Fast Analysis of Glucose Content in Food Samples Using a Chitosan-GOx/TiO2NTAs/Ti Amperometric Biosensor †

by
Margalida Artigues-Cladera
,
Jordi Abella
and
Sergi Colominas
*
Electrochemical Methods Laboratory, Department of Analytical and Applied Chemistry, ETS Institut Químic de Sarrià, Universitat Ramon Llull, 08017 Barcelona, Spain
*
Author to whom correspondence should be addressed.
Presented at the 5th International Symposium on Sensor Science (I3S 2017), Barcelona, Spain, 27–29 September 2017.
Proceedings 2017, 1(8), 831; https://doi.org/10.3390/proceedings1080831
Published: 20 November 2017
Among all the analytes that are monitored for quality control in the food sector, glucose is assuming a priority role because it has been associated with chronic health consequences. Numerous methods have been reported for glucose analysis in food, however, most of the adopted methods are time consuming or expensive. The development of fast, cheap, practical and selective methods for detecting glucose in food is still a popular research area. In this context, enzymatic-based biosensors can be used to produce easy-to-use, compact and inexpensive analytical tools.
In the present work, an amperometric glucose biosensor based on glucose oxidase (GOx) immobilization onto highly ordered titanium dioxide nanotube arrays (TiO2NTAs) was used to determine the glucose content of commercial food samples (soft drinks and soy sauces). The fundamental analytical parameters of the proposed biosensor were evaluated in order to guarantee that the analytical methodology is accurate, specific, reproducible and robust. The obtained results proved sufficient repeatability (RSD = 1.9%), reproducibility (RSD = 2.5%), accuracy (recovery between 95% and 105%) and robustness (RSD = 3.3%). Finally, the glucose content of different samples was measured using the biosensor and compared with the HPLC value. In the worst case scenario, a deviation smaller than 7% was obtained from the 11 samples evaluated.

Share and Cite

MDPI and ACS Style

Artigues-Cladera, M.; Abella, J.; Colominas, S. Fast Analysis of Glucose Content in Food Samples Using a Chitosan-GOx/TiO2NTAs/Ti Amperometric Biosensor. Proceedings 2017, 1, 831. https://doi.org/10.3390/proceedings1080831

AMA Style

Artigues-Cladera M, Abella J, Colominas S. Fast Analysis of Glucose Content in Food Samples Using a Chitosan-GOx/TiO2NTAs/Ti Amperometric Biosensor. Proceedings. 2017; 1(8):831. https://doi.org/10.3390/proceedings1080831

Chicago/Turabian Style

Artigues-Cladera, Margalida, Jordi Abella, and Sergi Colominas. 2017. "Fast Analysis of Glucose Content in Food Samples Using a Chitosan-GOx/TiO2NTAs/Ti Amperometric Biosensor" Proceedings 1, no. 8: 831. https://doi.org/10.3390/proceedings1080831

Article Metrics

Back to TopTop