4.1. Simulation Scenario and Evaluation Metrics
To verify the performance of multi-frame sequential priors on HRRP reconstruction under CIS modulation, a continuous observation scenario of an airborne target by a broadband radar was simulated. The radar operates in the X-band with a transmitted signal bandwidth of
B = 500 MHz, yielding a range resolution of
. The range dimension was discretized into
N = 256 cells, corresponding to a range window of 76.8 m. Under short-time continuous observation, the dominant scattering centers exhibit only small frame-to-frame variations. The position of the
i-th scattering center is modeled as
where a position change occurs with probability 0.35 and is limited to one range cell (0.3 m) between adjacent frames. The amplitudes follow a first-order smooth evolution with a smoothing coefficient of 0.90 and a perturbation magnitude of 0.05, while the phases follow a random walk with an increment standard deviation of 0.08 rad. These settings produce correlated but nonidentical HRRP profiles across frames.
A five-point scattering-center model of an aircraft is assumed. The model parameters are listed in
Table 1.
To quantify HRRP reconstruction performance, the normalized mean squared error (NMSE) and tolerance-support recovery rate (Tol-SRR) were adopted as the primary metrics, with the peak-to-sidelobe ratio (PSLR) as an auxiliary metric. NMSE measures the overall error between the reconstructed HRRP and the true HRRP and is defined as
where
is the true HRRP of the
m-th frame, and
is the corresponding reconstruction result. A smaller NMSE indicates that the reconstructed result is closer to the true HRRP.
Tol-SRR evaluates the recovery of scattering-center locations. Let
be the set of true scattering-center locations,
the estimated set, and
the position tolerance. A true scattering center is considered correctly recovered if an estimated location exists within δ range cells. Tol-SRR is defined as
where
is the indicator function. A higher Tol-SRR indicates better recovery of true scattering-center locations.
The exact scattering-center recovery rate (Exact-SRR) is also defined as a variant of Tol-SRR with zero position tolerance (
δ = 0). Exact-SRR measures the fraction of true scattering centers whose recovered positions exactly match the true positions. While Exact-SRR is more stringent, it is sensitive to discretization effects and is used primarily for parameter sensitivity analysis (
Section 4.5). For general performance evaluation, Tol-SRR with
δ = 1 range cell is adopted.
To further measure the separability between true scattering centers and sidelobe or false peaks, the target-peak region is defined as the neighborhood of the true scattering centers.
Based on this region, the PSLR of the reconstructed HRRP in the
m-th frame is defined as
where
is the reconstructed amplitude at the
q-th range cell, and
is a small positive constant to avoid division by zero. A higher PSLR indicates that the target peaks are more prominent relative to non-target peaks.
To evaluate the search efficiency of CI-OMP, the average search cardinality ratio (ASCR) is adopted. Let
denote the set of range cells searched during the
t-th iteration of the
m-th frame,
the number of iterations for the
m-th frame, and
the number of continuously observed frames. Then,
where
is the total number of range cells. For standard OMP, at the
t-th iteration, the search space excludes the already selected
atoms, so its cardinality is
. Thus, the denominator in (25) exactly equals the total search cost of standard OMP, yielding
for standard OMP. For CI-OMP, a smaller
indicates a greater reduction in the search scale achieved by the candidate-interval strategy. Equivalently, this corresponds to a search space compression of
compared with standard OMP.
All reported results are averaged over Monte Carlo trials. Error bars in the figures denote the 95% confidence interval, computed using the paired-sample t-distribution.
4.2. Reconstruction Performance Under Different Duty Ratios
To evaluate the influence of the duty ratio on HRRP reconstruction under CIS, the SNR was fixed at 0 dB, and the duty ratio was varied from 0.08 to 0.25. This low-to-medium duty-ratio range was selected to represent the challenging CIS conditions relevant to LPI enhancement, under which only a small proportion of the echo samples are retained. In this regime, the reduced number of valid samples makes conventional sparse reconstruction less reliable, thereby allowing the benefit of the multi-frame sequential prior to be evaluated more clearly. The target HRRP contained five dominant scattering centers at range cells 70, 95, 125, 160, and 198. For CI-OMP, the historical-window length, candidate-interval half-width, and Tol-SRR tolerance were set to , , and range cell, respectively. Matched filtering (MF), standard OMP, a warm-started OMP (WS-OMP) baseline, and CI-OMP were compared using representative reconstruction results and Monte Carlo statistics. WS-OMP restricts atom search to the neighborhood of the previous frame’s support and serves as a single-frame prior baseline to isolate the benefit of multi-frame statistics. MF directly applies pulse compression to the interrupted echoes without sparse reconstruction and is included as a conventional radar-processing baseline to demonstrate the benefit of sparse reconstruction in suppressing sidelobes and false peaks.
Figure 5 presents representative HRRP reconstruction results under different duty ratios. In the figure, the gray stems denote the matched-filtering output, the black markers indicate the true scattering-center locations, and the blue and red stems represent the reconstructed HRRPs obtained by OMP and CI-OMP, respectively. At low duty ratios, only a small portion of the echo samples is retained, resulting in pronounced sidelobe and false-peak components in the matched-filtering output. In this case, standard OMP tends to select locally strong false peaks, leading to missed scattering centers or range-position deviations. In contrast, CI-OMP provides more stable reconstructions, with the recovered peaks more concentrated around the true scattering-center locations and fewer obvious false peaks.
As the duty ratio increases, more valid echo samples become available, and the reconstruction performance of both OMP and CI-OMP improves. However, CI-OMP still exhibits better agreement with the true HRRP. In the medium-duty-ratio cases, OMP can recover part of the true scattering structure, but residual false peaks and position offsets remain. When the duty ratio reaches and , the reconstructed peaks of CI-OMP are highly consistent with the true scattering-center positions, indicating improved localization stability and peak separability.
To further verify the statistical reliability of the visual results, Monte Carlo experiments were conducted under different duty ratios. The results were averaged over 100 independent trials. The error bars in
Figure 6 represent 95% confidence intervals, confirming the statistical significance of the observed performance differences.
Figure 6 presents the NMSE, Tol-SRR, and PSLR results of MF, OMP, WS-OMP, and CI-OMP.
As presented in
Figure 6a, the NMSE values of all four methods, expressed in decibels, generally decrease as the duty ratio increases because more valid echo samples become available for HRRP reconstruction. At extremely low duty ratios, MF exhibits a lower NMSE than OMP, WS-OMP, and CI-OMP. In this severely undersampled regime, sparse support estimation becomes unstable, and incorrect atom selection may produce large localization and amplitude errors. In contrast, although the MF output contains broadened peaks and high sidelobes, its energy remains distributed around the true scattering-center locations, which can result in a smaller global NMSE. However, this does not indicate more accurate scattering-center recovery, as reflected by its lower Tol-SRR and PSLR in
Figure 6b,c. As the duty ratio increases, sparse reconstruction becomes more reliable, and both WS-OMP and CI-OMP achieve lower NMSE than standard OMP, with CI-OMP achieving the lowest NMSE owing to the use of multi-frame sequential priors. Specifically, as
increases from 0.08 to 0.25, the NMSE of OMP decreases from 2.4211 dB to 0.2633 dB, that of WS-OMP decreases from 2.1548 dB to 0.2011 dB, whereas that of CI-OMP decreases from 2.1440 dB to 0.1791 dB.
Figure 6b compares the Tol-SRR of all methods. CI-OMP consistently achieves the highest Tol-SRR, followed by WS-OMP, with standard OMP exhibiting the lowest Tol-SRR. Over the tested duty-ratio range, the Tol-SRR of OMP increases from 0.2029 to 0.8524, that of WS-OMP increases from 0.3280 to 0.9019, while that of CI-OMP increases from 0.2937 to 0.9171. The improvement is more evident in the medium-to-low duty-ratio region. For example, at
and
, CI-OMP improves Tol-SRR over OMP by 0.1445 and 0.1356, respectively, and over WS-OMP by 0.0879 and 0.0821, respectively. These results demonstrate that CI-OMP is more robust in locating true scattering centers than both standard OMP and WS-OMP when the current-frame echo samples are insufficient.
Figure 6c presents the PSLR results. CI-OMP achieves higher PSLR than both OMP and WS-OMP under different duty ratios. At
, the PSLR of OMP is 18.51 dB, that of WS-OMP is 21.38 dB, which is 7.56 dB higher than that of OMP. This result is consistent with the visual comparison in
Figure 5, where CI-OMP produces cleaner peak structures and fewer false peaks. When the duty ratio further increases to
, the performance gap between OMP, WS-OMP, and CI-OMP begins to decrease because the current-frame observations become increasingly sufficient for reliable reconstruction by standard OMP. Nevertheless, CI-OMP still maintains lower NMSE, higher Tol-SRR, and higher PSLR over the tested duty-ratio range.
Table 2 summarizes the performance gains of CI-OMP over standard OMP and WS-OMP at representative duty ratios. All reported improvements are statistically significant at the 0.05 level based on paired-sample 95% confidence intervals, confirming that the observed performance gains are not due to random variation. Compared with standard OMP, CI-OMP improves NMSE by 0.78–1.71 dB, increases Tol-SRR by 0.0647–0.1445, and improves PSLR by 4.57–7.56 dB. Compared with WS-OMP, CI-OMP improves NMSE by 0.82–3.34 dB, increases Tol-SRR by 0.233–0.335, and improves PSLR by 2.95–19.12 dB. The gains are most evident in the medium-duty-ratio range. For example, at
, CI-OMP improves the NMSE by 1.71 dB over OMP and by 3.06 dB over WS-OMP, and improves PSLR by 7.56 dB over OMP and by 16.43 dB over WS-OMP. These quantitative results confirm that CI-OMP improves both HRRP reconstruction accuracy and scattering-center localization when the available echo samples are sufficient to support meaningful reconstruction.
As the duty ratio continues to increase, the current-frame observations become increasingly sufficient for reliable support recovery, and the additional benefit provided by the multi-frame sequential prior is expected to gradually diminish. In the limiting case of , the difference between OMP, WS-OMP, and CI-OMP is expected to approach zero because all three methods rely on increasingly complete current-frame observations.
Overall, the duty-ratio experiments demonstrate that CI-OMP provides more accurate and stable HRRP reconstruction than both standard OMP and WS-OMP, particularly at low-to-medium duty ratios. The consistent improvements in NMSE, Tol-SRR, and PSLR verify the effectiveness of the proposed candidate-interval strategy, which outperforms the simpler warm-start approach by leveraging stable scattering-center priors extracted from a multi-frame historical window rather than directly propagating the previous frame’s support.
4.3. Reconstruction Performance Under Different SNRs
To evaluate the noise robustness of the proposed method, the duty ratio was fixed at
, and the SNR was varied from −8 dB to 5 dB. The same target HRRP and CI-OMP parameters as in
Section 4.2 were used.
Four methods were compared: MF, standard OMP, WS-OMP, and CI-OMP. The comparison isolates the benefit of different strategies for incorporating multi-frame sequential priors under varying noise levels.
Figure 7 presents representative HRRP reconstruction results under different SNRs;
Figure 8 presents the statistical results in terms of NMSE, Tol-SRR, and PSLR, and
Table 3 summarizes the performance gains of CI-OMP over standard OMP and WS-OMP at representative SNR levels. For the Monte Carlo evaluation, the performance was calculated from frames 6 to 20 to avoid the initial transient stage in which historical information is insufficient, and the results were averaged over 100 independent trials.
As presented in
Figure 7, at the very low SNR of −5 dB, neither OMP nor CI-OMP can completely recover the true HRRP structure. Both methods exhibit missed detections and incorrect peak selections, indicating that the current observations are too severely degraded to support reliable reconstruction. Nevertheless, CI-OMP recovers more true scattering centers and produces fewer prominent false peaks than OMP, showing a limited but observable improvement under this extremely noisy condition.
When the SNR increases to 0 dB, the historical sequential prior employed by CI-OMP begins to play an effective role, and the difference between the two methods becomes pronounced. Standard OMP still selects several non-target range cells and misses some true scattering centers. In contrast, CI-OMP successfully recovers all five scattering centers, achieving complete support recovery, and its reconstructed peaks are well aligned with the true scattering-center locations.
At an SNR of 5 dB, the current-frame observation becomes sufficiently reliable, and the true scattering-center responses can be effectively distinguished from the sidelobe and false-peak components in the MF output. Consequently, both OMP and CI-OMP recover all five scattering centers, and their reconstructed HRRP profiles are essentially identical. Overall,
Figure 7 shows that CI-OMP offers its most significant advantage at low SNR, provides only limited improvement when both methods fail at very low SNR, and converges to standard OMP at sufficiently high SNR.
As shown in
Figure 8a, the NMSE of all four methods decreases as the SNR increases, indicating that the HRRP reconstruction error is reduced when noise interference becomes weaker. The error bars represent 95% confidence intervals, demonstrating the statistical reliability of the observed performance differences. MF exhibits relatively high reconstruction error over most SNR conditions because it cannot effectively suppress the sidelobe and false-peak components introduced by interrupted sampling. Compared with standard OMP, both WS-OMP and CI-OMP achieve lower NMSE in the low-to-medium SNR range. CI-OMP consistently achieves the lowest NMSE across all tested SNRs. As summarized in
Table 3, the NMSE reductions of CI-OMP relative to OMP are 1.61 dB and 2.21 dB at −2 dB and 0 dB, respectively, and relative to WS-OMP are 2.41 dB and 4.28 dB at −2 dB and 0 dB, respectively. This confirms that CI-OMP improves reconstruction accuracy when the current-frame matching response is disturbed by noise.
Figure 8b compares the Tol-SRR results. CI-OMP achieves higher Tol-SRR than both OMP and WS-OMP from −8 dB to 3 dB, with the most noticeable improvements appearing at −5 dB, −2 dB, and 0 dB. In this SNR range, noise-induced peaks and sidelobe peaks may have amplitudes comparable to those of true scattering centers, making both standard OMP and WS-OMP more likely to select incorrect range cells. WS-OMP improves over standard OMP by initializing with the previous frame’s support, but it does not fully suppress false selections when the previous frame’s prior is inaccurate. As the SNR increases beyond 3 dB, the Tol-SRR values of OMP, WS-OMP, and CI-OMP gradually converge because the current-frame observation alone becomes sufficient for stable scattering-center localization.
Figure 8c presents the PSLR results. CI-OMP achieves higher PSLR than both OMP and WS-OMP under low-to-medium SNR conditions. According to
Table 3, the PSLR improvements of CI-OMP over OMP are 12.81 dB, 4.17 dB, and 8.01 dB at −5 dB, −2 dB, and 0 dB, respectively, and the PSLR improvements of CI-OMP over WS-OMP are 7.17 dB, 7.23 dB, and 13.63 dB at −8 dB, −2 dB, and 0 dB, respectively. These results are consistent with the representative HRRP reconstructions in
Figure 7, where CI-OMP produces cleaner peak structures and fewer non-target peaks. When the SNR increases to 3 dB and 5 dB, the PSLR improvement becomes smaller because standard OMP can already distinguish the main scattering centers from the background peaks under relatively high-quality observations.
Table 3 summarizes the performance gains of CI-OMP over OMP and WS-OMP at different SNRs. All reported improvements are statistically significant at the 0.05 level based on paired-sample 95% confidence intervals, confirming that the observed performance gains are not due to random variation. CI-OMP generally improves NMSE, Tol-SRR, and PSLR, with its clearest overall advantage appearing at low-to-medium SNRs. The maximum NMSE and Tol-SRR improvements over OMP are 2.21 dB at 0 dB and 0.142 at −2 dB, respectively. Compared with WS-OMP, CI-OMP achieves even larger gains: the maximum NMSE improvement is 4.28 dB at 0 dB, and the maximum Tol-SRR improvement is 0.358 at −2 dB. Although the PSLR improvement reaches 18.29 dB at −8 dB, the low absolute Tol-SRR values at −8 and −5 dB indicate that gains under these extremely noisy conditions should be interpreted as relative metric improvements rather than reliable recovery of the complete scattering-center structure. At high SNRs, the performance gap narrows as current-frame observations become more reliable, consistent with
Figure 8.
Overall, the SNR experiments demonstrate that CI-OMP outperforms both OMP and WS-OMP under low-to-medium SNR conditions, with consistent improvements in NMSE, Tol-SRR, and PSLR. The candidate-interval strategy, which leverages multi-frame historical priors, proves more effective than both the uninformed greedy search of OMP and the single-frame warm-start of WS-OMP. The advantage is most pronounced when noise-induced false peaks interfere with atom selection and diminishes at high SNRs when current-frame observations become sufficient.
4.4. Search Efficiency and Computational Complexity Analysis
The search efficiency of CI-OMP is evaluated using the ASCR, where standard OMP serves as the baseline with an ASCR of one. A lower ASCR indicates a smaller atom search scale and thus higher search efficiency. The search-reduction ratio is calculated as , where denotes the ASCR of CI-OMP.
Figure 9 presents the search-efficiency comparison between standard OMP and CI-OMP under different duty-ratio and SNR conditions. According to
Figure 9a, standard OMP maintains an ASCR of one because it searches over the full range-cell domain at each iteration. In contrast, CI-OMP consistently achieves a much lower ASCR by restricting atom matching to candidate range intervals constructed from historical scattering-center supports. As the duty ratio increases from 0.10 to 0.25, the ASCR of CI-OMP decreases from approximately 0.46 to 0.25. Accordingly, the atom-search-range reduction increases from approximately 54% to 75%, as illustrated in
Figure 9b. For example, at
, CI-OMP achieves an ASCR of approximately 0.30, corresponding to an atom-search-range reduction of approximately 70%. These results indicate that the proposed candidate-interval strategy can effectively compress the scattering-center search range under different sampling conditions.
The SNR-dependent results in
Figure 9c,d further verify the search-efficiency advantage of CI-OMP. With
fixed at 0.20, the ASCR of CI-OMP decreases from approximately 0.50 at an SNR of −10 dB to approximately 0.17 at 6 dB. The corresponding atom-search-range reduction therefore increases from approximately 50% to 83%, as presented in
Figure 9d. Even under the severe noise condition of −10 dB, CI-OMP still reduces nearly half of the atom search scale. At an SNR of 0 dB, the atom-search-range reduction is approximately 70%, and it reaches approximately 83% at 6 dB. This trend indicates that more reliable historical scattering-center priors allow CI-OMP to concentrate the atom search on a narrower and more informative range region.
From a computational-complexity perspective, ASCR reflects the reduction in the dominant atom-search cost. In terms of the atom-search stage, which dominates the overall computational cost of OMP-based reconstruction, the search complexity is reduced from approximately in standard OMP to in CI-OMP, excluding occasional fallback searches. The least-squares projection and residual update stages remain unchanged between the two methods.
4.5. Parameter Sensitivity and Ablation Analysis
The sensitivity of CI-OMP to its three principal parameters—the historical-window length L, the candidate-interval half-width Δ, and the historical-prior weight α—was investigated under a challenging coded interrupted-sampling condition. The HRRP dimension and sparsity level were set to N = 256 and K = 5, respectively. Each 25-frame sequence was tested at ρ = 0.20 and SNR = 0 dB. All results were averaged over 100 paired Monte Carlo trials using identical target realizations, noise samples, and sampling patterns for OMP and CI-OMP. During each one-factor-at-a-time scan, the remaining two parameters were fixed at L = 5, Δ = 4, and α = 1.
Regarding the ablation of the candidate-interval strategy, WS-OMP provides an effective ablation. WS-OMP reuses the previous frame’s support without constructing candidate intervals or employing historical weighting. The three methods form a progressive ablation hierarchy: standard OMP (no prior), WS-OMP (single-frame persistence without intervals), and CI-OMP (multi-frame intervals with weighting). The performance gap between WS-OMP and CI-OMP, as demonstrated in
Table 2 and
Table 3, quantifies the candidate-interval strategy’s contribution. The parameter sensitivity analysis below examines how
L, Δ, and
α influence CI-OMP’s performance.
Figure 10 presents the sensitivity of NMSE and Exact-SRR to the three parameters. Increasing
L from 1 to 5 reduces the NMSE from −5.31 to −5.67 dB and increases the Exact-SRR from 0.7757 to 0.7976. Further increasing
L provides only marginal improvement:
L = 10 achieves −5.68 dB, only 0.01 dB lower than
L = 5. This indicates that most useful historical information accumulates within approximately five frames. Therefore,
L = 5 is selected as a practical compromise between reconstruction accuracy and storage/processing costs.
The candidate-interval scan reveals a trade-off between amplitude reconstruction and strict support localization. Although Δ = 0 produces the highest Exact-SRR (0.8207), it yields a higher NMSE of −5.33 dB compared with Δ = 4, which achieves the lowest NMSE of −5.67 dB. Δ = 0 forces exact matching to historical scattering-center locations, maximizing the support hit rate when scattering centers remain stationary. However, scattering centers in our simulation migrate slowly due to aspect-angle changes (with a migration probability of 0.35 per frame, as described in
Section 4.1). When Δ = 0, these migrated scattering centers cannot be correctly localized, leading to amplitude estimation errors and consequently higher NMSE, despite the high Exact-SRR for stationary centers. In contrast, Δ = 4 provides sufficient tolerance to accommodate slow scattering-center migrations while still constraining the search space, achieving the best overall reconstruction fidelity. An excessively wide interval (e.g., Δ = 8) weakens the spatial constraint and allows more false-peak selections.
This result demonstrates that Exact-SRR alone is insufficient for evaluating reconstruction fidelity, and the optimal Δ = 4 represents the best trade-off between localization tolerance and false-peak suppression. Accordingly, Δ = 4 is selected for subsequent experiments.
The influence of
α further confirms the contribution of the historical weighting mechanism. When
α = 0, the historical weight is removed, but the search remains confined to the candidate intervals constructed from the multi-frame prior, thereby isolating the contribution of the weighting mechanism alone. As shown in
Table 4, increasing
α from 0 to 1 improves the NMSE from −3.15 to −5.67 dB and increases the Exact-SRR from 0.6906 to 0.8031, representing a 2.52 dB NMSE improvement attributable to the weighting mechanism. However, excessively large values of
α (e.g.,
α = 4) gradually degrade performance because an overly strong historical prior reduces the algorithm’s adaptability to current-frame variations.
Figure 11 provides a visual verification of these statistical trends. Compared with OMP, CI-OMP better preserves the dominant scattering-center locations and suppresses spurious background components. The
α = 0 case benefits only from the candidate-interval constraint, while
α = 1 fully leverages both the candidate-interval strategy and the historical weighting mechanism. Based on the combined quantitative and visual analyses,
L = 5, Δ = 4, and
α = 1 are adopted as the reference configuration for CI-OMP. Notably, these parameters are optimized for the current simulation conditions and may be adjusted under different operational scenarios, such as faster scattering-center migrations or more severe noise levels. Adaptive parameter-selection strategies will be explored in future work.