1. Introduction
Cytomegalovirus (CMV), a common human pathogen, can affect different organs, especially in immunocompromised patients [
1,
2]. The quantitation of CMV particles in the blood for pre-emptive and effective therapy is the treatment guideline to reduce morbidity and mortality [
3,
4,
5,
6]. However, CMV reactivation patients with non-DNAemia stage will delay the timing of medical intervention, especially in organ-related diseases such as pneumonia [
7], retinitis [
8], colitis [
9] and inflammatory mammary gland [
10]. Therefore, the accurate detection of CMV particles in various specimens is critical for clinical decision-making and prescribing appropriate treatment regimens.
The polymerase chain reaction (PCR) approach is widely used for rapid diagnosis of CMV infection [
11,
12,
13,
14]. However, commercial automated extraction combined with real-time PCR platforms (In Vitro Diagnostics, IVD), such as Roche Cobas AmpliPrep/TaqMan 48, Abbott m2000SP/2000RT and QIAGEN QIAsymphony/Rotor Gene-Q, have only been validated for the detection of CMV DNA in blood samples [
15,
16,
17]. Few assays for non-blood specimens have been validated and reported [
18,
19]. Considering the matrix complexity of non-blood specimens, the performance of CMV PCR testing of such specimens should be validated to ensure the quality of assays.
Although a manual PCR approach for the detection of viral nucleic acids has gradually been replaced with highly sensitive and straightforward automated platforms, most of these platforms require specific infrastructure, equipment, and trained staff. A fully integrated, automated platform for nucleic acid extraction and real-time PCR using the Becton Dickinson (BD) MAX system enables many commercial IVD assays for infectious agents [
20,
21,
22,
23,
24,
25,
26], with the possibility of creating user-defined protocols using open-system reagents to meet emerging diagnostic demands and address regional healthcare needs. The evaluation of laboratory developed assays using the BD MAX system for the detection of numerous microbes has recently been published, however, none of them pertain to CMV [
27,
28,
29,
30,
31,
32]. Herein, we designed a new QMT assay applied on the BD MAX system for the detection of CMV DNA in various clinical specimens, especially different respiratory tract specimens. In this study, we validated the performance of the QMT assay. In addition, the diagnostic performance and workflow of QMT assay, commercially available CMV assays and an in-house CMV PCR assay were compared.
2. Materials and Methods
2.1. Clinical Samples and Ethics Statement
A total of 1067 various clinical samples including 426 plasma, 293 respiratory tract specimens (RTS), 127 stool, 101 cerebral spinal fluid, 90 vitreous humour, 18 urine, 5 breast milk, 5 bone marrow and 2 lung tissue, suspected of being infected with CMV and collected at the Clinical Virology Laboratory of National Cheng Kung University Hospital (NCKUH), were analyzed retrospectively. Sample aliquots (stored at −70 °C) were simultaneously tested by the specified assays. The Institutional Review Board of NCKUH with a number of B-ER-106-316 approved the study. The demographic and clinical information for the patients was unlinked prior to analysis, and hence the patients’ informed consent was waived.
2.2. QMT Assay: QIAGEN Artus CMV RG PCR (Q-CMV PCR) Reagents Applied on BD MAX
The BD MAX system allows nucleic acid extraction and real-time PCR using a single instrument. We designed the following protocols for the new QMT assay. Sample preparation for BD MAX: first, 15 µL of internal positive control (IPC) from the artus CMV RG PCR (Q-CMV PCR) kit (QIAGEN GmbH, GERMANY) was added to each sample-processing tube of the ExK DNA-1 extraction kit (DNA-1 kit, ref: 442818; Becton Dickinson, Breda, the Netherlands). Next, 200 µL of liquid specimens was added to SPT, while 2 g of solid stool or 10 µL of liquid stool were used. Liquefy the viscous sputum or BAL: an equal volume of viral transport medium consisting of Eagle’s Minimum Essential Medium (EMEM), penicillin/streptomycin, and 0.5% gelatine was added to viscous respiratory specimens, which were then vortex-mixed for over 3 min. Five new designed different BD MAX extraction programmes were used for the different specimens (
Table 1).
The new reaction conditions of the Q-CMV PCR kit protocol on BD MAX realtime PCR instrument were modified as follows: 12.5 µL of DNA with 12.5 µL PCR master mix in each PCR reaction. Fluorescent signal gain for Cycling Green and Cycling Yellow was used for detecting CMV and IPC, respectively.
2.3. Validation of QMT Assay
Assay accuracy was determined using external quality control (College of American Pathologists Proficiency Testing (CAP-PT)) specimens. Precision was evaluated by testing the aliquots from pooled positive plasma samples and comparing the results from intra-runs and inter-runs. Clinical specimens containing herpes simplex virus, varicella-zoster virus, Epstein–Barr virus, hepatitis B virus, hepatitis C virus, influenza A and B virus, BK virus and enterovirus, were tested for analytical specificity.
2.4. RGQ Assay and in-House PCR Assay
The automated LabTurbo 48 Compact extraction system (Taigen Bioscience Corp., Taipei, Taiwan) with LabTurbo virus mini kit (cat. no. LVN480-300) were used for DNA extraction for RGQ and in-house PCR assays. RGQ assay using the QAIGEN artus CMV RG PCR (Q-CMV PCR) kit constituting a ready-to-use reagent for the detection of CMV DNA was validated for the Rotor-Gene-Q real-time PCR instrument, as suggested by the manufacturer. The in-house PCR assay was performed as previously described [
33].
2.5. Roche Cobas CAP/CTM CMV Real-Time PCR Assay (Roche Assay)
The Roche CMV assay relies on the Cobas AmpliPrep/Cobas® TaqMan system, which consists of a Cobas AmpliPrep system for automatic nucleic acid isolation with magnetic beads and Cobas® TaqMan system for real-time PCR (Roche Molecular Diagnostics, Pleasanton, CA, USA). In the current study, the Roche assay was only used for the analysis of plasma specimens, as per manufacturer’s recommendations.
2.6. Comparison of QMT Assay to the Others Assays
Diagnostic sensitivity and specificity were evaluated by comparing the performance of different CMV PCR assays with clinical specimens. A total of 205 various clinical specimens were simultaneously used for the evaluation of diagnostic performance of the in-house, RGQ, and QMT assays (as specified below). The workflow is shown in
Figure 1.
2.7. Statistical Analysis
Post hoc statistical analysis was used for the performance of QMT assay with various clinical specimens. The statistical analysis of CMV DNA detection rate in respiratory tract specimens and plasma of QMT assay was performed using Fisher’s exact test. The limit of detection (LOD) for the MAX assay was determined by analyzing dilutions of pooled CMV-positive plasma samples (quantified using the Roche CMV assay) in replicates of eight on three separate days. Assay LOD was determined by using Probit analysis (SPSS Statistics for Windows, v. 17.0., SPSS Inc., Chicago, IL, USA) with 95% probability within a 95% confidence interval (95% CI). Regression analysis correlating the different assays and agreement between viral loads was performed using the Bland–Altman method with MEDCALC
® statistical software (
https://www.medcalc.org/, accessed on 15 July 2020). Data were log
10-transformed prior to analysis.
4. Discussion
The accurate and rapid detection of CMV in various specimens is essential for the timely treatment of CMV-related disease. Here, we present a user-defined protocol using the fully automated BD MAX system for CMV detection. In the current study, we have successfully devised CMV nucleic acid extraction protocols for the BD MAX instrument for multiple non-blood samples (
Table 1). Although EDTA plasma was the most suitable sample for the artus CMV RG PCR kit as recommended by the manufacturer, we used this kit on BD MAX real-time PCR system with excellent test accuracy, precision, sensitivity, and specificity.
CMV is often found in various clinical specimens; however, blood is the only specimen that had been validated for use in these tests to date, save for the fully automated ELITe platform (
https://www.elitechgroup.com/product/cmv-elite-mgb-kit-ingenius, accessed on 15 July 2020), which has also been validated for non-blood specimens such as amniotic fluid, urine, CSF, and saliva. In our developed protocols, we had validated additional clinical specimen types (
Table 2) including respiratory tract specimens, stool, cerebral spinal fluid, vitreous humour, urine, breast milk, bone marrow and lung tissues using QMT assay.
According to a recent study, LOD for the commercial CMV DNA standard, detected using Roche CAP/CTM CMV platform in BAL, is lower than that in CSF and urine [
18]. Costa et al. (28th European Congress of Clinical Microbiology and Infectious Disease in 2018, abstract P0496) reported an LOD for CMV ELITe platform of 57 IU/mL for amniotic fluid, 151 IU/mL for urine, and 44 IU/mL for saliva. Consequently, LOD values for CMV detection in various clinical specimens by using different methods vary greatly. The LOD of QMT assay developed in this study (82 IU/mL) was lower than that of the Roche assay (<137 IU/mL) in plasma specimens. Hence, the LOD for QMT assay in other types of clinical specimens should be investigated further.
Precise detection of CMV in specimens from immunocompromised patients with clinical manifestations of disease is crucial for therapeutic regimes. However, the ability of various commercial kits to manually extract CMV DNA from spiked human specimens varies [
34]. Further, compared to other clinical specimens, respiratory tract specimens have been reported to account for the highest CMV PCR detection rate [
18,
19,
35], for which similar results were found in this study (
Table 2 and
Table 3). Comparison of the sensitivities of different extraction protocols in various clinical specimens for CMV PCR requires further investigation.
Further, we observed an abnormally high IPC invalid rate for MAX assay with bone marrow 80% (4/5) and stool specimens 25.2% (32/127) (
Table 2). Bone marrow, with high concentrated haemoglobin, yields high invalid PCR results and a high IC invalid rate in stools may be due to the presence of PCR inhibitors, including complex polysaccharides, and lipids [
36,
37]. The manufacturer has validated the BD MAX DNA-1 kit, used in this study, for the extraction of DNA from plasma, serum, or neat urine specimens, however, not from stool and bone marrow. Therefore, the BD MAX DNA-4 kit (validated for stool by the manufacturer) was tested to resolve the issue of IPC invalidity of the DNA-1 kit. Indeed, the invalid rate of IPC DNA in this study decreased to 9.5% (2/21) by the DNA-4 kit. Using an appropriate DNA extraction kit may serve to reduce the invalid rate in various clinical specimens.
Previous studies [
16,
38] reported a difference within 0.1 log
10 copies/mL of the mean CMV viral load for plasma samples analysed using two different automated or semi-automated extraction instruments in conjunction with the same PCR reagents. However, the differences between two automated systems exceeded 0.5 log
10 copies/mL when whole-blood or plasma specimens were tested [
15,
17,
39]. Of note, the current study is the first to compare the quantitation of CMV DNA in various clinical specimens by two automated extraction and detection systems.
Finally, the fully automated BD MAX assay entails less manual intervention than the semi-automated RGQ assay, especially during PCR reaction preparation. Hence, the QMT assay, with a short operating time, is more convenient than RGQ assay for the analysis of small batches of specimens (six or less) in the laboratory workflow analysis. Although a limited number of samples can be extracted in one batch in these instruments, the turnaround time of the QMT assay (2.5 h) is significantly shorter than that of the RGQ assay (6 h) (
Figure 1). Overall, these two assays provide reliable and comparable results.