4.1. Middle Euphrates Basin
The statistical methods described above were applied on each of eight streamflow gauging stations in the Middle Euphrates Basin, which were considered in turn as the target station. The donor station or stations based on physical similarity was selected to transfer daily streamflow data to the target station.
Table 6 shows the sequence of the donor stations for each target station in the Middle Euphrates Basin which was determined according to the similarity coefficient.
In order to estimate daily streamflow at the target stations, the DAR method was applied by using the most physically similar station to each target station. In order to test the applicability of the donor station selection criteria for the study area, the NSE values were determined for the donor stations identified by the physical similarity and compared with the NSE values obtained from the donor stations traditionally selected as the geographically nearest stations. On the other hand, the MDAR and ISW methods were applied by using the two and the three most physically similar stations to each target station. In order to test the effect of different power parameter selection in the use of the ISW method on the accuracy of daily streamflow estimation, the ISW method was applied with power parameters of 1, 2, and 3. In addition, the comparisons between the statistical methods with and without the MA were carried out to indicate the effectiveness of the MA-based preprocessing on the accuracy of daily streamflow estimation. Daily streamflow values estimated using both observed and smoothed data from the donor stations were compared with observed data at the target station. According to the selection criteria of the donor stations, the NSE values obtained for the target stations in the Middle Euphrates Basin are given in
Table 7 for DAR and MDAR and in
Table 8 for ISW with three different power parameters (1, 2, and 3). Performance evaluations of the best statistical method without and with the MA for each target station in the Middle Euphrates Basin are presented in
Table 9 and in
Table 10, respectively.
For 2 out of 8 target stations (i.e., E21A002 and E21A022), the geographically nearest and the most physically similar station were the same. For 3 out of the remaining 6 target stations (i.e., D21A169, E21A064, and E21A077), higher NSE values were obtained using DAR with the most physically similar station as the donor station. For 3 out of the target stations (i.e., D21A167, D21A213, and E21A058), higher NSE values were obtained using DAR with the geographically nearest station as the donor station. According to these results, for half of the target stations in the study area, the geographical distance seems to be a good selection criterion as the donor station; however, for the remaining half of the target stations, geographical distance cannot identify the best donor station. Therefore, donor station selection criteria can provide different estimated results that vary from basin to basin.
As can be seen in
Table 7, for all target stations other than E21A064, higher NSE values were obtained with MDAR compared with DAR. Especially for D21A167, E21A002, and E21A022, negative NSE values obtained with DAR using the most physically similar donor station improved considerably when the three most physically similar donor stations were used with MDAR. The performance of the DAR method was unsatisfactory for D21A167, E21A002, and E21A022. This was mostly due to the significant increase in the drainage area ratio between the donor and target stations. D21A213 has the smallest drainage area (65.3 km
2) in the Middle Euphrates Basin and was determined as the most physically similar donor station for both D21A167 (250 km
2) and E21A022 (5882.4 km
2). Moreover, E21A002 has the largest drainage area (25,515.6 km
2) in the Middle Euphrates Basin. Its drainage area is more than four times the next largest station. For all target stations other than D21A169 and E21A058, MDAR using of the three most physically donor stations produced better NSE values than that using the two most physically similar donor stations. For D21A169, the NSE value decreased from 0.852 to 0.595 when the three most physically similar donor stations were used instead of the two most physically similar donor stations (see
Table 7). In case of the use of the three most physically similar donor stations, the third most physically similar donor station for D21A169 was determined as D21A167. The NSE value obtained for D21A169 using the DAR method and utilizing D21A167 was lower than the NSE values obtained from the other two donor stations (i.e., E21A058 and E21A077). The drainage area of the donor station D21A167 is very close to the target station D21A169. On the other hand, the drainage areas of the other two donor stations are much larger than D21A169. Hence, the weight of donor station D21A167 for streamflow estimation of D21A169 is significantly larger compared to the other two. Consequently, the NSE value obtained for D21A169 using MDAR with the three most physically similar donor stations is predominantly influenced by donor station D21A167. Similarly, for E21A058, the decrease in the NSE (i.e., from 0.894 to 0.839) was due to the same reason as for D21A169.
As can be seen in
Table 8, in case of the use of the two most physically similar donor stations, the best performance results were obtained with ISW1 for 5 out of 8 target stations (i.e., D21A167, D21A213, E21A002, E21A022, and E21A058). On the other hand, in case of the use of the three most physically similar donor stations, the best performance results were obtained with ISW1 for 5 out of 8 target stations (i.e., D21A167, D21A213, E21A002, E21A022, and E21A077). In both cases, the most reasonable estimation results were mostly obtained when ISW1 was applied instead of ISW2 and ISW3. Moreover, the NSE values mostly improved when the three most physically similar donor stations were used instead of the two most physically similar donor stations.
As can be seen in
Table 9, for all target stations other than for E21A064, the MDAR and the ISW methods resulted in higher NSEs compared to the DAR method. For 6 out of 8 target stations, the results can be rated as “very good” for NSE according to the performance ratings in
Table 5. For 7 out of 8 target stations, the RSR values were considered satisfactory (i.e., less than 0.7) according to the performance ratings recommended by Moriasi et al. [
47]. The negative PBIAS values for D21A213, E21A002, E21A022, and E21A077 demonstrate that the method overestimated daily streamflow, while positive PBIAS values for D21A167, D21A169, E21A058, and E21A064 demonstrate underestimation. For all target stations, the statistical methods with the MA tend to achieve slightly higher NSE values. However, the PBIAS values of the target stations did not change when the statistical methods with the MA were used.
For the target station E21A058 as the example, the estimated daily streamflow values from the statistical methods without MA were compared to the observed (original) daily streamflow values in the hydrograph and scatter plots in
Figure 6. The remarkably better agreement between observed and estimated daily streamflow values by three statistical methods (i.e, DAR, MDAR, and ISW) was obtained for E21A058 compared to the other target stations in the Middle Euphrates Basin. ISW2 gave a coefficient of determination (R
2) of 0.91, which was higher than the R
2 values of 0.87 and 0.90 obtained by using the DAR and MDAR, respectively. The NSE values for these methods ranged from 0.814 to 0.907, and the best NSE value was achieved by ISW2. The best NSE performance for E21A058 was obtained using ISW2 with the three most physically similar donor stations.
On the other hand, for the target station E21A058 as the example, the estimated daily streamflow values from the statistical methods with MA were compared with observed-MA (smoothed) daily streamflow values in the hydrograph and scatter plots in
Figure 7. The statistical methods with MA performed slightly better for E21A058.
4.2. Upper Euphrates Basin
Using the same procedure applied for the Middle Euphrates Basin, the statistical methods were applied on each of seven streamflow gauging stations in the Upper Euphrates Basin for the purpose of estimating daily streamflow.
Table 11 shows the sequence of the donor stations for each target station in the Upper Euphrates Basin which was determined according to the similarity coefficient.
According to the selection criteria of the donor stations, the NSE values obtained for the target stations in the Upper Euphrates Basin were given in
Table 12 for DAR and MDAR and in
Table 13 for ISW with three different power parameters (1, 2, and 3). Performance evaluations of the best statistical method without and with the MA for each target station in the Upper Euphrates Basin are presented in
Table 14 and in
Table 15, respectively.
For 2 out of the 3 target stations (i.e., E21A054 and E21A056) for which the geographically nearest and the most physically similar stations were not the same, higher NSE values were obtained using DAR with the most physically similar station (see
Table 12). For E21A054 and E21A056, the NSE values improved considerably when the most physically similar station was used. These results indicate that the physical similarity may be a better selection criterion for the donor station in the study area compared to the geographical distance between the stations.
As can be seen in
Table 12, for all target stations other than E21A033, E21A054, and E21A056, higher NSE values were obtained with MDAR compared with DAR. Especially for E21A066, the NSE value improved considerably with MDAR as compared with DAR. For D21A001, E21A051, E21A054, and E21A056, the NSE values decreased when MDAR was applied using the three most physically similar donor stations instead of the two most physically similar donor stations (see
Table 12). In case of use of the three most physically similar donor stations, the third most physically similar donor stations for the target stations D21A001, E21A051, and E21A054 were determined as D21A193, E21A066, and E21A033, respectively. All NSE values obtained for these target stations using the DAR method utilizing their third most physically similar donor stations were negative. The drainage areas of these target stations and their third most physically similar donor stations are very close to each other. Therefore, the weight of their third most physically similar donor stations for streamflow estimation of these target stations is significantly larger. As a result, the NSE values obtained for the target stations D21A001, E21A051, and E21A054 using MDAR with three most physically similar donor stations are predominantly influenced by their third most physically similar donor stations. On the other hand, for E21A056, the reason is slightly different from the others. The NSE value obtained for E21A056 using the DAR method utilizing its third most physically similar donor station E21A033 was too low (i.e., −10.029). Although the contribution of the donor station E21A033 is not much more than the other two, this leads to poor estimation performance for E21A056 when MDAR was applied using the three most physically similar donor stations.
As can be seen in
Table 13, in case of the use of the two most physically similar donor stations, the best performance results were obtained with ISW1 for 4 out of 7 target stations (i.e., D21A001, D21A193, E21A051, and E21A066). On the other hand, in case of the use of the three most physically similar donor stations, the best performance results were obtained with ISW3 for all target stations other than D21A193 and E21A066. As the power parameter increased from 1 to 3, the NSE values mostly improved when the three most physically similar donor stations were used, whereas the NSE values mostly decreased when the two most physically similar donor stations were used. Moreover, the NSE values mostly improved when the two most physically similar donor stations were used instead of the three most physically similar donor stations.
As can be seen in
Table 14, for all target stations other than for E21A033, E21A054, and E21A056, the MDAR and the ISW methods resulted in higher NSE compared to the DAR method. For 4 out of 7 target stations, the results can be rated as “very good” for the NSE according to the performance ratings in
Table 5. For 5 out of 7 target stations, the RSR values were considered satisfactory (i.e., less than 0.7) according to the performance ratings recommended by Moriasi et al. [
47]. The negative PBIAS values for D21A193, E21A051, E21A054, and E21A066 demonstrate that the method overestimated daily streamflow, while positive PBIAS values for D21A001, D21A033, and E21A056 demonstrate underestimation. For all target stations other than D21A193, the statistical methods with MA tend to achieve slightly higher NSE values. However, the PBIAS values of the target stations did not change when the statistical methods with MA were used.
For the target station E21A051 as the example, the estimated streamflow values from the statistical methods without the MA were compared to observed (original) streamflow values in the hydrograph and scatter plots in
Figure 8. The remarkably better agreement between observed and estimated streamflow values by three statistical methods was obtained for E21A051 compared to the other stations in the Upper Euphrates Basin. Both MDAR and ISW1 gave a coefficient of determination (R
2) of 0.94, which was higher than the R
2 values of 0.89 by using DAR. The NSE values for these methods ranged from 0.883 to 0.932, and the best NSE value was achieved by MDAR. The best NSE performance for E21A051 was obtained using MDAR with the two most physically similar donor stations.
On the other hand, for the target station E21A051 as the example, the estimated streamflow values from the statistical methods with the MA were compared to observed-MA (smoothed) streamflow values in the hydrograph and scatter plots in
Figure 9. The statistical methods with the MA performed slightly better for E21A051.