Next Article in Journal
Soil Water Measurement Using Actively Heated Fiber Optics at Field Scale
Next Article in Special Issue
Design and Evaluation of an Aerosol Electrometer with Low Noise and a Wide Dynamic Range
Previous Article in Journal
Oil Motion Control by an Extra Pinning Structure in Electro-Fluidic Display
 
 
Font Type:
Arial Georgia Verdana
Font Size:
Aa Aa Aa
Line Spacing:
Column Width:
Background:
Article

Low-Cost GRIN-Lens-Based Nephelometric Turbidity Sensing in the Range of 0.1–1000 NTU

1
Institute for Applied Research, Faculty for Mechanical and Medical Engineering, Furtwangen University, 78054 Villingen-Schwenningen, Germany
2
Institute of Physical and Theoretical Chemistry, University of Tuebingen, 72076 Tuebingen, Germany
3
Department of Mathematics, Physics and Chemistry, Beuth University of Applied Sciences, 13353 Berlin, Germany
4
Process Analysis and Technology (PA&T), Reutlingen Research Institute, Reutlingen University, 72762 Reutlingen, Germany
*
Author to whom correspondence should be addressed.
Sensors 2018, 18(4), 1115; https://doi.org/10.3390/s18041115
Submission received: 21 February 2018 / Revised: 28 March 2018 / Accepted: 4 April 2018 / Published: 6 April 2018
(This article belongs to the Special Issue Sensors for Emerging Environmental Markers and Contaminants)

Abstract

:
Turbidity sensing is very common in the control of drinking water. Furthermore, turbidity measurements are applied in the chemical (e.g., process monitoring), pharmaceutical (e.g., drug discovery), and food industries (e.g., the filtration of wine and beer). The most common measurement technique is nephelometric turbidimetry. A nephelometer is a device for measuring the amount of scattered light of suspended particles in a liquid by using a light source and a light detector orientated in 90° to each other. Commercially available nephelometers cost usually—depending on the measurable range, reliability, and precision—thousands of euros. In contrast, our new developed GRIN-lens-based nephelometer, called GRINephy, combines low costs with excellent reproducibility and precision, even at very low turbidity levels, which is achieved by its ability to rotate the sample. Thereby, many cuvette positions can be measured, which results in a more precise average value for the turbidity calculated by an algorithm, which also eliminates errors caused by scratches and contaminations on the cuvettes. With our compact and cheap Arduino-based sensor, we are able to measure in the range of 0.1–1000 NTU and confirm the ISO 7027-1:2016 for low turbidity values.

Graphical Abstract

1. Introduction

Turbidity is the reduction of transparency due to the amount of scattered light that results from the interaction between a beam of light and matter. The intensity of scattered light from a single particle depends on the size and shape of the particle as well as on the wavelength of the incident light and the scattering angle, which can be described by a scattering coefficient [1]. If a mono-disperse particle solution is analyzed, the scattering intensity increases linearly over a wide range of (low) particle concentrations. As a consequence, the amount of scattered light can be used to determine the particle concentration. This fails with suspensions consisting of several particles with different sizes and shapes.
Turbidity can be analyzed by using turbidimeters, which normally consist of a collimated light source (mainly a laser or a light emitting diode (LED)), a sample chamber, and one or more detectors with a specific orientation to the illumination source [1,2]. The detected light intensity of every turbidimeter has to be converted to a turbidity unit, such as a Formazin Nephelometric Unit (FNU), a Formazin Turbidity Unit (FTU), or a Nephelometric Turbidity Unit (NTU), which depends on the sensor principle or setup, application field, and calibration [3]. Therefore, a calibration with different primary standards of known turbidity values must be performed. The basic setups of turbidimeters can be categorized by the type of the detection angle for the scattered light and the number of light detectors. Basically, three different turbidity sensing methods are used [1,2,3,4]: (i) a method where the transmitted intensity is analyzed with a turbidimeter with respect to the incident intensity, used for turbid samples with considerable turbidity values because a significant reduction of the incident light intensity is needed for precise results, (ii) a method by which a nephelometer measures the intensity of scattered light at 90°, which scales with the particle concentration in the sample, very common and very sensitive to a broad range of different particle sizes, the main advantage of which is the high accuracy at low turbidity levels, and (iii) the so-called ratio method, a combination of analyzing the transmitted light and the 90° scattered light with, normally, one or two more forward-scattered and back-scattered (e.g., 60° and 120°) signals, which is more precise than the others and allows for the measurement of a large range of turbidity values.
To analyze turbidity values in the most common range of 0.1–1000 NTU (e.g., for water analysis [1,5,6,7], process/analytical monitoring [8,9], and drug discovery [10,11]), the nephelometric concept is the best compromise due to the good price–performance ratio and its superlative cost-effectiveness. It combines a simple turbidity sensing setup with high precision, especially at lower turbidity levels. As a consequence, nephelometric turbidity measurements are one of the most common measurements used in the qualitative assessment of water suitability [12]. For drinking water, the turbidity value must not exceed 1.0 NTU ([13]), because the particles that cause turbidity are principal indicators for disinfection processes (distilled water: 0.08 NTU; tap water: 0.54 NTU; raw water: 3.52 NTU; water in rivers: up to 150 NTU; wastewater: about 1000 NTU [1,14,15]). Therefore, turbidity data is useful in drinking water treatment and production as well as wastewater and environmental monitoring [16,17,18].
Furthermore, nephelometric measurements are used in the chemical, pharmaceutical (filtration equipment monitoring [8,9,10,11]), and food industries (the filtration of wine and beer). For quality control or to distinguish between cloudy and clear juices, the turbidity measured with a nephelometer is also a key parameter [19,20,21].
Due to many possible applications, a considerable amount of peer-reviewed literature has emerged during the last few years describing new nephelometer designs and measurement setups. Sampedro et al. developed a new nephelometric turbidimeter in combination with an RGB sensor for obtaining additional information about the color of a sample [22]. Another low-cost nephelometer presented by Omar et al. uses also an IR-LED as a light source and detects within a range of 0–500 NTU [6]. Lambrou et al. presented a nephelometric turbidimeter using a red-laser for illumination, and a photodiode for detection measuring NTU values from 0 to 100 [2]. They also developed a low-cost system for measuring considerably high water quality parameters, including turbidity, temperature, pH, and conductivity [7]. Another new nephelometer applies an IR-LED and a light-to-frequency converter in combination with a microcontroller to detect the scattered light from 0 to 1000 NTU [18]. Tai et al. presented a further nephelometric sensor in the range from 0 to 100 NTU, using an IR-LED and a silicon photoelectric generator, and the sensor was calibrated with a self-synthesized formazin solution [23]. A further publication presented a combination of a temperature and turbidity measurement system for measurements of samples in an NTU range of 1–200 [24]. Hussain et al. developed a turbidimeter especially for analyzing water samples [25].
In this study, we present a new nephelometric turbidimeter, called GRINephy, for measuring within an NTU range of 0.1–1000 NTU, which conforms to the standard water analysis norm for low turbidity values ( < 400 NTU) [26]. The norm requires, for the NTU range of 0–400 NTU, a detection angle of 90° for the scattered light (nephelometric measurement) with respect to the incident beam, an aperture angle of a maximum of 30°, and an illuminating light ray at 860 nm with a limited convergency of 1.5°. For that reason, we used for illumination a fiber-coupled 860 nm SMD-LED, which was fused to the flat surface of a GRIN lens to emit collimated light. GRIN lenses achieve their focusing properties due to a continuous change of the refractive index within the lens material [27]. Therefore, the focus is on the plane surfaces of the lens, which allows for the possibility of glueing the optical fiber directly onto the lens, which is one of the main applications of GRIN lenses. Another GRIN lens collects scattered light and focuses it on a light-to-frequency sensor at a detection angle of 90°. GRIN lenses are also an alternative for standard objectives of obtaining optical properties with low costs. For detection, the aperture angle or the numerical aperture of the GRIN lens is reduced to 30° to conform to ISO 7027-1:2016 by adding an aperture between the GRIN lens and the sensor. According to this regulation, we calibrate and verify our GRINephy with a primary calibration standard called StablCal® (stabilized formazin) and compare the results to a commercially available nephelometer (2100Qis, Hach, Loveland, CO, USA). Our GRINephy provides results in the range of 0.1–1000 NTU that are comparable to the commercial nephelometer. With this compact and cheap GRINephy, we are able to measure with excellent reproducibility and precision at very low turbidity levels by rotating the sample. Hence, we obtain average turbidity values determined by about 50 measured positions of the cuvettes.

2. Materials and Methods

Our nephelometric sensor system was built using a cylindrical self-designed polyoxymethylene (POM) cuvette holder that uses 25-mm-round glass cuvettes (2434706, Hach) (see Figure 1). For illuminating the sample, an SMD LED emitting at 860 nm (SFH 4253, Osram, Munich, Germany)in the near infrared region, which observes the European turbidity norm ISO 7027-1:2016, was used. The LED is coupled into a 0.2-mm-diameter-core fiber (FP200URT, Thorlabs, Newton, NJ, USA)and decoupled by a 0.25 pitch GRIN lens (W40–S0250–063–SBC, GoFoton, Somerset, NJ, USA)housed in the cuvette holder to guarantee parallel light with a 4 mm diameter. A second GRIN lens, at 90° with respect to the decoupling lens, collects the scattered light and focuses it toward a light-to-frequency sensor (TSL237, ams, Premstaetten, Austria). To reduce the detection angle, according to the European turbidity norm, to 20°–30°, a 2 mm aperture between the GRIN lens (normally 55°) and the detector was implemented. An 8-bit, 16 MHz microprocessor (Arduino Uno, Arduino, Turin, Italy) sums pulses from the sensor, which triggers an electrical pulse train with its frequency corresponding to the intensity of the detected light. This is done in different time intervals depending on the light intensity (normally 1 s). The same microprocessor controls a stepper motor (QSH4218-35-10-027, Trinamic, Hamburg, Germany) below the cuvette holder, which enabled us to rotate the sample. Thus, the position of the cuvette is turned automatically to 50 positions (always 7.2° further). At each position, the number of pulses are measured and saved in internal storage. A self-developed C-based Arduino program recognizes statistical outliers using a Grubbs’ test [28,29]. If no outliers are detected, the average of the 50 measured data values is determined and displayed. If one or more outliers are recognized, the measurement is considered as faulty and restarted from a different position (a maximum of 5 times). After starting at 5 different positions, further outliers (a maximum of 5) are considered, resulting in an average of at least 45 values. As a final result, we get a mean frequency value with a corresponding standard deviation. Otherwise, an error will be displayed and supposed existing bubbles should be removed or the cuvette should be changed. For calibration, we used the primary standard formazin with NTU values 0.1, 20, 200, and 1000 (StablCal®, Hach) and compared the results with a commercially available nephelometer (2100Qis, Hach) for validation. For this comparison, further NTU standards, with NTU values of 1, 10, 100, and 800 and some dilutions of 4000 NTU, were used.

3. Results

3.1. Calibration of GRINephy

The calibration of our GRINephy was performed by analyzing different calibration standards (StablCal® Turbidity Standards Calibration Kit, 2662105, 0.1, 20, 200, and 1000 NTU, but not 4000 NTU, Hach). Therefore, each calibration solution was automatically measured on at least 45 positions on the cuvette by rotating the sample. As a final result, the mean values of the frequencies given by the light-to-frequency detector were determined, which is shown in Table 1 (middle column) and Figure 2. These frequency values were converted into the standard NTU values (0.1, 20, 200, and 1000 NTU). The total range of 0.1–1000 NTU was divided into the three sub-regions 0.1–20 NTU, 20–200 NTU, and 200–1000 NTU. For each sub-region, a linear regression was performed (see Figure 2). The x-axis values in Figure 2 correspond to the respective standard NTU values, and the y-axis represents the frequency values (gray, yellow, and blue dots) of the light-to-frequency detector and the converted NTU values (orange crosses).

3.2. Validation of GRINephy

The performance and operational capability of GRINephy was examined by comparing turbidity values measured and determined with our GRINephy with a commercially available nephelometer (Hach 2100Qis). Therefore, four solutions observing StablCal® turbidity standards with NTU values of 1, 10, 100, and 800, and dilutions of the 4000 NTU standard calibration solution, were analyzed. The frequency values measured with GRINephy were converted into NTU values using the linear regression of the four calibration standards (see Figure 2). Table 2 summarizes the results for the measured solutions given by Hach2100Qis and our GRINephy. In Figure 3, one can see the determined turbidity values as a function of the turbidity values given by the Hach2100Qis with a Pearson’s r correlation coefficient of 0.9969.

4. Discussion

To our knowledge, this paper is the first to report a low-cost nephelometer with excellent reproducibility and precision for small turbidity values ( < 400 NTU) conforming to the ISO 7027-1:2016 norm. The illumination of the sample was done by an LED (860 nm) in combination with a GRIN lens instead of a laser due to the lower price, with lower energy consumption, and without a need to comply with laser safety regulations. The application of GRIN lenses guarantees parallel light for illumination and a limitation of the detection angle in collecting scattered light. This scattered light was detected by a light-to-frequency sensor, which is a cheap and simple method with high accuracy and linear-intensity-dependent behavior. The implementation of a stepper motor enables measurements at many different places to reduce disturbing parameters such as scratches and fingerprints on the cuvette by a special algorithm. Therefore, the reproducibility and precision, especially for low turbidity values ( < 5 NTU), was improved. Due to the fact that no regulation on a nephelometric turbidimeter calibration routine exists, the calibration of GRINephy was achieved with primary standard StablCal® formazin solutions (with NTU values 0.1, 20, 200, and 1000 NTU). The choice of using three regression regions to characterize the calibration dataset presented in Figure 2 was motivated by a slight non-linearity in the dataset. To account for errors produced by the optical design, the measured values are divided into three sub-regions, which individually show an absolute linear behavior. Similar to Kelley et al., a buckling of linearity (at 200 NTU) was observed. The measured and analyzed data set of the GRINephy shows comparable results with respect to the commercially available device (see Table 2 and Figure 3). A plot of each mean value of all determined NTU values of the GRINephy against the mean NTU values recorded with the commercially available nephelometer shows an absolutely linear correlation, which indicates the applicability of GRINephy. The comparison of the standard deviation for each NTU value shows that the commercial device is slightly more precise and has a lower deviation (see Table 2). We developed a nephelometric sensor that measures in the range of 0.1–1000 NTU, as does the commercially available nephelometer. By measuring exclusively at 90°, we diverge from the norm for higher NTU values. Implementing a sensor at 0° the sensor to be further improved such that higher turbidity values conforming the norm can be achieved. Other low-cost nephelometers diverge partly from the norm and have further disadvantages, such as restricted accuracy for certain NTU ranges and the usage of a laser instead of an LED. Kelley et al. calibrated their device with a cutting oil solution instead of an accepted primary calibration standard such as formazin. The comparison between their nephelometer and the commercial one shows that there is a deviation in small turbidity values and no linearity in the range from 0 to 1000 NTU. A further nephelometric turbidity sensor presented by Sampredo et al. used only four self-made formazin solutions for calibration in the range of 0–500 NTU, and these solutions were not verified by independent measurement. Lambrou et al. used a laser for illumination and observed a deviation in their linearity. With this sensor, only values between 0 and 100 NTU could be analyzed. The sensor presented by Tai et al. was also only accurate within a range of 0–100 NTU. Hussein et al. claim that their developed sensor confirms to the ISO 7027-1:2016 norm, but they have no focusing element between the sample area and the detector. Therefore, the norm ISO 7027-1:2016 that stipulates that light be collected with angles less than 30° could not be fullfilled. In summary, the compact and cheap nephelometer presented here is optimal for measuring NTU values in the range of 0.1–1000 NTU with high repeatability and precision. This sensor has considerable scope and can be used by environmental organizations, public authorities, or private companies, or even by unaffiliated individuals. With small improvements and changes in the design, it is possible to take inline measurements, which would dramatically increase the scope of such types of sensors. In many turbidity sensing applications, continuous monitoring of the turbidity value by implementation of an inline system would simplify the process substantially.

Acknowledgments

This work was financially supported by the Baden-Württemberg Foundation as part of the project GRIN-SEN in the research program Optical Technologies. We thank the company Hach Lange GmbH for providing us with the opportunity to use the Hach2100Qis nephelometer for testing. The article processing charge was funded by the Baden-Württemberg Ministry of Science, Research and Culture in the funding programme Open Access Publishing.

Author Contributions

Marc Brecht, Andreas Modler, and Michael Metzger conceived the experiment; Michael Metzger and Marc Brecht designed and performed the experiments; Michael Metzger, Marc Brecht, and Alexander Konrad analyzed the data; Michael Metzger wrote the first draft of the manuscript; all authors contributed to the writing and editing of the manuscript.

Abbreviations

The following abbreviations are used in this manuscript:
NTUNephelometric Turbidity Unit
ISOInternational Organization for Standardization
FNUFormazin Nephelometric Unit
FTUFormazin Turbidity Unit
GRINgradient index

References

  1. Lawler, D.M. Encyclopedia of Analytical Science, 2nd ed.; Elsevier: New York, NY, USA, 2005; Volume 10, pp. 343–351. ISBN 978-0127641003. [Google Scholar]
  2. Lambrou, T.P.; Anastasiou, C.C.; Panayiotou, C.G. Anastasiou, C.C.; Panayiotou, C.G. A Nephelometric Turbidity System for Monitoring Residential Drinking Water Quality. In Sensor Applications, Experimentation, and Logistics, Proceedings of the First International Conference, SENSAPPEAL 2009, Athens, Greece, 25 September 2009; Komninos, N., Ed.; Springer: Berlin, Germany, 2010; pp. 43–55. [Google Scholar]
  3. Khairi, M.T.M.; Ibrahim, S.; Yunus, M.A.M.; Faramarzi, M. A review on the design and development of turbidimeter. Sensor Rev. 2015, 35, 98–105. [Google Scholar] [CrossRef]
  4. Bin Omar, A.F.; Bin MatJafri, M.Z. Turbidimeter Design and Analysis: A Review on Optical Fiber Sensors for the Measurement of Water Turbidity. Sensors 2009, 9, 8311–8335. [Google Scholar] [CrossRef] [PubMed]
  5. Barter, P.J.; Deas, T. Comparison of portable nephelometric turbidimeters on natural waters and effluents. N. Z. J. Mar. Freshw. 2003, 37, 485–492. [Google Scholar] [CrossRef]
  6. Omar, A.F.; MatJafri, M.Z. The Swift Turbidity Marker. Phys. Educ. 2011, 46, 90. [Google Scholar] [CrossRef]
  7. Lambrou, T.P.; Panayiotou, C.G.; Anastasiou, C.C. A low-cost system for real time monitoring and assessment of potable water quality at consumer sites. In Proceedings of the 2012 IEEE Sensors, Taipei, Taiwan, 28–31 October 2012; pp. 1–4. [Google Scholar]
  8. Chianese, A.; Bravi, M.; Fazio, E. Turbidimetry and Nephelometry. In Industrial Crystallization Process Monitoring and Control; Wiley-VCH Verlag GmbH & Co. KGaA: Hoboken, NJ, USA, 2012; pp. 51–57. ISBN 978-3527645206. [Google Scholar]
  9. Baker, K.N.; Rendall, M.H.; Patel, A.; Boyd, P.; Hoare, M.; Freedman, R.B.; James, D.C. Rapid monitoring of recombinant protein products: A comparison of current technologies. Trends Biotechnol. 2002, 20, 149–156. [Google Scholar] [CrossRef]
  10. Bevan, C.D.; Lloyd, R.S. A High-Throughput Screening Method for the Determination of Aqueous Drug Solubility Using Laser Nephelometry in Microtiter Plates. Anal. Chem. 2000, 72, 1781–1787. [Google Scholar] [CrossRef] [PubMed]
  11. Lipinski, C.A.; Lombardo, F.; Dominy, B.W.; Feeney, P.J. Experimental and computational approaches to estimate solubility and permeability in drug discovery and development settings. Adv. Drug Deliv. Rev. 2001, 46, 3–26. [Google Scholar] [CrossRef]
  12. Gholizadeh, M.H.; Melesse, A.M.; Reddi, L. A Comprehensive Review on Water Quality Parameters Estimation Using Remote Sensing Techniques. Sensors 2016, 16, 1298. [Google Scholar] [CrossRef] [PubMed]
  13. Gregory, J. Turbidity and beyond. Filtr. Sep. 1998, 35, 63–67. [Google Scholar] [CrossRef]
  14. Huey, G.M.; Meyer, M.L. Turbidity as an Indicator of Water Quality in Diverse Watersheds of the Upper Pecos River Basin. Water 2010, 2, 273–284. [Google Scholar] [CrossRef]
  15. Amosa, M.K.; Jami, M.S.; Alkhatib, M.F.R.; Tajari, T.; Jimat, D.N.; Owolabi, R.U. Turbidity and suspended solids removal from high-strength wastewater using high surface area adsorbent: Mechanistic pathway and statistical analysis. Cogent Eng. 2016, 3, 1162384. [Google Scholar] [CrossRef]
  16. LeChevallier, M.W.; Evans, T.M.; Seidler, R.J. Effect of turbidity on chlorination efficiency and bacterial persistence in drinking water. Appl. Environ. Microbiol. 1981, 1, 159–167. [Google Scholar]
  17. LeChevallier, M.W.; Norton, W.D.; Lee, R.G. Giardia and Cryptosporidium spp. in filtered drinking water supplies. Appl. Environ. Microbiol. 1991, 9, 2617–2621. [Google Scholar]
  18. Kelley, C.D.; Krolick, A.; Brunner, L.; Burklund, A.; Kahn, D.; Ball, W.P.; Weber-Shirk, M. An Affordable Open-Source Turbidimeter. Sensors 2014, 14, 7142–7155. [Google Scholar] [CrossRef] [PubMed]
  19. Dickinson, E. Colloidal aspects of beverages. Food Chem. 1994, 51, 343–347. [Google Scholar] [CrossRef]
  20. Dongare, M.L.; Buchade, P.B.; Awatade, M.N.; Shaligram, A.D. On-line turbidity measurement of clear juice. J. Opt. 2013, 42, 92–95. [Google Scholar] [CrossRef]
  21. Markowski, J.; Baron, A.; Mieszczakowska, M.; Plocharski, W. Chemical composition of French and Polish cloudy apple juices. J. Hortic. Sci. Biotechnol. 2009, 84, 68–74. [Google Scholar] [CrossRef]
  22. Sampedro, Ó.; Salgueiro, J.R. Turbidimeter and RGB sensor for remote measurements in an aquatic medium. Measurement 2015, 68, 128–134. [Google Scholar] [CrossRef]
  23. Tai, H.; Li, D.; Wang, C.; Ding, Q.; Wang, C.; Liu, S. Design and characterization of a smart turbidity transducer for distributed measurement system. Sens. Actuators A Phys. 2012, 175, 1–8. [Google Scholar] [CrossRef]
  24. Nuzula, N.I.; Sakinah, W. Manufacturing temperature and turbidity sensor based on ATMega 8535 microcontroller. AIP Conf. Proc. 2017, 1788, 030108. [Google Scholar] [CrossRef]
  25. Hussain, I.; Ahamad, K.; Nath, P. Water turbidity sensing using a smartphone. RSC Adv. 2016, 6, 22374–22382. [Google Scholar] [CrossRef]
  26. International Organization for Standardization. ISO 7027-1:2016 Water quality—Determination of Turbidity; International Organization for Standardization: Geneva, Switzerland, 2016. [Google Scholar]
  27. Moore, D.T. Gradient-index optics: A review. Appl. Opt. 1980, 19, 1035–1038. [Google Scholar] [CrossRef] [PubMed]
  28. Grubbs, F.E. Sample Criteria for Testing Outlying Observations. Ann. Math. Statist. 1950, 21, 27–58. [Google Scholar] [CrossRef]
  29. Grubbs, F.E. Procedures for Detecting Outlying Observations in Samples. Technometrics 1969, 11, 1–21. [Google Scholar] [CrossRef]
Figure 1. Sketch of GRINephy, which shows the main parts of the sensor including the illumination part, the detection unit, and the chamber parts.
Figure 1. Sketch of GRINephy, which shows the main parts of the sensor including the illumination part, the detection unit, and the chamber parts.
Sensors 18 01115 g001
Figure 2. Calibration of GRINephy: (top) Output frequency signals of our GRINephy given in counts/s for measuring formazin turbidity calibration standards with the NTU values 0.1, 20, and 200 (gray and yellow dots). (bottom) Sub-region of 200–1000 NTU with the output frequency of our GRINephy using 200 and 1000 NTU calibration standards. The orange crosses show in both figures the converted NTU values.
Figure 2. Calibration of GRINephy: (top) Output frequency signals of our GRINephy given in counts/s for measuring formazin turbidity calibration standards with the NTU values 0.1, 20, and 200 (gray and yellow dots). (bottom) Sub-region of 200–1000 NTU with the output frequency of our GRINephy using 200 and 1000 NTU calibration standards. The orange crosses show in both figures the converted NTU values.
Sensors 18 01115 g002
Figure 3. Validation of GRINephy: Values from the Hach2100Qis (x-axis) and the GRINephy (y-axis) using standard and diluted StablCal® solutions.
Figure 3. Validation of GRINephy: Values from the Hach2100Qis (x-axis) and the GRINephy (y-axis) using standard and diluted StablCal® solutions.
Sensors 18 01115 g003
Table 1. Mean and standard deviation of measured StablCal® standards with our GRINephy in counts per second (middle column) and the converted NTU values (right column).
Table 1. Mean and standard deviation of measured StablCal® standards with our GRINephy in counts per second (middle column) and the converted NTU values (right column).
StablCal® Standard (NTU)GRINephy (counts/s)GRINephy (NTU)
0.151.5 ± 0.10.10 ± 0.01
20150 ± 120.00 ± 0.2
200957 ± 3200 ± 1
10003105 ± 81000 ± 3
Table 2. Mean turbidity and standard deviation of measured standard StablCal® solutions and dilutions with the commercial Hach2100Qis nephelometer and our GRINephy.
Table 2. Mean turbidity and standard deviation of measured standard StablCal® solutions and dilutions with the commercial Hach2100Qis nephelometer and our GRINephy.
SampleHACH2100Qis (NTU)GRINephy (NTU)
Standard 1 NTU1.09 ± 0.010.993 ± 0.04
dilution 14.50 ± 0.023.94 ± 0.08
dilution 27.91 ± 0.027.57 ± 0.09
Standard 10 NTU10.3 ± 0.19.64 ± 0.1
dilution 315.8 ± 0.115.4 ± 0.1
dilution 442.6 ± 0.240.7 ± 0.2
dilution 568.6 ± 0.364.4 ± 0.3
Standard 100 NTU104 ± 198.7 ± 0.5
dilution 6119 ± 1114 ± 1
dilution 7298 ± 1325 ± 1
dilution 8405 ± 1456 ± 1
dilution 9486 ± 1532 ± 1
dilution 10607 ± 2644 ± 2
dilution 11671 ± 2724 ± 2
dilution 12793 ± 2831 ± 2
Standard 800 NTU821 ± 2836 ± 2
dilution 13917 ± 2912 ± 3
dilution 14978 ± 2972 ± 3

Share and Cite

MDPI and ACS Style

Metzger, M.; Konrad, A.; Blendinger, F.; Modler, A.; Meixner, A.J.; Bucher, V.; Brecht, M. Low-Cost GRIN-Lens-Based Nephelometric Turbidity Sensing in the Range of 0.1–1000 NTU. Sensors 2018, 18, 1115. https://doi.org/10.3390/s18041115

AMA Style

Metzger M, Konrad A, Blendinger F, Modler A, Meixner AJ, Bucher V, Brecht M. Low-Cost GRIN-Lens-Based Nephelometric Turbidity Sensing in the Range of 0.1–1000 NTU. Sensors. 2018; 18(4):1115. https://doi.org/10.3390/s18041115

Chicago/Turabian Style

Metzger, Michael, Alexander Konrad, Felix Blendinger, Andreas Modler, Alfred J. Meixner, Volker Bucher, and Marc Brecht. 2018. "Low-Cost GRIN-Lens-Based Nephelometric Turbidity Sensing in the Range of 0.1–1000 NTU" Sensors 18, no. 4: 1115. https://doi.org/10.3390/s18041115

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop