Positron Range Corrections and Denoising Techniques for Gallium-68 PET Imaging: A Literature Review

by
Prodromos Gavriilidis
1,2,3,
Michel Koole
3,
Salvatore Annunziata
4,
Felix M. Mottaghy
1,2,5,* and
Roel Wierts
1,2
1
Department of Radiology and Nuclear Medicine, Maastricht University Medical Center, 6229 HX Maastricht, The Netherlands
2
School for Oncology and Reproduction (GROW), Maastricht University, 6200 MD Maastricht, The Netherlands
3
Nuclear Medicine and Molecular Imaging, Katholieke Universiteit Leuven, 3000 Leuven, Belgium
4
Unit of Nuclear Medicine, TracerGLab, Department of Radiology, Radiotherapy and Hematology, Fondazione Policlinico Universitario A. Gemelli IRCCS, 00168 Rome, Italy
5
Department of Nuclear Medicine, RWTH University Hospital, D-52074 Aachen, Germany
*
Author to whom correspondence should be addressed.
Diagnostics 2022, 12(10), 2335; https://doi.org/10.3390/diagnostics12102335
Submission received: 8 August 2022 / Revised: 12 September 2022 / Accepted: 16 September 2022 / Published: 27 September 2022
(This article belongs to the Section Machine Learning and Artificial Intelligence in Diagnostics)

Abstract

:
Gallium-68 (68Ga) is characterized by relatively high positron energy compared to Fluorine-18 (18F), causing substantial image quality degradation. Furthermore, the presence of statistical noise can further degrade image quality. The aim of this literature review is to identify the recently developed positron range correction techniques for 68Ga, as well as noise reduction methods to enhance the image quality of low count 68Ga PET imaging. The search engines PubMed and Scopus were employed, and we limited our research to published results from January 2010 until 1 August 2022. Positron range correction was achieved by using either deblurring or deep learning approaches. The proposed techniques improved the image quality and, in some cases, achieved an image quality comparable to 18F PET. However, none of these techniques was validated in clinical studies. PET denoising for 68Ga-labeled radiotracers was reported using either reconstruction-based techniques or deep learning approaches. It was demonstrated that both approaches can substantially enhance the image quality by reducing the noise levels of low count 68Ga PET imaging. The combination of 68Ga-specific positron range correction techniques and image denoising approaches may enable the application of low-count, high-quality 68Ga PET imaging in a clinical setting.

1. Introduction

In recent years, 68Ga-labeled radiotracers have been increasingly used in clinical positron emission tomography (PET). The commercial availability of 68Ge/68Ga generators facilitates the in-house production of 68Ga-labeled radiotracers, with applications in oncological, cardiovascular, infection, and inflammation PET imaging [1,2,3,4]. Particularly, 68Ga-labeled somatostatin receptor PET imaging is extensively used for neuroendocrine tumor imaging while the 68Ga-labeled prostate-specific membrane antigen (PSMA) has been shown to provide valuable information for the clinical management of prostate cancer patients.
Compared to the most used PET radionuclide 18F, 68Ga emits high-energy positrons (68Ga: Emax = 1.899 MeV; 18F: Emax = 0.635 MeV) [5]. Consequently, 68Ga exhibits an increased positron range, which is the travelled distance between the location of the decaying parent nucleus and the location of positron annihilation. Therefore, the larger the positron range, the more the measured PET signal can become blurred. This degradation of the spatial resolution of PET images depends on the underlying tissue type with a more severe degradation for lower-density tissues [6,7,8,9,10]. Studies have confirmed a negative impact on the spatial resolution for radionuclides emitting high-energy positrons, while phantom experiment revealed lower recovery coefficients (RC) and a reduced quantitative accuracy for radionuclides emitting high-energy positrons [6,8,9,10,11,12].
As good manufacturing practice (GMP)-compliant 68Ge/68Ga generators and labeling kits are generally expensive, it is essential to maximize the cost effectiveness of 68Ga PET examinations. Therefore, in clinical practice, the reduction in both the administered activity and PET acquisition time is highly relevant to maximize patient throughput. Additionally, it is important from radiation exposure and patient comfort perspectives. However, this will result in PET acquisitions with lower count statistics and thus increased statistical noise. Along with spatial resolution, noise is another prominent factor that can degrade image quality, lesion detectability, and quantitative accuracy of PET images. In recent years, deep learning approaches have been successfully applied for denoising PET images acquired with lower administered activities or shorter acquisition times [13,14,15].
In the current paper, we aim to provide a literature review of recently developed software-based techniques to enhance the PET image quality and quantification by compensating for the positron range effect and reducing the noise, specifically for 68Ga-labeled radiotracers. The abbreviations used in this paper are summarized in Table 1.

2. Materials and Methods

A literature review was conducted using the search engines PubMed and Scopus. Different queries were created for the positron range correction techniques and denoising approaches. For the positron range correction, we created queries incorporating terms related to PET, 68Ga, and positron range, while for noise reduction, terms such as PET, 68Ga, noise, or low count were used. The queries of the positron range corrections techniques are presented in Table A1 and Table A2 in Appendix A for PubMed and Scopus, respectively. Table A3 and Table A4 in Appendix A describe the queries that were used to retrieve articles related to noise reduction from PubMed and Scopus, respectively. As we aim to provide an overview of recent literature, only articles published between January 2010 and 1 August 2022 were included. From the retrieved records, duplicates were removed, then title and abstract screening was performed. Lastly, full-text screening was performed. In Figure 1, the chart of the screening process is illustrated.

3. Results

The total number of research articles generated from the search queries was 1162, with 126 publications related to positron range correction and 1036 publications related to denoising. After the removal of duplicates and title and abstract screening, 38 articles in total remained and were screened based on the full-text. During this process, another 14 publications were excluded (n = 2: Positron range correction techniques, which have already been described in other included articles, n = 9: No noise reduction technique, n = 1: Not studied for 68Ga, n = 1: Not a novel technique, n = 1: No noise-related metric used for the analysis, where the term n depicts, in this case, the number of full-text articles excluded based on the corresponding reason). The total number of articles included in this literature review was 24, with 8 articles describing a positron range correction technique and 16 articles introducing a novel denoising technique. Figure 1 presents the workflow and results of the literature screening.
Four different categories of techniques were identified, consisting of articles reporting on reconstruction-based positron range correction (n = 5), post-reconstruction positron range correction (n = 3), reconstruction-based noise reduction (n = 11), and deep learning noise reduction (n = 5), as shown in Figure 2. First, positron range correction techniques will be described followed by image noise reduction techniques.

3.1. Positron Range Correction Techniques

Based on the included literature, a variety of approaches have recently been developed for 68Ga-specific positron range correction. These can be grouped into two approaches: (1) Incorporating positron range models in the reconstruction algorithm (reconstruction-based correction, n = 5) and (2) approaches applying post-reconstruction positron range correction (post-reconstruction correction, n = 3), as seen in Figure 2. Reconstruction-based corrections generally used blurring kernels to model the positron range during the iterative reconstruction process [16,17,18,19,20], while post-reconstruction corrections applied deblurring kernels or deep learning techniques on the reconstructed images [21,22,23].

3.1.1. Reconstruction-Based Correction

The reconstruction-based techniques correct for the positron range by incorporating positron range distribution models in the image reconstruction algorithm. In contrast to system response modeling techniques, in which position-dependent response modeling is performed based on scanner characteristics, positron range modeling typically requires patient-specific modeling depending on the underlying patient-specific tissue characteristics [6,7,8,9,10]. The reconstruction-based positron range corrections can be divided into three subcategories: Tissue-independent [16,17,18], homogeneous tissue-dependent [16,17,18,19], and heterogeneous tissue-dependent corrections [16,19,20]. Those corrections are applied in the forward projection step of the reconstruction algorithm, while a simplified positron range modeling is introduced in the back projection step, unless it is stated otherwise.

Homogeneous Tissue-Dependent Positron Range Correction

In clinical and preclinical practice, the scanned object comprises several different tissue types. As the positron range effect varies for different tissue types [6,7,8,9], a tissue-independent correction, as described in the previous section, is an oversimplification of the real situation. Therefore, techniques employing different blurring kernels based on different homogeneous tissues were developed [16,17,18,19]. Similar to the tissue-independent approach, the location of the decaying parent nucleus, the travelled distance of the positron, and the location of positron annihilation were assumed to occur in the same homogenous tissue; however, depending on the tissue (water, soft, lung, or bone tissue), different kernels were applied. For each tissue of interest, positron range distribution profiles of 68Ga were generated and used to create isotropic blurring kernels. For each voxel, a specific blurring kernel was applied depending on the respective tissue type of that voxel. The positron range distribution was obtained from Monte Carlo simulations or analytically [16,17,18,19].
Using phantom simulations, Bertolli et al. demonstrated that the activity recovery of the homogeneous tissue-dependent approach in lung tissue was approximately 90% of the simulated activity [17]. Compared to the non-corrected images, the homogeneous tissue-dependent approach achieved greater than double recovered activity for both lung and bone tissues [17,18]. Additionally, RC up to approximately 80% was obtained from reconstruction implementing homogeneous tissue-dependent correction, while the reconstruction without positron range correction achieved RC up to approximately 40% [18]. The tissue-dependent positron range correction also improved the full width half maximum (FWHM) up to 28% [18] and the full width tenth maximum (FWTM) up to 50% compared to the non-corrected reconstructions [16]. On top of that, PET data from mice injected with 68Ga-DOTATOC in a microPET/CT system revealed an increment in the tumor-to-background ratio (TBR) of approximately 80% compared to reconstruction without any correction [16]. Aside from that, it was demonstrated that implementation of the positron range kernels in both the forward- and backward projection steps resulted in similar image quality compared to the implementation of the kernel in only the forward projection step [18]. However, the first approach introduced a smoothing effect and reduced the convergence speed [18]. Simulating a PET/MRI scanner, Kraus et al. demonstrated that the tissue-dependent approach improved the image contrast up to approximately 50% and reduced the misplaced activity concentration in soft tissue compared to the reconstruction without positron range correction [19]. However, application of the tissue-dependent correction introduced Gibbs artifacts, noise was increased due to deblurring and even doubled in some cases, while the reconstruction time can also be doubled [16,17]. To achieve the same image noise between reconstructions with and without homogeneous tissue-dependent positron range correction, Cal-González et al. increased the number of image updates of the reconstruction without correction by a factor of 2.6 [16].

Heterogeneous Tissue-Dependent Positron Range Correction

It is possible that the decay of the parent nucleus and positron annihilation does not occur in the same tissue, meaning that the positron travels through different tissue types with different densities. Therefore, kernels assuming homogeneous tissues are only an approximation of the underlying situation. Consequently, more realistic heterogeneous tissue-dependent positron range corrections were developed [16,19,20].
Kraus et al. used Monte Carlo simulations to obtain the positron range distribution profiles for heterogeneous tissue configurations with different tissue borders [19]. Those simulated positron range data were used to create heterogeneous blurring kernels. This approach enhanced the image contrast to greater than 55% [19]. It further and further reduced the misplaced activity concentration in soft tissue compared to the homogeneous tissue approach [19]. Cal-González et al. also introduced a heterogeneous, tissue-dependent approach for PET/CT, in which the positron range profiles were calculated analytically [16]. Phantom simulations revealed an improvement of the FWTM up to 50% compared to the non-corrected images, similar to the homogeneous correction [16]. However, microPET imaging of mice injected with 68Ga-DOTATOC demonstrated that the heterogeneous correction increased TBR by more than 90% compared to reconstruction without correction [16]. Additionally, it produced better results in regions along tissue borders compared to homogeneous tissue-dependent techniques [16]. Another correction was proposed by Kertész et al. [20]. They merged parts of homogeneous tissue-dependent kernels to create a positron range kernel for heterogeneous tissues. The homogeneous tissue-dependent kernels were based on distribution profiles extracted from Monte Carlo simulations. Phantom measurements showed that the proposed correction resulted in an improvement of the RC by up to 33%, contrast recovery (CR) by up to 49%, and in some cases, more than doubled the contrast-to-noise ratio (CNR) compared to the non-corrected reconstruction [20].
Reported drawbacks of heterogeneous tissue-dependent correction techniques were Gibbs artifacts and increased reconstruction times [16,20]. Cal-González et al. observed a ninefold increment of the reconstruction time compared to the reconstruction without correction for the same system without any parallel implementation [16]. For a parallel implementation, the reconstruction time was increased by approximately 50% [16]. Kertész et al. reported a reconstruction time of up to 22 h for their parallel implementation [20]. Additionally, they incorporated the model in both the forward and backward projection, and they reported an image noise reduction of 11% [20]. However, Cal-González et al. observed enhancement of the image noise and, thus, increased the number of image updates of the reconstruction without correction by a factor of 2.6, achieving the same image noise levels between reconstructions with and without correction [16]. A summary of the reconstruction-based positron range correction techniques is given in Table 2.

3.1.2. Post-Reconstruction Approaches

Post-reconstruction approaches are correction techniques that are applied to the reconstructed PET image and, as such, are not part of the reconstruction algorithm itself. Rukiah et al. [21] proposed a positron range correction for 68Ga based on the Richardson–Lucy (RL) de-blurring method [24,25]. Monte Carlo simulations were performed to obtain the positron range distribution profiles of 68Ga for different homogeneous tissues from which kernels were created [21]. Simulations of phantom measurements were performed for both 68Ga and 18F. Rukiah et al. reported that spatial resolution of the corrected 68Ga PET images was inferior to 18F PET images by approximately 9%, 1%, and 56% for water, bone, and lung tissue, respectively [21]. Additionally, noise was enhanced in the corrected 68Ga PET images compared to the uncorrected 68Ga PET images by 159% in lung tissue, while for bone tissue and water, the noise increased by 13% and 17%, respectively [21].
Deep learning techniques were also used to correct for positron range effects. Herraiz et al. implemented a positron range correction based on convolutional neural networks (CNNs) [22]. They generated 68Ga and 18F PET images of mice models using Monte Carlo simulations. The 18F PET images were used as a reference. Data augmentation was performed to increase the amount and variety of training data. The corrected 68Ga PET images generated by this deep learning-based positron range correction method demonstrated increased RC up to approximately 80% and lower noise levels up to approximately 37% in comparison to the uncorrected 68Ga PET images [22]. Compared to the 18F PET images, the corrected 68Ga PET images could achieve comparable RC and noise levels [22]. When implemented on a graphics processing unit (GPU), the required time for positron range correction was only a few seconds [22]. Yang also implemented a deep learning correction method for 68Ga utilizing networks with different architectures [23]. For this study, Monte Carlo simulations of phantoms were performed using 68Ga and another source of gamma rays with no positron range effect. The CNNs were trained using the 68Ga PET images as inputs and the images produced from the gamma source as a reference. Yang reported that the corrected images had sharper boundaries, higher RC, and a higher spillover ratio (SOR) [23].

3.2. Noise Reduction Techniques

Two different approaches were identified for reducing the noise in 68Ga PET images and enabling low-count 68Ga PET measurements. The first approach reduces the noise during the image reconstruction process (reconstruction-based noise reduction approaches, n = 11) [26,27,28,29,30,31,32,33,34,35,36], whereas the second approach is based on neural networks (deep learning approaches for noise reduction, n = 5) [37,38,39,40,41], as seen in Figure 2.

3.2.1. Reconstruction-Based Noise Reduction Approaches

Ordered subset expectation maximization (OSEM) is a reconstruction algorithm that is generally used to reconstruct PET images [42]. It is an iterative expectation maximization (EM) algorithm that tries to find the maximum likelihood (ML) solution, which corresponds to the most likely image given the measured projection data. During each iteration, a subset of the measured data is employed to estimate a new image with a higher likelihood. However, with each update of OSEM, the noise in the reconstructed image increases because of the noisy projection data. Therefore, one approach is to limit the number of updates during reconstruction limiting the reconstructed image noise. However, this means that full convergence is not guaranteed, which may result in an underestimation of radioactivity concentrations [43,44]. Another approach is to ensure full convergence of the reconstructed image by applying enough updates during the reconstruction, while a low-pass filter is applied to the final reconstructed image to reduce the impact of noise.
An approach that allows full convergence is the Bayesian penalized likelihood (BPL) or regularized iterative reconstruction algorithm. In comparison to a regular OSEM algorithm, a noise regularization or penalty term is added to the objective function with a weighting or beta value (β value). Higher β values depict a higher weighting of the regularization term resulting in a more effective noise suppression during the reconstruction. However, including a penalty term in the objective function generally needs an adaption of the optimization scheme as well, with Block-Sequential Regularized Expectation Maximization (BSREM), an optimizer that is frequently used for BPL reconstruction algorithms [45,46]. Based on the studied literature two BPL reconstruction algorithms have been investigated with respect to 68Ga-labeled radiotracers.
One of those BPL reconstruction algorithms was introduced by GE Healthcare under the name Q.Clear, which includes a noise regularization term based on relative differences between neighboring voxels and uses BSREM as an optimizer. In addition, Q.Clear incorporates both time-of-flight (TOF) information and point spread function (PSF) modeling. A literature search including both BSREM and Q.Clear revealed multiple studies on the optimal use of Q.Clear for different 68Ga-labeled radiotracers using both patient and phantom studies. These studies demonstrated that increasing the β value resulted in a reduction in the image noise [26,27,28,29,30,31,32,33], an improvement in the signal-to-noise ratio (SNR) [26,27,29,30,32], and a reduction in Gibbs artifacts [33] at the cost of a reduction in the maximum standardized uptake value (SUVmax) [26,28,29,31], CR [26,28,30], RC [33], and signal-to-background ratio (SBR) [27,29,32]. Additionally, the mean standardized uptake value (SUVmean) was almost identical for different β values [26,27,28,29]. Increasing the number of iterations had a minor effect on image noise in BSREM [26]. It was further observed that the same β values have different impacts on SUVmax, CR, and SNR of lesions with different sizes and uptakes. Specifically, an increase in the β value enhanced the relative difference in SUVmax, CR, and SNR as the size and uptake of lesions decreased [26,29,32].
Quantitative image analysis and visual assessment by experts confirmed that BSREM can outperform OSEM with respect to image quality. However, these findings were only valid for a certain range of β values. Different studies investigated the potential of BSREM in comparison to OSEM for different 68Ga-labeled radiotracers, while using different imaging settings and measures of image quality. As a result, different ranges of recommended β values were reported, as presented in Table 3. BMI can also affect the optimal β value as Zanoni et al. suggested a rather large β value of 1600 for 68Ga-DOTANOC PET scanning of scan patients with 25 ≤ BMI < 30 or BMI ≥ 30 [34]. However, these ranges of β values need to be interpreted with care since this is a GE proprietary and unitless tuning parameter, which can also depend on the version of the reconstruction software. Nevertheless, the different ranges of optimal β values clearly demonstrate optimal noise regularization is highly dependent on the acquisition time, injected dose, and 68Ga-labeled radiotracer.
Additionally, studies demonstrated the potential to reduce the acquisition time using BSREM while maintaining an image quality that is still comparable to OSEM reconstructions using longer acquisition times [27,29,32]. Based on quantitative image analysis, up to a 75% reduction in acquisition time was achieved [27], while another study reported a potential acquisition time reduction of up to 33% based on a visual assessment [32]. Additionally, Svirydenka et al. demonstrated that the use of BSREM and a five-fold increase in the acquisition time resulted in a ten-fold reduction in the administered activity of 68Ga-PSMA in comparison to standard OSEM [31], therefore enabling ultra-low activity examinations.
In addition to Q.Clear, United Imaging Healthcare introduces a BPL iterative reconstruction algorithm based on total variation regularized expectation maximization TVREM (HYPER Iterative) with a penalization factor between 0 and 1 to adjust the total variation penalization of voxels of corresponding neighborhoods. TVREM was used for PET phantom and patient scanning with 68Ga-PSMA (20 patients) [35] and 68Ga-DOTATATE (17 patients) [36]. As the penalization factor α increased, image noise [35,36], CR [35], SUVmax [35,36], and TBR [36] were reduced, while SNR increased [36]. SUVmean remained similar when the penalization factor increased [35,36]. Additionally, it was observed that the penalization factor α had a greater effect as the size of the lesions decreased. Specifically, for lesions with a diameter ranging between 10 and 20 mm, the SUVmax was decreased as the penalization factor α increased, while for lesions with a diameter greater than 20 mm, the SUVmax was almost identical [35]. In addition to that, the relative differences in SNR and TBR were larger for lesions with a diameter lower than 10 mm than lesions with greater or equal [36].
Quantitative image analysis and visual assessment by experts demonstrated that TVREM was capable of enhancing image quality more compared to OSEM but only for a certain range of values [35,36]. Yang et al. demonstrated that for penalization factor α ranging between 0.07 and 0.28 for 68Ga-PSMA, a reduction in the acquisition time of 33% was achieved based on quantitative image analysis and visual assessment from experts [35]. Specifically, compared to OSEM, they reported improvements in the contrast up to 17%, the SUVmax up to 15%, and the image noise was reduced up to 32% [35]. Liu et al. reported that TVREM improved the overall SUVmax, SNR, and TBR compared to OSEM [36]. Additionally, based on visual assessment, they suggested that TVREM was capable of preserving the image quality for 68Ga-DOTATATE while achieving a reduction in the acquisition time by 33% for values between 0.14 and 0.35 [36].

3.2.2. Deep Learning Approaches for Noise Reduction

The potential of deep learning approaches, both supervised and unsupervised, to reduce the noise in PET images has been demonstrated in the literature. For a supervised approach, low- and high-count PET images are available for each scan to train the network. Liu et al. implemented a supervised deep learning approach in which cross-tracer and cross-protocol learning were performed [37]. They used PET images consisting of only 10% of the counts of the original single-bed 18F-FMISO (n = 12), single-bed 18F-FDG (n = 9), whole-body 18F-FDG (n = 12), and whole-body 68Ga-DOTATATE (n = 15) PET scans. Image quality enhancement was reported by comparison with the unenhanced, low-count PET images. Additionally, they reported the potential of employing neural networks trained based on 18F-FDG image data to reduce the image noise for other radiotracers such as 68Ga-DOTATATE and different scanning protocols [37]. Deng et al. used low-count 68Ga-PSMA PET images (41 patients) and data augmentation to train a supervised deep learning model for denoising [38]. Based on quantitative analysis and visual assessment, Deng et al. suggested a 50% count reduction as the optimal trade-off.
Additionally, unsupervised approaches were also investigated in the literature. In comparison to supervised approaches, training pairs are not required to train the unsupervised approaches, since there is no target output. Therefore, high-count PET images are unnecessary to train an unsupervised deep learning approach. Cui et al. investigated the potential of unsupervised deep learning [39,40]. An unsupervised neural network was validated using simulated 18F-FDG phantom data and evaluated using 68Ga-PRGD2 PET/CT (n = 10) and 18F-FDG PET/MRI patient data (n = 30) [40]. Compared to the original PET images, the mean CNR was improved by approximately 53% for 68Ga-PRGD2 PET/CT and approximately 47% for 18F-FDG PET/MRI [40]. However, they suggested radiotracer- and modality-independent effects since they reported no statistically significant differences between those improvements [40]. In a later publication [41], Cui et al. expanded their previous approach [40] by incorporating a second unsupervised neural network, which used, as initial parameters, the pre-trained parameters of the first network. In that study [41], the same dataset was used as in [40]. This new approach improved the mean CNR by approximately 71% for 68Ga-PRGD2 PET/CT and by approximately 58% for 18F-FDG PET/MRI compared to the original PET images. In addition, compared to the previous approach [40], it could better retain better structural image features [41]. Meanwhile, SUVmax and SUVmean values were comparable for both approaches [41].

4. Discussion

In the studied literature, the capability of improving the image quality of 68Ga was demonstrated by incorporating a 68Ga-specific positron range correction. In some cases, image quality comparable to 18F was reported [21,22]. The introduction of a 68Ga-specific positron range correction may increase the 68Ga image quality in clinical practice and improve the diagnostic value of these images. Despite that, none of these techniques have yet been applied or validated in a clinical setting. To the best of our knowledge, from the presented studies, only the techniques presented in [16] were applied in a preclinical setting (18 scans of 12 mice), where an improvement in the image quality in relation to the case of no correction was reported [47]. However, there is the need for 68Ga-specific positron range correction to be translated into the clinical setting after which thorough validation is required prior to clinical implementation.
Another aspect of the 68Ga-specific positron range corrections that needs to be considered for clinical applications is the computation time. Advanced and accurate reconstruction-based corrections have increased the computation time [16,20], which may be incompatible with clinical practice. Therefore, for the reconstruction-based corrections, one necessary criterion that needs to be fulfilled is to have a minor impact on the overall computational time.
The post-reconstruction corrections also showed promising results. Deep learning approaches have the potential to improve image quality while maintaining low noise levels. Aside from that, they seem to be computationally time efficient. It was demonstrated that a post-reconstruction deep learning correction implemented in GPU required only a few seconds [22]. However, for deep learning approaches, large datasets are required, including as many different cases as possible and reducing the potential of overfitting. Data augmentation techniques can increase the size of the dataset and reduce overfitting [48] and have already been applied [22]. However, for small datasets without enough representative training data, data augmentation techniques will not be able to correct the lack of insufficient data.
Noise is another important factor that diminishes image quality. Iterative reconstruction algorithms using penalizing factors have the potential to produce better image quality than OSEM and even reduce acquisition time. However, quantitative image analysis demonstrated a greater reduction in acquisition time in comparison to the reduction suggested through visual assessment. Additionally, the intrareader variability may be considered when performing a visual assessment.
In contrast to OSEM, a small effect on image noise was reported in Q.Clear as the number of iterations increased [26]. This will allow for better conversion of small structures in particular, which will enhance small lesion identification and quantification [49]. Increasing the β value resulted in higher relative differences in SUVmax, CR, and SNR for small lesions, while as the size and uptake increased, this effect became less prominent [26,29,32]. Q.Clear can enhance small lesion identification and quantification; however, for too-large β values, the noise suppression effect of Q.Clear may hinder the detection of small lesions in comparison to OSEM. Additionally, identifying the optimal values of the noise penalization factor where the generated images are better than OSEM is not a straightforward process. The optimal value of the noise penalization factor depends on various factors such as the radiotracer distribution, acquisition time, BMI, clinical indication, measures of image quality, and even the preference of the specialists or experts evaluating the images. Consequently, studies usually reported a range of recommended values. However, ideally, the noise penalization factor should be evaluated for each clinical application.
Deep learning techniques were also implemented to denoise data. In the literature, both supervised and unsupervised approaches were described. However, studies did not only explicitly focus on 68Ga-labeled radiotracers but also a combination of 68Ga-labeled and 18F-labeled radiotracers. It was demonstrated that supervised deep learning enabled networks trained from one radiotracer or scanning protocol to denoise PET images of other radiotracer or scanning protocols [37]. This was demonstrated using PET images consisting of only 10% of the counts of full-activity PET images [37]. However, Deng et al. showed that PET images generated from 50% of the counts using a deep learning approach revealed the same clinical result as the original data in patients with a suspected diagnosis of prostate cancer based on a clinical quantitative assessment by specialists [38]. Unsupervised deep learning approaches also demonstrated the potential of denoising. One advantage is the absence of high-quality PET images as training pairs. However, none of the included studies compared the image quality produced by unsupervised learning with respect to supervised learning.
One disadvantage of the reconstruction-based noise reduction approaches in comparison to the deep learning approaches is the need to define the optimal value of the noise penalization factor, which depends on various factors that are highly variable. Furthermore, it was demonstrated that using supervised deep learning networks can be trained from one radiotracer or scanning protocol and used for denoising other radiotracer or scanning protocols [37]. Despite that, one important disadvantage of deep learning approaches is the dependency on the training dataset. Deng et al. performed data augmentation to increase the dataset size and avoid overfitting [38]. However, data augmentation techniques will not be able to compensate if there are not sufficient representative training data.
In the studied literature, either positron range corrections or denoising approaches were investigated with respect to 68Ga. Ideally, in a clinical setting using 68Ga-labeled radiotracers, there is a need for both positron range correction and noise reduction techniques. The 68Ga PET image quality can be enhanced by implementing positron range correction and can even achieve comparable image quality as 18F PET. However, some positron range correction techniques were reported to also enhance image noise. Therefore, combining a denoising approach that will enable low-count PET imaging may compensate for the increased image noise introduced by the positron range correction. Low-count PET imaging can be translated into lower administered activities and/or acquisition time, which has the potential to reduce the cost of PET imaging, increase the patient throughput of the system, and increase patient comfort.

5. Conclusions

Positron range correction techniques have the potential to improve the 68Ga PET image quality, and it has been shown that image quality comparable to 18F can be achieved. However, none of the proposed approaches have been clinically evaluated. Therefore, prior to clinical implementation, thorough validation is necessary.
Different denoising approaches for 68Ga that were evaluated in both phantom and clinical data, based on both reconstruction and deep learning techniques, showed the potential to achieve a substantial reduction in administered radiotracer activity or acquisition time.
The combination of both positron range correction and image denoising techniques has great potential to significantly improve the 68Ga PET image quality and consequently enhance the clinical value of low-count PET using 68Ga-based radiotracers.

Author Contributions

Conceptualization, P.G., M.K., F.M.M. and R.W.; writing—original draft preparation, P.G.; supervision, M.K., F.M.M., S.A. and R.W. All authors have read and agreed to the published version of the manuscript.

Funding

This research was partially funded by an institutional grant to the Department of Radiology and Nuclear medicine, Maastricht by GE Healthcare.

Conflicts of Interest

F.M.M. is medical advisor for NanoMab Technology Ltd. and Advanced Accelerator Applications (AAA) GmbH and has recently received institutional grants from NanoMab Technology Ltd., Siemens, and GE Precision Healthcare LLC. The other authors declare no conflict of interest.

Appendix A

The queries of the positron range correction techniques in the search engines PubMed and Scopus as well as their results are presented in Table A1 and Table A2, respectively. In Table A3 and Table A4, the queries we employed to retrieve articles related to denoising approaches and their results for PubMed and Scopus, respectively, can be seen. The results are from January 2010 until 1 August 2022.
Table A1. The search query for positron range correction techniques in PubMed. In the column “Results”, the numbers of records are provided for the corresponding search terms from January 2010 until 1 August 2022.
Table A1. The search query for positron range correction techniques in PubMed. In the column “Results”, the numbers of records are provided for the corresponding search terms from January 2010 until 1 August 2022.
SetSearch TermsResults
#1“Positron-Emission Tomography” [Mesh] OR positron emission tomograph* [tiab] OR PET [tiab] OR PET/CT [tiab] OR PET-CT [tiab] OR PET/MR* [tiab] OR PET-MRI [tiab] OR MRI/PET [tiab] OR MR/PET [tiab]106,387
#2gallium-68* [tiab] OR 68Ga* [tiab] OR 68 Ga* [tiab] OR Ga-68* [tiab] OR “Gallium Radioisotopes*” [Mesh] OR “high positron energ*” [tiab] OR “high-energy positron*” [tiab]6230
#3positron range* [tiab] OR positron-range* [tiab] OR positron range* [Mesh] OR positron energ* [tiab] OR positron-energ* [tiab]122
#4#1 AND #2 AND #343
Table A2. The search query for positron range correction techniques in Scopus. In the column “Results”, the numbers of records are provided for the corresponding search terms from January 2010 until 1 August 2022.
Table A2. The search query for positron range correction techniques in Scopus. In the column “Results”, the numbers of records are provided for the corresponding search terms from January 2010 until 1 August 2022.
SetSearch TermsResults
#1TITLE-ABS-KEY((“Positron-Emission Tomography”) OR (“positron emission tomograph*”) OR (“PET”) OR (“PET/CT”) OR (“PET-CT”) OR (“PET/MR*”) OR (“PET-MRI”) OR (“MRI/PET”) OR (“MR/PET”))196,962
#2TITLE-ABS-KEY((gallium-68*) OR (68Ga*) OR (68 Ga*) OR (Ga-68*) OR (high positron energ*) OR (high energy positron*))44,335
#3TITLE-ABS-KEY((“positron range*”) OR (“positron energ*”))368
#4#1 AND #2 AND #383
Table A3. The search query for noise reduction techniques in PubMed. In the column “Results”, the numbers of records are provided for the corresponding search terms from between January 2010 until 1 August 2022.
Table A3. The search query for noise reduction techniques in PubMed. In the column “Results”, the numbers of records are provided for the corresponding search terms from between January 2010 until 1 August 2022.
SetSearch TermsResults
#1“Positron-Emission Tomography” [Mesh] OR positron emission tomograph* [tiab] OR PET [tiab] OR PET/CT [tiab] OR PET-CT [tiab] OR PET/MR* [tiab] OR PET-MRI [tiab] OR MRI/PET [tiab] OR MR/PET [tiab]106,387
#2* [tiab] OR 68Ga* [tiab] OR 68 Ga* [tiab] OR Ga-68* [tiab] OR “Gallium Radioisotopes*” [Mesh] OR “high positron energ*” [tiab] OR “high-energy positron*” [tiab]6230
#3nois* [tiab] OR nois* [Mesh] OR denois* [tiab] OR “low dose*” [tiab] OR “low-dose*” [tiab] OR dose reduction* [tiab] OR SNR [tiab] OR CNR [tiab] OR “variability*” [tiab] OR “low-count*” [tiab] OR “low count” [tiab]368,718
#4#1 AND #2 AND #3192
Table A4. The search query for noise reduction techniques in Scopus. In the column “Results”, the numbers of records are provided for the corresponding search terms from between January 2010 until 1 August 2022.
Table A4. The search query for noise reduction techniques in Scopus. In the column “Results”, the numbers of records are provided for the corresponding search terms from between January 2010 until 1 August 2022.
SetSearch TermsResults
#1TITLE-ABS-KEY((“Positron-Emission Tomography”) OR (“positron emission tomograph*”) OR (“PET”) OR (“PET/CT”) OR (“PET-CT”) OR (“PET/MR*”) OR (“PET-MRI”) OR (“MRI/PET”) OR (“MR/PET”))196,962
#2TITLE-ABS-KEY((gallium-68*) OR (68Ga*) OR (68 Ga*) OR (Ga-68*) OR (high positron energ*) OR (high energy positron*))44,335
#3TITLE-ABS-KEY((nois*) OR (denois*) OR (“low dose*”) OR (“low-dose*”) OR (dose reduction*) OR (“SNR”) OR (“CNR”) OR (“variability*”) OR (“low count*”) OR (“low-count*”))1,578,094
#4#1 AND #2 AND #3844

References

  1. Bozkurt, M.F.; Virgolini, I.; Balogova, S.; Beheshti, M.; Rubello, D.; Decristoforo, C.; Ambrosini, V.; Kjaer, A.; Delgado-Bolton, R.; Kunikowska, J.; et al. Guideline for PET/CT imaging of neuroendocrine neoplasms with 68Ga-DOTA-conjugated somatostatin receptor targeting peptides and 18F–DOPA. Eur. J. Nucl. Med. Mol. Imaging 2017, 44, 1588–1601. [Google Scholar] [CrossRef] [PubMed]
  2. Fendler, W.P.; Eiber, M.; Beheshti, M.; Bomanji, J.; Ceci, F.; Cho, S.; Giesel, F.; Haberkorn, U.; Hope, T.A.; Kopka, K.; et al. 68Ga-PSMA PET/CT: Joint EANM and SNMMI procedure guideline for prostate cancer imaging: Version 1.0. Eur. J. Nucl. Med. Mol. Imaging 2017, 44, 1014–1024. [Google Scholar] [CrossRef] [PubMed]
  3. Velikyan, I. Prospective of 68Ga Radionuclide Contribution to the Development of Imaging Agents for Infection and Inflammation. Contrast Media Mol. Imaging 2018, 2018, 9713691. [Google Scholar] [CrossRef] [PubMed]
  4. Banerjee, S.R.; Pomper, M.G. Clinical applications of Gallium-68. Appl. Radiat. Isot. 2013, 76, 2–13. [Google Scholar] [CrossRef] [PubMed]
  5. Jødal, L.; Le Loirec, C.; Champion, C. Positron range in PET imaging: An alternative approach for assessing and correcting the blurring. Phys. Med. Biol. 2012, 57, 3931–3943. [Google Scholar] [CrossRef]
  6. Sanchez-Crespo, A. Comparison of Gallium-68 and Fluorine-18 imaging characteristics in positron emission tomography. Appl. Radiat. Isot. 2013, 76, 55–62. [Google Scholar] [CrossRef] [PubMed]
  7. Kemerink, G.J.; Visser, M.G.W.; Franssen, R.; Beijer, E.; Zamburlini, M.; Halders, S.G.E.A.; Brans, B.; Mottaghy, F.M.; Teule, G.J.J. Effect of the positron range of 18F, 68Ga and 124I on PET/CT in lung-equivalent materials. Eur. J. Nucl. Med. Mol. Imaging 2011, 38, 940–948. [Google Scholar] [CrossRef]
  8. Carter, L.M.; Kesner, A.L.; Pratt, E.C.; Sanders, V.A.; Massicano, A.V.F.; Cutler, C.S.; Lapi, S.E.; Lewis, J.S. The Impact of Positron Range on PET Resolution, Evaluated with Phantoms and PHITS Monte Carlo Simulations for Conventional and Non-conventional Radionuclides. Mol. Imaging Biol. 2020, 22, 73–84. [Google Scholar] [CrossRef]
  9. Alva-Sánchez, H.; Quintana-Bautista, C.; Martínez-Dávalos, A.; A Ávila-Rodríguez, M.; Rodríguez-Villafuerte, M. Rodríguez-Villafuerte. Positron range in tissue-equivalent materials: Experimental microPET studies. Phys. Med. Biol. 2016, 61, 6307–6321. [Google Scholar] [CrossRef]
  10. Andreo, P.; Larsson, S.A. Positron flight in human tissues and its influence on PET image spatial resolution. Eur. J. Nucl. Med. Mol. Imaging 2004, 31, 44–51. [Google Scholar] [CrossRef]
  11. Liu, X.; Laforest, R. Quantitative small animal PET imaging with nonconventional nuclides. Nucl. Med. Biol. 2009, 36, 551–559. [Google Scholar] [CrossRef] [PubMed]
  12. Teuho, J.; Riehakainen, L.; Honkaniemi, A.; Moisio, O.; Han, C.; Tirri, M.; Liu, S.; Grönroos, T.J.; Liu, J.; Wan, L.; et al. Evaluation of image quality with four positron emitters and three preclinical PET/CT systems. EJNMMI Res. 2020, 10, 155. [Google Scholar] [CrossRef] [PubMed]
  13. Chaudhari, A.S.; Mittra, E.; Davidzon, G.A.; Gulaka, P.; Gandhi, H.; Brown, A.; Zhang, T.; Srinivas, S.; Gong, E.; Zaharchuk, G.; et al. Low-count whole-body PET with deep learning in a multicenter and externally validated study. Npj Digit. Med. 2021, 4, 127. [Google Scholar] [CrossRef]
  14. Arabi, H.; AkhavanAllaf, A.; Sanaat, A.; Shiri, I.; Zaidi, H. The promise of artificial intelligence and deep learning in PET and SPECT imaging. Phys. Med. 2021, 83, 122–137. [Google Scholar] [CrossRef] [PubMed]
  15. Xing, Y.; Qiao, W.; Wang, T.; Wang, Y.; Li, C.; Lv, Y.; Xi, C.; Liao, S.; Qian, Z.; Zhao, J. Deep learning-assisted PET imaging achieves fast scan/low-dose examination. EJNMMI Phys. 2022, 9, 7. [Google Scholar] [CrossRef]
  16. Cal-González, J.; Perez-Liva, M.; Herraiz, J.L.; Vaquero, J.J.; Desco, M.; Udias, J.M. Tissue-Dependent and Spatially-Variant Positron Range Correction in 3D PET. IEEE Trans. Med. Imaging 2015, 34, 2394–2403. [Google Scholar] [CrossRef] [PubMed]
  17. Bertolli, O.; Eleftheriou, A.; Cecchetti, M.; Camarlinghi, N.; Belcari, N.; Tsoumpas, C. PET iterative reconstruction incorporating an efficient positron range correction method. Phys. Med. 2016, 32, 323–330. [Google Scholar] [CrossRef]
  18. Cal-González, J.; Herraiz, J.; España, S.; Vicente, E.; Herranz, E.; Desco, M.; Vaquero, J.; Udías, J. Study of CT-based positron range correction in high resolution 3D PET imaging. Nucl. Instrum. Methods Phys. Res. 2011, 648, S172–S175. [Google Scholar] [CrossRef]
  19. Kraus, R.; Delso, G.; Ziegler, S.I. Simulation study of tissue-specific positron range correction for the new biograph mMR whole-body PET/MR system. IEEE Trans. Nucl. Sci. 2012, 59, 1900–1909. [Google Scholar] [CrossRef]
  20. Kertész, H.; Beyer, T.; Panin, V.; Jentzen, W.; Cal-Gonzalez, J.; Berger, A.; Papp, L.; Kench, P.L.; Bharkhada, D.; Cabello, J.; et al. Implementation of a Spatially-Variant and Tissue-Dependent Positron Range Correction for PET/CT Imaging. Front. Physiol. 2022, 13, 818463. [Google Scholar] [CrossRef]
  21. Rukiah, A.; Meikle, S.R.; Gillam, J.E.; Kench, P.L. An investigation of 68Ga positron range correction through de-blurring: A simulation study. In Proceedings of the 2018 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Sydney, Australia, 10–17 November 2018; pp. 1–2. [Google Scholar] [CrossRef]
  22. Herraiz, J.; Bembibre, A.; López-Montes, A. Deep-learning based positron range correction of pet images. Appl. Sci. 2021, 11, 266. [Google Scholar] [CrossRef]
  23. Yang, C.-C. Compensating Positron Range Effects of Ga-68 in Preclinical PET Imaging by Using Convolutional Neural Network: A Monte Carlo Simulation Study. Diagnostics 2021, 11, 2275. [Google Scholar] [CrossRef] [PubMed]
  24. Richardson, W.H. Bayesian-Based Iterative Method of Image Restoration*. J. Opt. Soc. Am. 1972, 62, 55–59. [Google Scholar] [CrossRef]
  25. Lucy, L.B. An iterative technique for the rectification of observed distributions. Astron. J. 1974, 79, 745. [Google Scholar] [CrossRef]
  26. Ter Voert, E.E.G.W.; Muehlematter, U.J.; Delso, G.; Pizzuto, D.A.; Müller, J.; Nagel, H.W.; Burger, I.A. Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical 68Ga-PSMA PET/MR. EJNMMI Res. 2018, 8, 70. [Google Scholar] [CrossRef]
  27. Lindström, E.; Velikyan, I.; Regula, N.; Alhuseinalkhudhur, A.; Sundin, A.; Sörensen, J.; Lubberink, M. reconstruction of digital time-of-flight 68Ga-PSMA-11 PET/CT for the detection of recurrent disease in prostate cancer patients. Theranostics 2019, 9, 3476–3484. [Google Scholar] [CrossRef]
  28. Baratto, L.; Duan, H.; Ferri, V.; Khalighi, M.; Iagaru, A. The Effect of Various β Values on Image Quality and Semiquantitative Measurements in 68Ga-RM2 and 68Ga-PSMA-11 PET/MRI Images Reconstructed with a Block Sequential Regularized Expectation Maximization Algorithm. Clin. Nucl. Med. 2020, 45, 506–513. [Google Scholar] [CrossRef]
  29. Lindström, E.; Lindsjö, L.; Sundin, A.; Sörensen, J.; Lubberink, M. Evaluation of block-sequential regularized expectation maximization reconstruction of 68Ga-DOTATOC, 18F-fluoride, and 11C-acetate whole-body examinations acquired on a digital time-of-flight PET/CT scanner. EJNMMI Phys. 2020, 7, 40. [Google Scholar] [CrossRef]
  30. Seo, Y.; Khalighi, M.M.; Wangerin, K.A.; Deller, T.W.; Wang, Y.-H.; Jivan, S.; Kohi, M.P.; Aggarwal, R.; Flavell, R.R.; Behr, S.C.; et al. Quantitative and qualitative improvement of low-count [68Ga] citrate and [90Y] microspheres PET image reconstructions using block sequential regularized expectation maximization algorithm. Mol. Imaging Biol. 2020, 22, 208–216. [Google Scholar] [CrossRef]
  31. Svirydenka, H.; Muehlematter, U.J.; Nagel, H.W.; Delso, G.; Ferraro, D.A.; Kudura, K.; Burger, I.A.; ter Voert, E.E.G.W. 68Ga-PSMA-11 dose reduction for dedicated pelvic imaging with simultaneous PET/MR using TOF BSREM reconstructions. Eur. Radiol. 2020, 30, 3188–3197. [Google Scholar] [CrossRef]
  32. Chicheportiche, A.; Goshen, E.; Godefroy, J.; Grozinsky-Glasberg, S.; Oleinikov, K.; Meirovitz, A.; Gross, D.J.; Ben-Haim, S. Can a penalized-likelihood estimation algorithm be used to reduce the injected dose or the acquisition time in 68Ga-DOTATATE PET/CT studies? EJNMMI Phys. 2021, 8, 13. [Google Scholar] [CrossRef] [PubMed]
  33. Krokos, G.; Pike, L.C.; Cook, G.J.R.; Marsden, P.K. Standardisation of conventional and advanced iterative reconstruction methods for Gallium-68 multi-centre PET-CT trials. EJNMMI Phys. 2021, 8, 52. [Google Scholar] [CrossRef] [PubMed]
  34. Zanoni, L.; Argalia, G.; Fortunati, E.; Malizia, C.; Allegri, V.; Calabrò, D.; Civollani, S.; Campana, D.; Fanti, S.; Ambrosini, V. Can Q.Clear reconstruction be used to improve [68Ga] Ga-DOTANOC PET/CT image quality in overweight NEN patients? Eur. J. Nucl. Med. Mol. Imaging 2022, 49, 1607–1612. [Google Scholar] [CrossRef] [PubMed]
  35. Yang, F.-J.; Ai, S.-Y.; Wu, R.; Lv, Y.; Xie, H.-F.; Dong, Y.; Meng, Q.-L.; Wang, F. Impact of total variation regularized expectation maximization reconstruction on the image quality of 68Ga-PSMA PET: A phantom and patient study. Br. J. Radiol. 2021, 94, 20201356. [Google Scholar] [CrossRef]
  36. Liu, L.; Liu, H.; Xu, S.; Zhang, S.; Tao, Y.; Mok, G.S.P.; Chen, Y. The Impact of Total Variation Regularized Expectation Maximization Reconstruction on 68Ga-DOTA-TATE PET/CT Images in Patients With Neuroendocrine Tumor. Front. Med. 2022, 9, 845806. [Google Scholar] [CrossRef]
  37. Liu, H.; Wu, J.; Lu, M.W.; Onofrey, J.A.; Liu, Y.-H.; Liu, C. Noise reduction with cross-tracer and cross-protocol deep transfer learning for low-dose PET. Phys. Med. Biol. 2020, 65, 185006. [Google Scholar] [CrossRef]
  38. Deng, F.; Li, X.; Yang, F.; Sun, H.; Yuan, J.; He, Q.; Xu, W.; Yang, Y.; Liang, D.; Liu, X.; et al. Low-Dose 68 Ga-PSMA Prostate PET/MRI Imaging Using Deep Learning Based on MRI Priors. Front. Oncol. 2022, 11, 818329. [Google Scholar] [CrossRef]
  39. Cui, J.; Gong, K.; Guo, N.; Kim, K.; Liu, H.; Li, Q. CT-guided PET parametric image reconstruction using deep neural network without prior training data. In Proceedings of the Medical Imaging 2019: Physics of Medical Imaging, San Diego, CA, USA, 17–20 February 2019; pp. 238–243. [Google Scholar] [CrossRef]
  40. Cui, J.; Gong, K.; Guo, N.; Wu, C.; Meng, X.; Kim, K.; Zheng, K.; Wu, Z.; Fu, L.; Xu, B.; et al. PET image denoising using unsupervised deep learning. Eur. J. Nucl. Med. Mol. Imaging 2019, 46, 2780–2789. [Google Scholar] [CrossRef]
  41. Cui, J.; Gong, K.; Guo, N.; Wu, C.; Kim, K.; Liu, H.; Li, Q. Populational and individual information based PET image denoising using conditional unsupervised learning. Phys. Med. Biol. 2021, 66, 155001. [Google Scholar] [CrossRef]
  42. Hudson, H.M.; Larkin, R.S. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans. Med. Imaging 1994, 13, 601–609. [Google Scholar] [CrossRef]
  43. Liow, J.-S.; Strother, S. The convergence of object dependent resolution in maximum likelihood based tomographic image reconstruction. Phys. Med. Biol. 1993, 38, 55–70. [Google Scholar] [CrossRef] [PubMed]
  44. Boellaard, R.; Van Lingen, A.; Lammertsma, A. Experimental and Clinical Evaluation of Iterative Reconstruction (OSEM) in Dynamic PET: Quantitative Characteristics and Effects on Kinetic Modeling. J. Nucl. Med. 2001, 42, 808–817. [Google Scholar] [PubMed]
  45. De Pierro, A.R.; Yamagishi, M.E.B. Fast EM-like methods for maximum “a posteriori” estimates in emission tomography. IEEE Trans. Med. Imaging 2001, 20, 280–288. [Google Scholar] [CrossRef] [PubMed]
  46. Ahn, S.; Fessler, J. Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms. IEEE Trans. Med. Imaging 2003, 22, 613–626. [Google Scholar] [CrossRef]
  47. Cal-Gonzalez, J.; Vaquero, J.J.; Herraiz, J.L.; Pérez-Liva, M.; Soto-Montenegro, M.L.; Peña-Zalbidea, S.; Desco, M.; Udías, J.M. Improving PET Quantification of Small Animal [68Ga] DOTA-Labeled PET/CT Studies by Using a CT-Based Positron Range Correction. Mol. Imaging Biol. 2018, 20, 584–593. [Google Scholar] [CrossRef]
  48. Shorten, C.; Khoshgoftaar, T.M. A survey on Image Data Augmentation for Deep Learning. J. Big Data 2019, 6, 60. [Google Scholar] [CrossRef]
  49. Van Der Vos, C.S.; Koopman, D.; Rijnsdorp, S.; Arends, A.J.; Boellaard, R.; Van Dalen, J.A.; Lubberink, M.; Willemsen, A.T.M.; Visser, E.P. Quantification, improvement, and harmonization of small lesion detection with state-of-the-art PET. Eur. J. Nucl. Med. Mol. Imaging 2017, 44, 4–16. [Google Scholar] [CrossRef] [Green Version]
Figure 1. Chart illustrating the screening process of the literature review.
Figure 1. Chart illustrating the screening process of the literature review.
Diagnostics 12 02335 g001
Figure 2. Categorization of the literature search with the numbers depicting the number of articles per category.
Figure 2. Categorization of the literature search with the numbers depicting the number of articles per category.
Diagnostics 12 02335 g002
Table 1. Summary of abbreviations in this paper.
Table 1. Summary of abbreviations in this paper.
AbbreviationDescription
18FFluorine-18
68GaGallium-68
68GeGermanium-68
BMIBody mass index
BPLBayesian penalized likelihood
BSREMBlock-sequential regularized expectation maximization
CNNConvolutional neural network
CNRContrast-to-noise ratio
CRContrast recovery
CTComputerized tomography
EMExpectation maximization
EmaxMaximum energy
FWHMFull width half maximum
FWTMFull width tenth maximum
GMPGood manufacturing practice
GPUGraphics processing unit
min/bpMinutes per bed position
MLMaximum likelihood
MRIMagnetic resonance Imaging
OSEMOrdered subset expectation maximization
PETPositron emission tomography
PSMAProstate-specific membrane antigen
RCRecovery coefficients
SBRSignal-to-background ratio
SNRSignal-to-noise ratio
SORSpill-over ratio
SUVmaxMaximum standardized uptake value
SUVmeanMean standardized uptake value
TBRTumor-to-background ratio
TOFTime-of-flight
TVREMTotal variation regularized expectation maximization
Table 2. Summary of the reconstruction-based corrections for 68Ga. The asterisk (*) depicts an improvement observed by incorporating the blurring kernel both in forward and backward projection steps.
Table 2. Summary of the reconstruction-based corrections for 68Ga. The asterisk (*) depicts an improvement observed by incorporating the blurring kernel both in forward and backward projection steps.
Positron Range CorrectionReported ImprovementsReported Drawbacks
Tissue-independent
  • Improve FWTM [16]
  • Improve activity recovery [17,18]
  • FWTM and activity recovery over-correction in the bone tissue [16,18]
  • FWTM under-correction in lung tissue [16]
  • Increased noise [17]
Homogeneous tissue-dependent
  • Improved FWTM [16] and FWHM [18]
  • Improved TBR [16]
  • Improve activity recovery [17,18]
  • Improved RC [18]
  • Improved contrast [19]
  • Reduced misplaced activity concentration in soft tissue [19]
  • Increased reconstruction time [16]
  • Increased noise [16,17]
  • Gibbs artifacts [16,17]
Heterogeneous tissue-dependent
  • Improved FWTM [16]
  • Improved TBR [16]
  • Improved contrast [19]
  • Reduced misplaced activity concentration in soft tissue [19]
  • Improved RC [20]
  • Improved CR [20]
  • Improved CNR [20]
  • Reduced image noise * [20]
  • Increased reconstruction time [16,20]
  • Increased noise [16]
  • Gibbs artifacts [16,20]
Table 3. Summary of the recommended β values based on the findings of studied literature. Those values were reported to have better overall performance than OSEM. (*) The respective studies did not report the median or mean injected dose. (**) The recommended β value for overweight patients as was suggested by the corresponding authors.
Table 3. Summary of the recommended β values based on the findings of studied literature. Those values were reported to have better overall performance than OSEM. (*) The respective studies did not report the median or mean injected dose. (**) The recommended β value for overweight patients as was suggested by the corresponding authors.
Recommended β ValuesRadiotracerNumber of PatientsAcquisition TimeInjected DoseReferences
400–55068Ga-PSMA252 min/bp128 MBq (median)[26]
400–90068Ga-PSMA202 min/bp2 MBq/kg (mean)[27]
800–100068Ga-PSMA201 min/bp2 MBq/kg (mean)[27]
1200–140068Ga-PSMA200.5 min/bp2 MBq/kg (mean)[27]
500–75068Ga-PSMA364 min/bp151.7 MBq (median)[28]
55068Ga-PSMA2515 min, single bed position15 MBq *[31]
500–75068Ga-RM2424 min/bp144.3 MBq (median)[28]
50068Ga-Citrate64–8 min/bp242.35 MBq (mean)[30]
40068Ga-DOTATOC132 min/bp2.3 MBq/kg (mean)[29]
53368Ga-DOTATOC131.5 min/bp2.3 MBq/kg (mean)[29]
1100–140068Ga-DOTATATE301.5 min/bp160.3 MBq (mean)[32]
1300–160068Ga-DOTATATE301 min/bp160.3 MBq (mean)[32]
1600 **68Ga-DOTANOC753 min/bp100–200 MBq *[34]
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Gavriilidis, P.; Koole, M.; Annunziata, S.; Mottaghy, F.M.; Wierts, R. Positron Range Corrections and Denoising Techniques for Gallium-68 PET Imaging: A Literature Review. Diagnostics 2022, 12, 2335. https://doi.org/10.3390/diagnostics12102335

AMA Style

Gavriilidis P, Koole M, Annunziata S, Mottaghy FM, Wierts R. Positron Range Corrections and Denoising Techniques for Gallium-68 PET Imaging: A Literature Review. Diagnostics. 2022; 12(10):2335. https://doi.org/10.3390/diagnostics12102335

Chicago/Turabian Style

Gavriilidis, Prodromos, Michel Koole, Salvatore Annunziata, Felix M. Mottaghy, and Roel Wierts. 2022. "Positron Range Corrections and Denoising Techniques for Gallium-68 PET Imaging: A Literature Review" Diagnostics 12, no. 10: 2335. https://doi.org/10.3390/diagnostics12102335

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop