An Artificial Sweating System for Sweat Sensor Testing Applications

1
School of Engineering and Technology, Central Michigan University, Mt Pleasant, MI 48859, USA
2
School of Computer Science and Engineering, Sacred Heart University, Fairfield, CT 06825, USA
*
Author to whom correspondence should be addressed.
Electronics 2019, 8(6), 606; https://doi.org/10.3390/electronics8060606
Submission received: 17 April 2019 / Revised: 13 May 2019 / Accepted: 28 May 2019 / Published: 30 May 2019
(This article belongs to the Section Bioelectronics)

Abstract

:
This research proposes a completely automated, computer-controlled fluid mixing and dispensing system, which is suitable for testing sweat sensing devices, as an alternative to requiring human trials during the development phase of a sweat sensor device. An arm mold was designed and implemented with dragon skin and pores to simulate sweating action. The relay controlled mixing tanks allow for the different concentration of fluid solutions at various rates of fluid dispensing through pores. The onboard single board computer controls a dozen electronic relays and it switches and presents an easy to use graphical user interface to allow end users to conduct the experiments with ease and not require further programming. With the recent advances in sweat sensors, this platform offers a unique way of testing sensing devices during development, allowing for researchers to focus on their design parameters one at a time before actual validation through human trials are conducted. The current device can provide sweat rates from 1 µL/min to 500 µL/min. Furthermore, concentrations of 10 mM up to 200 mM of salt concentrations were able to be repeatedly produced. In an ANOVA test with salt concentrations varying from 40–60 mM, a p-value of 0.365 shows that the concentration does not have any effect on the flow rate. Similarly, a p-value of 0.329 and 0.167 for different relative humidity and temperature shows that the system does not present a statistical difference. Lastly, when the interactions among all the factors were considered, a p-value of 0.416 clearly presents that the system performance is insensitive to different factors, thus validating the system reliability.

1. Introduction

Sweating is a natural human phenomenon that regulates core body temperature. Eccrine sweat glands that are distributed along the body are responsible for cooling the human body via thermoregulation. Constituents of sweat along with its mechanisms have been the interest of researchers since the early 20th century [1,2,3]. In the last few decades, sweat related research included sweat content detection, such as pH [4,5], Sodium [6,7,8], Lactate [9,10], conductivity [11], alcohol [12], as well as sweat rate detection [13,14]. Although sweat information can provide important information, this data would be more beneficial if it was provided in real-time. Therefore, similar to any physiological activity monitoring systems, such as heart rate or temperature, portable and real-time sweat monitoring devices were explored [11,15,16,17].
With the real-time sweat monitoring devices on the rise, sweat collection methods become more crucial. Traditional methods, such as whole body wash [18] or absorbent pads [19] cannot be used for these portable devices, where only microliters of sweat are necessary. Therefore, significant efforts were spent on first modifying the traditional sweat collection methods to fit into a smaller form-factor for portability, such as Macroduct [16,20] (a commercial sweat collection device that was developed for Cystic Fibrosis diagnosis on infants [21]), fabric wicking [6], and modified absorbent pads [22]. Currently, devices that are directly exposed to the skin were developed as tattoos [9,23] and flexible substrates [15,24]. Recently, a conductive thread-based textile sweat sensor was developed that has a novel approach of using an embedded conductivity sensing into the fabric; eliminating the need of sweat collection [25].
There has yet to be an established method of testing these devices despite the extensive device development attempts. Obviously, human subject tests are the golden standard to validate a device for real-time human sweat monitoring. However, using human subjects during the development of the device itself is time-consuming and characterization cannot be performed due to the variance of sweating mechanisms, not only from subject to subject, but also different regions or times of sweat collection from the same subject [26]. Most of the researchers settle with traditional methods of preparing artificial sweat in various concentrations in advance and running the sweat solution through a syringe pump to simulate the rate of sweat [16,27]. A standardized testing platform with an automatically controlled sweat concentration generation system would help to accelerate the development of sweat collection, concentration, and sweat rate devices. It should be noted here that this kind of platform does not eliminate the need for human trials. It is known that hundreds of preliminary testing would be needed before human trials during wearable device design and prototyping. A testing platform would significantly reduce the development testing phase of a sweat sensor device prototyping. The validation would still need to be done via human trials. However, the premature testing of sweat sensor prototypes on humans complicate the device optimization, as various parameters need to be taken into account before making sure that the device operates as intended. A standardized testing platform would allow for researchers to explore more ideas and prototype versions before validating the best design on humans. For example, Heikenfeld’s group has developed an artificial skin model that is useful in characterizing sweat collection devices [28]. This model is suitable for sweat collection research and it could potentially be used in a testing platform that is proposed in this article. Similarly, textile-based devices [25] could also be more realistically tested on a sweating platform. Furthermore, other flexible sensing prototypes, such as breathing rate devices based on moisture sensing, could also be tested on the sweating arm model [29].
Sweating occurs through sweat glands that are distributed throughout the human body [17,30]. Sweat is secreted in sweat coils under the skin and is brought to the skin surface through the sweat ducts that have pores of around 50 µm. The main purpose of eccrine sweat glands is to regulate body temperature through cooling. As the sweat evaporates through the skin, the temperature of the skin is lowered to help cool down the core body temperature. A number of sweat glands in a body region vary significantly between 150 glands/cm2 (forearm) to 600 glands/cm2 (hand) [31]. Therefore, testing a sweat sensing device in various regions of the body produces different results and it should be accounted for during validation.
In this article, we propose an automated and computer controlled testing environment for sweat monitoring devices. Through a simple graphical user interface and the close replication of a human arm, the proposed system allows for the testing of sweat monitoring devices in different shapes and sizes. With an emphasis on user intuitiveness, the system was designed, whereby the user is not required to program to test the device. Furthermore, extensive statistical experiments have been conducted to validate reliable system operation in various conditions.

2. Materials and Methods

The primary element used in the proposed system is the testing solution that was obtained by combining saline solution and water in a mixing tank, both of which are initially stored in individual 1.5-L buckets (Marinecolor® Acrylic Made Liquid Storage Bucket from Amazon (Seattle, WA, USA). Both the water and saline solutions are extracted from their respective tanks through a brushless DC motor (Anself Ultra-quiet Mini DC12V Micro Brushless Water Oil Pump 240 L/H 5W Lift 3M from Amazon (Seattle, WA, USA) and a 0.25 inch quick connect check valve (Hydronix from Amazon (Seattle, WA, USA) that is attached at bottom of each tank. The extracted liquids are mixed in two steps to obtain a uniform concentration of the testing solutions. In the first step, both the extracted liquids from the pumps are sprayed into the mixing tanks via multiple three-dimensional (3D) printed sprinkler modules (Stratasys Objet30 Pro and PolyJet VEROBLUE RGD840 as the printing material). This uniform spraying mixes both liquids as they gravitate to the bottom. In the second step, a peristaltic DC pump (Langer Instruments—Model: BQ50-1J-A, Tucson, AZ, USA) placed in the middle of the tank to extract and pump the solution on a continual basis to obtain the desired uniform concentration and flow rates of 0.25 µL/min to 20 mL/min in the testing solution. Once the testing has completed, the solution is drained through the brass liquid solenoid valves (Adafruit—product ID #996, New York City, NY, USA) to the bottom trash bin. The custom-made metal controller box (Diarco model 16–24 hand press brake was used) was used to place all of the electronic components where a metal frame was slid into the box.
The automation of the test setup was accomplished by utilizing a Raspberry Pi 3 single board computer (SBC) to control the onboard electronics, conduct data analysis, and present the user interface. The concentration of the mixed testing solution was tested on a continual basis while using the Atlas Scientific EZO-EC conductivity sensor (Amazon, Seattle, WA, USA). The analog voltage signal from the conductivity sensor was communicated to the SBC via the conductivity circuit. If the concentration of the testing solution is not at the desired level, the appropriate motors from a water tank or saline solution tank are activated to pump more liquid into the testing solution. An eight-channel DC 5 V relay module (JBtek 8 Channel DC 5 V Relay Module from Amazon (Seattle, WA, USA) was used to control each pump in order to provide sufficient current to drive the motors and electronically isolate the pumps from SBC. Reliable pump and sprinkler operation was ensured via a separate 110 V to 5 V (10 A) and 100 V to 12 V (10 A) switching power supplies, respectively. The analog signal that is required to control the pumps was generated via the MCP4725 12-bit digital to analog converter and the SBC. When the desired concentration of the testing solution is obtained, it is pumped to the arm model via a 0.25-inch microflow silicone tubing, which is optimum for the required flow speeds with the peristaltic pump that is used in this setup. This arm was developed using Dragon Skin 10, and body double fast set lifecasting platinum silicone accompanied by the universal mold release (Smooth-On, Lehigh, PA, USA). Dragon Skin 10 was chosen for its quick cure time (30 min) and flexible surface properties, which are close to human skin (shore A hardness of 10 A).
The program for the SBC was written in Python and the graphical user interface (GUI) was created by utilizing the general-purpose Tkinter module. The program allows for (a) manual control of all pumps, (b) the ability to continually check saline conductivity, (c) a cleaning function, and (d) the ability to run a trial at a desired conductivity and flow rate.

3. System Design

Figure 1a shows the overall view of the sweating arm prototype. There are three tanks that are responsible to mix the set amount of concentration and volume. Water tank only has DI (deionized) water, whereas salt tank has a predetermined, high concentration salt solution. The tank in the middle is the mixing tank in which the water and salt solutions are proportionally added, depending on the desired salt concentration. The amount of mixing solution (i.e., the sweat) can also be adjusted. The desired salt concentration and volume are controlled by turning on relays (R2 and R3) at certain time intervals to make sure the correct amount of solution was added. A control feedback loop was also incorporated into the mixing tank to make sure that the salt concentration is at the right level. A salinity sensor (illustrated as “probe” in Figure 1a) was used for this purpose; a lower than desired salt concentration triggers the relay from the salt tank in order to increase the salt concentration of the mixing tank, whereas the relay from the water tank was initiated if a higher than desired salt concentration was detected. Two mixing pumps that were placed in the bottom of the salt and mix tanks (R1 and R4) were used to continuously stir the solutions to prevent aggregation.
A peristaltic pump was used to bring the solution to the arm at a specified flow rate (i.e., the sweat rate) once the mixing solution is ready. The tubing from the peristaltic pump goes through a multiplier joint to come to the arm from close holes that represent the sweat pores on the skin. A two-step flushing system was designed and implemented. The tanks were first flushed by enabling the relays at the bottom of each tank to a water bin (R5 and R6). Once the flushing is done, the tanks were rinsed from the top by turning on R7, which allows water to run through the 3D-printed spray heads on the top, so as to make sure tank walls are not contaminated with residual salt water for another test.
Figure 1b gives the circuit schematic for the electronics. A peristaltic pump is controlled via a digital to analog converter (DAC), as the pump works with analog inputs. The probe information is brought to the SBC via a conductivity circuit. Both DAC and the conductivity circuit are connected through the I2C bus control on the SBC. An eight-bit parallel connection to the relay array controls the individual relays for aggregation (R1 and R4), transfer (R2 and R3), drain (R5 and R6), and sprinkle (R7) operations. An eighth valve (R8) was used to activate the cooling fan for the electronics in the box.
Figure 2a shows the overall physical view of the system. The arm was placed on a stand (C, M, and U letters, representing Central Michigan University) for easy access. The tanks were placed on an in-house made metal stand to accommodate flushing valves. Figure 2b shows a zoomed view with sprinkler heads that were used to rinse the salt and mixing tanks. The electronics and relays were placed in a metal-bent box to eliminate connection issues, as shown in Figure 2c. The box was designed by considering the electric codes. The system was housed on a cart for portability and was ready to be plugged into a keyboard, mouse, and monitor.
A Graphical User Interface (GUI) was built for the ease of operation as shown in Figure 3, which eliminates the need for programming by the end user. The interface allows for the manual control of all pumps, checking saline conductivity, a cleaning function, and run a trial at a specific concentration and flow rate.

4. Implementation & Results

In order to characterize the functionality of the sweating arm system, first, the peristaltic pump was tested. The pump speed in revolutions per minute (rpm) was changed to measure the DI flow rate that comes out the pores of the sweating arm. Letting the arm sweat for a certain time and collecting the resulting DI water and measuring the volume was done to collect the tests. These measurements were conducted in a way where the sweating tubes were collected in a polystyrene weighing dishes and the volume was measured from the weight of the dish as it is filled with sweat. A high resolution weighing scale was used and measurements were done in a closed chamber to prevent water evaporation. The pump speed was directly controlled from the GUI that is shown in Figure 3. The results were given in µL/min to put it into context with the standard notation of sweat research.
It can be seen from Figure 4 that the peristaltic pump provides a quadratic control over the flow rate in interest. A second order polynomial was fit (root mean square value of 0.8901, R2 = 0.9955, and standard deviation of 0.9614) to the data for the slow flow rates (FR) for varying pump speeds (rpm):
FR = 1.56 rpm2 + 7.63 rpm − 0.03
The pump speed can be increased to 50 rpm for high flow rates, as shown in Figure 4 inset, which would result in a mostly linear increase. However, it is important to note that the flow rate range for athlete sweat rates is less than 50 µL/min [12]. It is clear that the arm prototype is able to create fluid dispersion at a rate that is compatible with human sweat rate.
Another aspect of the sweating arm is the controlled concentration of liquids. The sweating arm was designed to provide different concentrations of salt as proof of concept. The concentration was measured via the salinity sensor and was calibrated real time to reach the desired concentration. The desired solution was generated using the feedback that was received from the salinity sensor, which was set to 3% error range. Therefore, the sweating arm system would stop the electroconductivity calibration once the set concentration and desired concentration values are within 3%. A commercial sweat salt checker (Sweat-Chek) was utilized in order to further validate the sweating arm-generated solutions. Once the sweating arm generated the desired solution through the input to the GUI, the concentration of these solutions was measured with the Sweat-Chek again. The percentage errors were calculated as (measured concentration—set concentration)/set concentration × 100 and are plotted in Figure 5. It is clear that the error in Sweat-Chek device was higher; 6.2% of an average with respect to 2.0% of an average for the sweating arm. We were unable to reach these margins with the commercial system that we calibrated, even though Sweat-Chek data sheet states the error should be within 2% for the range of 75–110 mM, w. Error margins for very low concentrations were much higher for both measurements due to the resolution limitations of both systems.
A recently developed sweat rate sensor [14,26] was tested on the arm in order to show the functionality of the dispensing capability of the system. The sensor was attached to the arm and tests were conducted. The smartphone shows the results of the sensor device (Figure 6). The detailed calibration of the sweat rate sensor is provided in [14]. As the proof-of-concept, the sweat rate was set to 10 µL/min, which translates to 2 µL/min/cm2 provided the area of the Macroduct sweat collector shown in Figure 6 is 5 cm2. It is clear that the desired flow rate that is set by the arm model was achieved at the sweat rate sensor device. This approach gives easy access to testing systems and eliminates the need for human subjects during the device development phase.
The platform has two main aspects: liquid mixing and dispensing. Liquid mixing was performed in a closed-loop system, where the desired concentration was adjusted through the feedback of a commercial conductivity sensor real time. Therefore, this aspect of the testing platform was considered to be reliable as the system uses a commercial product. On the other hand, the aspect of flow rate needs to be verified by taking the environmental conditions into account, since several manipulations to the system were made besides the commercial peristaltic pump, such as the arm cast, sweat holes on the arm, and additional tubing that are required for liquid deliver from/to the peristaltic pump. Therefore, statistical analysis using analysis of variance (ANOVA) was performed on the testing approach to determine how relative humidity (RH), temperature, and concentration (CN) would affect the flow rate (FR). Each parameter was changed and flow rate values were tested in order to test these interactions. Therefore, the following assumptions were made for statistical analysis purposes: First, it was assumed that variation in relative humidity (RH), temperature, and concentration (CN) will not produce diverse results with respect to flow rate (FR). Apart from these factors, various other factors might influence the flow rate. However, it is assumed that all of the other factors do not have a major effect on the flow rate, and so they have been blocked to suppress their respective effects. This includes the measurement precision suppressed by using the same measuring instrument for all the tests and human error that is a complex aspect to address, but was suppressed by making sure that only one person records all of the readings. Further, the experiment is randomized to prevent unspecified disturbances that are associated with time, location, or order of observation from being confounded with effects. This allowed for the assumption that errors are independently distributed random variables.
A hypothesis was made to determine the effect of RH and CN on the flow rate, if any, once the factors of interest were found. It was assumed that all of the RH and saline concentration results in the same flow rate in order to simplify the effects of RH and CN. With this assumption, a few hypotheses were formulated, including:
  • Overall
    Null Hypothesis (H0): The flow rate for different RH on different CN will not vary.
    Alternate Hypothesis (HA): Some RHs will result in higher FR than the other.
  • RH Effect
    Null Hypothesis (H0): The RH type does not have an effect on the FR. All of the treatments effects, in this case, are equal to zero. i.e., τ1 = τ2 = τ3 = 0
    Alternate Hypothesis (HA): At least one of RHs has a significant effect on the flow rate. i.e., at least one of τ ≠ 0.
  • Concentration Effect
    Null Hypothesis (H0): The CN does not have an effect on the FR. All the treatments effects, in this case, are equal to zero. i.e., β1 = β2 = β3 = 0
    Alternate Hypothesis (HA): At least one of CNs has a significant effect on the flow rate. i.e., at least one of β ≠ 0.
  • Temperature Effect
    Null Hypothesis (H0): The temperatures does not have an effect on the FR. All the treatments effects, in this case, are equal to zero. i.e., α1 = α2 = α3 = 0
    Alternate Hypothesis (HA): At least one of the temperatures has an effect. i.e., at least one of α ≠ 0.
  • Interaction Effects
    Null Hypothesis (H0): None of the interactions have any effect on the flow rate. i.e., (τβ)ij = (βα)jk = (τα)ik = (τβα)ijk = 0
    Alternate Hypothesis (HA): At least one interaction has a significant effect on the flow rate.
After all of the levels in factors of the study were found and the data collection strategy was formulated, the test was implemented to obtain the data using a random order. Temperature and RH tests were carried out in a walk-in, closed-loop environmental chamber. While the majority of the factors that might affect the flow rate were considered and precautions were taken, each test was replicated three times in order to account for any measurement error. The data analysis started with the assumption that the interaction of RH and CN does not have a significant effect on the FR. Additionally, it was assumed that different temperatures might have a significant effect on the FR. Accordingly, the temperature was held as constant and a Graeco Latin Square (GLS) design test was performed to observe whether the RH has a major effect on the FR. The saline concentration type was blocked and the RH used is randomized to find the effect of RH on FR. With the temperatures being included, the study had three factors to test for their significance. This can be found while using the GLS design. However, GLS helps in finding the significance of the main effects; it assumes that there is no interaction among the main effects. There is always a possibility that the interaction of the main factors might have a significant effect on the system output. A study was performed to identify the various methods by which the significance of interactions can be found to find this interaction effect. The Latin Square and Graeco Latin Square designs are ruled out, as they do not account for the interaction effects. The two-factorial design that was again ruled out as the study had more than two levels in each factor to test. Finally, the test method was determined to be the general factorial design, as it was possible to test for more than two levels in each factor along with accounting for the main factor interactions.
The flow rate data used to perform the general factorial test is shown in Table 1 for three different RH (30%, 40%, 50%), CN (40 mM, 50 mM, 60 mM), and temperatures (20 °C, 27.5 °C, 35 °C). The values that were chosen were the most common values in human sweating and environmental conditions. A combination of factorials and randomization methods were used to collect a total of 324 (12 per combination of RH, CN, and T) test observations. The potential error from data collection methodologies and human error are minimized by converging the 12 observations to three averages that are presented in each cell of Table 1. General factorial design test is performed on the above data and Table 2 depicts the ANOVA results. These results clearly demonstrate that all of the main factors of relative humidity, saline concentration, temperature, and all of their interactions do not have any significant effect on the system flow rate, validating the statistical reliability of the proposed testing system.
It is important to note that the flow rate in actual humans (i.e., sweat rate) is significantly correlated to environmental conditions and the physical condition of the body [11,14]. The aim of the statistical analysis in this work was to show that the environmental conditions of the test setup do not affect the accuracy of artificially generated flow rates, which is important for the consistent testing of a sweat sensor before the sensor is tested on humans.

5. Conclusions

In this article, we proposed a sweating arm prototype that can be used for sweat sensor development projects, which will make the development phase more efficient without the need for human trials. Several microcontroller controlled relays allow for different mixing ratios of salt solutions (similar to sweat concentrations). The entire system was contained and placed on a cart for portability. Sweat rates down to 1 µL/min were able to be generated for realistic human sweat rate trials. Concentrations of 10 mM up to 200 mM of salt concentrations were repeatedly produced for real-life sweat concentration values. The ANOVA tests showed that the concentration does not have any effect on the flow rate for the prototype. The relative humidity and temperature values were also varied and the results did not show a statistical difference. When the interactions among all factors were considered, the system performance was proven to be insensitive to different factors, validating the system reliability.
We believe that this approach will allow the sweat sensing research community to rapidly test their initial designs without any particular expensive equipment and a further advance the real-time sweat sensing devices that can ultimately be used for physiological activity monitoring of performance athletes. Even though the system is currently designed for steady state sweat concentration and sweat rate values, it can be modified to provide instantaneous sweat condition changes, which can then be characterized by the real-time sweat rate and concentration prototypes, such as [14]. In its current form, the testing platform cannot address the physical wearability challenges that would arise from limb mobility and various other complications that could arise from human trials, such as skin contact, body hair, skin oils and minerals, dirt, etc. However, it is known that a significant amount of time and testing are needed prior to human trials and standardized testing equipment would significantly help to focus on the design development and optimization by being able to control the outside parameters. The arm mold could be easily cast in different forms to replicate some of the human conditions and could be perhaps made a mobile by adding several degrees of freedom to its mechanics, which are outside the scope of this work.

Author Contributions

Formal analysis, A.B., K.B., T.W., W.H., Z.M., S.P., B.R., K.Y. and T.K.; Funding acquisition, K.Y.; Investigation, A.B., K.B., T.W., W.H., Z.M., S.P., B.R. and T.K.; Methodology, A.B., K.B., T.W., W.H., Z.M., S.P., B.R., K.Y. and T.K.; Project administration, T.K.; Supervision, T.K.; Validation, K.Y. and T.K.; Writing—original draft, K.Y. and T.K.

Funding

This work was funded in part by the National Science Foundation under Grant #1542368 and Central Michigan University’s School of Engineering and Technology.

Conflicts of Interest

Authors declare no conflict of interest.

References

  1. Licht, T.; Stern, M.; Shwachman, H. Measurement of the electrical conductivity of sweat; its application to the study of cystic fibrosis of the pancreas. Clin. Chem. 1957, 3, 37–48. [Google Scholar]
  2. Kuno, Y. Physiology of Human Perspiration; J. & A. Churchill LTI: London, UK, 1934. [Google Scholar]
  3. Robinson, S.; Robinson, A. Chemical composition of sweat. Physiol. Rev. 1954, 34, 202–220. [Google Scholar] [CrossRef]
  4. Coyle, S.; Lau, K.; Moyna, N.; O’Gorman, D.; Diamond, D.; Francesco, F.; Costanzo, D.; Salvo, P.; Trivella, M.; De Rossi, D.; et al. BIOTEX-biosensing textiles for personalized healthcare management. IEEE Trans. Inf. Technol. Biomed. 2010, 14, 364–370. [Google Scholar] [CrossRef] [PubMed]
  5. Caldara, M.; Colleoni, C.; Guido, E.; Re, V.; Rosace, G. Optical monitoring of sweat pH by a textile fabric wearable sensor based on covalently bonded litmus-3-glycidoxypropyltrimethoxysilane coating. Sens. Actuators B Chem. 2016, 222, 213–220. [Google Scholar] [CrossRef]
  6. Schazmann, B.; Morris, D.; Slater, C.; Beirne, S.; Fay, C.; Reuveny, R.; Moyna, N.; Diamond, D. A wearable electrochemical sensor for the real-time measurement of sweat sodium concentration. Anal. Methods 2010, 2, 342–348. [Google Scholar] [CrossRef] [Green Version]
  7. Alomari, M.; Sel, K.; Mueller, A.; Edwards, J.; Kaya, T. Detection of relative [Na+] and [K+] levels in sweat with optical measurements. J. Appl. Phys. 2014, 115, 203107. [Google Scholar] [CrossRef]
  8. Sempionatto, J.R.; Martin, A.; García-Carmona, L.; Barfidokht, A.; Kurniawan, J.F.; Moreto, J.R.; Tang, G.; Shin, A.; Liu, X.; Escarpa, A.; et al. Skin-Worn Soft Microfluidic Potentiometric Detection System. Electroanalysis 2019, 31, 239–245. [Google Scholar] [CrossRef]
  9. Jia, W.; Bandodkar, A.; Valdes-Ramirez, G.; Windmiller, J.; Yang, Z.; Ramirez, J.; Chan, G.; Wang, J. Electrochemical tattoo biosensors for real-time noninvasive lactate monitoring in human perspiration. Analy. Chem. 2013, 85, 6553–6560. [Google Scholar] [CrossRef]
  10. Onor, M.; Gufoni, S.; Lomonaco, T.; Ghimenti, S.; Salvo, P.; Sorrentino, F.; Bramanti, E. Potentiometric sensor for non invasive lactate determination in human sweat. Anal. Chim. Acta 2017, 989, 80–87. [Google Scholar] [CrossRef]
  11. Liu, G.; Ho, C.; Slappey, N.; Zhou, Z.; Snelgrove, S.; Brown, M.; Grabinski, A.; Guo, X.; Chen, Y.; Miller, K.; et al. A wearable conductivity sensor for wireless real-time sweat monitoring. Sens. Actuators B Chem. 2016, 227, 35–42. [Google Scholar] [CrossRef]
  12. Kim, J.; Sempionatto, J.R.; Imani, S.; Hartel, M.C.; Barfidokht, A.; Tang, G.; Campbell, A.S.; Mercier, P.P.; Wang, J. Simultaneous Monitoring of Sweat and Interstitial Fluid Using a Single Wearable Biosensor Platform. Adv. Sci. 2018, 5, 1800880. [Google Scholar] [CrossRef] [Green Version]
  13. Wei, P.; Morey, B.; Dyson, T.; McMahon, N.; Hsu, Y.; Gazman, S.; Klinker, L.; Ives, B.; Dowling, K.; Rafferty, C. A conformal sensor for wireless sweat level monitoring. In Proceedings of the 2013 IEEE Sensors, Baltimore, MD, USA, 3–6 November 2013. [Google Scholar]
  14. Brueck, A.; Iftekhar, T.; Stannard, A.; Yelamarthi, K.; Kaya, T. A Real-Time Wireless Sweat Rate Measurement System for Physical Activity Monitoring. Sensors 2018, 18, 533. [Google Scholar] [CrossRef] [PubMed]
  15. Gao, W.; Emaminejad, S.; Nyein, H.; Challa, S.; Chen, K.; Peck, A.; Fahad, H.; Ota, H.; Shiraki, H.; Kiriya, D.; et al. Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis. Nature 2016, 529, 509–514. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  16. Matzeu, G.; O’Quigley, C.; McNamara, E.; Zuliani, C.; Fay, C.; Glennon, T.; Diamond, D. An integrated sensing and wireless communications platform for sensing sodium in sweat. Anal. Methods 2016, 8, 64–71. [Google Scholar] [CrossRef]
  17. Bariya, M.; Nyein, H.Y.Y.; Javey, A. Wearable Sweat Sensors. Nat. Electron. 2018, 1, 160–171. [Google Scholar] [CrossRef]
  18. Vellar, O. Studies on sweat losses of nutrients I. Iron content of whole body sweat and its association with other sweat constituents, serum iron levels, hematological indices, body surface area, and sweat rate. Scand. J. Clin. Lab. Investig. 1967, 21, 157–167. [Google Scholar] [CrossRef]
  19. Pilardeau, P.; Harichaux, P.; Chalumeau, M.; Vasseur, B.; Vaysse, J.; Garnier, M. Sweat collection from athletes. Br. J. Sports Med. 1985, 19, 197–198. [Google Scholar]
  20. Ely, M.; Ely, B.; Chinevere, T.; Lacher, C.; Lukaski, H.; Cheuvront, S. Evaluation of the megaduct sweat collector for mineral analysis. Physiol. Meas. 2012, 33, 385–394. [Google Scholar] [CrossRef] [PubMed]
  21. Cole, D.; Boucher, M. Use of a new sample-collection device (Macroduct) in anion analysis of human sweat. Clin. Chem. 1986, 32, 1375–1378. [Google Scholar]
  22. Dam, V.; Zevenbergen, M.; van Schaijk, R. Toward wearable patch for sweat analysis. Sens. Actuators B Chem. 2016, 236, 834–838. [Google Scholar] [CrossRef]
  23. Bandodkar, A.; Molinnus, D.; Mirza, O.; Guinovart, T.; Windmiller, J.; Valdes-Ramirez, G.; Andrade, F.; Schoning, M.; Wang, J. Epidermal tattoo potentiometric sodium sensors with wireless signal transduction for continuous non-invasive sweat monitoring. Biosens. Bioelectron. 2014, 54, 603–609. [Google Scholar] [CrossRef]
  24. Rose, D.; Ratterman, M.; Griffin, D.; Hou, L.; Kelley-Loughnane, N.; Naik, R.; Hagen, J.; Papautsky, I.; Heikenfeld, J. Adhesive RFID sensor patch for monitoring of sweat electrolytes. IEEE Trans. Biomed. Eng. 2015, 62, 1457–1465. [Google Scholar] [CrossRef] [PubMed]
  25. Jia, J.; Xu, C.; Pan, S.; Xia, S.; Wei, P.; Noh, H.; Zhang, P.; Jiang, X. Conductive Thread-Based Textile Sensor for Continuous Perspiration Level Monitoring. Sensors 2018, 18, 3775. [Google Scholar] [CrossRef]
  26. Sato, K.; Kang, W.; Saga, K.; Sato, K. Biology of sweat glands and their disorders. II. Disorders of sweat gland function. J. Am. Acad. Dermatol. 1989, 20, 713–726. [Google Scholar] [CrossRef]
  27. Iftekhar, A.; Ho, J.C.T.; Mellinger, A.; Kaya, T. 3D modeling and characterization of a calorimetric flow rate sensor for sweat rate sensing applications. J. Appl. Phys. 2017, 121, 094505. [Google Scholar] [CrossRef]
  28. Hou, L.; Hagen, J.; Wang, X.; Papautsky, I.; Naik, R.; Kelley-Loughnane, N.; Heikenfeld, J. Artificial microfluidic skin for in vitro perspiration simulation and testing. Lab Chip 2013, 13, 1868–1875. [Google Scholar] [CrossRef] [PubMed]
  29. Zhu, Z.; Yang, G.; Li, R.; Pan, T. Photopatternable PEDOT: PSS/PEG hybrid thin film with moisture stability and sensitivity. Microsyst. Nanoeng. 2017, 3, 17004. [Google Scholar] [CrossRef]
  30. Kaya, T.; Liu, G.; Ho, J.; Yelamarthi, K.; Miller, K.; Edwards, J.; Stannard, A. Wearable Sweat Sensors: Background and Current Trends. Electroanalysis 2019, 31, 411–421. [Google Scholar] [CrossRef]
  31. Taylor, N.A.; Machado-Moreira, C.A. Regional variations in transepidermal water loss, eccrine sweat gland density, sweat secretion rates and electrolyte composition in resting and exercising humans. Extreme Physiol. Med. 2013, 2, 4. [Google Scholar] [CrossRef]
Figure 1. Sketch and schematic of the overall system (a) Diagram of the system. Tanks are connected together with several control valves. Desired solution concentration is generated at the mix tank and the proper rate of liquid flow is adjusted at the peristaltic pump. (b) A circuit schematic of the electronics. Relays are controlled via Raspberry Pi 3 and a relay module. The conductivity circuit and data converter are also connected to the microcontroller for calibration.
Figure 1. Sketch and schematic of the overall system (a) Diagram of the system. Tanks are connected together with several control valves. Desired solution concentration is generated at the mix tank and the proper rate of liquid flow is adjusted at the peristaltic pump. (b) A circuit schematic of the electronics. Relays are controlled via Raspberry Pi 3 and a relay module. The conductivity circuit and data converter are also connected to the microcontroller for calibration.
Electronics 08 00606 g001
Figure 2. Snapshots of the sweating arm prototype. (a) The arm itself was positioned on CMU three-dimensional (3d) printed letters (representing Central Michigan University). (b) Zoomed view of the tanks with sprinkler systems exposed for a clear view. Gray components that are attached to orange covers of the tanks are the 3d printed sprinkler heads for rinsing the salt and mix tanks. Stirring pumps in the tanks and flush valves under the tanks are visible. (c) Electronic components were placed on a metal board that slides into the controller box for easy access and debugging.
Figure 2. Snapshots of the sweating arm prototype. (a) The arm itself was positioned on CMU three-dimensional (3d) printed letters (representing Central Michigan University). (b) Zoomed view of the tanks with sprinkler systems exposed for a clear view. Gray components that are attached to orange covers of the tanks are the 3d printed sprinkler heads for rinsing the salt and mix tanks. Stirring pumps in the tanks and flush valves under the tanks are visible. (c) Electronic components were placed on a metal board that slides into the controller box for easy access and debugging.
Electronics 08 00606 g002
Figure 3. The user interface to control the sweating arm prototype functions. Automatic trials can be initiated by the top left side of the code by entering desired values and pressing start. Rest of the interface provides manual controls for individual components for characterization purposes of the system.
Figure 3. The user interface to control the sweating arm prototype functions. Automatic trials can be initiated by the top left side of the code by entering desired values and pressing start. Rest of the interface provides manual controls for individual components for characterization purposes of the system.
Electronics 08 00606 g003
Figure 4. The flow rate is controlled with the rpm of the peristaltic pump, which results in a quadratic correlation for slow flow rates. Subset shows the wider range for the peristaltic pump’s semi-linear function overall. Higher flow rates are not realistic with respect to actual human sweat ranges.
Figure 4. The flow rate is controlled with the rpm of the peristaltic pump, which results in a quadratic correlation for slow flow rates. Subset shows the wider range for the peristaltic pump’s semi-linear function overall. Higher flow rates are not realistic with respect to actual human sweat ranges.
Electronics 08 00606 g004
Figure 5. Error percentages of the proposed system (Raspberry Pi) as compared to the commercial product (Sweat Check). Very low concentrations result in an error of around 10% but the error is within 2% for the rest of the concentration range. This error margin could further be reduced by lowering the feedback thresholds resulting in slower response time of the system to reach the desired concentration value.
Figure 5. Error percentages of the proposed system (Raspberry Pi) as compared to the commercial product (Sweat Check). Very low concentrations result in an error of around 10% but the error is within 2% for the rest of the concentration range. This error margin could further be reduced by lowering the feedback thresholds resulting in slower response time of the system to reach the desired concentration value.
Electronics 08 00606 g005
Figure 6. The sweating arm was tested using a recently developed sweat rate system [14]. The phone was positioned on the arm for visual purposes. A wireless sweat rate testing watch was attached to the arm and sweat rate was increased to reach the threshold of the phone application.
Figure 6. The sweating arm was tested using a recently developed sweat rate system [14]. The phone was positioned on the arm for visual purposes. A wireless sweat rate testing watch was attached to the arm and sweat rate was increased to reach the threshold of the phone application.
Electronics 08 00606 g006
Table 1. Data used in the general factorial design test.
Table 1. Data used in the general factorial design test.
Temp20 °C Temp27.5 °C Temp35 °C
RH (%)Concentration (mM)
405060405060405060
H300.2110.18850.18550.1980.19250.1970.1990.1940.198
0.210.2040.20250.1930.019950.2020.20350.1990.1995
0.210.20550.20450.1950.2020.2010.20650.2040.1995
H400.19550.19250.19650.2040.1960.18350.1980.19850.19
0.19950.19850.2050.2080.1960.19750.20750.2010.1995
0.19950.2130.20650.20450.1990.15450.2040.20250.1565
H500.1960.19450.1920.18950.1960.1810.2040.2060.2125
0.2090.19650.20350.20150.1880.1980.20550.2090.1985
0.2050.20650.1950.20550.21950.230.19650.2090.217
Table 2. ANOVA table for the general factorial design.
Table 2. ANOVA table for the general factorial design.
SourceSSdfMSF0p-Value
SS-RH0.001102720.000551361.134578450.329101981
SS-Conc0.000995920.000497951.024671820.36578608
SS-Temp0.001798420.000899191.850334880.167012298
SS-BF0.003089340.000772331.589293510.190419753
SS-BH0.002016340.000504091.037298610.396560851
SS-FH0.001462640.000365660.75244030.560778912
SS-BFH0.004051480.000506421.042112010.416851276
SS-Error0.0262418540.00048596
SS-T0.040758580

Share and Cite

MDPI and ACS Style

Brueck, A.; Bates, K.; Wood, T.; House, W.; Martinez, Z.; Peters, S.; Root, B.; Yelamarthi, K.; Kaya, T. An Artificial Sweating System for Sweat Sensor Testing Applications. Electronics 2019, 8, 606. https://doi.org/10.3390/electronics8060606

AMA Style

Brueck A, Bates K, Wood T, House W, Martinez Z, Peters S, Root B, Yelamarthi K, Kaya T. An Artificial Sweating System for Sweat Sensor Testing Applications. Electronics. 2019; 8(6):606. https://doi.org/10.3390/electronics8060606

Chicago/Turabian Style

Brueck, Andrew, Kyle Bates, Trent Wood, William House, Zackary Martinez, Shannon Peters, Blain Root, Kumar Yelamarthi, and Tolga Kaya. 2019. "An Artificial Sweating System for Sweat Sensor Testing Applications" Electronics 8, no. 6: 606. https://doi.org/10.3390/electronics8060606

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop