1. Introduction
14C is a common radionuclide in our environment, due to its long half-life (5730 years) [
1], its natural origin in the atmosphere from the reaction between
14N and cosmic neutrons and its artificial origin from nuclear explosions and releases from the nuclear industry and medical and research activities [
2].
As
14C is a long-lived radionuclide, it remains in the environment for a long time and is integrated in the carbon cycle, mainly as CO
2. There, it can be transformed into organic carbon via the photosynthesis of plants and other organisms, such as cyanobacteria, and then into CO
2 via breathing or into inorganic carbon through the decomposition of bacteria and fungi. CO
2 can also be absorbed and released by oceans, where it is transformed into carbonic acid [
3,
4].
In both terrestrial and marine ecosystems, natural
14C tends to reach a stable ratio to the total carbon in different environmental compartments: air, water, soil, sediments, food and biota [
2].
Therefore, its measurement is very useful for the assessment of the impact caused by nuclear activities on abiotic and biotic compartments, as they increase their
14C content and the natural ratio due to
14C releases [
5,
6]. It is also very useful to assess the impact of the fossil fuel industry, which may cause the so-called Suess effect [
7], i.e., the decrease in the
14C to the total carbon natural ratio, as it releases carbon with low content of
14C into the atmosphere.
In order to control the level of 14C in air, water, soil, sediments, food and biota, the Laboratory of Low Activity Measurements (LMBA) of the University of the Basque Country (UPV/EHU), has used different methods to analyse this radionuclide for around 15 years. Most of them have been carried out within the framework of the Spanish National Radiological Surveillance Network for the environmental radiological monitoring of nuclear power plants and nuclear fuel cycle facilities.
As
14C is a medium-energy (E
max = 156.476 keV) beta emitter [
1], it is measured via liquid scintillation spectrometry (LSS). This radiometric technique allows the result of the
14C measurement to be expressed in activity unit (Bq), but it is usually expressed in activity concentration units (Bq kg
−1 or Bq m
−3).
Even if the radiometric technique is always LSS, the prior sample preparation depends on the type of sample and the detection limit required.
Aqueous matrices are usually analysed following the ISO (International Organization for Standardization) 13162 standard [
8], namely, via direct counting or bubbling and CO
2 absorption. Direct counting provides good detection efficiencies, the measurement of total
14C and rapid and inexpensive results. Therefore, it is very useful for screening purposes. However, it is limited by spectral interferers; by variable quench, depending on water purity; and by the maximum volume of sample to be measured, which leads to detection limits higher than 2 Bq L
−1.
The bubbling and CO
2 absorption method also enables the measurement of total
14C. As it is carbon-selective, there are no spectral interferences during the sample measurement. Moreover, as it supports water samples greater than 50 mL, detection limits become lower than 2 Bq L
−1. However, the method is limited by the maximum quantity of the CO
2 absorber, which is miscible with the scintillation cocktail, and it is strongly affected by its quench and chemiluminescence [
9], which lead to low overall efficiencies (around 30%).
Solid samples can be treated with a sample oxidizer, which also allows the determining of the total
14C [
10]. Its use is limited by the maximum quantity of sample that can be put into the sample oxidizer [
10], which leads to high detection limits, and also the CO
2 absorber, which is a quench agent [
9], leading to low overall efficiencies (around 40%), and in some cases, spectral interferences.
The LMBA has implemented a quality system in accordance with the ISO/IEC (International Electrotechnical Commission) 17025 [
11], in order to assure the quality of its results, involving analytical procedures, such as
14C activity determination, in any kind of sample. Therefore, analytical procedures should be validated and controlled, subjecting them to proficiency tests and interlaboratory comparisons using reference materials. This has permitted the laboratory to be accredited by ENAC, the Spanish national accreditation body, under the ISO/IEC 17025 standard [
11].
Today, there are many established validation methods for analytical methods, but not so many for the radioanalytical ones. Hence, the validation of radioanalytical methods still pose a challenge for many laboratories, which usually base it on those for analytical methods.
The fundamental items that make the difference between analytical and radioanalytical validation methods are:
- −
Radioanalytical methods are usually time-consuming (some of them may involve several days for sample preparation).
- −
Radioanalytical methods validation involves the use of radioactive substances.
Due to the impact that radioactive substances may cause on humans and the environment, most laboratories avoid preparing radioactive materials and validate their procedures externally—taking part in proficiency tests and interlaboratory comparison exercises or using certified reference materials—covering a wide range of activities.
However, in the case of
14C, validations still pose a challenge for laboratories due to the scarcity of intercomparison exercises, especially in solid samples. In the last 10 years, only two intercomparisons in solids have been located, both in biological (leaves) samples and in the same activity concentration range (100–1000 Bq kg
−1, dry weight) (IRSN (Institut de Radioprotection et de Sûrete Nucléaire), 2018 and CSN (Spanish Nuclear Safety Council), 2014). We participated in the second one and the result was “accepted” (
Table 1). Other intercomparisons in solids, in which we participated, did not yield conclusive results; either because our results were below the detection limit (NPL (National Physical Laboratory), 2008, and INSIDER (Improved Nuclear SIte characterisation for waste minimisation in Decommissioning and Dismantling operations under constrained EnviRonment), 2020) or because the exercise could not be evaluated due to the lack of a certified reference value and the small number of participating laboratories (CSN, 2010, in diet ashes and CSN, 2018, in milk power).
The accepted exercise only partially serves to validate our method, a validation that was able to be completed using reference materials. The IAEA (International Atomic Energy Agency) provides the C reference material series, whose 14C values are given in pMC (percent modern carbon); units that do not correspond to the radiometric units used in the radioactivity measurement laboratories (Bq m−3 or Bq kg−1). It is possible to theoretically transform pMC data into Bq m−3 data, but this implies the assumption of certain hypotheses that may cause the measured activity concentration value to not fit with the calculated one.
To overcome this problem, an intercomparison exercise to determine
14C activity concentrations in three of the nine IAEA reference materials, C2, C6 and C7, was conducted between 10 laboratories in the UK (United Kingdom). The laboratories performed the determinations using a number of techniques. The results obtained were very scattered, and those laboratories using radiometric methods provided results below the detection limit on several occasions, such as samples C2 and C7, which were those with the lowest pMC values [
12]. Sample C6, the sample with the highest pMC value and the best radiometrically characterised sample, is no longer available as reference material from the IAEA. So, we performed different determinations on the sample IAEA-C7. The results are presented in
Table 1 and can be considered as approved, although the reference value was theoretically calculated from its pMC value.
In the case of aqueous samples, it is easier to find intercomparisons. For example, NPL, IARMA (International Atomic Reference Material Agency), IRSN and CSN organise intercomparisons on aqueous matrices with some regularity. In the last ten years, we have participated in two of them (CSN 2015 and CSN 2019); in both cases, the evaluation was positive (see
Table 1).
Table 1.
14C activity concentration (), in Bq kg−1, in solid (biological) and water samples from different CSN (Spanish Nuclear Safety Council) proficiency tests (PT) and IAEA (International Atomic Energy Agency) reference material (Ref. Mat.). Uncertainty () is reported with a coverage factor of k = 2. Disp. means dispersion, in %, between laboratory (lab) and reference (ref) 14C activity concentration () values.
Table 1.
14C activity concentration (), in Bq kg−1, in solid (biological) and water samples from different CSN (Spanish Nuclear Safety Council) proficiency tests (PT) and IAEA (International Atomic Energy Agency) reference material (Ref. Mat.). Uncertainty () is reported with a coverage factor of k = 2. Disp. means dispersion, in %, between laboratory (lab) and reference (ref) 14C activity concentration () values.
PT | Matrix | | | | | Disp. (%) | Z-Score |
---|
CSN, 2014 [13] | Leaves | 116 | 16 | 124 | 17 | 7 | 0.4 |
CSN, 2015 [14] | Water | 32 | 6 | 38 | 3 | 17 | - |
CSN, 2019 [15] | Water | 28 | 6 | 29 | 3 | 4 | 0.2 |
Ref. Mat. | | | | | | | |
IAEA-C-6 [12] | Oxalic acid | 31 | 3 | 35 | 4 | 13 | - |
These exercises serve to show the general capability of our methods. However, it is necessary to perform determinations closest to the detection limits of some of the procedures considered in this paper. Moreover, the ranges of activity concentrations shown via the historical analysis of the intercomparison providers studied in this work are higher than those required for some of our procedures. In the case of water samples, reference materials are not available.
In addition, for both solid and water samples, in order to validate a method, it is necessary to obtain more parameters (such as accuracy, precision, repeatability and linearity) than those evaluated in the PT (proficiency test) exercises. If reference materials are not available for method validation, as in our case, certain traced samples can be produced by the laboratory to obtain the quality parameters of the methods to be validated.
Thus, the objectives of this work are to describe and to apply the proposed activities to internally carry out a complete validation for 14C activity determination, according to the requirements for accreditation and developing strategies fit for purpose. This validation method would be applicable to the determination of other radionuclides for which there is a lack not only of intercomparison exercises but also of suitable reference materials—a situation that is becoming more and more frequent in the framework of dismantling and decommissioning (D and D) projects.
Therefore, traced sample preparation, measurement and analysis carried out for each
14C activity determination method is presented in this paper, having previously defined the quality parameters and their acceptance criteria, according to the performance evaluation of the ISO 13528 standard [
16] and the Eurachem Laboratory Guide to Method Validation [
17].
Currently, these methods are usually applied to real samples of water, air, food and crops; the results obtained being within the range of values published by the IRSN [
2].
14C activity concentration in water is below the detection limit (
) required (300 Bq m
−3), whereas in air it varies from 0.03 to 0.07 Bq m
−3, the
required being 0.002 Bq m
−3. In food and crops,
14C activity concentration is in the range of 1 Bq wet kg
−1 (
required) to more than 80 Bq wet kg
−1, depending on the food product and crop.
In order to ensure the quality of these results, analyses are usually performed by two different laboratories at least: the main laboratory and the control one, which performs between 5 and 10% of the analyses. This allows laboratories to support their validation results, as a field validation, and control their radiochemical methods.
3. Results and Discussion
3.1. Direct Counting for the Determination of 14C Activity via LSS
Following the methods described in
Section 2.2., the measurement results of the set of water samples spiked with
14C standard solution and analysed according to the ISO 13162:2021 standard [
8], namely, via direct counting, are summarised in
Table 2.
As we can see in the table, at each activity concentration, reference and measured activity concentrations are very close to each other. The relative standard uncertainty of measured activity concentration is below 7% at low activity level and 2% at high activity level.
The results of the analysis of this set of samples are presented in
Table 3.
This table shows that all the studied parameters are in the acceptance range at both low and high activity levels: accuracy as is below 2.58 and precision and repeatability are below 10%. Therefore, and also taking into account the results of the intercomparison exercises in which we have participated, the method can be considered validated for water samples.
Despite some limitations, the results are acceptable, so the detection limit and uncertainty of the direct method can be defined as 2 Bq L−1 and 2–7%, respectively, for 10 mL of water, using 100 min for sample counting and 300 min for blank counting. The detection efficiency of the method is around 65% and linearity has been proven in an activity concentration range from 13 to 130 Bq L−1.
3.2. Bubbling Method for the Determination of 14C Activity via LSS
According to the methods described in
Section 2.2, measurement results in the set of water samples spiked with
14C standard solution and prepared via bubbling are presented in the following
Table 4.
As seen in the table, the relative standard uncertainty of measured activity concentration is below 10% at both low and high activity levels. Moreover, reference and measured activity concentrations are very similar, at each activity concentration, with the relative standard deviation of the results being below 5%.
The results of the analysis of this set of samples are presented in
Table 5.
This table shows that in the bubbling method for air and water samples, with low or high activity concentration, we also achieved good results regarding accuracy (<2.58), precision (<15%) and repeatability (<15%). Therefore, and also taking the results of the intercomparison exercises in which we have participated into account, the method can be considered validated for water samples.
Using this method, the sample volume can be arbitrarily large, although we routinely use 100 mL. This means that, even if this method consumes more time and resources than the previous method, detection limits can be orders of magnitude lower, making it more suitable when very low detection limits are to be achieved.
In view of the results obtained, the detection limit and the uncertainty of the bubbling method can be defined as 0.14 Bq L−1 and 5–10%, respectively, for 100 mL of water with 5 h of sample counting time and 30 h of blank counting time. The overall efficiency of the method is around 32% and linearity has been proven with activity concentrations between 0.9 to 9 Bq L−1.
3.3. Combustion Method for Determination of 14C Activity via LSS
Finally, measurement results of the set of soil samples spiked with
14C standard solution and prepared using a sample oxidizer are summarised in
Table 6.
As we can see in the above table, after the combustion of soil samples with organic 14C standard solution, reference and measured activity concentrations appear to be very close to each other, at either low or high activity concentrations. In this case, the relative uncertainty of measured activity concentration is below 6% at both low and high activity levels and the relative standard deviation of the results is below 5%.
It should be clarified that the absence of a result in the low activity concentration set is due to an outlier.
The results of the analysis of this set of samples are presented in
Table 7.
This table shows that in the combustion method for solid samples, with low or high activity concentration, we also achieved good results regarding accuracy (<2.58), precision (<15%) and repeatability (<15%). So, linearity was also proven.
Obtained results are acceptable, so the method can be considered validated in the interim, pending certified reference materials or intercomparison exercises. Its detection limit and uncertainty can be defined as 70 Bq kg−1 and 2–6%, respectively, for 0.25 g of solid sample, using 3 h for sample counting and 30 h for blank counting. The overall efficiency of the method is around 40% and linearity has been proven with activity concentrations of between 500 to 29,000 Bq kg−1.
In this method, the volume of the container of the combustion equipment limits the size of the sample. In the equipment employed in this work, this volume is 1 mL; the amount of sample in grams, and therefore, the detection limit in Bq g−1, will depend on the bulk density of the sample, being lower for soil samples and higher for organic samples.
3.4. Field Validation in Environmental Samples
After validating the methods described in
Section 2.2,
14C activity concentration has been determined in a set of air, water, soil, sediments, food and biota samples. Most of these belong to the Spanish National Radiological Surveillance Network for the environmental radiological monitoring of nuclear power plants and nuclear fuel cycle facilities.
In order to ensure the quality of the results obtained within the framework of the Radiological Surveillance Network, analyses are performed by two different laboratories at least: a main laboratory and a control one, which performs between 5 and 10% of the analyses. The Laboratory of Low Activity Measurements (LMBA) of the University of the Basque Country (UPV/EHU) is sometimes the main laboratory and sometimes the control one.
This allows our laboratory to carry out 14C field validation through radioanalytical methods in environmental samples, which supports internal laboratory validation and external validation through proficiency tests and interlaboratory comparisons.
In
Table 8, environmental samples utilized for the field validation and obtained relevant results are summarised.
As we can see in the table, a significant number of samples have been analysed using the 14C radioanalytical methods: 550 samples via direct counting, 652 samples via bubbling and 230 samples via combustion (i.e., 1432 samples in total).
All the results obtained are in accordance with those from the main or the control laboratory. Moreover, obtained efficiency and activity concentration relative uncertainty are very similar to those from the internal validation:
- −
Direct counting: around 65% of detection efficiency and 2–7% of uncertainty.
- −
Bubbling method: around 32% of overall efficiency and 5–10% of uncertainty.
- −
Combustion method: around 40% of overall efficiency and 2–6% of uncertainty.
However, detection limits are lower than the ones obtained in the internal validation, in order to achieve surveillance requirements. To lower them, we increased only the counting time.
4. Conclusions
14C is a meaningful radionuclide for the assessment of the impacts of nuclear activity or the fossil fuel industry on the environment. However, there are still many challenges in these procedures for determining 14C activity in environmental samples via liquid scintillation spectrometry (LSS).
Regarding the development and adaptation of the methods, we should take into account that the direct and the combustion methods described in this work must deal with spectral interference issues (beta-emitters with energy spectra overlapping those of 14C) and are limited by the small sample volumes (10 mL and 1 mL, respectively) that can be utilized in each method.
To solve spectral interferer issues, the bubbling method is an adequate approach, as it is more selective than the direct method but is limited to water and air samples.
With regard to the accreditation of the methods for
14C activity determination in environmental samples via LSS under the ISO/IEC 17025 [
11], the lack of reference materials and the few and limited proficiency tests and interlaboratory comparisons lead to laboratories having to design their own internal validation strategies and to prepare fit for purpose materials.
In this work, the validation strategy for three different radioanalytical methods to determine 14C activity in environmental samples via LSS—the direct counting and bubbling methods, both based on the ISO 13162:2021 standard, for water samples, and the combustion of solid samples using a sample oxidizer—has been described and analysed.
In all the methods, the studied quality parameters (accuracy, precision and repeatability) are within the acceptance range at two activity levels (around 10 to 100 times the detection limits), which enabled us to assess their detection limits, uncertainties and linearity, despite their limitations.
In the direct method for water samples, the detection limit obtained was 20 mBq for 10 mL of water; in the bubbling method for water and air samples the detection limit was 14 mBq for 100 mL of water or 2.6 m3 of air; and in the combustion method for solid samples the detection limit was 17.5 mBq when using a sample of 0.25 g.
In all the methods, relative uncertainty was below 10% and linearity was proven with a rather large range of activity concentrations (from around 5 to 100 times the detection limit), within those found in environmental samples. Regarding overall method efficiencies, the highest was that of the direct method (65%), as there was no chemical procedure and the water sample was simply mixed with the scintillation cocktail, whereas the lowest was obtained via the bubbling method (32%). Finally, regarding counting times, to achieve the detection limits found, the shortest time was required for the direct counting method (100 min), whereas longer times were required for the bubbling and combustion methods (5 and 3 h, respectively).
These outcomes are supported by a field validation conducted within the framework of the Spanish National Radiological Surveillance Network, as the results of each 14C radioanalytical method are in accordance with the ones from the other laboratory.
In conclusion, in spite of all the aforementioned difficulties, our analytical procedures for
14C activity determination via LSS have been validated and accredited under ISO/IEC 17025 [
11].