A main goal of this research was the comparison of the 14 devices in terms of accuracy. The study showed that all of the monitors tested were accurate for temperature and CO2, because their results were within the accuracy ranges identified in their respective product literature. Specifically for accuracy for CO2, the most accurate devices are listed as follows: Laser Egg + CO2 was the most accurate, then Airthings View Plus, Awair Omni, AirBird, and Aranet4 Home. It is worth noting that the published accuracy threshold ranges varied from product to product. For example, Laser Egg + CO2 had the smallest threshold range, ±30 ppm. When comparing with the same device for each product to see if they are consistent, AirBird and Aranet4 Home had the greatest discrepancy, because on more than four occasions, the results were not within the accuracy threshold. The devices were tested a second time for these two devices after analyzing the initial test results. In these second tests, AirBird and Aranet4 Home still displayed results on more than four occasions that were over the accuracy threshold. In terms of the relative humidity, only Airthings, Eve Room and Purple Air PA-I tested within the accuracy range; AirBird tested once within its range; Aranet4 Home and Laser Egg did not meet their respective accuracy ranges. For PM2.5, most of the devices did not meet the expectations for accuracy. All of the devices did not provide accurate PM2.5 data when compared to the research meter, Lighthouse Handheld-3016-IA. Only Purple Air PA-1 was able to meet its own threshold range.
4.1. Summary of Results for the Selected Devices
In the laboratory testing for CO2, the results at 15–20 min and 20–30 min were the significant time stamps to discuss and compare accuracy both between two of the same branded devices, and among the tested devices. This is due to the fact that the CO2 concentration should be relatively stable at around fifteen minutes. During the testing, the twenty-minute mark was where the injection of CO2 stopped and this was therefore a significant reference time to compare accuracy, as the data between two devices should gradually return to the threshold range. In the results, it was noted that the manufacturer’s accuracy thresholds varied among the devices, for example, Airthings View Plus CO2 is +/− 50 ppm and Laser Egg + CO2 is +/− 30 ppm.
The study compared the two devices for each product and compared the seven products together (
Table 5).
The comparison methods use the specific accuracy range for each product as a baseline. Each graph indicates two curved lines for the eight times CO
2 data was collected. The reference line indicates when there was no difference between two products, and the bias line shows the difference between each device at different tested times. When compared with other products, different accuracy ranges also varied and need to be discussed further. Note that Eve Room only tested temperature, relative humidity and TVOCs, therefore discussion about CO
2 did not include Eve Room.
Table 5 presents a statistical analysis from the actual vs fitted graph and the linear graph was generated from the statistical software based on the data from the first and second devices.
p-value indicated with ***, represents the statistical significance, which means the two devices are significantly correlated. The closer the coefficient number was to 1, the better the performance of two devices was. The closer R
2 (0 <= R
2 <= 1) approaches 1, the better the performance of the devices. The lower the constant value, the lower the accuracy difference (bias) between two devices, the better the device performance. Tests at the end were also beneficial and are to be discussed in the next section as the CO
2 concentration level should be stable and the accuracy range should be relatively reduced.
The results for PM2.5 are shown in
Table 6. The table shows the findings from the lab testing of each of the devices compared to the reference research-grade meter.
The following section will explain the accuracy results for each device. As stated previously, all devices tested CO2 readings within their accuracy threshold (the data is within the range at least for one tested time). However, this section reports on which device was the most accurate for each of the four tested parameters.
AirBird: According to the manufacturer specifications [
31], the range of CO
2 measured by the device is from 0 to 5000 ppm, and the accuracy is ±30 ppm or ±3% for each reading. For temperature, it is from −20 to +60 °C, with an accuracy of ±0.5 °C. RH measured by the device is from 1 to 99% with a ±3% accuracy. When the AirBird average accuracy performance was compared to other devices, there are four times when data between the two devices aligned with the accuracy range and performance graph (
Figure 2). Accuracy ranges varied among the devices, and therefore this needed to be considered. AirBird and Laser Egg + CO
2 shared the same range, which was +/−30 ppm and their CO
2 performance was the best since it was accurate across the eight tests. Statistical analysis was also conducted (
Table 5), and the two devices from AirBird indicated a neutral correlation between each other compared with other devices. The devices had a relatively lower constant number, 21.0552 (according to the formula), compared with the device accuracy difference of 30 ppm. Therefore, results from both the accuracy performance graph (
Figure 2) and statistical analysis (
Table 5) consistently indicate that the two devices performed a better average accuracy than other devices. However, one of the two devices was not within the accuracy threshold for temperature or relative humidity.
Airthings View Plus: According to the manufacturer specifications [
37], the range of CO
2 measured by the device is from 400 to 5000 ppm, and the accuracy is ±50 ppm. For temperature, it is from 10 to +35 °C, and the accuracy is ±0.5 °C. RH measured by the device is from 0 to 80% with ±3% accuracy. Based on the device performance graph (
Figure 3), Airthings View Plus devices were accurate when testing CO
2, temperature, and relative humidity; and comparison with other devices was presented in
Table 4. Based on the statistical analysis conducted (
Table 5), the two devices from Airthings View Plus indicated a strong correlation between each other compared with other devices. The devices had the lowest constant number, 2.928 (formula), compared with the device accuracy difference of 50 ppm.
Aranet4 Home: According to the manufacturer specifications [
40], the range of CO
2 measured by the device is 0 to 9999 ppm and the accuracy is ±30 ppm, or ±3% for each reading. For temperature, it is from 0 to 50 °C, with an accuracy of ±0.3 °C of reading. RH measured by the device is from 0 to 85% with a ±3% accuracy. When the Aranet4 Home average accuracy performance was compared to other devices, there were four times when data between the two devices did not meet the accuracy range. By contrast, Airthings Views Plus, Awair Omni and Laser Egg + CO
2 met the accuracy across all eight tests. Based on the statistical analysis conducted (
Table 5), the two devices from Aranet4 Home indicated a weaker correlation between each other compared with other devices. The devices had the highest constant number, 77.01 (formula), compared with the device accuracy difference of 30 ppm. Therefore, results from both the accuracy performance graph (
Figure 4) and statistical analysis (
Table 5) consistently indicated that the two devices performed with less average accuracy than other devices. In addition, the two devices were not within the accuracy range either for temperature or relative humidity.
Awair Omni: According to the manufacturer specifications [
42], the range of CO
2 measured by the device is 400 to 5000 ppm and the accuracy is ±75 ppm or ±10% for each reading. For temperature, it is from −40 to 125 C, with an accuracy of ±0.2 C. RH measured by the device is from 0 to 100% with a ±2% accuracy. Based on the device performance graph (
Figure 5), Awair Omni performed accurately when detecting CO
2 every time. Based on the statistical analysis conducted (
Table 5), the two devices from Awair Omni indicated a strong correlation between each other compared with other devices. The devices had the second lowest constant number, 8.48 (formula), compared with the device accuracy difference of 75 ppm. Therefore, results from both the accuracy performance graph (
Figure 5) and statistical analysis (
Table 5) consistently indicate that the two devices performed a good accuracy compared to other devices. However, one of the two tested devices was not within the accuracy range for temperature and for PM2.5. In addition, both devices did not meet the accuracy range for relative humidity.
Laser Egg + CO2: According to the manufacturer specifications [
51], the range of CO
2 measured by this device is from 400 to 5000 ppm (±30 ppm or 3%). For temperature, it is from −20 to 100 °C, with an accuracy of ±1 °C. RH measured by the device is from 0 to 99% with a +/−1% accuracy. PM2.5 range is from 0.3 to 10 µg/m
3. Based on the device performance graph (
Figure 6), Laser Egg + CO
2 devices performed an accurate measurement of CO
2 comparison between two devices. One of the two devices was not accurate when detecting temperature. Based on the statistical analysis conducted (
Table 5), the two devices from Laser Egg + CO
2 indicated a strong correlation between each other compared with other devices. The devices had a low constant number, −24.579 (formula), compared with the device accuracy difference of 30 ppm. Therefore, results from both the accuracy performance graph (
Figure 6) and statistical analysis (
Table 5) consistently indicate that the two devices perform with high accuracy compared to other devices. However, the two devices did not give results within the accurate range for relative humidity. In addition, the two devices were not within the accuracy range to detect PM2.5.
PurpleAir PA-I: According to the manufacturer specifications [
56], the accuracy threshold range of temperature is from −40 to 85 C. RH measured by the device is from 0 to 99% with a +/−3% accuracy. PM2.5 range is from 0.3 to 500 µg/m
3. This device does not test for CO
2. Based on the test conducted in the lab, PurpleAir PA-I performed relatively accurately for PM2.5 both in terms of the devices’ accuracy range, and also in terms of best correlation with the reference meter. The PM2.5 comparison data was extracted from the four-test average data in the lab environment. In addition, the two devices also tested accurately for temperature and relative humidity.
4.2. Field Studies
The study included field testing where three scenarios were used to show the devices’ responsiveness, visualization, and ease of use in residential spaces. These studies were not completed in a laboratory, and instead were carried out in a house as indicated in the methods section. The location of the devices inside their rooms was important because the devices must be placed near to where people would be breathing: thus, in the living room, they were placed on a coffee table; in the bedroom, they were on the floor near a yoga mat while a person exercised; and in the kitchen, they were placed on a table.
Scenario 1—Living Room: Snuffing out a candle
In scenario 1, there was one small window opened in the living space for outdoor ventilation without A/C. Two candles were placed in the center of the coffee table among the fourteen devices (see
Figure 7).
The findings relating to CO
2 showed that the levels did not change significantly after 30 min. The reason might be that the testing time was too short or that the size of the candles was relatively too small to provide enough CO
2 in the environment. Results of Scenario 1 are presented in
Table 7 and
Table 8.
Snuffing out the candles triggered a high concentration of PM2.5. The seven devices provided responses differently, and only Airthings View Plus pushed the notifications to the phone and indicated red warnings on the display. The light was indicated as orange/red on the Awair device, and its App indicated poor IAQ. In addition, Purple Air indicated red warnings on the device (
Figure 8), but there was no direct data access available via phone, and data can only be viewed on the airmap.
In the case of Laser Egg + CO2, it did not provide an accurate response as its display continued to indicate good air quality and no change occurred to reflect the higher PM2.5 concentration level.
Scenario 2—Bedroom: Exercising indoors for 30 min
Scenario 2 was located in the bedroom on the upper floor with one window open, and the door closed. The devices were placed beside the yoga mat and the initial air quality was measured as good at 440–460 ppm of CO
2 (See
Figure 9).
One person exercised for thirty minutes in the room. It was observed that at around 20 min, the CO
2 concentration in the room started to exceed 1000 ppm. Although the window was open to allow air circulation into the space, the release of CO
2 from a person breathing and moving around was relatively high showing that the ventilation system needed improvement. The results of Scenario 2 are presented in
Table 9 and
Table 10.
In Scenario 2, the Airthings View Plus had the most responsive phone notifications when the temperature and CO2 levels were elevated (
Figure 10). The related App also responded with the potential health problems associated with the poor indoor CO
2, temperature, relative humidity, and PM2.5. The Awair Omni also had orange/red notifications on the device at around 20 min (
Figure 11), and the changing CO
2 levels and temperature were shown on opening of the Smartphone Application. Each device communicated IAQ data, for example the Awair Omni App displayed colours to show IAQ data, from green to orange then dark red (1 to 2000 ppm) for CO
2 and to illustrate changing temperatures, 18–25 °C was indicated in green with good healthy IAQ and temperatures the ranges 9–18 °C and 25–34 C were indicated as poor IAQ. At the 20 min mark, Aranet4 Home indicated green (CO
2 below 1000 ppm) (
Figure 12) thus no notification was shown on either device or phones, and no health threshold was indicated in the App. Only one of the tested devices provided both IAQ data and information about the impacts. Awair Omni provided both data ranges and related potential health problems listed in the App to help users understand the impact of poor IAQ. This was a unique function that other devices did not provide.
Scenario 3—Kitchen: Cooking without using an exhaust fan
Based on a published study by Wang et al., this scenario tested three activities that take place in residential environments: frying garlic, steak and also using an air fryer to cook bacon [
35]. In scenario 3, a small window above the sink in the kitchen was open halfway to allow outside ventilation, but no exhaust fan or air conditioning was used. IAQ devices were placed on the counter island (at table height) that was across from the stove because there was not enough space to place fourteen devices on the counter beside the stove (
Figure 13).
The Scenario 3 observation occurred and data was recorded for (1) the initial setting, (2) during cooking (at 15 min) and (3) after cooking (at 30 min). The initial hypothesis was that there would be a substantial increase in CO
2 and PM2.5 levels during cooking. However, the CO
2 level did not change very much during the 30 min test.
Table 11 and
Table 12 summarize the results of Scenario 3.
The use of the air fryer had the highest impact on the PM2.5 levels. Airthings View Plus, Awair, Purple Air all responded quickly to the increasing PM2.5 concentration by changing the color code on the devices. In contrast, Laser Egg + CO2 did not respond, and when the other devices responded to the changing CO2 and PM2.5 levels, it still indicated good air quality inside the space.