Article

The Accuracy of Digital Face Scans Obtained from 3D Scanners: An In Vitro Study

by
Pokpong Amornvit
and
Sasiwimol Sanohkan
*
Department of Prosthetic Dentistry, Faculty of Dentistry, Prince of Songkla University, Hat Yai, Songkhla 90110, Thailand
*
Author to whom correspondence should be addressed.
Int. J. Environ. Res. Public Health 2019, 16(24), 5061; https://doi.org/10.3390/ijerph16245061
Submission received: 16 November 2019 / Revised: 4 December 2019 / Accepted: 5 December 2019 / Published: 12 December 2019
(This article belongs to the Section Digital Health)

Abstract

:
Face scanners promise wide applications in medicine and dentistry, including facial recognition, capturing facial emotions, facial cosmetic planning and surgery, and maxillofacial rehabilitation. Higher accuracy improves the quality of the data recorded from the face scanner, which ultimately, will improve the outcome. Although there are various face scanners available on the market, there is no evidence of a suitable face scanner for practical applications. The aim of this in vitro study was to analyze the face scans obtained from four scanners; EinScan Pro (EP), EinScan Pro 2X Plus (EP+) (Shining 3D Tech. Co., Ltd. Hangzhou, China), iPhone X (IPX) (Apple Store, Cupertino, CA, USA), and Planmeca ProMax 3D Mid (PM) (Planmeca USA, Inc. IL, USA), and to compare scans obtained from various scanners with the control (measured from Vernier caliper). This should help to identify the appropriate scanner for face scanning. A master face model was created and printed from polylactic acid using the resolution of 200 microns on x, y, and z axes and designed in Rhinoceros 3D modeling software (Rhino, Robert McNeel and Associates for Windows, Washington DC, USA). The face models were 3D scanned with four scanners, five times, according to the manufacturer’s recommendations; EinScan Pro (Shining 3D Tech. Co., Ltd. Hangzhou, China), EinScan Pro 2X Plus (Shining 3D Tech. Co., Ltd. Hangzhou, China) using Shining Software, iPhone X (Apple Store, Cupertino, CA, USA) using Bellus3D Face Application (Bellus3D, version 1.6.2, Bellus3D, Inc. Campbell, CA, USA), and Planmeca ProMax 3D Mid (PM) (Planmeca USA, Inc. IL, USA). Scan data files were saved as stereolithography (STL) files for the measurements. From the STL files, digital face models are created in the computer using Rhinoceros 3D modeling software (Rhino, Robert McNeel and Associates for Windows, Washington DC, USA). Various measurements were measured five times from the reference points in three axes (x, y, and z) using a digital Vernier caliper (VC) (Mitutoyo 150 mm Digital Caliper, Mitutoyo Co., Kanagawa, Japan), and the mean was calculated, which was used as the control. Measurements were measured on the digital face models of EP, EP+, IPX, and PM using Rhinoceros 3D modeling software (Rhino, Robert McNeel and Associates for Windows, Washington DC, USA). The descriptive statistics were done from SPSS version 20 (IBM Company, Chicago, USA). One-way ANOVA with post hoc using Scheffe was done to analyze the differences between the control and the scans (EP, EP+, IPX, and PM). The significance level was set at p = 0.05. EP+ showed the highest accuracy. EP showed medium accuracy and some lesser accuracy (accurate until 10 mm of length), but IPX and PM showed the least accuracy. EP+ showed accuracy in measuring the 2 mm of depth (diameter 6 mm). All other scanners (EP, IPX, and PM) showed less accuracy in measuring depth. Finally, the accuracy of an optical scan is dependent on the technology used by each scanner. It is recommended to use EP+ for face scanning.

Graphical Abstract

1. Introduction

The human face not only shows the physical anatomical landmarks of a person’s identity, but reveals the psychological make-up or personalization [1]. Facial morphology and analysis are important for various disciplines, such as craniofacial-maxillofacial surgery, orthodontics, prosthodontics, pedodontics, biometrics, and forensic odontologists [2,3,4]. The conventional methods of facial analysis include two-dimensional (2D) photographic, Vernier caliper, and bevel protractor measurements, to measure 2D projection distances and angles [2,5,6]. Recently, there has been a digital dental era due to the massive advancement and evolution in optical scanning and designing technology that has led a shift from 2D to three-dimensional (3D) technology, with the use of 3D leading to disruption in the treatment modality [7,8,9]. The facial landmarks can be recorded digitally from the 3D face scanning using the scanner and can be used for facial recognition, capturing facial emotions, facial cosmetic planning and surgery, maxillofacial rehabilitation, etc. [10,11,12,13,14]. The computer aided design/computer aided manufacturing (CAD/CAM) system, milling systems, rapid prototyping, three-dimensional (3D) scanning, and 3D printing have revolutionized and created new modalities in medicine and dentistry, which improved the speed and accuracy of treatment [15,16,17]. Currently, CAD/CAM is widely used in dentistry for the fabrication of inlays, onlays, veneers, crowns, implant prosthesis, and full-mouth reconstruction [16,18,19]. In addition, in dentistry, 3D imaging using dental cone beam computed tomography (CBCT) offers volumetric data on jaw bones and teeth, which helps in presurgical diagnosis, preoperative planning, and transferring a preoperative plan for actual treatment outcome for oral rehabilitation [20,21,22]. Similarly, augmented reality is also an emerging technology in medicine and dentistry, and this includes an accurate display of either static or dynamic diagnostic images via the use of a visor or specific glasses, which is useful, especially in implant surgery [23].
For aesthetic and full mouth rehabilitation, although the semi-adjustable articulator and face bow are commonly used, they are based on the average values and cannot be individualized unless the fully adjustable articulator is used [24]. However, due to high technical skills requirements and errors from the users, the measurements may have questionable accuracy. Hence, the 3D face scan and intraoral scan can be integrated with a CBCT scan for the facial analysis, occlusion analysis, and even digital/virtual face bow transfer and full mouth rehabilitation [22,25,26,27,28,29,30,31,32]. The data from CBCT, intraoral scans, and facial scans can be superimposed to create a 3D “virtual patient” for better diagnosis, treatment planning, and patient outcomes [22]. Mangano et al. [32] found that the combination of intraoral and face scans allowed them to successfully restore fully edentulous patients with maxillary overdentures supported by four implants and a CAD/CAM PEEK bar. The facially driven design is a technique which considers the facial esthetics, facial profile, proportions, and harmony for the oral rehabilitation, and uses digital smile design to make aesthetically pleasing faces. This approach is important for aesthetic dental restorations considering hard and soft tissues, which results in enhanced smiling, self-esteem, and self-confidence of patients [1,33,34,35,36]. Higher accuracy improves the quality of the data recorded from the face scanner, which ultimately improve the outcome [37]. Although there are various face scanners available in the market, there is no evidence of a suitable scanner for practical applications.
Hence, the aim of this in vitro study was to analyze the face scans obtained from four scanners:—EinScan Pro (EP), EinScan Pro 2X Plus (EP+) (Shining 3D Tech. Co., Ltd. Hangzhou, China), iPhone X (IPX) (Apple Store, Cupertino, CA, USA), and Planmeca ProMax 3D Mid (PM) (Planmeca USA, Inc. IL, USA)—and to compare scans obtained from various scanners with the control (measured from Vernier caliper). This helped to identify the appropriate scanner for face scanning.

2. Materials and Methods

The overview of the study is shown in Figure 1 and the method can be divided into three parts: scanning, measurements, and comparison.

2.1. Scanning

At first, a face model was designed in Rhinoceros 3D modeling software (Rhino, Robert McNeel and Associates for Windows, Washington DC, USA) with shape, size, and ratios close to the human face. Reference points marked on the model and various measurements along the x axis (length), y axis (length), and z axis (depth) are shown in Figure 2. The face model was printed from polylactic acid using the resolution of 200 microns on x, y, and z axes (Figure 3).
Then, the master face model was 3D scanned with four scanners five times according to the manufacturer’s recommendations: EinScan Pro (Shining 3D Tech. Co., Ltd. Hangzhou, China), EinScan Pro 2X Plus (Shining 3D Tech. Co., Ltd. Hangzhou, China) using Shining Software, iPhone X (Apple Store, Cupertino, CA, USA) using the Bellus3D Face Application (Bellus3D, version 1.6.2, Bellus3D, Inc. Campbell, CA, USA), and Planmeca ProMax 3D Mid (PM) (Planmeca USA, Inc. IL, USA) (Figure 4). Scan data files were saved as stereolithography (STL) files for the measurements. From the STL files, digital face models were created in the computer using Rhinoceros 3D modeling software (Rhino, Robert McNeel and Associates for Windows, Washington DC, USA).

2.2. Measurement

Various measurements were measured five times from the reference points on three axes (x, y, and z) using a digital Vernier caliper (VC) (Mitutoyo 150 mm Digital Caliper, Mitutoyo Co., Kanagawa, Japan), and the mean was calculated, which is shown in Figure 2. The same measurements as measured before were measured on the digital face models of EP, EP+, IPX, and PM using Rhinoceros 3D modeling software (Rhino, Robert McNeel and Associates for Windows, Washington DC, USA).

2.3. Comparison

The measurements of the various scanners were compared with the Vernier caliper. The descriptive statistics were done from SPSS version 20 (IBM Company, Chicago, USA). One-way ANOVA with post hoc using Scheffe was done to analyze the difference between the control and the scans (EP, EP+, IPX, and PM). The significance level was set at p = 0.05.

3. Results

3.1. Scanning Time

Table 1 shows the scanning time, processing time, and total time for the scanning process of four scanners (EP, EP+, IPX, and PM). The scanning process consists of scanning and data processing. Scanning time is the time taken for scanning the face. Data processing includes data editing, generating points, meshing, removing artifacts, making a solid model, and finishing the face model. The fastest scan was done by IPX with Bellus3D (0.57 ± 0.03 min), and was followed by PM (0.7 ± 0.05 min), EP (6.77 ± 0.15 min), and finally, EP+ (9.4 ± 0.21 min).

3.2. Scanning Accuracy

The best scanning accuracy was shown by EP+, which was followed by EP, IPX, and finally, PM (Figure 3). EP showed that medium scanning accuracy and the least accuracy was shown by IPX and PM. Regarding the depth, EP+ showed accuracy of 2 mm in depth (diameter 6 mm). All other scanners (EP, IPX, and PM) showed less scanning accuracy when measuring depth.
Table 2 shows the mean measurements along the x-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of various scans from the VC.
For the EP, the X3, X4, and X5 measurements showed significant differences (p < 0.01) compared to VC. For the EP+, all the measurements on the z-axis showed no significant difference to the VC. For the IPX, the X4 measurement showed a significant difference (p < 0.01) compared to the VC. The X1 measurement could not be measured because scanning the point for the measurement of X1 could not be captured in the IPX scan. Hence, from the results in x-axis, it can be implied that the EP showed inaccuracy in capturing the length of more than 50 mm. However, EP+ showed accuracy in recording the length until 120 mm. IPX showed accuracy in capturing the length from 10 mm until 50 mm, but failed to record the details and had difficulties while measuring length (X1). Similarly, Table 3 shows the mean measurements in the y-axis of face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of various scans from the VC.
For the EP, it showed that the Y3, Y4, Y5, and Y6 measurements showed significant differences (p < 0.01) compared to the VC. For the EP+, all the measurements showed no significant difference to the VC. For the IPX, Y6 showed significant difference (p < 0.01) compared to the VC. For the PM, Y1, Y3, Y5, and Y6 showed significant difference (p < 0.01) compared to the VC. The X1, Y1, and Y2 measurements could not be measured because scanning the point for the measurement of Y1 could not be captured in the IPX scan. Similarly, from the results in y-axis, the EP showed inaccuracy in capturing the length more than 50 mm (similar to X-axis). In addition, EP+ showed accuracy in recording the length until 150 mm. IPX showed accuracy in capturing the length from 50 mm until 120 mm, but failed to record the details and had difficulties while measuring length (Y1 and Y2).
Finally, Table 4 shows the mean measurements in z-axis of face model from the caliper (VC) and various scans (EP, EP+, IPX, and PM) and the comparisons of scans from the VC.
For both the EP and EP+, it showed that all the measurements in z-axis showed significant differences (p < 0.01) compared to the VC. For IPX, all the measurements in z-axis could not be measured because scanning the point for the measurement of Z-axis could not be captured in the IPX scan. Hence, all the scanners were unable to record clearly, the depth of more than 2–4 mm with a diameter of 6 mm. The IPX showed the least accuracy in recording the depth of more than 2 mm.
Figure 5 shows the mean measurements along the x-axis (length), y-axis (length), and z-axis (depth) of VC, and three scanners. Discontinuity in a line shows the inability in measurements. Figure 6 shows the mean differences of various measurements in the x-axis (length), y-axis (length), and z-axis (depth) of three scanners from the caliper. It showed that the mean difference of the measurements increased as the distance increased (from 2 to 120 mm in the x-axis, from 2 to 150 mm in the y-axis, and from 4 to 10 mm in the z-axis). The factors affecting the accuracy in the z-axis, or depth, are light intensity, focus, distance of objects from the scanner, and deviation of light.

4. Discussion

Facial and dental measurements analyses help in smile design and esthetic rehabilitation [38,39,40,41,42]. With the huge advancements in digital technology, it has been widely applied in medicine and dentistry for diagnoses, therapeutics, artificial intelligence, and augmented reality [43]. The face scanning and digital impression-making of teeth are widely used now a days [44]. The advantages of scanners are simplified clinical procedure; time efficiency; patient comfort; no use of plaster/dental stone, as a digital model is prepared in the computer; better patient communication and motivation; simplifying clinical procedures for both the dentist and the laboratory technician; ease of communication with technician; and ease of fabrication of prosthesis [19,44,45,46,47].
Accuracy is key in all clinical applications in prosthesis and scanners should be able to detect an accurate impression [48,49,50]. In this study, a face model was fabricated using the fused deposition modeling technique and this study analyzed the details and accuracy of face scans obtained from four scanners (EP, EP+, IPX, and PM) and compared the scans obtained from various scanners with the control (measured from Vernier caliper). This helped to identify the appropriate scanner for face scanning. In this study, EP+ showed the highest accuracy (accuracy until 150 mm of length). EP showed medium scanning accuracy (accurate until 10 mm of length). IPX and PM were the least accurate (accuracy from 10 mm to 120 mm in length). For IPX, various measurements (X1, Y1, and Y2) could not be measured because they could not be captured in the scan, as IPX and PM showed. In addition, IPX showed accuracy in capturing the length from 10 mm until 50 mm in the x-axis and from 50 mm until 120 mm in the y-axis, with failure to record the details and difficulty while measuring. This might be due the ability of the video capturing capacity of the IPX. The inaccuracy of PM scanner may be due scanning from a short distance due to the lesser field of vision. Our study is similar to the study done by Zhao et al. [2], wherein they compared the practical accuracy of optical facial scanners for facial deformity patients using a high-accuracy industrial “line-laser” scanner (Faro), stereophotography (edMD), and a “structured light” facial scanner (FaceScan). The respective 3D accuracy of stereophotography and structured light facial scanners obtained for facial deformities were 0.58 ± 0.11 mm and 0.57 ± 0.07 mm. The 3D accuracies of different facial partitions were inconsistent; the measurements at the middle face had the showed the highest accuracy. Although the accuracies of two facial scanners were lower than their nominal accuracies, they all met the requirements for oral clinic use. However, from the results of our study, only EP+ is suitable for face scanning.
Mangano et al. [48] compared the accuracy of five different intraoral scanners (IOSs) in the impressions of single and multiple implants. They used plaster models representative of a partially edentulous maxilla to be restored with a single crown and a partial prosthesis, and a totally edentulous maxilla to be restored with a full-arch. They concluded that the IOSs showed significant differences between them, both in trueness and in precision. The mathematical error increased in the transition from single crown to partial prosthesis up to full-arch, both in trueness and in precision. A similar study was done by Braian et al. [51], wherein they studied the trueness and precision under repeatable conditions for different IOSs when scanning fully edentulous arches with multiple implants, and they found low precision when scanning fully edentulous arches with multiple implants. This result is similar to our study: as the scanning length increased, the accuracy decreased. They mentioned that the accuracy can be enhanced by reducing the span of scanning, and ensuring the scanned surfaces exhibit minimal irregularities [49]. Furthermore, Braian et al. [52] studied the accuracies of IOSs for scanning dentate and edentulous casts and they found that the significant differences were found in scanning edentulous and dentate scans for short arches and complete arches. Trueness for complete-arch scans was <193 μm (edentulous scans), and was <150 μm for dentate scans. Trueness for short-arch scans was <103 μm (edentulous scans) and <56 μm for dentate scans.
Similarly, for the depth measurement, all the optical scanners showed less accuracy; i.e., were unable to record clearly, a depth of more than 2 mm for diameter 6 mm. The IPX showed the least accuracy in recording the depth of more than 2 mm. Inaccuracy in recording the depth might be due to failure of passing the light into the depth while scanning. For the depth measurement in z-axis, none of the scanners were able to capture it accurately. In this study, a hole of 2 mm was used, which was too narrow to pass the light to the bottom of the hole. The scanners used in this study (EP, EP+, IPX, and PM) are optical scanners. The light from the scanner’s projects in various patterns to the surface and records the 3D picture. The software analyzes and creates the 3D model. The EP projects the visible line pattern, the EP+ projects a QR code-like pattern, the IP projects one QR code-like with infrared rays, and the PM projects X-rays. Light or optical scanning shows it is also acceptable but may have distortion from the light intermittent. IPX uses infrared for face scanning using the Bellus3D Face application to capture hard and soft tissue. In another study, it was found that there was no error in recording the tissue of depth of 5 mm for diameter 4.5 mm by a TRIOS scanner Model S1P (3Shape Trios A/S, Copenhagen, Denmark) for the fabrication of polyether ether ketone (PEEK) abutment for the implant retained finger prosthesis [53].
Van der Meer et al. [54] studied the accuracy of the IOSs by scanning high precision PEEK cylinders and they found the errors and it increased in distance and/or angulations in arch due to an accumulation of registration errors of the patched 3D surfaces. The registration errors may vary in magnitude depending on the scanning technology and the registration algorithms [55]. This all can be eliminated with advancement in the technology. In addition, the scanners may be difficult when scanning shiny, reflective, or transparent objects. The calibration is also done to compensate the errors that have occurred during scanning.
In addition, the fastest scan was done by IPX (0.57 min). PM: 0.7 min; EP, 6.7 min; and EP+, 9.4 min. The faster scanning of IPX and PM might be due to lower accuracy of the scanning and resolution of the scanned file. It was found that there was no significant difference between EP and EP+, which might be due to similar accuracy of the scanning and resolution of the scanned file. In this study, the descending order of the rendering of the 3D scan files was EP+, EP, IPX, PM. In the clinical situation, if the capture details are required, such as the teeth surfaces, auditory canal, or nostril, more accuracy of the scanner or technique of the scan is required. In addition, in a procedure such as digital face bow recording, accuracy is important for accurate recording of an occlusion to the face. IOS can be combined with a facial scanner for oral rehabilitation purposes. Mangano et al. [32] combined intraoral and face scanning for the CAD/CAM fabrication of implant-supported bars for maxillary overdentures. Hence, the scanners which were used in our study can be combined with the IOSs for the esthetic oral rehabiliation. The results from this study can be implemented in capturing landmarks, as shown in Table 5. The capturing difficulty of the facial structures can be classified as follows.
Many factors influence the accuracy of the 3D scanner, such as ability to record details, accuracy, scanning principles, span of scanning, size of scanning area, arch length, surface irregularities, temperature, relative humidity, and illumination [56,57]. The 3D scanning uses one of the various scanning technologies—laser triangulation, structured light, photogrammetry, contact-based, and laser pulse (time of flight or lidar) [58]. Contact-based is the best for the surface scanning but it depends on the probe size. The disadvantages of scanners include costs for the machines, difficulty in handing, technique-sensitive difficulties when capturing the deeper tissues, and rendering [45,58,59,60]. The future of digital scanning is expected to involve wide availability of scanners at lower costs with high quality and accuracy for various dental and medical applications.

5. Conclusions

The following conclusions can be drawn from this study:
  • The accuracy of a 3D scanner is affected by the scanning length and pattern of scanning.
  • The accuracy of an optical scan is dependent on the technology used by each scanner.
  • Among the scanners evaluated, EinScan Pro 2X Plus (EP+) showed the highest accuracy (accuracy until 150 mm of length). EinScan Pro (EP) showed moderate accuracy (accurate until 10 mm of length). iPhone (IPX) and ProMax 3D Mid (PM) showed the least accuracy (accuracy from 10 mm to 120 mm in length).
  • In addition, EP+ showed accuracy measuring the 2 mm of depth (diameter 6 mm). All other scanners (EP, IPX, and PM) showed less accuracy measuring depth.
  • Hence, it is recommended to use EinScan Pro 2X Plus for the face scan for facial driven design and other scanning purposes.
  • For measuring the depth of more than 2 mm, these scanners are not recommended. Further development of the scanners is needed for accurately measuring depth.

Author Contributions

Conceptualization and supervision, S.S.; data curation, formal analysis, investigation, visualization, and writing—original draft, P.A.; validation, S.S.; writing—review and editing, P.A. and S.S.

Acknowledgments

This research was supported by the Maxillofacial Prosthodontics and Dental Materials Research Unit, Faculty of Dentistry, Prince of Songkla University, and the Scholarship Award of Thai Ph.D. Student under Thailand’s Education Hub for Southern Region of ASEAN countries. Prince of Songkla University.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Ahmad, I. Anterior dental aesthetics: Facial perspective. Br. Dent. J. 2005, 199, 15–21. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Zhao, Y.J.; Xiong, Y.X.; Wang, Y. Three-Dimensional Accuracy of Facial Scan for Facial Deformities in Clinics: A New Evaluation Method for Facial Scanner Accuracy. PLoS ONE 2017, 12, e0169402. [Google Scholar] [CrossRef] [PubMed]
  3. Li, C.; Barreto, A.; Chin, C.; Zhai, J. Biometric identification using 3D face scans. Biomed. Sci. Instrum. 2006, 42, 320–325. [Google Scholar] [PubMed]
  4. Xiao, N.G.; Perrotta, S.; Quinn, P.C.; Wang, Z.; Sun, Y.H.P.; Lee, K. On the facilitative effects of face motion on face recognition and its development. Front. Psychol. 2014, 5, 633. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  5. Berlin, N.F.; Berssenbrügge, P.; Runte, C.; Wermker, K.; Jung, S.; Kleinheinz, J.; Dirksen, D. Quantification of facial asymmetry by 2D analysis—A comparison of recent approaches. J. Cranio Maxillofac. Surg. 2014, 42, 265–271. [Google Scholar] [CrossRef]
  6. Kim, J.Y.; Jung, H.D.; Jung, Y.S.; Hwang, C.J.; Park, H.S. A simple classification of facial asymmetry by TML system. J. Cranio Maxillofac. Surg. 2014, 42, 313–320. [Google Scholar] [CrossRef] [PubMed]
  7. Zeike, S.; Choi, K.E.; Lindert, L.; Pfaff, H. Managers’ Well-Being in the Digital Era: Is it Associated with Perceived Choice Overload and Pressure from Digitalization? An Exploratory Study. Int. J. Environ. Res. Public Health 2019, 16, 1746. [Google Scholar] [CrossRef] [Green Version]
  8. Berssenbrügge, P.; Berlin, N.F.; Kebeck, G.; Runte, C.; Jung, S.; Kleinheinz, J.; Dirksen, D. 2D and 3D analysis methods of facial asymmetry in comparison. J. Cranio Maxillofac. Surg. 2014, 42, e327–e334. [Google Scholar] [CrossRef]
  9. Toma, A.; Zhurov, A.; Playle, R.; Ong, E.; Richmond, S. Reproducibility of facial soft tissue landmarks on 3D laser-scanned facial images. Orthod. Craniofac. Res. 2009, 12, 33–42. [Google Scholar] [CrossRef]
  10. Ko, B.C. A Brief Review of Facial Emotion Recognition Based on Visual Information. Sensors 2018, 18, 401. [Google Scholar] [CrossRef]
  11. Zuo, K.J.; Saun, T.J.; Forrest, C.R. Facial Recognition Technology: A Primer for Plastic Surgeons. Plast. Reconstr. Surg. 2019, 143, 1298e–1306e. [Google Scholar] [CrossRef] [PubMed]
  12. Hibler, B.P.; Schwitzer, J.; Rossi, A.M.; Hible, B.P. Assessing Improvement of Facial Appearance and Quality of Life after Minimally-Invasive Cosmetic Dermatology Procedures Using the FACE-Q Scales. J. Drugs Dermatol. 2016, 15, 62–67. [Google Scholar] [PubMed]
  13. Lee, J.G.; Jung, S.J.; Lee, H.J.; Seo, J.H.; Choi, Y.J.; Bae, H.S.; Park, J.T.; Kim, H.J. Quantitative anatomical analysis of facial expression using a 3D motion capture system: Application to cosmetic surgery and facial recognition technology. Clin. Anat. 2015, 28, 735–744. [Google Scholar] [CrossRef] [PubMed]
  14. Huang, Y.H.; Seelaus, R.; Zhao, L.; Patel, P.K.; Cohen, M. Virtual surgical planning and 3D printing in prosthetic orbital reconstruction with percutaneous implants: A technical case report. Int. Med Case Rep. J. 2016, 9, 341–345. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. Beuer, F.; Schweiger, J.; Edelhoff, D. Digital dentistry: An overview of recent developments for CAD/CAM generated restorations. Br. Dent. J. 2008, 204, 505–511. [Google Scholar] [CrossRef] [PubMed]
  16. Davidowitz, G.; Kotick, P.G. The use of CAD/CAM in dentistry. Dent. Clin. N. Am. 2011, 55, 559–570. [Google Scholar] [CrossRef]
  17. Ferrini, F.; Sannino, G.; Chiola, C.; Capparé, P.; Gastaldi, G.; Gherlone, E.F. Influence of Intra-Oral Scanner (I.O.S.) on The Marginal Accuracy of CAD/CAM Single Crowns. Int. J. Environ. Res. Public Health 2019, 16, 544. [Google Scholar] [CrossRef] [Green Version]
  18. Solaberrieta, E.; Garmendia, A.; Brizuela, A.; Otegi, J.R.; Pradíes, G.; Szentpétery, A. Intraoral Digital Impressions for Virtual Occlusal Records: Section Quantity and Dimensions. BioMed Res. Int. 2016, 2016, 1–7. [Google Scholar] [CrossRef] [Green Version]
  19. Richert, R.; Goujat, A.; Venet, L.; Viguie, G.; Viennot, S.; Robinson, P.; Farges, J.C.; Fages, M.; Ducret, M. Intraoral Scanner Technologies: A Review to Make a Successful Impression. J. Health Eng. 2017, 2017, 8427595. [Google Scholar] [CrossRef]
  20. Jacobs, R.; Salmon, B.; Codari, M.; Hassan, B.; Bornstein, M.M. Cone beam computed tomography in implant dentistry: Recommendations for clinical use. BMC Oral Health 2018, 18, 88. [Google Scholar] [CrossRef] [Green Version]
  21. Scarfe, W.; Li, Z.; Aboelmaaty, W.; Scott, S.; Farman, A. Maxillofacial cone beam computed tomography: Essence, elements and steps to interpretation. Aust. Dent. J. 2012, 57, 46–60. [Google Scholar] [CrossRef] [PubMed]
  22. Mangano, C.; Luongo, F.; Migliario, M.; Mortellaro, C.; Mangano, F.G. Combining Intraoral Scans, Cone Beam Computed Tomography and Face Scans: The Virtual Patient. J. Craniofac. Surg. 2018, 29, 2241–2246. [Google Scholar] [CrossRef]
  23. Pellegrino, G.; Mangano, C.; Mangano, R.; Ferri, A.; Taraschi, V.; Marchetti, C. Augmented reality for dental implantology: A pilot clinical report of two cases. BMC Oral Health 2019, 19, 158. [Google Scholar] [CrossRef] [PubMed]
  24. Kumar, M.; D’Souza, S.J. Comparative Evaluation of Two Techniques in Achieving Balanced Occlusion in Complete Dentures. Med. J. Armed Forces India 2010, 66, 362–366. [Google Scholar] [CrossRef] [Green Version]
  25. Hassan, B.; Greven, M.; Wismeijer, D. Integrating 3D facial scanning in a digital workflow to CAD/CAM design and fabricate complete dentures for immediate total mouth rehabilitation. J. Adv. Prosthodont. 2017, 9, 381–386. [Google Scholar] [CrossRef] [PubMed]
  26. Hassan, B.; Gimenez Gonzalez, B.; Tahmaseb, A.; Greven, M.; Wismeijer, D. A digital approach integrating facial scanning in a CAD-CAM workflow for complete-mouth implant-supported rehabilitation of patients with edentulism: A pilot clinical study. J. Prosthet. Dent. 2017, 117, 486–492. [Google Scholar] [CrossRef] [PubMed]
  27. AlGhazzawi, T.F. Advancements in CAD/CAM technology: Options for practical implementation. J. Prosthodont. Res. 2016, 60, 72–84. [Google Scholar] [CrossRef] [PubMed]
  28. Solaberrieta, E.; Garmendia, A.; Minguez, R.; Brizuela, A.; Pradies, G. Virtual facebow technique. J. Prosthet. Dent. 2015, 114, 751–755. [Google Scholar] [CrossRef]
  29. Solaberrieta, E.; Otegi, J.R.; Minguez, R.; Etxaniz, O. Improved digital transfer of the maxillary cast to a virtual articulator. J. Prosthet. Dent. 2014, 112, 921–924. [Google Scholar] [CrossRef]
  30. Lee, H.; Cha, J.; Chun, Y.-S.; Kim, M. Comparison of the occlusal contact area of virtual models and actual models: A comparative in vitro study on Class I and Class II malocclusion models. BMC Oral Health 2018, 18, 109. [Google Scholar] [CrossRef] [Green Version]
  31. Stanley, M.; Paz, A.G.; Miguel, I.; Coachman, C. Fully digital workflow, integrating dental scan, smile design and CAD-CAM: Case report. BMC Oral Health 2018, 18, 134. [Google Scholar] [CrossRef] [PubMed]
  32. Mangano, F.; Mangano, C.; Margiani, B.; Admakin, O. Combining Intraoral and Face Scans for the Design and Fabrication of Computer-Assisted Design/Computer-Assisted Manufacturing (CAD/CAM) Polyether-Ether-Ketone (PEEK) Implant-Supported Bars for Maxillary Overdentures. Scanning 2019, 2019, 4274715. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  33. Jamayet, N.B.; Viwattanatipa, N.; Amornvit, P.; Pornprasertsuk, S.; Chindasombatjaroen, J.; Alam, M.K. Comparison of Crown Width/Length Ratio of Six Maxillary Anterior Teeth between Different Facial Groups in Bangladeshi Population. Int. Med. J. 2014, 21, 49–54. [Google Scholar]
  34. Sitthiphan, P.; Viwattanatipa, N.; Amornvit, P.; Shrestha, B.; Srithavaj, M.T.; Alam, M.K. Comparison of Maxillary Anterior Teeth Crown Ratio (Width/Length) between Gender in Laotian Population. Int. Med. J. 2014, 22, 199–205. [Google Scholar]
  35. Afroz, S.; Rathi, S.; Rajput, G.; Rahman, S.A. Dental Esthetics and Its Impact on Psycho-Social Well-Being and Dental Self Confidence: A Campus Based Survey of North Indian University Students. J. Indian Prosthodont. Soc. 2013, 13, 455–460. [Google Scholar] [CrossRef]
  36. Esfahanizadeh, N.; Daneshparvar, N.; Askarpour, F.; Akhoundi, N.; Panjnoush, M. Correlation between Bone and Soft Tissue Thickness in Maxillary Anterior Teeth. J. Dent. Tehran 2016, 13, 302–308. [Google Scholar]
  37. Rudolph, H.; Salmen, H.; Moldan, M.; Kühn, K.; Sichwardt, V.; Wöstmann, B.; Luthardt, R.G. Accuracy of intraoral and extraoral digital data acquisition for dental restorations. J. Appl. Oral Sci. 2016, 24, 85–94. [Google Scholar] [CrossRef] [Green Version]
  38. Rokaya, D.; Kitisubkanchana, J.; Wonglamsam, A.; Santiwong, P.; Srithavaj, T.; Humagain, M. Nepalese Esthetic Dental (NED) Proportion in Nepalese Population. Kathmandu. Univ. Med. J. 2015, 51, 249–254. [Google Scholar] [CrossRef] [Green Version]
  39. Rokaya, D.; Humagain, M.; Amornvit, P. Maxillary Anterior Teeth Proportions for Creating Esthetically Pleasuring Smile in Nepalese Patients. J. Nepal Health Res. Counc. 2018, 16, 362. [Google Scholar] [CrossRef]
  40. Zagar, M.; Knezović Zlatarić, D. Influence of esthetic dental and facial measurements on the Caucasian patients’ satisfaction. J. Esthet. Restor. Dent. 2011, 23, 12–20. [Google Scholar] [CrossRef]
  41. Rokaya, D.; Bhattarai, B.P.; Suttagul, K.; Kafle, D.; Humagain, M. Mentolabial sulcus: An esthetic-based classification. J. Datta Meghe Inst. Med. Sci. Univ. 2018, 13, 16. [Google Scholar] [CrossRef]
  42. Rokaya, D.; Suttagul, K.; Sapkota, B.; Maharjan, S.K. Gender Based Comparison of the Relationships of Maxillary Anterior Teeth and Facial Measurements. J. Int. Dent. Med. Res. 2018, 11, 465–469. [Google Scholar]
  43. Cowie, J.; Calveley, E.; Bowers, G.; Bowers, J. Evaluation of a Digital Consultation and Self-Care Advice Tool in Primary Care: A Multi-Methods Study. Int. J. Environ. Res. Public Health 2018, 15, 896. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  44. Lawson, N.C.; Burgess, J.O. Clinicians reaping benefits of new concepts in impressioning. Compend. Contin. Educ. Dent. 2015, 36, 152–153. [Google Scholar] [PubMed]
  45. Mangano, F.; Gandolfi, A.; Luongo, G.; Logozzo, S. Intraoral scanners in dentistry: A review of the current literature. BMC Oral Health 2017, 17, 149. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  46. Schepke, U.; Meijer, H.J.; Kerdijk, W.; Cune, M.S. Digital versus analog complete-arch impressions for single-unit premolar implant crowns: Operating time and patient preference. J. Prosthet. Dent. 2015, 114, 403–406.e1. [Google Scholar] [CrossRef]
  47. Goracci, C.; Franchi, L.; Vichi, A.; Ferrari, M. Accuracy, reliability, and efficiency of intraoral scanners for full-arch impressions: A systematic review of the clinical evidence. Eur. J. Orthod. 2016, 38, 422–428. [Google Scholar] [CrossRef]
  48. Mangano, F.G.; Hauschild, U.; Veronesi, G.; Imburgia, M.; Mangano, C.; Admakin, O. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: A comparative in vitro study. BMC Oral Health 2019, 19, 101. [Google Scholar] [CrossRef] [Green Version]
  49. Abduo, J.; Elseyoufi, M. Accuracy of Intraoral Scanners: A Systematic Review of Influencing Factors. Eur. J. Prosthodont. Restor. Dent. 2018, 26, 101–121. [Google Scholar]
  50. Medina-Sotomayor, P.; Pascual-Moscardó, A.; Camps, I. Correction: Accuracy of four digital scanners according to scanning strategy in complete-arch impressions. PLoS ONE 2018, 13, e0209883. [Google Scholar] [CrossRef] [Green Version]
  51. Mutwalli, H.; Braian, M.; Mahmood, D.; Larsson, C. Trueness and Precision of Three-Dimensional Digitizing Intraoral Devices. Int. J. Dent. 2018, 2018, 5189761. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  52. Braian, M.; Wennerberg, A. Trueness and precision of 5 intraoral scanners for scanning edentulous and dentate complete-arch mandibular casts: A comparative in vitro study. J. Prosthet. Dent. 2019, 122, 129–136.e2. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  53. Amornvit, P.; Rokaya, D.; Sanohkan, S. Applications of PEEK in Implant Retained Finger Prosthesis. J. Int. Dent. Med. Res. 2019, 12. in press. [Google Scholar]
  54. Van Der Meer, W.J.; Andriessen, F.S.; Wismeijer, D.; Ren, Y. Application of Intra-Oral Dental Scanners in the Digital Workflow of Implantology. PLoS ONE 2012, 7, e43312. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  55. Besl, P.; McKay, H. A method for registration of 3-D shapes. IEEE Trans. Pattern Anal. Mach. Intell. 1992, 14, 239–256. [Google Scholar] [CrossRef]
  56. Park, H.N.; Lim, Y.J.; Yi, W.J.; Han, J.S.; Lee, S.P. A comparison of the accuracy of intraoral scanners using an intraoral environment simulator. J. Adv. Prosthodont. 2018, 10, 58–64. [Google Scholar] [CrossRef]
  57. Matta, R.E.; Adler, W.; Wichmann, M.; Heckmann, S.M. Accuracy of impression scanning compared with stone casts of implant impressions. J. Prosthet. Dent. 2017, 117, 507–512. [Google Scholar] [CrossRef]
  58. You, Y.; Shen, Y.; Zhang, G.; Xing, X. Real-Time and High-Resolution 3D Face Measurement via a Smart Active Optical Sensor. Sensors 2017, 17, 734. [Google Scholar] [CrossRef] [Green Version]
  59. Lee, S.J.; MacArthur, R.X.; Gallucci, G.O. An evaluation of student and clinician perception of digital and conventional implant impressions. J. Prosthet. Dent. 2013, 110, 420–423. [Google Scholar] [CrossRef]
  60. Aragón, M.L.C.; Pontes, L.F.; Bichara, L.M.; Flores-Mir, C.; Normando, D. Validity and reliability of intraoral scanners compared to conventional gypsum models measurements: A systematic review. Eur. J. Orthod. 2016, 38, 429–434. [Google Scholar] [CrossRef]
Figure 1. Study overview. The study involved scanning, measurements, and comparison.
Figure 1. Study overview. The study involved scanning, measurements, and comparison.
Ijerph 16 05061 g001
Figure 2. Reference points marked on the model and various measurements. X-axis (length), y-axis (length), and z-axis (depth).
Figure 2. Reference points marked on the model and various measurements. X-axis (length), y-axis (length), and z-axis (depth).
Ijerph 16 05061 g002
Figure 3. 3D face model rendering (A) and 3D printed master face model (B).
Figure 3. 3D face model rendering (A) and 3D printed master face model (B).
Ijerph 16 05061 g003
Figure 4. Vernice caliper and various face scanners used in this study. EinScan, EinScan Pro 2X Plus, iPhone X, and Planmeca ProMax 3D Mid.
Figure 4. Vernice caliper and various face scanners used in this study. EinScan, EinScan Pro 2X Plus, iPhone X, and Planmeca ProMax 3D Mid.
Ijerph 16 05061 g004
Figure 5. Mean measurements of caliper and three scanners in three axes: (A) x-axis (length), (B) y-axis (length), and (C) z-axis (depth). Discontinuity in a line shows the inability in measurements.
Figure 5. Mean measurements of caliper and three scanners in three axes: (A) x-axis (length), (B) y-axis (length), and (C) z-axis (depth). Discontinuity in a line shows the inability in measurements.
Ijerph 16 05061 g005
Figure 6. Mean difference of various measurements of various scanners from the caliper in three axes: (A) x-axis (length), (B) y-axis (length), and (C) z-axis (depth). Discontinuity in a line shows the inability in measurements.
Figure 6. Mean difference of various measurements of various scanners from the caliper in three axes: (A) x-axis (length), (B) y-axis (length), and (C) z-axis (depth). Discontinuity in a line shows the inability in measurements.
Ijerph 16 05061 g006
Table 1. Scanning and processing times of various scans (EinScan Pro (EP), EinScan Pro 2X Plus (EP+), iPhone X (IPX), and Planmeca ProMax 3D Mid (PM)).
Table 1. Scanning and processing times of various scans (EinScan Pro (EP), EinScan Pro 2X Plus (EP+), iPhone X (IPX), and Planmeca ProMax 3D Mid (PM)).
ScannersScanning Time (min)Data Processing Time (min)Total Time for Scanning Process (min)
EP2.14 ± 0.034.63 ± 0.116.77 ± 0.15
EP+2.12 ± 0.017.28 ± 0.29.4 ± 0.21
IPX0.29 ± 0.010.28 ± 0.040.57 ± 0.03
PM0.22 ± 0.010.48 ± 0.050.7 ± 0.05
Table 2. Measurements in x-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of various scans from the VC.
Table 2. Measurements in x-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of various scans from the VC.
MeasurementsGroupsMean ± SD (mm)Comparison from Vernier Caliper (VC) Using One-Way ANOVA
VC vs. EPVC vs. EP+VC vs. IPXVC vs. PM
X1VC2.11 ± 0.040.820.62NA<0.001 *
EP2.2 ± 0.15
EP+2.25 ± 0.37
IPXNA
PM5.15 ± 1.04
X2VC10.02 ± 0.050.210.940.39<0.001 *
EP9.04 ± 0.44
EP+10.29 ± 0.25
IPX10.8 ± 1.18
PM15.36 ± 2.15
X3VC50.48 ± 0.25<0.001 *10.96<0.001 *
EP42.41 ± 0.51
EP+50.51 ± 0.4
IPX50.23 ± 1.36
PM59.92 ± 3.34
X4VC100.28 ± 0.06<0.001 *0.79<0.01 *0.99
EP84.55 ± 0.51
EP+99.82 ± 0.63
IPX98.43 ± 1.11
PM100.08 ± 0.94
X5VC120.18 ± 0.05<0.001 *0.970.770.064
EP116.3 ± 0.25
EP+120.02 ± 0.127
IPX119.83 ± 0.99
PM117.36 ± 2.88
SD = standard deviation; NA = not available. * Significant at p < 0.05.
Table 3. Measurements in the y-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of various scans from the VC.
Table 3. Measurements in the y-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of various scans from the VC.
MeasurementsGroupsMean ± SD (mm)Comparison from Vernier Caliper (VC) Using One-Way ANOVA
VC vs. EPVC vs. EP+VC vs. IPXVC vs. PM
Y1VC1.97 ± 0.150.480.25NA<0.001 *
EP2.15 ± 0.09
EP+2.22 ± 0.389
IPXNA
PM4.1 ± 0.46
Y2VC10.08 ± 0.050.04 *0.82NA0.966
EP9.06 ± 0.43
EP+10.24 ± 0.48
IPXNA
PM9.90 ± 0.90
Y3VC50.21 ± 0.05<0.001 *0.620.990.027 *
EP42.60 ± 0.24
EP+50.70 ± 0.81
IPX50.13 ± 0.8
PM48.93 ± 1.49
Y4VC100.52 ± 0.67<0.001 *0.140.951
EP86.28 ± 0.22
EP+101.76 ± 0.96
IPX100.81 ± 0.99
PM100.41 ± 0.57
Y5VC121.30 ± 0.18<0.001 *0.41.650.029 *
EP102.56 ± 0.27
EP+121.98 ± 0.71
IPX120.37 ± 0.93
PM119.53 ± 1.18
Y6VC151.49 ± 0.11<0.001 *0.62<0.001 *0.028 *
EP128.22 ± 0.54
EP+151.86 ± 0.32
IPX150.30 ± 0.60
PM150.15 ± 0.92
SD = standard deviation; NA = not available. * Significant at p < 0.05.
Table 4. Measurements in z-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of scans from the VC.
Table 4. Measurements in z-axis of the face model from the Vernier caliper (VC) and various scans (EP, EP+, IPX, and PM), and the comparisons of scans from the VC.
MeasurementsGroupsMean ± SD (mm)Comparison from Vernier Caliper (VC) Using One-Way ANOVA
VC vs. EPVC vs. EP+VC vs. IPXVC vs. PM
Z1VC2.02 ± 0.080.03 *<0.001 *NANA
EP1.9 ± 0.06
EP+2.23 ± 0.18
IPX and PMNA
Z2VC4.06 ± 0.03<0.001 *<0.001 *NANA
EP3.48 ± 0.09
EP+4.39 ± 0.13
IPX and PMNA
Z3VC5.99 ± 0.05<0.001 *<0.001 *NANA
EP3.3 ± 1.48
EP+1.85 ± 0.07
IPX and PMNA
Z4VC8.04 ± 0.02<0.001 *<0.001 *NANA
EP2.7 ± 0.44
EP+4.69 ± 0.19
IPX and PMNA
Z5VC10 ± 0.05<0.001 *<0.001 *NANA
EP2.46 ± 0.05
EP+1.95 ± 0.05
IPX and PMNA
SD = standard deviation; NA = not available. * Significant at p < 0.05.
Table 5. The classification of the capturing difficulty of the facial structures.
Table 5. The classification of the capturing difficulty of the facial structures.
Capturing DifficultnessLandmarks
EasyForehead, cheek, and chin
MediumEar lobe and eye lids
HardTeeth, extra auditory canal, and nostril

Share and Cite

MDPI and ACS Style

Amornvit, P.; Sanohkan, S. The Accuracy of Digital Face Scans Obtained from 3D Scanners: An In Vitro Study. Int. J. Environ. Res. Public Health 2019, 16, 5061. https://doi.org/10.3390/ijerph16245061

AMA Style

Amornvit P, Sanohkan S. The Accuracy of Digital Face Scans Obtained from 3D Scanners: An In Vitro Study. International Journal of Environmental Research and Public Health. 2019; 16(24):5061. https://doi.org/10.3390/ijerph16245061

Chicago/Turabian Style

Amornvit, Pokpong, and Sasiwimol Sanohkan. 2019. "The Accuracy of Digital Face Scans Obtained from 3D Scanners: An In Vitro Study" International Journal of Environmental Research and Public Health 16, no. 24: 5061. https://doi.org/10.3390/ijerph16245061

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop