter
0.3–3 mm Cu
* For pulsed radiography mode.
Table 2. Acquisition and reconstruction parameters used with three imaging devices (Airo TruCT, Loop-X, and O-Arm).
Table 2. Acquisition and reconstruction parameters used with three imaging devices (Airo TruCT, Loop-X, and O-Arm).
Airo TruCTLoop-XO-Arm
Acquisition parameterskVp120120120
Temps de rotation (s)1507.5
Pitch factor1.415--
Tube current (mAs)257480298
Acquisition field of view (SFOV) cm380350350
CTDIvol (mGy)/CTDIw2726.726.2
Reconstruction parametersReconstruction algorithmFBPFBPFBP
Displayed field of view (DFOV) cm256250250
Slice thickness (mm)11.20.833
KernelSoftStdStd
Table 3. Noise magnitude, average, and peak frequency results obtained for the three devices (Airo TruCT, Loop-X, and O-Arm).
Table 3. Noise magnitude, average, and peak frequency results obtained for the three devices (Airo TruCT, Loop-X, and O-Arm).
Noise Magnitude (HU)fav (mm−1)fpeak (mm−1)
Airo TruCT7.60.220.26
Loop-X12.310.170.03
O-Arm28.80.470.36
Table 4. Task transfer function at 50% (TTF50) for the three devices (Airo TruCT, Loop-X, and O-Arm).
Table 4. Task transfer function at 50% (TTF50) for the three devices (Airo TruCT, Loop-X, and O-Arm).
TTF50
TeflonDelrin
Airo TruCT0.320.34
Loop-X0.390.37
O-Arm0.490.48
Table 5. Detectability index values of the three simulated lesions with for the three imaging devices (Airo TruCT, Loop-X, and O-Arm).
Table 5. Detectability index values of the three simulated lesions with for the three imaging devices (Airo TruCT, Loop-X, and O-Arm).
d’
Lytic ScleroticBone
Airo TruCT7.3725.8064.22
Loop-X5.4018.8348.84
O-Arm1.715.9815.30
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Dabli, D.; Salvat, C.; Fitton, I.; Van Ngoc Ty, C.; Palanchon, P.; Beregi, J.-P.; Greffier, J.; Hadid-Beurrier, L. Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study. Sensors 2024, 24, 6883. https://doi.org/10.3390/s24216883

AMA Style

Dabli D, Salvat C, Fitton I, Van Ngoc Ty C, Palanchon P, Beregi J-P, Greffier J, Hadid-Beurrier L. Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study. Sensors. 2024; 24(21):6883. https://doi.org/10.3390/s24216883

Chicago/Turabian Style

Dabli, Djamel, Cécile Salvat, Isabelle Fitton, Claire Van Ngoc Ty, Peggy Palanchon, Jean-Paul Beregi, Joël Greffier, and Lama Hadid-Beurrier. 2024. "Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study" Sensors 24, no. 21: 6883. https://doi.org/10.3390/s24216883

APA Style

Dabli, D., Salvat, C., Fitton, I., Van Ngoc Ty, C., Palanchon, P., Beregi, J. -P., Greffier, J., & Hadid-Beurrier, L. (2024). Image Quality Comparison of Three 3D Mobile X-Ray Imaging Guidance Devices Used in Spine Surgery: A Phantom Study. Sensors, 24(21), 6883. https://doi.org/10.3390/s24216883

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop