Article

Development and Validation of Targeted Metabolomics Methods Using Liquid Chromatography–Tandem Mass Spectrometry (LC-MS/MS) for the Quantification of 235 Plasma Metabolites

by
Kangkang Xu
1,2,3,
Franz Berthiller
1,2,
Barbara U. Metzler-Zebeli
2,4 and
Heidi E. Schwartz-Zimmermann
1,2,*
1
Institute of Bioanalytics and Agro-Metabolomics, Department of Agricultural Sciences, BOKU University, 3430 Tulln, Austria
2
Christian Doppler Laboratory for Innovative Gut Health Concepts of Livestock, 1210 Vienna, Austria
3
Austrian Competence Centre for Feed and Food Quality, Safety and Innovation (FFoQSI), 3430 Tulln, Austria
4
Unit Nutritional Physiology, Department of Biomedical Sciences, University of Veterinary Medicine, Veterinaerplatz 1, 1210 Vienna, Austria
*
Author to whom correspondence should be addressed.
Molecules 2025, 30(3), 706; https://doi.org/10.3390/molecules30030706
Submission received: 20 December 2024 / Revised: 15 January 2025 / Accepted: 23 January 2025 / Published: 5 February 2025
(This article belongs to the Special Issue The Application of LC-MS in Pharmaceutical Analysis)

Abstract

:
Plasma contains metabolites with diverse physicochemical properties, ranging from highly polar to highly apolar, and concentrations spanning at least nine orders of magnitude. Plasma metabolome analysis is valuable for monitoring health and evaluating medical interventions but is challenging due to the metabolome’s diversity and complexity. This study aims to develop and validate targeted LC-MS/MS methods for quantifying 235 mammalian metabolites from 17 compound classes in porcine plasma without prior derivatization. Utilizing reversed-phase and hydrophilic interaction liquid chromatography coupled with tandem mass spectrometry, each analyte is identified and quantified using two selected reaction monitoring (SRM) transitions. Fast polarity switching and scheduled SRM enhance the metabolome coverage and throughput, enabling the analysis of one sample in about 40 min. A simple “dilute and shoot” sample preparation protocol was employed, with samples injected at two dilution levels to align metabolite concentrations within calibration curve ranges. Validation in porcine plasma included assessments of carryover, linearity, detection and quantification limits, repeatability and recovery. The method was further applied to plasma samples from various animal species, demonstrating its applicability to human and animal studies. This study establishes two robust LC-MS/MS methods for comprehensive porcine plasma metabolome quantification, advancing large-scale targeted metabolomics in biomedical research.

1. Introduction

Metabolomics is the comprehensive study of a wide range of low-molecular-weight compounds, called metabolites, in a biological system. Metabolomics provides a snapshot of the metabolic products of an organism during biological processes. Metabolomics facilitates an enhanced understanding of biochemical and biological mechanisms in intricate systems [1]. It has been commonly utilized as a tool for biomarker discovery and metabolic pathway research [2,3].
Unlike untargeted metabolomics, which aims at measuring all the metabolites in a sample without prior information and provides only relative analyte abundances, targeted metabolomics allows researchers to identify and quantity specific sets of known metabolites [4]. A necessity to determine a wide range of metabolites is the use of a universal detection method, such as mass spectrometry (MS) or nuclear magnetic resonance (NMR). MS- or tandem mass spectrometry (MS/MS)-based metabolomics allows for the profiling of a large number of metabolites over a wide concentration range in a single analysis—a feature that less sensitive metabolomics assays based on NMR cannot compete with. In targeted metabolomics, a triple-quadrupole mass spectrometer (QqQ) is commonly used in selected reaction monitoring (SRM) mode, which is more affordable, sensitive and robust than high-resolution mass spectrometry [5]. Absolute quantification is then realized on the basis of calibration curves established from authentic reference compounds or by comparison with a specified quantity of isotopically labeled internal standards.
Flow injection analysis with tandem mass spectrometry (FIA-MS/MS), one of the formats in targeted metabolomics, works without an analytical column [6,7,8]. The disadvantages of FIA-MS/MS are the occurrence of matrix effects due to the lack of a chromatographic separation step and the potential confusion of isomeric and even isobaric compounds, which can lead to the incorrect identification and quantification of the analyte if no unique fragmentation patterns are available [9,10]. In contrast, the combination of liquid chromatography with tandem mass spectrometry (LC-MS/MS) separates analytes in complex samples efficiently according to the diverse physicochemical properties of the metabolites. This decreases matrix effects, improves limits of detection and quantification and greatly reduces the misidentification of isomeric compounds. In this respect, hydrophilic interaction chromatography (HILIC) has demonstrated its ability to separate polar and charged metabolites in biological matrices, which are hardly retained in reversed-phase liquid chromatography (RPLC) [11,12,13]. As one single analytical method is insufficient to cover the entire metabolome, a combination of chromatographic separation methods can significantly increase metabolite coverage [14,15].
Metabolomics analysis has been extensively applied to study biofluids like blood, plasma, serum, saliva and urine, as reviewed in [16,17]. The blood metabolome is used to evaluate and monitor internal and external perturbations of an organism’s metabolism [18,19], providing insights into its physiological and health conditions [20]. Plasma is the supernatant obtained by centrifuging blood that has been treated with an anticoagulant (e.g., EDTA, sodium citrate or heparin) to prevent clotting, thereby separating it from the cellular components. Plasma comprises metabolites with diverse physicochemical properties, from highly polar to highly apolar, and their concentrations span nine orders of magnitude [21,22,23,24]. The complexity of the plasma metabolome poses a challenge for the choice of an analysis platform. In recent years, large-scale targeted metabolomics methods for the quantification of hundreds of human or rodent plasma metabolites from diverse compound classes were established, e.g., [13,25,26,27,28,29]. For instance, McMillen et al. [27] developed two large-scale, targeted LC-MS/MS methods based on RPLC and HILIC, respectively, to (semi)quantify 540 metabolites in human serum. Likewise, Lee and co-workers [26] presented two RPLC-MS/MS and HILIC-MS/MS methods for the quantification of over 200 metabolites related to core metabolic pathways in a wide range of biological samples, including serum, tissues and cells from humans and mice. Li et al. [13], Cao et al. [25] and Nandania and co-workers [29] developed and used single HILIC-MS/MS methods for the quantitative determination of 610 [13], 428 [25] and 102 [29] metabolites in various biological matrices including plasma or serum. Wei et al. [28] determined 205 polar metabolites in rat plasma by the parallel use of three analytical columns with different separation mechanisms and using a sophisticated system of sequential elution and MS/MS detection. However, most of these assays show notable limitations, such as lack of certain validation parameters or poor validation results [27]. Apparent recoveries that are crucial for accurate quantification were provided only by McMillen et al. [27] and Nandania and co-workers [29]. Limits of detection and quantification were only stated by McMillen [27], Wei et al. [28] and Cao et al. [25], and only partly for a selected number of analytes (96 metabolites for [27], 20 amino acids for [25]). Only Cao et al. [25] and Lee et al. [26] provided two SRM transitions for each analyte to ensure correct compound identification. Additionally, to date, there are no fully validated LC-MS/MS-based targeted metabolomics methods available for porcine plasma. Instead, ready-to-use commercial metabolomic kits relying on the derivatization of amino acids and amines have been widely utilized for the quantification of porcine plasma metabolites, e.g., [20,30,31,32,33]. In addition, a set of three RP-HPLC-MS/MS methods was developed for the quantitative determination of 176 porcine metabolites from various compound classes. These methods also employed the derivatization of amino acids, amino acid-related compounds and biogenic amines to improve the ionization and separation of analytes [34] but were not validated for porcine plasma. Moreover, the derivatization step prolongs and complicates sample preparation, which can lead to impaired repeatability. Likewise, some of the derivatized analytes have limited stability, necessitating immediate analysis or the repetition of sample work-up. Another common challenge in targeted LC-MS/MS methods is managing the trade-off between analysis time and metabolome coverage. Multiple methods are required to achieve a high metabolome coverage, which prolongs the analysis time and increases analysis costs.
In this study, we aim to develop and thoroughly validate two targeted multi-analyte LC-MS/MS methods for the determination of 235 mammalian metabolites from 17 compound classes in porcine plasma samples. Both RPLC and HILIC coupled to QqQ-MS will be used after simple sample preparation without derivatization. To further improve the correct identification of analytes, two SRM transitions (a quantifier and a qualifier) are to be measured. Fast polarity switching and scheduled SRM (sSRM) will be employed to simultaneously quantify metabolites in both polarities. The analytical performance of these methods, including limits of detection and quantification, linearity, repeatability, apparent recovery, carryover and trueness, will be determined. Furthermore, we will test the applicability of the methods to plasma samples from other species. Our large-scale targeted metabolomics methods will overcome current limitations in porcine plasma metabolomics, including metabolite misidentification, speed of analysis and accuracy of quantification.

2. Results and Discussion

2.1. Method Development

2.1.1. Targeted Metabolites

A total of 235 metabolites were included in our two LC-MS/MS methods, selected based on the commercial availability of standards and their suitability for reliable quantification under LC conditions. The metabolites were grouped into 17 compound classes of biological significance to human and animal health, with a particular focus on inflammation and gut dysbiosis. These classes included polar compounds such as amino acids, sugars and biogenic amines, as well as signaling lipids like fatty acids, phospholipids and sterols [34,35]. The physicochemical diversity of the metabolites in this assay is demonstrated by the range of calculated log p values that partially reflect the compounds’ lipophilicity (Figure 1).
The initial assignment of metabolites to an LC method was based on retention factors determined in our preceding study [14]. Metabolites with a retention factor > 1 in the bare-silica HILIC method were chosen for HILIC method development, while those well retained in the RP-C18 method or under lipidomics conditions were selected for the RP method. Metabolites potentially suitable for both methods were originally measured under both conditions. Eventually, metabolites detected in both polarities and/or with both methods were assigned to the method and polarity in which they showed the highest signal intensity. With only a few exceptions, each metabolite was measured only in one polarity per method.

2.1.2. LC Conditions

One of the major challenges in plasma metabolomics is separating highly diverse classes of metabolites, which cannot be achieved using just a single chromatographic method. Therefore, we initially selected the RP-C18 method and the bare-silica HILIC method from our previous study to enhance the metabolic coverage [14]. Mobile phase composition, flow rate and column oven temperature were modified to achieve optimal separation performance. To minimize the potential carryover of highly hydrophobic lipids, the solvent strength of mobile phase B in the RP method was increased by replacing pure acetonitrile (ACN) with 47.5/47.5/5 isopropanol (IPA)/ACN/water (v/v/v). Additionally, the isocratic wash period of the RP run was extended to 11.5 min. To enhance the removal of highly polar compounds in the HILIC method, the percentage of the aqueous mobile phase was increased from the original 40% to 60% in the isocratic wash period. Furthermore, the flow rate was doubled during the wash periods to accelerate the methods, while the oven temperature was increased to ensure that the column pressure remained within an acceptable range. To avoid issues like pump pressure fluctuations, the organic mobile phase of both methods contained 5% water, ensuring the solubility of solvent additives.
Chromatograms showing the majority of the metabolites detected in the porcine quality control (QC) plasma sample are shown in Figure 2 (HILIC) and Figure 3 (RP). The metabolites analyzed by HILIC-MS/MS were eluted mainly between 1 and 8 min, whereas the compounds targeted by the RP method were eluted within the gradient from 2 to 15 min. Notably, seven phospholipids had broader peaks, as exemplified by peak 26 and peak 29 in Figure 3. As discussed in our previous study [14], we observed similarly broad peaks for other phosphorylated metabolites, such as nucleotides and sugar phosphates. This can be explained by the partial adsorption of phosphate-containing analytes onto metal surfaces within the LC-ESI-MS system, such as stainless-steel components used in the injector, column housing, tubing, connectors and similar parts. Additionally, these seven phospholipids had slightly different retention times in porcine plasma compared to those in pure solvent standards. This indicates the presence of isomers with different double bond positions in the samples, which caused retention time shifts. Therefore, the detection window for these seven metabolites was increased to 60 s to account for this variability.
A significant source of interference arises from isomeric or isobaric metabolites that exhibit similar chromatographic behaviors and potentially have similar primary fragmentation patterns [28]. The separation of these compounds, while concurrently resolving a broad spectrum of other metabolites, represents an inherent challenge in targeted metabolomics. Remarkably, certain isomeric metabolites can be effectively baseline separated. For instance, α-aminobutyric acid (AABA), β-aminobutyric acid (BABA) and γ-aminobutyric acid (GABA) represent three isomers of aminobutyric acids, which were independently eluted with retention time differences of 0.05 min and 0.22 min between each pair. An LC discrimination between symmetric N,N′-dimethylarginine (SDMA) and asymmetric N,N-dimethylarginine (ADMA) was also achieved. Other isomeric pairs, e.g., valine and norvaline, were partially separated, requiring deconvolution for peak integration. In contrast, leucine and isoleucine could not be separated in our HILIC method. Instead, they were differentiated by their specific MS/MS fragmentation patterns. The unique transition of m/z 132.1 → 69 enabled the quantification of isoleucine, while another transition, m/z 132.1 → 89, where both compounds had similar peak areas at equal concentrations, was used to estimate the sum concentration of leucine and isoleucine. The leucine concentration was then calculated by subtracting the isoleucine concentration from the sum concentration.
Nonetheless, certain isobaric compounds remained unresolved in the current methods. For example, hexoses, pentoses, alanine, β-alanine and sarcosine could be separated neither chromatographically nor by mass spectrometry. Consequently, the provided concentrations of hexoses, pentoses and alanine isobaric compounds are sum concentrations.

2.1.3. MS/MS Conditions

Our methods use two SRM transitions (quantifier and qualifier), enabling the calculation of the ion ratio, which can be used with the retention time for unambiguous compound identification. The best-suited quantifier and qualifier transitions were selected by measuring authentic standards with both LC methods in both polarities using all SRM transitions provided in our previous work [14]. For each metabolite, the two SRM transitions showing the highest signal-to-noise ratios were chosen for the final methods. Exceptions included cholesteryl myristate, cholesteryl palmitoleate and 3-hydroxyphenylacetic acid, for which only one reliable SRM transition was available. Additionally, six metabolites, namely serine, myo-inositol and four bile acids, were measured in both polarities because it was impossible to identify two reliable SRM transitions in a single polarity. As a result, a total of 467 SRM transitions were incorporated into our targeted LC-MS/MS methods.
We applied fast polarity switching in our LC-MS/MS methods to measure in both positive and negative polarity within one run, reducing two separate injections to one. However, impaired repeatability is a common concern when simultaneously targeting a large number of metabolites in both negative and positive ion modes. This is attributed to fewer data points within one chromatographic peak when accelerating large-scale targeted metabolomics assays, which in return diminish accurate and precise quantification [36]. This disadvantage was tackled by further employing scheduled SRM, where each SRM transition is monitored within a narrow retention time window to ensure an optimal dwell time and number of data points for higher levels of SRM multiplexing [37].
Dwell times in our methods were optimized and generated by SciexOS. Modern QqQ instruments have demonstrated reliable quantification with dwell times as short as 3 ms [38]. As shown in Figure 4, despite a large number of compounds eluting around 5 min and 8 min, the lowest dwell times of the transitions in our HILIC and RP methods were 7 and 10 ms, respectively.

2.2. Method Validation

2.2.1. Linearity

The determination of the limits of detection (LODs), lower limits of quantification (LLOQs) and upper limits of quantification (ULOQs) was conducted using 14 calibration levels of standard solutions. Analytes with poor linearity (extremely high LLOQs) due to weak ionization efficiency were excluded from our methods. Examples of these compounds included keto- and aromatic carboxylic acids and several cholesterol esters.
The linear range varied among metabolites, with 64% of the metabolites exhibiting a wide linear response over two orders of magnitude, compared to 33% with a narrower linear range of one to two orders of magnitude. Metabolites displaying linearity below one order of magnitude, for instance dopamine, dehydroergosterol and myristic acid, also possessed high LLOQs. Overall, our method demonstrated good linearity for most compounds. Moreover, 99% of the porcine metabolites quantified in the applicability test had a coefficient of determination (r2) exceeding 0.900; 64% of the quantified porcine metabolites showed a coefficient of determination ≥ 0.995; and 30% had a coefficient of determination ≥ 0.999.

2.2.2. Carryover

Carryover can adversely impact the accurate qualitative and quantitative determination of compounds and may restrict the dynamic range of quantitation methods based on LC-MS [39]. The calculated carryover percentage for each analyte is provided in Table S1. In total, 203 out of 235 metabolites had a carryover rate of less than 5%, of which 152 metabolites exhibited a carryover below 1%. Only 6% of the metabolites covered by HILIC and RP, respectively, exceeded 5% carryover. Notably, the majority of metabolites with high carryover in the RP method were long-chain fatty acids, such as oleic acid and arachidic acid. The generally low carryover can be attributed to our systematic approach to counteracting analyte adsorption, including a rigorous autosampler cleaning plan involving washes with multiple column volumes of strong eluent at the end of the gradient and blank injections throughout the sequence.

2.2.3. Recovery

Recovery experiments, aimed at evaluating analyte losses during sample preparation in addition to assessing mass spectrometric matrix effects, involve adding defined amounts of standard compounds to the sample prior to work-up [40]. Apparent recoveries (RAs) were assessed by the standard addition method, which partly requires spiking very high concentrations of metabolites into plasma samples. The limited solubility of some compounds makes it impossible to simply spike solutions into plasma samples without significantly diluting the sample. As such, for method validation only, stock solutions had to be evaporated before taking up metabolites in plasma or pure solvents. Thus, we comprehensively tested various solvent mixtures with decreasing polarity for the uptake of RP metabolites, namely ACN/water (80/20, v/v, polarity index = 6.8), methanol (MeOH) (polarity index = 6.6), ACN (polarity index = 6.2), acetone/water (80/20, v/v, polarity index = 6.1), ACN/IPA (50/50, v/v, polarity index = 5.3), IPA/water (80/20, v/v, polarity index = 5.2), acetone/IPA (50/50, v/v, polarity index = 4.9) and pure IPA (polarity index = 4.3) (detailed results not shown). IPA/water (80/20, v/v) yielded the highest recovery of 91% of total RP metabolites. To verify if IPA/water (80/20, v/v) was also suitable for the uptake of HILIC compounds, it was compared to ACN/water (80/20, v/v), which is routinely used as extraction solvent for HILIC metabolites. As both solvent mixtures yielded similar results, IPA/water (80/20, v/v) was used as the only extraction solvent for the work-up of samples and for the validation experiments for both methods.
The standard addition protocol was applied to all 235 metabolites, including those that were not detected in our QC porcine plasma sample. This approach was chosen because any of these metabolites could potentially occur in plasma and serve as biomarker, depending on the design of animal trials and/or the animal species analyzed. The standard addition plan involved spiking six levels into porcine QC plasma in triplicate. This, however, was impossible for certain undetected metabolites exhibiting high LLOQs, primarily hydrophobic fatty acids, cholesterol ester and sterols. For these 31 metabolites, only one high level was spiked into plasma, as excessive concentrations could negatively impact solubility and subsequent reconstitution. Recovery data for both dilutions of plasma (1:5, 1:100, v:v) are presented in Table S2.
Recovery rates between 70% and 130% are considered acceptable [40]. Any deviations outside of the acceptable range indicate that a metabolite requires further investigation to ensure accurate quantification. Metabolites showing low and inconsistent RAs were eliminated from our methods. For instance, five hydrophobic metabolites (e.g., 1,2-dioleoyl-sn-glycerol, cholesteryl arachidonate and C18 galactosyl(β) ceramide) originating from glycerides, cholesterol esters and phospholipids exhibited poor and erratic linearity in both spiked samples and standards in pure solvent solutions. Low and inconsistent recoveries of certain metabolites, particularly associated with hydrophobic compounds, pose a challenge for LC-MS/MS method development and validation [41]. These issues arise from analyte losses during sample preparation and/or from matrix effects during analysis [41]. Furthermore, no single pre-treatment adequately covers the entire metabolome [42]. In particular, large-scale LC-MS/MS methods target diverse metabolite classes with varied physio-chemical properties, and one single set of sample preparation could compromise the recoveries of some analytes. This can be exemplified by cholesteryl arachidate and C18 galactosyl(β) ceramide. Both had a poor recovery of 49% and 208%, respectively, using IPA/water (80/20, v/v) as the extraction solvent and therefore yielded low or variable RAs.
As shown in Figure 5 and Figure 6, the average RAs across all compound classes met the acceptance criteria, regardless of plasma dilution. However, single analytes had lower or, rarely, higher RAs, most likely due to matrix effects. In general, matrix effects can be minimized by injecting smaller volumes or diluting the samples, provided the compounds are still above the LLOQ [43]. For five compound classes (carboxylic acids, eicosanoids, ethanolamides, nucleotides and sugars), only the 1:5 plasma dilution yielded reliable apparent recoveries, as a 1:100 dilution reduced their concentrations below the LLOQs. For certain compound classes where both plasma dilutions were feasible, the RAs and intra-class consistency improved upon 1:100 dilution, presumably due to reduced matrix effects. For example, the average recovery of biogenic amines increased from 77% to 91%, and the average recovery of amino acids improved from 72 to 83%. Additionally, the coefficient of variation (CV%) decreased significantly for some compound classes after dilution, with glycerides improving from 30% to 8% and biogenic amines from 22% to 7%.
Glucose was excluded from the 1:100 dilution analysis due to its high concentration in QC plasma (>400 mg/L, estimated by 1:500 dilution) and relatively low ULOQ (3 mg/L in measurement solution). Despite the dilution (1:100, v:v), glucose had an RA of only 29% with poor repeatability (105%). Because of the high plasma concentration and the poor validation results at a 1:100 dilution, glucose was excluded from the method. A 1:500 dilution would be required for the validation and accurate quantification of glucose.

2.2.4. Repeatability

Repeatability was expressed as the relative standard deviation (RSD) of the recoveries determined at six spiking levels in triplicate (n = 18). A total of 94% of metabolites exhibited consistent RSD values of ≤25%. Previous studies have linked poor repeatability or reproducibility to low peak intensity, low recovery or poor chromatography [28,29]. This aligns with our findings for some metabolites with high RSD values, such as cholesteryl docosapentaenoate and C24 dihydroceramide (d18:0/24:0), showing low peak intensity and low recovery despite high spiking concentrations.

2.2.5. Trueness and Applicability to Plasma from Other Animal Species

Due to the absence of a certified reference material for porcine plasma, the human plasma NIST standard reference material (SRM) 1950 was analyzed as a surrogate to assess the trueness of the method, given the physiological similarities of the two species. The concentrations obtained by our method were in line with the NIST reference values. Overall, the concentrations of 11 out of 15 NIST metabolites were within 70–130% of the certified values, demonstrating the excellent trueness of our analytical approach (Table 1).
Both cholecalciferol and calciferol were below the LODs in our methods. These compounds had low concentrations in the NIST reference sample, where dedicated methods had been used for quantification. Glycine was removed from our HILIC method due to a significant mismatch with the NIST reference values. The NIST value calculated by our HILIC method was eleven times higher than the reference value. As glycine did not fragment in MS/MS analysis, two “pseudo-molecular” SRM transitions were selected, with SRM transitions from the protonated/deprotonated precursor ions to the respective precursor ions. The non-specific SRM transitions could be more susceptible to matrix interference, leading to an overestimation of the measured values in the NIST sample.
In total, our methods quantified 124 out of 235 metabolites in the QC porcine plasma sample. Besides porcine plasma, the method applicability was also tested on plasma from other species, including human, rat, bovine and chicken plasma. A comprehensive compilation of metabolite concentrations is reported in Table S3. Overall, we quantified 159 out of 235 metabolites in the analyzed human and animal plasma samples. Among the 76 metabolites that were not detected, 30 possessed an LOQ ≤ 30 µg/L in measurement solution, indicating at best only low concentrations of these metabolites in plasma. Moreover, 46 undetected metabolites, primarily keto- and aromatic carboxylic acids, fatty acids, cholesterol esters and sterols, showed higher LOQs, most likely due to mediocre ionization under the electrospray conditions used. However, these undetected metabolites were still kept in our methods, as they may be present in larger quantities and serve as potential biomarkers depending on the design of the animal trial and/or the animal species. Additionally, the use of another ionization technique like atmospheric pressure chemical ionization or of a different mass spectrometer with a different ion source may result in improved LODs and LOQs. To address the challenge of high LODs of compounds naturally present in low concentrations, injecting larger volumes or utilizing different ionization techniques or derivatization could be viable alternatives.
The pattern of plasma metabolites differed between animal species, which highlights differences in metabolic pathways among species. One notable example is the presence of urea: while mammals like pigs, humans, rats and cattle excrete nitrogen primarily as urea, chickens predominantly excrete nitrogenous end products as uric acid due to their unique nitrogen metabolism [44]. Furthermore, chickens were found to have lower concentrations of plasma creatinine compared to mammals. The difference arises because birds excrete creatine without its conversion to creatinine, leading to reduced plasma creatinine levels [45]. In general, our compilation of plasma metabolite concentrations serves as a valuable reference for the validation and quantification of metabolites in plasma from diverse animal sources.

2.2.6. Comparison to Other Large-Scale Targeted Metabolomics Methods

Given the complexity of porcine plasma, one analytical method is insufficient to cover the entire metabolome. Using one single LC technique for targeted large-scale metabolomics studies results in poor peak shapes and the loss of some compounds [13]. If only RP is used, additional sample preparation steps are required to further improve the metabolomic coverage, i.e., the derivatization of polar compound classes like carboxylic acids [46,47], amino acids, biogenic amines [34,46]. This in turn complicates the workflow and compromises the reproducibility. To enhance the metabolic coverage, we developed and combined two complementary LC-MS/MS methods: one RP and one HILIC method. Fast polarity switching and sSRM were employed to accelerate the methods and increase the throughput for multi-analyte analysis. The sample preparation was optimized as a simple “dilute and shoot” approach.
Many studies or commercially available metabolomics kits acquire only one SRM transition per analyte, e.g., [13,27,28,29]. This approach carries the risk of interference from a confounding isobaric compound that produces isobaric fragment ions. To counteract the limitation, high-quality targeted metabolomics assays demand at least two transitions for accurate identification and quantitation [48]. In our methods, the two best transitions for each metabolite were carefully selected to minimize the risk of false positives. Retention times, as well as ion ratios (qualifier to quantifier), are given in Table S4 for further confirmation.
Notably, some studies either omitted essential validation parameters, such as recovery data [13,25,26,28], LOD or LOQ [13,26,29], or reported calibration curves with poor correlation coefficients [27]. Without RA data, analyte losses cannot be assessed and, if required, corrected, which affects the accurate quantification of metabolites. LODs and LOQs are essential to assess a method’s capacity to detect and precisely quantify low levels of analytes, while poor linearity remains a primary source of quantification errors [49]. In contrast, our methods underwent a comprehensive validation protocol, demonstrating minimal carryover, excellent linearity, good repeatability and high accuracy.
Additionally, our methods are more affordable and more robust compared to commercial metabolomic kits, which mostly rely on the derivatization of amino acids, amino acid-related compounds and biogenic amines. As our methods include only transitions of compounds available as reference standards to ensure correct quantification, they cover a smaller metabolite number compared to other large-scale plasma metabolomics studies [13,26,27]. Expanding the metabolic coverage by introducing a third LC method, such as anion-exchange chromatography, could improve the detection of gut-related carboxylic acids and phosphate-containing metabolites like sugar phosphate or nucleotides, but this is out of the scope of the presented study.

3. Materials and Methods

3.1. Chemicals

ACN (LC-MS grade) was purchased from VWR International GmbH (Vienna, Austria) and both IPA (LC-MS grade) and formic acid (FA) (MS grade) from Honeywell (Vienna, Austria). Acetone was obtained from Sigma-Aldrich (Vienna, Austria), while heptane was purchased from J.T. Baker (Deventer, The Netherlands). Aqueous ammonium hydroxide solution (25%, MS grade) and MeOH (LC-MS grade) were procured from Merck (Darmstadt, Germany). LC-MS-grade water was produced using a Arium® Pro water purification system (Sartorius, Göttingen, Germany).
A detailed list providing CAS numbers, sum formulas, SMILES codes, exact masses, log p values and providers of the compounds used in this study is provided in Table S5. Stock solutions were prepared mainly at a concentration of 1000 mg/L and stored at −80 °C.

3.2. Plasma Samples

Porcine plasma was obtained from the pig facility (Vetfarm) of the University of Veterinary Medicine Vienna. The samples were collected from 48 weaned pigs (Large White × Piétrain, 24 males and 24 females) on day 14 after weaning (on days 39 to 42 of life). Prior to blood sampling, the piglets were anesthetized into the ear vein with azaperone (Stresnil 40 mg/mL, Elanco Tiergesundheit AG, Basel, Switzerland) and ketamine (Narketan 100 mg/mL, Vetoquinol Österreich GmbH, Vienna, Austria). Blood samples were collected by cardiac puncture into plasma tubes (Vacuette® Röhrchen K3E K3EDTA; Greiner Bio-One International GmbH, Kremsmünster, Austria), which were then inverted and stored on ice until centrifugation at 3000× g for 20 min at 4 °C (centrifuge 5810 R, Eppendorf, Hamburg, Germany). The obtained plasma was aliquoted and stored at −80 °C until analysis.
A QC plasma sample for method validation was prepared by pooling equal aliquots of all 48 plasma samples. Human, rat, bovine and chicken plasma was purchased from Sigma-Aldrich (Vienna, Austria) as lyophilized powder and reconstituted in the indicated volume of purified water. In addition, SRM 1950–Metabolites in Frozen Human Plasma, originating from the National Institute of Standards and Technology (NIST, Portland, OR, USA), was obtained as a certified reference material to test the method trueness.

3.3. Work-Up of Plasma Samples

Sample preparation consisted of protein precipitation by mixing 200 µL of plasma with 800 µL of cold (4 °C) IPA and centrifuging at 14,350× g and 4 °C for 10 min. The supernatants were further diluted 1 + 19 (v + v) with IPA/water (80/20, v/v). Both samples (1:5 and 1:100 dilution in total) were either measured directly or stored at −80 °C before LC-MS/MS measurement.

3.4. Preparation of Calibration Standards

Three sets of standards were prepared to establish calibration curves. For all standard sets, standard mixes were prepared by dilution from the original 1000 mg/L solutions. For all three mixes, 7 concentration levels were prepared that covered up to three orders of magnitude for each individual compound. The actual concentrations of the individual compounds were selected based on the LLOQ, ULOQ (see Table S1), and, following preliminary measurements of porcine plasma samples, the expected concentration range in porcine plasma. For HILIC compounds, the lowest calibration level was 0.03 µg/L, while the highest calibration level was 25,000 µg/L. Due to the limited solubility of some RP compounds, two sets of standards had to be prepared for RP measurements. Set RPAqu contained polar to mid-polar compounds, while set RPHep contained mid-polar to nonpolar substances. The sets used are indicated in Table S1. The lowest calibration level in RP mixes was 0.1 µg/L, while the highest level was 50,000 µg/L.

3.5. LC-MS/MS Analysis

LC-MS/MS analyses were carried out on an Agilent 1290 Infinity II series UHPLC system (Waldbronn, Germany) coupled to a 6500 triple-quadrupole mass spectrometer equipped with an IonDrive Turbo V® source (Sciex, Foster City, CA, USA).
RP chromatography was carried out on a Kinetex C18 column (150 × 2.1 mm, 2.6 μm, Phenomenex, Aschaffenburg, Germany). The RP column was protected by a SecurityGuard ULTRA pre-column of the same stationary phase (Phenomenex). For RP, mobile phase A consisted of water modified with 1 mM ammonia and 0.1% FA (v/v), while mobile phase B consisted of 47.5/47.5/5 IPA/ACN/water (v/v/v), also modified with 1 mM ammonia and 0.1% FA (v/v). The gradient elution started with 5% solvent B from 0.0 to 0.5 min, continued with a linear increase to 100% B between 0.5 and 8.0 min and an isocratic wash period at 100% B between 8.0 and 19.5 min and ended with re-equilibration at 5% B from 19.5 to 23.0 min. The flow rate was 400 µL/min, except during the wash phase of 100% B between 9.5 and 19.5 min, where the flow rate was increased to 800 µL/min to shorten the wash-out period. The injection volume was 1 µL, and separations were carried out at 50 °C for RP.
For HILIC measurements, a Kinetex HILIC column (150 × 2.1 mm, 2.6 μm, Phenomenex) protected by a SecurityGuard ULTRA pre-column of the same stationary phase was used. For HILIC, mobile phase A consisted of 95/5 ACN/water (v/v) modified with 1 mM ammonia and 0.1% FA (v/v), while mobile phase B was pure water modified with 1 mM ammonia and 0.1% FA (v/v). The gradient for the HILIC method was as follows: 0.0–0.5 min: 100% A, 0.5–6.5 min: linear decrease to 63.2% A (60% ACN), 6.5–6.8 min: linear decrease to 42.1% A, 6.8–12.0 min: 42.1% A (40% ACN), 12.0 to 12.8 min: linear increase to 100% A, 12.8–16.5 min: 100% A. The flow rate was 400 µL/min, except during the re-equilibration phase of 100% A between 13.3 and 15.8 min, where the flow rate was increased to 1000 µL/min to accelerate column re-equilibration. The injection volume was 1 µL, and separations were carried out at 40 °C.
Mass spectrometric detection was conducted in sSRM mode using fast polarity switching. The electrospray ionization parameters were electrospray voltage 4500 V for positive mode and −4500 V for negative mode, source temperature 450 °C, curtain gas 30 psi and ion source gases 1 and 2 at 50 psi each. The settling time for polarity switching was 15 ms. The target cycle time was 1000 ms, and the pause time was 5 ms. The detection window width was mainly 30 s in both polarity modes. sSRM transitions, retention times, detection windows, calculated dwell times and ion ratios are shown in Table S4, the data for which can be directly copied into SciexOS (version 3.0.0.3339, Sciex, Foster City, CA, USA).

3.6. Data Processing

SciexOS was used for peak integration and to establish linear or quadratic 1/x weighted calibration curves. Peak areas were used for quantification. Instrument drift was corrected by using bracket calibration curves or by normalizing with QC samples throughout the sequence. Statistical analysis and plot generation were performed in MS Excel 2019 (Microsoft, Redmond, WA, USA) and in R 4.2.2 (Bell Laboratories, Murray Hill, NJ, USA).

3.7. Method Validation

3.7.1. Spiking Experiments

A preliminary experiment was carried out with 2 mg/L metabolite mixtures in order to determine (a) the potential losses of metabolites during evaporation to dryness (e.g., stability, volatility) and (b) whether compounds could be completely dissolved after evaporation. Three sets of standards (HILIC, RPAqu and RPHep) were individually evaporated under a stream of nitrogen and reconstituted with various solvents. The tested solvents were ACN/water (80/20, v/v), MeOH, ACN, acetone/water (80/20, v/v), ACN/IPA (50/50, v/v), IPA/water (80/20, v/v), acetone/IPA (50/50, v/v) and pure IPA. The reconstituted standard mixtures were then analyzed together with the original standard mixtures as controls. Evaporation recovery was calculated by dividing the peak area of each analyte in the reconstituted samples by that in the untreated control.

3.7.2. Limits of Detection and Quantification, Linearity, Ion Ratios and Carryover Rate

LODs and LLOQs were established according to the EURACHEM guide [50]. As such, calibration levels of 0, 0.0001, 0.0003, 0.001, 0.003, 0.01, 0.03, 0.1, 0.3, 1, 3, 10, 25 and 50 mg/L were prepared for all 3 sets of standards in neat solvents. The concentration of the lowest calibration standard showing a signal-to-noise ratio of at least 3 was set as the LOD, while a signal-to-noise ratio of at least 10 was required for the LLOQ. Due to unavoidable carryover at very low concentrations, several compounds showed peaks that did not increase in height at the next higher concentration level. In those cases, the LOD could not be determined, and the LLOQ was set to the lowest level at which the peak areas significantly increased and the signal-to-noise ratio was at least 10. The ULOQ was determined by visual inspection of the calibration curve. An r2 value greater than 0.995 was mandatory to determine linearity. Additionally, the determined ULOQ value needed to be within 95–105% of the actual concentration. LODs, LLOQs and ULOQs are provided in Supplementary Table S1. Ion ratios were obtained by averaging the peak area ratios of the qualifier and quantifier transition at three representative calibration levels within the linear range.
Analyte carryover was determined by injecting a pure solvent blank after each 1 mg/L of standard. For compounds with LLOQs higher than 1 mg/L, the peak area of the highest validation level was compared with that of the blank measured directly afterward. The carryover percentage was calculated as indicated in Equation (1):
Carryover [%] = Area (blank)/Area (standard) × 100

3.7.3. Determination of Apparent Recoveries and Repeatabilities in the Matrix

Apparent recoveries were evaluated using the standard addition method. As such, native concentrations of metabolites in pooled QC plasma were determined before spiking. Concentrations corresponding to 0.5, 1, 1.5, 2, 2.5 and 3 times the predetermined analyte concentration were spiked into the QC plasma. For metabolites < LOD in the QC sample, a concentration corresponding to 12.5 times the LLOQ (in pure solvents) was designated as the lowest spiking level of plasma. Additionally, spiking levels of 25, 37.5, 50, 62.5 and 75× LLOQ were prepared. The standard addition protocol consisted of drying down mixed standard solutions under a gentle stream of nitrogen, reconstituting the residue into 200 µL of plasma and 800 µL of cold IPA (−20 °C), vigorous vortexing, transfer to Eppendorf reaction tubes and centrifugation at 14,350× g at 4 °C for 10 min. One aliquot of the supernatants was further diluted 1:20 (v:v). For comparison, equal volumes of spiking solutions were processed in the same manner, except that 1 mL of IPA/water (80/20, v/v) was used for reconstitution of analytes.
Spiking was conducted in triplicate for each level. Two dilutions of plasma (1:5, 1:100, v:v) were prepared, and samples were measured by HILIC-MS/MS and RP-MS/MS as described above. For metabolites showing good linearity (r2 > 0.995), average RAs were calculated as the ratios of the slopes of the regression lines from the spiked samples and the standards in neat solvents. For analytes showing quadratic responses, RAs were calculated as indicated in Equation (2). The metabolite concentrations in the spiked samples (csp) were calculated based on quadratic calibration curves established with pure solvent standards. Afterward, the concentration determined in the non-spiked sample (cNS) was subtracted, and the resulting difference was divided by the spiking concentration of the analyte (cA) at the respective level. The mean RA was calculated by averaging the RAs obtained from six levels prepared and measured in triplicates (n = 18).
RA [%] = (csp − cNS)/cA × 100
To determine the repeatability, RAs were calculated at each spiking level using Equation (2). The repeatability was represented by the RSDs of the recoveries determined at six spiking levels in triplicate (n = 18).

3.8. Applicability to Plasma Samples from Different Animal Species

In order to test the applicability of the method, human, rat, bovine and chicken plasma samples were worked up as described above and analyzed. To minimize instrument bias, samples were measured in duplicate in a randomized sequence. The same two dilutions (1:5 and 1:100 in total) were measured to obtain the metabolite concentrations in animal samples. Recovery correction was applied only when the RA values obtained from porcine plasma were below 70% or higher than 130%. In this case, measured analyte concentrations were corrected by the RAs according to Equation (3).
c corrected = c calculated from calibration curve × 100/RA [%]
To assess the method trueness, the metabolite concentrations in human plasma NIST SRM 1950 were compared to the certified values.

4. Conclusions

In this paper, we present the first fully validated, large-scale, targeted metabolomics method for the accurate and reliable quantification of porcine plasma metabolites. Two complementary LC techniques, namely RP and HILIC, were coupled to tandem mass spectrometry to cover 235 porcine plasma metabolites from 17 compound classes. Each analyte was determined using two SRM transitions to guarantee accurate identification. Fast polarity switching and sSRM were utilized to accelerate the methods and increase the throughput capacity, with a total analysis time of 39.5 min per sample. Sample preparation was simplified as a “dilute and shoot” approach, avoiding the derivatization of polar compound classes. The fully validated methods demonstrated minimal carryover, excellent linearity and repeatability and high accuracy. The inclusion of recovery data ensures robust quantification by correcting matrix effects and potential analyte losses during sample preparation [41]. The applicability test provides valuable concentration references for the validation and quantification of metabolites in plasma from other animal species. Future studies should focus on validating these methods in plasma from other animal species and/or other biological matrices. In conclusion, this study presents two comprehensively validated LC-MS/MS methods for the accurate quantification of 235 porcine plasma metabolites, setting a new standard for large-scale targeted metabolomics in porcine plasma.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/molecules30030706/s1, Table S1: Sensitivity, linearity and carryover of metabolites; Table S2: Apparent recoveries and repeatability. Compounds with RAs calculated based on quadratic or single-level calibration curves are marked with an asterisk; Table S3: Metabolite concentrations in plasma from diverse species; Table S4: Optimized sSRM transitions along with retention times and ion ratios; Table S5: Library of metabolites containing CAS numbers, sum formulas, SMILES codes, exact masses, log p values and providers.

Author Contributions

Conceptualization, F.B., H.E.S.-Z. and K.X.; methodology, F.B., H.E.S.-Z. and K.X.; software, K.X.; validation, F.B., H.E.S.-Z. and K.X.; formal analysis, K.X.; investigation, K.X.; resources, B.U.M.-Z. and F.B.; data curation, K.X.; writing—original draft preparation, K.X.; writing—review and editing, B.U.M.-Z., F.B., H.E.S.-Z. and K.X.; visualization, K.X.; supervision, F.B. and H.E.S.-Z.; project administration, F.B. and H.E.S.-Z.; funding acquisition, F.B. and H.E.S.-Z. All authors have read and agreed to the published version of the manuscript.

Funding

The research was funded by the Austrian Federal Ministry for Digital and Economic Affairs; the Austrian National Foundation for Research, Technology and Development; as well as BIOMIN Holding GmbH, which is part of dsm-firmenich, through support of the Christian Doppler Laboratory for Innovative Gut Health Concepts of Livestock. The LC-MS/MS equipment was provided by the InfraCore project, funded by the Lower Austrian Government and co-funded by the European Regional Development Fund. Additional financial support was received from the Austrian Competence Centre for Feed and Food Quality, Safety and Innovation (FFoQSI).

Institutional Review Board Statement

All procedures involving animal handling and treatment were approved by the institutional ethics committee of the University of Veterinary Medicine Vienna and the national authority according to the Law for Animal Experiments, Tierversuchsgesetz (BMWFW 2022-0855.741).

Informed Consent Statement

Not applicable.

Data Availability Statement

Data are contained within the article and Supplementary Materials. Selected MS instrument files are available at https://doi.org/10.5281/zenodo.14652873, accessed on 15 January 2025. Additional inquiries can be forwarded to the corresponding author.

Acknowledgments

The authors acknowledge the very appreciated technical support of Marco Reiter.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  1. Fiehn, O. Combining genomics, metabolome analysis, and biochemical modelling to understand metabolic networks. Comp. Funct. Genom. 2001, 2, 155–168. [Google Scholar] [CrossRef] [PubMed]
  2. Johnson, C.H.; Ivanisevic, J.; Siuzdak, G. Metabolomics: Beyond biomarkers and towards mechanisms. Nat. Rev. Mol. Cell. Biol. 2016, 17, 451–459. [Google Scholar] [CrossRef] [PubMed]
  3. Zhang, A.; Sun, H.; Yan, G.; Wang, P.; Wang, X. Mass spectrometry-based metabolomics: Applications to biomarker and metabolic pathway research. Biomed. Chromatogr. 2016, 30, 7–12. [Google Scholar] [CrossRef]
  4. Roberts, L.D.; Souza, A.L.; Gerszten, R.E.; Clish, C.B. Targeted Metabolomics. Curr. Protoc. Mol. Biol. 2012, 98, 30.2.1–30.2.24. [Google Scholar] [CrossRef]
  5. Ferrario, M.; Cambiaghi, A.; Brunelli, L.; Giordano, S.; Caironi, P.; Guatteri, L.; Raimondi, F.; Gattinoni, L.; Latini, R.; Masson, S.; et al. Mortality prediction in patients with severe septic shock: A pilot study using a target metabolomics approach. Sci. Rep. 2016, 6, 20391. [Google Scholar] [CrossRef]
  6. Dankers, J.; van den Elshout, J.; Ahr, G.; Brendel, E.; van der Heiden, C. Determination of nifedipine in human plasma by flow-injection tandem mass spectrometry. J. Chromatogr. B 1998, 710, 115–120. [Google Scholar] [CrossRef]
  7. Boscaro, F.; Pieraccini, G.; Marca, G.l.; Bartolucci, G.; Luceri, C.; Luceri, F.; Moneti, G. Rapid quantitation of globotriaosylceramide in human plasma and urine: A potential application for monitoring enzyme replacement therapy in Anderson-Fabry disease. Rapid Commun. Mass Spectrom. 2002, 16, 1507–1514. [Google Scholar] [CrossRef]
  8. Griffiths, W.J.; Koal, T.; Wang, Y.; Kohl, M.; Enot, D.P.; Deigner, H.P. Targeted Metabolomics for Biomarker Discovery. Angew. Chem. Int. Ed. 2010, 49, 5426–5445. [Google Scholar] [CrossRef] [PubMed]
  9. Nanita, S.C.; Kaldon, L.G. Emerging flow injection mass spectrometry methods for high-throughput quantitative analysis. Anal. Bioanal. Chem. 2016, 408, 23–33. [Google Scholar] [CrossRef] [PubMed]
  10. Casas-Ferreira, A.M.; del Nogal-Sánchez, M.; Rodríguez-Gonzalo, E.; Moreno-Cordero, B.; Pérez-Pavón, J.L. Determination of leucine and isoleucine/allo-isoleucine by electrospray ionization-tandem mass spectrometry and partial least square regression: Application to saliva samples. Talanta 2020, 216, 120811. [Google Scholar] [CrossRef]
  11. Buszewski, B.; Noga, S. Hydrophilic interaction liquid chromatography (HILIC)—A powerful separation technique. Anal. Bioanal. Chem. 2012, 402, 231–247. [Google Scholar] [CrossRef]
  12. Tolstikov, V.V.; Fiehn, O. Analysis of highly polar compounds of plant origin: Combination of hydrophilic interaction chromatography and electrospray ion trap mass spectrometry. Anal. Biochem. 2002, 301, 298–307. [Google Scholar] [CrossRef]
  13. Li, K.; Naviaux, J.C.; Bright, A.T.; Wang, L.; Naviaux, R.K. A robust, single-injection method for targeted, broad-spectrum plasma metabolomics. Metabolomics 2017, 13, 122. [Google Scholar] [CrossRef]
  14. Floros, D.J.; Xu, K.; Berthiller, F.; Schwartz-Zimmermann, H. Comparison of chromatographic conditions for the targeted tandem mass spectrometric determination of 354 mammalian metabolites. J. Chromatogr. A 2023, 1697, 463985. [Google Scholar] [CrossRef]
  15. Ovbude, S.T.; Sharmeen, S.; Kyei, I.; Olupathage, H.; Jones, J.; Bell, R.J.; Powers, R.; Hage, D.S. Applications of chromatographic methods in metabolomics: A review. J. Chromatogr. B 2024, 1239, 124124. [Google Scholar] [CrossRef]
  16. Goldansaz, S.A.; Guo, A.C.; Sajed, T.; Steele, M.A.; Plastow, G.S.; Wishart, D.S. Livestock metabolomics and the livestock metabolome: A systematic review. PLoS ONE 2017, 12, e0177675. [Google Scholar] [CrossRef] [PubMed]
  17. Zhang, A.; Sun, H.; Wang, P.; Han, Y.; Wang, X. Recent and potential developments of biofluid analyses in metabolomics. J. Proteom. 2012, 75, 1079–1088. [Google Scholar] [CrossRef]
  18. Kell, D.B.; Brown, M.; Davey, H.M.; Dunn, W.B.; Spasic, I.; Oliver, S.G. Metabolic footprinting and systems biology: The medium is the message. Nat. Rev. Microbiol. 2005, 3, 557–565. [Google Scholar] [CrossRef] [PubMed]
  19. Manzoni, C.; Kia, D.A.; Vandrovcova, J.; Hardy, J.; Wood, N.W.; Lewis, P.A.; Ferrari, R. Genome, transcriptome and proteome: The rise of omics data and their integration in biomedical sciences. Brief. Bioinform. 2018, 19, 286–302. [Google Scholar] [CrossRef]
  20. Bovo, S.; Schiavo, G.; Galimberti, G.; Fanelli, F.; Bertolini, F.; Dall’Olio, S.; Pagotto, U.; Fontanesi, L. Comparative targeted metabolomic profiles of porcine plasma and serum. Animal 2023, 17, 101029. [Google Scholar] [CrossRef]
  21. Kiseleva, O.; Kurbatov, I.; Ilgisonis, E.; Poverennaya, E. Defining Blood Plasma and Serum Metabolome by GC-MS. Metabolites 2022, 12, 15. [Google Scholar] [CrossRef] [PubMed]
  22. García-Sevillano, M.; Contreras-Acuña, M.; García-Barrera, T.; Navarro, F.; Gómez-Ariza, J. Metabolomic study in plasma, liver and kidney of mice exposed to inorganic arsenic based on mass spectrometry. Anal. Bioanal. Chem. 2014, 406, 1455–1469. [Google Scholar] [CrossRef] [PubMed]
  23. Psychogios, N.; Hau, D.D.; Peng, J.; Guo, A.C.; Mandal, R.; Bouatra, S.; Sinelnikov, I.; Krishnamurthy, R.; Eisner, R.; Gautam, B. The human serum metabolome. PLoS ONE 2011, 6, e16957. [Google Scholar] [CrossRef] [PubMed]
  24. Dunn, W.B.; Bailey, N.J.C.; Johnson, H.E. Measuring the metabolome: Current analytical technologies. Analyst 2005, 130, 606–625. [Google Scholar] [CrossRef]
  25. Cao, G.; Song, Z.; Hong, Y.; Yang, Z.; Song, Y.; Chen, Z.; Chen, Z.; Cai, Z. Large-scale targeted metabolomics method for metabolite profiling of human samples. Anal. Chim. Acta 2020, 1125, 144–151. [Google Scholar] [CrossRef] [PubMed]
  26. Lee, H.-J.; Kremer, D.M.; Sajjakulnukit, P.; Zhang, L.; Lyssiotis, C.A. A large-scale analysis of targeted metabolomics data from heterogeneous biological samples provides insights into metabolite dynamics. Metabolomics 2019, 15, 103. [Google Scholar] [CrossRef]
  27. McMillen, T.S.; Leslie, A.; Chisholm, K.; Penny, S.; Gallant, J.; Cohen, A.; Drucker, A.; Fawcett, J.P.; Pinto, D.M. A large-scale, targeted metabolomics method for the analysis and quantification of metabolites in human plasma via liquid chromatography-mass spectrometry. Anal. Chim. Acta 2023, 1279, 341791. [Google Scholar] [CrossRef]
  28. Wei, R.; Li, G.; Seymour, A.B. High-Throughput and Multiplexed LC/MS/MRM Method for Targeted Metabolomics. Anal. Chem. 2010, 82, 5527–5533. [Google Scholar] [CrossRef]
  29. Nandania, J.; Peddinti, G.; Pessia, A.; Kokkonen, M.; Velagapudi, V. Validation and Automation of a High-Throughput Multitargeted Method for Semiquantification of Endogenous Metabolites from Different Biological Matrices Using Tandem Mass Spectrometry. Metabolites 2018, 8, 44. [Google Scholar] [CrossRef] [PubMed]
  30. Luise, D.; Bovo, S.; Bosi, P.; Fanelli, F.; Pagotto, U.; Galimberti, G.; Mazzoni, G.; Dall’Olio, S.; Fontanesi, L. Targeted metabolomic profiles of piglet plasma reveal physiological changes over the suckling period. Livest. Sci. 2020, 231, 103890. [Google Scholar] [CrossRef]
  31. Solberg, R.; Enot, D.; Deigner, H.-P.; Koal, T.; Scholl-Bürgi, S.; Saugstad, O.D.; Keller, M. Metabolomic analyses of plasma reveals new insights into asphyxia and resuscitation in pigs. PLoS ONE 2010, 5, e9606. [Google Scholar] [CrossRef] [PubMed]
  32. Newman, M.A.; Zebeli, Q.; Eberspächer, E.; Grüll, D.; Molnar, T.; Metzler-Zebeli, B.U. Transglycosylated Starch Improves Insulin Response and Alters Lipid and Amino Acid Metabolome in a Growing Pig Model. Nutrients 2017, 9, 291. [Google Scholar] [CrossRef] [PubMed]
  33. Metzler-Zebeli, B.U.; Ertl, R.; Klein, D.; Zebeli, Q. Explorative study of metabolic adaptations to various dietary calcium intakes and cereal sources on serum metabolome and hepatic gene expression in juvenile pigs. Metabolomics 2015, 11, 545–558. [Google Scholar] [CrossRef]
  34. Metzler-Zebeli, B.U.; Lerch, F.; Yosi, F.; Vötterl, J.C.; Koger, S.; Aigensberger, M.; Rennhofer, P.M.; Berthiller, F.; Schwartz-Zimmermann, H.E. Creep Feeding and Weaning Influence the Postnatal Evolution of the Plasma Metabolome in Neonatal Piglets. Metabolites 2023, 13, 214. [Google Scholar] [CrossRef]
  35. Metzler-Zebeli, B.U.; Eberspächer, E.; Grüll, D.; Kowalczyk, L.; Molnar, T.; Zebeli, Q. Enzymatically modified starch ameliorates postprandial serum triglycerides and lipid metabolome in growing pigs. PLoS ONE 2015, 10, e0130553. [Google Scholar] [CrossRef]
  36. Zhou, J.; Yin, Y. Strategies for large-scale targeted metabolomics quantification by liquid chromatography-mass spectrometry. Analyst 2016, 141, 6362–6373. [Google Scholar] [CrossRef]
  37. Guo, B.; Chen, B.; Liu, A.; Zhu, W.; Yao, S. Liquid chromatography-mass spectrometric multiple reaction monitoring-based strategies for expanding targeted profiling towards quantitative metabolomics. Curr. Drug. Metab. 2012, 13, 1226–1243. [Google Scholar] [CrossRef] [PubMed]
  38. Yuan, M.; Breitkopf, S.B.; Yang, X.; Asara, J.M. A positive/negative ion–switching, targeted mass spectrometry–based metabolomics platform for bodily fluids, cells, and fresh and fixed tissue. Nat. Protoc. 2012, 7, 872–881. [Google Scholar] [CrossRef] [PubMed]
  39. Jogpethe, A.; Jadav, T.; Rajput, N.; Kumar Sahu, A.; Tekade, R.K.; Sengupta, P. Critical strategies to pinpoint carryover problems in liquid chromatography-mass spectrometry: A systematic direction for their origin identification and mitigation. Microchem. J. 2022, 179, 107464. [Google Scholar] [CrossRef]
  40. Alseekh, S.; Aharoni, A.; Brotman, Y.; Contrepois, K.; D’Auria, J.; Ewald, J.C.; Ewald, J.; Fraser, P.D.; Giavalisco, P.; Hall, R.D.; et al. Mass spectrometry-based metabolomics: A guide for annotation, quantification and best reporting practices. Nat. Methods 2021, 18, 747–756. [Google Scholar] [CrossRef] [PubMed]
  41. Kumar, D.; Gautam, N.; Alnouti, Y. Analyte recovery in LC-MS/MS bioanalysis: An old issue revisited. Anal. Chim. Acta 2022, 1198, 339512. [Google Scholar] [CrossRef]
  42. Yang, Y.; Cruickshank, C.; Armstrong, M.; Mahaffey, S.; Reisdorph, R.; Reisdorph, N. New sample preparation approach for mass spectrometry-based profiling of plasma results in improved coverage of metabolome. J. Chromatogr. A 2013, 1300, 217–226. [Google Scholar] [CrossRef] [PubMed]
  43. Kruve, A.; Leito, I.; Herodes, K. Combating matrix effects in LC/ESI/MS: The extrapolative dilution approach. Anal. Chim. Acta 2009, 651, 75–80. [Google Scholar] [CrossRef]
  44. Wright, P.A. Nitrogen excretion: Three end products, many physiological roles. J. Exp. Biol. 1995, 198, 273–281. [Google Scholar] [CrossRef]
  45. Sturkie, P. Kidneys, extrarenal salt excretion, and urine. In Avian Physiology; Springer: Berlin/Heidelberg, Germany, 1986; pp. 359–382. [Google Scholar]
  46. Zheng, J.; Zhang, L.; Johnson, M.; Mandal, R.; Wishart, D.S. Comprehensive targeted metabolomic assay for urine analysis. Anal. Chem. 2020, 92, 10627–10634. [Google Scholar] [CrossRef]
  47. Schwartz-Zimmermann, H.E.; Hündler, M.; Reiterer, N.; Ricci, S.; Rivera-Chacon, R.; Castillo-Lopez, E.; Zebeli, Q.; Berthiller, F. Comparison of LC-MS-based methods for the determination of carboxylic acids in animal matrices. Anal. Bioanal. Chem. 2024, 416, 1199–1215. [Google Scholar] [CrossRef]
  48. Lehotay, S.J.; Sapozhnikova, Y.; Mol, H.G. Current issues involving screening and identification of chemical contaminants in foods by mass spectrometry. TrAC Trends Anal. Chem. 2015, 69, 62–75. [Google Scholar] [CrossRef]
  49. Meesters, R.; Voswinkel, S. Bioanalytical Method Development and Validation: From the USFDA 2001 to the USFDA 2018 Guidance for Industry. J. Appl. Bioanal. 2018, 4, 67–73. [Google Scholar] [CrossRef]
  50. Peris-Vicente, J.; Esteve-Romero, J.; Carda-Broch, S. Validation of analytical methods based on chromatographic techniques: An overview. Anal. Sep. Sci. 2015, 5, 1757–1808. [Google Scholar] [CrossRef]
Figure 1. Properties of metabolites covered in the targeted metabolomics assay. (a) Relative proportion of compound classes; (b) calculated octanol–water partition coefficients (log p) of 235 metabolites.
Figure 1. Properties of metabolites covered in the targeted metabolomics assay. (a) Relative proportion of compound classes; (b) calculated octanol–water partition coefficients (log p) of 235 metabolites.
Molecules 30 00706 g001
Figure 2. Extracted ion chromatogram of the QC plasma sample measured by HILIC-MS/MS. The peak intensity of betaine (peak 34) was scaled (divided by 3) for better visualization. 1: 2′-deoxythymidine; 2: uridine; 3: xanthine; 4: urea; 5: hypoxanthine; 6: inosine; 7: guanine; 8: cytidine; 9: 2′-deoxycytidine; 10: taurine; 11: cytosine; 12: creatinine; 13: tryptophan; 14: phenylalanine; 15: pyroglutamic acid; 16: tyrosine; 17: leucine; 18: pyrrolidine; 19: methionine; 20: valine; 21: N-methylalanine; 22: threonine; 23: alanine; 24: glutamic acid; 25: trans-4-hydroxyproline; 26: proline; 27: trimethylamine-N-oxide; 28: glutamine; 29: creatine; 30: propionylcarnitine; 31: acetylcarnitine; 32: citrulline; 33: trigonelline; 34: betaine; 35: carnitine; 36: stachydrine; 37: arginine; 38: histamine; 39: ornithine; 40: lysine; 41: N,N-dimethylarginine; 42: 3-methylhistidine; 43: 1-methylhistidine.
Figure 2. Extracted ion chromatogram of the QC plasma sample measured by HILIC-MS/MS. The peak intensity of betaine (peak 34) was scaled (divided by 3) for better visualization. 1: 2′-deoxythymidine; 2: uridine; 3: xanthine; 4: urea; 5: hypoxanthine; 6: inosine; 7: guanine; 8: cytidine; 9: 2′-deoxycytidine; 10: taurine; 11: cytosine; 12: creatinine; 13: tryptophan; 14: phenylalanine; 15: pyroglutamic acid; 16: tyrosine; 17: leucine; 18: pyrrolidine; 19: methionine; 20: valine; 21: N-methylalanine; 22: threonine; 23: alanine; 24: glutamic acid; 25: trans-4-hydroxyproline; 26: proline; 27: trimethylamine-N-oxide; 28: glutamine; 29: creatine; 30: propionylcarnitine; 31: acetylcarnitine; 32: citrulline; 33: trigonelline; 34: betaine; 35: carnitine; 36: stachydrine; 37: arginine; 38: histamine; 39: ornithine; 40: lysine; 41: N,N-dimethylarginine; 42: 3-methylhistidine; 43: 1-methylhistidine.
Molecules 30 00706 g002
Figure 3. Extracted ion chromatogram of the QC plasma sample measured by RP-MS/MS. 1: 3-indoxylsulfate; 2: hippuric acid; 3: phenylacetylglycine; 4: 3-indoleacetic acid; 5: 3-indolepropionic acid; 6: taurohyodeoxycholic acid; 7: glycoursodeoxycholic acid; 8: glycocholic acid; 9: glycochenodeoxycholic acid; 10: chenodeoxycholic acid; 11: 1-myristoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 12: 1-palmitoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 13: 1-oleoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 14: 1-heptadecanoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 15: 1-stearoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 16: linoleic acid; 17: oleic acid; 18: cholesterol; 19: 24α-methylcholesterol; 20: N-palmitoyl-D-sphingosin; 21: 1,2-dipalmitelaidoyl-sn-glycero-3-phosphatidylcholine; 22: 1-palmitoyl-2-myristoyl-sn-glycero-3-phosphatidylcholine; 23: 1-oleoyl-2-stearoyl-sn-glycero-3-phosphatidylcholine; 24: 1-O-hexadecyl-2-arachidonoyl-sn-glycero-3-phosphatidylcholine; 25: C16 sphingomyelin; 26: 1-stearoyl-2-arachidonoyl-sn-glycero-3-phosphatidylcholine; 27: 1,2-distearoyl-sn-glycero-3-phosphatidylcholine; 28: 1,2-dilinoleoyl-3-palmitoylglycerol; 29: 1,2-dielaidoyl-sn-glycero-3-phosphatidylcholine; 30: 1-1(Z)-hexadecenyl-2-palmitoyl-sn-glycero-3-phosphatidylcholine; 31: C16-18:1 PC; 32: 1,3-dioleyl-2-palmitoylglycerol; 33: 1,3-dipalmitoyl-2-oleoylglycerol.
Figure 3. Extracted ion chromatogram of the QC plasma sample measured by RP-MS/MS. 1: 3-indoxylsulfate; 2: hippuric acid; 3: phenylacetylglycine; 4: 3-indoleacetic acid; 5: 3-indolepropionic acid; 6: taurohyodeoxycholic acid; 7: glycoursodeoxycholic acid; 8: glycocholic acid; 9: glycochenodeoxycholic acid; 10: chenodeoxycholic acid; 11: 1-myristoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 12: 1-palmitoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 13: 1-oleoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 14: 1-heptadecanoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 15: 1-stearoyl-2-hydroxy-sn-glycero-3-phosphatidylcholine; 16: linoleic acid; 17: oleic acid; 18: cholesterol; 19: 24α-methylcholesterol; 20: N-palmitoyl-D-sphingosin; 21: 1,2-dipalmitelaidoyl-sn-glycero-3-phosphatidylcholine; 22: 1-palmitoyl-2-myristoyl-sn-glycero-3-phosphatidylcholine; 23: 1-oleoyl-2-stearoyl-sn-glycero-3-phosphatidylcholine; 24: 1-O-hexadecyl-2-arachidonoyl-sn-glycero-3-phosphatidylcholine; 25: C16 sphingomyelin; 26: 1-stearoyl-2-arachidonoyl-sn-glycero-3-phosphatidylcholine; 27: 1,2-distearoyl-sn-glycero-3-phosphatidylcholine; 28: 1,2-dilinoleoyl-3-palmitoylglycerol; 29: 1,2-dielaidoyl-sn-glycero-3-phosphatidylcholine; 30: 1-1(Z)-hexadecenyl-2-palmitoyl-sn-glycero-3-phosphatidylcholine; 31: C16-18:1 PC; 32: 1,3-dioleyl-2-palmitoylglycerol; 33: 1,3-dipalmitoyl-2-oleoylglycerol.
Molecules 30 00706 g003
Figure 4. Dwell time plots. (a) HILIC-MS/MS with a minimal dwell time of 7 ms; (b) RP-MS/MS with a minimal dwell time of 10 ms.
Figure 4. Dwell time plots. (a) HILIC-MS/MS with a minimal dwell time of 7 ms; (b) RP-MS/MS with a minimal dwell time of 10 ms.
Molecules 30 00706 g004
Figure 5. Box plot showing apparent recoveries of metabolite classes for 1:5 dilution.
Figure 5. Box plot showing apparent recoveries of metabolite classes for 1:5 dilution.
Molecules 30 00706 g005
Figure 6. Box plot showing apparent recoveries of metabolite classes for 1:100 dilution.
Figure 6. Box plot showing apparent recoveries of metabolite classes for 1:100 dilution.
Molecules 30 00706 g006
Table 1. Metabolite concentrations in NIST reference plasma and method trueness (concentrations determined by HILIC-MS/MS and RP-MS/MS relative to the NIST reference values).
Table 1. Metabolite concentrations in NIST reference plasma and method trueness (concentrations determined by HILIC-MS/MS and RP-MS/MS relative to the NIST reference values).
Compound Namec in NIST Reference Plasma (µg/L)Measured c in NIST Plasma (µg/L)Trueness (%)
Methionine3325214665
Leucine13,158903869
Proline20,29815,85978
Valine21,31817,05380
Threonine14,21911,91584
Alanine26,72422,44984
Serine10,067921892
Isoleucine7273682394
Tyrosine10,373977094
Phenylalanine8364828799
Lysine20,40023,637116
Arginine14,16823,863168
Creatinine67897672113
Urea234,500224,88896
Cholesterol1,514,000645,65543
Cholecalciferol25
Calciferol1.0
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Xu, K.; Berthiller, F.; Metzler-Zebeli, B.U.; Schwartz-Zimmermann, H.E. Development and Validation of Targeted Metabolomics Methods Using Liquid Chromatography–Tandem Mass Spectrometry (LC-MS/MS) for the Quantification of 235 Plasma Metabolites. Molecules 2025, 30, 706. https://doi.org/10.3390/molecules30030706

AMA Style

Xu K, Berthiller F, Metzler-Zebeli BU, Schwartz-Zimmermann HE. Development and Validation of Targeted Metabolomics Methods Using Liquid Chromatography–Tandem Mass Spectrometry (LC-MS/MS) for the Quantification of 235 Plasma Metabolites. Molecules. 2025; 30(3):706. https://doi.org/10.3390/molecules30030706

Chicago/Turabian Style

Xu, Kangkang, Franz Berthiller, Barbara U. Metzler-Zebeli, and Heidi E. Schwartz-Zimmermann. 2025. "Development and Validation of Targeted Metabolomics Methods Using Liquid Chromatography–Tandem Mass Spectrometry (LC-MS/MS) for the Quantification of 235 Plasma Metabolites" Molecules 30, no. 3: 706. https://doi.org/10.3390/molecules30030706

APA Style

Xu, K., Berthiller, F., Metzler-Zebeli, B. U., & Schwartz-Zimmermann, H. E. (2025). Development and Validation of Targeted Metabolomics Methods Using Liquid Chromatography–Tandem Mass Spectrometry (LC-MS/MS) for the Quantification of 235 Plasma Metabolites. Molecules, 30(3), 706. https://doi.org/10.3390/molecules30030706

Article Metrics

Back to TopTop