1. Introduction
Hospitalized patients are at high risk of morbidity and even mortality from hospital-acquired infections (HAIs), in addition to their primary illness. According to the European Centre for Disease Prevention and Control, 8.9 million European patients acquire infections in hospitals and long-term care facilities every year [
1]. HAIs cause over 90,000 deaths per year in European countries [
2], more than any other infectious disease under surveillance. Hospital-acquired pneumonia is the second most common nosocomial infection and one of the most common causes of death from HAI in critically ill patients [
3]. Ventilator-associated pneumonia (VAP) is a serious hospital-acquired pneumonia which is defined as lung infection acquired by patients mechanically ventilated in the intensive care unit for at least 48 h [
4]. Patients with VAP have increased length of hospitalization and consequently increased healthcare cost [
5].
Endotracheal tube microbial colonization is a major risk factor for VAP and a multispecies biofilm of commensal microorganisms forms in 90% of the tubes used in mechanical ventilation [
6]. This biofilm presents a clinical challenge since pathogens can attach to the microbial community formed on the inner surface of the endotracheal tube. Carried by airway pressure, they can then bypass normal barriers and reach the pulmonary parenchyma or infect the lungs. In a prospective observational study that included patients from 27 European intensive care units (ICUs),
Staphylococcus aureus is reported as the most common Gram-positive isolate in VAP patients and
Pseudomonas aeruginosa,
Acinetobacter baumannii, and
Escherichia coli are among the most common Gram-negative isolates [
7]. Mortality rates can vary in different patient groups and depend on underlying conditions and disease severity. In a meta-analysis using randomized trials and considering a homogeneous group of patients, the overall attributable mortality associated with VAP was 13% [
8].
To address this high risk of mortality from VAP, several strategies have been investigated, such as modifying endotracheal tubes in order to make their surface antifouling and/or preventing biofilm formation by incorporating substances with antibacterial properties during the manufacturing process [
9]. The limited number of available solutions is due to the complex process of developing reliable devices, starting from idea up to the market. For a successful development, an interdisciplinary approach is essential, requiring drug development, material science, engineering, clinically relevant animal models and, finally, well-designed clinical studies. Thus, the availability of tools to produce testing samples for each step of characterization and validation is key to achieving the required results. In this study, we focus on producing tubes for preclinical tests with a VAP mouse model.
The main manufacturing process used to produce tubing is hot melt extrusion. In this process, a thermoplastic polymer is melted and extruded through a specially designed nozzle (commonly known by the term “extrusion die” in the relevant literature) in the shape of the product [
10]. Via this process, a wide range of products can be manufactured, from plastic profiles used in buildings, to tubes used for medical applications. Moreover, there are several examples of the latter application such as tubes and catheters being used in chemotherapy and neonates. The nozzles used in such processes are custom-made and composed of multiple components. An alternative approach to producing such nozzles can be via additive manufacturing (AM) technologies. These technologies have recently been used in the production of tools in different sectors [
11,
12]. They offer the unique possibility of layered manufacturing in producing complex objects in a cost- and time-effective manner. Moreover, AM has been used for the production of extrusion nozzles capable of extruding multiple materials with a coaxial complex design [
13,
14,
15].
In this work, we propose a method for producing test tubes to validate a VAP mouse model via the outstanding capabilities of AM technologies. An extrusion nozzle was designed using computer-aided design software, and was 3D printed to be used as a tool for hot melt extruding tubes that can meet the demands of the mouse VAP model. A strict outer and inner diameter of 1.00 mm and 0.60 mm, respectively, is required to fit within the mouse trachea. The use of a mini single screw extrusion system was key to having a feasible system that allows testing different formulations. On the other hand, solvent casting was used to produce drug-loaded films. Ciprofloxacin (CPX), which is a conventional broad-spectrum antibiotic widely used to treat VAP infections, was chosen as an antibacterial and antibiofilm compound [
16]. Thermoplastic polyurethane (TPU) was selected as the carrier polymer since it is widely used in many medical tubing applications [
17].
The tubes were characterized for their thermophysical properties, release of CPX, and cytotoxicity. The antibacterial and antibiofilm properties of the tubes were assessed by antimicrobial activity testing, bacterial adherence evaluation, and scanning electron microscopy. The strain of choice was
S. aureus ATCC 25923, an established biofilm-forming strain used in previous standardization studies and in the screening of new and repurposed drugs [
18]. The insertion procedure of the tubes was based on a previously described
P. aeruginosa model used to investigate treatment of VAP [
19].
The main aim of the presented study is to establish a proof of concept for the development of a novel 3D-printed nozzle, which allows the extrusion of customized drug-loaded tubes which meet the demands of an S. aureus VAP mouse model. Our findings support the proposal that the manufacturing approach described can successfully produce CPX-loaded tubes that allow us to validate the developed VAP mouse model. Proof of concept is key for further use of this novel technology in the evaluation of VAP prevention by other promising antibacterial and antibiofilm compounds incorporated in hot melt extruded polymeric tubes.
2. Materials and Methods
2.1. Materials Used
The TPU used was Tecoflex EG-60D, Lubrizol, purchased from VELOX (Laren, The Netherlands). The polymer contains a soft segment and hard segment in a ratio of 3:1. It has a glass transition temperature of 23 °C, and a melting temperature of 63 °C. Its molecular weight is 7.89 kDa [
20]. CPX was purchased from Acros Organics (Geel, Belgium). Some preliminary testing was performed using polylactic acid (PLA) filament which was purchased from 3d4makers (Haarlem, The Netherlands). It has a glass temperature around 50 °C and a melting range between 170 °C and 230 °C [
21].
TPU was loaded with CPX using the solvent casting approach adapted from previous work by Shaqour et al. [
22] (
Figure 1A). To prepare the TPU-CPX film, 5% (
w/w) of CPX was suspended in an organic solvent (chloroform) (Chempur, Piekary Śląskie, Poland) and sonicated for 30 min. Then, a magnetic stirrer (VELP Scientifica Srl, Usmate, Italy) was used for 8 h to homogeneously distribute the drug particles in the solvent. Afterwards, 95% (
w/w) TPU was added to the system while stirring, and was left overnight to dissolve the polymer. The solution was then poured on a Petri dish (200 mm in diameter) to evaporate the solvent completely under the fume hood, resulting in solid TPU-CPX (95:5
w/w) films. Finally, the films were vacuum-dried for 3 days at 25 °C and 50 mbar. To prepare the TPU-only films, the same procedure was performed excluding the addition of CPX.
2.2. Nozzle Design and 3D Printing
A novel nozzle with the ability of extruding tubes was designed using Creo Parametric (v.5.0.4, PTC, Boston, MA, USA) (
Figure 2A). Our nozzle was designed in a more simplified structure and taking into consideration the capabilities of AM technologies. The inlet of the nozzle has a diameter of 2.4 mm. The outlet of the nozzle has an outer diameter of 1 mm with a concentric coaxial rod with an outer diameter of 0.5 mm. The final design was sent to an external manufacturer (i.materialize, Leuven, Belgium). Metal 3D binder printing technology was used to produce the nozzle [
23] (
Figure 2B). In this technology, stainless steel fine powder is used as the raw material. A thin layer of powder is spread on top of the printing platform. Subsequently, the printing head deposits a binding material on selected parts of the layer in order to bind the metallic particles. Then, the printing platform is lowered, and the process is repeated. Finally, the objects produced in this process are placed in a sintering oven to decompose the binding material and fuse the metallic particles.
Microcomputer tomography scanning (μCT) measurements were performed using a SKYSCAN 1172 (Bruker, Kontich, Belgium) in order to check the internal geometry of the 3D-printed nozzles (
Figure 2C). The scan of nozzles was performed at 100 kV and 100 μA over 180° with a rotation step of 0.42° and exposure time of 600 ms. A Cu+Al filter was used during the scanning process. The image pixel size was 7.95 μm. The obtained planar images were reconstructed and analyzed with the instrument software—NRecon and CTAn (v1.10.1.0, Bruker micro-CT, Kontich, Belgium).
2.3. Fibers and Tubes Extrusion
In order to test the ability of extruding tubes with complex cross section using the new nozzle design, several shapes were used (star, cross, and circular) in different nozzles (
Figure 2(D1–D3), respectively). Preliminary testing for the 3D-printed nozzle was conducted using a PLA filament and extrusion temperature of 200 °C. This was done as PLA is widely available in the filament form and has suitable rheological properties when melted. The nozzle was fitted on a normal filament extruder usually used in fused filament fabrication 3D printers. Several extrusion experiments were conducted at different speeds.
TPU and TPU-CPX fibers and tubes were extruded using an in-house developed single screw extrusion set up (
Figure 1B) attached to the designed coaxial nozzle (
Figure 2(D3)) or an e3d v6 stainless steel nozzle (E3D-Online, Oxford, UK) with a 0.8 mm diameter. Verstraete et al. [
20] conducted a screening for a wide variety of TPU grades and considered 180 °C suitable for processing the polymer. Thus, the extrusion was done at a temperature of 180 °C and the screw extrusion speed was set at 15 rpm. TPU and TPU-CPX films were cut into strips and manually fed into the extruder. The extruded fibers were cut into 40 mm pieces, while the extruded tubes were then cut into 10 mm pieces.
Tubes were inspected using a Leica s9i microscope (Leica, Machelen, Belgium). Pictures were analyzed using ImageJ software (v.1.52A, NIH, Bethesda, MD, USA) by calculating the equivalent diameter of the measured area for the inner and outer circles for each tube (n = 6 for each group).
2.4. Mechanical Testing
In order to investigate the effect of CPX loading on the mechanical properties of the TPU polymer, tensile tests were performed on an AGS 5 kND tensile testing machine (Shimadzu, Duisburg, Germany). The tensile rate was set to 20 mm/min and the distance between grips was 20 mm. For simplicity, we used a simple geometry (fibers) to study the tensile properties of the CPX-loaded versus non-loaded TPU. Extruded fibers from a 0.8 mm nozzle were used for this test (n = 5 for each group). The tensile elastic modulus (EM) was calculated from the force-displacement curve generated by the machine’s software. A Matlab (v.R2020b, The Math-works, Inc., Natick, MA, USA) code was written to calculate the stress-strain curve and the slope of the best fit line between 50% and 150% strain.
2.5. Thermal Analysis
To study the thermal stability and the temperature of decomposition for the extruded materials, thermogravimetric analysis (TGA) was conducted using a Q5000 Analyzer (TA Instruments, New Castle, DE, USA). Tubes of around 10 mg were placed on a platinum pan and then a dynamic heating ramp of 20 °C/min with a resolution of 3.00 °C was performed up to 500 °C under a nitrogen flow of 60 mL/min.
Moreover, the thermal stability of the CPX was tested by heating its powder using the TGA machine for 5 min at 180 °C. Afterwards, non-heated and heated samples were compared to test whether the processing temperature affects the activity of the drug or not. The 5 min period was selected as it is close to but slightly longer than the expected residence time of the drug in the extruder.
2.6. Fourier-Transform Infrared Spectroscopy (FTIR)
CPX powder before and after heating to 180 °C was examined using FTIR spectroscopy in order to study the effect of heat on the drug’s molecular structure. Moreover, in order to study the interactions between the drug and the TPU molecules, FTIR analysis was conducted on the produced tubes. The FTIR spectrometer used (Nicolet 8700, ThermoScientific, Madison, WI, USA) is equipped with a diamond attenuated total reflectance (ATR) accessory. All ATR-FTIR spectra were recorded at room temperature in the 400–4000 cm−1 range. The spectral resolution and accuracy were 4 cm−1 and ±1 cm−1, respectively.
2.7. In Vitro Cell Proliferation Assay
A cell proliferation assay was performed on the CPX powder heated at 180 °C and not heated to evaluate the cytotoxic effects after heating. The L929 mouse fibroblast cell line was purchased from European Collection of Authenticated Cell Cultures (ECACC, Salisbury, UK) and cultured in cell culture flasks containing Dulbecco’s modified Eagle’s medium (DMEM, Thermofisher, Walthman, MA, USA) supplemented with an additional 10% of fetal bovine serum (FBS) (Thermofisher, Walthman, MA, USA) and 1% of penicillin-streptomycin (Thermofisher, Walthman, MA, USA). Cells were incubated at 37 °C, 5% CO2 and humidity. Cells were detached with trypsin-EDTA (0.25% trypsin, 1 mM EDTA, Thermofisher, Walthman, MA, USA) and counted for the seeding. Per well, 100 µL containing 1 × 103 cells was seeded in a 96-well cell culture plate and cultured for 24 h at 37 °C and 5% CO2. The next day, the medium was removed and 200 µL of CPX (non-heated control or heated at 180 °C) at 10 µg/mL in DMEM was added. DMEM with 1% of penicillin and streptomycin was used as a positive control and DMSO at 10% was used as negative control. The cells were incubated for 24, 48, or 72 h at 37 °C and in 5% CO2. At each time point, the medium was removed, and cells were washed with 200 µL of phosphate buffered saline (PBS, Thermofisher, Walthman, MA, USA). To assess the percentage of viable cells, 100 µL DMEM and 20 µL of the reagent of Cell Titer 96 Aqueous One Solution Cell Proliferation Assay (Promega, Madison, WI, USA) were added to each well and incubated for 2 h at 37 °C and then the absorbance was measured at 490 nm.
2.8. In Vitro Antimicrobial Susceptibility Assay
To assess the minimal inhibitory concentration (MIC) and the minimal bactericidal concentration (MBC) of CPX, S. aureus ATCC 25923 was cultured to mid-logarithmic growth phase in tryptic soy broth (TSB) (Difco Laboratories Inc., New York, NY, USA) at 37 °C and 120 rpm (Heidolph Instrument, Schwabach, Germany) and diluted to approximately 1 × 106 CFU/mL in TSB, based on the optical density of the suspension at 620 nm. From this bacterial inoculum, 10 μL was added to 90 µL of TSB containing CPX which had been heated at 180 °C for 5 min and cooled to room temperature, or non-heated CPX (control). CPX was serially diluted from 128 µg/mL to 0.125 µg/mL in TSB and cultured with the bacteria in 96-well plates with flat bottom (Greiner bio-one, Monroe, NC, USA) overnight at 37 °C, 120 rpm, in a box with a humidified atmosphere. As a control for bacterial growth, 10 µL of the inoculum was incubated in TSB without CPX. The MIC is defined as the lowest CPX concentration without visual growth. For the MBC, 2 × 10 µL of the undiluted samples, from the wells without visible growth and the well with the highest concentration that had visible growth (as a growth control), were plated on blood agar (Oxoid, Basingstoke, UK) and cultured at 37 °C overnight. The MBC is defined as the lowest CFX concentration that killed ≥99.9% of bacteria.
2.9. In Vitro Drug Release Assay
In order to measure the amount of CPX released from the extruded tubes both drug-loaded (n = 3) and non-loaded (n = 3), TPU tubes were cut into 3 mm length and weighted on a microbalance. Then, each tube was placed in a 1.5 mL Eppendorf tube (SafeSeal tube, Sarstedt Ag&Co, Nümbrecht, Germany) with 500 µL of PBS (Merck, PA, USA) and incubated at 37 °C and 120 rpm. The solution was changed at every time point (1 h, 6 h, 1 d, 2 d, 7 d, 14 d, 18 d, 24 d, 30 d, and 37 d). To measure the CPX concentration at each time point, 300 µL aliquots from each time point were placed in a well of a 96-well plate with flat bottom (Greiner bio-one, Monroe, NC, USA) and the absorbance at 330 nm was measured using a multi-well plate reader (Synergy H1, Biotek, Winooski, VT, USA). A calibration curve was plotted for CPX to estimate the concentration of the drug released from the tubes. This curve ranged from 1 to 50 µg/mL with R2 equal to 0.9998. The cumulative drug release was calculated based on the total loading amount (5% w/w) present in the tubes.
2.10. Agar Diffusion Assay
A modified Kirby–Bauer agar diffusion assay [
24] was performed to determine the zone of inhibition (ZOI) of the 5% (
w/w) CPX-loaded TPU and non-loaded TPU extruded tubes against
S. aureus ATCC 25923.
S. aureus was cultured to mid-logarithmic growth phase, in TSB at 37 °C and 120 rpm and diluted to 1 × 10
6 CFU/mL. Two hundred microliters of the inoculum were spread on blood agar plates under sterile conditions. The 3 mm tube segments were inserted vertically in the inoculated agar plates in triplicates and the agar plates were incubated at 37 °C overnight. The next day, the tubes were transferred onto freshly inoculated blood agar plates; this step was repeated for 10 days. The resulting zones of growth inhibition were measured in mm.
2.11. In Vitro Bacterial Adhesion Assay
An in vitro bacterial adhesion assay was performed to quantify the numbers of bacteria that adhered to the extruded tubes. S. aureus was cultured to mid-logarithmic growth phase in TSB at 37 °C and 120 rpm and diluted in TSB to 1 × 106 CFU/mL. The extruded tubes were incubated in 0.5 mL of this suspension in 1.5 mL Eppendorf tubes overnight at 37 °C and 120 rpm. The tubes were gently washed twice with demi water, placed in 1.5 mL Eppendorf tubes with 500 μL of PBS, vortexed 30 s and sonicated at 35 kHz for 15 min in a water bath sonicator (Elma Transsonic T460, Elma, Singen, Germany). This procedure does not affect the viability of the bacteria but releases them from the surface. The sonicates were serially diluted and the number of viable bacteria was determined by quantitative culture on blood agar plates.
2.12. Scanning Electron Microscopy (SEM)
SEM was performed to visualize the bacteria that adhered to the inside of the tubes. The set up was the same as in
Section 2.11 until the 2 washing steps with demi water. Prior to SEM, samples were fixed in 4% (
v/v) paraformaldehyde and 1% (
v/v) glutaraldehyde (Merck, Kenilworth, NJ, USA) overnight at room temperature. Samples were rinsed twice with demi water for 10 min and dehydrated in a graded ethanol concentration series from 50% to 100% of ethanol. To reduce the sample surface tension, samples were immersed in hexamethyldisilazane (Polysciences Inc., Warrington, FL, USA) overnight and air-dried. Before imaging, samples were mounted on aluminum SEM stubs and sputter-coated with a 4 nm platinum–palladium layer using a Leica EM ACE600 sputter coater (Microsystems, Wetzlar, Germany). Images were acquired at 3 kV using a Zeiss Sigma 300 SEM (Zeiss, Oberkochen, Germany) at the Electron Microscopy Center Amsterdam (Amsterdam UMC, Amsterdam, the Netherlands). Of each tube, 8–10 fields were viewed and photographed at magnifications of 100× and 500×.
2.13. Preparation of the Pre-Incubated Tube and Determination of Bacterial Adhesion
After overnight growth, S. aureus ATCC 25923 cultures were centrifuged at 5251× g for 10 min and the resulting pellet was resuspended in 10 mL of TSB resulting in a suspension of approximately 1 × 108 CFU/mL, the maximum start inoculum possible, in order to ensure successful establishing lung infection. The bacterial concentration was confirmed by viable plate count. Tube segments of 3 mm in length were incubated in 1 mL of this suspension for 4 h. The viability of the bacteria that adhered to the tubes was measured. For this assay, each tube was rinsed in TSB to remove planktonic bacteria and placed in 1 mL of sterile 0.5% Tween 20—TSB solution. For bacterial detachment, the tubes were vortexed for 20 s, sonicated for 5 min at 35 kHz at 20 °C, and vortexed for 20 s again. The number of viable bacteria was determined by quantitative culture of this sonicate.
2.14. Validation of the Anti-Infective Effect In Vivo
Twenty female SWISS-CD1 mice (Janvier Labs, Le Genest-Saint-Isle France) of approximately 25 g and 6 weeks old were kept in individually ventilated cages with HEPA filters as a barrier system, a constant temperature of 20–25 °C, average humidity of 50–60%, 12 h dark–light cycle, and food and drinking water ad libitum. Experiments were conducted when the mice were older than 12 weeks and approximately 45 g since at this age mice are considered adults [
25]. This model was developed for
S. aureus ATCC 25923 infection, since this bacterial strain is virulent enough to cause a chronic infection, without leading to extensive burden to the mice.
TPU tube segments were pre-incubated as described in the previous section and kept in an Eppendorf tube with 1 mL of PBS until insertion in the right main bronchus of the mouse by tracheotomy. The TPU-CPX tubes and TPU tubes were placed in the right main bronchus of the lungs, according to the protocol described by Yanagihara et al. [
26] that has been used to study the treatment of
P. aeruginosa endotracheal tube-related infection, with the addition of tracheotomy, as described by Facchini et al. [
27].
In brief, mice were anesthetized with 50 mg/mL ketamine (Nimatek 100 mg/mL, Dechtra Veterinary Products NV, Lille, Belgium) and 5 mg/mL xylazine (Rompum 2%, Bayer SA-NV, Machelem, Belgium) administered at 0.002 mL/g bodyweight in 0.9% saline solution by intraperitoneal injection. Then, the mice were placed in a supine position and the coat was disinfected with 70% ethanol. The trachea was exposed by a vertical cut of the skin and an opening was created in the exposed trachea with a 22G needle. Then, the tube was inserted through this opening with the help of a sterile, flexible 22G 0.9 mm × 25 mm intravenous catheter (Becton Dickinson Medical, Temse, Belgium) with a blunted needle and its outer sheath (
Figure 3A) and the attached tube at the tip. The tube was advanced into the trachea until it reached the desired location in the right main bronchus. The inner needle was then pulled out, followed by a gentle push of the outer sheath to place the tube into the main bronchus (
Figure 3B–D). The infected mice were observed for clinical signs (i.e., coat quality, posture, ambulation, and hydration) after surgery and had had unlimited access to feed and water. Body weight was monitored daily since mice that lose ≥20% bodyweight are considered in morbid status and must be euthanized.
After 3 days, the animals were euthanized by cervical dislocation, the left lung excised, weighed, and homogenized in 1 mL of PBS to determine the bacterial burden (CFU/g). For that, 100 μL of the homogenate was 10-fold serially diluted in PBS, and quadruplicate 10 µL droplets of each dilution were plated onto triple soy agar plates and incubated at 37 °C overnight. The bacterial burden was expressed per gram of lung tissue. The right lung was fixated in 4% paraformaldehyde for 24 h and kept in 60% isopropanol until the moment of processing for histology. Sections of murine lungs were stained with hematoxylin and eosin (H&E) and inflammation was scored by 2 independent observers at 400× magnification using an EVOS FL Auto microscope (Life Technologies, Belgium), according to Cigana et al. [
28]. Briefly, neutrophil and macrophage counts classified the infection as absent (score of 0) when there were no or fewer than 19 cells per field, mild (score of 1) for 20 to 49 cells per field, moderate (score of 2) for 50 to 99 cells per field, and severe (score of 3) for 100 or more per field.
2.15. Statistical Analysis
Statistical analysis was performed with Graphpad prism 8 (GraphPad Software Inc., La Jolla, CA, USA.). Unpaired
t-tests were performed in
Section 2.11,
Section 2.13 and
Section 2.14. *** indicates
p < 0.001 and * indicates
p < 0.05.
4. Discussion and Conclusions
AM technologies are fundamental to produce an extrusion nozzle that allows us to hot melt extrude drug-loaded tubes to be used in a VAP mouse model. The production of micro tubing for medical applications has been intensively explored in the scientific community and the industry for preclinical use in small animal models and clinical use in chemotherapy and for neonates [
33]. Typically, tubes are manufactured via hot melt extrusion techniques, allowing either single or multiple lumen(s) with outer diameters starting from 100 μm [
33]. The nozzle designing process should take into consideration manufacturability with conventional processes and ease of assembly. It is pivotal to have a good understanding of the polymer’s melt behavior during extrusion, so that a tube with the required dimensions is produced [
34]. Most extrusion systems are equipped with a channel to supply air within the lumen of the extruded tube, to avoid its collapse and adjust its final dimensions [
35]. The manufacturability of extrusion nozzles with novel designs is still a limitation as subtractive technologies such as machining are mostly used. This implies the need for designing the nozzle in several parts that should be assembled during installation [
36,
37].
The advancement in AM technologies allows the production of highly complex designs with internal geometries that were not possible using conventional techniques. It permits the design and production of the nozzle as one single part and reduces the manufacturing cost. In our work, we have successfully produced a nozzle with a specially designed internal geometry for extruding tubes with a single lumen. Moreover, our design proves that an air stream within the tube’s lumen while extruding can be eliminated without loss of the internal cavity. Jin et al. [
37] optimized the design of an extrusion nozzle based on a conventional design and manufacturing approach. Their nozzle assembly consisted of two parts: the nozzle (die) with an output diameter of 1.0 mm and a mandrel with an outer diameter of 0.7 mm. The mandrel section included a central hole with a diameter of 0.4 mm for supplying air flow to the inner cavity of the extruded tube during the extrusion process. When comparing their design with ours, the concentric rod in the middle of our nozzle’s output represents the mandrel part. Our nozzle has 1.0 mm and 0.5 mm outer and inner diameters, respectively, which represent the 1.0 mm and 0.7 mm ones in their design, respectively. In their experiments, an outer diameter that ranged between 0.6 mm and 1.6 mm and an inner diameter that ranged between 0.4 mm and 1.0 mm were achieved. The main technology that allowed these small diameters was the use of a pulling device. The purpose of this device was to pull the extruded tube before its total solidification. In our work, we produced tubes with an inner and outer diameter of 0.51 ± 0.03 mm and 0.90 ± 0.06 mm, respectively. These small dimensions, which were required for the VAP animal model, were possible without the need of a pulling device.
The aim of our study was to develop a test system for evaluating novel antimicrobial compounds in a VAP mouse model. Therefore, the tubes produced with this novel design were loaded with CPX, a well-known antibiotic for treating
S. aureus VAP. The advantages of incorporating CPX in the tubes are its thermal stability and low cost. Additionally, incorporating CPX at 5% (
w/w) did not alter the original mechanical properties of TPU fibers. Although fibers are solid and tubes are hollow, the fact that the stiffness of the bulk polymer in the fibers did not change, means that the overall stiffness of the tubes will likely not be affected by the CPX addition [
38]. From a clinical perspective, CPX should not be used prophylactically, since it is the drug of choice for treatment of VAP in patients [
16]. The tubes were not cytotoxic, and their antibacterial properties were demonstrated by zone of inhibition and quantification of the planktonic bacteria. Their antibiofilm properties were also confirmed by quantification and visualization of the attached bacteria.
Animal models of device-related infections can be divided into site-specific and subcutaneous models. The VAP mouse model is a site-specific model and allows the evaluation of the host response to endotracheal tubes located in the same position as in the clinic [
39]. Here, AM technologies allowed us to produce test samples that fit in the main bronchus of the mouse. Other animal species have been used to study biofilm-related infections in endotracheal tubes, such as sheep, pigs, and dogs [
40,
41,
42]. Mice have the advantages of their small size, relatively high reproductive rate, and the availability of antibodies, increasing the chance to bring new technologies and drugs from bench to bedside. We used fully mature adult mice to ensure that the tubes fit in the trachea. Outbred SWISS-CD1 mice are used in all areas of biomedical research and are more resistant to infections than inbred mice as reported in a
P. aeruginosa VAP model [
25].
S. aureus is the most common cause of VAP among Gram-positive bacteria [
7,
43] and the PRINT-AID consortium has identified antibiofilm leads against
S. aureus that could be incorporated into endotracheal tubes [
44]. The challenge to establish a VAP mouse model lies not only in the technical, but also in the biological aspects. The selection of a bacterial strain is fundamental for the success of murine models [
39]. The benefit of our animal model is the relative ease to establish a chronic infection (≥3 days) [
45] with
S. aureus ATCC 25923. The limitation is that the delivery of bacteria at implantation of the device is initially different between the CPX and the TPU tubes, since the bacterial viable numbers are reduced during the pre-incubation step due to eluting CPX. However, in the clinical situation an endotracheal tube is placed in the trachea in contact with the fluid lining of the tracheal epithelium. In this wet environment, CPX elution will also start at the same time as the first contaminating bacteria will reach the tube. Thus, it might be argued that the exposure of the bacteria to CPX in the pre-incubation step represents the situation that may occur in vivo. In future studies, it would be interesting to focus on pathophysiology by inserting non-colonized tubes and subsequently exposing mice to the bacteria. Other bacterial strains and even species could be used for this purpose, such as
P. aeruginosa, since chronicity of infection has been previously obtained with the appropriate strain and methodology [
26,
46].
Although CPX-loaded tubes will not be used in the clinic, they were highly functional to validate our in vivo VAP model. The tubes were successfully manufactured using the hot melt extrusion process and the novel nozzle. The developed approach can be adopted for investigating other promising antibacterial and antibiofilm compounds and formulations incorporated in experimental endotracheal tubes. Moreover, our newly developed nozzle may be used for producing micro tubing in the medical and veterinary fields.