Article
Peer-Review Record

Evaluation of Load on Cervical Disc Prosthesis by Imposing Complex Motion: Multiplanar Motion and Combined Rotational–Translational Motion

Bioengineering 2024, 11(8), 857; https://doi.org/10.3390/bioengineering11080857 (registering DOI)
by Hossein Ansaripour 1,*, Stephen J. Ferguson 1 and Markus Flohr 2
Reviewer 1: Anonymous
Reviewer 2: Anonymous
Bioengineering 2024, 11(8), 857; https://doi.org/10.3390/bioengineering11080857 (registering DOI)
Submission received: 26 July 2024 / Revised: 16 August 2024 / Accepted: 19 August 2024 / Published: 22 August 2024
(This article belongs to the Special Issue Spine Biomechanics)

Round 1

Reviewer 1 Report

Comments and Suggestions for Authors

“Five generic zirconia-toughened alumina (BIOLOX® delta, CeramTec, Germany) ball and socket samples (shape: cylindrical body, height: 20 mm, and diameter: 11 mm) were tested” Is this a relevant implant being used currently?

 

 

Figure 2, remove the wording “Command” from the legend.

 

Section 3.1, add a table to show the values, instead of two paragraphs being used.

 

Figure 8. Replace the images a) and b) by 3 projections of the curve in the 3 planes. It’s not possible to know the evolution of the curves only by the 3d curves shown.

 

Are the interfaces being continuously lubricated?

Author Response

Please see the attachment.

Author Response File: Author Response.pdf

Reviewer 2 Report

Comments and Suggestions for Authors

Line 71-75 The research objective is too general. Give specific research questions, what you want to measure (force, rotation over time, etc.).

Line 241-243 "The first objective of this work was to compare the loads and moments in multiplanar motion tests ..." I did not notice such an objective in the introduction.

Line 266-267 "This study also investigated subluxation , which is one of the most important clinical concerns following disc replacement." Include this, for example, in the research question in the introduction.

Author Response

"Please see the attachment."

Author Response File: Author Response.pdf