Implementation of an Automated Cerebrospinal Fluid Drainage System for Early Mobilization in Neurosurgical Patients

by
Sebastian Arts
*,
Martine van Bilsen
,
Erik J. van Lindert
,
Ronald HMA Bartels
,
Rene Aquarius
and
Hieronymus D. Boogaarts
Department of Neurosurgery, Radboud University Medical Center, 6525GA Nijmegen, The Netherlands
*
Author to whom correspondence should be addressed.
Brain Sci. 2021, 11(6), 683; https://doi.org/10.3390/brainsci11060683
Submission received: 25 April 2021 / Revised: 15 May 2021 / Accepted: 18 May 2021 / Published: 22 May 2021
(This article belongs to the Section Neurosurgery and Neuroanatomy)

Abstract

:
Background: Automated cerebrospinal fluid (CSF) drainage systems allow for the mobilization of patients with an external CSF drain. The aim of this study is to describe the implementation of an automated CSF drainage system in neurosurgical patients with external CSF drains. Methods: A feasibility study was performed using an automated CSF drainage system (LiquoGuard®7, Möller Medical GmbH, Fulda, Germany) in adult neurosurgical patients treated with external lumbar or external ventricular drains between December 2017 and June 2020. Limited mobilization was allowed—patients were allowed to adjust their inclined beds, sit in chairs and walk under the supervision of a nurse or physical therapist. The primary outcome was the number of prematurely terminated drainage sessions. Results: Twenty-three patients were included. Drainage was terminated prematurely in eight (35%) patients. In three (13%) of these patients, drainage was terminated due to signs of hydrocephalus. Pressure-controlled drainage in patients with external lumbar drains (ELD) showed inaccurate pressure curves, which was solved by using volume-controlled drainage in ELD patients. Conclusion: The implementation of an automated CSF drainage system (LiquoGuard®7) for CSF drainage allows for early mobilization in a subset of patients with external CSF drains. External lumbar drains require volume-based drainage rather than differential pressure-dependent drainage.

1. Introduction

Temporary treatment of hydrocephalus using external lumbar or ventricular drainage is common in neurosurgical practice. Conventional hydrostatic pressure-controlled drainage with dripping chambers are the gold standard [1]. However, using dripping chambers restricts patient mobility and can cause unintentional over- or underdrainage during Valsalva manoeuvres or patient movement. Patients are therefore immobilized during drainage.
It has been shown that early mobilization decreases the complications associated with immobilization, results in a shorter length of hospital stay and seems to be safe for patients with an external cerebrospinal fluid (CSF) drain [2,3,4,5,6,7]. To prevent over- and underdrainage, external drains are clamped during mobilization and require continuous intracranial pressure monitoring (ICP), which limits mobilization time [5,6,7].
The LiquoGuard®7 (Möller Medical GmbH, Fulda, Germany), is an automated cerebrospinal fluid (CSF) drainage system that has been available since 2011. It acts as a self-adjusting, pressure-based resistor that allows CSF to drain in a controlled manner, and is thus intended to prevent over- or underdrainage [1]. The system is designed to enable early mobilization of patients. Its use is described in patients requiring CSF drainage due to thoracoabdominal aortic aneurysm repair and in neurosurgical patients requiring ICP pressure regulation [1,8,9,10]. However, to our knowledge, its capacity to facilitate the mobilization of neurosurgical patients has not been previously described.
The aim of this study is to describe the implementation of the LiquoGuard®7 in mobilized hydrocephalus patients with external CSF drains.

2. Materials and Methods

This feasibility study was performed in accordance with the ethical standards of the institutional board review (NL62883.091.17). From December 2017 until June 2020, adult neurosurgical patients (>18 years) treated with an external ventricular drain (EVD, CODMAN® BACTISEAL® EVD Catheter, Johnson & Johnson, Raynham, MA, USA) or an external lumbar drain (ELD, Perifix® epidural catheter, B Braun, Melsungen AG, Germany) were included.
Two LiquoGuard®7 systems (Möller Medical GmbH, Fulda, Germany) were used. The device is an automated pressure-based CSF drainage system connected to the EVD or ELD catheter. It measures ICP by using a pressure sensor that is fixed by an electrocardiogram sticker on the skin over the temporal bone at the height of the foramen of Monro (Figure 1) [9].
The manufacturer gave instruction on the theoretical background of the device and its practical use to all caregivers. The first author (S.A.) and a nurse practitioner were available to answer remaining questions during the implementation and to provide assistance to daily caregivers. User manuals, in the form of flowcharts, were also attached to each system.
The placement of the EVDs and ELDs was performed according to the institutional standard insertion protocol. At the intensive care unit (ICU), patients were treated with a conventional drainage system (Duet™ External Drainage and Monitoring System, Medtronic, Minneapolis, MN, USA). After transfer to the neurosurgical ward, the LiquoGuard®7 was installed if written and verbal informed consent was obtained from patients, or their legally authorised representatives. Patients were included in the study on a first-come, first-served basis, due to restricted availability of the system.
Two drainage modes are available for the LiquoGuard®7—pressure-regulated drainage and volume-regulated drainage. In this study, pressure-regulated drainage was used.
In pressure-regulated drainage, the system drains CSF as soon as intracranial pressure exceeds the pre-set pressure value. The device produces an audible alarm when pre-set alarm limits are exceeded. Our conventional system is calibrated at 10 cm above the foramen of Monro per institutional protocol. Accordingly, the used pre-set pressure value for the LiquoGuard®7 was set to 10 cmH2O, with an alarm range between −10 cmH2O and 20 cmH2O. In volume-regulated drainage, the system drains a pre-set volume of CSF per hour, regardless of intracranial pressure. According to the instructions for use, the diameter of the lumen of the EVD and ELD catheters used needed to be more than 0.7 mm in order to measure CSF pressure correctly.
Mobilization was encouraged as much as was clinically possible. Patients received physical therapy once a day and were seated for at least thirty minutes, three times a day. Furthermore, patients were allowed to adjust their inclined bed and walk under the supervision of a nurse.
The primary outcome was the number of prematurely terminated drainage sessions with the LiquoGuard®7. Reasons for termination were described. Complications were included as secondary outcome measures and were registered if they occurred between the placement of the first drain and 30 days after removal of the last drain.
Complications were defined according to our earlier study regarding complications of external CSF drainage in aneurysmal subarachnoid haemorrhage (aSAH) [11], and were divided into direct and indirect complications. Direct complications were defined as complications with a direct relation to the external CSF drain (i.e., drain dislodgement, drain occlusion and meningitis). Indirect complications were defined as complications that could not be directly related to the external CSF drainage. These indirect medical complications were divided into four subgroups: infection, delirium, pressure injuries and thromboembolic complications. The number of internal shunts after external drain placement was also registered. Thromboembolic processes were subdivided into deep-vein thrombosis and pulmonary embolisms. Deep-vein thrombosis consisted of a confirmed diagnosis by echo-Doppler, while for pulmonary embolisms a confirmed diagnosis by spiral-CT scan was required. Delirium was registered as a complication when patients had clinical signs of delirium in accordance with the Delirium Observation Scale, for which haloperidol was given [12]. Pressure injuries were defined by the pressure injury grading score as stated by the National Pressure Ulcer Advisory Panel in 2016 [13]. Infections were only registered as complications if antibiotic treatment was started. An infection was detected by monitoring the clinical condition of the patient combined with a rising C-reactive protein and leukocyte count or positive cultures. The EVDs or ELDs were considered to be dislodged when the drains were inadvertently partially or entirely removed. Occlusion was registered as a complication if drain reimplantation was needed.
Patient-specific data were retrieved from the digital patient information system (Epic Systems Corporation (2014), Madison, WI, USA). Information regarding demographics, drainage period, drain type, length of hospital stay, destination after discharge and complications were prospectively registered.
All data were analysed using IBM SPSS Statistics for Windows (Version 22.0. IBM Corp., Armonk, NY, USA). Continuous data were presented as median and range. Categorical data were presented as counts and percentages.

3. Results

Twenty-three patients were included (Table 1). Nineteen EVDs and five ELDs were connected to the LiquoGuard®7 (one patient received an ELD after the EVD was removed). The median time from drain placement until mobilization (connection to the LiquoGuard®7) was three (0–20) days, with a median mobile period of six days (Table 1).
Twenty-eight complications occurred in 14 patients (61%). A total of nine direct complications and 19 indirect complications occurred in these patients, with a median of one (0–4) complication per patient (Figure 2, Table 2).
One drain dislodgement occurred while connected to the LiquoGuard®7. LiquoGuard®7′s tubing set occluded twice. In these cases, the LiquoGuard®7 was removed instead of being replaced.
The LiquoGuard®7 was disconnected prematurely in eight patients (35%) (Table 3). The first patient treated with the system became clinically unstable due to multi-organ failure and was reconnected to the conventional system. Any causal link with the device was excluded.
Although the lumen of the ELD tubing set used was more than 0.7 mm, it was observed that the pressure curves were not accurate in the first patient with an ELD. No pulsation was measured, and the pressures shown were constantly negative, even in supine position. As a result, the device did not drain the desired amount of CSF, which lead to underdrainage. After contacting a technical expert from the manufacturer, the drainage mode in patients treated with an ELD was set to volume-regulated with a pre-set volume of 10 mL/h, which solved the problem.
Despite pressure or volume measurements within limits and normal pulsations, three patients neurologically deteriorated while being drained with the LiquoGuard®7 (Table 4). All three patients showed clinical improvement after changing to a conventional drainage system.

4. Discussion

Neurological deterioration due to hydrocephalus occurred three times despite drainage systems being functional. This suggested that settings different from a normal overflow system, depending on hydrostatic pressure, might be required.
Patients with conventional external CSF drainage systems tend to moderately over-drain CSF, which occurs especially during manoeuvres that cause a short increase in intracranial pressure, for example coughing, leg elevation, positional changes or speaking. This overdrainage is prevented by the LiquoGuard®7, which may contribute to the observed neurological deterioration in some patients. Measuring ICP exclusively is possibly insufficient for maintaining CSF homeostasis, due to normal pressure hydrocephalus in aSAH patients [14,15]. Since the majority of patients in this series suffered from aSAH, normal pressure hydrocephalus might have contributed to the relatively high number of premature disconnections.
It may be that standardized pressure settings in these patients are insufficient. To overcome this problem, lowering the pressure settings stepwise or switching to volume-controlled drainage may help to drain sufficient amounts of CSF.
Additionally, when patients are in an upright position the physiological intracranial pressure can become negative, varying with exercise or movement [16,17]. This may result in underdrainage during mobilization, as pressure settings remain unchanged during mobilization. One would expect this relative underdrainage to be corrected during bed rest. However, this correction did not occur. This supports the idea that the regulation of CSF dynamics is complex and cannot always be limited exclusively to pressure or flow control in a mobilized patient. Continuous flow may be very important for maintaining an equilibrium in CSF homeostasis.
Although inaccurate pressure curves are a well-known phenomenon, the reason for inaccurate pressure curves in patients treated with an ELD is unknown. It is remarkable that previous literature regarding thoracic surgery shows adequate pressure measurements in lumbar drainage [8,10]. Although not described, it may be the case that, in these studies, pressure sensors were attached near the ELD insertion place. Since patients were mobilized in our study, the pressure sensor was attached at the height of the foramen of Monro, so pressure waves might not have been conducted well through the longer catheter. A second possible explanation is that intracranial haemorrhage compartmentalization of CSF may occur, which results in a discrepancy between measured spinal CSF pressure and intracranial CSF pressure.
Previous studies describe that the LiquoGuard®7 shows inappropriate curves and drains an inadequate amount of fluid from patients with slit ventricles [1,9]. In this study, no patients suffered from slit ventricles and, therefore, no experience was gained on this potential problem.
It is unclear which patient will benefit the most from automated CSF drainage with the LiquoGuard®7. Pressure and volume settings seem patient specific and underlying factors are currently unknown, which hampers patient selection for automated CSF drainage.
The incidence of meningitis is relatively high and not related to drainage time [18]; however, due to the small number of patients, no conclusions can be drawn concerning the number of complications. However, we have reported on all types of complications for transparency reasons. Theoretically, using the LiquoGuard®7 could lead to a decline in complications since mobilized patients might be less prone to secondary complications [2,3,5,6,7].
While shunt dependency in this study seemed slightly higher in comparison to our earlier series regarding complications in patients treated with a conventional EVD or ELD system (22% in this study versus 15%) [11], future studies should compare patients that receive automated CSF drainage and drainage with a conventional system. The sample size of the present study is small; however, the presented results are beneficial for other neurosurgical departments planning to implement the LiquoGuard®7 in their clinical practise.

5. Conclusions

The implementation of an automated CSF drainage system (LiquoGuard®7) for CSF drainage allows early mobilization in a subset of patients with an external CSF drain. External lumbar drains require a volume-based drainage rather than a differential pressure-dependent drainage.

Author Contributions

Conceptualization, H.D.B. and E.J.v.L.; methodology, S.A., R.A. and H.D.B.; investigation, S.A.; data curation, S.A.; writing—original draft preparation, S.A., R.A. and H.D.B.; writing—review and editing, all authors; visualization, M.v.B.; supervision, R.H.B.; funding acquisition, H.D.B. All authors have read and agreed to the published version of the manuscript.

Funding

This study was co-funded by the Dutch Healthcare Insurance Company, VGZ. The authors do not have any personal or institutional financial interest.

Institutional Review Board Statement

The study was conducted according to the guidelines of the Declaration of Helsinki and approved by the Institutional Review Board (or Ethics Committee) of Radboud University Medical Center (NL62883.091.17).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Data can be requested by sending an email to the corresponding author.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Linsler, S.; Schmidtke, M.; Steudel, W.I.; Kiefer, M.; Oertel, J. Automated intracranial pressure-controlled cerebrospinal fluid external drainage with LiquoGuard. Acta Neurochir. 2013, 155, 1589–1594. [Google Scholar] [CrossRef]
  2. Dittmer, D.K.; Teasell, R. Complications of immobilization and bed rest. Part 1: Musculoskeletal and cardiovascular complications. Can. Fam. Physician 1993, 39, 1428–1432, 1435–1437. [Google Scholar]
  3. Teasell, R.; Dittmer, D.K. Complications of immobilization and bed rest. Part 2: Other complications. Can. Fam. Physician 1993, 39, 1440–1442, 1445–1446. [Google Scholar]
  4. Karic, T.; Roe, C.; Nordenmark, T.H.; Becker, F.; Sorteberg, W.; Sorteberg, A. Effect of early mobilization and rehabilitation on complications in aneurysmal subarachnoid hemorrhage. J. Neurosurg. 2017, 126, 518–526. [Google Scholar] [CrossRef] [Green Version]
  5. Yataco, R.A.; Arnold, S.M.; Brown, S.M.; David Freeman, W.; Carmen Cononie, C.; Heckman, M.G.; Partridge, L.W.; Stucky, C.M.; Mellon, L.N.; Birst, J.L.; et al. Early Progressive Mobilization of Patients with External Ventricular Drains: Safety and Feasibility. Neurocrit. Care 2018. [Google Scholar] [CrossRef]
  6. Moyer, M.; Young, B.; Wilensky, E.M.; Borst, J.; Pino, W.; Hart, M.; LoBreglio, J.; Zaleski, D.; Leonor, I.; Kung, D.; et al. Implementation of an Early Mobility Pathway in Neurointensive Care Unit Patients with External Ventricular Devices. J. Neurosci. Nurs. 2017, 49, 102–107. [Google Scholar] [CrossRef]
  7. Young, B.; Moyer, M.; Pino, W.; Kung, D.; Zager, E.; Kumar, M.A. Safety and Feasibility of Early Mobilization in Patients with Subarachnoid Hemorrhage and External Ventricular Drain. Neurocrit. Care 2019, 31, 88–96. [Google Scholar] [CrossRef]
  8. Kotelis, D.; Bianchini, C.; Kovacs, B.; Muller, T.; Bischoff, M.; Bockler, D. Early experience with automatic pressure-controlled cerebrospinal fluid drainage during thoracic endovascular aortic repair. J. Endovasc. Ther. 2015, 22, 368–372. [Google Scholar] [CrossRef]
  9. Kwon, Y.S.; Lee, Y.H.; Cho, J.M. Early Experience of Automated Intraventricular Type Intracranial Pressure Monitoring (LiquoGuard(R)) for Severe Traumatic Brain Injury Patients. Korean J. Neurotrauma 2016, 12, 28–33. [Google Scholar] [CrossRef] [Green Version]
  10. Tshomba, Y.; Leopardi, M.; Mascia, D.; Kahlberg, A.; Carozzo, A.; Magrin, S.; Melissano, G.; Chiesa, R. Automated pressure-controlled cerebrospinal fluid drainage during open thoracoabdominal aortic aneurysm repair. J. Vasc. Surg. 2017, 66, 37–44. [Google Scholar] [CrossRef] [Green Version]
  11. Arts, S.; van Lindert, E.J.; Aquarius, R.; Bartels, R.; Boogaarts, H.D. Complications of external cerebrospinal fluid drainage in aneurysmal subarachnoid haemorrhage. Acta Neurochir. 2021. [Google Scholar] [CrossRef]
  12. Schuurmans, M.J.; Shortridge-Baggett, L.M.; Duursma, S.A. The Delirium Observation Screening Scale: A screening instrument for delirium. Res. Theory Nurs. Pract. 2003, 17, 31–50. [Google Scholar] [CrossRef]
  13. Edsberg, L.E.; Black, J.M.; Goldberg, M.; McNichol, L.; Moore, L.; Sieggreen, M. Revised National Pressure Ulcer Advisory Panel Pressure Injury Staging System: Revised Pressure Injury Staging System. J. Wound Ostomy Cont. Nurs. 2016, 43, 585–597. [Google Scholar] [CrossRef] [Green Version]
  14. Yamada, S.; Ishikawa, M.; Yamamoto, K.; Ino, T.; Kimura, T.; Kobayashi, S.; Japan Standard Stroke Registry Study, G. Aneurysm location and clipping versus coiling for development of secondary normal-pressure hydrocephalus after aneurysmal subarachnoid hemorrhage: Japanese Stroke DataBank. J. Neurosurg. 2015, 123, 1555–1561. [Google Scholar] [CrossRef] [Green Version]
  15. Gruber, A.; Reinprecht, A.; Bavinzski, G.; Czech, T.; Richling, B. Chronic shunt-dependent hydrocephalus after early surgical and early endovascular treatment of ruptured intracranial aneurysms. Neurosurgery 1999, 44, 503–509. [Google Scholar] [CrossRef]
  16. Haykowsky, M.J.; Eves, N.D.; DE, R.W.; Findlay, M.J. Resistance exercise, the Valsalva maneuver, and cerebrovascular transmural pressure. Med. Sci. Sports Exerc. 2003, 35, 65–68. [Google Scholar] [CrossRef]
  17. Brimioulle, S.; Moraine, J.J.; Norrenberg, D.; Kahn, R.J. Effects of positioning and exercise on intracranial pressure in a neurosurgical intensive care unit. Phys. Ther. 1997, 77, 1682–1689. [Google Scholar] [CrossRef]
  18. Jamjoom, A.A.B.; Joannides, A.J.; Poon, M.T.; Chari, A.; Zaben, M.; Abdulla, M.A.H.; Roach, J.; Glancz, L.J.; Solth, A.; Duddy, J.; et al. Prospective, multicentre study of external ventricular drainage-related infections in the UK and Ireland. J. Neurol. Neurosurg. Psychiatry 2018, 89, 120–126. [Google Scholar] [CrossRef] [Green Version]
Figure 1. Set-up of the LiquoGuard®7 in a patient treated with an EVD. The pressure sensor is attached to the skin by an electrocardiogram sticker at the height of the foramen of Monro.
Figure 1. Set-up of the LiquoGuard®7 in a patient treated with an EVD. The pressure sensor is attached to the skin by an electrocardiogram sticker at the height of the foramen of Monro.
Brainsci 11 00683 g001
Figure 2. The number of complications per patient.
Figure 2. The number of complications per patient.
Brainsci 11 00683 g002
Table 1. Demographics data are presented as counts or median and range.
Table 1. Demographics data are presented as counts or median and range.
Total (n)
Number of Patients23
Gender (M:F) *9:14
Age59 (26–76)
DiagnosisSAH15 (65%)
Tumour1 (4.3%)
ICH *5 (22%)
Ventriculitis After VPD Placement2 (8.7)
Type of DrainEVD17
ELD4
Both2
Number of Drains1 (1–2)
Total Drainage Period (days)13 (7–41)
Drainage Period LG (days)6.0 (1–16)
Drainage Period before Connected to LG (days)3.0 (0–20)
Length of Stay (days)18 (12–45)
ASA *2 (1–3)
VPD *5
* M: male; F: Female; ICH: intracerebral haemorrhage; LG: LiquoGuard®7; ASA: American Society of Anaesthesiology score; VPD: ventriculoperitoneal drain.
Table 2. Complications.
Table 2. Complications.
ComplicationGroupCounts (%)
Direct (Drain-Related)Meningitis5 (13%)
Dislodgement3 (7.9%)
Occlusion1 (2.6%)
Indirect (Medical)Infection10 (26%)
Delirium2 (5.3%)
Pressure Injuries5 (13%)
Thromboembolic Process2 (5.3%)
Total28 (100%)
Table 3. Reasons for LiquoGuard®7 removal.
Table 3. Reasons for LiquoGuard®7 removal.
NumberPercentage
Successful Challenge1357%
Clinically Instable14.3%
Technical Problems14.3%
Signs of Hydrocephalus313%
Distal Occlusion28.7%
Drain Dislodgement14.3%
Other *28.7%
* Other consists of one VPD insertion and one discharge to another department.
Table 4. The characteristics of patients that showed signs of hydrocephalus during drainage with the LiquoGuard®7.
Table 4. The characteristics of patients that showed signs of hydrocephalus during drainage with the LiquoGuard®7.
PatientAgeDiagnoseWFNSEVD/ELDNumber of DrainsDP *DP * LG *SymptomsNumber of Complications
165aSAH3EVD1196Bradyphrenia, apraxia and ventriculomegaly 2
261aSAH1ELD21916Headache and CSF leakage along drain3
372PMH *N/AELD181Headache0
* DP: drainage period; LG: LiquoGuard®7; PMH: perimesencephalic haemorrhage.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Arts, S.; van Bilsen, M.; van Lindert, E.J.; Bartels, R.H.; Aquarius, R.; Boogaarts, H.D. Implementation of an Automated Cerebrospinal Fluid Drainage System for Early Mobilization in Neurosurgical Patients. Brain Sci. 2021, 11, 683. https://doi.org/10.3390/brainsci11060683

AMA Style

Arts S, van Bilsen M, van Lindert EJ, Bartels RH, Aquarius R, Boogaarts HD. Implementation of an Automated Cerebrospinal Fluid Drainage System for Early Mobilization in Neurosurgical Patients. Brain Sciences. 2021; 11(6):683. https://doi.org/10.3390/brainsci11060683

Chicago/Turabian Style

Arts, Sebastian, Martine van Bilsen, Erik J. van Lindert, Ronald HMA Bartels, Rene Aquarius, and Hieronymus D. Boogaarts. 2021. "Implementation of an Automated Cerebrospinal Fluid Drainage System for Early Mobilization in Neurosurgical Patients" Brain Sciences 11, no. 6: 683. https://doi.org/10.3390/brainsci11060683

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop