Next Article in Journal
Chimeric Oncolytic Adenovirus Armed Chemokine Rantes for Treatment of Breast Cancer
Previous Article in Journal
Matrix Metalloproteases from Adipose Tissue-Derived Stromal Cells Are Spatiotemporally Regulated by Hydrogel Mechanics in a 3D Microenvironment
 
 
Correction published on 19 August 2024, see Bioengineering 2024, 11(8), 844.
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

The Effect of an Energy Window with an Ellipsoid Phantom on the Differential Defect Contrast on Myocardial SPECT Images

by
Ammar A. Oglat
1,
Mohannad Adel Sayah
2,* and
Norlaili A. Kabir
3
1
Department of Medical Imaging, Faculty of Applied Medical Sciences, The Hashemite University, Zarqa 13133, Jordan
2
Department of Radiography, Princess Aisha Bint Al-Hussein College of Nursing & Health Sciences, Al-Hussein Bin Talal University, P.O. Box 20, Ma’an 71111, Jordan
3
School of Physics, Universiti Sains Malaysia, Penang 11800, Malaysia
*
Author to whom correspondence should be addressed.
Bioengineering 2022, 9(8), 341; https://doi.org/10.3390/bioengineering9080341
Submission received: 6 June 2022 / Revised: 16 July 2022 / Accepted: 22 July 2022 / Published: 26 July 2022 / Corrected: 19 August 2024
(This article belongs to the Section Nanobiotechnology and Biofabrication)

Abstract

:
Good quality single-photon emission computed tomography (SPECT) images are required to achieve a perfect diagnosis and determine the severity of defects within the myocardial wall. There are many techniques that can support the diagnosis of defect formations in acquired images and contribute to avoiding errors before image construction. The main aim of this study was to determine the effect of energy width (15%, 20%, and 25%) on defect contrast in myocardial SPECT images correlated with the decentralization of positioning of a phantom. A phantom of polyethylene plastic was used to mimic the myocardial wall of the left ventricle. The phantom consists of two chambers, inner and outer. Two rectangular pieces of plastic were placed in anterior and inferior locations in the mid-region of the myocardial phantom to simulate myocardial infarction (defects). The average defect contrast for all phantom positions using 15% to 20% energy was (1.2, 1.6) for the anterior region and (1.1, 2) for the inferior region, respectively. Additionally, the energy window width was >25% with a large displacement of the positioning off center, leading to loss of the defect contrast in myocardial SPECT images, particularly in the inferior region. The study showed decreasing defect contrast in both locations, anterior and inferior, with increasing energy window width correlated with eccentricity positioning of the phantom on an imaging table.

1. Introduction

SPECT has been widely used in cardiac nuclear medicine imaging for nearly four decades. This non-invasive modality technique allows the examination of patients with known or suspected coronary artery abnormalities [1,2,3]. Over time, the use of this technique for the acquisition of cardiac images has become well established. The American Society of Nuclear Cardiology (ASNC) has published several guidelines for cardiac SPECT imaging [4,5,6]. Nowadays, better imaging properties have been achieved by technetium-99m (Tc-99m)-labeled radiotracers and new gamma cameras without any increase in scanning time [7,8,9].
SPECT can be utilized to define a patient’s pathologic status when cardiac symptoms cannot be explained by structural findings [10,11]. This disconnect between radiologic and clinical findings is frequently seen. SPECT has been routinely used to detect myocardial infarction in myocardial perfusion imaging [12].
Gamma cameras are used in SPECT imaging to obtain images from different angles from a patient on an imaging table injected with isotopes [13,14]. For SPECT images to be of perfect quality, the camera heads must avoid non-uniformity of isotope distribution within organs [15]. Variations in energy and uniformity of tomography reconstruction require more calibration for the center of rotation [16,17,18]. In the acquired SPECT image, limited energy causes scattering, and the scattering must be removed before correction [19,20]. Center of rotation error is one of the common reasons for artifacts that cause the non-uniformity of SPECT images [21]. The patient’s position on the imaging table should align the infected organ below the rotating gamma camera heads [22]. Many studies [23,24,25] in nuclear SPECT imaging focused on proper patient body positioning on an imaging table with gamma camera rotation.
SPECT with Tc-99m uses parameters to obtain optimized imaging. Performance can be enhanced for visualizing perfusion distribution through a careful selection of parameters such as energy windows and positioning on an imaging table. This study seeks to test the different energy windows of Tc-99m on the defect contrast of myocardial SPECT images correlated with the positioning on an imaging table [23,24].

2. Methods

2.1. Phantom Design and Configuration

Fabricated ellipsoid phantoms of polyethylene plastic (Zeff = 7.5) were used for this study. The Zeff of polyethylene can be calculated using Equation (1) as follows:
Z e f f = ( a 1   Z 1   3 + a 2   Z 2 3 + ) 1 / 3  
where
α is the fractional content of electrons of the ith element and
Z is the atomic number of the ith element.
On the basis of chemical composition (carbon and hydrogen) and Equation (1), the effective atomic number for polyethylene ( C 6 12 2   H 1 1 4) is calculated as follows:
Z e f f =   3   2 × 6 3 + 4 Z e f f = 7.5
Moreover, the number of electrons per gram ( N 0 ) on a compound polyethylene can be calculated using Equation (2) as follows:
N 0 N A · Z   A W    
N 0 = N A · Z ( C , H ) A W   ( C , H ) N 0 = 6.023 × 10 23 [ 2 × 12 + 4 × 1 ] [ 2 × 12 + 4 × 1 ] N 0 = 3.44 × 10 23   electrons / g .
where
N A is Avogadro’s number
A W   is the atomic weight and
Z is the atomic number.
Notably, the   N 0   for water is 3.343 × 10 23   electrons/g (Attix, 2008), whereas that for the cardiac muscle is 3.312 × 10 23   . Considering the effective atomic number (Zeff) and electron densities (N0) of polyethylene, water, and cardiac muscle, it is concluded that polyethylene is a suitable material to simulate LV myocardial wall tissue.
The atomic number and density of the phantom material corresponded to the atomic number and density of the myocardial wall muscle (number of electrons per gram = 0.55, density = 1 g/cm3). The fabricated phantom was meant to simulate tissue from the myocardial wall of the left ventricle. Measurements of the fabricated phantoms of two chambers (inner and outer) were designed to mimic the left ventricle in both stages (end diastole and end systole). Non-perfused defects with a 5-mm thickness were placed in the space between the two chambers [25,26]. These defects were fixed in the mid-region of the myocardial phantom at interior and inferior locations [Figure 1]. The inner chamber of the phantom was filled with water, and the space between the two chambers was also filled with a thoroughly mixed technetium-99m solution (0.43, 0.31) (mci) (1.147 MBq) for end diastole and end systole, respectively. Moreover, a myocardial phantom was placed at four positions on an imaging table at the center and off-center (5 cm, 10 cm, and 15 cm) to mimic the normal heart position of the patient’s body [Figure 2].

2.2. Image Acquisition

A dual-head gamma camera model (Discovery NM/CT670 Pro) with low energy and high resolution and a parallel holes’ collimator was used in this study. At Tc-99m (140 KeV), energy windows of 15%, 20%, and 25% were used. These values were selected because they are proper energy values for clinical left ventricular measurements. An angle of 180° circular orbit from the right anterior oblique to the left posterior oblique was used. The total acquisition time was 13 min using step-shot mode. The matrix size was 64 × 64. A Butterworth filter (order 10, 0.4 frequency) was used. All acquired SPECT images of the vertical axis view were saved in DICOM format. The acquisition of SPECT images using software Image J 1.48v was quantified to filter the images during reconstruction.
For each energy window, at four positions of the phantom, six acquired SPECT slices were obtained to calculate the mean pixels. The mean pixels in the center of the defect (the coldest spot) and 10 points adjacent to the defect (5 points on each side (left–right) of the defect) were calculated. The defect contrast calculations from the short axis SPECT profile were obtained and analyzed as myocardial wall defect “hottest spot to coldest spot”. The contrast was calculated using the relation: [M-D/M], where [M] is the mean pixels’ adjacent defect and [D] is the mean pixels at the center of the defect.
Theoretically, the variability of pixels’ distribution in SPECT images should be 0%. The distribution of pixel intensity in the reconstructed SPECT images was quantitatively analyzed as the percentage as follows:
Non uniformity (%) = [(maxvalue − minvalue)/(maxvalue + minvalue)] × 100
T-test, sigma plot was used for two group comparisons between energy window image acquisition for defect contrast, which correlated with different positioning on the imaging table. Probability values of <0.05 were considered statistically significant.
The mean percentage of measurements was calculated by averaging the percentage of three repeated SPECT acquisitions-test; sigma plot was used for two groups’ comparison for end diastole and end systole image acquisitions. Probability values of <0.05 were considered statistically significant.

3. Results and Discussion

Myocardial perfusion imaging (MPI) using SPECT has poorer contrast for sub-endocardial defects within the myocardial wall of the left ventricle. In this study, a 5 mm thickness equals 42% of the myocardial width (12 mm) to represent the sub-endocardial defect. Results showed differences in the defect contrast value in myocardial SPECT images when different widths of the energy window (15%, 20%, and 25%) were used. Furthermore, it was found that defect contrast in anterior and inferior locations of acquired SPECT images had different values in each width of the energy window, which was used with correlation to determine the effect of the positioning on the imaging (Table 1 and Table 2).
Figure 3, Figure 4 and Figure 5 represent the defect contrast using 3 energy windows at 4 positions of the phantom on an imaging table in a range of (0–15 cm) with an increment of 5 cm from the center. Visually, in the reconstruction of myocardial SPECT images, when the positioning of the myocardial phantom was at the center, the defect contrast was the greatest, whether this defect was in an anterior or inferior region within the myocardial wall phantom. Results show that the quality of the acquired images was the best when the MI phantom was positioned at the center of the imaging table, regardless of whether the defects were located in the inferior or anterior regions.
At the central position, results showed that the greatest defect contrast value of the 3 energy windows was (0.31) in the anterior location for a 15% energy window, while in the same window, it was (0.25) for the inferior region. Such a difference in defect contrast between anterior and inferior locations within myocardial SPECT images may be attributed to the loss of counts in the inferior region compared to the anterior [27]. When the phantom was placed off center, and the 3 energy windows were used (15%, 20%, and 25%), SPECT images showed a reduction in defect contrast in linear relation with the increase in the phantom displacement on the imaging table (R1, R2, and R3 = 0.99, anterior region) and (R1 = 0.98, R2 = 0.95, R3 = 0.97, inferior region), respectively [Figure 6 and Figure 7]. This reduction can be attributed to the decentralization of the positioning on the imaging table, which produces artifacts. This finding is in agreement with results that showed that the position of the heart within the orbit is very important and that an off-center event causes artifacts.
The reduction in defect contrast was proportional to the eccentricity of the positioning and became significant (p < 0.001, sigma plot) as energy windows were increased from 15% to 25% in all energy windows and positions of the myocardial phantom. The results showed a differential defect contrast between anterior and inferior locations within myocardial SPECT images, where the defect contrast was better in anterior than in inferior locations. Herein, the effects of window width (15%, 20%, and 25%) and the decentralization of the positioning on defect contrast were investigated. Moreover, the optimal defect contrast in the anterior and inferior regions of the myocardial SPECT image was improved by using energy window width with positioning. This optimal defect contrast may contribute to the improvement of SPECT protocol procedures before imaging.
If a physician could predict image distortion due to set-up error, he would be able to designate a distortion location due to blood flow insufficiency. Set-up errors cause insufficient data when the patient has been imaged sub-optimally. Therefore, another injection of a radiopharmaceutical is necessary to obtain a diagnosable image. In a clinical setting, involuntary motions (respiratory and heart motion) have a much greater impact on image quality than set-up error. In addition, SPECT image contrast is poor if the phantom is positioned off-center, even if energy-efficient windows are used.

4. Conclusions

The width of the wall of the MI phantom and the position on the imaging table are two important parameters in the quality assurance of myocardial SPECT imaging when the SPECT system is used to measure the thickness of MI. The results showed that the minimum detectable defect thickness in both stages (end diastole and end systole) varied in different regions in MI phantom and positions on the imaging table. Using an appropriate protocol for myocardial SPECT imaging may assess defect contrast to be at its best when an energy window of 15% is used. In conjunction with the positioning on an imaging table, the energy window width > 25% with a large displacement of phantom positioning off-center leads to loss of the defect contrast in myocardial SPECT images, particularly in an inferior region.

Author Contributions

Conceptualization, A.A.O. and M.A.S.; methodology, M.A.S. and N.A.K.; software, M.A.S.; validation, A.A.O.; formal analysis, A.A.O.; investigation, M.A.S. and N.A.K.; resources, M.A.S.; data curation, A.A.O.; writing—original draft preparation, M.A.S.; writing—review and editing, A.A.O.; visualization, A.A.O.; supervision, M.A.S.; project administration, M.A.S.; funding acquisition, M.A.S. and N.A.K. All authors have read and agreed to the published version of the manuscript.

Funding

The support of this work under financial Grant USM RUI GRANT 1001/PFIZIK/811345 by Universiti Sains Malaysia (USM) is gratefully acknowledged.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Acknowledgments

We would like to thank the researchers at Department of Medical Imaging, Faculty of Applied Medical Sciences, The Hashemite University, Zarqa, Jordan, and Higher Institute Sciences Medical, Elkhomes, Libya for their efforts to make this work readable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Ljungberg, M.; Pretorius, P.H. SPECT/CT: An update on technological developments and clinical applications. Br. J. Radiol. 2018, 91, 20160402. [Google Scholar] [CrossRef] [PubMed]
  2. Ritt, P. Recent developments in SPECT/CT. In Seminars in Nuclear Medicine; Elsevier: Amsterdam, The Netherlands, 2022. [Google Scholar]
  3. Bocher, M.; Blevis, I.M.; Tsukerman, L.; Shrem, Y.; Kovalski, G.; Volokh, L. A fast cardiac gamma camera with dynamic SPECT capabilities: Design, system validation and future potential. Eur. J. Nucl. Med. Mol. Imaging 2010, 37, 1887–1902. [Google Scholar] [CrossRef] [PubMed]
  4. Abbott, B.G.; Case, J.A.; Dorbala, S.; Einstein, A.J.; Galt, J.R.; Pagnanelli, R.; Bullock-Palmer, R.P.; Soman, P.; Wells, R.G. Contemporary cardiac SPECT imaging—Innovations and best practices: An information statement from the American society of nuclear cardiology. Circ. Cardiovasc. Imaging 2018, 11, e000020. [Google Scholar] [CrossRef] [PubMed]
  5. Hendel, R.C.; Berman, D.S.; Di Carli, M.F.; Heidenreich, P.A.; Henkin, R.E.; Pellikka, P.A.; Pohost, G.M.; Williams, K.A. ACCF/ASNC/ACR/AHA/ASE/SCCT/SCMR/SNM 2009 appropriate use criteria for cardiac radionuclide imaging: A report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the American Society of Nuclear Cardiology, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the Society of Cardiovascular Computed Tomography, the Society for Cardiovascular Magnetic Resonance, and the Society of Nuclear Medicine Endorsed by the American College of Emergency Physicians. J. Am. Coll. Cardiol. 2009, 53, 2201–2229. [Google Scholar] [PubMed]
  6. Brindis, R.G.; Douglas, P.S.; Hendel, R.C.; Peterson, E.D.; Wolk, M.J.; Allen, J.M.; Patel, M.R.; Raskin, I.E.; Hendel, R.C.; Bateman, T.M. ACCF/ASNC appropriateness criteria for single-photon emission computed tomography myocardial perfusion imaging (SPECT MPI) a report of the American College of Cardiology Foundation Quality Strategic Directions Committee Appropriateness Criteria Working Group and the American Society of Nuclear Cardiology endorsed by the American Heart Association. J. Am. Coll. Cardiol. 2005, 46, 1587–1605. [Google Scholar]
  7. Liu, Y.-H.; Lam, P.T.; Sinusas, A.J.; Wackers, F.J.T. Differential effect of 180 and 360 acquisition orbits on the accuracy of SPECT imaging: Quantitative evaluation in phantoms. J. Nucl. Med. 2002, 43, 1115–1124. [Google Scholar]
  8. Borges-Neto, S.; Pagnanelli, R.A.; Shaw, L.K.; Honeycutt, E.; Shwartz, S.C.; Adams, G.L.; Coleman, R.E. Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies. J. Nucl. Cardiol. 2007, 14, 555–565. [Google Scholar] [CrossRef]
  9. Khaw, B.A.; Strauss, H.W.; Carvalho, A.; Locke, E.; Gold, H.K.; Haber, E. Technetium-99m labeling of antibodies to cardiac myosin Fab and to human fibrinogen. J. Nucl. Med. 1982, 23, 1011–1019. [Google Scholar]
  10. Sciagrà, R.; Giaccardi, M.; Porciani, M.C.; Colella, A.; Michelucci, A.; Pieragnoli, P.; Gensini, G.; Pupi, A.; Padeletti, L. Myocardial perfusion imaging using gated SPECT in heart failure patients undergoing cardiac resynchronization therapy. J. Nucl. Med. 2004, 45, 164–168. [Google Scholar]
  11. Nagueh, S.F.; Bierig, S.M.; Budoff, M.J.; Desai, M.; Dilsizian, V.; Eidem, B.; Goldstein, S.A.; Hung, J.; Maron, M.S.; Ommen, S.R. American Society of Echocardiography clinical recommendations for multimodality cardiovascular imaging of patients with hypertrophic cardiomyopathy: Endorsed by the American Society of Nuclear Cardiology, Society for Cardiovascular Magnetic Resonance, and Society of Cardiovascular Computed Tomography. J. Am. Soc. Echocardiogr. 2011, 24, 473–498. [Google Scholar]
  12. Jaszczak, R.J. The early years of single photon emission computed tomography (SPECT): An anthology of selected reminiscences. Phys. Med. Biol. 2006, 51, R99. [Google Scholar] [CrossRef] [PubMed]
  13. Iskandrian, A.S.; Heo, J.; Askenase, A.; Segal, B.L.; Helfant, R.H. Thallium imaging with single photon emission computed tomography. Am. Heart J. 1987, 114, 852–865. [Google Scholar] [CrossRef]
  14. Buechel, R.R.; Herzog, B.A.; Husmann, L.; Burger, I.A.; Pazhenkottil, A.P.; Treyer, V.; Valenta, I.; von Schulthess, P.; Nkoulou, R.; Wyss, C.A. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur. J. Nucl. Med. Mol. Imaging 2010, 37, 773–778. [Google Scholar] [CrossRef] [PubMed]
  15. Baron, J.M.; Chouraqui, P. Myocardial single-photon emission computed tomographic quality assurance. J. Nucl. Cardiol. 1996, 3, 157–166. [Google Scholar] [CrossRef]
  16. Fahey, F.H.; A Harkness, B.; Keyes, J.W.; Madsen, M.T.; Battisti, C.; Zito, V. Society of Nuclear Medicine. Sensitivity, resolution and image quality with a multi-head SPECT camera. J. Nucl. Med. 1992, 33, 1859–1863. [Google Scholar]
  17. Lim, C.B.; Walker, R.; Pinkstaff, C.; Kim, K.I.; Anderson, J.; King, S.; Janzso, J.; Shand, D.; Valentino, F.; Coulman, K. Triangular SPECT system for 3-D total organ volume imaging: Performance results and dynamic imaging capability. IEEE Trans. Nucl. Sci. 1986, 33, 501–504. [Google Scholar] [CrossRef]
  18. Zwarts, F.; Dobbeleir, A.A.; te Kruibeke, B. Quantification and Data Optimisation of Heart and Brain Studies in Conventional Nuclear Medicine; University of Groningen: Groningen, The Netherlands, 2006. [Google Scholar]
  19. Ljungberg, M.; Strand, S.E. Scatter and attenuation correction in SPECT using density maps and Monte Carlo simulated scatter functions. J. Nucl. Med. 1990, 31, 1560–1567. [Google Scholar]
  20. Ogawa, K.; Harata, Y.; Ichihara, T.; Kubo, A.; Hashimoto, S. A practical method for position-dependent Compton-scatter correction in single photon emission CT. IEEE Trans. Med. Imaging 1991, 10, 408–412. [Google Scholar] [CrossRef]
  21. Young, K.; Kouris, K.; Awdeh, M.; Abdel-Dayem, H.M. Reproducibility and action levels for gamma camera uniformity. Nucl. Med. Commun. 1990, 11, 95–101. [Google Scholar] [CrossRef]
  22. Movahed, A.; Gnanasegaran, G.; Buscombe, J.; Hall, M.J.A.S.V. Integrating Cardiology for Nuclear Medicine Physicians; Springer: Berlin/Heidelberg, Germany, 2009. [Google Scholar]
  23. Knešaurek, K.; Machac, J.; Muzinic, M.; DaCosta, M.; Zhang, Z.; Heiba, S. Quantitative comparison of yttrium-90 (90 Y)-microspheres and technetium-99m (99m Tc)-macroaggregated albumin SPECT images for planning 90 Y therapy of liver cancer. Technol. Cancer Res. Treat. 2010, 9, 253–261. [Google Scholar] [CrossRef]
  24. Garcia, E.V.; Cooke, C.; Van Train, K.F.; Folks, R.; Peifer, J.; DePuey, E.; Maddahi, J.; Alazraki, N.; Galt, J.; Ezquerra, N.; et al. Technical aspects of myocardial SPECT imaging with technetium-99m sestamibi. Am. J. Cardiol. 1990, 66, E23–E31. [Google Scholar] [CrossRef]
  25. Vallejo, E. New radiotracers in cardiac imaging: Principles and applications taillefer and Tamaki. J. Nucl. Cardiol. 1999, 6, 549. [Google Scholar] [CrossRef]
  26. Ye, J.; Liu, L.; Braymer, W.; Buskard, T. Improvement in myocardial defect contrast and uniformity for SPECT in an anthropomorhphic phantom using attenuation correction and resolution recovery for Tc-99m and Tl-201. J. Nucl. Med. 1998, 39, 180. [Google Scholar]
  27. Lowe, V.J.; Greer, K.L.; Hanson, M.W.; Jaszczak, R.J.; Coleman, R.E. Cardiac phantom evaluation of simultaneously acquired dual-isotope rest thallium-201/stress technetium-99m SPECT images. Soc. Nucl. Med. 1993, 34, 1998–2006. [Google Scholar]
Figure 1. Schematic phantom to simulate myocardial wall of the left ventricle with two solid plastics.
Figure 1. Schematic phantom to simulate myocardial wall of the left ventricle with two solid plastics.
Bioengineering 09 00341 g001
Figure 2. Schematic of four myocardial phantom positions on an imaging table.
Figure 2. Schematic of four myocardial phantom positions on an imaging table.
Bioengineering 09 00341 g002
Figure 3. Defect in interior and inferior region of myocardial SPECT image using 15% energy window at 4 positions, where the rows (A, B, C, and D) at (the center, 5 cm, 10 cm, and 15 cm), respectively.
Figure 3. Defect in interior and inferior region of myocardial SPECT image using 15% energy window at 4 positions, where the rows (A, B, C, and D) at (the center, 5 cm, 10 cm, and 15 cm), respectively.
Bioengineering 09 00341 g003
Figure 4. Defect in interior and inferior region of myocardial SPECT image using 20% energy window at four positions, where the rows (A, B, C, and D) at (the center, 5 cm, 10 cm, and 15 cm), respectively.
Figure 4. Defect in interior and inferior region of myocardial SPECT image using 20% energy window at four positions, where the rows (A, B, C, and D) at (the center, 5 cm, 10 cm, and 15 cm), respectively.
Bioengineering 09 00341 g004
Figure 5. Defect in interior and inferior region of myocardial SPECT image using 25% energy window at four positions, where the rows (A, B, C, and D) at (the center, 5 cm, 10 cm, and 15 cm), respectively.
Figure 5. Defect in interior and inferior region of myocardial SPECT image using 25% energy window at four positions, where the rows (A, B, C, and D) at (the center, 5 cm, 10 cm, and 15 cm), respectively.
Bioengineering 09 00341 g005
Figure 6. Effect of the positioning on defect contrast in the anterior region of myocardial SPECT image in different energy windows.
Figure 6. Effect of the positioning on defect contrast in the anterior region of myocardial SPECT image in different energy windows.
Bioengineering 09 00341 g006
Figure 7. Effect of the positioning on defect contrast in the inferior region of myocardial SPECT image in different energy windows.
Figure 7. Effect of the positioning on defect contrast in the inferior region of myocardial SPECT image in different energy windows.
Bioengineering 09 00341 g007
Table 1. Values of defect contrast at the anterior location in myocardial SPECT image in different positions of the phantom for three energy windows.
Table 1. Values of defect contrast at the anterior location in myocardial SPECT image in different positions of the phantom for three energy windows.
Energy WindowABCD
15%0.310.260.180.12
20%0.260.210.140.10
25%0.190.150.110.08
Table 2. Values of defect contrast at an inferior location in SPECT image in different positions of the phantom for three energy windows.
Table 2. Values of defect contrast at an inferior location in SPECT image in different positions of the phantom for three energy windows.
Energy WindowABCD
15%0.250.200.120.10
20%0.230.190.080.07
25%0.140.100.050.05
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Oglat, A.A.; Sayah, M.A.; Kabir, N.A. The Effect of an Energy Window with an Ellipsoid Phantom on the Differential Defect Contrast on Myocardial SPECT Images. Bioengineering 2022, 9, 341. https://doi.org/10.3390/bioengineering9080341

AMA Style

Oglat AA, Sayah MA, Kabir NA. The Effect of an Energy Window with an Ellipsoid Phantom on the Differential Defect Contrast on Myocardial SPECT Images. Bioengineering. 2022; 9(8):341. https://doi.org/10.3390/bioengineering9080341

Chicago/Turabian Style

Oglat, Ammar A., Mohannad Adel Sayah, and Norlaili A. Kabir. 2022. "The Effect of an Energy Window with an Ellipsoid Phantom on the Differential Defect Contrast on Myocardial SPECT Images" Bioengineering 9, no. 8: 341. https://doi.org/10.3390/bioengineering9080341

APA Style

Oglat, A. A., Sayah, M. A., & Kabir, N. A. (2022). The Effect of an Energy Window with an Ellipsoid Phantom on the Differential Defect Contrast on Myocardial SPECT Images. Bioengineering, 9(8), 341. https://doi.org/10.3390/bioengineering9080341

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop