1. Introduction
The implantation of the optimal stem size is important for the implantation of a total hip arthroplasty (THA) in order to reduce complications. Digital templating of a THA is a helpful tool for preoperative planning of surgery regarding implant type, implant size, reconstruction of the center of rotation, offset, and limb lengths [
1]. This tool assists in order to receive a good outcome after a THA and to avoid complications due to anatomic variations. It helps to reduce the incidence of undersizing the femoral implant with the associated risk of increased micromotions, subsidence, and early aseptic loosening [
2,
3,
4] and the incidence of oversizing associated with the risk of femoral fractures. Furthermore, it is helpful to avoid limb length discrepancies, instability, or excessive tension of the periarticular soft tissues by inaccurate reconstruction of the center of rotation of the hip [
3,
5,
6,
7,
8,
9,
10].
Meanwhile, preoperative planning with plastic templates on radiographs or digital templating as a valid method to determine the correct size of the femoral stem and acetabular component belongs to the standard preoperative procedures prior to a THA [
5,
11,
12]. Digital templating of a THA is usually performed on the anteroposterior (a.p.) pelvic overview X-ray [
13]. Some software programs allow for digital templating also use the lateral view of the proximal femur for templating the stem. Although efforts have been made to further improve the accuracy of preoperative templating by using CT-based three-dimensional computerized models [
14], the simple method using only the pelvic overview X-ray is still the standard procedure in clinical routine due to its high practicability on the one hand and the high radiation exposure when using CT scans on the other. The fixation of short stem THA (STHA) with predominantly metaphyseal anchorage is different from conventional THA (CTHA). While the conventional stem has a metaphyseal-diaphyseal anchorage with an intramedullary fixation within the proximal third of the femur [
11], STHA with predominantly metaphyseal anchorage adapts its orientation to the femoral neck and thus to its ante torsion (
Figure 1).
This study aimed to determine whether there is a difference in accuracy when templating THA stems using not only the a.p. view of the pelvis but also the lateral view of the proximal femur. Furthermore, we analyzed whether there is a difference in the rate of correctly planned sizes when templating CTHA stems in comparison to STHA implants with a predominantly metaphyseal anchorage.
2. Materials and Methods
One hundred fifty patients with a good radiological outcome after implantation of a THA were selected: 50 after implantation of a THA using a conventional femoral stem (CTHA group) and 100 after implantation of a THA using a short stem with primary metaphyseal fixation (STHA group). A good radiological outcome was judged on the postoperative X-rays by three senior surgeons according to the following criteria: under- or oversizing, early subsidence, and intra- or postoperative fractures.
Undersizing was defined as either a too low fit of the implant that could have been avoided by choosing a bigger size or by the presence of a visible distance between the cortical bone and the implant in a region of bone that is crucial for the anchorage of the implant (calcar region and lateral cortex for CTHA and cortical ring of the resected femoral neck basis as well as lateral cortex for STHA). Oversizing was defined as a too high fit of the implant that could have been avoided by choosing a smaller size, indicated by the shoulder of the implant exceeding the highest point of the greater trochanter for CTHA or the shoulder of the implant exceeding the cortical ring of the femoral neck basis by more than 10 mm for STHA. Early subsidence after full load-bearing, which was allowed for all patients, was defined as an unambiguous (>2 mm) migration of the implant within the surrounding bone in a follow-up X-ray examination 3 months after surgery. The occurrence of intra- or postoperative fractures were retrieved from the surgery reports and patients’ charts.
For these patients, who were included consecutively and underwent THA between 2010 and 2011, preoperative and postoperative digital X-rays in pairs were selected from the database of the department of orthopedic surgery of a university hospital. All these patients underwent standardized X-ray examinations, including an a.p. pelvic overview and an lateral view (Lauenstein) of the proximal femur. These X-rays existed for the preoperative and postoperative state and with a follow-up of at least 3 months. All X-rays were taken with a standardized film-focus-distance of 100 cm. An external calibration marker of 25 mm diameter was positioned between the legs of each patient for the preoperative pelvic overview.
Digital templating was performed by three senior surgeons experienced in THA. When conducting digital templating, the examiners were only able to see the preoperative, but not the postoperative X-rays.
The software used for digital templating was OrthoView (OrthoView Orthopaedic Digital Planning Version 6.3, Meridian Technique Limited, Southampton, UK). For this software tool, templates for the conventional BICONTACT stem (CTHA) and the METHA short stem (STHA; both Aesculap B. Braun, Tuttlingen, Germany) were available (
Figure 2 and
Figure 3).
One hundred patients received a cementless METHA prosthesis with a caput-collum-diaphysis (CCD) angle of 130°, 135°, or 140°. All these stems are available in a size range of 0 to 7. The median stem size of the implanted METHA short stems was 3.
Fifty patients received a cementless conventional BICONTACT stem. These stems are monobloc-types with different offset variations: S for standard, H for High offset, and SD for dysplastic femurs. In our study, we templated and implanted only S and H types. BICONTACT stems are available in stem sizes 10–21. From the 50 truly implanted hip implants, the median stem size was 14.
Statistics
First, the results of all three raters were described with the cumulative rate of consistency between the templated and the implanted stem size. The difference was described as size derivation from the actual stem size. In addition, the planned stem sizes were presented in scatterplots. The intra-class correlation (ICC) 3,1 for consistency and 3,k for conformity for all raters was measured and quantified using Fleiss Kappa and Kendall’s coefficient of concordance. For the correlation between the templated stem size and the actual implanted stem size, a Spearman’s rank correlation coefficient was determined. Analysis was conducted in R (Version 4.0.3) [
15], and figures were produced using the package ggplot2 [
16].
4. Discussion
Total hip arthroplasty using short stem implants has grown up to a popular technique beside the successful conventional procedure. Short stem devices, such as the METHA stem, show promising short-term results [
18,
19]. Prior to implantation of the stem, digital preoperative templating is a common and routinely used tool that has not yet been determined with regards to its benefit and exactitude [
11,
13]. So far, nothing is known about the benefit in terms of the rate of correctly planned sizes when using additional X-rays in a second radiological plane.
This study confirms on the one hand that there is no significant difference between templating STHA in comparison to CTHA, as former studies have shown [
11,
13,
20,
21]. On the other hand, the current study indicates that the exactitude can possibly be increased by using a second radiological plane for templating. Although this effect did not reach statistical significance for neither the CTHA nor the STHA group, the tendency of the effect could be observed especially for the short stem implant.
Matching exactly the same stem size as predicted by preoperative templating occurred in 44.4% ± 6.8% with CTHA when using one plane radiographs. Other studies on conventional implants have shown an exact rate of correctly planned sizes between 31% and 58% of cases [
6,
11,
13,
22,
23]. By tolerating a variation of +/− one size between templating and implantation, the data of this study showed a predictability of 88.4% ± 6.0%, again in good conformity to other studies that have reported a predictability of 71.0%–94.0% under these circumstances.
By the use of a second radiological plane, the lateral view, according to Lauenstein, in addition to the a.p. pelvic overview, a comparable rate of exact matching between templated and implanted size of 40.0% ± 20.8% was determined. With an allowed measuring tolerance of +/− one stem size, a slightly increased accordance of 89.4% ± 9.8% was reached, although the difference was not significant. A variation of more than +/−2 stem sizes between templating and implanted size occurred in 0.7% of the cases when using two radiological planes compared to 6.2% when using only one plane (p = 0.18). This reduced range of dispersion indicates a higher accuracy when using two planes, although the difference was not significant.
Results analyzing the rate of correctly planned size of preoperative templating of short stem implants are less documented in the literature. There are only two studies dealing with this question: on the one hand, for the METHA short stem by Schmidutz et al. [
11] and on the other hand, for the MAYO short stem (Zimmer, Warsaw, IN, USA) by Wedemeyer et al. [
24]. Both of these short stem designs aim at a metaphyseal anchorage. Wedemeyer et al. reported an exact match without size tolerance in 38% of the cases. Schmidutz et al. found for the METHA short stem an exactitude of 49% of preoperative templating. In our study, we found similar results with up to 36.2% ± 16.9%. The MAYO device showed predictability of 95%, Schmidutz et al. reported 89% for the METHA stem, and our study reports a comparable matching rate of 86.3% ± 9.5% when tolerating a difference of one size.
In analogy to the CTHA group, the second plane (lateral view) led to a smaller range of dispersion, matching the exact size in 50% ± 27.5%, and in 89.9% ± 12% with a tolerance of +/−1 stem size.
When comparing the STHA and CTHA groups, adding the second radiological plane increased the probability of matching the exact size in the STHA group (50.0% vs. 36.2%), but not in the CTHA group (40.0% vs. 44.4%). When tolerating a difference of +/− one stem size, templating with two planes revealed comparable results between the groups that were not statistically significant (89.9% for STHA vs. 89.4% for CTHA). Although none of these results reached statistical significance, these data and the graphical illustration (
Figure 4) imply that using two planes for preoperative templating is more likely to have a beneficial effect with regard to a short stem than to a conventional stem.
In general, preoperative templating is important with regard to the risk of an intraoperative fracture of the femur that is a common complication in a non-cemented THA [
25]. On the other hand, the undersizing of the implant may lead to early implant migration [
26] that may, in turn, impair osseointegration and lead to early aseptic loosening [
27].
Less experienced surgeons especially benefit from a profound preparation by means of preoperative templating that helps them to improve their orientation and to feel safer during bone preparation. The information derived from this study that the planned size tends to be too big for small implant sizes and too small for big implant sizes might be especially helpful for them.
In this study, all templating was performed by digital techniques, but there are alternative analog techniques using plastic templates. Higher exactitude for digital templating was described in studies of The et al. and Widdon et al. [
12,
28]. On the other hand, there are also studies that come to the conclusion that analog templating might have a higher rate of correctly planned size [
23,
29]. All studies showed that templating is a beneficial tool to avoid intraoperative complications.
In particular, templating in two radiological planes may also be very helpful with regard to the preoperative detection of possible contraindications for a certain implant, e.g., a retro torsion or an excessive ante torsion (
Figure 5) of the femoral neck when planning the implantation of a short stem device with primary metaphyseal anchorage along the femoral neck.
This study has some limitations. Although the external calibration marker had been positioned in a standardized way for all patients, an imprecise positioning may influence the correct scaling of the X-rays and the results of THA templating [
30]. A high body mass index (BMI) that was not recorded in this study may complicate the exact positioning of the calibration marker. However, the correlation between BMI and magnification has been shown to be weak so that a major bias can be ruled out [
31].
Second, due to the retrospective design, the study was not able to analyze whether templating with two radiological planes would have led to a measurable effect on complication rates within this particular study population.
Furthermore, the study had an asymmetrical distribution between the CTHA and STHA groups (50 vs. 100 cases). This rather unconventional study design was chosen due to the fact that the rate of correctly planned size of digital templating of conventional stems has already been well analyzed by other studies, whereas there are very little data about short stems that were put into the focus of this study.