Article

The Impact of Excitation Periods on the Outcome of Lock-In Thermography

Faculty of Mechanical Engineering, University of Zilina, Univerzitná 8215/1, 010 26 Žilina, Slovakia
*
Author to whom correspondence should be addressed.
Materials 2023, 16(7), 2763; https://doi.org/10.3390/ma16072763
Submission received: 13 February 2023 / Revised: 17 March 2023 / Accepted: 24 March 2023 / Published: 30 March 2023
(This article belongs to the Special Issue Newer Paradigms in Advanced Materials Characterisation)

Abstract

:
Thermal imaging is a non-destructive test method that uses an external energy source, such as a halogen lamp or flash lamp, to excite the material under test and measure the resulting temperature distribution. One of the important parameters of lock-in thermography is the number of excitation periods, which is used to calculate a phase image that shows defects or inhomogeneities in the material. The results for multiple periods can be averaged, which leads to noise suppression, but the use of a larger number of periods may cause an increase in noise due to unsynchronization of the camera and the external excitation source or may lead to heating and subsequent damage to the sample. The phase image is the most common way of representing the results of lock-in thermography, but amplitude images and complex images can also be obtained. In this study, eight measurements were performed on different samples using a thermal pulse source (flash lamp and halogen lamp) with a period of 120 s. For each sample, five phase images were calculated using different number of periods, preferably one to five periods. The phase image calculated from one period was used as a reference. To determine the effect of the number of excitation periods on the phase image, the reference phase image for one period was compared with the phase images calculated using multiple periods using the structural similarity index (SSIM) and multi-scale SSIM (MS-SSIM).

1. Introduction

Infrared thermography is a powerful and widely used method of non-destructive testing (NDT) and evaluation that utilizes an external energy source to create a temperature difference between areas of a sample that are defective and those that are not, in order to identify any defects. This method is known as active thermography, as opposed to passive thermography, in which defects are identified through naturally occurring thermal contrasts.
One of the key advantages of infrared thermography is that it allows for the detection of defects that may not be visible to the naked eye, such as those found inside a material or component. Additionally, it can be used to detect defects in a wide range of materials and industries, including aerospace, automotive, and electrical engineering. As a non-destructive testing method, Infrared thermography can be used to identify defects without causing any damage to the sample, making it a cost-effective and efficient method of testing.
Lock-in thermography is a popular type of non-destructive testing that uses an external source to induce temperature changes in a material or component, and then uses an infrared camera to measure the resulting thermal response. Lock-in thermography is often used to inspect composite materials such as aerospace structural elements [1,2,3,4,5], wind turbine blade parts [6,7,8], and other advanced composites such as composites containing nanotubes [9,10] as well as sandwich structures [11]. Lock-in thermography is widely used for detecting subsurface defects, such as cracks [12,13], delamination [14,15,16], impact damage [17,18], and corrosion [19,20]. It is considered a highly versatile NDT method because it can be used for a wide range of materials and components such as metal [21], composites [22], wood [23], ceramics [24], plastics [25], and concrete [26]. It can be used in laboratory and in-field settings, and it can be used to inspect both single samples or large areas.

2. Lock-In Method

The lock-in principle is the technique of choice, if signals have to be extracted from statistical noise. Prerequisite to using this technique is that the primary signal, can be periodically pulsed or amplitude-modulated with a certain frequency called “lock-in frequency” [27].
It uses a reference signal, known as a local oscillator that is synchronized with the signal of interest and is used to “lock-in” or extract the small signal from the background noise. This can be done by multiplying the input signal with the reference signal and then applying a low-pass filter to the product. The result is a signal that is proportional to the original small signal, but with much-reduced noise. This method is commonly used in experimental physics, chemistry, and other scientific fields.
The concept of this method is that the response of the system, when stimulated by a signal with a sinusoidal character and lock-in frequency f L , will also be in the form of a sinusoid with the same frequency and with amplitude A anf phase shift φ . This means that the response s(t) can be represented as a function of this sinusoidal character in the time t:
s ( t ) = A s i n   ( 2 π f L t + φ ) = a c o s   ( 2 π f L t ) + b s i n ( 2 π f L t ) ,
a = s i n   ( φ ) ,   b = c o s ( φ ) .
Given that the amplitude of the response s ( t ) is very small, and the signal S ( t ) being measured also contains a much higher level of noise N ( t ) , the relationship between the two can be described as:
S ( t ) = s ( t ) + N ( t ) .
If S ( t ) is expressed by Fourier’s series than:
S ( t ) = A 0 2 + i = 1 A i s i n   ( 2 π i f L t + φ i ) = = A 1 s i n ( 2 π f L + φ 1 ) + [ A 0 2 + i = 2 A i s i n ( 2 π i f L t + φ i ) ] = A 1 ( 2 π f L t + φ 1 ) + N ( t ) .  
If A = A 1 and φ = φ 1 , then the coefficients a ,   b in Equation (2) can be estimated by using the relationships of discrete Fourier transformation:
a = k = 1 n F k   c o s ( 2 π k n f L ) ,
b = k = 1 n F k   s i n ( 2 π k n f L ) ,
when n is number of frames recorded by infrared camera in one period, F k is the k-th frame recorded in time t k with sample rate of IR camera. Then the phase shift of the response s ( t ) is:
φ = a t a n a b .
The phase image is formed by the phase shift values for each single pixel of IR camera.

3. Structural Similarity Index Measure (SSIM)

The Structural Similarity Index (SSIM) is a tool for measuring the similarity between two images. It was first proposed by Wang et al. in 2004 [28], to quantify the perceived quality of images and videos. The SSIM metric is based on a comparison of the structural information in the image, as well as the luminance and contrast of the image. It is calculated as a value between −1 and 1, with values closer to 1 indicating greater similarity between the two images. The SSIM metric has been widely used in image and video processing, and it has been found to correlate well with human perception of image quality.
It was designed to improve upon traditional methods for measuring the quality of images, such as the peak signal-to-noise ratio (PSNR) and mean squared error (MSE), which rely solely on the absolute error as a metric. The SSIM was developed to address this limitation by considering three independent factors, luminance, contrast, and structure, in its calculation, providing a more comprehensive measure of image similarity.
The general form of the Structural Similarity (SSIM) Index between x and y signals is defined as [28]:
S S I M ( x ,   y ) = [ l ( x ,   y ) ]   α · [ c ( x ,   y ) ] β   · [ s ( x ,   y ) ] γ
l ( x ,   y ) = 2 μ x μ y + C 1 μ x 2 + μ y 2 + C 1 ,
c ( x ,   y ) = 2 σ x σ y + C 2 σ x 2 + σ y 2 + C 2 ,
s ( x ,   y ) = σ x y + C 3 σ x σ y + C 3 ,
where μx, μy are the local means, σx, σy are standard deviations and σxy is cross-covariance for images x, y. If α = β = γ = 1 (the default values for exponents), and C3 = C2/2 (default value of C3) the index simplifies to [28]:
S S I M ( x ,   y ) = ( 2 μ x μ y + C 1 ) ( 2 σ x y + C 2 ) ( μ x 2 + μ y 2 + C 1 ) ( σ x 2 + σ y 2 + C 2 ) ,
which satisfies the following conditions [28]:
  • Symmetry: SSIM(x, y) = SSIM(y, x);
  • Boundedness: SSIM(x, y) < 1;
  • Unique maximum: SSIM(x, y) = 1 if and only if x = y.

Multi-Scale SSIM Index (MS-SSIM)

Multi-scale method is a convenient way to incorporate image details at different resolutions. Experiments show that with an appropriate parameter setting, the multi-scale method outperforms the best single-scale SSIM model as well as state-of-the-art image quality metrics [28].
Multi-scale SSIM is calculated by first converting the two images to grayscale, if they are not already. Next, the images are down-sampled to multiple scales using a Gaussian pyramid. For each scale, the SSIM index is calculated between the two images. The final MS-SSIM value is obtained by averaging the SSIM values across all scales. Finally, the MS-SSIM value is a single value between −1 and 1, where 1 indicates a perfect match and −1 indicates no similarity.

4. Materials and Methods

4.1. Test Samples

Eight samples, each measuring 120 mm × 120 mm × 5 mm and containing defects, were printed using a 3D printer (Prusa i3 MK3S+, Prusa Research a.s., Partyzánská 188/7A 17000 Praha 7, Czech Republic) with PET-G (polyethylene terephthalate glycol, Table 1 [29]) material and FDM (fused deposition modeling) technology. FDM 3D printers work by extruding thermoplastic filaments through a heated nozzle, melting the material and layering it onto a build platform until the part is complete.
The defects, in the form of blind holes, had known dimensions, geometry, and location. For this study, eight samples were made: S1, S2, S3, and S4 for the samples with circular cross-section defects, and S5, S6, S7, and S8 for the samples with square cross-section defects. Defects were placed 1 mm and 0.5 mm under the measured surface and the midpoints of the defects closest to the edges of the samples had a distance of 30 mm. The samples are shown in Figure 1, where the defects with square cross-sections have an edge length of 20 mm and the defects with circular cross-sections have a diameter of 20 mm. On the measured surface of the test samples a black color with the high emissivity (0.96) was applied to reduce the reflection of the surface.
The additively manufactured specimens were built with full density.

4.2. Experimental Setup

Test samples were measured using lock-in thermography. In lock-in thermography, various types of excitations are used, such as laser, halogen lamp, ultrasound sonotrode, flash lamp, etc. In this experiment, two sources of excitation were used: a halogen lamp (Kaiser Videolight 4, Kaiser Fototechnik GmbH & Co. KG, Buchen, German) and a flash lamp (Godox QS 1200 II, GODOX Photo Equipment Co.,Ltd., Shenzhen, China). Using the National Instruments module NI 9481, the halogen and flash lamps were connected to the PC and a control script was written in MATLAB. In the case of the halogen lamp, the device was turned on and off every 1 min, and in the case of the flash lamp, it was fired every 2 min. The halogen lamp controlled by the NI 9481 module was used to generate a one-minute thermal wave, and the flash lamp generated a flash with duration 1/800 s. Every period of excitation was 2 min because the lock-infrequency was set to 1/120 Hz. Five periods were performed for each measurement to investigate the effect of the excitation periods number on the results. Table 2 provides a summary of the setup parameters used for the experiment.
For collecting the data (sequences of thermograms), we used an infrared camera (FLIR SC7500, FLIR Systems Advanced Thermal Solutions SA, Croissy-Beaubourg, France) with thermal sensitivity less than 20 mK, a spectral range 1.5–5.1 µm and 50 mm lens (objective). The cooled detector is a focal plane array (FPA) with resolution of 320 × 256, made of InSb. Emissivity ε = 0.96 was adopted for the tests because the black color with the same emissivity value were placed on the samples. Figure 2 shows a schematic representation of the experimental setup.
The defects are not located inside the specimens but on one side of the specimen so that they are located 1 mm and 0.5 mm below the surface, respectively. For the measurements, the samples were turned towards the camera so that the camera could see the side of the sample without defects, so that from the camera’s point of view the defects were below the surface.
Control signals for switching on and off the halogen lamp and also for switching on the flash lamp were generated in Matlab through the NI 9481 control unit.

4.3. Data Processing and Phase Image Calculation

Each measurement consisted of 3000 thermograms recorded over a 10-min period. The camera was positioned 0.5 m away from the sample’s surface being measured. The data was collected using the ResearchIR software and exported to a Matlab file. The exported data included sequences of thermograms and files containing information about the date and time each thermogram was taken. We created a script in Matlab to load this data and calculate the phase image (using Formula (7)) for one period (600 thermograms) up to five periods (3000 thermograms). The results were then compared using the SSIM and MS-SSIM indices. SSIM method was chosen because it provides a reliable and accurate measure of the structural similarity between two images, and SSIM is well-established and trusted in many image and video processing applications. The MS-SSIM was used as a second comparative method and because it outperforms the single-scale SSIM method.

5. Results

In this paper, two excitation methods were used: flash lamp and halogen lamp. For each method, eight measurements were performed. The samples were designed so that the defects occupied different areas on the measured surface, ranging from approximately 8% for three defects to 25% for nine defects.
Figure 3 and Figure 4 present a sample thermogram from the recorded sequences of thermograms on the left side. The mean temperature value for the highlighted area is plotted on the graph labeled ”Output” located on the right side. The graph labeled ”Input” shows the input signal The signal was generated in Matlab and used to control the flash lamp and halogen lamp through the NI 9481 control unit.

5.1. Lock-In Thermography Using Flash Lamp Defects Placed 1 mm under the Measured Surface

Figure 5 and Figure 6 display phase images calculated for all eight samples for one to five periods. The reference phase images were calculated using data from a 2-min record, whereas the phase image for five periods was calculated using data from a 10-min record.
The phase images calculated for different periods of time, ranging from one to five, are very similar in appearance. To measure the similarity between these images, the SSIM index and MS-SSIM index were used. The reference phase images, calculated from one period of time, were used as the benchmark to compare against the other phase images calculated for different periods of time. The results, as shown in Table 3, demonstrate that the difference between the reference phase images and the other phase images calculated for different periods of time is quite small. In fact, the difference, as measured by the SSIM index, is less than 10% and, when measured by the MS-SSIM index, is less than 5%. This indicates a high level of similarity between the phase images calculated for different periods of time, providing confidence in the accuracy and reliability of the calculations performed.
The Figure 7 shows graphically the SSIM and MS-SSIM index values for the flash lamp. A decrease in the index values with increasing number of periods was observed. We assume that the extreme values of the decrease are due to the increasing noise in the processed data.

5.2. Lock-In Thermography Using Halogen Lamp Defects Placed 1 mm under the Measured Surface

The data processing and phase image calculation procedures are similar to those used in the case of lock-in thermography using a flash lamp, as described in Section 4.1.
The Figure 8 shows graphically the SSIM and MS-SSIM index values for a halogen lamp. A decrease in the index values with increasing number of periods was observed. We assume that the extreme values of the decrease are due to the increasing noise in the processed data.
Figure 9 and Figure 10 present phase images calculated for all eight samples for periods ranging from one to five. It can be observed from the figures that the phase images are similar, but not as clear as in the previous case. With an increase in the number of periods, the phase images become more blurred, as seen in the comparison between the S3 reference and S3 for four periods. Results in Table 4 indicate that, using both the SSIM and MS-SSIM indices, the difference between the reference phase images and phase images calculated for two periods is less than 10%. However, as the number of periods increases, the percentage differences also increase.

5.3. Lock-In Thermography Using Halogen Lamp and Flash Lamp. Defects Placed 0.5 mm under the Measured Surface

The next measurement that was performed was on samples that had a defect depth of 0.5 mm below the surface of the measured area. The sample was extruded from PET-G material. In this case, a sample with eight square holes, also marked as S8, is chosen. The measurement setup is the same as for the 1 mm thickness samples. The measurement was performed for both the flash lamp and the halogen lamp excitation types. The phase image from the halogen lamp excitation can be observed in Figure 11 and the phase image from the flash lamp excitation can be observed in Figure 12. The representations are made for five periods. The comparison of the SSIM and MS-SSIM indexes is shown in Table 5.
The graph indicating the dependence of the MS-SSIM and SSIM indices on the number of periods is on the left side for flash lamp excitation and on the right side for halogen lamp excitation. The graphs are made for the comparison of sample S8 wall thickness of the measured area of 1 mm and 0.5 mm (Figure 13).

6. Discussion

The case of excitation by flash lamp. Increasing the number of periods used in the lock-in processing does not increase the value of the MS-SSIM index or the SSIM for the 1 mm and 0.5 samples This observation is represented by the triangular points in Figure 13 for the flash lamp.
We consider the slower decline of the MS-SSIM index relative to the SSIM index to be due to the lower sensitivity of the MS-SSIM to period growth.
The case of excitation by halogen lamp. Increasing the number of periods used in the lock-in data processing does not significantly increase the value of the MS-SSIM index or the SSIM index for the 1 mm sample, and the index values shown are lower for higher numbers of periods. However, in the case of a 0.5 mm sample, increasing the number of periods used in the data processing by the lock-in method increases significantly the value of the MS-SSIM index and not the SSIM for the 0.5 mm sample. In this case, we cannot claim that one period is sufficient.
The presented results for the halogen lamp excitation indicate a possible dependence of the MS-SSIM and SSIM index values on the depth below the surface in which the defect is located. We assume that the halogen lamp heated the sample more than the flash lamp, and then the increase of the indices at 0.5 mm sample may be due to this phenomenon. However, the rising temperature when measuring using a halogen lamp may cause sample damage if the sample is excited for a large number of periods.

7. Conclusions

In this paper, two excitation methods, namely flash lamp and halogen lamp, were used to perform eight measurements on samples. The phase images were calculated for periods ranging from one to five, and the similarity index was calculated between the reference phase image and others. The data processing and phase image calculation procedures were similar to those used in the case of lock-in thermography using a flash lamp.
For the measurements on the 1 mm and 0.5 mm depth of the defect surfaces, the same lamp power was set for both the halogen lamp and the flash lamp, and the flash power was also set to be the same. The same lock-in periods, camera frame rates, and camera integration times were also used, and the same distance of the samples from the excitation sources was maintained.
Using a flash lamp, it was found that increasing the number of periods that were used in the data processing by the lock-in method did not increase the value of the MS-SSIM index or the SSIM for the 1 mm sample. For the 0.5 mm sample, the increase in value is not demonstrable.
If we subjectively or visually evaluate (0.5 mm sample) the images for one to five periods (see Figure 13), then the image for five periods appear blurrier than for one period, although the values of both indices increase with increasing number of periods.
Based on the above, in future research we will want to observe the influence of other parameters on the value of the above indices as well as the influence of other parameters, such as the power of the halogen reflector or the temperature of the sample during the test, and to make these measurements for other defect depths below the surface as well.
In the case of excitation of an object using a flash lamp, we recommend to use only one period, and we cannot make a recommendation in the case of using a halogen lamp.

Author Contributions

Conceptualization, M.S. and V.D.; methodology, V.D.; software, M.S.; validation, V.D., P.K. and L.J.; formal analysis, Z.Š.; investigation, M.S.; resources, P.K.; data curation, M.S.; writing—original draft preparation, M.S.; writing—review and editing, L.J.; visualization, M.S.; supervision, V.D.; project administration, Z.Š.; funding acquisition, V.D. All authors have read and agreed to the published version of the manuscript.

Funding

This work has been supported by Projects VEGA 1/0510/20, KEGA 011ZU-4/2022.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon request.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Kostroun, T.; Dvořák, M. Application of the pulse infrared thermography method for nondestructive evaluation of composite aircraft adhesive joints. Materials 2021, 14, 533. [Google Scholar] [CrossRef]
  2. Ciampa, F.; Mahmoodi, P.; Pinto, F.; Meo, M. Recent advances in active infrared thermography for non-destructive testing of aerospace components. Sensors 2018, 18, 609. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  3. Meola, C.; Boccardi, S.; Carlomagno, G.M. Infrared Thermography in Evaluation of Aerospace Composite Materials; Woodhead Publishing: Cambridge, UK, 2017; pp. 25–56, 98–103. [Google Scholar]
  4. Deane, S.; Avdelidis, N.P.; Castanedo, I.C.; Zhang, H.; Nezhad, H.Y.; Williamson, A.A.; Mackley, T.; Davis, M.J.; Maldague, X.; Tsourdos, A. Application of NDT thermographic imaging of aerospace structures. Infr. Phys. Technol. 2019, 97, 456–466. [Google Scholar] [CrossRef] [Green Version]
  5. Moustakidis, S.; Anagnostis, A.; Karlsson, P.; Hrissagis, K. Non-destructive inspection of aircraft composite materials using triple IR imaging. IFAC-PapersOnLine 2016, 49, 291–296. [Google Scholar] [CrossRef]
  6. Manohar, A.; di Scalea, F.L. Detection of defects in wind turbine composite blades using statistically enhanced Lock-in Thermography. Struct. Health Mon.-Inter. J. 2013, 12, 566–574. [Google Scholar] [CrossRef]
  7. Manohar, A.; di Scalea, F.L. A fast lock-in infrared thermography implementation to detect defects in composite structures like wind turbine blades. Rev. Prog. Quan. Nondest. Eval. Vols. 32a 32b 2013, 1511, 563–570. [Google Scholar]
  8. Sanati, H.; Wood, D.; Sun, Q. Condition monitoring of wind turbine blades using active and passive thermography. App. Sci. Bas. 2018, 8, 2004. [Google Scholar] [CrossRef] [Green Version]
  9. Montinaro, N.; Fustaino, M.; Pantano, A. Carbon nanotubes dispersion assessment in nanocomposites by means of a pulsed thermographic approach. Materials 2020, 13, 5649. [Google Scholar] [CrossRef]
  10. Montinaro, N.; Fustaino, M.; Bellisario, D.; Quadrini, F.; Santo, L.; Pantano, A. Testing the dispersion of nanoparticles in a nanocomposite with an ultra-low fill content using a novel non-destructive evaluation technique. Materials 2022, 15, 1208. [Google Scholar] [CrossRef]
  11. Chung, Y.; Shrestha, R.; Lee, S.; Kim, W. Binarization mechanism evaluation for water ingress detectability in honeycomb sandwich structure using lock-inthermography. Materials 2022, 15, 2333. [Google Scholar] [CrossRef]
  12. Colom, M.; Rodríguez-Aseguinolaza, J.; Mendioroz, A.; Salazar, A. Sizing the depth and width of narrow cracks in real parts by laser-spot lock-inthermography. Materials 2021, 14, 5644. [Google Scholar] [CrossRef] [PubMed]
  13. Kim, G.; Hong, S.; Kim, G.H.; Jhang, K.Y. Evaluation of subsurface defects in fiber glass composite plate using lock-in technique. Int. J. Precis. Eng. Manuf. 2012, 13, 465–470. [Google Scholar] [CrossRef]
  14. Toscano, C.; Riccio, A.; Camerlingo, F.; Meola, C. On the use of lock-inthermography to monitor delamination growth in composite panels under compression. Sci. Engine Comp. Mater. 2014, 21, 485–492. [Google Scholar] [CrossRef]
  15. Gu, J.; Choi, N.S. Evaluation of delamination of dental composite restoration using infrared lock-inthermography. Comp. Res. 2012, 25, 236–240. [Google Scholar]
  16. Meola, C.; Carlomango, G.M.; Annicchiarico, D.; Giordano, M.; Zarrelli, M. Detection of delamination in carbond-fibre-reinforced polymers with lock-inthermography. Proc. Inst. Mech. Eng. Part G-J. Aerosp. Eng. 2010, 224, 1219–1227. [Google Scholar] [CrossRef]
  17. Opara, L.K.; Klarin, B.; Neves, P.; Domazet, Z. Thermal imaging and Thermoelastic Stress Analysis of impact damage of composite materials. Eng. Fail. Anal. 2011, 18, 713–719. [Google Scholar] [CrossRef]
  18. Liu, T.; Zhang, Y.; Netzelmann, U.; Guo, J.G.; Zhang, W. Inspection of Impact Damage on CFRP by Lock-In Induction Thermography. In Eleventh International Conference on Information Optics and Photonics (CIOP 2019); SPIE: Cergy Pontoise, France, 2019; Volume 11209. [Google Scholar]
  19. Grosso, M.; Margarit-Mattos, I.C.P.; Pereira, G.R. Pulsed thermography inspection of composite anticorrosive coatings: Defect detection and analysis of their thermal behaviour through computational simulation. Materials 2020, 13, 4812. [Google Scholar] [CrossRef]
  20. Rodriguez-Aseguinolaza, J.; Colom, M.; Gonzalez, J.; Mendioroz, A.; Salazar, A. Quantifying the width and angle of inclined cracks using laser-spot lock-inthermography. NDT E Inter. 2021, 122, 102494. [Google Scholar] [CrossRef]
  21. Nolte, P.W.; Malvisalo, T.; Wagner, F.; Schweizer, S. Thermal diffusivity of metals determined by lock-inthermography. Quant. Inf. Therm. J. 2017, 14, 218–255. [Google Scholar] [CrossRef]
  22. D’Accardi, E.; Palumbo, D.; Galietti, U. A comparison among different ways to investigate composite materials with lock-inthermography: The multi-frequency approach. Materials 2021, 14, 2525. [Google Scholar] [CrossRef]
  23. Kucharska, M.; Jaskowska-Lemańska, J. Active Thermography in Diagnostics of Timber Elements Covered with Polychrome. Materials 2021, 14, 1134. [Google Scholar] [CrossRef] [PubMed]
  24. Sakagami, T.; Kubo, S. Applications of pulse heating thermography and lock-inthermography to quantitative nondestructive evaluations. Inf. Phys. Tech. 2002, 43, 211–218. [Google Scholar] [CrossRef]
  25. Moskovchenko, A.I.; Švantner, M.; Vavilov, V.P.; Chulkov, A.O. Characterizing depth of defects with low size/depth aspect ratio and low thermal reflection by using pulsed ir thermography. Materials 2021, 14, 1886. [Google Scholar] [CrossRef] [PubMed]
  26. Sakagami, T.; Kubo, S.; Nakamura, S.; Kawashima, Y.; Komiyama, T. Application of lock-in data processing for thermographic NDT of concrete structures. Therm. XXIV 2002, 4710, 552–557. [Google Scholar]
  27. Breitenstein, O.; Warta, W.; Langenkamp, M. Lock-In Thermography, 2nd ed.; Springer: Berlin/Heidelberg, Germany; Dordrecht, The Netherlands; London, UK; New York, NY, USA, 2010; pp. 18–19. [Google Scholar]
  28. Wang, Z.; Simoncelli, E.P.; Bovik, A.C. Multiscale structural similarity for image quality assessment. Thirty-Seventh Asilomar Conf. Signals Syst. Comput. 2003, 2, 1398–1402. [Google Scholar]
  29. Rigid.ink PETG Data Sheet. Available online: https://devel.lulzbot.com/filament/Rigid_Ink/PETG%20DATA%20SHEET.pdf (accessed on 17 March 2023).
Figure 1. Test samples.
Figure 1. Test samples.
Materials 16 02763 g001
Figure 2. Experimental configuration of a lock-in thermography setup.
Figure 2. Experimental configuration of a lock-in thermography setup.
Materials 16 02763 g002
Figure 3. Halogen lamp input and output signal.
Figure 3. Halogen lamp input and output signal.
Materials 16 02763 g003
Figure 4. Flash lamp input and output signal.
Figure 4. Flash lamp input and output signal.
Materials 16 02763 g004
Figure 5. Phase images for the samples with the circular cross section defects. A flash lamp was used as a source of excitation. S1–S4 denotes configuration of the circle holes.
Figure 5. Phase images for the samples with the circular cross section defects. A flash lamp was used as a source of excitation. S1–S4 denotes configuration of the circle holes.
Materials 16 02763 g005
Figure 6. Phase images for the samples with the square cross section defects. A flash lamp was used as a source of excitation. S5–S8 denotes configuration of the square holes.
Figure 6. Phase images for the samples with the square cross section defects. A flash lamp was used as a source of excitation. S5–S8 denotes configuration of the square holes.
Materials 16 02763 g006
Figure 7. Variation of the SSIM and MS-SSIM index values as a function of the number of periods for the flash lamp.
Figure 7. Variation of the SSIM and MS-SSIM index values as a function of the number of periods for the flash lamp.
Materials 16 02763 g007
Figure 8. Variation of the SSIM and MS-SSIM index values as a function of the number of periods for the halogen lamp.
Figure 8. Variation of the SSIM and MS-SSIM index values as a function of the number of periods for the halogen lamp.
Materials 16 02763 g008
Figure 9. Phase images for the samples with the circular cross section defects. A halogen lamp was used as a source of excitation. S1–S4 denotes configuration of the circle holes.
Figure 9. Phase images for the samples with the circular cross section defects. A halogen lamp was used as a source of excitation. S1–S4 denotes configuration of the circle holes.
Materials 16 02763 g009
Figure 10. Phase images for the samples with the square cross section defects. A halogen lamp was used as a source of excitation. S5–S8 denotes configuration of the square holes.
Figure 10. Phase images for the samples with the square cross section defects. A halogen lamp was used as a source of excitation. S5–S8 denotes configuration of the square holes.
Materials 16 02763 g010
Figure 11. Phase images for the samples with the square cross section defects. A halogen lamp was used as a source of excitation. Defects were placed 0.5 mm under the measured surface.
Figure 11. Phase images for the samples with the square cross section defects. A halogen lamp was used as a source of excitation. Defects were placed 0.5 mm under the measured surface.
Materials 16 02763 g011
Figure 12. Phase images for the samples with the square cross section defects. A flash lamp was used as a source of excitation. Defects were placed 0.5 mm under the measured surface.
Figure 12. Phase images for the samples with the square cross section defects. A flash lamp was used as a source of excitation. Defects were placed 0.5 mm under the measured surface.
Materials 16 02763 g012
Figure 13. Comparison of the dependence of the MS-SSIM and SSIM indices on the number of periods on the left side for flash lamp excitation and on the right side for halogen lamp excitation.
Figure 13. Comparison of the dependence of the MS-SSIM and SSIM indices on the number of periods on the left side for flash lamp excitation and on the right side for halogen lamp excitation.
Materials 16 02763 g013
Table 1. PETG, the selected parameters of mechanical and thermal properties (injection molded), ISO method, [29].
Table 1. PETG, the selected parameters of mechanical and thermal properties (injection molded), ISO method, [29].
Parameter Test MethodValue
DensityISO 1183, Method
D
1.27 g/cm3
Tensile stress @ breakISO 52728 MPa
Tensile stress @ yieldISO 52750 MPa
Elongation @ breakISO 527100%
Tensile modulusISO 5272100 MPa
Deflection temperature
@ 0.455 MPaD 64870 °C
@ 1.82 MPaD 64864 °C
Vicat softening temperatureD 152585 °C
Glass transition temperature (Tg)DSC80 °C
Coefficient of linear thermal
expansion (−30 °C to 40 °C)
D 6965.1 × 10−5/°C (mm/mm·°C)
Table 2. Specification of the parameters used for lock-in thermography.
Table 2. Specification of the parameters used for lock-in thermography.
Parameter Value
Distance sample to IR camera0.5 m ± 0.1 m
Distance sample to halogen lamp0.5 m ± 0.1 m
Distance sample to flash lamp0.5 m ± 0.1 m
IR camera frame rate5 Hz
Lock-in frequency1/120 Hz
Number of periods5
Power of halogen lamp1 kW
Power of flash lamp1.2 kW
Table 3. Comparation of phase images for all measurement in the case of a flash lamp as source of excitation.
Table 3. Comparation of phase images for all measurement in the case of a flash lamp as source of excitation.
SampleSSIM Index
MS-SSIM Index
2 Periods3 Periods4 Periods5 Periods
S10.96
0.99
0.94
0.98
0.93
0.97
0.93
0.97
S20.95
0.99
0.95
0.98
0.94
0.98
0.94
0.98
S30.91
0.99
0.95
0.99
0.94
0.98
0.94
0.99
S40.97
0.99
0.95
0.99
0.94
0.99
0.94
0.99
S50.95
0.98
0.92
0.98
0.92
0.98
0.91
0.97
S60.95
0.99
0.94
0.99
0.92
0.98
0.92
0.98
S70.95
0.99
0.94
0.98
0.92
0.97
0.91
0.97
S80.96
0.99
0.96
0.99
0.95
0.99
0.95
0.99
Table 4. Comparation of phase images for all measurement in the case of halogen lamp as source of excitation.
Table 4. Comparation of phase images for all measurement in the case of halogen lamp as source of excitation.
SampleSSIM Index
MS-SSIM Index
2 Periods3 Periods4 Periods5 Periods
S10.93
0.91
0.81
0.83
0.90
0.83
0.88
0.79
S20.97
0.96
0.95
0.95
0.97
0.96
0.95
0.93
S30.96
0.95
0.97
0.94
0.78
0.77
0.86
0.79
S40.94
0.92
0.95
0.93
0.75
0.75
0.81
0.75
S50.97
0.97
0.91
0.88
0.81
0.71
0.85
0.81
S60.95
0.98
0.91
0.90
0.79
0.82
0.89
0.88
S70.89
0.90
0.86
0.88
0.73
0.78
0.51
0.52
S80.92
0.95
0.93
0.95
0.93
0.93
0.79
0.76
Table 5. Comparation of phase images for measurement in the case of the halogen lamp and the flash lamp as source of excitation for defects place 0.5 mm under measured surface.
Table 5. Comparation of phase images for measurement in the case of the halogen lamp and the flash lamp as source of excitation for defects place 0.5 mm under measured surface.
ExcitationSSIM Index
MS-SSIM Index
2 Periods3 Periods4 Periods5 Periods
Flash lamp0.93
0.97
0.93
0.98
0.8
0.95
0.87
0.97
Halogen lamp0. 85
0. 92
0.9
0.96
0.93
0.97
0.94
0.99
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Sapieta, M.; Dekýš, V.; Kopas, P.; Jakubovičová, L.; Šavrnoch, Z. The Impact of Excitation Periods on the Outcome of Lock-In Thermography. Materials 2023, 16, 2763. https://doi.org/10.3390/ma16072763

AMA Style

Sapieta M, Dekýš V, Kopas P, Jakubovičová L, Šavrnoch Z. The Impact of Excitation Periods on the Outcome of Lock-In Thermography. Materials. 2023; 16(7):2763. https://doi.org/10.3390/ma16072763

Chicago/Turabian Style

Sapieta, Milan, Vladimír Dekýš, Peter Kopas, Lenka Jakubovičová, and Zdenko Šavrnoch. 2023. "The Impact of Excitation Periods on the Outcome of Lock-In Thermography" Materials 16, no. 7: 2763. https://doi.org/10.3390/ma16072763