3.1. GSI Response and Magnetoelastic Coupling Analysis
Firstly, the GSI response of the ribbons under axial (tensile and compressive, see
Figure 1b) stresses at
H = 0 was characterized. As can be seen in
Figure 3, both samples display a clear different magnetoelastic response. Firstly, the sample
P1 shows a maximum for the unstressed state (
σ = 0 MPa), leading to a clear diminution in
Z when both tensile and compressive stresses are applied. However, different behavior is observed for
P2, where the maximum impedance does not occur at the unstressed state. Starting from
σ = 0 MPa, the application of compressive stresses causes a decrease in
P2 impedance. On the contrary, the application of tensile stresses leads to an increase of
Z and, consequently, of the GSI ratio, obtaining its maximum value close to
σ = 300 MPa. Further application of tensile stresses leads to a large region where GSI ratio slowly decreases until the most intense tensile stress (
σ = 850 MPa) is applied. This effect can be ascribed to the appearance of internal stresses in the sample during sample fabrication procedure (melting spinning technique) and the occurrence of a different intrinsic magnetoelastic contribution in both samples.
To evaluate this effect, namely, the contribution of the intrinsic magnetoelastic anisotropy, the magnetostriction constant,
λs, was evaluated in both samples through the analysis of
Z(
H) at different applied
σ.
Figure 4 shows the impedance ratio,
, versus
H as a function of
σ (compressive and tensile) for the (a)
P1 and (b)
P2 sample. In this case and similarly to
Figure 3, the value of
was normalized with respect to the
DC electric resistance, resulting in the expression
.
Initially, the sample
P1 in the unstressed state undergoes a continuous decrease in
Z with
H (see
Figure 4a). This GMI behavior is known as single peak (
SP). The application of compressive stresses (
σ < 0) promotes the appearance of a double maximum in the impedance curve, known as double peak (
DP). This impedance evolution indicates a positive value of the magnetostriction constant (
λs > 0), that is, a reinforcement of the transverse anisotropy field under
σ < 0. Thus, under the effect of tensile stresses (
σ > 0), the
SP behavior remains but a decrease in
values is detected as a consequence of the reinforcement of the longitudinal easy axis, leading to a decrease in the effective transverse magnetic permeability involved in the GMI effect. Particularly for
H = 0, a similar trend to that in
Figure 3 is obtained, that is, a decrease in
Z with respect to the unstressed state when tensile or compressive stresses were applied.
On the other hand, the sample
P2 exhibits an opposite behavior (see
Figure 4b), with the initial unstressed state characterized by a
DP behavior. In this case, the application of tensile stresses (
σ > 0) causes the displacement of the maximum of
Z to higher
H field values together with an increase in the maximum impedance variations, indicating a negative value of the magnetostriction constant (reinforcement of the transverse anisotropy field under
σ > 0). An opposite effect is found for compressive stresses (
σ < 0), namely, the displacement of the maximum toward lower
H field values and a progressive decrease in the maximum impedance variations. Again, at
H = 0, the results are consistent with
Figure 3, where an increase in
Z is produced with respect to the unstressed state when axial tensile stresses were applied (and vice versa, decrease in Z for
σ < 0).
The actual value of
can be calculated from the evolution of the estimated
through the axial
H at which Z displays a maximum value. It is important to note that the components of the impedance,
Z = R + jX, (with
R and
X the real and imaginary part) evolve differently with
Ipp. While the maximum in the curve
X(H) stays nearly constant around
, independently of the value of
Ipp, a dependence is found for
R(H) curve [
25]. Since, in this study, only the modulus of
Z was measured, the samples (see
Figure 4) were excited at lower electric current amplitude (
Ipp = 40 mA) to assure a low-exciting transverse magnetic field and, therefore, according to the literature, an accurate estimation of
through this procedure [
25].
Figure 5 shows the obtained evolution of
with
σ. As expected, a polynomial second-order dependence among both parameters is found,
, that is correlated to the magnetostriction constant as follows [
26,
27]:
Values of
λs0 = 1.4 × 10
−7,
k = 3.0 × 10
−10 (MPa
−1), and
λs0 = −4.6 × 10
−7,
k = −2.6 × 10
−10 (MPa
−1) for
P1 and
P2, respectively, are obtained (where
λs0 is the unstressed valued of the saturation magnetostriction and
k is the measured slope of the linear dependence of
λs with stress). Similar results were reported for nearly zero magnetostrictive amorphous alloys (
λs0 ~ 10
−7 and
k ~ 10
−10 (MPa
−1)) employing the small angle magnetization rotation (SAMR) [
27] technique and GSI effect [
26].
The described behavior of Z in both samples can be understood in terms of the magnetic anisotropies present in the samples. The amorphous nature of the synthesized ribbons indicates the lack of magnetocrystalline anisotropy. Thus, the presence of a magnetoelastic anisotropy resulting from rapid solidification of the samples (coupling between the quenched-in stresses and
λs) and the magnetoelastic coupling with the externally applied mechanical stresses determines the magnetic behavior and, consequently, the GMI (or GSI) effect. In the unstressed state, the sample
P1 shows an impedance dependence with just one maximum (
SP), which could be indicative of a not well-defined magnetic anisotropy distribution in the sample, as previously published for other melt-spun ribbons [
28]. Furthermore, when axial tensile stresses were applied, the
SP behavior remains with a noticeable reduction in the impedance values. On the contrary, the application of compressive stresses leads to the induction of a transverse magnetic anisotropy in the sample, clearly visible as the appearance of a double peak (
DP) in the
Z curve. Thus, the impedance stress evolution is mainly governed by the positive value of the saturation magnetostriction of the sample. Conversely, the ribbon
P2 displays a well-defined transverse magnetoelastic anisotropy for
σ = 0 MPa, which can only be due to the internal stresses during the cooling down procedure and the coupling with the negative magnetostriction constant (see
Figure 4b). This transverse anisotropy is reinforced when tensile stresses were exerted, observing a displacement of
Z maxima to higher
H values. The opposite trend is observed for compressive stresses (shift toward lower
H). Thus, the evolution of the GSI response of the samples (
H = 0, see
Figure 3) is the result of the counterbalance of the internal quenched-in magnetoelastic anisotropy and the magnetoelastic contribution associated with the externally applied stress (tensile or compressive).
It is relevant to mention that, although the magnetostriction constant value of both samples is quite low (nearly zero), the analyzed ribbons can be successfully applied as GSI magnetoelastic sensors, as shown in the next section. Their amorphous structure together with the initial large value of the transverse magnetic permeability, µ, favors a large GSI effect that enables the characterization of stresses on the samples through the variations in the ribbon impedance.
Nevertheless, the previous results permit the election of the optimum ribbon to be employed as the sensor nucleus for the final magnetoelastic devices. On one hand, P2 shows a univocal Z(σ) response for −200 MPa < σ < 300 MPa at H = 0, allowing the determination of both sign and strength (magnitude) of the applied mechanical stress within this interval. However, further application of tensile stresses leads to an indetermination given by the fact that the same Z value is observed for different applied tensile stresses on the sample. On the other hand, P1 exhibits a symmetric Z(σ) that hinders the determination of the stress sign but favors the univocal characterization when just compressive or tensile stresses are applied.
3.2. Magnetoelastic Sensor for Cross-Section Evaluation
Figure 6 shows the response of both ribbons under the mechanical tests. The main purpose is to analyze their suitability as sensors for the evaluation of the variations in the cross-section dimensions of pieces (see
Section 2.2.1). As can be seen, the relative impedance variation (
) displays, in both analyzed samples, enough sensitivity to measure the micrometric variation of the diameter of the probes (Δ
Dm). The effective stress applied to the samples during mechanical tests,
σ, can be calculated assuming that the strain in the ribbon,
ε, is the same as the relative changes in the cylinder diameter (
). Thus, applying Hooke’s law,
σ, it can be calculated as
[
19]. Thus, maximum tensile stresses of 500 and 250 MPa are applied to the samples for
Dm = 20 and 40 mm, respectively, assuming
Er = 100 GPa and Δ
Dm = 0.1 mm. Notice that these maximum tensile stresses values are similar to those applied in the initial magnetoelastic characterization (see
Figure 3). Thus, according to the
curve, the application of tensile stresses, as those applied in this characterization, would give rise to a decrease and an increase in Z for
P1 and
P2 samples, respectively, as a consequence of the different sign in magnetostriction constant, as stated in
Figure 6 (keep in mind that
).
Moreover, as previously reported [
18], the initial bent configuration of the sensing ribbons along the cylindrical probe induces a distribution of tensile and compressive stresses whose strength increases for lower probe diameters. Therefore, this pre-stressed state should display, for
Dm = 20 mm, lower impedance variations in comparison with the impedance evolution for
Dm = 40 mm. However, as
Figure 6 shows, the impedance displays an opposite trend, that is, slightly lower maximum
ratios for the larger diameter (
Dm = 40 mm). This difference can be explained on the basis of the dependence of the GMI effect on the sample length. In fact, as explained in the experimental section, ribbon lengths (
Lr = 5 and 11 cm) were chosen for covering the whole cylinder perimeter (
Dm). As reported, maximum GMI ratios are found for optimal sample lengths [
29]. For the employed samples, this value is around
Lr ~ 5 cm. Thus, although, due to initial bending stresses, a lower maximum
ratio should be expected for the smallest diameter (
Dm = 20 mm), the larger GMI ratio in this shorter ribbon (
Lr = 5 cm) dominates the sensor response and provides a higher maximum ratio [
19].
Nevertheless, it should be noted that both samples can be employed as sensitive micrometric cross-section sensors, displaying the following sensitivities to the diameter variations: P1 −98 and −62%/mm−1; P2 64 and 38%/mm−1 for Dm = 20 and 40 mm, respectively. Notice the highest sensitivity of the P1 sample under both cylinder diameters. Therefore, this larger sensitivity together with the fact that the proposed application is based on the application of just tensile stresses justifies the election of the sample P1 for the design of the proposed magnetoelastic sensors.
Furthermore, to properly characterize the response of the selected
P1 ribbon for the proposed magnetoelastic sensor, the mutual effect of temperature and the fixing agent was examined. Its relevance relies on the fact that changes in temperature may alter the properties of the adhesive and, therefore, the transmission of the mechanical stresses together with the impedance itself. At first, the evolution of the relative variation of impedance with temperature was examined fixing the
P1 ribbon on a cylindrical methacrylate probe of
Dm = 20 mm. This relative variation was normalized employing the expression
. As
Figure 7 shows,
is markedly larger when the ribbon
P1 was fixed with cyanoacrylate. The best performance employing Araldite was also confirmed during the transmission of mechanical stresses.
Figure 8 shows
for the mechanical tests performed at different temperatures. As
Figure 8 shows, no relevant changes in
are observed when the mechanical tests were performed at different temperatures with Araldite. Maximum differences in
around 1.5% are obtained when measuring Δ
Dm at different temperatures (i.e Δ
Dm = 0.1 mm). However, the cyanoacrylate adhesive strongly affects the behavior of the impedance, showing significant changes in the maximum diminution of
with Δ
Dm (from 7.2% to 2.4% at room temperature and 60 °C, respectively), invalidating its use in these magnetoelastic sensors. These results are in good concordance with
Figure 7. This effect could be related to the higher coefficient of thermal expansion of the cyanoacrylate. Thus, when temperature changes, the larger variations of dimensions displayed by cyanoacrylate may exert more stresses on the ribbon, leading to changes in Z even under zero applied external stresses.
As the main conclusion, it can be pointed out that the proposed magnetoelastic sensor based in soft magnetic (nearly zero magnetostrictive) amorphous ribbons demonstrated its capability to characterize micrometric variations of the cross-section dimensions. This sensor prototype exhibits several advantages such as high Young modulus and tensile strength, parameters that favor an accurate stress characterization. Furthermore, these amorphous alloys can be easily adapted to the whole perimeter of the pieces to characterize them even if they offer an irregular geometry and/ or small size. In this sense, although the employed probes in this analysis displayed a smooth and regular surface, the proposed magnetoelastic sensor was successfully employed in systems with irregular and rough surfaces [
30]. Precisely, this adaptability permits integrating the stresses around the whole perimeter, enabling us to obtain a mean variation of the cross-section (even for irregular shapes) where all points around the piece perimeter can contribute to the general stress characterization. In contrast, piezoresistive strain gauges (see
Figure 1a) can only compute the stresses in a local point, hindering the mean estimation along the whole perimeter and favoring the loss of information especially in irregularly shaped pieces.
3.3. Development of A Flow Meter
The magnetoelastic ribbon was located perpendicularly to the water flow inside the pipe (see
Section 2.2.2) which exerted variable bending (tensile) stresses on the ribbon. The principle of operation of the proposed flow meter relies on the changes of impedance experimented by the
P1 ribbon when the water flow changes. To obtain significant variations of
Z, the sample required a certain pre-tightened state; otherwise, no variations of impedance were observed.
The relative variations of impedance,
, were studied as a function of the water flow in the pipe. In order to study reproducibility and hysteretic behavior of the sensor response, the water flow was gradually increased and subsequently decreased, repeating this process in different attempts. The impedance changes were registered simultaneously with the controlled variation of the water flux, measured by the commercial dispositive. As an example,
Figure 9 shows
versus water flow for a scan of measurements. It can be seen that, despite the low impedance variations, the changes in the water flow can be properly determined. Since no linear response was obtained, a standard sensitivity cannot be calculated. In consequence, the sensor response was fitted to a polynomial second-order curve (
, where
F is the water flow) for the set of performed measurements, both increasing and decreasing
F.
Table 1 shows the mean values of the estimated polynomial second-order parameters, together with the standard error of the mean. It should be pointed out that there is a slightly higher uncertainty for the decreasing flux measurements. Concerning the sensor hysteresis, only small differences can be observed in impedance evolution under increased and decreased water flow. To quantify this effect, a “hysteretic coefficient (
HC)” can be calculated. This parameter can be defined as the relative maximum variation of the magnitude output (in this case,
) between increasing and decreasing water flow changes. In this case, as
Figure 9 shows, the relative variation of the
a parameter (
F = 0) can be employed in the hysteresis estimation. Thus, a value of
is obtained.
Eventually, it is also relevant to note that the suggested sensor can be employed not only for water flow measurements, permitting the generalization of its employment for other liquids. Finally, in comparison with other flow meter devices, the main advantages of the proposed sensor rely on its low cost and its capacity of being scalable to different pipe diameters and sizes, thereby broadening its applicability. However, some future work should be addressed, such as the design of a holder for fixing the sensor firmly inside the pipe, an analysis of the lifetime of the sensor based on the number of cycles with a linear response, and the effect of aging on its response.